use of the labnet micropipettor
DESCRIPTION
Use of the Labnet Micropipettor. Purpose of the Pipettor. Biotechnologists use a pipettor to transfer precise amounts of liquids from one container to another. The measurements are in microliters 1 ml=1000 µl. Use of the Labnet Micropipettor. Care of the Pipettor. - PowerPoint PPT PresentationTRANSCRIPT
Use of the Labnet Micropipettor
Purpose of the Pipettor
• Biotechnologists use a pipettor to transfer precise amounts of liquids from one container to another.
• The measurements are in microliters 1 ml=1000 µl.
Use of the Labnet Micropipettor
Care of the Pipettor
• Misuse of the pipettor can result in incorrect measurements affecting your investigation.
• Do not drop the pipettor.• Do not adjust the volume
to less than 2 µl.
Use of the Labnet Micropipettor
Parts of the Pipettor
• 1) yellow pushbutton• 2) calibration knob• 3) digital volume display• 4) barrel• 5) disposable pipette tip• 6) tip ejector
3
5
6
1
4
6
1
62
3
4
5
Use of the Labnet Micropipettor
Volume AdjustmentStep 1
• In the display window are three vertical numbers.
• The top number is tens µl (10.0).
• The middle number is ones µl (02.0).
• The bottom number tenths µl (00.9).
Use of the Labnet Micropipettor
Volume AdjustmentStep 2
• Set the pipettor to the desired volume by turning the calibration knob until the volume display reads the desired numbers.
• 7.5 µl would read from top to bottom in the display window: 0 - tens 7 - ones .5 - tenths
Use of the Labnet Micropipettor
Transferring LiquidsStep 1
• Adjust the pipettor to the desired volume.
• Hold the pipettor so the display is facing the user and the thumb can easily depress the yellow pushbutton.
Use of the Labnet Micropipettor
Transferring LiquidsStep 2
• Open the box of yellow pipette tips supplied by the Office of Biotechnology.
• Place the pipette cone (end) into one of the tips and press down firmly.
• NEVER USE THE PIPETTOR WITHOUT A TIP!
Use of the Labnet Micropipettor
Transferring LiquidsStep 3
• Lift the pipettor and attached tip away from the tip box.
• Depress the pushbutton to the first stop before placing the pipette tip in the liquid.
• Place the tip in the liquid.
Use of the Labnet Micropipettor
Transferring LiquidsStep 4
• Slowly release the pushbutton and draw the liquid into the tip.
• Remove the tip from the liquid and visually check for liquid in the tip.
Use of the Labnet Micropipettor
Transferring LiquidsStep 5
• Place the tip into the container which is to receive the liquid.
• Slowly depress the pushbutton as far as it can go.
• With the pushbutton fully depressed, withdraw the tip.
Use of the Labnet Micropipettor
Transferring LiquidsStep 6
• Press down on the tip ejector button to eject the tip.
• Dispose of the tip in an appropriate container.