using a biocel system and assaymap technology in phage ... · using a biocel system and assaymap...
TRANSCRIPT
![Page 1: Using a BioCel System and AssayMap Technology in Phage ... · Using a BioCel System and AssayMap Technology in Phage Display and Antibody Screening Jason Graves September 20, 2011](https://reader030.vdocument.in/reader030/viewer/2022041004/5ea898b42c18fd6170081695/html5/thumbnails/1.jpg)
Using a BioCel System and AssayMap Technology in Phage Display and Antibody Screening
Jason Graves
September 20, 2011
AAS User Group Meeting
![Page 2: Using a BioCel System and AssayMap Technology in Phage ... · Using a BioCel System and AssayMap Technology in Phage Display and Antibody Screening Jason Graves September 20, 2011](https://reader030.vdocument.in/reader030/viewer/2022041004/5ea898b42c18fd6170081695/html5/thumbnails/2.jpg)
My Background
• 15 Years of Automation and Related Experience • Parke-Davis/Pfizer
• Used automation to support High Throughput Screening
• Implemented new technologies from companies such as Aurora and Rosys
• Developed and implemented database tools for data interpretation, upload and retrieval to and from corporate databases
• Rosetta (Merck) • Developed and implemented new automation processes to
support lab operations
• Quality control of automation and instrumentation
• Troubleshooting processes and equipment
![Page 3: Using a BioCel System and AssayMap Technology in Phage ... · Using a BioCel System and AssayMap Technology in Phage Display and Antibody Screening Jason Graves September 20, 2011](https://reader030.vdocument.in/reader030/viewer/2022041004/5ea898b42c18fd6170081695/html5/thumbnails/3.jpg)
My Background
• 15 Years of Automation and Related Experience
• Novo Nordisk Inflammation Research Center
• Evaluate processes for automation
• Identify and acquire necessary equipment
• Develop methods and supporting tools
• Maintain equipment
• Used many different platforms and equipment: Agilent (V11), Aurora, Beckman Coulter, BioTek, Hamilton, Matrix, Perkin Elmer, Rosys, Tecan, Thermo Scientific, TomTec, Zymark
![Page 4: Using a BioCel System and AssayMap Technology in Phage ... · Using a BioCel System and AssayMap Technology in Phage Display and Antibody Screening Jason Graves September 20, 2011](https://reader030.vdocument.in/reader030/viewer/2022041004/5ea898b42c18fd6170081695/html5/thumbnails/4.jpg)
Novo Nordisk – First Big Task
• “Integrated System that can do Tissue Culture Work”
• 24-Well Plate Transfections
• Incubator
• Liquid Handler(s) (96 wells and Individual Channels)
• Sterile Environment
• Plate and Tip Capacity
![Page 5: Using a BioCel System and AssayMap Technology in Phage ... · Using a BioCel System and AssayMap Technology in Phage Display and Antibody Screening Jason Graves September 20, 2011](https://reader030.vdocument.in/reader030/viewer/2022041004/5ea898b42c18fd6170081695/html5/thumbnails/5.jpg)
BioCel #1
![Page 6: Using a BioCel System and AssayMap Technology in Phage ... · Using a BioCel System and AssayMap Technology in Phage Display and Antibody Screening Jason Graves September 20, 2011](https://reader030.vdocument.in/reader030/viewer/2022041004/5ea898b42c18fd6170081695/html5/thumbnails/6.jpg)
BioCel #1
![Page 7: Using a BioCel System and AssayMap Technology in Phage ... · Using a BioCel System and AssayMap Technology in Phage Display and Antibody Screening Jason Graves September 20, 2011](https://reader030.vdocument.in/reader030/viewer/2022041004/5ea898b42c18fd6170081695/html5/thumbnails/7.jpg)
BioCel #1
![Page 8: Using a BioCel System and AssayMap Technology in Phage ... · Using a BioCel System and AssayMap Technology in Phage Display and Antibody Screening Jason Graves September 20, 2011](https://reader030.vdocument.in/reader030/viewer/2022041004/5ea898b42c18fd6170081695/html5/thumbnails/8.jpg)
Phage display selections are an iterative process
• A selectable phenotype is linked to a genotype by inserting a displayed protein into the phage genome. In this case a large library of random antibody sequences. (Click any key to continue)
VL VH RBS VH CH GIII
• The library is queried by exposure to the antigen of interest. Antibodies which stick, carry phage that code for them. Others are washed away, enriching the remaining phage for specific binding.
• The genotypes of the antibodies are rescued by infecting E. coli with the phage.
• New Phage are produced and the cycle is repeated
• After 2-3 rounds, the antibodies are produced in soluble form, assayed, and hits are chosen
• An antigen is immobilized to a surface.
![Page 9: Using a BioCel System and AssayMap Technology in Phage ... · Using a BioCel System and AssayMap Technology in Phage Display and Antibody Screening Jason Graves September 20, 2011](https://reader030.vdocument.in/reader030/viewer/2022041004/5ea898b42c18fd6170081695/html5/thumbnails/9.jpg)
First Protocol – 24 Well Transfections
• IgG conversions of antibodies from phage display
• DNA combined with fectin reagent and media on Bravo
• Plate incubates in plate hotel
• Cells (1ml) come out of incubator and go to Hamilton
• DNA is distributed to four 24 well plates on Hamilton
• Cells get put back into incubator
• Process repeats
• Once finished all cells go into a shaking incubator for five days
![Page 10: Using a BioCel System and AssayMap Technology in Phage ... · Using a BioCel System and AssayMap Technology in Phage Display and Antibody Screening Jason Graves September 20, 2011](https://reader030.vdocument.in/reader030/viewer/2022041004/5ea898b42c18fd6170081695/html5/thumbnails/10.jpg)
Second Protocol – 24 Well Harvest
• Cells come out of shaking incubator and go back into BioCel incubator after five days
• Four 24 well plates get reformatted back into a 96 well deep well block on Hamilton
• 96 well deep well block gets spun down in centrifuge
• Supernatant gets pulled off into a new storage plate on Bravo
• All plates get put away
• Process repeats
![Page 11: Using a BioCel System and AssayMap Technology in Phage ... · Using a BioCel System and AssayMap Technology in Phage Display and Antibody Screening Jason Graves September 20, 2011](https://reader030.vdocument.in/reader030/viewer/2022041004/5ea898b42c18fd6170081695/html5/thumbnails/11.jpg)
First Pass Yields – 24 Well Transfection
Automated vs Manual Transfection Yield Comparison
0.00
10.00
20.00
30.00
40.00
50.00
60.00
70.00
80.00
90.00
G5_BP1 G5_BP2 G5_BP3 G5_BP4 G5_M0hr G5_M3hr G5_M6hr
![Page 12: Using a BioCel System and AssayMap Technology in Phage ... · Using a BioCel System and AssayMap Technology in Phage Display and Antibody Screening Jason Graves September 20, 2011](https://reader030.vdocument.in/reader030/viewer/2022041004/5ea898b42c18fd6170081695/html5/thumbnails/12.jpg)
Sequence and Yield Correlation
• Sequences that binned produced similar expression levels as IgG
• Wells with no antibody have sequences with stop codons
1 2 3 4 5 6 7 8 9 10 11 12
A 256 243 245 249 252 255 254 267 277 276 282 5
B 265 243 146 236 181 255 233 264 264 258 11 199
C 278 287 242 260 279 273 264 246 287 271 304 265
D 272 274 246 339 218 246 278 252 98 268 187 274
E 231 292 277 283 249 267 260 261 285 291 282 298
F 306 317 298 276 294 266 295 285 282 304 308 188
G 301 249 321 317 299 325 309 315 335 318 329 345
H 5 350 328 328 327 307 319 310 330 304 325 349
1 2 3 4 5 6 7 8 9 10 11 12
A 12 0 13 15 67 104 0 0 0 0 27 0
B 0 1.4 0 0 0 0 0 0 13 12 0 0
C 0 51 0 0 12 11.8 0 0 0 0 0 0
D 0 0 0 32 25 0 5.4 0 12 12 0 7.3
E 0 0 0 0 0 53.2 0 5.9 62 24 0 12
F 0 0 7.5 0 0 63.9 12 8.4 0 0 6 0
G 0 0 4.1 0 2.6 6.97 0 0 9.5 0 11 5.6
H 0 23 14 0 14 13.3 0 11 0 0 0 9.4
DNA Concentration (ug/mL)
Antibody Concentration (ug/mL)
![Page 13: Using a BioCel System and AssayMap Technology in Phage ... · Using a BioCel System and AssayMap Technology in Phage Display and Antibody Screening Jason Graves September 20, 2011](https://reader030.vdocument.in/reader030/viewer/2022041004/5ea898b42c18fd6170081695/html5/thumbnails/13.jpg)
Yield Comparison
Format vs Yield
0
50
100
150
200
250
96 Well 48 Well 24 Well 30 mL
![Page 14: Using a BioCel System and AssayMap Technology in Phage ... · Using a BioCel System and AssayMap Technology in Phage Display and Antibody Screening Jason Graves September 20, 2011](https://reader030.vdocument.in/reader030/viewer/2022041004/5ea898b42c18fd6170081695/html5/thumbnails/14.jpg)
Transfection and Harvest Numbers
• We have the capacity to process up to 5,000 clones a week (~8 hour day)
• ~3.5 hours/10 DNA plates transfection
• ~2.5 hours/10 DNA plates harvest
• We’ve processed just over 30,000 clones in the first year of production
![Page 15: Using a BioCel System and AssayMap Technology in Phage ... · Using a BioCel System and AssayMap Technology in Phage Display and Antibody Screening Jason Graves September 20, 2011](https://reader030.vdocument.in/reader030/viewer/2022041004/5ea898b42c18fd6170081695/html5/thumbnails/15.jpg)
BioCel #2
![Page 16: Using a BioCel System and AssayMap Technology in Phage ... · Using a BioCel System and AssayMap Technology in Phage Display and Antibody Screening Jason Graves September 20, 2011](https://reader030.vdocument.in/reader030/viewer/2022041004/5ea898b42c18fd6170081695/html5/thumbnails/16.jpg)
BioCel #2
![Page 17: Using a BioCel System and AssayMap Technology in Phage ... · Using a BioCel System and AssayMap Technology in Phage Display and Antibody Screening Jason Graves September 20, 2011](https://reader030.vdocument.in/reader030/viewer/2022041004/5ea898b42c18fd6170081695/html5/thumbnails/17.jpg)
BioCel #2
![Page 18: Using a BioCel System and AssayMap Technology in Phage ... · Using a BioCel System and AssayMap Technology in Phage Display and Antibody Screening Jason Graves September 20, 2011](https://reader030.vdocument.in/reader030/viewer/2022041004/5ea898b42c18fd6170081695/html5/thumbnails/18.jpg)
High Throughput Protein Purification
• Unpurified IgG supernatants are fine for some ELISA and flow cytometry assays.
• Purified IgGs are required to run in most functional screens
• There is a desire to have the capability to purify large numbers of antibodies on a smaller volume scale than what is normally done by the protein purification group
![Page 19: Using a BioCel System and AssayMap Technology in Phage ... · Using a BioCel System and AssayMap Technology in Phage Display and Antibody Screening Jason Graves September 20, 2011](https://reader030.vdocument.in/reader030/viewer/2022041004/5ea898b42c18fd6170081695/html5/thumbnails/19.jpg)
What do we need?
• We need the ability to purify up to 5 mL of IgG or Fab supernatants with a minimum output of 10 μg.
• The method should be plate based in 96 well format to be able to be truly high throughput.
• The method should be endotoxin free.
• The method should be reasonably inexpensive.
![Page 20: Using a BioCel System and AssayMap Technology in Phage ... · Using a BioCel System and AssayMap Technology in Phage Display and Antibody Screening Jason Graves September 20, 2011](https://reader030.vdocument.in/reader030/viewer/2022041004/5ea898b42c18fd6170081695/html5/thumbnails/20.jpg)
What have we tried?
Purification Tips
Purification Plates
Magnetic Beads
Cost ~$150/96w ~$750/96w ~$1200/96w
% Recovery 20-30% 10-15% 20-30%
Automatable Yes Not really Semi
Notes >20μg input -- >20μg input
![Page 21: Using a BioCel System and AssayMap Technology in Phage ... · Using a BioCel System and AssayMap Technology in Phage Display and Antibody Screening Jason Graves September 20, 2011](https://reader030.vdocument.in/reader030/viewer/2022041004/5ea898b42c18fd6170081695/html5/thumbnails/21.jpg)
Lab Automation 2011
![Page 22: Using a BioCel System and AssayMap Technology in Phage ... · Using a BioCel System and AssayMap Technology in Phage Display and Antibody Screening Jason Graves September 20, 2011](https://reader030.vdocument.in/reader030/viewer/2022041004/5ea898b42c18fd6170081695/html5/thumbnails/22.jpg)
First Run – Minimal Optimization
• 300 μL/well of IgG clone split over two plates
• Equilibrate Cartridges
• Load Plate 1
• Load Plate 2
• Wash
• Elute
• Neutralize
• Strip Cartridges
• Equilibrate Cartridges
![Page 23: Using a BioCel System and AssayMap Technology in Phage ... · Using a BioCel System and AssayMap Technology in Phage Display and Antibody Screening Jason Graves September 20, 2011](https://reader030.vdocument.in/reader030/viewer/2022041004/5ea898b42c18fd6170081695/html5/thumbnails/23.jpg)
First Run – Minimal Optimization
• Quantified on the Octet Red
• Input per well – 62 μg
• Average Output – 52 +/- 3 μg (n=62)
• Average % Recovery – 84%
![Page 24: Using a BioCel System and AssayMap Technology in Phage ... · Using a BioCel System and AssayMap Technology in Phage Display and Antibody Screening Jason Graves September 20, 2011](https://reader030.vdocument.in/reader030/viewer/2022041004/5ea898b42c18fd6170081695/html5/thumbnails/24.jpg)
Head to Head with Magnetic Beads
Magnetic Beads AssayMap
Clone Input Output %Rec Input Output %Rec
Clone #1 675 179 27 243 194 80
Clone #2 760 166 22 273 158 58
Clone #3 775 143 18 279 184 66
Clone #4 675 94 14 243 141 58
Clone #5 925 89 10 333 241 72
Clone #6 770 139 18 277 173 62
Clone #7 865 157 18 311 218 70
Avg 18 Avg 67
• Note: Post buffer-exchange
![Page 25: Using a BioCel System and AssayMap Technology in Phage ... · Using a BioCel System and AssayMap Technology in Phage Display and Antibody Screening Jason Graves September 20, 2011](https://reader030.vdocument.in/reader030/viewer/2022041004/5ea898b42c18fd6170081695/html5/thumbnails/25.jpg)
Low Input Purification
Input (μg) Avg Recov (μg) %Recov N
20 15.8 78.8 24
10 7.3 73.3 24
5 3.3 65.0 32
![Page 26: Using a BioCel System and AssayMap Technology in Phage ... · Using a BioCel System and AssayMap Technology in Phage Display and Antibody Screening Jason Graves September 20, 2011](https://reader030.vdocument.in/reader030/viewer/2022041004/5ea898b42c18fd6170081695/html5/thumbnails/26.jpg)
Cartridge Use Limit
0
2
4
6
8
10
12
14
16
18
0 2 4 6 8 10 12 14 16
Plate Number
Avg
Ou
tpu
t (u
g)
![Page 27: Using a BioCel System and AssayMap Technology in Phage ... · Using a BioCel System and AssayMap Technology in Phage Display and Antibody Screening Jason Graves September 20, 2011](https://reader030.vdocument.in/reader030/viewer/2022041004/5ea898b42c18fd6170081695/html5/thumbnails/27.jpg)
AssayMap Purification
• Price
• Reasonable (~$170/96 wells – 10-15 uses per box)
• Automation
• Fully automated platform
• Input volumes
• up to 200 μL per load step
• Endotoxin
• None from cartridges
• Recovery
• Excellent!! ~80% consistently