vanguard “a” operator’s manual - prosites, inc.c1-preview.prosites.com/106982/wy/docs/porter...

13
VANGUARD “A” OPERATOR’S MANUAL FM-952 Rev. E 7/07 PORTER Parker Hannifin Corporation Porter Instrument Division 245 Township Line Rd. P.O. Box 907 Hatfield, PA 19440-0907 USA (215) 723-4000 / fax (215) 723-5106

Upload: hoangnga

Post on 10-Mar-2018

220 views

Category:

Documents


2 download

TRANSCRIPT

Page 1: VANGUARD “A” OPERATOR’S MANUAL - ProSites, Inc.c1-preview.prosites.com/106982/wy/docs/Porter Vanguard Alarm User... · Trouble Shooting Information ... The alarm should indicate

VANGUARD “A”

OPERATOR’S MANUAL

FM-952 Rev. E 7/07

PORTER

Parker Hannifin Corporation Porter Instrument Division 245 Township Line Rd. P.O. Box 907 Hatfield, PA 19440-0907 USA (215) 723-4000 / fax (215) 723-5106

Page 2: VANGUARD “A” OPERATOR’S MANUAL - ProSites, Inc.c1-preview.prosites.com/106982/wy/docs/Porter Vanguard Alarm User... · Trouble Shooting Information ... The alarm should indicate

2

Please read this manual carefully before operating the Vanguard Manifold System. Remove product from package and inspect for damage. If there is any damage, do not use (Contact your dealer.). Complete Warranty Card and return to Porter. To assure safe operation and conformation to local fire codes, all Porter Instrument Company Manifold Systems are designed to be used with sedation delivery systems mounted inside walls and they meet or exceed the guidelines established by the National Fire Protection Association for Nonflammable Medical Gas Systems, NFPA 99. Copies of NFPA 99 or portions thereof may be obtained by writing to: National Fire Protection Association Batterymarch Park Quincy, MA 02269-9904 Or call: 1-800-344-3555 These warnings and precautions are to help you to understand how to safely operate or troubleshoot the Porter Vanguard A Manifold System. A WARNING alerts you to a possible hazard to people. A CAUTION alerts you to the possibility of equipment damage.

WARNING Dental workers are exposed to N2O during administration of N2O / O2 conscious sedation analgesia. NIOSH has recommended that exposures should be minimized. Contact NIOSH (1-800-35-NIOSH) to receive NIOSH Publications on Control of Nitrous Oxide in Dental Operatories. Exposure can be minimized by effective controls. National Institute for Occupational Safety and Health (NIOSH) publications state that controls, including System Maintenance, Ventilation and Work Practices can effectively reduce N2O concentrations in dental operations. Your Porter Scavenger System is an important part of the system of controls.

CAUTIONS Static Damage Hazard: Computer and other electronic components can be permanently damaged by static electric discharge. Use caution when working with the molded manifold cover off, especially in the area of the RESET1 button. Use no Oil: Never use oils, greases, organic lubricants or any combustible materials on or near this product. Power Outage: During any power outage, remember to turn OFF the flowmeter and manually turn OFF the tank valves. When the power is restored, the Vanguard will come back to its status prior to the power outage. If gas was flowing when the power went out, gas will be flowing when the power is restored.

Page 3: VANGUARD “A” OPERATOR’S MANUAL - ProSites, Inc.c1-preview.prosites.com/106982/wy/docs/Porter Vanguard Alarm User... · Trouble Shooting Information ... The alarm should indicate

3

Table of Contents

Warnings and Precautions ...................................................................................................2 4222 Vanguard Manifold System Diagram ..........................................................................4 Vanguard “A” Alarm Diagram ...............................................................................................5 4222 Vanguard Manifold Schematic ....................................................................................5 Important Features to Know About the 4222 Manifold System ........................................... 6 – 7

How System Operates .............................................................................................. 6

Failsafe in Manifold ....................................................................................................6

Pressure Tested at 150 PSIG ....................................................................................7

Wiring in Manifold System .........................................................................................7

Shut-off Operation .....................................................................................................7

Silence Button ...........................................................................................................7

Normal Operation Indications for Alarm Station .........................................................7

Manifold System Testing and Start-up .................................................................................8 150 PSIG Nitrogen Pressure Test .......................................................................................9 Function Testing Changeover Manifold ...............................................................................9 Steps to Check Tank Switching Process .............................................................................10 Trouble Shooting Information ..............................................................................................11 Warranty, Orders and Return Policies .................................................................................12

Page 4: VANGUARD “A” OPERATOR’S MANUAL - ProSites, Inc.c1-preview.prosites.com/106982/wy/docs/Porter Vanguard Alarm User... · Trouble Shooting Information ... The alarm should indicate

4

4222

Page 5: VANGUARD “A” OPERATOR’S MANUAL - ProSites, Inc.c1-preview.prosites.com/106982/wy/docs/Porter Vanguard Alarm User... · Trouble Shooting Information ... The alarm should indicate

5

VANGUARD A ALARMStock Number 6251A

(WALL) (DESK)Stock Number 6201A

BUTTONGAS/OFF

BUTTONALARM SILENCETANK STATUS

INDICATORS(#1 & #2)

HIGH

NORMAL

LOW

LINE PRESSURE INDICATORS

NORM NORM

DI

GAS / OFF

SILENCE

1 2

LOWE

1 2

LOW

PORTER

T

S

OX

R

UO

HIGH

NI

AVANGUARD

YG

NE

OX

HIGH

LINE PRESSURE

4222 VANGUARD MANIFOLD

2N 0

- SYSTEM RESET 1

* TO BE REMOVEDAFTER 150 PSIG 24 HR.NITROGEN TEST

(RIGHT BUTTON)

LOCATEDUNDER COVER

- TANK RESET(LEFT BUTTON)

2N O DISS INLETFITTINGS

N 02

4 CONDUCTORROUND CABLE

INDICATORSTANK STATUS

N 02

1

432

ACKNOWLEDGEBUTTON

INDICATORS

GAS SERVICE

RED

BLACKWHITEGREEN

SOLENOIDS02

2SWITCH

(N O / O )

HI/LO PRESSURE

PRESSURE RELIEF

LINE PRESSUREGAUGE

O DISS INLET2FITTINGS VALVE (2)

20

2

* RED PLUGS (4)

(COPPER TUBING)

PUSH BUTTON

42

1

3

02

GAS PIPE LINE

Page 6: VANGUARD “A” OPERATOR’S MANUAL - ProSites, Inc.c1-preview.prosites.com/106982/wy/docs/Porter Vanguard Alarm User... · Trouble Shooting Information ... The alarm should indicate

6

PORTER 4222

AUTOMATIC CHANGEOVER MANIFOLD SYSTEM

The system that monitors and switches to your reserve tank when your primary tank is empty, all of which is accomplished automatically with only the gas serviceman maintaining the gas supply and tank status light entry condition by pushing the gas service acknowledge button located on the manifold. IMPORTANT FEATURES TO KNOW ABOUT THE PORTER AUTOMATIC CHANGEOVER MANIFOLD HOW THE SYSTEM OPERATES Make sure oxygen and nitrous oxide tanks are hooked-up and turned ON. Regulators must be set at 50 PSIG regulated pressure to insure normal operation. At alarm station PUSH (GAS / OFF) button to start up system. The alarm should indicate normal line pressure condition lights and the tank status light should lock on tank position 1 or 2 after a few seconds.

NOTE: It is important to know that on initial start-up of your manifold system, when there is no pressure in the lines, it may require the service acknowledge button on the manifold to be depressed several times to allow the manifold to switch back and forth between tanks to initially fill piping system to the 50 PSIG level.

If pressure in the system is at 40 PSIG or below, the alarm will indicate a low-pressure light and audible tone. If a reserve tank is available, the manifold will switch over automatically after 10-20 seconds of low status to the reserve tank. The low-pressure light and audible tone will change to the normal light when the pressure reaches the safety zone. Normal line pressure is 50 PSIG (gauge observable on the manifold). The tank that is now empty and off-line will indicate a flashing light signal at the alarm and manifold. This condition will remain until the service acknowledge button on manifold is depressed to turn OFF flashing signal. This should be done only when the gas supply is replenished by the gas service man. System must be ON to reset tank at service acknowledge button. Now your tank status lights should indicate one light ON and one light OFF indicating primary tank is in use and reserve tank is available for back-up. FAILSAFE IN MANIFOLD The 4222 manifold has a built-in failsafe feature. For your patients’ safety, the manifold will not deliver nitrous oxide through to the piping system if oxygen supply pressure is depleted or turned off.

Page 7: VANGUARD “A” OPERATOR’S MANUAL - ProSites, Inc.c1-preview.prosites.com/106982/wy/docs/Porter Vanguard Alarm User... · Trouble Shooting Information ... The alarm should indicate

7

PRESSURE TESTED AT 150 PSIG The 4222 manifold is made to be tested for leaks at 150 PSIG for 24 hours with dry nitrogen. After completion of this test, remove red test plugs on top of each solenoid at the manifold for proper operation of system. WIRING IN MANIFOLD SYSTEM The alarm connection to the system is accomplished by a six foot length modular male connection cable that connects to the signal interconnect assembly. From this point through to the manifold, all connections for transferring DC power and signal is carried through the four-conductor alarm cable. The colored wires in the alarm cable must terminate to the same terminal throughout the system.

• Terminal 1 carries DC power to system. • Terminals 2 and 3 carry signal for changeover. • Terminal 4 is a power ground. • Voltage reading (12 Vdc ± 2 Vdc) should be taken with voltmeter negative ( - ) on #4 ground,

voltmeter positive ( + ) on #1. • Voltage at terminal #2 is nominally +1.5 Vdc and at terminal #3 is nominally +0.4 Vdc.

SHUT-OFF OPERATON Depress the OFF button. Alarm beeps 3 times and all lights flash before shutting down. SILENCE BUTTON The silence button is used to turn OFF the audible alarm condition. NORMAL OPERATION INDICATIONS FOR ALARM STATION 1. The yellow LED flashing and one short beep indicates that the oxygen or nitrous oxide line

pressure(s) is below 40 PSIG. 2. The red LED flashing and two short beeps indicate that the oxygen or nitrous oxide line

pressure(s) is above 60 PSIG. 3. Four beeps indicate that the system has a hardware problem, usually signifying that a pressure

switch has failed. 4. Five beeps indicate that the alarm has lost communications with the manifold and that the alarm

is no longer indicating the status of the oxygen and nitrous oxide tanks. 5. The green LED light indicates that the line pressure is at normal pressure.

Page 8: VANGUARD “A” OPERATOR’S MANUAL - ProSites, Inc.c1-preview.prosites.com/106982/wy/docs/Porter Vanguard Alarm User... · Trouble Shooting Information ... The alarm should indicate

8

Page 9: VANGUARD “A” OPERATOR’S MANUAL - ProSites, Inc.c1-preview.prosites.com/106982/wy/docs/Porter Vanguard Alarm User... · Trouble Shooting Information ... The alarm should indicate

9

150 PSIG NITROGEN PRESSURE TEST Using the Porter Nitrogen Test Fixture Follow These Steps: 1. Connect regulator to nitrogen tank. 2. Turn selector valve handle to closed position (no flow or pressure coming out ends if tank would

be ON). 3. Turn tank on slowly and set line pressure gauge at 150 PSIG. 4. Insert male Quick Connect end of test hose (O2 or N2O) into pressure female Quick Connect. 5. Connect O2 or N2O manifold diameter indexed safety system (DISS) hose, which is sent with

manifold, to test hose male DISS fitting end and attach to appropriate inlet DISS connection on manifold.

6. Turn selector valve to pressure position, SLOWLY. 7. Turn ON Vanguard alarm. 8. Read gauge on manifold. Setting should be 150 PSIG. 9. Disconnect test hose from pressure Quick Connect, then DISS manifold hose from inlet fitting on

manifold. 10. Turn OFF Vanguard Alarm to turn off high indicator lights / LEDs. Pressure should hold. 11. Examine for leaks for 24 hours. (Reference NFPA 99 Requirements.) 12. After completion of 150 PSIG leak test, remove red plugs on top of solenoid and remove the

plugs on the pressure relief valves. FUNCTION TESTING CHANGEOVER MANIFOLD Initial Start-up and Testing Procedure 1. Make sure oxygen and nitrous oxide tanks are hooked up and turned ON. Regulators must be

set at 50 PSIG regulated pressure to insure normal operation. 2. At Alarm Station – PUSH – (GAS / OFF) button to start up system. 3. The alarm should indicate normal line pressure condition lights and the tank status light

should lock on tank position 1 or 2.

NOTE: It is important to know that on initial start-up of your manifold system, when there is no pressure in the lines, it may require the service acknowledge button on the manifold to be depressed several times to allow the manifold to switch back and forth between tanks to initially fill piping system to the 50 PSIG level.

4. Now that the piping system is filled and the manifold has conducted its initial start-up function, it must be checked for its automatic changeover capability.

Page 10: VANGUARD “A” OPERATOR’S MANUAL - ProSites, Inc.c1-preview.prosites.com/106982/wy/docs/Porter Vanguard Alarm User... · Trouble Shooting Information ... The alarm should indicate

10

5. Steps to Check Tank Switching Process

5.1. Turn down set pressure on primary tank regulator to 35 PSIG (note how much of a turn is made on regulator adjusting stem) while bleeding down system pressure through dental meter at 5 L/min rate of flow.

5.2. At 40 PSIG the manifold will announce an audible and visual alarm condition for 10 to

20 seconds and switch to reserve tank. The manifold will shut the primary solenoid and open the reserve solenoid, which will deliver gas at 50 PSIG and signal the normal light. After changeover has occurred, reset regulator on Tank #1 to original condition position 50 PSIG.

5.3. Manifold display panel and alarm will show a blinking light at the primary tank

location.

5.4. Push service acknowledge button on manifold to turn OFF blinking light at Tank 1 location.

5.5. Reverse the procedure to switch back to Tank #1. Turn down pressure to 35 PSIG on

reserve regulator while continuing to flow gas at 5 L/min downstream at a dental meter. Manifold will again announce alarm condition and switch back to primary tank. Reset Tank #2 regulator back to 50 PSIG setting.

5.6. Push service acknowledge button on manifold to turn OFF blinking light at Tank 2

location.

5.7. Repeat this process for both oxygen and nitrous oxide.

Page 11: VANGUARD “A” OPERATOR’S MANUAL - ProSites, Inc.c1-preview.prosites.com/106982/wy/docs/Porter Vanguard Alarm User... · Trouble Shooting Information ... The alarm should indicate

11

TROUBLE SHOOTING Symptom Possible Cause / Remedy

1. Tanks will not changeover. Defective pressure switch will not create a low-pressure signal to produce changeover response.

2. Tanks change over before tanks are empty.

This may result from piping system having a leak and bleeding down pressure over night. Tank pressure will remain the same due to the manifold and alarm system being turned OFF and holding cylinder pressure stable. Inspect line pressure in the morning for setting to remain between 45 to 55 PSIG.

3. System has no lights. Remove cover on Power/Signal Junction Assembly (Refer to page 4) and check if Red LED indicator is on to show power supply is working. Defective OFF / Button, or electronics. Contact Porter for further assistance.

4. Tanks change over, but pressure does not come up to 50 PSIG.

Check regulator for correct pressure setting. DISS inlet fitting has a check valve stuck shut restricting pressure. Depress or hold service acknowledge button to fill system to bring up line pressure. Defective solenoid prevents pneumatic circuit from opening for delivering line pressure.

5. The yellow and red LEDs are flashing and the unit is beeping four beeps.

The pressure switch is defective on the gas for which the yellow and red LEDs are flashing. Replace the pressure switch.

CAUTION Static Damage Hazard: Computer and other electronic components can be permanently damaged by static electric discharge. Use caution when working with the molded manifold cover off, especially in the area of the RESET1 button.

6. Lights Flashing; locked into a random flashing pattern.

Caution, carefully depress Reset1 button only on manifold. (Refer “Caution” above and page 6.) Defective electronics or timing problem. Contact Porter for further assistance.

7. Alarm will not silence. Caution, carefully depress Reset1 button only on manifold. (Refer to “Caution” above and page 6.) Defective electronics or timing problem. Contact Porter for further assistance.

8. Cannot turn OFF system Caution, carefully depress Reset1 button only on manifold. (Refer to “Caution” above and page 6.)

Page 12: VANGUARD “A” OPERATOR’S MANUAL - ProSites, Inc.c1-preview.prosites.com/106982/wy/docs/Porter Vanguard Alarm User... · Trouble Shooting Information ... The alarm should indicate

12

WARRANTY

THIS WARRANTY IS GIVEN IN PLACE OF ALL OTHER WARRANTIES, EXPRESS OR IMPLIED, OF MERCHANTABILITY, FITNESS FOR A PARTICULAR PURPOSE OR OTHERWISE. Under no circumstances shall Parker Hannifin Corporation be liable for incidental or consequential damages as those terms are defined in the uniform commercial code. Parker Hannifin Corporation, Porter Instrument Division warrants that each product or part shall be free from defects in workmanship and materials, under normal use and with appropriate maintenance, for one (1) year from the date of delivery to customer unless otherwise specified in writing. All rubber and plastic parts and accessories are warranted under the same conditions for a period of ninety (90) days from date of purchase. No statement or claim about the product by any employee, agent, representative, or dealer of Parker Hannifin Corporation shall constitute a warranty by Parker Hannifin Corporation or give to rise to any liability or obligation of Parker Hannifin Corporation. Parker Hannifin Corporation shall not be liable for any damage, injury or loss arising out of the use of the product, whether as a result of a defect in the product or otherwise, if, prior to such damage, injury or loss, the product was (1) damaged or misused; (2) repaired, altered or modified by persons other than Parker Hannifin Corporation; (3) not installed in strict compliance with applicable codes and ordinances; or (4) not installed by an authorized Parker Hannifin Corporation dealer. Parker Hannifin Corporation's obligation for breach of this warranty, or for negligence or otherwise, shall be strictly and exclusively limited to the repair or replacement of the product or part. This warranty shall be void on any product on which the serial number has been altered, defaced or removed. ORDERS All orders are to be made through authorized Parker Hannifin Corporation distributors. All billing will be done through said distributors. Direct orders will be handled through the authorized local dealer as determined by Parker Hannifin Corporation. RETURNS No returns will be accepted unless authorized in writing by Porter Instrument Division, and accompanied by a properly completed return goods authorization. All returns are subject to a re-stocking and possible rework charges to be determined by Porter Instrument Division. Policies subject to change without notice.

Page 13: VANGUARD “A” OPERATOR’S MANUAL - ProSites, Inc.c1-preview.prosites.com/106982/wy/docs/Porter Vanguard Alarm User... · Trouble Shooting Information ... The alarm should indicate

DATE: 01/15/02 ECN: 02-0045 DR: ALE CH: BTT AP: JPB Page 13 of 13 NUMBER: 0110010E REV.

Vanguard “A” Operator’s Manual FM-952 E

REVISION HISTORY

REV. DATE ECN # DESCRIPTION OF CHANGES DR CH APPR 0 01/15/02 02-0045 New document release. ALE BTT JPB

A 01/20/04 04-0033 p. 4 – B-1805-002 was B-1805-001; p. 10 Quick was Quik (3 places)

ALE KKS ALE

B 9/03/04 04-0769 p. 8 Voltage at #2 nom +1.5 and #1 nom

at +0.4 Vdc was Voltage at #1 is 12 Vdc and #2 is 1.5 Vdc and #3 is 0.5 Vdc. “short” was “quick” (4 places); p. 10 Step 11 remove the plug on the pressure relief valve was bleed system; p. 11 remove note on 5.1 Do not shut off valve to cylinder will drain piping system and not simulate changeover; p. 11 step 5.2 add 50 PSIG; p. 12 Step 4 re-arrange order of possible causes.

ALE WHK ALE

C 6/13/05 05-0374 Page 6: Showed correct positions of reset

buttons. Page 8: Under Wiring In Manifold Systems section “#3 nominally +0.4 Vdc” was “#1 nominally +0.4 Vdc” Page 10 under 150 PSIG Nitrogen Pressure Test section: added number 10.

MAM WHK ALE

D 4/6/06 06-0188 Updated warranty statement on page 13. MAM WHK TGT

E 7/2/07 07-0403 Table of Contents: removed “CXA” &”MXVA”. Page 4: removed “CXA” & “MXVA”. Combined pages 5 & 6, renumbered pages. Page 6: removed “CXA/DXA”, added “after a few seconds” & “10 to”. Page 7: “3 times” was “1 time”, notes 3 & 4; removed “short” Page 10, section 5.2: added “10 to”. Page 11, symptom 5: removed “short.

MAM