BCM350N Operator’s Manual 1
P/N : BCM350N- EN-OPM Bistos Co.,Ltd. R01(2013.11.13)
Table Contents Pr ior to beginn ing ………………………………………………………… 2
BCM350N Minimum system requirements ……………………………………… 3
Precautions on install and use BCM350N ………………………………………… 3
Chpater 1 General ……………………………………………………… 4
1.1 Introduction of BCM350N ……………………………………………… 4
1.2 BCM350N installation …………………………………………………… 6
Chpater 2 BCM350N Description ………………………………………… 10
2.1 Main window …………………………………………………………… 10
2.2 Patient basic registration window ……………………………… 13
2.3 Doctor registration widow ……………………………………………… 15
2.4 NST Review window …………………………………………… 15
2.5 Bed view setting window …………………………………………… 16
2.6 Language setup window …………………………………………… 16
2.7 Setting(Configuration) window …………………………………………… 17
2.8 C lose ……………………………………………………………… 17
Chpater 3 How to use BCM350N ………………………………………… 18
3.1 How to setup language ………………………………………… 18
3.2 How to setup communication port ………………………………… 19
3.3 How to register patient basic information ………………………… 20
3.4 How to assign a patient to the relevant bed ………………… 22
3.5 How to start diagnosis …………………………………………… 24
3.6 How to stop diagnosis ………………………………………… 25
3.7 How to reset bed information …………………………………… 27
3.8 How to setup bed view ……………………………………………… 27
3.9 How to listen fetal monitor sound ………………………………………… 29
3.10 How to control fetal monitor ……………………………………………… 29
3.11 How to review existing diagnosis ………………………………………… 30
3.12 How to save to image and print ………………………………………… 33
Chpater 4 BCM350N Uninstallation ……………………………………… 38
Chpater 5 Specification ………………………………………………… 39
Appendix 1. Wired communication system ………………………………………… 40
Appendix 2. Wireless communication system ……………………………………… 42
BCM350N Operator’s Manual 2
P/N : BCM350N- EN-OPM Bistos Co.,Ltd. R01(2013.11.13)
Prior to beginning
In this document, all information for correct using and operation are
included.
If you need a help for instruction or questions, please contact Bistos Co.,
Ltd.
Bistos Co.,Ltd. and the authorized agency do not take responsibility for
the harmful cases that are happened by user’s obvious careless or
incorrect uses.
In this document, the program indicates BCM350N. And the company and
we indicate Bistos Co.,Ltd.
May 2013
Rev 01
Copyright © 2013 Bistos Co., Ltd. All rights reserved.
Information and descriptions contained in this manual are the property of
Bistos Co.,Ltd. and may not be copied, reproduced, disseminated, or
distributed without express written permission from Bistos Co.,Ltd. The
materials in this manual can contain typographic errors and be subject to
change without notice.
Printed in Korea
BCM350N Operator’s Manual 3
P/N : BCM350N- EN-OPM Bistos Co.,Ltd. R01(2013.11.13)
BCM350N Minimum system requirements
To install BCM350 to PC, minimum system requirements are as follow.
1. CPU clock speed : 2.3 GHz
2. Memory (RAM): 4 Giga Bytes
3. Display resolution : 1024 * 768 Graphic LCD
4. Operating system : Windows XP or Window 7
5. Sound card and armored speaker
6. Ethernet adaptor or RS-232C port
Precautions on install and use BCM350N
1. When install BCM350N, a user must follow the regular install procedure. If
not, it is possible that BCM350N does not work properly.
2. Before use BCM350N, please read this manual closely. To operate some
functions what user wants, please read and follow this manual. If not, it is
possible that BCM350N does not work properly.
3. If you need a help for instruction or questions during installing or using
BCM350N, please contact Bistos Co., Ltd. or the authorized agency. Bistos
Co.,Ltd. and the authorized agency do not take responsibility for the harmful
cases that are happened by using BCM350N with problem.
BCM350N Operator’s Manual 4
P/N : BCM350N- EN-OPM Bistos Co.,Ltd. R01(2013.11.13)
Chapter 1 General
1.1 Instruction of BCM350N
BCM350N is an integrated program of NST(non-stress test) central monitoring
system and networking system. That is, the networking function is added to
existing NST central monitoring system. So it is possible to access to the Server
PC through the Local Area Network (LAN) and so multiple monitoring also is
possible at the same time. The concept diagram is shown as below.
<BCM350N concept diagram>
The maximum number of accessible clients at the same time is limited 8 to
reduce the server overflow.
Server mode and Client mode is selectable in [Configuration] window in the
program. In the client mode operation, the server IP address should be
registered and the relevant directory (DB and Date) of the server must be
shared to enable to remote access.
BCM350N Operator’s Manual 5
P/N : BCM350N- EN-OPM Bistos Co.,Ltd. R01(2013.11.13)
The useable devices in this system are BT-300, BT-330, BT-350 and Biocare
FM-1 which are fetal monitors manufactured by Bistos Co.,Ltd.
Using PC serial port, it is possible to connect from 4 to 32 fetal monitors to this
system. Default is 8 devices in upper of [Configuration] window.
Using System Base corporation’s multi-port product named ‘Multi-8’, it is
possible to comprise wire communication (refer to Appendix 1).
Using Sena Technology corporation’s Bluetooth module named ‘MSP-1000’
and ‘Parani-SD1000’, it is possible to comprise wireless communication (refer
to Appendix 2). To comprise wireless communication, it needs to create a
virtual port in PC. To create a virtual port, it needs to install extra program. The
install method of this extra program is described in Appendix 2.
Appendix 1. Wire communication system
Appendix 2. Wireless communication system
Please run BCM350N after check out there is no problem in wire or wireless
communication system.
To run the program, select as below;
‘C:\programfiles\bcm350n’ or ‘C:\programfiles(x86)\bcm350n’
When the program installation is finished, the ‘Sort cut icon’ is created in
wallpaper. If the ‘Short cut icon’ is not created, we recommend that create
‘Short cut icon’ manually and use that icon.
BCM350N Operator’s Manual 6
P/N : BCM350N- EN-OPM Bistos Co.,Ltd. R01(2013.11.13)
1.2 BCM350N installation
1. Run BCM-350N [Setup.exe].
2. Select the language for installation. (Korean / English)
BCM350N Operator’s Manual 7
P/N : BCM350N- EN-OPM Bistos Co.,Ltd. R01(2013.11.13)
3. Installation Wizard will come out. Click [Next] button.
4. Specify the path of the installed directory for BCM350N by clicking
[Browse…] button. Select a destination folder and click [Install] button. We
recommend that using the default path.
BCM350N Operator’s Manual 8
P/N : BCM350N- EN-OPM Bistos Co.,Ltd. R01(2013.11.13)
5. Please wait while the software is being installed.
If ‘Microsoft. Net Framework 3.5’ is not installed in your desktop, it is installed
automatically during the installation of BCM350N. In this case, it may take few
minutes.
6. Click [Finish] button to complete the installation. (Check the small box if
you want to start the BCM350N immediately.)
BCM350N Operator’s Manual 9
P/N : BCM350N- EN-OPM Bistos Co.,Ltd. R01(2013.11.13)
7. ‘BCM350N Short cut icon’ is created in desktop.
Run BCM350N and the initial window will be shown up.
8. The initial window is set with English. If you want to change language,
please click [Language] button on the toolbar and change language what
you want.
9. Initial network of BCM350N is ‘Server’. Change network mode to ‘Client’ as
follow.
[Setting] button -> [Configuration] window -> Check ‘Run as client’ of
Communication region -> Application restart required. -> Input Server IP
And then run BCM350 again. You can find the client mode main window.
BCM350N Operator’s Manual 10
P/N : BCM350N- EN-OPM Bistos Co.,Ltd. R01(2013.11.13)
Chapter 2 BCM350N Description
2.1 Main window
[Figure 1] BCM350N main window
Such as [Figure 1], the main window has a toolbar. The toolbar is located in
upper region of the main window. Also, there are 8 control panels for remote
control 8 fetal monitors in the left side of the main window. And there are Patient
Information region and NST chart region on the right of the control panels.
1. Toolbar button
There are several buttons for dealing with program on the upper region of the
main window.
These buttons are listed in the below table.
Patient Registration
New patient can be registered by using the
button. If you click the button, the [Patient
Registration] window will come out.
Doctor Registration
New doctor can be registered by using the button.
If you click the button, the [Doctor Registration]
small window will come out.
BCM350N Operator’s Manual 11
P/N : BCM350N- EN-OPM Bistos Co.,Ltd. R01(2013.11.13)
Review
If you click the button, the [Review] window will
come out. You can review and print the saved
NST chart. and convert the NST chart to jpg type
of image file.
1,2,3,4,6,8 Bed If you click each of the buttons, you can see the
NST chart according to number of bed view.
Bed View setting
If you click the button, the [Multi Bed View
Setting] window will come out. You can select the
bed that you want to see at the relevant bed view
respectively.
Language
If you click this button, available languages are
lists on pop-up menu. You can change the
language by choosing one of them.
Setting
If you click the button, the [Configuration]
window will come out. You can change the
necessary setting (or configuration).
Close To terminate NST Program.
(1) Control panel
Every bed has its own control panel which is located on the left of the main
window. The below figure is an example control panel of “Bed 1”. The functions
are starting or stopping diagnosis, turning off fetal monitor, starting or stopping
print and disabling alarm.
The patient’s name will be displayed in bed button region(①). If the user clicks
this button, the relevant graph region will be shown up and the graph will be
displayed. Also, The FHR1, FHR2 and UC will be displayed in value button
region(②) in real time.
In the case of the program is connected to fetal monitor, the connection
indicator on the left of the bed button is changed to green from red. When the
program starts to receive diagnosis data, the connection indicator is blinking.
The follows are functional description of each button in control panel.
① ②
BCM350N Operator’s Manual 12
P/N : BCM350N- EN-OPM Bistos Co.,Ltd. R01(2013.11.13)
Button Name Functional description
( ) Start / Stop To start or stop NST diagnosis
( ) Sound Listen On or Off the real time fetal heart sound.
Device Printer To control the device printer
Device Power Fetal monitor power off function
Device Alarm
Sound
You can control the alarm enable and disable
function, start or stop printing function.
Assign Patient To assign the patient
Reset When clicked, the bed configuration is reset.
Connected If connected, the connection indicator is changed
to green from red. Disconnected
Printer During print, it blinks.
Saving While the patient is assigned, if diagnosis is
ongoing, It means real time saving and blinks.
(2) Patient basic profile
In this region, patient basic information is displayed. Displaying parameters are
shown as below.
※ GA : Gestational Age
(3) NST chart region
In this region, FHR1, FHR2 and UC are displayed.
Use the scroll bar If you want to see other bed’s NST chart.
BCM350N Operator’s Manual 13
P/N : BCM350N- EN-OPM Bistos Co.,Ltd. R01(2013.11.13)
2.2 Patient registration window
Click [Patient Registration] button. Then ‘Patient Registration’ window will be
shown as below.
This window is composed of patient search region, patient information region,
guardian information region and medical information. Also the number of
registered patient is displayed.
[Figure 2] Patient Registration window
BCM350N Operator’s Manual 14
P/N : BCM350N- EN-OPM Bistos Co.,Ltd. R01(2013.11.13)
(1) Searching existing patient.
To use this function, insert the patient’s ID or name and press ‘Enter’ key. Also
click [+] region of tree structure and search the patient directly. The tree
structure is classified between two methods, registration date and name.
(2) New patient’s basic information registration region: [New]->[OK]
To register a new patient, you can insert ID, name, SSN(Social Security number),
zip code, address, home phone number, cell phone, birthday, attending doctor
and registration date in the ‘Patient Information’ region.
Also you can insert guardian’s name, contact and relation in the ‘Guardian
Information’ region.
To register a new patient, you must insert ID and name. The other items are
optional. Then, click [OK] button to complete and save it.
(3) Medical information region :
In this region, a doctor’s comment and delivery history such as LMP, EDC, GA
and Parity information is recorded. Then, press [OK] button to complete and
save it.
※ LMP : Last Menstrual Period EDC : Estimated Day of Confinement
GA : Gestational Age Parity : The number of pregnancies
If you want to delete it, select the item of patient and click [Delete] button.
Note : Same name can be registered, but Same ID cannot be registered. (When
you try to register the patient, if there is same ID, message comes out.)
BCM350N Operator’s Manual 15
P/N : BCM350N- EN-OPM Bistos Co.,Ltd. R01(2013.11.13)
2.3 Doctor registration window
Click [Doctor Registration] button. Then ‘Doctor Registration’ window will be
shown as below.
[Figure 3] Doctor Registration window
To register a new doctor, you can insert name and telephone number and press
[Save] button. If you want to delete it, select the item and use [Delete] button.
2.4 NST Review window
To review saved data, click [Review] button. Then [Review] window will be
shown as below.
[Figure 4] Review window
BCM350N Operator’s Manual 16
P/N : BCM350N- EN-OPM Bistos Co.,Ltd. R01(2013.11.13)
The NST Review window is composed of Direct / Tree search region, patient
information region, history region, and graph display region. Using direct or tree
search, you can search and select the patient. If the patient is selected, the
patient’s information is displayed in the patient information region. Also saved
diagnosis list is displayed in history region. When you click one item of the
diagnosis list, the relevant graph is displayed in the graph display region. In the
Print option region, you can select the range of printing. If you want to print the
test result as soon as the test is finished, you can select the relevant function
in After test print region. The printing paper size is fixed to A4.
2.5 Bed view setting window
[Figure 5] Bed view setting window
You can select beds what you want to view at the same time in this window.
Select beds to want to see when the relevant bed view is clicked. And then click
[Save] button.
2.6 Language setup window
Click [Language] button. Then the popup menu comes out. The default
language is English. BCM350N supports English, Korean, Japanese and
Spanish. You can change another language after registering.
BCM350N Operator’s Manual 17
P/N : BCM350N- EN-OPM Bistos Co.,Ltd. R01(2013.11.13)
2.7 Setting(Configuration) window
To change the configuration, click [Setting] button. Then [Configuration]
window will be shown as below.
[Figure 6] Configuration window
In this window, you can select the port to use and wave grid of graph mode.
Also you can modify the path of data base and image.
Note : You can change the path of database(DB), data, and image, but we
recommend that you hardly change them.
2.8 Close
To finish BCM350N, click [Close] button in the main window.
BCM350N Operator’s Manual 18
P/N : BCM350N- EN-OPM Bistos Co.,Ltd. R01(2013.11.13)
Chapter 3. How to use BCM350
After installing communication module and setting communication program, run
BCM 350N as follow path. <Start Programs BCM350N BCM350N > Then
the main window comes out as Figure 1. Operate BCM350N according to the
following procedure.
1. Language setting: Select the language what you want.
2. Port setting: Select the ports relevant to each bed and the monitor type.
(The baud rate is set automatically.) (See 6. 2.)
3. Register a new patient.
4. Select the bed what you want to connect in the control panel.
5. Move the mouse pointer to the graph display region and click on the right
mouse button. And then the popup menu comes out. Select [Patient Select]
item in the popup menu.
6. [Patient Basic Registration] window comes out. Then select the patient
using [Direct Search] or [Tree Search].
7. The selected patient’s name will be presented in the bed button region of
the control panel which is located in the left of the main window.
8. Click [Start] button in the control panel and start the diagnosis.
9. Click [Stop] button in the control panel and stop the diagnosis.
3.1 How to setup language
1 [Language] Click the Language setup button in the main
window.
2
The popup menu comes out such as left. The
present language is checked. Select the
language what you want. Default language is
English. Supporting language are Korean,
English, Japanese and Spanish. If you want to
add another language, please contact us.
BCM350N Operator’s Manual 19
P/N : BCM350N- EN-OPM Bistos Co.,Ltd. R01(2013.11.13)
3.2 How to setup communication port
1 [Setting] Click [Setting] button in the
main window.
2 Refer to [Figure 6].
The [Configuration] window
will be shown such as
[Figure 6].
3
Click [Serial port Setup]
button.
4
The [Port Setup] popup
window comes out. Select a
bed number and select
connect serial port number.
(If the connected device to
the bed uses 1.6k protocol
such as BT-300, check
“1.6k”)
☞ When you select serial port, if the port is not connected, selectable port
may do not appears.
5 [Save] When the setting is finished,
click [Save] button.
☞ [Cancel]
If you want to exit without
change, click [Cancel]
button.
BCM350N Operator’s Manual 20
P/N : BCM350N- EN-OPM Bistos Co.,Ltd. R01(2013.11.13)
3.3 How to register patient basic information
1 [Patient Registration] Click [Patient Registration] button in the main
window.
2 Refer to [Figure 2]. The [Patient Registration] window will be
shown such as [Figure 2].
3 [New]
If you want to register the new patient, click
[New] button. (Button to make all of the
blanks)
3
(1) Input the patient’s information. To register a new patient, you must
input the patient’s ‘ID’ and ‘name’. If not, it is impossible to register a
new patient. ‘Reg. Date’ in the Patient Information region means the date
of the first registration date.
(2) When you input all information, click [OK] button.
4
The dialogue box comes out. Click [OK]
button. Then, new patient is added.
BCM350N Operator’s Manual 21
P/N : BCM350N- EN-OPM Bistos Co.,Ltd. R01(2013.11.13)
☞
or
When the registration is
completed, the directory named
by the patient name(Name) and
registered date(Reg. Date) is
created. You can confirm the
patient’s information.
3.3.1 How to edit existing patient registration information
1 [Patient Registration]
Click [Patient Registration]
button in the main window.
2 Refer to [Figure 2].
The [Patient Registration]
window will be shown such as
[Figure 2].
4
Input the patient’s name or ID
what you want to find in the
Direct Search region and press
enter. For example, input ‘Lee’
and press enter.
☞ If there is no patient to find, any information is not appeared.
4-1
or
Also using the directory named
by the patient name(Name) and
registered date(Reg. Date), you
can search the patient what you
want to find. Refer to the
registration date and click [+]
button or patient’s name directly
and select the patient.
5
BCM350N Operator’s Manual 22
P/N : BCM350N- EN-OPM Bistos Co.,Ltd. R01(2013.11.13)
In this window, edit the item what you want, and then click [OK]
button. (In this window, the Home Phone number is modified.)
6
The [Update] popup window
comes out. If you want to edit
and save, click [Yes] button.
7
When the update or edit is
completed, the dialogue box
comes out one more time.
Click [OK] button.
☞
If you want to delete the existing patient registration information, click
[Delete] button in the [Patient Registration] window after selecting the
patient what you want.
3.4 How to assign a patient to the relevant bed.
There are three ways to assign a patient to a bed. First, click the [Assign] icon
in the control panel of left menu.
Second, move the mouse pointer to FHR grid of the graph region and click on
the right button. Then the popup menu comes out. Select <Assign Patient>.
Finally, Select a patient from [Patient Registration] window, and click [Assign
to Bed] button directly.
OK
BCM350N Operator’s Manual 23
P/N : BCM350N- EN-OPM Bistos Co.,Ltd. R01(2013.11.13)
1-1
Click the [Assign] icon in the control
panel of left menu.
1-2
Click the bed button and move the
mouse pointer to the graph region and
click on the right button.
The popup menu comes out. Select
<Assign Patient>.
※ <Start>/<Stop> has same function with [Start/Stop] icon in the
control panel.
※ <Reset> has same function with [Reset] icon in the control panel.
1-3 [Patient Registration] Click [Patient Registration] button in the
main window.
2 Refer to [Figure 2]. The [Patient Registration] window will
come out such as [Figure 2].
3
Input the patient’s name or ID what you
want to find and press enter. For
example, input ‘Lee’ and press enter.
☞ If there is no patient to find, any information is not appeared.
4
Selected patient’s information is presented.
Then click [Assign to Bed] button. And this window disappears.
BCM350N Operator’s Manual 24
P/N : BCM350N- EN-OPM Bistos Co.,Ltd. R01(2013.11.13)
5
The [Assign Patient] popup window
comes out. Select the bed number.(For
example, select <Bed 1>.)
And click [OK] button.
6
The selected patient’s name is
presented in the control panel.
Example) ‘Bed 1’ is connected to the
patient ‘Lee’.
3.5 How to start diagnosis
1-1
Click [Start] button in the control panel.
Then the diagnosis starts.
1-2
Click the bed button and move the mouse
pointer to the graph region and click on
the right button.
The popup menu comes out. Select
<Start>.
2
If the communication port is normal, the
signal indicator changes to green from
red. The FHR and UC are displayed.
Also FHR and UC are displayed in the
graph region.
The graph of FHR and UC are displayed.
(FHR1 and FHR2: ①, UC: ②)
①
②
BCM350N Operator’s Manual 25
P/N : BCM350N- EN-OPM Bistos Co.,Ltd. R01(2013.11.13)
☞
If the communication port is abnormal, the
dialogue box such like the left comes out.
To clear this box, click [OK] button.
In this case, exit program and check the
communication port setting program.
3.5.1 Display of information during diagnosis
During diagnosis, various information is displayed in the graph region window.
The meaning are as follows.
Patient’s Information
HR1 HR2 UC Diagnosis
time Functions
HR1 FHR value from Doppler 1 probe.
HR2 FHR value from Doppler 1 probe.
UC UC value from UC probe.
Note : [Save to JPG] and [Print] buttons for function appears after diagnosis
completion.
3.6 How to stop diagnosis
1-1
Click [Stop] button in the control panel.
OK
OK
BCM350N Operator’s Manual 26
P/N : BCM350N- EN-OPM Bistos Co.,Ltd. R01(2013.11.13)
1-2
Click the bed button and move the mouse
pointer to the graph region and click on
the right button.
The popup menu comes out. Select
<Stop>.
2
The [Stop] popup window comes out. If
you like to stop, click [Yes] button
3
If the diagnosis is stopped, the signal
indicator is change to red.
☞ If a patient is not assigned to the bed, the following dialogue
box comes out.
1
The [Save] popup window comes out. If
you like to save data, click [Yes] button
2
Yes
The [Patient Registration] window comes out.
☞
If you want to save diagnosis data, assign the patient
to the bed.
[Exit]
The [Discard] popup window comes out. . If you sure
to discard the data , click [Yes] button
No Stop monitoring without saving.
☞ Even if you do not save the patient information, diagnosis data still remain.
BCM350N Operator’s Manual 27
P/N : BCM350N- EN-OPM Bistos Co.,Ltd. R01(2013.11.13)
3.7 How to reset bed information
If you want to disconnect a bed and patient, click this button.
1-1
In the status of the diagnosis is
stopped, click the [Reset] icon in the
control panel of left menu.
1-2
Click the bed button and move the
mouse pointer to the graph region and
click on the right button.
The popup menu comes out. Select
<Reset>.
☞
This function does not work during
diagnosis. In this case, the dialogue
box such like the left comes out. To
clear this box, click [OK] button.
3.8 How to setup bed view
When BCM350N starts, there is one bed in the graph region of the main window.
Using this button, you can setup bed view. The number of bed you can view at
the same time is 2, 3, 6 and 8. The default view is one bed view. If you want to
view several beds, you have to setup bed view first. Follow next procedure to
setup bed view.
1 [Bed View Setting] Click [Bed View Setting] button in the
main window.
2
Select the beds to display and click [Save] button.
OK
BCM350N Operator’s Manual 28
P/N : BCM350N- EN-OPM Bistos Co.,Ltd. R01(2013.11.13)
☞
If there is a checking error, If there is a
checking error, the error message window
comes out. Click [OK]. Then check the
error and re-check beds. And click [OK]
button.
3
[1 Bed], [2 Bed], [3 Bed],
[4 Bed], [6 Bed], [8 Bed]
If you click each of the buttons, you can
see the NST chart according to number of
bed view.
☞ If you want to see only one, double click the necessary one bed on the
multi view screen in the main display.
Follow next procedure to change the bed number in the control panel.
1 [Setting]
Click [Setting] button in the main window.
Then, The [Configuration] window will
come out.
2
Click [Bed Number Setting] button. And this window disappears.
OK
BCM350N Operator’s Manual 29
P/N : BCM350N- EN-OPM Bistos Co.,Ltd. R01(2013.11.13)
3
The [Configuration] window will come out and you can modify the bed
number you want in the control panel. To apply the setting, click [Save]
button.
3.9 How to listen fetal heart sound
BCM350N support the hearing the real time fetal heart sound function.
1
Choose the bed what you want to hear
fetal heart sound and click [Sound Listen]
button in the control panel. The other
panel’s [Sound Listen] buttons are
deactivated.
2
To deactivate this button, click again this
button. In this case, this button is
activated like a procedure 1.
☞ In the case of the [Sound Listen] button is activated, you can hear fetal
heart sound from the device or the PC speaker.
3.10 How to control fetal monitor
3.10.1 Device Printer
1
Click [Device Printer] button in the control
panel.
BCM350N Operator’s Manual 30
P/N : BCM350N- EN-OPM Bistos Co.,Ltd. R01(2013.11.13)
3.10.2 Device Alarm Sound
3.10.3 Device Power
3.11 How to review existing diagnosis
1 [Review] Click [Review] button in the main
window.
2 Refer to [Figure 4]. The [Review] window comes out.
4
Input the patient’s name or ID
what you want to find and press
enter. For example, input ‘Lee’
and press enter.
☞
If there is no patient to find, the
error message comes out. To
clear this message, click [OK]
button.
☞ During printing, [Device Printer] icon blinks.
1
Click [Device Alarm Sound] button in the
control panel.
☞ You can hear the alarm sound from the device or the PC speaker. Alarm
on and off can be checked on the device only.
1
If you want to turn off the device power,
Click [Device Power] button in the control
panel.
☞ This function is only available during diagnosis. It can be only set the
device off.
OK
BCM350N Operator’s Manual 31
P/N : BCM350N- EN-OPM Bistos Co.,Ltd. R01(2013.11.13)
4-1
Or using Tree Search, you can
search the patient what you want
to find. Refer to the registration
date and select the patient.
5
The existing diagnosis list is
displayed in the upper right
region. Click the list what you
want to review.
6
Then the saved chart and relevant information are displayed.
☞
You can save to JPG or print 20 minutes of the stored data you want. In
the [Review] window, there is a track bar on the displayed chart. The
position of the track bar is point to save to JPG or print.
7
BCM350N Operator’s Manual 32
P/N : BCM350N- EN-OPM Bistos Co.,Ltd. R01(2013.11.13)
When you place the mouse pointer on the NST chart, time appears. At
that time, if you double click, it is possible to enter a record of the text.
After completing the input, click the [Save] button. Then text is saved.
☞
Double click to delete the text
entered.
Click the [Yes] button.
☞ If you want to delete the existing diagnosis in upper right, press [Delete]
key after selecting diagnosis information of the patient what you want.
Note: Not only you can read and save the USB data that is stored in the device
to PC(desktop or notebook), but you can save to JPG or print it.
☞
After connecting the USB memory stick to coputer, enter the [Review]
window and USB drive is shown on the left top.
BCM350N Operator’s Manual 33
P/N : BCM350N- EN-OPM Bistos Co.,Ltd. R01(2013.11.13)
3.12 How to save to image and print
3.12.1 Save to JPG
1-1
In the graph region window, click the [Save to JPG] button.
1-2 Refer to [Figure 4]. Click the [Save to JPG] button in the lower
left of [Review] window.
2.
According to the path designated in the path item of [Configuration]
window, the image is stored by ‘.jpg’ format.
3.12.2 Print
1-1
In the graph region window, click the [Print] button.
BCM350N Operator’s Manual 34
P/N : BCM350N- EN-OPM Bistos Co.,Ltd. R01(2013.11.13)
1-2 Refer to [Figure 4]. Click the [Print] button in the lower left of
[Review] window.
2
The [Print Preview] popup window comes out.
3 Click [Print] button.
3.12.3 Setting to JPG and Print
When saving to JPG or printing the stored data, information of each page and
graph line of 10 units is shown.
BCM350N Operator’s Manual 35
P/N : BCM350N- EN-OPM Bistos Co.,Ltd. R01(2013.11.13)
☞
<Detailed Grid>
According to the printing of [Configuration] window, you can change
the printing mode as ‘Landscape’ or ‘Portrait’. Also, if you check
‘Detailed Grid’, you can print graph line of 10 units
BCM350N Operator’s Manual 36
P/N : BCM350N- EN-OPM Bistos Co.,Ltd. R01(2013.11.13)
Note : When the connection is lost with the device, the reconnection is tried
automatically and it is marked with dotted line.
☞
Note: If you have two DOP1, DOP2 graph, ‘offset 20’ applied like a device when
they save to JPG print only.
☞
BCM350N Operator’s Manual 37
P/N : BCM350N- EN-OPM Bistos Co.,Ltd. R01(2013.11.13)
Chapter 4. BCM350N Uninstallation
Two ways to uninstall the BCM350N
- 1st
1. Run [uninst.exe].
2. The dialogue box comes out. Click [Yes] button.
- 2nd
1. Use the ‘Control panel’ of your desktop.
BCM350N Operator’s Manual 38
P/N : BCM350N- EN-OPM Bistos Co.,Ltd. R01(2013.11.13)
Chapter 5. Specification
BT-350 fetal monitor specification
Physical Chracteristics
Size 9.6 cm H x 32.6cm W x 27.6cm D
Weight About 5.5kg(except battery)
Safety
EN60601-1, EN60601-1-1, EN60601-1-2 Standard applied
Class I equipment & Internal powered equipment
Type BF applied parts
Continuous Operation
Power
Internal Battery
Ni-MH rechargeable battery
14 hours to full recharge during monitoring
14 hours to full recharge when monitor is
off
Using time : about 30 minutes(Dop 1/2
on[vol4], Print on[3cm/min, FM print on])
Power consumption : max. 10watt
External AC power AC 100~240V, 50/60Hz, 80VA maximum
Environmental
Operating temperature 10°C ~ 40°C (50°F ~ 104°F)
Storage temperature –20°C ~ 60°C (–4°F ~ 140°F)
Relative humidity 20% to 90%, non-condensing
Altitude 0 -3048m (0 -10,000 ft)
BCM350N Operator’s Manual 39
P/N : BCM350N- EN-OPM Bistos Co.,Ltd. R01(2013.11.13)
FHR measuring
Item Value
Probe surface IR(SATA) ≤ 10 mW/cmP
2P
Incidence beam area 19.95 cmP
2P, circle type
Ultrasound frequency 0.985 MHz ±0.5%
Pulse lasting time 97.65625 microseconds
Pulse repeat frequency 3.2 kHz
Range of FHR measuring 30-240 BPM
Accuracy ±2% of range
Leakage current <10 µA @ 264 VAC applied to transducer
Voltage withstand >4 kV RMS, Type BF applied part
UC measurement
Item Value
UC Range 0-99, relative unit
Resolution 1 Count
Accuracy ±1% relative unit
Leakage current <10 µA @ 264 VAC applied to transducer
Voltage withstand >4 kV RMS, Type BF applied part
Paper
Pack style Z-Fold.
Pack size 150 mm x 90 mm x 15 mm
End of pack Mark along paper edge
Loading Side-door, slide-in
Paper Detector Paper out
Door open
Paper speed
Normal 1, 2, and 3 cm/min ±1%
High speed 10 cm/min (only in Trend
mode)
Paper tracking accuracy : ±1% (exclusive of paper accuracy)
BCM350N Operator’s Manual 40
P/N : BCM350N- EN-OPM Bistos Co.,Ltd. R01(2013.11.13)
Appendix 1. Wired communication system
1. How to making connection cable
1.1 In case of connecting a fetal monitor (BT-300, BT-350 etc.) to PC
directly
You can get RS-232C cross cable in the market. Both side of
connector is D-SUB9 (Female). Or you can make a cable with the
connection information as below.
(Please secure the enough length of cable considering setting place
and distance between PC and a fetal monitor.)
1.2 In case of connecting monitors to the central monitoring PC through
multi-port. The cable information is as follow;
RXD
TXD
GND
BCM350N Operator’s Manual 41
P/N : BCM350N- EN-OPM Bistos Co.,Ltd. R01(2013.11.13)
2. Specification of Multi-port
Use the SystemBase Co.,Ltd.’s product - Multi-8/ Multi-8/USB Series.
SystemBase Co.,Ltd.
www.sysbas.com
BCM350N Operator’s Manual 42
P/N : BCM350N- EN-OPM Bistos Co.,Ltd. R01(2013.11.13)
Appendix 2. Wireless communication system
Table contents
1. Introduction of wireless communication system …………… 48
2. Checking IP address of installing PC ………………………… 48
3. How to set MSP-1000 IP address ………………………………… 51
4. How to set MSP-1000 ……………………………………………… 57
5. How to connect SD-1000 …………………………………………… 64
6. How to set Serial IP address ……………………………………… 70
BCM350N Operator’s Manual 43
P/N : BCM350N- EN-OPM Bistos Co.,Ltd. R01(2013.11.13)
1. Introduction of wireless communication system
MSP-1000 is communication device. Using this device, it is possible to
use bluetooth connection through Ethernet. Also this device supports
simultaneous access of bluetooth device. Parani-SD1000 is a Bluetooth
serial adaptor. Using MSP-100 and SD-1000, it is possible to comprise
wireless communication system.
2. Checking IP address of installing PC
2.1. Pre-installation before checking IP address
2.1. Connect a DC 5V power adapter to MSP-1000. And connect
internet LAN cable to MSP-1000 “EHT0” port. And connect MSP-1000
“ETH1” port and PC LAN port using LAN cable. Using serial cable,
connect MSP-1000 and PC.
2.2. Checking IP address
2.2.1. Run [Run] as below.
Input [cmd] and click [OK].
BCM350N Operator’s Manual 44
P/N : BCM350N- EN-OPM Bistos Co.,Ltd. R01(2013.11.13)
2.2.2. MS-DOS window comes out as below.
2.2.3. Input [ipconfig/all] and press [Enter]. Then below window comes out.
2.1.4. The red box in the upper window is IP address.
Make a note IP addresses in the blue box.
※ For example, make a note as follow.
1) IP Adress : 192. 168. 131. 65
2) Subent Mask : 255. 255. 255. 0
3) Default Gateway : 192. 168. 131. 254
4) DHCP Server : 192. 168. 131. 254
5) DNS Servers : 61. 97. 112. 155
61. 97. 113. 250
BCM350N Operator’s Manual 45
P/N : BCM350N- EN-OPM Bistos Co.,Ltd. R01(2013.11.13)
2.3. Check whether IP address is occupied or not
To install MSP-1000, you should be allowed to use IP address. If you use IP
address without checking whether that is occupied or not, it may occurs the
communication problem because of IP collision. Therefore you must check
whether IP address is occupied or not as follows.
2.3.1. In the case of IP address from a hospital is [192. 168. 131. 200].
1) Input [ping 192.168.131.200] in the MS-DOS window.
2) And then press [Enter].
3) The results are as below.
4) In the case of IP address is occupied, you can check the message like
as “Reply from 192.168.10.65: byte=32 time<1ms TTL=128”. In the case
of IP address is not occupied, you can find the message like as
“ Request timed out.”. In the case of IP is not occupied, you can use IP
from a hospital.
IP is occupied.
IP is not occupied.
BCM350N Operator’s Manual 46
P/N : BCM350N- EN-OPM Bistos Co.,Ltd. R01(2013.11.13)
3. How to set MSP-1000 IP address
3.1. How to connect MSP-1000 and PC
3.1.1. Connect MSP-1000 and PC using a serial cable.
3.1.2. Run [Hyperterminal]. The path is [Start →All Programs→Accessories→
Communications→Hyper Terminal].
3.1.3. After run [Hyper Terminal], the window as below comes out.
Input name (all value is available) and click [OK].
BCM350N Operator’s Manual 47
P/N : BCM350N- EN-OPM Bistos Co.,Ltd. R01(2013.11.13)
※ If [New connection] window does not come out and [Location Information]
window comes out, follow the next procedure.
Click 1 →2 →3.
3.1.4. Return to 3.1.3 and select COM port(in this manual, COM3 is used) for
[Connect using] and click [OK].
2
1
1
1
1
2
3
BCM350N Operator’s Manual 48
P/N : BCM350N- EN-OPM Bistos Co.,Ltd. R01(2013.11.13)
3.1.5. [COM3 Properties] window comes out. Then follow the below procedure.
Click [Restore Defaults] and [OK]. Then the basic values are changed.
And click [OK] and complete this step.
<Before run [Resotre Defaults]> <After run [Resotre Defaults]>
3.1.6. When there is no sign in the hypertermial window, press enter. Then
logging in message comes out. Input “root” for ID and password. You must
input in small characters. After logging in, input “editconf” and press enter.
1
1 2
1 3
1
BCM350N Operator’s Manual 49
P/N : BCM350N- EN-OPM Bistos Co.,Ltd. R01(2013.11.13)
3.1.7. If all steps are completed normally, the below messages come out.
3.1.8. Input ‘1’ and press enter. Repeat this step until the below messages
come out.
BCM350N Operator’s Manual 50
P/N : BCM350N- EN-OPM Bistos Co.,Ltd. R01(2013.11.13)
3.1.9. After complete this step like as upper window, check ‘2. IP address’. If IP
address differs from IP address what you received from the hospital, input ‘2’
and press enter. Then input IP address what you received from the hospital and
press enter. You can change IP address through this step. And check items
from 3 to 6. If IP addresses are different, change IP address using the same
changing method of the item ‘2. IP address’.
3.1.10. After change MSP-1000 IP address, press [ESC] 3 times and return to
the initial window.
Input received IP address
= 192.168.10.200
BCM350N Operator’s Manual 51
P/N : BCM350N- EN-OPM Bistos Co.,Ltd. R01(2013.11.13)
3.1.12. In this window, input ‘8’ and press [ENTER]. Then two questions come
out step by step. Input ‘Y’ twice.
3.1.13. Input ‘Editconf’ and press [ENTER]. Then check IP address.
3.1.14. If there is no problem in IP address, input ‘8’ and press [ENTER]. Then
two questions come out step by step. Input ‘Y’ twice. And disconnect serial
cable from MSP-1000 and PC.
BCM350N Operator’s Manual 52
P/N : BCM350N- EN-OPM Bistos Co.,Ltd. R01(2013.11.13)
4. How to set MSP-1000
4.1. Set MSP-1000
4.1.1. Run Internet Explore.
4.1.2. Input IP address in the web browser and press [ENTER].
For example) Input 192.168.131.200 then the login box comes out.
4.1.3. Input ‘root’ for User ID and Password. Then the window such as below
comes out. In this window, you can find IP address and change IP
address.
Click [Save & apply] button.
BCM350N Operator’s Manual 53
P/N : BCM350N- EN-OPM Bistos Co.,Ltd. R01(2013.11.13)
4.1.4. How to connect MSP-1000 and SD-1000.
Click 1 → 2 → 3.
4.1.5. Then the dialogue box comes out as below.
4.1.6. Click [Ok] and complete the configuration.
4.1.7. In this window, you can set the number of SD-1000. Click [Add] button.
1
2
3
Check after click 1.
Ok Cancel
BCM350N Operator’s Manual 54
P/N : BCM350N- EN-OPM Bistos Co.,Ltd. R01(2013.11.13)
If you want to add more SD-1000, click [Add] button again.
4.1.8. The below is the case of the number of SD-1000 is 3. In this case, you
should click [Add] button three times.
4.1.9. In the upper window, you should change each MAC address. Follow
next steps.
4.1.10. When you select a SD-1000 what you want to change MAC address, the
BCM350N Operator’s Manual 55
P/N : BCM350N- EN-OPM Bistos Co.,Ltd. R01(2013.11.13)
below window comes out. In this window, you should change MAC
address.(MAC address is printed in the SD-1000 label.)
4.1.11. Input SD-1000’s MAC address.
The values for the first 6 are default values. They are fixed to [00:01:95].
You should input the rest values.
00:01:95:08:ab:f4
Each SD-1000 has different MAC address.
Please check and input the right address.
BCM350N Operator’s Manual 56
P/N : BCM350N- EN-OPM Bistos Co.,Ltd. R01(2013.11.13)
4.1.12. Input MAC address and click [Save & apply]. Repeat these steps
according to the number of SD-1000.
4.1.13. When you complete these steps, the dialogue box comes out. Then
click [OK] and save configurations.
4.2. How to set Port
4.2.1. Click [Network Service] and input the port number.
4.2.2. Click [Network service] and add port by clicking [Add] button.
OK
BCM350N Operator’s Manual 57
P/N : BCM350N- EN-OPM Bistos Co.,Ltd. R01(2013.11.13)
4.2.3. In the upper window, click “Listening on 7000” and change port number.
4.2.4. MSP-1000’s “Local port number” is start from 7001.
Repeat these steps to change port number according to the number of
SD-1000. For example, the port numbers are 7002, 7003 and so on.
After change the port numbers, click [Save & apply] button and save
configurations.
4.2.5. After you click [Save & apply] button, click [Port configuration]. (Serial
BCM350N Operator’s Manual 58
P/N : BCM350N- EN-OPM Bistos Co.,Ltd. R01(2013.11.13)
Port Profile Port configuration) Then the below window comes out.
Check the MAC addresses are correct.
4.2.6. When you complete configurations, click [Apply changes] and [Logout].
Then exit program.
BCM350N Operator’s Manual 59
P/N : BCM350N- EN-OPM Bistos Co.,Ltd. R01(2013.11.13)
5. How to set SD-1000
5.1. Install SD-1000 program and configuration
5.1.1. Setup SD-1000 program.
Run Setup_sd1000_conf-v2.01.exe.
5.1.2. The below window comes out and click [Finish] button.
BCM350N Operator’s Manual 60
P/N : BCM350N- EN-OPM Bistos Co.,Ltd. R01(2013.11.13)
5.1.3. Run SD-1000 program.
5.1.4. After running SD-1000 program, check the communication status.
Connect ‘USB power cable’ to SD-1000. Then connect this cable to
PC‘USB-Port’. And connect SD-1000 to PC’s serial port. Check that
Power led is on and Connect led is blinking with green.
5.1.5. The default Baud Rate value of SD-1000 is 9600. Change the Baud Rate
to 115200 and click [OK].
BCM350N Operator’s Manual 61
P/N : BCM350N- EN-OPM Bistos Co.,Ltd. R01(2013.11.13)
5.2. How to initialize SD-1000
5.2.1. Click [Device Setting] in the main window.
5.1.2. To initialize SD-1000, click [Hard Reset] and [Apply].
BCM350N Operator’s Manual 62
P/N : BCM350N- EN-OPM Bistos Co.,Ltd. R01(2013.11.13)
5.1.3. After complete SD-1000 initialization, click [Connection].
And click [Search].
5.1.4. After searching process, click [MSP-1000-0] and [Connect].
1
2
BCM350N Operator’s Manual 63
P/N : BCM350N- EN-OPM Bistos Co.,Ltd. R01(2013.11.13)
5.1.5. If MSP-1000 is connected successfully, click [OK].
5.1.6. After complete previous steps, click [Disconnect] and [OK].
2
1
Connection completed
Connection dropped
BCM350N Operator’s Manual 64
P/N : BCM350N- EN-OPM Bistos Co.,Ltd. R01(2013.11.13)
5.2. How to set SD-1000 connection
5.2.1. Click [Device Setting]. And follow the next step. Click 1 → 2 → 3 → 4.
5.2.2. Disconnect SD-1000 from PC. Then connect another SD-1000 to PC.
Follow clause from 5.1.4 to 5.2.1.
1
3
2
4
H/W flow control
BCM350N Operator’s Manual 65
P/N : BCM350N- EN-OPM Bistos Co.,Ltd. R01(2013.11.13)
6. How set Serial IP
6.1. Install program
6.1.1. Run Setup_Sena_serialip-v4.8.4 and install this program.
6.1.2. Follow the below steps.
BCM350N Operator’s Manual 66
P/N : BCM350N- EN-OPM Bistos Co.,Ltd. R01(2013.11.13)
6.1.3. Click [Finish] and complete installing.
6.2. How to set SD-1000 Port
6.2.1. Run the Serial/IP Control Panel software (Start → Programs → Sena →
Serial IP → Control Panel). And click [Select Ports].
BCM350N Operator’s Manual 67
P/N : BCM350N- EN-OPM Bistos Co.,Ltd. R01(2013.11.13)
6.2.2. Select COM ports according to the number of SD-1000.
We recommend that you should select COM number under 20.
6.2.3. The below window shows that the method of setting IP address and port
number when two COM ports are used. Note that you must input IP
address which is received from a hospital and Port number. (Ex. IP
address from a hospital is 192.168.10.200 and SD-1000 port numbers
are 7001~700X.)
BCM350N Operator’s Manual 68
P/N : BCM350N- EN-OPM Bistos Co.,Ltd. R01(2013.11.13)
6.3. Checking the connection status
6.3.1. Click [Configuration Wizard...] and check the connection status.
6.3.2. The below windows show that the normal connection status and
abnormal connection status.
To test the communication
status, click [Start] button.
SD-1000 must be connected
to the monitor.
Normal connection
Abnormal connection / Disconnected
BCM350N Operator’s Manual 69
P/N : BCM350N- EN-OPM Bistos Co.,Ltd. R01(2013.11.13)
Warranty Product name Central monitoring & Networking system
Model name BCM350N
Permission No.P
1) N/A
Serial No.
Warranty Period 2 years from purchasing date
Date of purchase
C
U
S
T
O
M
E
R
Hospital :
Address :
Name :
Contact :
Agent / Dealer
Manufacturer Bistos Co.,Ltd.
※ Thanks for your purchasing our product.
※ This product is manufactured and passed through strict quality control system and
inspection.
※ Compensation standard concerning repair, replacement, refund of the product
complies with “Consumer’s protection law” noticed by Economic Planning Dept..
※ If an abnormal condition is happened with normal use in the term of guarantee,
the service center will offer free repair service.
※ If there is problem in the warranty period, please inform us model name, serial
number, purchasing date, problem and so on.
Note1) ‘Permission No.’ is a identify number by Korean Food and Drug Administration.
Manufacturer Information
Manufacturer
Head office /
Manufacturing
site
7P
thP Fl. A Bldg., Woolim Lions Valley 5cha,
144-3, Sangdaewon-dong, Jungwon-gu,
Seongnam-si, Gyeonggi-do, Korea
Telephone 82-31-750-0340
Tele fax 82-31-750-0344
Web site www.bistos.co.kr
e-mail [email protected]
BCM350N Operator’s Manual 70
P/N : BCM350N- EN-OPM Bistos Co.,Ltd. R01(2013.11.13)
Contact to Technical support
Telephone: +82 31 750 0340
Fax: +82 31 750 0344
Bistos Co., Ltd.
7P
thP Fl. A Bldg., Woolim Lions Valley 5cha, 302,
Galmachi-ro, Jungwon-gu, Seongnam-si,
Gyeonggi-do, Korea
2TUwww.bistos.co.krU2T
Model Name: BCM350N
EC Representative: Medical Econet Gmbh
High-t'Park Mainstr. 6c-6d D-45768 Marl/Germany
Telephone: +49 2365 92 437 - 0
Fax: +49 2365 92 437 – 55