RADMON meeting, 22 July 2004 Use of TLD at CERN 1

Use of TLD at CERN

Marco Silari, Jacques Wolf (SC/RP)

• Working principles• Evaluating the exposure• TLD response to gammas and neutrons• The TLD service of the RP group• Experimental procedure

Page 2: Use of TLD at CERN

RADMON meeting, 22 July 2004 Use of TLD at CERN 2

Exposure to radiation

• Thermoluminescent dosimeters are inorganic crystals exhibiting high concentration of trapping centers in the gap between valence and conduction bands

• Electrons are elevated from the valence to the conduction band by the incident radiation and are then captured at one of the trapping centers

• Holes can also be trapped in an analogous process• The TLD material functions as an integrating

detector in which the number of trapped electrons and holes is a measure of the number of electron-hole pairs formed by the radiation exposure

Page 3: Use of TLD at CERN

RADMON meeting, 22 July 2004 Use of TLD at CERN 3

Evaluating exposure• After the exposure, the trapped

carriers can be measured by heating the TLD

• At a temperature that is determined by the energy level of the trap, the trapped electrons can pick up enough thermal energy so that they are re-excited back to the conduction band

• The liberated electrons then migrate to near a trapped hole, where they can recombine with the emission of a photon

• Alternatively, if the holes are released at a lower temperature, they may migrate to a trapped electron and their recombination also results in a radiated photon

Page 4: Use of TLD at CERN

RADMON meeting, 22 July 2004 Use of TLD at CERN 4

Evaluating exposure• The emitted photons are in the visible region of the e-m

spectrum and their total number is an indication of the original number of electron-hole pairs created by the radiation

• TLD readers derive a signal by heating the dosimeter and reading the light output via a photomultiplier. The light yield is recorded as a function of sample temperature in a glow curve

Page 5: Use of TLD at CERN

RADMON meeting, 22 July 2004 Use of TLD at CERN 5

TLD response vs dose

• The response curve of a TLD material as a function of absorbed dose shows three regions• a region in which the response is linear • a region of supralinearity, in which the

sensibility increases (but the reliability decreases)

• a saturation region (defines maximum detectable dose = 80% of saturation dose)

Page 6: Use of TLD at CERN

RADMON meeting, 22 July 2004 Use of TLD at CERN 6

Lithium Fluoride

• Almost negligible fading at room temperature• Low atomic number, which does not differ

much from that of air or tissue energy deposited in LiF is correlated to gamma dose

• Constant response over a wide range of photon energies

• Because natural lithium contains 7.4% Li-6, LiF dosimeters are sensitive to slow neutrons via (n,α) reaction

• The response to neutrons can be enhanced by using lithium enriched in Li-6, or suppressed by using lithium consisting entirely of Li-7

Page 7: Use of TLD at CERN

RADMON meeting, 22 July 2004 Use of TLD at CERN 7

Energy response to photons

Page 8: Use of TLD at CERN

RADMON meeting, 22 July 2004 Use of TLD at CERN 8

Energy response to neutrons

Li-6: σth = 945 barn, Li-7: σth = 0.033 barn

Page 9: Use of TLD at CERN

RADMON meeting, 22 July 2004 Use of TLD at CERN 9

The TLD service of the RP group

At CERN we use Harshaw TLD 600 (95.6% Li-6) and TLD 700 (0.01% Li-6)

Size of the dosimeter: 3.1 mm x 3.1 mm x 0.9 mm (23 mg)

Dose range: 10 µGy – 100 Gy (in practice: 10 µGy – a few mGy)

Page 10: Use of TLD at CERN

RADMON meeting, 22 July 2004 Use of TLD at CERN 10

The TLD service of the RP group

One automatic reader, ALNOR (10 µGy – 2 Gy)

One semi-automatic reader, HARSHAW, for high doses

Cross-calibration of TLDs with the environmental monitoring stations

Page 11: Use of TLD at CERN

RADMON meeting, 22 July 2004 Use of TLD at CERN 11

The TLD service of the RP group

Number of measurements in 2003:

211 gamma measurement locations (1688 TLD readings)

442 neutron measurement locations (10608 TLD readings)

Page 12: Use of TLD at CERN

RADMON meeting, 22 July 2004 Use of TLD at CERN 12

Experimental procedure (1)

• 2 readings with ALNOR by heating the TLDs with a jet of N gas at 340 degrees

• thermal treatment of 16 hours at 80 degrees to reduce fading

• installation, exposure to radiation, recuperation

• thermal treatment of one hour at 80 degrees to eliminate instable peaks at low temperature and short-lived

• first reading with ALNOR at 340 degrees• second reading with ALNOR at 340 degrees for

background subtraction

Page 13: Use of TLD at CERN

RADMON meeting, 22 July 2004 Use of TLD at CERN 13

Experimental procedure (2)

• thermal treatment of 16 hours at 80 degrees

• irradiation with Cs-137 source in the calibration lab, 1 mSv

• thermal treatment of 1 hour at 80 degrees

• reading with ALNOR at 340 degrees for background subtraction

To correct the individual gamma response of each TLD, the following procedure is applied:

Page 14: Use of TLD at CERN

RADMON meeting, 22 July 2004 Use of TLD at CERN 14

Conclusions

• How many measurement positions are needed by the LHC experiments? If the number is significant (i.e., above a few percent of our present workload), the required manpower and resources have to be addressed

• To be used for high-level dosimetry, some investigation on their response to the various types of radiation is required

• For neutron monitoring in the environment, the TLDs are used within a polyethylene moderator, whilst in the LHC experiments they should be used bare

• The RP group still has to take a decision on whether or not to maintain the TLD service or to outsource it