Validation of the Cleaning and Sterilization Process

used for Medical Instruments

Dr. Michelle J. Alfa Ph.D., FCCMMedical Director Clinical Microbiology, DSM,

Winnipeg, Canada

Page 2: Validation of the Cleaning and Sterilization Process used ...novo.sobecc.org.br/programacao/congresso/material_congresso_2_30.pdf · Validation of the Cleaning and Sterilization Process

Overview:

n How to assess adequacy of instrument cleaning

n Process monitors for sterilization- steam

n What are the issues?

Page 3: Validation of the Cleaning and Sterilization Process used ...novo.sobecc.org.br/programacao/congresso/material_congresso_2_30.pdf · Validation of the Cleaning and Sterilization Process

Reprocessing of Medical Devices: Who is responsible for What??

n Manufacturers validate that an instrument can be reliably cleaned and sterilized/disinfected and is therefore re-usable.

n Users verify that cleaning/sterilization equipment is working and that in-hospital cleaning/sterilization protocols are consistently performed.

Page 4: Validation of the Cleaning and Sterilization Process used ...novo.sobecc.org.br/programacao/congresso/material_congresso_2_30.pdf · Validation of the Cleaning and Sterilization Process

Cleaning verification: Users

n Visually inspect the instruments after cleaning

n Monitor the automated washer

Page 5: Validation of the Cleaning and Sterilization Process used ...novo.sobecc.org.br/programacao/congresso/material_congresso_2_30.pdf · Validation of the Cleaning and Sterilization Process

Check Instruments after cleaning

n Visual inspection: good lighting, magnifying conditions- Do routinely for all surgical instruments- Not adequate alone: lumens, hinged areas etc.

n Take samples and test for residuals- protein, hemoglobin, carbohydrate- cutoff for acceptable cleaning??

Page 6: Validation of the Cleaning and Sterilization Process used ...novo.sobecc.org.br/programacao/congresso/material_congresso_2_30.pdf · Validation of the Cleaning and Sterilization Process

Taking Samples from Instruments

n Swab of surface

n Elute from instrument by immersion and sonication in sterile Reverse Osmosis water

Page 7: Validation of the Cleaning and Sterilization Process used ...novo.sobecc.org.br/programacao/congresso/material_congresso_2_30.pdf · Validation of the Cleaning and Sterilization Process

Rapid User Tests: SURFACE TESTING in-hospital cleaning assessment

n ProtTEST Check (unknown LD):Protein (MediSafe, UK)swab device àààà assess

n Protec Swab test (unknown LD):Protein or ATP (Biotrace) swab can be tested. (commercially available)

n Ninhydrin Swab test (2.5 µg/swab):Protein: ISO/CEN method evaluated for users;swab method (deBruin 2002)

Page 8: Validation of the Cleaning and Sterilization Process used ...novo.sobecc.org.br/programacao/congresso/material_congresso_2_30.pdf · Validation of the Cleaning and Sterilization Process

Study: Assess Surgical Instruments Before and After Cleaning to determine level of organic material

n Swab same area (1 cm2) on each instrument

n Elute swab

n Test fluid à protein, hemoglobin, carbohydrate and endotoxin

Quantitative test

Page 9: Validation of the Cleaning and Sterilization Process used ...novo.sobecc.org.br/programacao/congresso/material_congresso_2_30.pdf · Validation of the Cleaning and Sterilization Process

Plastic surgery instruments: patient-used [Avg of 5 patient procedures]

0

100

200

300

400

Ca

rbo

hyd

rate

ug

/cm

2

Before Cleaning After Cleaning

Carbohydrate on instrument

Curved forcep Fine needle driver

Curved Iris scissors

Page 10: Validation of the Cleaning and Sterilization Process used ...novo.sobecc.org.br/programacao/congresso/material_congresso_2_30.pdf · Validation of the Cleaning and Sterilization Process

Plastic surgery instruments: patient-used [Avg of 5 patient procedures]

0

5000

10000

15000

20000

25000

En

do

tox

in E

U/cm

2

Before Cleaning After Cleaning

Endotoxin on instrument

Curved forcep Fine needle driver

Curved Iris scissors

Page 11: Validation of the Cleaning and Sterilization Process used ...novo.sobecc.org.br/programacao/congresso/material_congresso_2_30.pdf · Validation of the Cleaning and Sterilization Process

Conclusions: Testing Instruments

n Level of Carbohydrate and Endotoxinincreased AFTER cleaning

n Level of Protein and Hemoglobin reduced after cleaning

n Likely due to final rinse water contaminated with biofilm [has high carbohydrate and endotoxin]

n Water used for final rinse needs to be monitored periodically.

Page 12: Validation of the Cleaning and Sterilization Process used ...novo.sobecc.org.br/programacao/congresso/material_congresso_2_30.pdf · Validation of the Cleaning and Sterilization Process

What are cutoffs for adequate cleaning??

n OPA method: 0.01 µg/device; (Verjat 1999)

n Ninhydrin method: 2.5 µg/swab; (deBruin 2002)

n Biuret method: 5.5 µg/cm2; (Kruger 1997)

n BCI method: 6.4 ug/cm2; (Alfa 2002)

Stainless Steel Surgical Instruments: Protein

Still controversial – No International Standard

Page 13: Validation of the Cleaning and Sterilization Process used ...novo.sobecc.org.br/programacao/congresso/material_congresso_2_30.pdf · Validation of the Cleaning and Sterilization Process

What Cleaning monitors for Washers are available (Surface vs Lumen)?

n TOSI:

n Browne’s Soil and washer monitor (dye)

Medisafe Lumen check: Laparoscopic device lumen Flexi check: Endoscope lumen

HealthMark USA, Medisafe, UK TOSI; Washer TOSI Lumchek

Page 14: Validation of the Cleaning and Sterilization Process used ...novo.sobecc.org.br/programacao/congresso/material_congresso_2_30.pdf · Validation of the Cleaning and Sterilization Process

Guidance documents: Cleaning of Surgical Instruments

n ISO15883-1: Washer-disinfectors –General requirements, terms and definitions and tests

n ISO15883-5: Test soils and methods for demonstrating cleaning efficacy

n AAMI TIR34:2007: Water for reprocessing of medical devices

Page 15: Validation of the Cleaning and Sterilization Process used ...novo.sobecc.org.br/programacao/congresso/material_congresso_2_30.pdf · Validation of the Cleaning and Sterilization Process

Monitors for Sterilization of Surgical Instruments

n Europe: Parametric release; [Based on standardized loading of sterilizer]

n North America: Biological and Chemical Indicators[Not able to ensure standardized loading of sterilizer]

Page 16: Validation of the Cleaning and Sterilization Process used ...novo.sobecc.org.br/programacao/congresso/material_congresso_2_30.pdf · Validation of the Cleaning and Sterilization Process

Steam Sterilization

n Gravity Displacement

n Dynamic Air-removal Steam sterilizere.g. Pre-vacuum steam sterilizer

Amsco/Steris Getinge

Page 17: Validation of the Cleaning and Sterilization Process used ...novo.sobecc.org.br/programacao/congresso/material_congresso_2_30.pdf · Validation of the Cleaning and Sterilization Process

Pre-Vac Steam Sterilization: Routine Hospital cycles

3 minutes4 minutesUnwrapped nonporous & porous items in a mixed load

3 minutes3 minutesUnwrapped nonporous items (e.g.instruments)

3 minutes4 minutesWrapped instruments

Exposure time at 135oC (275oF)

Exposure time at 132oC (270oF)

Dry Times not indicated, as these vary widely

AAMI ST79 Steam Sterilization Guidance document 2006

Page 18: Validation of the Cleaning and Sterilization Process used ...novo.sobecc.org.br/programacao/congresso/material_congresso_2_30.pdf · Validation of the Cleaning and Sterilization Process

Extended Steam Sterilization Cycle

n Instrument manufacturer recommended steam sterilization cycles that go beyond the times listed for routine hospital cycles

Page 19: Validation of the Cleaning and Sterilization Process used ...novo.sobecc.org.br/programacao/congresso/material_congresso_2_30.pdf · Validation of the Cleaning and Sterilization Process

Two main reasons for manufacturer recommended

Extended-Steam cycles

n Acetabular reamer system:- 18 mins PreVac at 132oC

n Orthopedic Set- 132oC for 40 mins

“Prion” related cycle:European instruments- alternatives may be given my manufacturer

Large mass/density of medical device or instrument set- steam penetration & heating issues

Page 20: Validation of the Cleaning and Sterilization Process used ...novo.sobecc.org.br/programacao/congresso/material_congresso_2_30.pdf · Validation of the Cleaning and Sterilization Process

Medical Device Reprocessing:User’s Responsibilities

n Provide Appropriate Process Monitoring for Steam sterilization cycles used.

n Both CI and BI have a role in process monitoring if parametric release is not used.

Page 21: Validation of the Cleaning and Sterilization Process used ...novo.sobecc.org.br/programacao/congresso/material_congresso_2_30.pdf · Validation of the Cleaning and Sterilization Process

Monitoring of Standard Pre-vac Steam sterilization cycles

n Biological Indicator (BI)- run 1/day (minimum 1/week) for every

steam cycle used that day- required for every load with implant(s)quarantine implant until results of BI known

n Chemical Indicator (CI)- Class I to VI (Canada)- external,internal and in PCD

BI: only used in Process Control Device (PCD)CI: used in PCD or in/on actual instrument package

Page 22: Validation of the Cleaning and Sterilization Process used ...novo.sobecc.org.br/programacao/congresso/material_congresso_2_30.pdf · Validation of the Cleaning and Sterilization Process

BI: Monitor Steam Cycle

n Spores

n Spores in self-contained BI

n Spores in BI in Process Challenge Device (PCD)

BI in PCD used for routine monitoring of steam and ETO cycles in Healthcare

Page 23: Validation of the Cleaning and Sterilization Process used ...novo.sobecc.org.br/programacao/congresso/material_congresso_2_30.pdf · Validation of the Cleaning and Sterilization Process

Chemical Indicators:

n Class 1: Process Indicators (changes color when exposed to steam)- differentiate processed vs unprocessed products. - use on the outside of ALL packages sterilized (e.g. autoclave tape).

n Class 2: Bowie-Dick- ONLY for pre-vacuum sterilizers to show adequate air removal

n Class 3: Single Parameter Indicator (e.g. temp, or steam, or ETO)- monitors single variable- place inside instrument package or in Process Challenge Device

n Class 4: Two or more critical variables (e.g. temp & sterilant)- monitors multiple variables- place inside instrument package or in Process Challenge Device

Page 24: Validation of the Cleaning and Sterilization Process used ...novo.sobecc.org.br/programacao/congresso/material_congresso_2_30.pdf · Validation of the Cleaning and Sterilization Process

Chemical Indicators:

n Class 5 & 6: “Integrators”: ALL critical variables for cycle(e.g. correct temp, & sterilant for specified period of time)

- tightly controlled must achieve success for ALL variables evaluate

- place inside instrument package or in Process Challenge Device

- Loads can be released based on these CIs(except implants which still need result of BI prior to release)

- Class 5 correlate with BI characteristics, - Class 6 “emulating” indicator correlate with stated values not necessarily with BI characteristics

Page 25: Validation of the Cleaning and Sterilization Process used ...novo.sobecc.org.br/programacao/congresso/material_congresso_2_30.pdf · Validation of the Cleaning and Sterilization Process

Why are BIs needed when CIsare more tightly controlled??

n Can chemicals or enzymes repair damage to regain function?

n Can microorganisms repair damage to regain function (e.g. Potential for infection)

NO

YES

Page 26: Validation of the Cleaning and Sterilization Process used ...novo.sobecc.org.br/programacao/congresso/material_congresso_2_30.pdf · Validation of the Cleaning and Sterilization Process

Extended Cycle Monitoring

n INAPPROPRIATE to use a BI (OR CI) in a test pack that has been validated for a standard hospital cycle (e.g. 4 min @ 135oC) to monitor an extended cycle (e.g. 18 min @ 132oc).

A self-contained BI in a “Test Pack” (i.e. PCD) is validated for a specific type of sterilization method and for a specific set of cycle parameters

Page 27: Validation of the Cleaning and Sterilization Process used ...novo.sobecc.org.br/programacao/congresso/material_congresso_2_30.pdf · Validation of the Cleaning and Sterilization Process

What can go wrong when manufacturer recommends extended steam cycle?

n Routine BI & CI in extended cycle:- not a proper challenge to ensure cycle parameters are adequate

n Standard steam cycle using appropriate routine BI:- inadequate sterilizationà instruments may not be sterile despite BI that passes

Page 28: Validation of the Cleaning and Sterilization Process used ...novo.sobecc.org.br/programacao/congresso/material_congresso_2_30.pdf · Validation of the Cleaning and Sterilization Process

What can Users do??

n Ask medical device manufacturer if extended cycles have been validated

n Limited user-testing to ensure steam penetration: - test BEFORE set is used for first time- place 3 BIs inside each layer of set- if any BIs fail; break set down into smaller tray sets and re-test

Page 29: Validation of the Cleaning and Sterilization Process used ...novo.sobecc.org.br/programacao/congresso/material_congresso_2_30.pdf · Validation of the Cleaning and Sterilization Process

Steam Sterilization: Multi-layer Surgical Sets

n If failure of any BIs: report to medical device manufacturer

n Break down set to smaller tray sets and re-test

Page 30: Validation of the Cleaning and Sterilization Process used ...novo.sobecc.org.br/programacao/congresso/material_congresso_2_30.pdf · Validation of the Cleaning and Sterilization Process

Issues that need Site specific attention:

n Ensure proper BIs and CIs used for steam cycles

n Ensure Device manufacturer cycle parameters (e.g. extended steam cycles) are followed unless written instructions otherwise

n Consolidate extended cycles to one sterilizer unit and ensure training and competency of staff

n May verify steam penetration using BI in tray-set as PCD

Page 31: Validation of the Cleaning and Sterilization Process used ...novo.sobecc.org.br/programacao/congresso/material_congresso_2_30.pdf · Validation of the Cleaning and Sterilization Process

What is being done to address this issue?

n Manufacturers working with AMMI and CSA to harmonize to ONLY two extended cycles (10 and 20 minutes suggested).

n All instruments and sets will be validated for existing routine steam cycles or for one of the two extended cycles

n Manufacturers to develop appropriate PCD for BI/CI monitoring of the two extended cycles(Note; there are CIs cleared for some extended steam cycles)

Page 32: Validation of the Cleaning and Sterilization Process used ...novo.sobecc.org.br/programacao/congresso/material_congresso_2_30.pdf · Validation of the Cleaning and Sterilization Process

SUMMARY

n BIs or CIs recommended for routine cycle monitoring when parametric release not used.

n Do NOT process devices/sets in routine cycles if extended cycles recommended by manufacturer

n Heavy, complex instrument sets à suggest limited pre-testing using instrument tray set as PCD until more appropriate ones available

Page 33: Validation of the Cleaning and Sterilization Process used ...novo.sobecc.org.br/programacao/congresso/material_congresso_2_30.pdf · Validation of the Cleaning and Sterilization Process

References:

n Alfa MJ et al A Survey of reprocessing methods, residual viable bioburden and soil levels in patient-ready ERCP duodenoscopes used in Canadian centers. Infect Control Hosp Epidemiol 2002;23:198-206.

n Moses FM, J Lee. Surveillance cultures to mnitor quality of gastrointestinal endoscope reprocessing Am J Gastroenterology 2003;98:77-81.

n Obee PC et al Real-time monitoring in managing the decontamination of flexible gastrointestinal endoscopes Am J Infect Control 2005;33:202-6.

n Pineau L et al Endoscope drying/storage cabinet: interest and efficacy. J Hosp Infect 2007;doi:10.1016/j.jhin.2007.10.012

n Sciortina CV, E L Xia, A Mozee Assessment of a novel approach to evaluate theoutcome of endoscope reprocessing. Infect Control and Hosp Epidemiol2004;25:284-290

n Orzechowski, ACP de Bruijn, C. Wassenaar Validation of a cleaning test for flexible endoscopes. Zentr Steril 2003;11:165-178

n Mehta AC et al American College of Chest physicians and American Association for Bronchology consensus Statement: Prevention of Flexible Bronchoscopy-associated infection. 2005;128;1742-1755

n Alfa M, C Landers, S Hadfield, C Hancock, L Jakeman, L Kingsbury,k P McCornick, D Trudeau User Alert: Problems with Process Monitors for Extended Steam Sterilization Cycles. Can J Infect Control Sept 2006