user manual - transcat · us, canada, latin america ametek tci at 1-800-527-9999 europe, africa,...

80
User Manual Advanced Signal Calibrator JOFRA ASC321

Upload: others

Post on 11-Aug-2020

2 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: User Manual - Transcat · US, Canada, Latin America AMETEK TCI at 1-800-527-9999 Europe, Africa, Middle East AMETEK Denmark A/S at + 45 4816 8000 Asia AMETEK Singapore Pte. Ltd. at

User Manual

Advanced Signal Calibrator

JOFRA ASC321

Page 2: User Manual - Transcat · US, Canada, Latin America AMETEK TCI at 1-800-527-9999 Europe, Africa, Middle East AMETEK Denmark A/S at + 45 4816 8000 Asia AMETEK Singapore Pte. Ltd. at

1. Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11.1 Contacting Ametek . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1

1.2 Standard Equipment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1

1.3 Safety information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2

2. Calibrator Interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 52.1 Main Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7

2.2 Menu Bar . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8

3. Using Measure Modes (Lower Display) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 133.1 Measuring volts and frequency . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13

3.2 Measuring mA. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13

3.3 Measuring Temperature . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14

3.4 Measuring Pressure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15

4. Using Source Modes (Lower Display) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 174.1 Setting 0% and 100% Output Parameters . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17

4.2 Using Auto Output Functions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18

4.3 Sourcing mA . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18

4.4 Simulating a Transmitter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19

4.5 Sourcing volts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19

4.6 Sourcing frequency . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19

4.7 Sourcing a pulse train . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20

4.8 Sourcing Thermocouples . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21

4.9 Sourcing Ohms/RTDs. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21

5. Using Isolated Measure Modes (Upper Display) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 235.1 Measuring volts and mA. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23

5.2 Measuring current with loop power . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23

5.3 Measuring Pressure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24

6. Using the Upper and the Lower Display for Calibration and Testing . . . . . . . . . . . . . . . 256.1 Performing a Switch Test. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25

6.2 Testing an Input or Indicating Device. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29

6.3 Calibrating an I/P Device . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30

6.4 Calibrating a Transmitter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31

6.5 Calibrating a Pressure Transmitter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31

6.6 Using Scaled Current or Voltage when testing or calibrating a Transmitter. . . . . . . . . . . . . . . . . . . . . . . . . 32

6.7 Using Percent Error when testing or calibrating a Transmitter. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34

6.8 Leak Test. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35

7. Remote Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 378. Document Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37

8.1 Introduction. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37

8.2 As Found Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 48

8.3 Adjustment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 62

8.4 As Left Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 64

8.5 Viewing Test Results . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 66

8.6 Erasing Work Orders . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 68

8.7 Extended Information Display. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 68

8.8 Setting the Date and Time . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 69

9. Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7010. Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 75

10.1 Replacing Batteries . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 75

10.2 Cleaning the Unit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 75

10.3 Service Center Calibration or Repair . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 75

10.4 Replacement Parts & Accessories. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 76

Page 3: User Manual - Transcat · US, Canada, Latin America AMETEK TCI at 1-800-527-9999 Europe, Africa, Middle East AMETEK Denmark A/S at + 45 4816 8000 Asia AMETEK Singapore Pte. Ltd. at

1

1. Introduction

The AMETEK ASC321 Multifunction Process Calibrator is a handheld, battery-operated instrument that measures and sources electrical and physical parameters. The calibrator has the following features and functions:

• A dual display. The upper display is used for the measurement of volts, current, pressure, %error, scaling, switch test and pressure leaktest. The lower display can be used to measure volts, current, pressure, resistance temperature detectors (RTDs), thermocouples, frequency, and resistance, and to source pulse trains

• A thermocouple (TC) input/output terminal with automatic reference-junction temperature compensation.

• An interactive menu

• Complete RS232 interface for remote control

• Isolated read back for transmitter calibration.

• Extended and comprehensive pressure measurement capabilities with JOFRA APM

• Documenting capability for up to 32 work orders

1.1 Contacting AmetekUS, Canada, Latin America AMETEK TCI at 1-800-527-9999 Europe, Africa, Middle East AMETEK Denmark A/S at + 45 4816 8000 Asia AMETEK Singapore Pte. Ltd. at + 65 (64) 842 388

1.2 Standard EquipmentCheck to see if your calibrator is complete. It should include: ASC321 Calibrator Instruction Manual Test Leads Carrying Case

Page 4: User Manual - Transcat · US, Canada, Latin America AMETEK TCI at 1-800-527-9999 Europe, Africa, Middle East AMETEK Denmark A/S at + 45 4816 8000 Asia AMETEK Singapore Pte. Ltd. at

2

1.3 Safety information

Symbols Used

The following table lists the International Electrical Symbols. Some or all of these symbols may be used on the instrument or in this manual.

Symbol Description

AC (Alternating Current)

AC-DC

Battery

CE Complies with European Union Directives

DC

Double Insulated

Electric Shock

Fuse

PE Ground

Hot Surface (Burn Hazard)

Read the User’s Manual (Important Information)

Off

On

Canadian Standards Association

This calibrator must be recycled or disposed of properly (2002/96/EC).

Page 5: User Manual - Transcat · US, Canada, Latin America AMETEK TCI at 1-800-527-9999 Europe, Africa, Middle East AMETEK Denmark A/S at + 45 4816 8000 Asia AMETEK Singapore Pte. Ltd. at

3

The following definitions apply to the terms “Warning” and “Caution”.

• “Warning” identifies conditions and actions that may pose hazards to the user.

• “Caution” identifies conditions and actions that may damage the instrument being used.

Use the calibrator only as specified in this manual, otherwise injury and damage to the calibrator may occur.

WarningThe ASC321 is designed to calibrate and measure low voltage process signals. To ensure the safety of the operator, DO NOT connect the ASC321 to inputs greater than 30 Volts.

WarningTo avoid possible electric shock or personal injury:

• Do not apply more than the rated voltage. See specifications for supported ranges.

• Follow all equipment safety procedures.

• Never touch the probe to a voltage source when the test leads are plugged into the current terminals.

• Do not use the calibrator if it is damaged. Before you use the calibrator, inspect the case. Look for cracks or missing plastic. Pay particular attention to the insulation surrounding the connectors.

• Select the proper function and range for your measurement.

• Make sure the battery cover is closed and latched before you operate the calibrator.

• Remove test leads from the calibrator before you open the battery door.

• Inspect the test leads for damaged insulation or exposed metal. Check test leads continuity. Replace damaged test leads before you use the calibrator.

• When using the probes, keep your fingers away from the probe contacts. Keep your fingers behind the finger guards on the probes.

• Connect the common test lead before you connect the live test lead. When you disconnect test leads, disconnect the live test lead first.

• Do not use the calibrator if it operates abnormally. Protection may be impaired. When in doubt, have the calibrator serviced.

• Do not operate the calibrator around explosive gas, vapor, or dust.

Page 6: User Manual - Transcat · US, Canada, Latin America AMETEK TCI at 1-800-527-9999 Europe, Africa, Middle East AMETEK Denmark A/S at + 45 4816 8000 Asia AMETEK Singapore Pte. Ltd. at

• When using a pressure module, make sure the process pressure line is shut off and depressurized before you connect it or disconnect it from the pressure module.

• Disconnect test leads before changing to another measure or source function.

• When servicing the calibrator, use only specified replacement parts.

• To avoid false readings, which could lead to possible electric shock or personal injury, replace the battery as soon as the battery indicator appears.

• To avoid a violent release of pressure in a pressurized system, shut off the valve and slowly bleed off the pressure before you attach the pressure module to the pressure line.

• To avoid personal injury or damage to the calibrator, use only the specified replacement parts and do not allow water into the case.

CautionTo avoid possible damage to calibrator or to equipment under test:

• Disconnect the power and discharge all high-voltage capacitors before testing resistance or continuity.

• Use the proper jacks, function, and range for your measurement or sourcing application.

• If the message changes to "OL" the range limit is exceeded and the pressure source must immediately be removed from the APM to prevent damage to the pressure transducer inside.

• Maximum torque allowed is 13.5 Nm/10 ftlbs. NEVER exceed the torque allowed.

• To avoid damaging the plastic lens and case, do not use solvents or abrasive cleansers.

Clean the calibrator with a soft cloth dampened with water or water and mild soap.

4

Page 7: User Manual - Transcat · US, Canada, Latin America AMETEK TCI at 1-800-527-9999 Europe, Africa, Middle East AMETEK Denmark A/S at + 45 4816 8000 Asia AMETEK Singapore Pte. Ltd. at

5

2. Calibrator InterfaceFigure 1 shows the location of the input and output connections on the calibrator, while Table 1 describes their use.

Figure 1. Input/Output Terminals

Table 1: Input and Output Terminals

No. Name Description

1, 2 Measure Isolated V, Input terminals for measuring current, voltage, and sup-plying loop power.

3 TC input/output Terminal for measuring, or simulating thermocouples. Accepts miniature polarized thermocouple plugs with flat in-line blades spaced 7.9 mm (0.312 in) center to center.

4,5 Source/Measure Terminals for sourcing and measuring voltage, frequen-cy, pulse train, and RTDs

6,7 Source/Measure mA terminals, 3W 4W

Terminals for sourcing and measuring current, and per-forming RTD measurements with 3-wire or 4-wire setups.

8 Serial port Connects calibrator to a PC for remote control.

9 Pressure module connector

Connects calibrator to a pressure module for pressure measurements.

10 Charger Connector Connector for optional battery charger

F1 F2 F3

CE

1

ENTER

HOME

2 3

4 5 6

7 8 9

- 0 .

+V

mALoop

+V

Hz–

Ω3W

mA+

4WmA-

TC

7

8

910

6 53

4

1

2

Page 8: User Manual - Transcat · US, Canada, Latin America AMETEK TCI at 1-800-527-9999 Europe, Africa, Middle East AMETEK Denmark A/S at + 45 4816 8000 Asia AMETEK Singapore Pte. Ltd. at

Figure 2 shows the location of the keys on the calibrator. Table 2 lists the functions of each key.

F1 F2 F3

CE

1

ENTER

HOME

2 3

4 5 6

7 8 9

- 0 .

1

2

3

5

4

Table 2. Key Functions

Figure 2. Keypad

No. Name Function

1 Function Keys F1, F2, F3

Used to operate the menu bar at the bottom of the calibra-tor display. F1 is used for selecting options in the left box, F2 for the center box, and F3 for the right box.

2 Arrow Keys Used to adjust output value and for setting 0 and 100% for step and ramp

3 Home Returns to home menu on the menu bar.

4 Power Turns calibrator on and off.

5 Numeric Keypad Allows user to enter Numeric values.

6

Page 9: User Manual - Transcat · US, Canada, Latin America AMETEK TCI at 1-800-527-9999 Europe, Africa, Middle East AMETEK Denmark A/S at + 45 4816 8000 Asia AMETEK Singapore Pte. Ltd. at

7

2.1 Main Display

mA IN LOOP

0.000 mA-25.00%

RTD IN 4W P10(90) 385

MENU-OL °C

1

78

4

5

62

3 LIGHT

Figure 3. Display

The display of the calibrator, shown in Figure 3, is divided into three main sections: the upper display, the lower display, and the menu bar.

The upper display is used for measuring dc voltage, dc current with and without loop power, pressure, percent error, scaled value, switch test and leaktest.

The lower display can be used for both measuring and sourcing.

The menu bar is used to setup both the upper and the lower display to perform the desired function.

Table 3 explains the different parts of the display:

Table 3: Display Functions

No. Name Description

1 Primary Parameters

Determine what parameter is going to be measured or sourced. The available options for the upper display are:VOLTS IN, PRESSURE, mA IN, SCALED, %ERROR, SW TEST and LEAKTEST. The available options for the lower display are:VOLTS, TC (ther-mocouple), RTD, FREQ (frequency), PULSE, PRESSURE, mA, and mA 2W SIM.

2 Input/Output control

Switches the lower display between input mode (read), and output mode (source).

3 Additional Settings

Available only for TC (thermocouple), and RTD measurements. For TC this setting turns the CJC (Cold Junction Connection) on and off. For RTD measure [RTD IN], his setting sets the number of wires used in the measurement (2-wire, 3-wire, or 4-wire)

Page 10: User Manual - Transcat · US, Canada, Latin America AMETEK TCI at 1-800-527-9999 Europe, Africa, Middle East AMETEK Denmark A/S at + 45 4816 8000 Asia AMETEK Singapore Pte. Ltd. at

88

4 Secondary Upper Display

Shows where in the preset span the measured value falls. Fixed for mA at 4 (0%) and 20 (100%). Also shows mA value for percent error, mA or Volts value for scaling and timer for leak test.

5 Units Shows what unit the measurement or source value is in. Available options are for RTD and TC (°C or °F), and for FREQ and PULSE (CPM, Hz, or KHz)

6 Sensor Types Allow for measurements to be made for different types of RTDs and TCs. All types are shown in the Specifications. Also, dis-plays the amplitude of the pulse and frequency source, and pressure units.

7 Numeric Displays

Display the numeric values of the signal being measured, or sourced. An “OL” reading indicates an out of range or overload condition.

8 Loop Power Indicator

Indicates whether loop power is on or off in mA, Scaled mA and %Error.

2.2 Menu BarThe parameters on the display are controlled by the menu bar, which is located at the bottom of the LCD. The function keys (F1, F2, and F3) are used to navigate through all the levels and choices of the menu bar. Refer to the menu tree for a clarification on the layout of all the levels.

The top level of the menu is the home menu. It can be accessed anytime by pressing the HOME key. There are multiple variations of the home menu: the input home menu, the output home menu, the leaktest home menu, and the pulse home menu.

There are two options for the input home menu. They are MENU and LIGHT. The MENU option is used to enter the main menu. Press the corresponding function key (F1) to enter the main menu. The LIGHT option (F2) turns the LCD backlight on of off.

MENU LIGHT

In the first output home menu there are three active options, [MENU], [LIGHT] and [STEP or RAMP]. [MENU] and [LIGHT] serve the same purpose as in the input home menu. STEP or RAMP either steps or ramps the output.

MENU STEPLIGHT

The Leak Test home menu has three active options, [START],[LIGHT] and [MORE]. The [START] option begins the test, the [LIGHT] option runs the

Page 11: User Manual - Transcat · US, Canada, Latin America AMETEK TCI at 1-800-527-9999 Europe, Africa, Middle East AMETEK Denmark A/S at + 45 4816 8000 Asia AMETEK Singapore Pte. Ltd. at

9

backlight and the [MORE] option serves the same purpose of the [MENU] option in the other home menus.

START LIGHT MORE

The pulse home menu also has three active options, [MENU], [TRIG], and [COUNTS]. The [TRIG] and [COUNTS] options are used for pulse simulation. The function of these options is explained in Section 4.7 (Sourcing a Pulse).

The next level of the menu bar is the main menu. The levels under the main menu depend on what mode the calibrator is in.

The main menu has three active options [UPPER], [LOWER], and [MORE].

Choosing [UPPER] calls up the parameter selection menu for the upper display. Choosing [LOWER] calls up the parameter selection menu for the lower display. [MORE] enters the next menu level.

The Document Mode selection menu is next. Its options are [DOCUMENT], [NEXT], and [DONE]. Choosing [DOCUMENT] enters the document mode menu system described in section 8 of this manual, [NEXT] proceeds to the next menu level, and [DONE] returns to the home menu.

DOCUMENT NEXT DONE

The contrast menu is usually the next menu level. Its options are [CONTRAST], [NEXT], and [DONE]. The [CONTRAST] option is used to adjust contrast. [NEXT] proceeds to the auto off main menu, and [DONE] returns to home menu. Contrast is adjusted using the arrow options, which are available after choosing [CONTRAST].

The auto off main menu contains the options [AUTO OFF], [NEXT], and [DONE].

The [AUTO OFF] option is used to turn the automatic shutoff on and off and set the amount of time the unit needs to stay dormant to shut off. [NEXT] allows access to the Hart Main Menu and [DONE] returns to the home menu.

Page 12: User Manual - Transcat · US, Canada, Latin America AMETEK TCI at 1-800-527-9999 Europe, Africa, Middle East AMETEK Denmark A/S at + 45 4816 8000 Asia AMETEK Singapore Pte. Ltd. at

10

HART NEXT DONE

The Hart main menu is used to insert a 250Ω resistor into the isolated loop power circuit and the mA source circuit.

CLOCK NEXT DONE

The Clock menu is the next menu displayed in the [MORE] menu sequence. Its options are [CLOCK], [NEXT], and [DONE]. Choose [CLOCK] to set the calibrator date and time as described in section 8.8 of this manual, [NEXT] to proceed to the terminal emulation menu, and [DONE] to return to the home menu.

LOOP POWER NEXT DONE

The loop power menu is added when mA IN, %ERROR or SCALED mA is selected on the upper display. The [LOOP POWER] option allows access to the loop power control.

% ERROR NEXT DONE

The percent error main menu is added when %ERROR is selected on the upper display. The [%ERROR] option allows access to the percent error parameters.

SCALING NEXT DONE

The scaling main menu is added when SCALED is selected on the upper display. The [SCALING] option allows access to the parameters used for scaling the value on the upper display.

LT TIMER NEXT DONE

The leak test timer main menu is added when LEAK TEST is selected on the upper display. The [LT TIMER] option allows access to set the leak test timer.

Page 13: User Manual - Transcat · US, Canada, Latin America AMETEK TCI at 1-800-527-9999 Europe, Africa, Middle East AMETEK Denmark A/S at + 45 4816 8000 Asia AMETEK Singapore Pte. Ltd. at

AUTO FUNC NEXT DONE

When the lower display is in any source mode except pulse mode, the auto output main menu is added. The options available on this menu are [AUTO FUNC], [NEXT] and [DONE]. The [AUTO FUNC] option allows you to turn the RAMP/STEP functions on or off. [NEXT] is used to access the contrast main menu and done returns to the home menu.

FREQ LEVEL NEXT DONE

When the lower display is in the frequency or pulse mode, the frequency level menu is added after the auto output main menu. The options available in this menu are [FREQ LEVEL], [NEXT], and [DONE]. The [FREQ LEVEL] option is used to adjust the amplitude of the wave. [NEXT] is used to access the contrast main menu, and [DONE] returns to the home menu.

SET CUSTOM NEXT DONE

When the calibrator is in RTD CUSTOM mode, the RTD custom setup menu, is inserted after the main menu. Options [SET CUSTOM], [NEXT], and [DONE] are available. [SET CUSTOM] is used to enter a custom PRT into the calibrator. Refer to Section 4.9-1 for instructions. [NEXT] is used to enter the contrast main menu, and [DONE] to return to the home menu.

ZERO NEXT DONE

When pressure is selected on either display the pressure zeroing menu is inserted after the main menu. It has the options [ZERO ], used to zero pressure, [NEXT] and [DONE], which have the same function as above. Refer to the Section 5.3 for instructions on zeroing.

SELECT NEXT DONE

The parameter selection menu is called up when [UPPER] or [LOWER] is selected from the main menu. It contains the following options: [SELECT], [NEXT], and [DONE]. When the display is selected, a parameter will start to flash. Use the [SELECT] option to change the parameter, and the [NEXT] option to switch to another variable. [DONE] returns to the home menu and enables the selected mode.

11

Page 14: User Manual - Transcat · US, Canada, Latin America AMETEK TCI at 1-800-527-9999 Europe, Africa, Middle East AMETEK Denmark A/S at + 45 4816 8000 Asia AMETEK Singapore Pte. Ltd. at

12

UPPERSelection M

enuLOW

ERM

ORE

SELECTParam

eter SelectionNEXT

DONE

FREQ LEVELFrequency LevelNEXT

DONESET CUSTOMRTD Custom

Menu

If calibrator is inFrequency Outor Pulse m

ode

NEXTDONE

ZEROPressure Zeroing

NEXTDONE

CONTRASTContrast M

enuNEXTDONE

AUTO OFFAuto Off M

enuNEXTDONE If calibrator is in

RTD CUSTOMm

ode

If calibrator is inPressure m

ode

If calibrator is inSource m

odeIf calibrator is inM

easure mode

AUTO FUNCAuto Output M

ain Menu

NEXTDONE

%ERROR

Percent ErrorNEXT

DONE

If calibrator is inPercent Errorm

ode

SCALINGScaling M

ain Menu

NEXTDONE

If calibrator is inScale Inputm

ode

LT TIMER

Leaktest TimerNEXT

DONE

If calibrator is inLeaktest m

ode

HARTHart M

ain MenuNEXT

DONE

MENU

Home M

enuLIGHT

MENU

Home M

enuLIGHT

MENU

Home M

enuLIGHTM

ENUHom

e MenuLIGHT

MENU

Home M

enuLIGHTM

ENUHom

e MenuLIGHT

MENU

Home M

enuLIGHT

MENU

Home M

enuLIGHT

MENU

Home M

enuLIGHT

MENU

Home M

enuLIGHT

MENU

Home M

enuLIGHT

DOCUMENT

NEXTDONEDONE

CLOCKNEXT

DONE

Document Selection M

enu

Clock Selection Menu

Figu

re 4. Sim

plified

Men

u Tree

Page 15: User Manual - Transcat · US, Canada, Latin America AMETEK TCI at 1-800-527-9999 Europe, Africa, Middle East AMETEK Denmark A/S at + 45 4816 8000 Asia AMETEK Singapore Pte. Ltd. at

13

3. Using Measure Modes (Lower Display)

3.1 Measuring volts and frequencyElectrical parameters volts and frequency can be measured using the lower display. To make the desired measurements, follow these steps:

1. Switch to the lower display [LOWER] from Main Menu.

2. Select the desired parameter for measurement.

3. Connect leads as shown in Figure 5.

VOLTS IN

-0.000 VVOLTS IN

MENU5.000 V

LIGHT

+V

mALoop

+V

Hz–

Ω

3WmA+

4WmA-

TC

+

Figure 5. Measuring Volts and Frequency with Input/Output Terminals

3.2 Measuring mATo measure mA follow these steps:

1. Switch to lower display and select mA.

2. Make sure the input/output control is set to IN.

3. Connect leads as shown in Figure 6.

mA IN

0.000 mA-25.00%

mA IN

MENU5.000 mA

LIGHT

+V

mALoop

+V

Hz–

Ω

3WmA+

4WmA-

TC–

+

Figure 6. Measuring mA with Input/Output Terminals

Page 16: User Manual - Transcat · US, Canada, Latin America AMETEK TCI at 1-800-527-9999 Europe, Africa, Middle East AMETEK Denmark A/S at + 45 4816 8000 Asia AMETEK Singapore Pte. Ltd. at

14

3.3 Measuring Temperature

3.3.1 Using Thermocouples mA IN

0.000 mA-25.00%

TC IN CJC OFF B

MENUOL °C

LIGHT

The calibrator supports the following thermocouple types: B, C, E, J, K, L, N, R, S, T, U, BP, and XK. The characteristics of all the types are described in Specifications section. The calibrator also has a Cold Junction Compensation (CJC) function. Normally this function should be ON and the actual temperature of the thermocouple will be measured. With CJC OFF, the calibrator will measure the difference between the thermocouple at the junction and at its TC input terminal.

Note: CJC off mode should only be used when calibration is being done using an external ice bath.

To use the thermocouple to measure temperature, follow these steps:

1. Attach the thermocouple leads to the TC miniplug, and insert the plug into the input/output of the calibrator, as in Figure 7.

Note: For best accuracy wait 2 to 5 minutes for the temperature between the miniplug and the calibrator to stabilize before any measurements are taken.

2. Switch to lower display from Main Menu.

3. Select TC from the primary parameters. Choose [IN] in the input/output control, and than the thermocouple type from the sensor types. The temperature unit may also be changed from Celsius to Fahrenheit.

The calibrator can also measure the mV of a Thermocouple, which can be used along with a table in case the corresponding TC type is not supported by the calibrator. To do so, proceed as above and choose mV from sensor types.

+V

mALoop

+V

Hz–

Ω

3WmA+

4WmA-

TC

+–

TC PLUG

Figure 7. Measuring Temperature Using Thermocouple Terminals

Note: The TC wire used must match the thermocouple type being calibrated.

Page 17: User Manual - Transcat · US, Canada, Latin America AMETEK TCI at 1-800-527-9999 Europe, Africa, Middle East AMETEK Denmark A/S at + 45 4816 8000 Asia AMETEK Singapore Pte. Ltd. at

15

3.3.2 Using Resistance-Temperature-Detectors (RTDs)

The supported types of RTDs are shown in Section 9. Specifications. RTDs are characterized by their 0°C resistance, R0. The calibrator accepts two, three, and four wire inputs, with four wire input being the most accurate.

To use the RTD option, apply the following steps:

1. Switch to lower display [LOWER] from Main Menu.

2. Select RTD from the primary parameters. Select [IN] from input/output control.

3. Choose 2, 3, or 4-wire connection [2W, 3W, 4W]. (4-wire allows for the most precise measurement)

4. Select RTD type from the sensor types.

5. Attach RTD leads as shown in Figure 8.

+V

mALoop

+V

Hz–

Ω

3WmA+

4WmA-

TC

+

+

+

+

+

+

RTD2W

3W

4W

RTD

RTD

mA IN

0.000 mA-25.00%

RTD IN 4W P10(90) 385

MENU25.0 °C

LIGHT

Figure 8. Measuring Temperature with RTD Connection

Resistance can also be measured using this function. To do so, use the above procedure and choose OHMS from the sensor types. This option can be used along with a table to measure an RTD which is not programmed into the calibrator.

3.4 Measuring Pressure

Warning!To avoid a violent release of pressure in a pressurized system, shut off the valve and slowly bleed off the pressure before you attach the pressure module to the pressure line.

Page 18: User Manual - Transcat · US, Canada, Latin America AMETEK TCI at 1-800-527-9999 Europe, Africa, Middle East AMETEK Denmark A/S at + 45 4816 8000 Asia AMETEK Singapore Pte. Ltd. at

16

CautionTo avoid mechanically damaging the pressure module, never apply more than 13.5 Nm/10 ftlbs. of torque between the pressure module fittings, or between the fittings an the body of the module.

To avoid damaging the pressure module from overpressure, never apply pressure above the rated maximum printed on the module.

To avoid damaging the pressure module from corrosion, use it only with specified materials. Refer to the pressure module documentation for material compatibility.

To measure pressure, follow these steps:

1. Connect the pressure module to the calibrator as shown in Figure 9.

The calibrator can measure pressure on both the upper and the lower display. This makes it possible to measure pressure in two different units at the same time.

2. Switch to either upper or lower display from the Main Menu.

3. Select [PRESSURE] from the primary parameters.

4. Select the desired measuring unit.

5. Zero the pressure module. The zero function on the calibrator can be found in the pressure zeroing menu.

PRESSUREMODULE

VALVE

F1 F2 F3

CE

1

ENTER

HOME

2 3

4 5 6

7 8 9

- 0 .

Figure 9. Connections for Measuring Pressure

3.4.1 Zeroing with Absolute Pressure Modules.

To zero, adjust the calibrator to read a known pressure, such as barometric pressure.

Page 19: User Manual - Transcat · US, Canada, Latin America AMETEK TCI at 1-800-527-9999 Europe, Africa, Middle East AMETEK Denmark A/S at + 45 4816 8000 Asia AMETEK Singapore Pte. Ltd. at

To adjust the calibrator, follow these steps:

1. Enter the pressure zeroing menu.

2. Select [ZERO ]. [SET REFERENCE ABOVE] will appear. Enter the pressure using the keypad.

3. The calibrator stores the Barometric zero offset in non-volatile memory.

The zero offset is stored for one absolute pressure module at a time. If a new absolute module is connected this process must be repeated.

PRESSURE

-------psi

PRESSURE mbar

MENU

-------LIGHT

4. Using Source Modes (Lower Display)The calibrator can generate calibrated signals for testing and calibrating process instruments. It can source voltages, currents, resistances, frequencies, pulses, and the electrical output of RTD and thermocouple temperature sensors.

4.1 Setting 0% and 100% Output Parameters To set the 0% and 100% points, use the following steps:

1. Select the lower display [LOWER] from Main Menu, and choose the desired primary parameter.

2. Select output [OUT] from the input/output control, and enter the desired value. For example select [VOLTS OUT].

VOLTS IN

0.000 VVOLTS OUT

0%2.500 V

25% 100%

3. Enter 5 V with the keypad. Press the right arrow key to bring up alternate functions for the function keys. Press and hold the 0% key until 0% flashes on the screen.

17

Page 20: User Manual - Transcat · US, Canada, Latin America AMETEK TCI at 1-800-527-9999 Europe, Africa, Middle East AMETEK Denmark A/S at + 45 4816 8000 Asia AMETEK Singapore Pte. Ltd. at

18

4. Enter 20 V with the keypad. Press the right arrow key to bring up alternate functions for the function keys. Press and hold the 100% key until 100% flashes on the screen.

5. Use the F2 key to cycle 5 V and 20 V in either 10% or 25% increments, depending on which is selected.

4.1-1 Manually stepping the current output

To use the 25% function with mA output, follow these steps:

1. Select the lower display from the Main Menu, and choose mA.

2. Use the 25% key to cycle between 4 mA and 20 mA in 25% intervals.

Note: You can also configure the unit to output in 10% steps.

4.2 Using Auto Output FunctionsThere are two Auto Output functions; [STEP] and [RAMP]. When the [STEP] function is selected the calibrator steps from the 0% set source value to the 100% set source value in 10% or 25% increments. The word [STEP] turns to [STEPPING] when the [STEP] function is active. The [RAMP] function ramps the calibrator output from the 0% to 100% set source values. When the [RAMP] function is active the word [RAMP] turns to [RAMPING]. Both the [STEP] and [RAMP] functions will continue from 0% to 100% and back again until a key is pressed.

Note: Auto Output options are not available for the Pulse Output function.

4.3 Sourcing mA To source a current, follow these steps:

1. From the Main Menu select lower display [LOWER]. Choose [mA] from the primary parameters.

2. Switch to input/output control, and select output [OUT].

3. Connect the leads to the mA terminals, as shown in Figure 10.

4. Enter the desired current using the keypad.

mA IN LOOP

0.000 mA-25.00 %

mA OUT

MENU5.000 mA

LIGHT

+V

mALoop

+V

Hz–

Ω

3WmA+

4WmA-

TC

+

1000 ohms max.

UUT

Figure 10. Connections for Sourcing Current

Page 21: User Manual - Transcat · US, Canada, Latin America AMETEK TCI at 1-800-527-9999 Europe, Africa, Middle East AMETEK Denmark A/S at + 45 4816 8000 Asia AMETEK Singapore Pte. Ltd. at

19

4.4 Simulating a TransmitterTo have the calibrator supply a variable test current to a loop in place of a transmitter, follow these steps:

1. Select lower display from the Main Menu.

2. Choose mA simulation from the primary parameters [mA 2W SIM], and enter the desired current.

3. Connect the 24V loop as shown in Figure 11.

mA IN LOOP

0.000 mA-25.00%

mA 2W SIM

MENU4.000 mA

LIGHT

+V

mALoop

+V

Hz–

Ω

3WmA+

4WmA-

TC

+

+LOOPPOWERSUPPLY

30 VDCMAX

UUT

Figure 11. Connections for Simulating a Transmitter

4.5 Sourcing voltsTo source volts follow these steps:

1. Select lower display from the Main Menu.

2. Choose [VOLTS] from the primary parameters. Switch to input/output control and select output [OUT].

3. Connect the leads for the voltage source terminals, as shown in Figure 12.

4. Enter the voltage using the keypad.

mA IN LOOP

0.000 mA-25.00%

VOLTS OUT

MENU6.000 V

LIGHT

+V

mALoop

+V

Hz–

Ω

3WmA+

4WmA-

TC

+

UUT

Figure 12. Connections for Sourcing Voltage and Frequency

4.6 Sourcing frequencyTo source a signal use these steps:

1. Switch to the lower display and select frequency from the primary parameters.

Page 22: User Manual - Transcat · US, Canada, Latin America AMETEK TCI at 1-800-527-9999 Europe, Africa, Middle East AMETEK Denmark A/S at + 45 4816 8000 Asia AMETEK Singapore Pte. Ltd. at

20

2. Select output, and than choose the frequency units.

3. Connect the leads to the frequency output terminals as shown in Figure 12.

4. Enter the desired frequency using the keypad.

5. To change the amplitude, select [FREQ LEVEL] from frequency level menu.

6. Enter the amplitude.

4.7 Sourcing a pulse trainThe calibrator can produce a pulse train with an adjustable number of pulses at a desired frequency. For example, setting the frequency to 60Hz and the number of pulses to 60 would produce 60 pulses for a period of 1 second. To source a pulse, use the same connection as for frequency, and proceed as follows:

1. Switch to the lower display and select pulse from the primary parameters.

2. Choose the desired unit and enter the frequency using the keypad.

3. Select the [COUNTS] function from the home menu to enter the number of pulses. Use [TRIG] to start and stop the signal.

4. The amplitude of the pulse can be adjusted in the same manner as for frequency.

mA IN LOOP

0.000 mA-25.00%

PULSE 5.0 Vpp

MENU5.0 kHzTRIG COUNTS

mA IN LOOP

0.000 mA-25.00%

FREQ OUT 5.0 Vpp

MENU4.0 kHz

LIGHT

+V

mALoop

+V

Hz–

Ω

3WmA+

4WmA-

TC–+

+–

TC PLUGNote: TC wire used must matchthe thermocouple type being calibrated.

Figure 13. Connections for Outputting Thermocouples

Page 23: User Manual - Transcat · US, Canada, Latin America AMETEK TCI at 1-800-527-9999 Europe, Africa, Middle East AMETEK Denmark A/S at + 45 4816 8000 Asia AMETEK Singapore Pte. Ltd. at

21

4.8 Sourcing ThermocouplesTo source a thermocouple use the following steps:

1. Connect the thermocouple leads to the appropriate polarized TC miniplug, and insert the plug into the TC terminals on the calibrator, as shown in Figure 13.

2. Select lower display from the Main Menu, and choose thermocouple [TC] from the primary parameters.

3. Choose output [OUT] from the input/output control.

4. Select the desired thermocouple type from the sensor types.

5. Enter the temperature using the keypad.

mA IN

0.000 mA-25.00%

TC OUT CJC OFF E

MENU100.0 °C

LIGHT

+V

mALoop

+V

Hz–

Ω

3WmA+

4WmA-

TC

Figure 14. Connections for Outputting RTDs

4.9 Sourcing Ohms/RTDsTo source an RTD, follow these steps:

1. Select lower display from the Main Menu, and choose [RTD] from the primary parameters.

2. Choose output [OUT] from the input/output control, and select RTD type from the sensor types.

3. Connect the calibrator to the instrument being tested, as in Figure 14.

4. Enter the temperature or resistance using the keypad.

Page 24: User Manual - Transcat · US, Canada, Latin America AMETEK TCI at 1-800-527-9999 Europe, Africa, Middle East AMETEK Denmark A/S at + 45 4816 8000 Asia AMETEK Singapore Pte. Ltd. at

22

+

+

+

+

+

+

3W 4W

1 1

2 2

3 3

4 4

Use stackabletest leads for

shared connections

Figure 15. Using a 3- or 4-wire Connection for RTDs

mA IN LOOP

0.000 mA-25.00%

RTD OUT P500(90)385

MENU10.0 °CLIGHT

Note: The calibrator simulates a 2-wire RTD. To connect 3- or 4-wire transmitter, use stacking cables, as shown in Figure 15.

4.9.1 Custom RTD

A custom curve-fit PRT may be entered into the calibrator for sourcing and measuring. To do so follow these steps:

1. Switch to lower display. Select RTD and set sensor type to CUSTOM.

2. Enter the RTD custom setup main menu, and select [SET CUSTOM].

3. Using the keypad, enter the values that the calibrator prompts for: minimum temperature, maximum temperature, R0, and the values for each of the temperature coefficients.

Page 25: User Manual - Transcat · US, Canada, Latin America AMETEK TCI at 1-800-527-9999 Europe, Africa, Middle East AMETEK Denmark A/S at + 45 4816 8000 Asia AMETEK Singapore Pte. Ltd. at

The custom function uses the Calendar-Van Dusen equation for outputting and measuring custom RTDs. The coefficient C is only used for temperatures below 0°C. Only A and B coefficients are needed for the range above 0°C, so coefficient C should be set to 0. The R0 is the resistance of the probe at 0°C. The ITS 90 coefficients for PT385, PT3926, and PT3616 are shown in Table 4.

Table 4. RTD Coefficients

RTD Range(°C) R0 Coefficient A Coefficient B Coefficient C

PT385 -260 - 0 100 3.9083x10-3 -5.775x10-7 -4.183x10-12

PT385 0 - 630 100 3.9083x10-3 -5.775x10-7 ---

PT3926 Below 0 100 3.9848x10-3 -5.87x10-7 -4x10-12

PT3926 Above 0 100 3.9848x10-3 -5.87x10-7 ---

PT3916 Below 0 100 3.9692x10-3 -5.8495x10-7 -4.2325x10-12

PT3916 Above 0 100 3.9692x10-3 -5.8495x10-7 ---

5. Using Isolated Measure Modes (Upper Display)

5.1 Measuring volts and mAUse the following steps to measure the voltage or current output of a transmitter.

1. Select the upper display from the Main Menu.

2. Select the desired primary parameter to be measured. Connect the leads to the isolated inputs of the calibrator, as in Figure 16.

+V

mALoop

+V

Hz–

Ω

3WmA+

4WmA-

TC

+

Figure 16. Isolated Input Connection

5.2 Measuring current with loop powerTo test a 2-wire, loop powered transmitter that is disconnected from wiring, use the loop power function. This function activates a 24V supply in series with the current measuring circuit. To use this option proceed as follows:

23

Page 26: User Manual - Transcat · US, Canada, Latin America AMETEK TCI at 1-800-527-9999 Europe, Africa, Middle East AMETEK Denmark A/S at + 45 4816 8000 Asia AMETEK Singapore Pte. Ltd. at

1. Select [mA] as primary parameter in the upper display.

2. Use the menu to turn on loop power. Connect the calibrator to transmitter current loop terminals, as shown in Figure 17.

+V

mALoop

+V

Hz–

Ω

3WmA+

4WmA-

TC

–+

Figure 17. Connection Using Current Loop

mA IN LOOP

0.000 mA-25.00%

mA IN

MENU-0.001 mA

LIGHT

5.3 Measuring Pressure

Warning!To avoid a violent release of pressure in a pressurized system, shut off the valve and slowly bleed off the pressure before you attach the pressure module to the pressure line.

CautionTo avoid mechanically damaging the pressure module, never apply more than 13.5 Nm/10 ftlbs. of torque between the pressure module fittings, or between the fittings an the body of the module.

To avoid damaging the pressure module from overpressure, never apply pressure above the rated maximum printed on the module.

To avoid damaging the pressure module from corrosion, use it only with specified materials. Refer to the pressure module documentation for

24

Page 27: User Manual - Transcat · US, Canada, Latin America AMETEK TCI at 1-800-527-9999 Europe, Africa, Middle East AMETEK Denmark A/S at + 45 4816 8000 Asia AMETEK Singapore Pte. Ltd. at

material compatibility.

To measure pressure, follow these steps:

1. Connect the pressure module to the calibrator as shown in Figure 18.

The calibrator can measure pressure on both the upper and the lower display. This makes it possible to measure pressure in two different units at the same time.

2. Switch to either upper or lower display from the Main Menu.

3. Select [PRESSURE] from the primary parameters.

4. Select the desired measuring unit.

5. Zero the pressure module. The zero function on the calibrator can be found in the pressure zeroing menu. For zeroing absolute modules see 3.4.1.

PRESSUREMODULE

VALVE

F1 F2 F3

CE

1

ENTER

HOME

2 3

4 5 6

7 8 9

- 0 .

Figure 18. Measuring Pressure Transmitter

6. Using the Upper and the Lower Display for Calibration and Testing

6.1 Performing a Switch TestThe calibrator can stimulate a switch and capture the results in any source/simulate mode (except pulse) and the calibrator can also monitor a switch and capture the results in any read mode based on the input requirements of the switch. Examples are given below for a temperature switch with an external temperature source and a pressure switch.

25

Page 28: User Manual - Transcat · US, Canada, Latin America AMETEK TCI at 1-800-527-9999 Europe, Africa, Middle East AMETEK Denmark A/S at + 45 4816 8000 Asia AMETEK Singapore Pte. Ltd. at

26

6.1.1 Performing a Temperature Switch Test:

RTD Probe

TemperatureSwitch

+V

mALoop

+V

Hz–

Ω

3WmA+

4WmA-

TC

Figure 19. Temperature Switch Test

1. Use the menu to select SW TEST on the upper display.

2. Connect the calibrator to the switch using the switch terminals. The polarity of the terminals does not matter.

P100-385

SW TEST CLOSE

RTD IN 4W

100.0 °C

------ °C

MENU LIGHT

3. The top of the display will read “CLOSE”.

4. Raise the temperature slowly until the switch opens.

5. Once the switch is open, “OPEN” will be displayed, lower the temperature slowly until the switch closes.

Page 29: User Manual - Transcat · US, Canada, Latin America AMETEK TCI at 1-800-527-9999 Europe, Africa, Middle East AMETEK Denmark A/S at + 45 4816 8000 Asia AMETEK Singapore Pte. Ltd. at

27

P100-385

SW TEST OPENED AT

RTD IN 4W

NEW TEST100.0 °C

NEXT

100.0 °C

DONE

6. At the top of the display it will now read, “SW OPENED AT” and give you the temperature that the switch opened at.

P100-385

SW TEST OPEN

RTD IN 4W

MENU100.0 °C

LIGHT

------ °C

7. Press the “NEXT” option to view when the switch closed, and the dead band.

8. Press the “NEW TEST” option to get back to the main menu, clear the data and perform another test.

Important NOTE: The previous example uses a normally closed switch. The basic procedure is still the same for a normally open switch, the display will just read “OPEN” instead of “CLOSE”.

Page 30: User Manual - Transcat · US, Canada, Latin America AMETEK TCI at 1-800-527-9999 Europe, Africa, Middle East AMETEK Denmark A/S at + 45 4816 8000 Asia AMETEK Singapore Pte. Ltd. at

28

6.1.2 Performing a pressure switch test:

HAND PUMP

PRESSUREMODULE

Pressure

Switch F1 F2 F3

CE

1

ENTER

HOME

2 3

4 5 6

7 8 9

- 0 .

Figure 20. Pressure Switch Test

1. Use the menu to select SW TEST on the upper display.

2. Connect the calibrator to the switch using the pressure switch terminals. The polarity of the terminals does not matter. Then connect the pump to the calibrator and the pressure switch.

3. Make sure the vent on the pump is open. Zero the calibrator if necessary. Close vent after zeroing the calibrator.

4. The top of the display will read “CLOSE”.

psi

SW TEST CLOSE

MENU0.000

LIGHT

------ psi

5. Apply pressure with the pump slowly until the switch opens.

Important NOTE: In the switch test mode the display update rate is increased to help capture changing pressure inputs. Even with this enhanced sample rate pressurizing device under test should be done slowly to ensure accurate readings.

Page 31: User Manual - Transcat · US, Canada, Latin America AMETEK TCI at 1-800-527-9999 Europe, Africa, Middle East AMETEK Denmark A/S at + 45 4816 8000 Asia AMETEK Singapore Pte. Ltd. at

29

6. Once the switch is open, “OPEN” will be displayed, bleed the pump slowly until the pressure switch closes.

psi

SW TEST OPEN

MENU28.000

LIGHT

------ psi

7. At the top of the display it will now read, “SW OPENED AT” and give you the pressure that the switch opened at.

psi

SW OPENED AT

MENU28.000

LIGHT

30.000 psi

8. Press the “NEXT” option to view when the switch closed, and the dead band.

Important NOTE: The previous example uses a normally closed switch. The basic procedure is still the same for a normally open switch, the display will just read “OPEN” instead of “CLOSE”.

6.2 Testing an Input or Indicating DeviceTo test and calibrate actuators, recording, and indicating devices using the source functions, follow these steps:

1. Select the lower display and choose the correct primary parameter.

2. Switch to [OUT] in the input/output control.

3. Connect the leads to the instrument and the calibrator as shown in Figure 21.

Page 32: User Manual - Transcat · US, Canada, Latin America AMETEK TCI at 1-800-527-9999 Europe, Africa, Middle East AMETEK Denmark A/S at + 45 4816 8000 Asia AMETEK Singapore Pte. Ltd. at

30

+V

mALoop

+V

Hz–

Ω

3WmA+

4WmA-

TC

1 mA max. current

Figure 21. Connections for Testing an Output Device

6.3 Calibrating an I/P DeviceThe following steps show how to calibrate a device that controls pressure:

1. Select upper display from the Main Menu, and select pressure from the primary parameters.

2. Switch to lower display from the Main Menu, and select current source [mA out] from the primary parameters.

3. Connect the calibrator to the device as shown in Figure 22. The calibrator will simulate the transmitter current and measure the output pressure.

4. Enter a current using the keypad.

+V

mALoop

+V

Hz–

Ω

3WmA+

4WmA-

TC

–+

SIG

NA

L

TES

T

Figure 22. Calibrating an I/P Device

Page 33: User Manual - Transcat · US, Canada, Latin America AMETEK TCI at 1-800-527-9999 Europe, Africa, Middle East AMETEK Denmark A/S at + 45 4816 8000 Asia AMETEK Singapore Pte. Ltd. at

31

6.4 Calibrating a Transmitter To calibrate a transmitter both the upper and the lower displays will be used; one for measuring and the second a source. This section covers all but the pressure transmitters. A thermocouple temperature transmitter is used in this example.

The following steps show how to calibrate a temperature transmitter:

1. From the Main Menu select upper display, and choose current loop [mA]. Use the LOOP POWER menu to turn loop power on.

2. Switch to lower display from the Main Menu, and select [TC] from the primary parameters. Choose output [OUT] from the input/output control, and select TC type.

3. Set the 0% and 100% span points using the keypad and the 0% and 100% keys (refer to Setting 0% and 100% Parameters section).

4. Connect the calibrator to the transmitter as shown in Figure 21.

5. Test transmitter at 0- 25- 50- 75- 100% using the 25% step function.

Adjust the transmitter as necessary.

To calibrate a different transmitter, follow the above steps with the exception of choosing TC on the lower display. Replace TC with the correct parameter for the transmitter.

+V

mALoop

+V

Hz–

Ω

3WmA+

4WmA-

TC

Figure 23. Calibrating a Transmitter

6.5 Calibrating a Pressure TransmitterTo calibrate a pressure transmitter, use these steps:

1. Select upper display from the Main Menu, and choose current [mA LOOP] from the primary parameters. Return to Main Menu.

2. Select lower display, and choose [PRESSURE] from the primary parameters.

3. Connect the calibrator to the transmitter and the pressure module as in Figure 24.

4. Zero the pressure module.

Page 34: User Manual - Transcat · US, Canada, Latin America AMETEK TCI at 1-800-527-9999 Europe, Africa, Middle East AMETEK Denmark A/S at + 45 4816 8000 Asia AMETEK Singapore Pte. Ltd. at

32

5. Test the transmitter at 0% and 100% of the span, and adjust as necessary.

+V

mALoop

+V

Hz–

Ω

3WmA+

4WmA-

TC

–+

HAND PUMP

PRESSUREMODULE

Figure 24. Calibrating a Pressure Transmitter

6.6 Using Scaled Current or Voltage when testing or calibrating a Transmitter.

The calibrator has the ability to read current or voltage on the upper display that is scaled to and displayed in the same units of the lower display.

1. Select the [SCALED] option on the upper display. The unscaled value of either voltage or current will be displayed in the top right corner. The main display will read out the value of the current or voltage scaled to the lower display.

2. On the lower display select the appropriate function.

3. From the Scaling Main Menu set the 0% and 100% point for the lower display. Select mA or Volts for the upper display. Set the 0% and 100% for the upper display.

SCALED

0.0 °C4.000mA

Enter 0% of Scale

For Scaling Function0.0 °C

Page 35: User Manual - Transcat · US, Canada, Latin America AMETEK TCI at 1-800-527-9999 Europe, Africa, Middle East AMETEK Denmark A/S at + 45 4816 8000 Asia AMETEK Singapore Pte. Ltd. at

SCALED

0.0 °C4.000mA

Enter 100% of Scale

For Scaling Function200 °C

SCALED

0.0 °C4.000mA

Upper Function

VOLTS mA NEXT

mA

SCALED

0.0 °C4.000mA

Enter 0% Current

4.000 mAFor Scaling Function

SCALED

0.0 °C4.000mA

Enter 100% Current

20.000 mAFor Scaling Function

Applications of the scaling function include the calibrations described above. In most cases the transmitter output will be adjusted so the upper and lower display read the same value.

33

Page 36: User Manual - Transcat · US, Canada, Latin America AMETEK TCI at 1-800-527-9999 Europe, Africa, Middle East AMETEK Denmark A/S at + 45 4816 8000 Asia AMETEK Singapore Pte. Ltd. at

34

+V

mALoop

+V

Hz–

Ω

3WmA+

4WmA-

TC

SCALED LOOP

100.0 °C0.000mA

TC OUT CJC ON J

MENU100.0 °CLIGHT STEP

Figure 25.

6.7 Using Percent Error when testing or calibrating a Transmitter.

The calibrator features a function which can calculate signal or pressure vs. milliamp error as a percentage of the 4 to 20 mA loop span.

1. Select the [%ERROR] option on the upper display. The mA value will be displayed in the top right corner. The main display will read out the error percentage.

2. On the lower display select the appropriate function.

3. From the %ERROR Main Menu set the 0% and 100% point for the lower display.

% ERROR

-25.00 %0.000mA

Enter 0% Value

0.0 °CFor % Error Function

% ERROR

-25.00 %0.000mA

Enter 100% Value

1000.0 °CFor % Error Function

You can now do the calibration using %Error.

Page 37: User Manual - Transcat · US, Canada, Latin America AMETEK TCI at 1-800-527-9999 Europe, Africa, Middle East AMETEK Denmark A/S at + 45 4816 8000 Asia AMETEK Singapore Pte. Ltd. at

35

+V

mALoop

+V

Hz–

Ω

3WmA+

4WmA-

TC

%ERROR LOOP

0.00 %0.000mA

TC OUT CJC ON J

0%75.0 °C

25% 100%

Figure 26.

6.8 Leak TestThe calibrator features a leak test function that works with a pressure module to calculate leak rate. The timer can be set from 5 to 120 seconds, regardless of the set time the leak rate is calculated in leak per minute. The function gives a good and repeatable expression for the leak of a pressure system. This feature can be used before calibration to document / indicate leak rate.

1. Select Pressure on the lower display.

2. Connect the calibrator to the transmitter (or whatever is being tested) and the pressure module as in Figure 24.

LEAK TEST

----100.00

------S

PRESSUREpsi

START LIGHT MORE

psi

LEAK TEST

30 Sec

------S

Enter Leak Test Period

Set Leak Test Timer

---- psi

Page 38: User Manual - Transcat · US, Canada, Latin America AMETEK TCI at 1-800-527-9999 Europe, Africa, Middle East AMETEK Denmark A/S at + 45 4816 8000 Asia AMETEK Singapore Pte. Ltd. at

3. Select [LEAK TEST] on the upper display.

4. If necessary set the timer through the Leak Test Timer Main Menu.

5. Use the START key to start the test. When the test begins the initial pressure will be shown on the upper display.

LEAK RATE

-0.05 /M99.00

------S

PRESSURE

NEW TEST NEXT DONE

psi

INITIAL

99.00

------S

PRESSURE

NEW TEST NEXT DONE

100.00 psi

psi

FINAL

99.75

99.00

------S

PRESSURE

NEW TEST NEXT DONE

psi

psi

6. If you want to end the test early use the STOP key

7. When the test is complete use the NEXT key to toggle through the leak rate, the initial and final pressures.

8. Use the NEW TEST or DONE key to exit the leak test recall mode and start a new test.

36

Page 39: User Manual - Transcat · US, Canada, Latin America AMETEK TCI at 1-800-527-9999 Europe, Africa, Middle East AMETEK Denmark A/S at + 45 4816 8000 Asia AMETEK Singapore Pte. Ltd. at

37

7. Remote OperationThe calibrator can be remotely controlled using a PC terminal, or by a computer program running the calibrator in an automated system. It uses an RS232 serial port connection for remote operation. With this connection the user can write programs on the PC, with Windows languages like Visual Basic to operate the calibrator, or use a Windows terminal, such as Hyper Terminal, to enter single commands. Typical RS232 remote configurations are shown in Figure 27.

Optional RS-232 or USB Cable

(See section 7.4)

COM PORT USB PORT

OR

F1 F2 F3

CE

1

ENTER

HOME

2 3

4 5 6

7 8 9

- 0 .

Figure 27. Calibrator-to-Computer Connection

8. Document Mode

8.1 IntroductionDocument mode allows an instrumentation technician to download predefined calibration tests from a PC database for up to 32 work orders. Tests may be defined for switch instruments and analog instruments.

For analog instruments, each test consists of 1 to 21 user selectable test points.

For switch instruments, each test consists of 1 to 5 user defined ramp operations to determine the switch trip and reset points.

For both instrument types, a test may be repeated as many times as necessary to complete the adjustment and calibration of the instrument. Two sets of test results are stored for each work order, an initial As Found set stored during the first test performed, and a final As Left set stored during the last test performed. If only a single test is required to confirm an instrument’s calibration, both sets of test results are the same.

To perform an initial As Found test, the technician selects the work order to be tested from a list on the display. The calibrator cycles through the test

Page 40: User Manual - Transcat · US, Canada, Latin America AMETEK TCI at 1-800-527-9999 Europe, Africa, Middle East AMETEK Denmark A/S at + 45 4816 8000 Asia AMETEK Singapore Pte. Ltd. at

38

points one at a time automatically generating the predefined output values in the case of a calibrator source type, or displaying the predefined values to be set externally for a pressure or manual source type. Once the input has settled, the technician presses a function key to save the actual output and actual input values and step to the next point. Once the last point has been saved, the As Found test automatically ends and the technician is prompted for changes to the work order identification data (any field but work order number can be changed). After that the technician has the option of displaying the As Found test results. Finally, the technician has the option of saving the As Found results into the As Left storage and ending the instrument calibration, of entering an adjustment mode to monitor physical adjustments to the instrument prior to starting an As Left test, or of directly entering the As Left test.

During adjustment mode, the technician can step up and down the list of test points and the calibrator automatically generates the corresponding predefined output value in the case of a calibrator source type, or displays the predefined value to be set externally for a pressure or manual source type. In this way any changes to the predefined test points resulting from physical adjustments to the instrument may be evaluated. The technician can also enter source values different from the test points and monitor the corresponding input values. When the adjustment is complete, a function key is pressed to enter the As Left test.

During the As Left test, the calibrator cycles through the test points one at a time automatically generating the predefined output values in the case of a calibrator source type, or displaying the predefined values to be set externally for a pressure or manual source type. Once the input has settled, the technician presses a function key to save the actual output and actual input values and step to the next point. Once the last point has been saved, the As Left test automatically ends and the technician has the option of displaying the As Left test results. Finally, the technician has the option of ending the instrument calibration, of entering an adjustment mode to monitor physical adjustments to the instrument prior to starting a further As Left test, or of directly entering a further As Left test.

Additional As Left tests may be performed after instrument calibration has ended as long as the original work order has not been erased from storage. A menu item is available for selecting any work order and starting another As Left test.

Review functions are available to view test results on the calibrator display.

Individual or all work orders may be erased from storage using menu functions.

Page 41: User Manual - Transcat · US, Canada, Latin America AMETEK TCI at 1-800-527-9999 Europe, Africa, Middle East AMETEK Denmark A/S at + 45 4816 8000 Asia AMETEK Singapore Pte. Ltd. at

39

8.1.1 Document Mode Menu Tree

Document Menu

Function Selection Menu TEST REVIEW EXIT Home Menu

Review Menu

Test Selection Menu AS FOUND AS LEFT EXIT

As Left Menu

Information Selection SELECT

EXIT

Document Menu

As Found Tag Selection Menu SELECT INFO EXIT

Document Menu

As Found Test

Information Display

EXIT

Page 42: User Manual - Transcat · US, Canada, Latin America AMETEK TCI at 1-800-527-9999 Europe, Africa, Middle East AMETEK Denmark A/S at + 45 4816 8000 Asia AMETEK Singapore Pte. Ltd. at

40

Pulse mode? Yes

No

AF data collection – pulse trigger TRIGGER READY

AF data collection – pulse stop STOP

Manual Input? Yes

No

AF data collection – manual entry NEXT ABORT REPEAT

Manual input entry DONE BACK

LEFT KEY

A

AF data collection – save test point SAVE ABORT REPEAT LEFT KEY

A

A

Last data point?

Yes

No

Save As Found test data and time stamp

As Found Test

Pressure Zero Menu ZERO DONE

Either display in pressure mode? Yes

No

A

B

Switch? Yes

No

AF data collection – switch start START ABORT REPEAT

Switch settle and ramp STOP On switch

trip or ramp time out

Page 43: User Manual - Transcat · US, Canada, Latin America AMETEK TCI at 1-800-527-9999 Europe, Africa, Middle East AMETEK Denmark A/S at + 45 4816 8000 Asia AMETEK Singapore Pte. Ltd. at

41

Validating results?

Yes

No

Display test details? YES NO

View test results DONE

B

Tag ID field selection SELECT NEXT SAVE

Tag field data entry DONE NEXT SAVE

Save Tag ID data changes

C

Save AF results as AL? YES NO

Adjust Instrument? YES NO

Document Menu

D

Page 44: User Manual - Transcat · US, Canada, Latin America AMETEK TCI at 1-800-527-9999 Europe, Africa, Middle East AMETEK Denmark A/S at + 45 4816 8000 Asia AMETEK Singapore Pte. Ltd. at

42

C

Pressure Zero Menu ZERO DONE

Either display in pressure mode? Yes

Pulse mode? Yes

No

Adjust – pulse trigger TRIGGER READY

Adjust – pulse stop STOP

No

Adjust – monitor/step/enter UP DOWN AS LEFT

D

Page 45: User Manual - Transcat · US, Canada, Latin America AMETEK TCI at 1-800-527-9999 Europe, Africa, Middle East AMETEK Denmark A/S at + 45 4816 8000 Asia AMETEK Singapore Pte. Ltd. at

43

As Left Menu

As Left tag selection SELECT INFO EXIT

Document Menu

Already has AL results? Yes

No

Overwrite AL results? YES NO

E

Pressure Zero Menu ZERO DONE

Either display in pressure mode? Yes

No

D

Information Selection SELECT

EXIT

Information Display

EXIT

Page 46: User Manual - Transcat · US, Canada, Latin America AMETEK TCI at 1-800-527-9999 Europe, Africa, Middle East AMETEK Denmark A/S at + 45 4816 8000 Asia AMETEK Singapore Pte. Ltd. at

44

Pulse mode? Yes

No

AL data collection – pulse trigger TRIGGER READY

AL data collection – pulse stop STOP

Manual Input? Yes

No

AL data collection – manual entry NEXT ABORT REPEAT

Manual input entry DONE BACK

LEFT KEY

E

AL data collection – save test point SAVE ABORT REPEAT LEFT KEY

E

E

Last data point?

Yes

No

Save As Left data and time stamp

F

Switch? Yes

No

AL data collection – switch start START ABORT REPEAT

Switch settle and ramp STOP On switch

trip or ramp time out

Page 47: User Manual - Transcat · US, Canada, Latin America AMETEK TCI at 1-800-527-9999 Europe, Africa, Middle East AMETEK Denmark A/S at + 45 4816 8000 Asia AMETEK Singapore Pte. Ltd. at

45

Validating results?

Yes

No

Display test details? YES NO

View test results DONE

AL complete, adjustment required? YES NO

Document Menu C

F

Page 48: User Manual - Transcat · US, Canada, Latin America AMETEK TCI at 1-800-527-9999 Europe, Africa, Middle East AMETEK Denmark A/S at + 45 4816 8000 Asia AMETEK Singapore Pte. Ltd. at

46

Review Menu

Review Function Select VIEW ERASE EXIT

View Tag Select SELECT INFO EXIT

View Tag DONE EXIT

Review Menu

Document Menu

Review Menu

Information Selection SELECT

EXIT

Information Display

EXIT

G

Page 49: User Manual - Transcat · US, Canada, Latin America AMETEK TCI at 1-800-527-9999 Europe, Africa, Middle East AMETEK Denmark A/S at + 45 4816 8000 Asia AMETEK Singapore Pte. Ltd. at

47

Review Menu

G

Erase Tag Select SELECT ERASE ALL EXIT

Confirm Tag Erase YES NO

Confirm Erase All Tags YES NO

G

G

Final Confirm Erase All Tags YES NO

G

Page 50: User Manual - Transcat · US, Canada, Latin America AMETEK TCI at 1-800-527-9999 Europe, Africa, Middle East AMETEK Denmark A/S at + 45 4816 8000 Asia AMETEK Singapore Pte. Ltd. at

48

8.2 As Found Test

8.2.1 Work Order Selection

At the main menu, press the MENU function key, followed by the MORE function key, followed by the DOCUMENT function key to display the first level document mode menu.

TEST REVIEW EXIT

Press the TEST function key to display the test selection menu.

AS FOUND AS LEFT EXIT

Press the AS FOUND function key to display the As Found work order Selection menu which shows all uncalibrated work orders.

U TIC-221-5U LI-22

SELECT INFO EXIT

Work order numbers are displayed on multiple pages, with up to six work orders to a page. The letter U indicates an uncalibrated work order. Press the UP ARROW and DOWN ARROW keys to scroll the work order selector highlight one work order at a time, scrolling from page to page, cycling from top to bottom of the list and vice versa. Use the LEFT ARROW and RIGHT ARROW keys to scroll the work order selector highlight one page at a time, cycling from top to bottom of the list and vice versa.

When the desired work order number is highlighted, press the SELECT function key to proceed to data collection.

Press the INFO function key to display the extended information for the highlighted work order as described in section 8.7.

Press the EXIT function key to return to the first level document mode menu without selecting a work order.

8.2.2 Data Collection, Analog Instruments

The test for each point consists of setting the calibrator output to the predefined value, waiting for the input to settle, and saving the data. This

Page 51: User Manual - Transcat · US, Canada, Latin America AMETEK TCI at 1-800-527-9999 Europe, Africa, Middle East AMETEK Denmark A/S at + 45 4816 8000 Asia AMETEK Singapore Pte. Ltd. at

49

process is repeated until all test points are completed. For ease of analysis, it is suggested that the test points be evenly spaced across the test range and that they be entered in ascending order, descending order, ascending followed by descending, or descending followed by ascending.

If the test includes a pressure input or output, the pressure module is validated and zeroed first as described in section 8.2.2.5. After the module has been zeroed, proceed as described below.

The prompt sequence displayed during the test depends on the input and output types selected. Any combination which includes a pulse train calibrator output or a manual calibrator input has its own unique prompt sequence. All other combinations are covered by a generic test prompt sequence.

8.2.2.1 Generic Test Prompt Sequence

A prompt sequence similar to the following is displayed for each test point when the input/output combination does not include a pulse train calibrator output or a manual calibrator input.

mA IN

4.000 mAAF 1

VOLTS OUT

SAVE1.000 V

ABORT REPEAT

The current test type, AF for As Found, and point number are shown in the upper right corner of the display. Note that the point number is the next point to be tested and data has not been saved for it yet.

If the calibrator output is pressure, the prompt SAVE is replaced by the prompt NEXT. Press the NEXT function key to display both the pressure value to be set externally and the present pressure reading. Once the pressure has stabilized, press the DONE function key on that display to return to the above display.

If the calibrator output is manual, set the external source to the displayed predefined value. Once the external source has stabilized continue as described below.

In all cases, once the displayed calibrator input value has settled, press the SAVE function key to save the data and step to the next point. The point number in the upper right increases to show that the test data is saved.

Once the SAVE function key is pressed for the last test point, the process

Page 52: User Manual - Transcat · US, Canada, Latin America AMETEK TCI at 1-800-527-9999 Europe, Africa, Middle East AMETEK Denmark A/S at + 45 4816 8000 Asia AMETEK Singapore Pte. Ltd. at

5050

proceeds to the identification data entry step described in section 8.2.4.

If for any reason the data collection is to be terminated without saving any data, press either the HOME key or the ABORT function key. A prompt is displayed requesting confirmation that the test data accumulated so far is to be abandoned. Press the NO function key to return to the data collection process at the point where it was interrupted. Press the YES function key to abandon the data collected.

Press the REPEAT function key to restart the test sequence at the first test point.

8.2.2.2 Pulse Train Test Prompt Sequence

When the input/output combination includes a pulse train calibrator output, a prompt similar to the following is displayed for each test point.

mA IN

4.000 mAAF 1

PULSE OUT

TRIG10.0 Hz

READY

5.0 Vpp

The current test type, AF for As Found, and point number are shown in the upper right corner of the display. Note that the point number is the next point to be tested and data has not been saved for it yet.

Press the TRIG function key to generate the pulse train. The lower line changes to the following while the pulse train is being generated.

STOP

Press the STOP function key to abort the pulse train before it is finished. When the pulse train is complete, or is aborted, the lower line changes back to its original state.

Once the calibrator input value has settled, and the results are acceptable, press the READY function key to proceed to the prompts for saving the data.

If the calibrator input is manual, the lower line changes to the following prompts.

NEXT ABORT REPEAT

Page 53: User Manual - Transcat · US, Canada, Latin America AMETEK TCI at 1-800-527-9999 Europe, Africa, Middle East AMETEK Denmark A/S at + 45 4816 8000 Asia AMETEK Singapore Pte. Ltd. at

5151

Press the NEXT function key to enter the manual input value as described in the section 8.2.2.4. See the following paragraphs for a description of the ABORT and REPEAT function keys.

If the calibrator input is not manual, or after the manual data has been entered, the lower line changes to the following prompts.

SAVE ABORT REPEAT

If the data displayed is correct, press the SAVE function key to save the data and step to the next point. The point number in the upper right increases to show that the test data is saved and the prompts change back to TRIG/READY. If the pulse train or manual data must be corrected before saving it, press the LEFT ARROW key to change back to the TRIG/READY prompts and repeat the test.

Once the SAVE function key is pressed for the last test point, the process proceeds to the identification data entry step described in section 8.2.4.

If for any reason the data collection is to be terminated without saving any data, press either the HOME key or the ABORT function key. A prompt is displayed requesting confirmation that the test data accumulated so far is to be abandoned. Press the NO function key to return to the data collection process at the point where it was interrupted. Press the YES function key to abandon the data collected.

Press the REPEAT function key to restart the test sequence at the first test point.

8.2.2.3 Manual Calibrator Input Test Prompt Sequence

When the calibrator input is manual, and does not include a pulse train calibrator output, a prompt similar to the following is displayed for each test point. See section 8.2.2.2 if the calibrator output is pulse train.

MANUAL IN

----- lbsAF 1

VOLTS OUT

NEXT1.000 V

REPEATABORT

The current test type, AF for As Found, and point number are shown in the upper right corner of the display. Note that the point number is the next point to be tested and data has not been saved for it yet.

Page 54: User Manual - Transcat · US, Canada, Latin America AMETEK TCI at 1-800-527-9999 Europe, Africa, Middle East AMETEK Denmark A/S at + 45 4816 8000 Asia AMETEK Singapore Pte. Ltd. at

5252

If the calibrator output is pressure, press the NEXT function key to display both the pressure value to be set externally and the present pressure reading. Once the pressure has stabilized, press the DONE function key on that display to prompt for a manual calibrator input value.

If the calibrator output is manual, set the external source. Once the external source has stabilized, continue as described below.

In all cases, wait for the external measuring device reading to settle and then press the NEXT function key to enter the manual input value as described in section 8.2.2.4. Upon return from the manual entry prompt, the NEXT function key label changes to SAVE.

If the data displayed is correct, press the SAVE function key to save the data and step to the next point. The point number in the upper right increases to show that the test data is saved. If the manual data must be corrected before saving it, press the LEFT ARROW key to change the SAVE function key back to NEXT, and press it to repeat the manual entry(s).

Once the SAVE function key is pressed for the last test point, the process proceeds to the identification data entry step described in section 8.2.4.

If for any reason the data collection is to be terminated without saving any data, press either the HOME key or the ABORT function key. A prompt is displayed requesting confirmation that the test data accumulated so far is to be abandoned. Press the NO function key to return to the data collection process at the point where it was interrupted. Press the YES function key to abandon the data collected.

Press the REPEAT function key to restart the test sequence at the first test point.

8.2.2.4 Manual Input Data Entry Prompt

A prompt similar to the following is displayed for entry of manual inputs.

AF 1

Enter manual input for Upper Display

DONE- - - - - TONS

BACK

Enter the manual value using the numeric keys, followed by ENTER. If the value is correct, press the DONE function key to continue. The BACK function key may be pressed to return to the previous prompt display.

Page 55: User Manual - Transcat · US, Canada, Latin America AMETEK TCI at 1-800-527-9999 Europe, Africa, Middle East AMETEK Denmark A/S at + 45 4816 8000 Asia AMETEK Singapore Pte. Ltd. at

5353

8.2.2.5 Pressure Module Validation and Zero

If no module is connected, or the connected module does not have the predefined model description, a prompt is displayed indicating the model description of the required pressure module.

----- psi

Connect module: MODELx

ABORT

Press the ABORT function key to cancel the test or adjustment. Otherwise insert a pressure module with the indicated model type.

The next display depends on whether the pressure module is for absolute or gauge pressure.

Gauge Pressure Module

Once a correct gauge pressure module has been connected, the following display appears showing the present pressure reading.

-0.008 psi

Zero pressure reading

DONEZERO

Vent or set the pressure module input to zero and press the ZERO function key to zero the pressure reading.

Press the DONE function key to continue with the test or adjustment.

Absolute Pressure Module

Once a correct absolute pressure module has been connected, the following display appears showing the present pressure reading on the upper line and the present barometric pressure setting on the lower line.

Page 56: User Manual - Transcat · US, Canada, Latin America AMETEK TCI at 1-800-527-9999 Europe, Africa, Middle East AMETEK Denmark A/S at + 45 4816 8000 Asia AMETEK Singapore Pte. Ltd. at

54

14.017 psi

Enter referencepressure

DONE

14.017 psi

Enter a new barometric pressure if required.

Press the DONE function key to continue with the test or adjustment.

8.2.3 Data Collection, Switch Instruments

The full switch test consists of up to five independent ramp tests. Each test has four parameters, a start value, a start settle time, an end value, and a start to end ramp time. For an ascending ramp the end value is greater than the start value. For a descending ramp the end value is less than the start value. Ascending and descending ramps may be mixed in any order.

A single ramp test consists of setting the calibrator output to the start value, waiting for the start settle time to elapse, and ramping the calibrator output to the end value while monitoring the switch for a state change. The test ends when the switch changes state or the end value is reached. If the switch changes state, the calibrator output at that moment is recorded as the trip point. If the switch never changes state, the trip point is recorded as NONE.

The start value, end value, and start to end ramp time determine the size of each output step during the ramp. There are two output steps per second. This gives the switch time to react to each step change in the output. To reduce the output step size, either increase the start to end ramp time or reduce the difference between the start value and the end value.

If the calibrator output is pressure, setting the start value and ramping the pressure must be done manually with an external device. Before the first ramp test, the pressure module is validated and zeroed as described in section 8.2.2.5.

8.2.3.1 Start Step

A prompt similar to the following display is the first step in a ramp test.

Page 57: User Manual - Transcat · US, Canada, Latin America AMETEK TCI at 1-800-527-9999 Europe, Africa, Middle East AMETEK Denmark A/S at + 45 4816 8000 Asia AMETEK Singapore Pte. Ltd. at

55

SWITCH STRT OP AF 1

VOLTS OUT

SAVE1.000 V

ABORT REPEAT

F1 to start test

The text STRT on the top line indicates that the procedure is at the test start step. The next item indicates the present state of the switch contact, OP for open or CL for closed.

Press the START function key to begin the settling step described in section 8.2.3.2.

If for any reason the data collection is to be terminated without saving any data, press either the HOME key or the ABORT function key. A prompt is displayed requesting confirmation that the test data accumulated so far is to be abandoned. Press the NO function key to return to the data collection process at the point where it was interrupted. Press the YES function key to abandon the data collected.

Press the REPEAT function key to restart the test sequence at the first ramp test.

8.2.3.2 Settling Step

The display content and operation depend on whether the calibrator output is pressure or not.

Other than Pressure

The display changes to the following when the calibrator output is not pressure.

SWITCH SETL OP AF 1

VOLTS OUT

STOP1.000 V

10 sec. remain

Page 58: User Manual - Transcat · US, Canada, Latin America AMETEK TCI at 1-800-527-9999 Europe, Africa, Middle East AMETEK Denmark A/S at + 45 4816 8000 Asia AMETEK Singapore Pte. Ltd. at

56

The text SETL on the top line indicates that the procedure is at the settling step. The next item indicates the present state of the switch contact, OP for open or CL for closed.

The calibrator output is set to the start value and the second line indicates the time remaining in the settling period. When the time remaining reaches zero, the procedure automatically proceeds to the ramp step described in section 8.2.3.3.

Press the STOP function key to restart this ramp test at the start step described in section 8.2.3.1.

Pressure

The display changes to the following when the calibrator output is pressure.

SWITCH SETL OP AF 1

PRESSURE OUT

STOP1.000

Set 1.000 psi8 of 30 sec.

NEXT

psi

The text SETL on the top line indicates that the procedure is at the settling step. The next item indicates the present state of the switch contact, OP for open or CL for closed.

The second line indicates the pressure start value to be set externally. The third line indicates the time remaining in the suggested settling time. The settling time display is for reference only and a zero value does not automatically end the settling step.

Once the pressure measurement has settled, press the NEXT function key to proceed to the ramp step described in section 8.2.3.3. This key may be used before the suggested settling time has elapsed, but that is not a recommended procedure.

Press the STOP function key to restart this ramp test at the start step described in section 8.2.3.1.

8.2.3.3 Ramp Step

The display content and operation depend on whether the calibrator output is pressure or not.

Other than Pressure

The display changes to the following when the calibrator output is not pressure.

Page 59: User Manual - Transcat · US, Canada, Latin America AMETEK TCI at 1-800-527-9999 Europe, Africa, Middle East AMETEK Denmark A/S at + 45 4816 8000 Asia AMETEK Singapore Pte. Ltd. at

57

SWITCH RAMP OP AF 1

VOLTS OUT

STOP1.000 V

10 sec. remain

The text RAMP on the top line indicates that the procedure is at the ramp step. The next item indicates the present state of the switch contact, OP for open or CL for closed.

The calibrator output ramps to the end value. The second line indicates the time remaining in the ramp period. When switch changes state or the time remaining reaches zero, the procedure automatically proceeds to the save step described in section 8.2.3.4.

Press the STOP function key to restart this ramp test at the start step described in section 8.2.3.1.

Pressure

The display changes to the following when the calibrator output is pressure.

SWITCH RAMP CL AF 1

PRESSURE OUT

STOP1.000

Ramp to 5.000 psi8 of 30 sec.

SAVE

psi

The text RAMP on the top line indicates that the procedure is at the ramp step. The next item indicates the present state of the switch contact, OP for open or CL for closed.

The second line indicates the pressure end value to be ramped externally. The third line indicates the time remaining in the suggested ramp period. The ramp time display is for reference only and a zero value does not automatically end the ramp step.

When switch changes state, the procedure automatically proceeds to the save step described in section 8.2.3.4.

If the switch does NOT trip and this condition is to be recorded, press the

Page 60: User Manual - Transcat · US, Canada, Latin America AMETEK TCI at 1-800-527-9999 Europe, Africa, Middle East AMETEK Denmark A/S at + 45 4816 8000 Asia AMETEK Singapore Pte. Ltd. at

58

SAVE function key to proceed to the save step described in section 8.2.3.4.

Press the STOP function key to restart this ramp test at the start step described in section 8.2.3.1.

8.2.3.4 Save Step

The display changes to the following for saving the test result.

SWITCH TRIP CL

3.472 VAF 1

VOLTS OUT

SAVE3.472 V

ABORT REPEAT

The text TRIP on the top line indicates that the procedure is at the save results step. The next item indicates the changed state of the switch contact, OP for open or CL for closed. If the switch does not change state during the ramp time these two items are replaced with the text NO TRIP.

Press the SAVE function key to save the test results and proceed to the start of the next ramp test as described in section 8.2.3.1.

Press the LEFT ARROW key to restart this ramp test at the start step described in section 8.2.3.1.

Once the SAVE function key is pressed for the last ramp test, the process proceeds to the identification data entry step described in section 8.2.4.

If for any reason the data collection is to be terminated without saving any data, press either the HOME key or the ABORT function key. A prompt is displayed requesting confirmation that the test data accumulated so far is to be abandoned. Press the NO function key to return to the data collection process at the point where it was interrupted. Press the YES function key to abandon the data collected.

Press the REPEAT function key to restart the test sequence at the first ramp test.

8.2.4 Identification Data Entry

When the last test point is completed at the data collection step, the first page of the following two page identification data display is shown. The present value of each field is shown to the right of its description.

NOTE: The identification data can only be changed during an As Found test.

Page 61: User Manual - Transcat · US, Canada, Latin America AMETEK TCI at 1-800-527-9999 Europe, Africa, Middle East AMETEK Denmark A/S at + 45 4816 8000 Asia AMETEK Singapore Pte. Ltd. at

59

WO# FI-2451MFG companyMDL# model

SELECT NEXT SAVE

SER# 6789245TECH George K.TEMP 23CHUMD 38%OTHR other

SELECT NEXT SAVE

These two pages provide space to display eight phrases of up to 16 or 32 characters each. The phrases labeled WO#, MFG, and SER# can be up to 32 characters, while the rest can be up to 16 characters long. Initially they provide predefined instructions to the technician for the test. At the end of the As Found test all but the WO# phrase may be changed by the technician to describe the test that has been performed. The first seven phrases have fixed usage: Work Order Number, Instrument Manufacturer, Instrument Model Number, Instrument Serial Number, Technician Name or ID number, Temperature, and Humidity. The last phase, Other is for client use.

Press the NEXT function key to switch between the two pages.

Press the SAVE function key to save the data and proceed to the test results display step described in section 8.2.5.

Press the UP ARROW and DOWN ARROW keys to move the field selector highlight from one field to the next.

Press the SELECT function key to enter data into the highlighted field. A prompt similar to the following is displayed showing the field name and current value on the top line.

Page 62: User Manual - Transcat · US, Canada, Latin America AMETEK TCI at 1-800-527-9999 Europe, Africa, Middle East AMETEK Denmark A/S at + 45 4816 8000 Asia AMETEK Singapore Pte. Ltd. at

60

TAG# FT-384 7ABC 8DEF 9GHI 4JKL 5MNO 6PQR 1STU 2VWX 3YZ -+ 0#%_ .:,

DONE NEXT SAVE

Characters are entered into the field using a telephone style entry method. The four lower lines provide a mapping of the allowed alphanumeric characters to the numeric keypad keys. To enter a character press the corresponding numeric key multiple times to cycle through the mapped characters until the desired character is displayed. For example, to enter E, press the 8 key 3 times. If the next character is on a different numeric key, simply press that key to advance the cursor and display the first mapped character. If the next character is on the same key, press ENTER to advance the cursor before pressing the same key. To insert a space, press ENTER twice. To erase characters press CE once for each character. To switch a character from lower to upper case, or vice versa, press the UP ARROW or DOWN ARROW.

Press the DONE function key to return to the two page identification data display.

Press the NEXT function key to display and edit the next data field in sequence.

Press the SAVE function key to save the data and proceed to the test results display step described in section 8.2.5.

8.2.5 Test Results Display

After the identification data has been saved, a prompt is displayed asking whether test results are to be displayed or not. If either pass/fail tolerance has been specified, the prompt also shows the overall pass/fail status of the entire test, and each test point on the detail display is assigned a pass or fail status.

Page 63: User Manual - Transcat · US, Canada, Latin America AMETEK TCI at 1-800-527-9999 Europe, Africa, Middle East AMETEK Denmark A/S at + 45 4816 8000 Asia AMETEK Singapore Pte. Ltd. at

61

PASSED

YES NO

Display test details?

The overall test status is displayed on the top line. If neither pass/fail tolerance has been specified, the status display is NONE. Otherwise, if all points passed, the status display is PASSED, or, if one or more points have failed, the status display is FAILED.

Press the NO function key to bypass the detail results display and proceed to the test conclusion step described in section 8.2.6.

Press the YES function key to display the test details. The content and operation of this display is identical to that described in section 8.5 on viewing test results. When viewing is complete, press the DONE function key to proceed to the test conclusion step described in section 8.2.6.

8.2.6 Test Conclusion

A prompt is displayed asking whether the testing for this work order is complete and the As Found test results are to be saved for the As Left test results.

If the test has been successful and no further testing is required, press the YES function key to return to the first level document mode menu.

If further testing is required, press the NO function key and a prompt is displayed asking whether an adjustment step is required before proceeding to the first As Left test.

If an adjustment is required, press the YES function key to proceed to the adjustment step described in section 8.3.

If no adjustment is required, press the NO function key to proceed to the As Left data collection step described in section 8.4.

Page 64: User Manual - Transcat · US, Canada, Latin America AMETEK TCI at 1-800-527-9999 Europe, Africa, Middle East AMETEK Denmark A/S at + 45 4816 8000 Asia AMETEK Singapore Pte. Ltd. at

62

8.3 AdjustmentDuring the adjustment process, the predefined output values can be generated by stepping up or down through the test points. If the output type is sourced by the calibrator, the user can enter a value that is different from the predefined test points.

If the test includes a pressure input or output, the pressure module is validated and zeroed first as described in section 8.2.2.5. After the module has been zeroed, proceed as described below.

The prompt displayed during the adjustment depends on the input and output types selected. A switch instrument test, or any other combination which includes pulse train output has its own unique prompt sequence. All other combinations are covered by a generic prompt.

8.3.1 Generic Adjustment Prompt

A prompt similar to the following is displayed for each test point of an analog instrument when the calibrator output is not pulse train.

mA IN

4.000 mAPT 1

VOLTS OUT

UP1.000 V

DOWN AS LEFT

The PT in the upper right corner of the display indicates that this is the adjustment display. If a predefined test point is being generated, its number is displayed. If a user entered value is being generated, two dashes are displayed.

Press the UP or DOWN function key to cycle to the next predefined test point. If the output type is not pressure or manual, the calibrator generates it. If the output type is pressure or manual, the required output value is displayed so that it can be set on the external source.

If the calibrator output type is not pressure or manual, a value can be entered using the numeric keys, followed by ENTER. In addition, the output value can be ramped with the UP ARROW and DOWN ARROW keys. Press the LEFT ARROW or RIGHT ARROW key to change the least significant digit being ramped.

If the calibrator output type is pressure, the present measurement of the external pressure source is displayed. It can not be changed via the calibrator. All source value changes are done using the external source.

Page 65: User Manual - Transcat · US, Canada, Latin America AMETEK TCI at 1-800-527-9999 Europe, Africa, Middle East AMETEK Denmark A/S at + 45 4816 8000 Asia AMETEK Singapore Pte. Ltd. at

63

If the calibrator output type is manual, the required output value for the test point is displayed. It can not be changed via the calibrator. All source value changes are done using the external source.

Press the AS LEFT function key to proceed to the As Left data collection step described in section 8.4.

8.3.2 Pulse Train Adjustment

A prompt similar to the following is displayed for each test point of an analog instrument when the calibrator output is pulse train.

VOLTS IN

- - - - - VPT 1

PULSE OUT

TRIG60.0 Hz

READY

1.0 VpP

The PT in the upper right corner of the display indicates that this is the adjustment display. If a predefined test point is selected, its number is displayed. If a user entered value is selected, two dashes are displayed.

Use the numeric keys, followed by ENTER, to enter a user value for generation, or press the READY function key to proceed to the next menu where the UP and DOWN function keys can be used to cycle through the predefined test points.

Press the TRIG function key to generate the pulse train. The lower line changes to the following while the pulse train is being generated.

STOP

Press the STOP function key to abort the pulse train before it is finished. When the pulse train is complete, or is aborted, the lower line changes back to its original state.

Press the READY function key to change the lower line to the prompts for cycling through the predefined test points, or proceeding to As Left data collection.

UP DOWN AS LEFT

Press the UP or DOWN function key to cycle to the next predefined test point. The prompt changes back to TRIG/READY.

Page 66: User Manual - Transcat · US, Canada, Latin America AMETEK TCI at 1-800-527-9999 Europe, Africa, Middle East AMETEK Denmark A/S at + 45 4816 8000 Asia AMETEK Singapore Pte. Ltd. at

64

Press the AS LEFT function key to proceed to the As Left data collection step described in section 8.4.

Press the LEFT ARROW key to return to the trigger menu and generate a new pulse train. The prompt changes back to TRIG/READY.

8.3.3 Switch Instrument Adjustment

When the instrument is a switch, a prompt similar to the following is displayed.

SWITCH ADJ CL PT --

VOLTS OUT

1.000 VAS LEFT

The text ADJ on the top line indicates that this is an adjustment procedure. The next item indicates the present state of the switch contact, OP for open or CL for closed.

If the calibrator output type is not pressure, a value can be entered using the numeric keys, followed by ENTER. In addition, the output value can be ramped with the UP ARROW and DOWN ARROW keys. Press the LEFT ARROW or RIGHT ARROW key to change the least significant digit being ramped.

If the calibrator output type is pressure, the present measurement of the external pressure source is displayed. It can not be changed via the calibrator. All source value changes are done using the external source.

8.4 As Left TestAn As Left test can immediately follow an As Found test or a previous As Left test, or it can be initiated at a later time for any work order stored in memory. If it immediately follows another test, no work order selection step is required and data collection starts right away. If it is initiated at a later time, the work order must be selected first.

8.4.1 Work order Selection

At the main menu, press the MENU function key, followed by the MORE function key, followed by the DOCUMENT function key to display the first level document mode menu.

Page 67: User Manual - Transcat · US, Canada, Latin America AMETEK TCI at 1-800-527-9999 Europe, Africa, Middle East AMETEK Denmark A/S at + 45 4816 8000 Asia AMETEK Singapore Pte. Ltd. at

65

TEST REVIEW EXIT

Press the TEST function key to display the test selection menu.

AS FOUND AS LEFT EXIT

Press the AS LEFT function key to display the work order selection prompt which shows all previously calibrated work orders available for recalibration.

C TIC-221-5C LI-22

SELECT INFO EXIT

Work order numbers are displayed on multiple pages, with up to six work orders to a page. The letter C indicates a calibrated work order. Press the UP ARROW and DOWN ARROW keys to scroll the work order selector highlight one work order at a time, scrolling from page to page, cycling from top to bottom of the list and vice versa. Use the LEFT ARROW and RIGHT ARROW keys to scroll the work order selector highlight one page at a time, cycling from top to bottom of the list and vice versa.

When the desired work order number is highlighted, press the SELECT function key to proceed to data collection.

Press the INFO function key to display the extended information for the highlighted work order as described in section 8.7.

Press the EXIT function key to return to the first level document mode menu without selecting a work order.

8.4.2 Data Collection

If the test includes a pressure input or output, the pressure module is validated and zeroed first as described in section 8.2.2.5. After the module has been zeroed, proceed as described below.

The As Left data collection is identical to the As Found data collection except for these minor differences.

1) The AF in the upper right corner of the display is changed to AL to indicate that this is an As Left test.

Page 68: User Manual - Transcat · US, Canada, Latin America AMETEK TCI at 1-800-527-9999 Europe, Africa, Middle East AMETEK Denmark A/S at + 45 4816 8000 Asia AMETEK Singapore Pte. Ltd. at

66

2) Once the SAVE function key is pressed for the last test point, the process automatically proceeds to the test results step described in section 8.4.3. The test identification data can not be changed during an As Left test.

8.4.3 Test Results

A display showing the overall pass/fail status of the test is displayed with a prompt asking whether details should be displayed.

Press the NO function key to skip the details and proceed to the test conclusion step described in section 8.4.4.

Press the YES function key to display the details. The content and operation of this display is identical to that described in section 8.5 on viewing test results. When viewing is complete, press the DONE function key to proceed to the test conclusion step described in section 8.4.4.

8.4.4 Test Conclusion

A prompt is displayed asking whether further adjustment and testing is required.

Press the YES function key to proceed to the adjustment procedure described in section 8.3.

Press the NO function key to proceed to the first level document mode menu.

8.5 Viewing Test ResultsAt the main menu, press the MENU function key, followed by the MORE function key, followed by the DOCUMENT function key to display the first level document mode menu.

TEST REVIEW EXIT

Press the REVIEW function key to display the review selection menu.

VIEW ERASE EXIT

Press the VIEW function key to display the work order selection prompt.

Page 69: User Manual - Transcat · US, Canada, Latin America AMETEK TCI at 1-800-527-9999 Europe, Africa, Middle East AMETEK Denmark A/S at + 45 4816 8000 Asia AMETEK Singapore Pte. Ltd. at

67

C FT-384U TIC-221-5U LI-22C PI-884

SELECT INFO EXIT

The letter U indicates an uncalibrated work order, and a C indicates a calibrated work order. Work order numbers are displayed on multiple pages, with up to six work orders to a page. Press the UP ARROW and DOWN ARROW keys to scroll the work order selector highlight one work order at a time, scrolling from page to page, cycling from top to bottom of the list and vice versa. Use the LEFT ARROW and RIGHT ARROW keys to scroll the work order selector highlight one page at a time, cycling from top to bottom of the list and vice versa.

Press the EXIT function key to return to the review selection menu without selecting a work order.

Press the INFO function key to display the extended information for the highlighted work order as described in section 8.7.

When the desired work order number is highlighted, press the SELECT function key to display the first page of data showing the work order number, calibration date and time, and the first few identification fields.

Press the UP ARROW and DOWN ARROW keys to scroll through the remaining pages of data, cycling from the last page to the first page and vice versa. The second page contains the remaining identification fields. The third page contains the instrument input setup. The fourth page contains the instrument output setup. The fifth page contains the pass/fail tolerances and the input and output ranges. The sixth and subsequent pages display the test results. On these pages only, use the LEFT ARROW and RIGHT ARROW keys to switch between the As Found results and the As Left results.

Note that the input/output orientation of the data on these pages is with respect to the instrument under test. This is in contrast to the orientation during testing, which was with respect to the calibrator.

Press the DONE function key to return to the work order selection menu.

Press the EXIT function key to return to the review selection menu.

Page 70: User Manual - Transcat · US, Canada, Latin America AMETEK TCI at 1-800-527-9999 Europe, Africa, Middle East AMETEK Denmark A/S at + 45 4816 8000 Asia AMETEK Singapore Pte. Ltd. at

68

8.6 Erasing Work OrdersAt the main menu, press the MENU function key, followed by the MORE function key, followed by the DOCUMENT function key to display the first level document mode menu.

TEST REVIEW EXIT

Press the REVIEW function key to display the review selection menu.

VIEW ERASE EXIT

Press the ERASE function key to display the work order selection prompt and erase menu options.

SELECT ERASE ALL EXIT

To erase all work orders from the calibrator memory, press the ERASE ALL function key. Additional prompts are displayed to confirm the request. Press the YES function key in response to every prompt to proceed with erasing all work orders. To abort the erase request, press the NO function key in response to any prompt.

To erase a single work order from the calibrator memory, highlight the desired work order as described under viewing test results in section 8.5. Press the SELECT function key to erase the work order. An additional prompt is displayed to confirm the request. Press the YES function key to erase the work order, or the NO function key to abort the request. To erase additional work orders, highlight each in turn and press the SELECT function key. When all desired work orders have been erased, press the EXIT function key to return to the review selection menu.

8.7 Extended Information DisplayThe information category list is displayed as follows.

Instrument InputCompanySensorsExternal SourceASC321Reference Sensor

SELECT EXIT

Page 71: User Manual - Transcat · US, Canada, Latin America AMETEK TCI at 1-800-527-9999 Europe, Africa, Middle East AMETEK Denmark A/S at + 45 4816 8000 Asia AMETEK Singapore Pte. Ltd. at

69

Press the UP ARROW and DOWN ARROW keys to scroll the category list selector highlight one item at a time, scrolling from the bottom to the top of the list and vice versa.

Press the EXIT function key to return to the menu from which the extended information display was called.

When the desired category is highlighted, press the SELECT function key to proceed to a detail item display similar to the following.

Identity: ASC321 ident

Serial Number: ASC321 sn

EXIT

Items are displayed on multiple pages, with two items per page. Press the UP ARROW and DOWN ARROW keys to scroll through the pages, scrolling from the last page to the first page and vice versa.

Press the EXIT function key to return to category list.

8.8 Setting the Date and TimeAt the main menu, press the MENU function key, followed by the MORE function key, followed by the NEXT function multiple times to get to the clock selection menu.

CLOCK NEXT DONE

Press the CLOCK function key to display the clock data entry screen.

2006/03/28 15:21:14

Enter year

BACK2006

SAVENEXT

Initially the date and time values displayed on the top line are the precise clock setting when the screen was first displayed. After that they are static

Page 72: User Manual - Transcat · US, Canada, Latin America AMETEK TCI at 1-800-527-9999 Europe, Africa, Middle East AMETEK Denmark A/S at + 45 4816 8000 Asia AMETEK Singapore Pte. Ltd. at

70

values representing any data changes made. The clock is not updated until the SAVE function key is pressed. For the most accurate setting it is recommended that all values but the seconds be entered first, then enter a seconds value slightly ahead of the current time, and finally press the SAVE function key exactly when the entered seconds is reached.

Press the NEXT and BACK function keys to switch between the individual date and time fields. The upper part of the display changes to reflect the current value for that field. Use the numeric keys followed by ENTER to change the value. The year must be 2006 to 2100, the month 1 to 12, the day 1 to the number of days in the specified month, the hour 0 to 23, the minute 0 to 59, and the second 0 to 59. If the day is made invalid by changing to a month with fewer days, the day value is automatically changed to the maximum number of days valid for the new month.

Press the SAVE function key to set the clock and return to the home menu.

Press the HOME key to exit the data entry screen without changing the clock.

9. SpecificationsOne year at 23°C ± 5°C, unless specified otherwise. Outside of this range the stability of the measurements is ± 0.003% of FS/°C.

Table 9: General Specifications

Operating Temperature -10°C to 50°C

Storage Temperature -20°C to 60°C

Power 4 X AA batteries; Alkaline or optional rechargeable (Charge current 60 to 85 mA, dependent on the cells and the state of charge.) Use only NiMH cells with capability larger than 1700 mA/h (trickle charge C/20)

Battery Life Approximately 20 hours

Serial Communications ASCII; RS232 or USB depending on the communications cable used

CE - EMC EN61326: 2006, ANNEX B Criterion A

Safety EN/IEC 61010-1:2010 3rd Edition (Low Voltage Directive)

Size 9.3" x 3.8" x 2.3" (23.5cm x 9.7cm x 5.7cm)

Weight Approximately 21oz. (0.6kg)

Page 73: User Manual - Transcat · US, Canada, Latin America AMETEK TCI at 1-800-527-9999 Europe, Africa, Middle East AMETEK Denmark A/S at + 45 4816 8000 Asia AMETEK Singapore Pte. Ltd. at

71

Table 10: DC Voltage Measurement/Source

Range Accuracy (% of reading ± floor)

Read: Isolated(Upper Display) 0.000V - 30.000V 0.01% ± 2mV

Read: non-Isolated(Lower Display) 0.000V - 20.000V 0.01% ± 2mV

Source 0.000V - 20.000V 0.01% ± 2mV

Maximum current output in voltage ranges is 3mA with an output impedance of <= 1Ω.

Table 11: DC mA Measurement/Source

Range Accuracy (% of reading ± floor)

Read: Isolated(Upper Display) 0.000mA - 24.000mA 0.01% ± 2µA

Read: non-Isolated(Lower Display) 0.000mA - 24.000mA 0.01% ± 2µA

Source 0.000mA - 24.000mA 0.01% ± 2µA

Maximum load on mA source is 1000Ω. Voltage input range on simulate mode 5V - 30V.

Table 12: Frequency Measurement/Source

Range Accuracy (% of reading ± floor)

Read 2.0CPM - 600.0CPM 0.05% ± 0.1CPM

1.0Hz - 1000.0Hz 0.05% ± 0.1Hz

1.00KHz - 10.00KHz 0.05% ± 0.01KHz

Source 2.0CPM - 600.0CPM 0.05%

1.0Hz - 1000.0Hz 0.05%

1.00KHz - 10.00KHz 0.250%

Input voltage amplitude range on frequency is 1V to 20V zero based square wave only.Output amplitude is adjust-able from 1V to 20V, and is a square wave with 50% duty cycle. For output frequency, a slight negative offset of approximately -0.1V is present to assure zero crossing.

Table 13: Resistance Measurement

Range Accuracy (% of reading ± floor)

Ohms low 0.00Ω - 400.0Ω 0.015% ± 0.03Ω

Ohms high 401.0Ω - 4000.0Ω 0.015% ± 0.3Ω

Page 74: User Manual - Transcat · US, Canada, Latin America AMETEK TCI at 1-800-527-9999 Europe, Africa, Middle East AMETEK Denmark A/S at + 45 4816 8000 Asia AMETEK Singapore Pte. Ltd. at

72

Table 14: Resistance Source

Range Excitation Current Accuracy (% of reading ± floor)

Ohms low 5.0Ω - 400.0Ω 0.1mA - 0.5mA 0.015% ± 0.1Ω

5.0Ω - 400.0Ω 0.5mA - 3mA 0.015% ± 0.03Ω

Ohms high 400Ω - 1500Ω 0.05mA - 0.8mA 0.015% ± 0.3Ω

1500Ω - 4000Ω 0.05mA - 0.4mA 0.015% ± 0.3Ω

Note: Unit is compatible with smart transmitters and PLCs. Frequency response is <= 5ms.

Table 15: Thermocouple Measurement/Source

Range Accuracy

(% of reading ± floor)

Read (mV) -10.000mV - 75.000mV 0.015% ± 10µV

Source (mV) -10.000mV - 75.000mV 0.015% ± 10µV

Maximum current output in voltage ranges is 1mA with an output impedance of <= 1Ω

Page 75: User Manual - Transcat · US, Canada, Latin America AMETEK TCI at 1-800-527-9999 Europe, Africa, Middle East AMETEK Denmark A/S at + 45 4816 8000 Asia AMETEK Singapore Pte. Ltd. at

73

Table 16: Thermocouple Read and Source (errors in °C)

Range (°C) Accuracy (°C)

TC Type Minimum Maximum CJC OFF CJC ONJ -210.0 -150.0 0.4 0.6

-150.0 1200.0 0.2 0.4

K -200.0 -100.0 0.5 0.7

-100.0 600.0 0.2 0.4

600.0 1000.0 0.3 0.5

1000.0 1372.0 0.4 0.6

T -250.0 -200.0 1.5 1.7

-200.0 0.0 0.5 0.7

0.0 400.0 0.2 0.4

E -250.0 -200.0 1.0 1.2

-200.0 -100.0 0.3 0.5

-100.0 1000.0 0.2 0.4

R1 0.0 200.0 1.7 1.9

200.0 1767.0 1.0 1.2

S1 0.0 200.0 1.7 1.9

200.0 1767.0 1.1 1.3

B1 600.0 800.0 1.5 1.7

800.0 1000.0 1.2 1.4

1000.0 1820.0 1.0 1.2

C 0.0 1000.0 0.5 0.7

1000.0 2316.0 1.5 1.7

XK -200.0 800.0 0.2 0.4

BP 0.0 800.0 1.9 2.1

800.0 2500.0 0.6 0.8

L -200.0 900.0 0.2 0.4

U -200.0 0.0 0.4 0.6

0.0 600.0 0.2 0.4

N -200.0 -100.0 0.8 1.0

-100.0 1300.0 0.3 0.5

CJC error outside of 23 ± 5°C is 0.05°C/°C

1. For thermocouple measure mode on B, R and S; round the specification up or down accordingly as there is no resolution past the decimal point.

Page 76: User Manual - Transcat · US, Canada, Latin America AMETEK TCI at 1-800-527-9999 Europe, Africa, Middle East AMETEK Denmark A/S at + 45 4816 8000 Asia AMETEK Singapore Pte. Ltd. at

74

Table 17: RTD Read and Source

RTD Type Range (°C) Accuracy (°C)

Minimum Maximum Measure 4W/Source

Measure3W

P10(90)385 -200 100 0.85 2.20

100 400 1.00 2.50

400 800 1.20 2.80

P50(90)385 -200 100 0.20 0.50

100 400 0.30 0.60

400 800 0.40 0.70

P100(90)385 -200 100 0.15 0.25

100 400 0.20 0.35

400 800 0.30 0.45

P200(90)385 -200 100 0.40 0.50

100 630 0.50 0.60

P400(90)385 -200 100 0.20 0.25

100 630 0.25 0.30

P500(90)385 -200 100 0.20 0.25

100 630 0.30 0.35

P1K(90)385 -200 100 0.15 0.20

100 630 0.20 0.25

P50(90)391 -200 100 0.20 0.50

100 400 0.30 0.60

400 800 0.40 0.70

P100(90)391 -200 100 0.15 0.25

100 400 0.20 0.35

400 800 0.30 0.45

P100(90)392 -200 100 0.10 0.25

100 630 0.20 0.35

M10(90)427 -100 260 0.75 2.00

M50(90)428 -180 200 0.15 0.40

M100(90)428 -180 200 0.10 0.20

H120(90)672 -80 260 0.10 0.15

P100(90)JIS -200 100 0.10 0.25

YSI(90)400 15 50 0.10 0.15

Page 77: User Manual - Transcat · US, Canada, Latin America AMETEK TCI at 1-800-527-9999 Europe, Africa, Middle East AMETEK Denmark A/S at + 45 4816 8000 Asia AMETEK Singapore Pte. Ltd. at

7575

10. Maintenance

10.1 Replacing BatteriesReplace batteries as soon as the battery indicator turns on to avoid false measurements. If the batteries discharge too deeply the calibrator will automatically shut down to avoid battery leakage.

Note: Use only AA size alkaline batteries or optional rechargeable batteries.

CautionDo not use charger with alkaline batteries

10.2 Cleaning the Unit

WarningTo avoid personal injury or damage to the calibrator, use only the specified replacement parts and do not allow water into the case.

CautionTo avoid damaging the plastic lens and case, do not use solvents or abrasive cleansers.

Clean the calibrator with a soft cloth dampened with water or water and mild soap.

10.3 Service Center Calibration or Repair Only qualified service personnel should perform calibration, repairs, or servicing not covered in this manual. If the calibrator fails, check the batteries first, and replace them if needed.

Verify that the calibrator is being operated as explained in this manual. If the calibrator is faulty, send a description of the failure with the calibrator. Be sure to pack the calibrator securely, using the original shipping container if it is available.

Page 78: User Manual - Transcat · US, Canada, Latin America AMETEK TCI at 1-800-527-9999 Europe, Africa, Middle East AMETEK Denmark A/S at + 45 4816 8000 Asia AMETEK Singapore Pte. Ltd. at

76

10.4 Replacement Parts & Accessories

SPK-ASC-007 Operating Manual

124720 Mains adapter 9VDC/200mA - 230VAC/115VAC

SPK-HHC-002 Soft carrying case for the large handheld models including shoulder strap

124004 Shoulder strap

2206011 Wire adapter kit - Type K thermocouple

2206012 Wire adapter kit - Type T thermocouple

104203 Test lead set

123958 RS232 cable 2 m / 6 ft (Stereo Jack to 9 pol D-sub)

121985 Extension Cable for Pt 100 Sensor - 5 m

121983 Extension Cable for Type K - 5 m

122523 Extension Cable for Type N - 5 m

120519 Thermocouple Male Plug - Type Cu-Cu - White

120518 Thermocouple Male Plug - Type R / S - Green

120517 Thermocouple Male Plug - Type K - Yellow

120516 Thermocouple Male Plug - Type J - Black

120515 Thermocouple Male Plug - Type T - Blue

120514 Thermocouple Male Plug - Type N - Orange

Page 79: User Manual - Transcat · US, Canada, Latin America AMETEK TCI at 1-800-527-9999 Europe, Africa, Middle East AMETEK Denmark A/S at + 45 4816 8000 Asia AMETEK Singapore Pte. Ltd. at

77SPK-ASC-007 Rev. C 1/14

Page 80: User Manual - Transcat · US, Canada, Latin America AMETEK TCI at 1-800-527-9999 Europe, Africa, Middle East AMETEK Denmark A/S at + 45 4816 8000 Asia AMETEK Singapore Pte. Ltd. at

Information in this document is subject to change without notice. ©2013, by AMETEK, Inc., www.ametek.com. All rights reserved.

www.ametekcalibration.com

AMETEK Test & Calibration InstrumentsA business unit of AMETEK Measurement &

Calibration Technologies Division offering the following industry leading brands for test and

calibration instrumentation.

JOFRA Calibration InstrumentsTemperature Calibrators

Portable dry-block calibrators, precision thermometers and liquid baths. Temperature

sensors for industrial and marine use.Pressure Calibrators

Convenient electronic systems ranging from -25 mbar to 1000 bar - fully temperature-

compensated for problem-free and accurate field use.

Signal InstrumentsProcess signal measurement and simulation for easy control loop calibration and measurement

tasks.

M&G Dead Weight Testers & PumpsPneumatic floating-ball or hydraulic piston dead

weight testers with accuracies to 0.015% of reading. Pressure generators delivering up to

1,000 bar.

Crystal PressureDigital pressure gauges and calibrators that are

accurate, easy-to-use and reliable. Designed for use in the harshest environments; most products

carry an IS, IP67 and DNV rating.

Lloyd Materials TestingMaterials testing machines and software that

guarantees expert materials testing solutions. Also covering Texture Analysers to perform rapid, general food testing and detailed texture analysis

on a diverse range of foods and cosmetics.

Davenport Polymer Test Equipment Allows measurement and characterization of

moisture-sensitive PET polymers and polymer density.

Chatillon Force MeasurementThe hand held force gauges and motorized

testers have earned their reputation for quality, reliability and accuracy and they represent the de

facto standard for force measurement.

Newage Hardness TestingHardness testers, durometers, optical systems and software for data acquisition and analysis.

United KingdomTel +44 (0)1243 833 [email protected]

FranceTel +33 (0)1 30 68 89 [email protected]

GermanyTel +49 (0)2159 9136 [email protected]

DenmarkTel +45 4816 [email protected]

USAFloridaTel +1 (800) 527 [email protected]

CaliforniaTel +1 (800) 444 [email protected]

IndiaTel +91 22 2836 [email protected]

Singapore Tel +65 6484 [email protected]

ChinaShanghaiTel +86 21 5868 5111

BeijingTel +86 10 8526 2111

GuangzhouTel +86 20 8363 [email protected]