validation and optimization of adiabatic t and t for
TRANSCRIPT
1
Validation and optimization of adiabatic T1ρ and T2ρ for
quantitative imaging of articular cartilage at 3T
Victor Casula 1,2, Joonas Autio 3, Mikko J. Nissi 1,4-6, Edward J. Auerbach 4, Jutta Ellermann
4, Eveliina Lammentausta 7, and Miika T. Nieminen 1,2,7
1 Research Unit of Medical Imaging, Physics and Technology, University of Oulu, Oulu,
Finland,
2 Medical Research Center, University of Oulu and Oulu University Hospital, Oulu, Finland,
3 Center for Life Science and Technologies, RIKEN, Kobe, Japan
4 Center for Magnetic Resonance Research, Department of Radiology, University of
Minnesota, Minneapolis, United States
5 Department of Applied Physics, University of Eastern Finland, Kuopio, Finland,
6 Diagnostic Imaging Center, Kuopio University Hospital
7 Department of Diagnostic Radiology, Oulu University Hospital, Oulu, Finland
Address correspondence and reprint requests to: V. Casula, Research Unit of Medical
Imaging, Physics and Technology, University of Oulu, POB 50, FI-90029 Oulu, Finland Tel:
+358-505904448. E-mail address: [email protected] (V.Casula).
Word count: 4803
Running title: Validation and optimization of adiabatic T1ρ and T2ρ for quantitative imaging
of articular cartilage at 3T
Keywords: articular cartilage; T1ρ; T2ρ; spin-lock; adiabatic; relaxation; in vivo.
2
Abstract
Purpose: The aim of the present work was to validate and optimize adiabatic T1ρ and T2ρ
mapping for in vivo measurements of articular cartilage at 3T.
Methods: Phantom and in vivo experiments were systematically performed on a 3 T clinical
system to evaluate the sequences using hyperbolic secant HS1 and HS4 pulses. R1ρ and R2ρ
relaxation rates were studied as a function of agarose and chondroitin sulfate concentration and
pulse duration. Optimal in vivo protocol was determined by imaging the articular cartilage of
two volunteers and varying the sequence parameters, and successively applied in eight
additional subjects. Reproducibility was assessed in phantoms and in vivo.
Results: Relaxation rates depended on agarose and chondroitin sulfate concentration. The
sequences were able to generate relaxation time maps with pulse length of 8 ms and 6 ms for
HS1 and HS4, respectively. In vivo findings were in good agreement with the phantoms. The
implemented adiabatic T1ρ and T2ρ sequences demonstrated regional variation in relaxation
time maps of femorotibial cartilage. Reproducibility in phantoms and in vivo was good to
excellent for both adiabatic T1ρ and T2ρ.
Conclusions: The findings indicate that sequences are suitable for quantitative in vivo
assessment of articular cartilage at 3T.
Keywords: articular cartilage; T1ρ; T2ρ; spin-lock; adiabatic; relaxation; in vivo.
3
Introduction
The biological applications of rotating frame relaxation techniques extend the capabilities of
quantitative MRI to study the biophysical properties of tissues, due to the high sensitivity of
the techniques for slow molecular motion. Relaxation during a continuous wave spin-lock pulse
for evaluation of articular cartilage has so far been extensively investigated [1]. T1ρ has been
shown to be sensitive to cartilage degradation [2-5], particularly to glycosaminoglycan (GAG)
content [4-8]. However, dependence on the properties of the collagen network has also been
reported [9]. Recent studies proposed novel imaging techniques where trains of adiabatic fast
passage pulses [10-13] have been adopted for spin-locking [14-18]. The main advantage of
using adiabatic pulses for in vivo implementation of rotating frame experiments resides in their
ability to overcome some of the limitations intrinsically associated with continuous wave spin-
lock sequences. Most notably, adiabatic pulses are insensitive to tilt angle inaccuracies by
definition [13], and hence less prone to the undesirable effects of non-uniform magnetic fields
compared to continuous wave spin-lock pulses [19]. Moreover, the adiabatic pulse approach
allows more flexibility in the pulse design and thus benefits moderation of the RF power
deposition [13], which instead significantly hampers the application of continuous wave pulses
in the clinical realm [20]. Furthermore, through an extended range of correlation times
effectively involved in the relaxation during adiabatic pulses, more of the physicochemical
mechanisms underlying pathological changes in tissues may be monitored to elucidate
information potentially relevant for diagnostic purposes.
Adiabatic fast passage pulses have been successfully applied for assessing longitudinal and
transverse relaxation times in the rotating frame, namely adiabatic T1ρ (AdT1ρ) and adiabatic
T2ρ (AdT2ρ), and their ability to probe slow molecular motion has been proven at different field
strengths [14, 16-18, 21]. For articular cartilage, preclinical studies on AdT1ρ and AdT2ρ have
demonstrated evidence of improved sensitivity for detection of early degenerative changes in
both animal and human specimens over rotating frame relaxation times assessed with
conventional spin lock experiments or free precession relaxation times [22-25]. Furthermore
AdT1ρ has been reported less sensitive to the magic angle effect compared to T2 and continuous
wave T1ρ [26]. Such ability in characterizing degeneration of cartilage could be exploited to
provide valuable and practical diagnostic application for clinical use. The rationale of this study
was to validate and optimize adiabatic T1ρ and adiabatic T2ρ sequences on a 3 T clinical scanner
for in vivo quantitative measurements of cartilage.
4
5
Methods
MR pulse sequences
AdT1ρ and AdT2ρ sequences were implemented on a 3 T clinical system (Skyra, Siemens,
Erlangen, Germany) using the IDEA sequence programming environment (Version
VD13A/SP4, Siemens). Adiabatic T1ρ mapping was performed as described previously [18, 21,
22, 24, 25] using a preparation block which consisted, for both phantom and in vivo
experiments unless otherwise stated, of a train of 0, 4, 8, 12 and 16 adiabatic full passage (AFP)
hyperbolic secant pulses of the HSn family (n = 1,4) with R = 10 and phase cycled with MLEV4
[13, 27, 28] (Fig. 1a). For AdT2ρ, the AFP pulses were placed between two adiabatic half
passage pulses (AHP), and two spatial pre-saturation slabs were applied above and below the
FOV to correct for off-resonance effects. A 15-channel transmit/receive knee coil (single
channel transmit) was used. For both AdT1ρ and AdT2ρ, the peak RF amplitude was set to γB1max
= 800 Hz (highest achievable with the selected coil). The preparation block was followed by a
segmented gradient recalled echo (FLASH) readout.
Phantom measurements
Eleven gel tubes were prepared with different concentrations of agarose (0.50 - 4.00% w/v) to
yield transverse relaxation times within the range of biological tissue, and served as a test
phantom for evaluating the feasibility, performance, stability, reproducibility and operational
limits of the adiabatic T1ρ and T2ρ sequences. Imaging parameters were adjusted for the
phantom (TR / TE = 20 s / 3.75 ms, flip angle = 15°, matrix = 256 x 256, FOV = 120 x 120
mm2, slice thickness = 7 mm, 2 averages), and AdT1ρ and AdT2ρ were measured with five
different AFP pulse durations (Tp), i.e. 3, 6, 8, 10 and 12 ms. Additionally, T2 relaxation time
of the phantom was measured with a multi-echo spin echo sequence (TR = 10 s, 32 TE between
10.7 and 342.4 ms, matrix = 256 x 256, FOV = 120 x 120 mm2, slice thickness = 7 mm, 2
averages). To assess the homogeneity of B1, flip angle maps were measured using B1 mapping
sequence provided by the scanner manufacturer (Siemens). The relaxation times were
investigated as relaxation rates R2 (= 1/T2), R1ρ (= 1/T1ρ) and R2ρ (= 1/T2ρ) due to the more
straightforward analysis. Furthermore relaxivities r2, r1ρ and r2ρ as a function of agarose
concentration were determined.
6
Additionally, AdT1ρ and AdT2ρ experiments were performed for the three lowest agarose
concentration tubes and for a supplementary distilled water phantom using longer total spin
lock durations (trains of 0, 4, 8, 12, 16, 20, 24, 28, 32, 36 and 40 pulses, Tp = 8 ms for HS1 and
Tp = 6 ms for HS4) to better observe the relaxation behavior when approaching pure water
relaxation.
To specifically study the dependence of adiabatic relaxation time measurements on cartilage
proteoglycans, chondroitin sulfate (CS) phantoms with various concentrations (0, 2, 4, 6, 8 %
w/w) were prepared from chondroitin sulfate A (CS; Sigma-Aldrich, Schnelldorf, Germany) in
PBS containing 0.2 mM MnCl2 to reduce T1, with pH of the solutions titrated to 7.1, which has
been shown to be the pH of clinically normal human cartilage [29]. R1ρ and R2ρ were measured
(TR / TE = 5 s / 3.75 ms, flip angle = 15°, matrix = 256 x 256, FOV = 120 x 120 mm2, slice
thickness = 7 mm, 2 averages) and studied as functions of CS concentration with four different
Tps (6, 8, 10 and 12 ms). For reference, R2 was also measured (TR = 5 s, 32 TE between 10.7
and 342.4 ms, matrix = 256 x 256, FOV = 120 x 120 mm2, slice thickness = 7 mm, 2 averages).
Relaxivities r2, r1ρ and r2ρ were calculated as a function of the chondroitin sulfate concentration.
In vivo measurements
To demonstrate the feasibility of AdT1ρ and AdT2ρ measurements in the human knee joint, ten
asymptomatic adult volunteers (age range 25-35 y, mean age 27.4 y; six male, four female)
were included in this study. Additionally, three subjects with radiographic signs of
osteoarthritis (Kellgren-Lawrence (KL) grade ≥ 2) were enrolled: a 52 y male (KL = 2), a 62 y
female (KL = 3) and a 57 y male (KL = 4). Approval from the local institutional review board
on human research ethics and informed consents from the subjects were obtained prior to the
study.
To optimize the relevant sequence parameters, including flip angle, repetition time and pulse
duration, the first two subjects were scanned with the following protocols:
1. Single sagittal slices of right medial condyle were taken with no T1ρ-preparation; the
flip angle was varied (6, 10, 15, 20, 30, 45, 60 and 90 degrees) and the optimal value
was evaluated in terms of SNR and absence of artifacts.
7
2. At the same anatomical location, adiabatic T1ρ relaxation maps were obtained with four
different TRs (1, 2.5, 4, 6 s) and analyzed to find the shortest suitable TR. Flip angle
was taken from the previous step and the preparation pulse lengths were the shortest
fulfilling the adiabatic condition [13, 28] on the basis of the phantom findings, i.e. Tp,HS1
= 8 ms and Tp,HS4 = 6 ms.
3. AdT1ρ and AdT2ρ experiments were finally repeated with three different Tps (3, 6, 8, 10
and 12 ms), using the determined optimal TR and flip angle from steps 1 and 2.
In this adopted approach, the optimal parameters were determined by evaluating the relaxation
time maps and by minimizing the scanning time and RF power deposition while maximizing
the quality of the images in terms of signal to noise ratio, stability and absence of artifacts. The
other acquisition parameters for all the in vivo experiments were TE = 3.36 ms, FOV = 180 x
180 mm2, matrix = 256 x 256, and slice thickness = 3 mm. In addition, parallel imaging
acceleration factor = 2 (GRAPPA) and segmented k-space acquisition (23 acquisitions per
preparation) were used to decrease the scanning time to be clinically suitable with no
significant reduction of image quality. Similar to phantom experiments, the uniformity of B1
was evaluated by acquiring the flip angle maps using the B1 mapping sequence described
above.
The optimized protocols were then used to image the knees of all the ten volunteers. T1ρ and
T2ρ data were acquired and maps computed in slices covering the centers of medial and lateral
condyles. To determine the in vivo reproducibility, measurements with the optimized protocols
were repeated three times in four subjects, with knee repositioning between consecutive scans.
For comparison purposes, T2 measurements were performed scanning the same joint locations
with a multi-echo spin echo sequence (TR = 1.680 s, five TEs between 13.8 and 69 ms, matrix
= 384 x 384, FOV = 160 x 160 mm2, slice thickness = 3 mm).
To explore the potential of the optimized protocol in clinical setting, we also scanned three OA
patients. The pulse type with the shortest pulse duration Tp based on the phantom and in vivo
measurements (here HS4), less demanding in terms of power deposition and thus more suitable
for clinical examinations, was employed. The relaxation time constants of articular cartilage
were determined in two slices positioned at the centers of the lateral and medial condyles.
8
Data processing
Phantom and in vivo image analyses were executed with an in-house MATLAB-based
software. Longitudinal and transverse relaxation time constants in the rotating frame were
calculated by mono-exponential fitting of the signal intensity decays on a pixel-by-pixel basis.
In all the in vivo images, articular cartilage was manually segmented by a single operator (V.C.)
and 18 ROIs were considered: each condyle was divided into six superficial and six deep ROIs,
and each tibial plateau into three superficial and three deep ROIs (Fig. 1b).
Statistical analysis
Coefficient of determination (R2) and root-mean-square deviation (RMSD) values were
calculated between the relaxation rates and agarose or chondroitin sulfate concentrations. The
reproducibility of the techniques was assessed with the agarose tubes by calculating the average
root-mean-square coefficient of variation (CVRMS) of three measurements repeated within one
month using Tp = 8 ms and Tp = 6 ms for HS1 and HS4, respectively. To determine the in vivo
reproducibility of AdT1ρ and AdT2ρ, CVRMS of three consecutive scans in four subjects was
calculated for bulk cartilage and for each ROI.
9
Results
Agarose gel and chondroitin sulfate phantoms
R1ρ and R2ρ were strongly linearly correlated with agarose concentration (Fig. 2 and Sup. Fig.
S1, R2 = 0.98 – 0.99, RMSD = 0.24 – 1.25). Using Tp = 6 ms with HS1, linear correlation
between relaxation rates and agarose concentration was smaller (R2 = 0.93 - 0.94, RMSD =
0.69 – 1.39) and the maps were less homogeneous (Fig. 3). For shortest pulse length (Tp = 3
ms) relaxation rate values exhibited notable errors, the correlation with concentration was
reduced (R2 = 0.70 – 0.85, RMSD = 3.02 – 18.67) and the relaxation time maps were largely
inhomogeneous (Fig. 3). Table 1 reports the relaxivities as a function of agarose concentration.
Relaxivity r1ρ was smaller than r2ρ. Furthermore, r1ρ increased with pulse duration and was
greater with HS4. On the contrary, r2ρ decreased with pulse duration and was greater with HS1.
In the agarose phantom, relaxation rates showed a tendency to converge at lower concentrations
(Fig. 2), except for HS1 with Tp = 6 ms. This convergence was better shown for R1ρ,HS1 and
R2ρ,HS1 with Tp = 8 ms and R1ρ,HS4 and R2ρ,HS4 with Tp = 6 ms, when measured using a larger
number of and including longer duration spin-lock times in the low concentration samples
(Sup. Fig. S1). Note that all the relaxation times in the reference water phantom were very close
to each other (Sup. Fig. S1).
For the agarose phantom, the mean relative error of the flip angle in the imaged region was
3.6%. The reproducibility assessed with agarose phantom (the CVRMS) for the AdT1ρ sequence
was 2.2% with HS1 and 1.3% with HS4. For the AdT2ρ sequence the reproducibilities were
2.9% and 2.3% with HS1 and HS4, respectively.
Although relaxation rates for the CS phantom were generally smaller than those of the agarose
tubes, all of the aforementioned findings were confirmed with the CS phantom as well (Fig. 4).
R1ρ and R2ρ were strongly linearly correlated with CS concentration (R2 = 0.98 – 0.99, RMDS
= 0.08 – 0.22). Table 1 shows the relaxivities for CS phantoms. For CS, r2ρ was greater than
r1ρ.
In vivo measurements of articular cartilage
10
The SNR of the FLASH images in cartilage region showed a maximum between flip angles of
20 to 30 degrees (SNR = 68 - 98) and monotonically decreased to a minimum at 90 degrees
(SNR = 28 - 48). At larger flip angles the images were visibly affected by aliasing artifact,
which increased towards greater angles. At 15 degrees the SNR was almost doubled compared
to 6 degrees (SNR = 23 - 30) and the images were free of artifacts. For the FLASH readout
sequence utilized, flip angle of 15 degrees was found to be the best compromise between good
SNR (~ 54 - 60) and preventing the aliasing artifacts that occurred at larger flip angles.
Table 2 shows the T1ρ relaxation times of the entire femoral and tibial AC as obtained using
different TR values. Relaxation with the shortest TR (= 1 s) was considerably faster as
compared to those measured with longer TRs (mean difference = 27.2 %). Mean difference
between relaxation times obtained with TR of 4 and 6 s was minimal (2.2 %) and smaller than
the mean difference of T1ρ between TRs of 2.5 and 6 s (4.8 %). Therefore repetition time of 4
s was considered the best compromise between scan time and relaxation dependency on TR.
The variability in T1ρ and T2ρ relaxation times with HS1 and HS4 pulse duration is reported in
Table 3 and summarized below. Very short and incongruous relaxation values were measured
with Tp of 3 ms irrespective of the pulse type, in addition to atypical variances across T1ρ maps
caused by signal dropout and distortion affecting the images and increasing with spin-lock
time. Such features, indicating violation of the adiabaticity, were still observed, although
largely mitigated in T1ρ maps acquired with HS1 pulses of 6 ms. With longer pulse durations
for HS1, limited dependency on pulse duration was noted for relaxation time. Using HS4, T1ρ
revealed a maximum with pulse duration of 6 ms and then a small tendency to decrease with
increasing Tps. Conversely, T2ρ appeared to slightly increase with increasing pulse duration.
Based on these findings, Tp of 8 and 6 ms were the shortest pulse lengths satisfying the adiabatic
condition for HS1 and HS4, respectively. The scanning time for obtaining full relaxation data
necessary for fitting of the time constants with the final protocol was 2 min and 22 s per slice.
In the in vivo scans, the mean error of the flip angle was 3.2% in femoral cartilage and 3.5%
in tibial cartilage. Supporting Table S1 reports the in vivo reproducibilities. CVRMS was lower
than 5% in bulk ROIs of lateral and medial compartments for both T1ρ and T2ρ, except in medial
tibia for T1ρ,HS1.
Figure 5 shows representative T1ρ and T2ρ relaxation time maps of articular cartilage obtained
using the most appropriate settings found in this study (flip angle = 15 degrees, TR = 4 s, Tp =
11
8 ms for HS1 and Tp = 6 ms for HS4). Table 4 reports the mean and standard deviation of T2,
T1ρ and T2ρ for each ROI. Anterior trochlea could not be analyzed in the medial compartment
due to the slice orientation. In all the ROIs, T1ρ,HS1 was always longer than T1ρ,HS4. T2ρ,HS1 was
always shorter than T2ρ,HS4, except in deep central tibia. On average, relaxation times were
longer in femur than in tibia and longer in the superficial zone than in the deep cartilage. The
shortest relaxation values were found in deep lateral central tibia. The longest values of T2, T1ρ
and T2ρ were found among superficial trochlear and central femoral ROIs. Figure 6 shows a
comparison of relaxation time maps between a representative healthy subject and three patients
with radiographic OA. On average, T1ρ and T2ρ relaxation times of femorotibial cartilage with
OA were found respectively 13-52% and 18-68% longer compared to healthy control mean
values, with a trend for increasing values with increasing KL score.
12
Discussion
Relaxation in the rotating frame can probe the long time scale component of molecular
interactions; as a result the influence of processes such as chemical exchange becomes more
relevant as compared to free relaxation where the contribution of direct dipolar-dipolar
interactions is prevailing. This has been proposed as an explanation for the sensitivity of
continuous wave T1ρ to cartilage proteoglycans [30, 31] and reduced sensitivity to the magic
angle effect as compared to T2 [32]. To diminish the magic angle effect in continuous wave
T1ρ, it has been calculated that spin lock frequencies of 2 kHz or higher would be necessary
[33], far beyond the power accessible in the clinical realm. The significant advantages of AdT1ρ
are the simultaneous sensitivity to more than one time scale [16, 17] and the fact that it is
essentially unaffected by susceptibility and magic angle effect [26]. Previously, T1ρ dependence
on CS and proteoglycans of articular cartilage (of which CS is a major component) has been
shown [6, 7, 20, 34]. Nonetheless in literature there is no consensus regarding the T1ρ relaxation
mechanisms in cartilage. A recent study has showed that T1ρ does not correlate with
glycosaminoglycan [35] and another study suggested that the contribution of scalar relaxation
becomes relevant only at high fields, whilst at clinical fields dipolar interactions are the
dominant relaxation mechanism in T1ρ [36]. The relaxivities found in this study are greater than
those reported in previous studies at clinical fields and comparable to r1ρ at higher fields [34,
37]. This suggests that AdT1ρ has better sensitivity to CS as compared to continuous wave T1ρ.
AdT2ρ has been reported to be sensitive to early cartilage degeneration in an animal model of
OA and advanced OA in human tissue [22, 25, 38]. Although the biophysical mechanisms of
T2ρ contrast in cartilage remain unknown, it has been proposed that the contribution of proton
exchange is more effective as compared to T1ρ. Thus T2ρ relaxation process should be more
sensitive to cartilage proteoglycans than T1ρ. This is consistent with the r2ρ relaxivities found
in this study, which were greater than the r1ρ relaxivities (Table 1). Since continuous wave T2ρ
is particularly sensitive to B1 inhomogeneities and requires high spin-lock fields, its utilization
in clinical studies has been limited by SAR constrains. Therefore AdT2ρ, thanks to the
insensitivity of adiabatic pulses to B1 inhomogeneities, provides a unique means for measuring
transverse relaxation in the rotating frame in vivo.
In this study the feasibility of AdT1ρ and AdT2ρ on a clinical MRI system was documented via
systematic phantom and in vivo tests. Through phantom experiments it was verified that both
13
of the sequences functioned as expected and produced results consistent with previous studies.
Specifically, in agarose phantoms the relaxation rates increased with increasing gel density [39,
40] and converged to the pure water relaxation (R1ρ ≈ R2ρ ≈ R2) towards lower concentrations.
Furthermore, R1ρ,HS1 was smaller than R1ρ,HS4 [24] and conversely R2ρ,HS1 was greater than
R2ρ,HS4 in the agarose phantom [21], in agreement with previous studies and the in vivo
measurements. Relaxation was also found to be dependent on the pulse duration. Increasing
the length of an AFP pulse corresponds to decreasing of the frequency sweep amplitude [13],
which in turn yielded faster R1ρ and slower R2ρ, as expected [21]. The aforementioned results
were further confirmed by CS experiments, except for approximating the pure water relaxation,
effect not visible here since phantoms were paramagnetically doped.
Phantom and in vivo measurements were congruent indicating the shortest suitable pulse
duration Tp for HS1 and HS4. With 3 ms pulses, the images exhibited severe random artifacts
suggesting violation of the adiabatic condition [13], wherefrom the weak association of
relaxation rates with agarose and CS content. With 6 ms pulses using HS1 shape, mild errors
were observed in the relaxation maps. The same inhomogeneities were found in vivo and not
imputable to the physiological variation within the tissue, rather revealing a sub-adiabatic
regime. This explains the unexpected behavior of the relaxation rates obtained using HS1 with
6 ms pulse duration. Conversely, using HS4 pulses with Tp of 6 ms resulted in homogenous
maps and in shorter R1ρ and greater R2ρ as compared to longer Tps. Such dissimilarity is
attributed to be a direct consequence of the different “stretching” factors of the HSn pulses.
HS4 pulses are de facto less demanding compared to HS1 in terms of peak RF power
requirements. The peak power required to fully satisfy the adiabatic condition with 6 ms HS1
pulses exceeded the capabilities of the RF transmit chain.
The reproducibility of AdT1ρ and AdT2ρ in phantom, as indicated by the CVRMS of agarose
tubes, was excellent for the shortest suitable pulse durations, i.e. 8 ms for HS1 and 6 ms for
HS4. Previous studies have reported in vivo T1ρ reproducibility to range from 1.7-19% [41-
43]. In vivo AdT1ρ reproducibility ranged from 1.1% to 10.8%, was excellent in over 70% of
the ROIs and was greater than 10% in only one ROI (Sup. Table S1). In vivo AdT2ρ
reproducibility ranged from 1.4% to 15.3%, was excellent in over 44% of the ROIs and was
14
greater than 10% in only three ROIs (Sup. Table S1). The relatively small CVRMS, while
providing supplementary evidence that sequences are correctly working, demonstrated in
addition their appropriateness for accurate and longitudinally stable quantitative analyses.
The experiments with human subjects demonstrated the in vivo feasibility of the sequences and
indicated the most appropriate protocol for reproducible and reliable relaxation time
quantification. Concerning the optimal flip angle and TR, the choice was mostly led by
conservative arguments. While 15 degrees was not the absolute best flip angle in terms of
signal-to-noise ratio, it produced artifact free images while still retaining excellent SNR; a
smaller flip angle is also conducive to mitigating SAR and was thus considered as the optimal
choice. The need for minimizing the acquisition time with short TRs must meet the minimum
length required to avoid excessive formation of steady-state if one wishes to retain quantitative
relaxation coefficients. Using a TR of 1 second, the relaxation was too short as compared to
longer TRs, suggesting that the signal reduction was rather due to partial saturation. Although
no strong variations in relaxation times were observed using the other considered TRs, a closer
look revealed that the least difference in average was between values obtained with 4 and 6s
TR, hence the former appears to be a more cautious selection than 2.5 s, yet preferable to 6 s
for decreasing the acquisition time. Different delay times between two consecutive RF spin-
lock pulses have been used in previous T1ρ cartilage studies. Witschey et al., for example, used
a conservative value of 6 s [44]. A delay time, shorter than that empirically determined in the
present study has been employed in combination with RF cycling [42]. The TR used in this
study with FLASH readout appeared as the optimal compromise between acceptable clinical
acquisition times and avoiding formation of a complex steady-state.
Regarding the durations of the HS1 and HS4 pulses, the most important results were the
shortest possible Tps, which were discussed above. The scanning time with the optimal protocol
was appropriate for clinical applications, allowing reliable quantification of T1ρ and T2ρ of two
slices. The small error in the flip angle both in vivo and in phantoms indicated presence of only
a mild B1 inhomogeneity at 3 T in the setup used, which could be compensated by the use of
adiabatic pulses. In vivo experiments showed variation in relaxation times depending on
topographical location (medial or lateral compartment; femur or tibia; anterior, central or
posterior region), depth (superficial or deep zone), pulse used (HS1 or HS4), and contrast (T1ρ
or T2ρ). The findings were consistent with those reported in previous studies. Specifically
longer T1ρ values were obtained with HS1 as compared to HS4, with a mean difference of about
15
40% [17, 21, 24]; superficial ROIs had longer relaxation times as compared to deep ROIs [23];
T1ρ were longer than T2ρ [23]; T2 were shorter than T1ρ [3, 45]. T1ρ, T2ρ and T2 exhibited
differences in regional distribution, attributable to the different information regarding the tissue
provided by the three parameters. These results confirm the ability of the sequences to probe
differences in the structure and composition of AC also at clinically accessible fields.
The increased adiabatic T1ρ and T2ρ observed in cartilage of patients with OA is in agreement
with previous reports that the techniques are sensitive biomarkers for cartilage degeneration
[22, 23, 25]. Further studies are warranted to compare adiabatic T1ρ and T2ρ in asymptomatic
subjects with early OA patients.
Certain limitations restrain the generalizability of some results of this study. For instance,
accuracy of the sequences is likely to change with systems of different manufacturers, in which
also the sequence stability has to be verified. Optimal TR and flip angle may also vary, since
they both are, for example, not independent of B0. Finally the minimum pulse lengths required
for achieving adiabaticity for HS1 and HS4 may differ using different RF coils and amplifiers,
since that depends on the attainable RF power. Implementation-wise, the adiabatic T1ρ and T2ρ
sequences require an introduction of a train of adiabatic inversion pulses in front of a readout
sequence, which could be FLASH, EPI or other appropriate sequence, and should not be a
major limiting factor. However, a limitation of the present setup is the relatively long scan time,
i.e. data for only two slices could be acquired within a reasonable time frame. The TR could
be shortened significantly with RF cycling [46] and further studies are needed to test the
method with trains of HSn pulses. The acquisition could also be sped up by other choices, such
as increasing the portion of k-space acquired per preparation or using partial acquisition [47],
by increasing the parallel acquisition factor or utilizing multiband excitation [48] or by
applying other techniques for speeding up the readout. T1ρ sequences are generally prone to
SAR issues due to the RF energy required to manipulate the contrast before acquisition [20],
the adiabatic pulses utilized here provide flexibility overcoming the SAR issues and changes
in the TR and factors speeding up the acquisition may be utilized to further alleviate SAR.
Longitudinal magnetization recovery and transient signal evolution during acquisition may
degrade T1ρ and T2ρ contrasts. RF cycling [46] and variable flip angle techniques may be tested
in combination with AFP pulse trains to reduce inaccuracies [42, 44].
16
Another possible source of inaccuracies is inherent in the quantitation method of transverse
relaxation time. In this study, T2 values of some phantoms or ROIs in in vivo data were longer
than the corresponding T2ρ values, which in theory should not occur. Rautiainen et al. recently
reported discrepancies in T2 as measured with different sequences on a 9.4 T system and mean
values of T2 longer than T2ρ, depending on the sequence used for measuring T2 [22]. Variability
of T2 has also been observed in phantoms and in vivo measurements carried out on different
clinical scanners and with different pulse sequences [49, 50]. In this study, the higher than
expected T2s were likely due to the properties of the T2-mapping sequence provided by the
system manufacturer.
In conclusion, adiabatic T1ρ and adiabatic T2ρ appear feasible for in vivo quantitative analysis
of cartilage tissue at 3 T within acceptable scanning times for clinical application. Previous
investigations have shown the superior sensitivity of these contrasts for the assessment of
articular cartilage as opposed to other relaxation time measurements [22-24] in the ex vivo
setting and thus with the demonstrated feasibility, the parameters appear promising for in vivo
as well. The potential of these techniques in detecting early osteoarthritis will be investigated
in subsequent studies.
17
Acknowledgments
The Center for Magnetic Resonance Research, University of Minnesota is gratefully
acknowledged for providing the adiabatic T1ρ and T2ρ sequences. The authors would like to
thank Silvia Mangia and Shalom Michaeli for constructive discussions, Anna-Leena Manninen
and Abdul Wahed Kajabi for their advice and assistance with phantom preparation, and Ute
Goerke for the original Siemens implementation of the adiabatic T1ρ sequence. This work was
supported by Academy of Finland (grants 260321, 285909 and 293970).
18
References
1. Wang L, Regatte RR. T1rho MRI of human musculoskeletal system. Journal of Magnetic
Resonance Imaging 2015;41(3):586-600.
2. Regatte RR, Akella SV, Borthakur A, Kneeland JB, Reddy R. In vivo proton MR three-
dimensional T1rho mapping of human articular cartilage: Initial experience. Radiology
2003 Oct;229(1):269-74.
3. Li X, Benjamin Ma C, Link TM, Castillo DD, Blumenkrantz G, Lozano J, Carballido-
Gamio J, Ries M, Majumdar S. In vivo T(1rho) and T(2) mapping of articular cartilage
in osteoarthritis of the knee using 3 T MRI. Osteoarthritis Cartilage 2007 Jul;15(7):789-
97.
4. Keenan KE, Besier TF, Pauly JM, Han E, Rosenberg J, Smith RL, Delp SL, Beaupre GS,
Gold GE. Prediction of glycosaminoglycan content in human cartilage by age, T1rho and
T2 MRI. Osteoarthritis Cartilage 2011 Feb;19(2):171-9.
5. Wheaton AJ, Casey FL, Gougoutas AJ, Dodge GR, Borthakur A, Lonner JH, Schumacher
HR, Reddy R. Correlation of T1rho with fixed charge density in cartilage. J Magn Reson
Imaging 2004 Sep;20(3):519-25.
6. Akella SV, Regatte RR, Gougoutas AJ, Borthakur A, Shapiro EM, Kneeland JB, Leigh JS,
Reddy R. Proteoglycan-induced changes in T1rho-relaxation of articular cartilage at 4T.
Magn Reson Med 2001 Sep;46(3):419-23.
7. Duvvuri U, Kudchodkar S, Reddy R, Leigh JS. T1ρ relaxation can assess longitudinal
proteoglycan loss from articular cartilage in vitro. Osteoarthritis and Cartilage 2002
11;10(11):838-44.
8. Wang N, Xia Y. Depth and orientational dependencies of MRI T(2) and T(1rho)
sensitivities towards trypsin degradation and gd-DTPA(2-) presence in articular cartilage
at microscopic resolution. Magn Reson Imaging 2012 Apr;30(3):361-70.
9. Menezes NM, Gray ML, Hartke JR, Burstein D. T2 and T1rho MRI in articular cartilage
systems. Magn Reson Med 2004 Mar;51(3):503-9.
10. Tannús A, Garwood M. Adiabatic pulses. NMR Biomed 1997;10(8):423-34.
11. Norris DG. Adiabatic radiofrequency pulse forms in biomedical nuclear magnetic
resonance. Concepts Magn Reson 2002;14(2):89-101.
12. De Graaf RA, Nicolay K. Adiabatic rf pulses: Applications to in vivo NMR. Concepts
Magn Reson 1997;9(4):247-68.
13. Garwood M, DelaBarre L. The return of the frequency sweep: Designing adiabatic pulses
for contemporary NMR. J Magn Reson 2001 Dec;153(2):155-77.
19
14. Mangia S, Traaseth NJ, Veglia G, Garwood M, Michaeli S. Probing slow protein
dynamics by adiabatic R 1ρ and R 2ρ NMR experiments. J Am Chem Soc
2010;132(29):9979-81.
15. Mangia S, Carpenter AF, Tyan AE, Eberly LE, Garwood M, Michaeli S. Magnetization
transfer and adiabatic T1 MRI reveal abnormalities in normal-appearing white matter of
subjects with multiple sclerosis. Multiple Sclerosis 2014;20(8):1066-73.
16. Michaeli S, Grohn H, Grohn O, Sorce DJ, Kauppinen R, Springer CS,Jr, Ugurbil K,
Garwood M. Exchange-influenced T2rho contrast in human brain images measured with
adiabatic radio frequency pulses. Magn Reson Med 2005 Apr;53(4):823-9.
17. Michaeli S, Sorce DJ, Springer CS,Jr, Ugurbil K, Garwood M. T1rho MRI contrast in the
human brain: Modulation of the longitudinal rotating frame relaxation shutter-speed
during an adiabatic RF pulse. J Magn Reson 2006 Jul;181(1):135-47.
18. Michaeli S, Sorce DJ, Garwood M. T2ρ and T1ρ adiabatic relaxations and contrasts.
Current Analytical Chemistry 2008;4(1):8-25.
19. Witschey II WRT, Borthakur A, Elliott MA, Mellon E, Niyogi S, Wallman DJ, Wang C,
Reddy R. Artifacts in T1ρ-weighted imaging: Compensation for B1 and B0 field
imperfections. Journal of Magnetic Resonance 2007 5;186(1):75-85.
20. Borthakur A, Mellon E, Niyogi S, Witschey W, Kneeland JB, Reddy R. Sodium and
T1rho MRI for molecular and diagnostic imaging of articular cartilage. NMR Biomed
2006 Nov;19(7):781-821.
21. Mangia S, Liimatainen T, Garwood M, Michaeli S. Rotating frame relaxation during
adiabatic pulses vs. conventional spin lock: Simulations and experimental results at 4 T.
Magn Reson Imaging 2009 Oct;27(8):1074-87.
22. Rautiainen J, Nissi MJ, Liimatainen T, Herzog W, Korhonen RK, Nieminen MT.
Adiabatic rotating frame relaxation of MRI reveals early cartilage degeneration in a
rabbit model of anterior cruciate ligament transection. Osteoarthritis and Cartilage 2014
2014/09;22(10):1444-52.
23. Rautiainen J, Nissi MJ, Salo E, Tiitu V, Finnilä MAJ, Aho O, Saarakkala S, Lehenkari P,
Ellermann J, Nieminen MT. Multiparametric MRI assessment of human articular
cartilage degeneration: Correlation with quantitative histology and mechanical
properties. Magnetic Resonance in Medicine 2015;74(1):249-59.
24. Ellermann J, Ling W, Nissi MJ, Arendt E, Carlson CS, Garwood M, Michaeli S, Mangia
S. MRI rotating frame relaxation measurements for articular cartilage assessment. Magn
Reson Imaging 2013;31(9):1537-43.
25. Nissi MJ, Salo EN, Tiitu V, Liimatainen T, Michaeli S, Mangia S, Ellermann J, Nieminen
MT. Multi-parametric MRI characterization of enzymatically degraded articular
cartilage. J Orthop Res 2015. doi: 10.1002/jor.23127.
20
26. Nissi MJ, Mangia S, Michaeli S, Nieminen MT. Orientation anisotropy of rotating frame
and T2 relaxation parameters in articular cartilage. . Proc Intl Soc Mag Reson Med
2013;21:3552.
27. Warnking JM, Pike GB. Bandwidth-modulated adiabatic RF pulses for uniform selective
saturation and inversion. Magnetic Resonance in Medicine 2004;52(5):1190-9.
28. Silver MS, Joseph RI, Hoult DI. Highly selective and p pulse generation. Journal of
Magnetic Resonance (1969) 1984;59(2):347-51.
29. Konttinen YT, Mandelin J, Li T-, Salo J, Lassus J, Liljeström M, Hukkanen M, Takagi
M, Virtanen I, Santavirta S. Acidic cysteine endoproteinase cathepsin K in the
degeneration of the superficial articular hyaline cartilage in osteoarthritis. Arthritis
Rheum 2002;46(4):953-60.
30. Duvvuri U, Reddy R, Patel SD, Kaufman JH, Kneeland JB, Leigh JS. T1rho-relaxation in
articular cartilage: Effects of enzymatic degradation. Magn Reson Med 1997
Dec;38(6):863-7.
31. Duvvuri U, Goldberg AD, Kranz JK, Hoang L, Reddy R, Wehrli FW, Wand AJ,
Englander SW, Leigh JS. Water magnetic relaxation dispersion in biological systems:
The contribution of proton exchange and implications for the noninvasive detection of
cartilage degradation. Proc Natl Acad Sci U S A 2001 Oct 23;98(22):12479-84.
32. Li X, Cheng J, Lin K, Saadat E, Bolbos RI, Jobke B, Ries MD, Horvai A, Link TM,
Majumdar S. Quantitative MRI using T1rho and T2 in human osteoarthritic cartilage
specimens: Correlation with biochemical measurements and histology. Magn Reson
Imaging 2011 Apr;29(3):324-34.
33. Akella SVS, Regatte RR, Wheaton AJ, Borthakur A, Reddy R. Reduction of residual
dipolar interaction in cartilage by spin-lock technique. Magnetic Resonance in Medicine
2004;52(5):1103-9.
34. Akella SV, Regatte RR, Borthakur A, Reddy R. T1ρ relaxation rate and dispersion of
cartilage at 1.5 and 4T. Proc Intl Soc Mag Reson Med 2003;10:1503.
35. van Tiel J, Kotek G, Reijman M, Bos PK, Bron EE, Klein S, Nasserinejad K, van Osch,
Gerjo J. V. M., Verhaar JAN, Krestin GP, et al. Is T1ρ mapping an alternative to delayed
gadolinium-enhanced MR imaging of cartilage in the assessment of sulphated
glycosaminoglycan content in human osteoarthritic knees? an in vivo validation study.
Radiology 2015 11/20; 2015/12:150693.
36. Mlynarik V, Szomolanyi P, Toffanin R, Vittur F, Trattnig S. Transverse relaxation
mechanisms in articular cartilage. J Magn Reson 2004 Aug;169(2):300-7.
37. Regatte RR, Akella SV, Borthakur A, Reddy R. Proton spin-lock ratio imaging for
quantitation of glycosaminoglycans in articular cartilage. J Magn Reson Imaging 2003
Jan;17(1):114-21.
21
38. Jokivarsi KT, Niskanen J-, Michaeli S, Gröhn HI, Garwood M, Kauppinen RA, Gröhn
OH. Quantitative assessment of water pools by T1ρ and T 2ρ MRI in acute cerebral
ischemia of the rat. Journal of Cerebral Blood Flow and Metabolism 2009;29(1):206-16.
39. Blumenkrantz G, Li X, Han ET, Newitt DC, Crane JC, Link TM, Majumdar S. A
feasibility study of in vivo T1rho imaging of the intervertebral disc. Magn Reson
Imaging 2006 Oct;24(8):1001-7.
40. Li X, Han ET, Ma CB, Link TM, Newitt DC, Majumdar S. In vivo 3T spiral imaging
based multi-slice T(1rho) mapping of knee cartilage in osteoarthritis. Magn Reson Med
2005 Oct;54(4):929-36.
41. Li X, Pedoia V, Kumar D, Rivoire J, Wyatt C, Lansdown D, Amano K, Okazaki N, Savic
D, Koff MF, et al. Cartilage T1ρ and T2 relaxation times: Longitudinal reproducibility
and variations using different coils, MR systems and sites. Osteoarthritis and Cartilage
2015 12;23(12):2214-23.
42. Li X, Han ET, Busse RF, Majumdar S. In vivo T(1rho) mapping in cartilage using 3D
magnetization-prepared angle-modulated partitioned k-space spoiled gradient echo
snapshots (3D MAPSS). Magn Reson Med 2008 Feb;59(2):298-307.
43. Mosher TJ, Zhang Z, Reddy R, Boudhar S, Milestone BN, Morrison WB, Kwoh CK,
Eckstein F, Witschey WR, Borthakur A. Knee articular cartilage damage in
osteoarthritis: Analysis of MR image biomarker reproducibility in ACRIN-PA 4001
multicenter trial. Radiology 2011 Mar;258(3):832-42.
44. Witschey WR, Borthakur A, Elliott MA, Fenty M, Sochor MA, Wang C, Reddy R.
T1rho-prepared balanced gradient echo for rapid 3D T1rho MRI. J Magn Reson Imaging
2008 Sep;28(3):744-54.
45. Li X, Pai A, Blumenkrantz G, Carballido-Gamio J, Link T, Ma B, Ries M, Majumdar S.
Spatial distribution and relationship of T1rho and T2 relaxation times in knee cartilage
with osteoarthritis. Magn Reson Med 2009 Jun;61(6):1310-8.
46. Wright GA, Brittain JH, Stainsby JA. Preserving T1 or T2 contrast in magnetization
preparation sequences. Proc Intl Soc Mag Reson Med 1996;4:1474.
47. Wheaton AJ, Borthakur A, Reddy R. Application of the keyhole technique to T1rho
relaxation mapping. Journal of Magnetic Resonance Imaging 2003;18(6):745-9.
48. Feinberg DA, Setsompop K. Ultra-fast MRI of the human brain with simultaneous multi-
slice imaging. Journal of Magnetic Resonance (San Diego, Calif.: 1997) 2013
02/13;229:90-100.
49. Matzat SJ, McWalter EJ, Kogan F, Chen W, Gold GE. T2 relaxation time quantitation
differs between pulse sequences in articular cartilage. J Magn Reson Imaging 2015
Jul;42(1):105-13.
22
50. Lammentausta E, Multanen J, Nieminen MT. Differences in T2 values of knee cartilage
measured with different scanners. Proc Int Soc Mag Reson Med 2009;17:3986.
23
Figure 1. (a) Schematic diagram of adiabatic T1ρ and T2ρ sequences. The preparation block
consisted in a variable train of adiabatic full passage (AFP) pulses (HS1 or HS4), preceding a
FLASH readout. To measure T2ρ, an adiabatic half passage (AHP) was placed prior to the
AFP train and a reverse AHP after the AFP pulses, before the imaging readout. (b) Articular
cartilage ROI segmentation and nomenclature.
24
Figure 2. R1ρ and R2ρ dependence on agarose concentration at different Tp (6, 8, 10 and 12 ms)
for HS1 (a) and HS4 (b). R2 is also shown for reference (red dots).
25
Figure 3. T1ρ maps (a) and T2ρ maps (b) of three representative agarose gel tubes (2.0, 2.25 and
2.75 % w/v).
26
Figure 4. R1ρ and R2ρ versus chondroitin sulfate concentration at Tp 6, 8, 10 and 12 ms for HS1
(a) and HS4 (b). R2 (red circles) is shown for reference.
27
Figure 5. T1ρ,HS1 (a), T1ρ,HS4 (b), T2ρ,HS1 (c) and T2ρ,HS4 (d) relaxation time constant maps of
articular cartilage
28
Figure 6. T1ρ and T2ρ maps and mean relaxation time values in femoral and tibial cartilage of
a representative asymptomatic subject and three patients with radiographic osteoarthritis
(KL=2-4). In the representative subject, mean T1ρ and T2ρ of the asymptomatic volunteer group
are shown.
29
Table 1. Relaxivity (r) as a function of agarose concentration and chondroitin sulfate concentration with 95% confidence interval (C.I.), coefficient of determination (R2) and root mean square
deviation (RMSD). Values affected by artifacts resulting from violation of the adiabatic condition are marked with a gray background.
AGAROSE CHONDROITIN SULFATE
PULSE
TYPE
Tp
[ms]
r
[%-1 s-1] 95% C.I. R2 RMSD
r
[%-1 s-1] 95% C.I. R2 RMSD
r1ρ HS1 3 27.06 (13.86, 40.27) 0.705 18.666
6 2.29 (1.81, 2.78) 0.926 0.689 0.26 (0.09, 0.42) 0.889 0.333
8 1.99 (1.82, 2.16) 0.987 0.243 0.17 (0.12, 0.21) 0.980 0.089
10 2.19 (2.00, 2.37) 0.988 0.260 0.18 (0.14, 0.22) 0.984 0.083
12 2.35 (2.13, 2.58) 0.984 0.319 0.19 (0.15, 0.23) 0.986 0.082
HS4 3 6.66 (4.53, 8.80) 0.847 3.020
6 4.20 (3.83, 4.56) 0.987 0.516 0.25 (0.20, 0.30) 0.989 0.095
8 4.79 (4.37, 5.21) 0.987 0.593 0.28 (0.23, 0.34) 0.989 0.108
10 5.25 (4.73, 5.77) 0.983 0.731 0.30 (0.25, 0.36) 0.990 0.112
12 5.33 (4.85, 5.81) 0.986 0.683 0.31 (0.25, 0.37) 0.990 0.116
r2ρ HS1 3 28.05 (16.82, 39.27) 0.780 15.868
6 5.37 (4.39, 6.36) 0.944 1.393 0.48 (-0.04, 1.00) 0.738 1.042
8 7.81 (6.93, 8.70) 0.978 1.249 0.63 (0.52, 0.74) 0.991 0.214
10 7.08 (6.44, 7.73) 0.986 0.906 0.62 (0.51, 0.73) 0.991 0.219
12 6.61 (5.86, 7.35) 0.978 1.054 0.61 (0.51, 0.72) 0.992 0.207
HS4 3 12.61 (7.58, 17.64) 0.781 7.114
6 6.91 (6.21, 7.62) 0.982 1.002 0.55 (0.48, 0.63) 0.995 0.149
8 6.76 (6.14, 7.38) 0.985 0.876 0.54 (0.47, 0.60) 0.996 0.129
10 6.46 (5.84, 7.08) 0.984 0.881 0.52 (0.46, 0.59) 0.996 0.127
12 6.36 (5.77, 6.95) 0.985 0.830 0.51 (0.43, 0.59) 0.993 0.159
r2 6.71 (6.05, 7.37) 0.983 0.935 0.50 (0.47, 0.53) 0.999 0.053
30
Table 2. In vivo T1ρ measurements (ms) of articular cartilage using different TRs. Values that demonstrated significant TR dependence are marked with a gray background.
PULSE
TYPE ROI
TR = 1.0 S TR = 2.5 S TR = 4.0 S TR = 6.0 S
SBJ #1 SBJ #2 SBJ #1 SBJ #2 SBJ #1 SBJ #2 SBJ #1 SBJ #2
HS1 femur 73.0 75.1 88.3 95.9 90.0 99.0 92.1 104.4
tibia 64.2 75.0 78.2 91.7 79.7 97.1 80.1 94.9
HS4 femur 53.1 51.0 63.1 65.1 67.2 66.0 67.3 70.2
tibia 46.3 51.9 58.9 65.5 59.5 64.9 62.3 65.1
ROI = Region of interest. SBJ = subject.
31
Table 3. In vivo T1ρ and T2ρ measurements (ms) of articular cartilage with different pulse lengths. Values affected by artifacts resulting from violation of the adiabatic condition are marked with
a gray background.
PULSE
TYPE ROI
Tp = 3 ms Tp = 6 ms Tp = 8 ms Tp = 10 ms Tp = 12 ms
SBJ #1 SBJ #2 SBJ #1 SBJ #2 SBJ #1 SBJ #2 SBJ #1 SBJ #2 SBJ #1 SBJ #2
T1ρ HS1 femur 18.1 17.5 69.0 79.9 90.0 99.0 95.6 99.7 93.2 96.9
tibia 18.2 16.5 51.2 74.9 79.7 97.1 83.8 94.8 79.0 91.8
HS4 femur 42.0 43.8 67.2 66.0 64.6 63.4 62.1 61.9 62.8 63.8
tibia 34.6 37.9 59.5 64.9 58.5 60.3 56.7 56.7 54.7 56.6
T2ρ HS1 femur 21.8 17.1 44.2 37.4 55.3 37.9 54.0 40.9 58.4 41.5
tibia 27.2 15.1 49.0 30.4 64.0 38.1 60. 38.9 62.9 38.9
HS4 femur 30.3 26.9 48.4 42.3 50.3 45.2 60.5 46.4 64.1 48.0
tibia 34.4 34.4 54.8 38.9 53.3 35.1 64.5 40.8 67.5 41.1
ROI = Region of interest. SBJ = subject.
.
32
Table 4. Means and standard deviations (SD) of T1ρ, T2ρ and T2 relaxation time values (ms) in superficial and deep articular cartilage of human subjects (N = 10).
COMPARTMENT ROI
T1ρ T2ρ T2
HS1 HS4 HS1 HS4
SUPERFICIAL DEEP SUPERFICIAL DEEP SUPERFICIAL DEEP SUPERFICIAL DEEP SUPERFICIAL DEEP
LATERAL
a Tr 110.6 (11.0) 95.4 (13.0) 78.8 (7.4) 68.5 (10.5) 39.5 (7.0) 31.6 (6.2) 50.3 (4.5) 39.7 (4.4) 60.0 (6.3) 45.1 (6.9)
p Tr 110.8 (11.1) 85.4 (9.1) 76.9 (5.0) 61.6 (8.2) 40.0 (6.2) 33.6 (7.3) 48.0 (7.8) 40.8 (3.9) 53.4 (6.8) 34.8 (11.4)
a CF 116.3 (19.4) 83.3 (11.2) 80.5 (11.2) 59.8 (10.0) 44.9 (9.1) 36.4 (8.1) 51.5 (13.9) 44.1 (9.9) 54.5 (7.4) 35.5 (7.1)
p CF 105.3 (12.6) 84.5 (5.0) 74.7 (7.9) 60.7 (8.0) 40.3 (6.8) 40.7 (9.5) 50.1 (6.0) 47.8 (5.5) 56.5 (7.6) 43.7 (8.6)
a PF 104.1 (11.5) 76.7 (13.6) 71.1 (11.6) 52.9 (7.2) 35.5 (9.3) 38.8 (7.4) 41.0 (5.1) 41.5 (6.7) 53.6 (12.2) 40.7 (7.5)
p PF 87.6 (29.3) 70.3 (30.0) 60.3 (20.7) 45.2 (15.8) 30.2 (11.9) 34.8 (14.7) 35.5 (12.8) 37.3 (13.8) 44.9 (6.3) 35.7 (5.5)
AT 92.5 (12.2) 78.5 (20.8) 66.3 (10.2) 56.2 (8.2) 35.2 (6.5) 31.5 (9.0) 42.9 (8.4) 37.3 (4.7) 49.5 (11.8) 31.0 (4.1)
CT 92.3 (12.3) 64.6 (11.0) 66.4 (9.3) 42.2 (2.6) 31.4 (4.1) 24.7 (7.7) 40.5 (5.8) 26.8 (11.8) 42.0 (5.9) 26.0 (5.0)
PT 97.3 (11.7) 75.0 (13.8) 68.7 (11.8) 52.3 (7.7) 36.1 (8.1) 29.3 (5.9) 45.6 (6.6) 36.9 (4.7) 52.7 (5.2) 37.1 (5.1)
MEDIAL
p Tr 90.1 (32.7) 84.9 (32.5) 65.4 (32.7) 61.5 (29.8) 35.8 (17.7) 38.3 (17..8) 44.6 (21.0) 44.0 (20.4) 62.7 (33.3) 54.5 (29.0)
a CF 105.6 (13.8) 87.7 (11.8) 78.3 (11.0) 64.7 (9.0) 37.1 (6.7) 34.8 (6.4) 47.7 (4.9) 44.9 (7.0) 59.1 (10.3) 37.6 (9.6)
p CF 103.1 (7.9) 82.6 (5.5) 73.3 (5.8) 62.8 (7.6) 37.3 (6.2) 38.5 (8.2) 47.9 (4.6) 48.9 (10.5) 52.3 (6.3) 34.5 (7.5)
a PF 101.3 (11.8) 77.5 (7.1) 71.0 (10.0) 57.2 (7.9) 34.9 (5.8) 35.7 (5.3) 43.5 (4.3) 39.6 (9.4) 54.1 (4.9) 39.7 (9.0)
p PF 92.0 (30.3) 76.3 (26.9) 64.9 (22.8) 54.5 (19.0) 31.5 (11.2) 32.6 (11.3) 39.1 (12.8) 36.9 (12.8) 46.5 (4.7) 35.3 (8.7)
AT 79.6 (35.7) 84.7 (49.6) 55.6 (25.6) 61.7 (35.6) 32.5 (16.4) 34.9 (11.3) 32.2 (15.2) 38.8 (22.4) 44.3 (23.8) 34.4 (18.8)
CT 113.7 (12.3) 71.5 (9.7) 87.7 (9.9) 48.1 (5.5) 42.8 (7.7) 34.9 (20.3) 51.6 (5.0) 33.8 (6.6) 56.3 (6.6) 26.8 (4.5)
PT 105.5 (15.0) 78.0 (11.4) 77.0 (10.9) 54.0 (8.9) 37.1 (9.1) 31.0 (5.4) 46.2 (7.6) 37.5 (5.2) 52.4 (6.9) 34.0 (6.6)
ROI = Region of interest. a = anterior; p = posterior; Tr = troclea; CF = central femur; PF = posterior femur. AT = anterior tibia; CT = central tibia; PT = posterior tibia.