vad design consideration 1. fit of the system decide who the device intended for. device fit in...
TRANSCRIPT
VAD Design Consideration1. FIT OF THE SYSTEM Decide who the device intended for. Device fit in patient. Cause minimal or no pathologic conditions. Consider volume, mass, dimension, location
of tubes, conduits, connectors. Physical attributes – hard, soft, smooth,
rough, sharp corners damage tissues. Should not project heat. Effect of device movement and vibration. Acceptable sound levels at various
frequencies. Meet EM interference, susceptibility standard.
VAD Design Consideration2. PUMP PERFORMANCE Performance specified in cardiac output range. Cardiac output performance obtained at physiologic
pressure (left atrial pressure ~7mmHg, aortic pressure~100mmHg etc).
Control of device must be included. Device must respond to patient’s cardiac output
requirement.3. BIOCOMPATIBILITY Device must not cause excessive damage to biologic
system. Minimally thrombogenic and minimally hemolytic. Minimal effect on immune system. Should not promote infection, calcification, tissue
necrosis.
VAD Design Consideration4. RELIABILITY Design specification must have :
Target reliability for device (e.g 80% confidence for 2 yrs) Clarification on components that could be changed if
necessary Deal with any service that device may require.
Example : device design life is 5 yrs but battery replacement at 2 yrs interval is allowed
5. QUALITY OF LIFE Designer must specify what is a satisfactory quality of life. Quality of life must be considered in relation to patient’s
quality of life without the device. How much weight older patient can carry? How often power source required to recharge? What sound level is acceptable?