DEPARTMENT OF BIOMEDICAL ENGINEERING
MGM COLEGE OF ENGINEERING AND TECHNOLOGY,
NAVI MUMBAI
PROJECT REPORT
ON
PC BASED PESPIRATION RATE
&
OXYGEN SATURATION MEASUREMENT
BY
VIREN PATEL
APOORVA MULYE
AMANPREET SINGH SAINI
B.E BIOMEDICAL
SEMESTER VIII
UNDER THE GUIDANCE OF
Prof.
-
8/6/2019 Viren Pc Based
2/21
MAHATMA GANDHI MISSION'S
COLLEGE OF ENGINEERING & TECHNOLOGY
KAMOTHE,NAVI MUMBAI
CERTIFICATE
This is to certify that following students have satisfactorily completed
in Project work in Stage 1 of Bachelors Degree in Biomedical Engineering
Course conducted by University of Mumbai.
1)Viren D. Patel
2)Apoorva D. Mulye
3)Amanpreet Singh G. Saini
GUIDE EXAMINER
HEAD OF DEPARTMENT PRINCIPAL
-
8/6/2019 Viren Pc Based
3/21
ACKNOWLEDGEMENT
An in house project provides challenges quite unlike done in industry.
While the focus in industry is on efficiency and maximization of available
resources, a project done in the college under the watchful eyes of our
project guide ensures that we explore all possible permutations and
combinations in our quest to make a successful project. The focus is on
practical implementation of the theoretical knowledge gained over the last
four years.
In our quest, our project guide Mrs. Aarti Bokade deserves a specialthanks who allowed a group of imaginative students to experiment at will
and produce a project that is thereby a symbol of work put in the laboratory.
Though the duration of our project, the entire department of teachers took a
keen interest and helped us along. We would like to mention the co-
operation extended by Mrs. Aarti Bokade who allowed us a complete use of
the laboratory facilities and his valuable time and insights.
Last but not the least I thank everybody those who directly or indirectly
helped us in making this project a grand success.
-
8/6/2019 Viren Pc Based
4/21
ABSTRACT
OXYGEN SATURATIONThis report deals with the different types of pulse oximeter and
the design considerations taken when building high-end portablebedside monitors or mid and low-end battery-powered models. The
discussion covers light absorption, sensor calibration, and detailsspecifics within the transmit path leading to in-depth discovery of
photodiode, interface, signal conversion, and data conversioncharacteristics.
RESPIRATION RATEThis report explains a simple method for respiration rate measurement
using displacement transducer. This meter is used to monitor the respiration
rate, pulse rate and heart rate. It responds fast and is cost-effective comparedto conventional medical equipment. by using this, respiration rate can be
measured in the range 0-999 respirations/minute.
-
8/6/2019 Viren Pc Based
5/21
-
8/6/2019 Viren Pc Based
6/21
CHAPTER 1
INTRODUCTION
1.1 PULSE OXIMETRY
Pulse oximeters noninvasively measure or continuously monitor oxygen saturation in
arterial blood to ensure that there is sufficient oxygenation. Typical applications includepeople with respiratory problems who are under anesthesia, neonates, and critically ill
patients. In a pulse oximetry system, a clip with optical electronics is usually attached to
a finger, toe, or ear so that light can be transmitted through the skin from one side of the
clip and received on the other side with a photodiode. Good arterial blood flow isrequired to measure oxygen saturation.
The principle advantage of optical sensors for medical applications is their intrinsic safety
since there is no electrical contact between the patient and the equipment.
For patients at risk of respiratory failure, it is important to monitor the efficiency of gas
exchange in the lungs, i.e how well the arterial blood is oxygenated (as opposed to
whether or not air is going in and out of the lungs). Preferably, such information shouldbe available to clinicians of a continuous basis (rather than every few hours). Both of
these requirements can be met noninvasively with the technology of pulse oximetry.
The technique is now well established and is in regular clinical use during anaesthesiaand intensive care (especially neonatal intensive care since many premature infants
undergo some form of ventilator therapy). Pulse oximetry is also being used in the
monitoring of pulmonary
disease in adults and in the investigation of sleep disorders.
1.2 PRINCIPLE OF OPERATION
Operation is based on the light-absorption characteristics of hemoglobin in the blood.Oxygenated hemoglobin absorbs more infrared light than red light, and deoxygenated
hemoglobin absorbs more red light than infrared light.Thus, red and infrared LEDs in the
oximeter alternately transmit light, and a photodiode receives the light that is notabsorbed. The ratio of the red and infrared light received by the photodiode is used to
calculate the percentage of oxygen in the blood. Based on the pulsatile nature of arterialblood flow, the pulse rate and strength are also determined and displayed during the
measurement cycle.
1.3 Normal Findings
-
8/6/2019 Viren Pc Based
7/21
Basically, a saturation of 97% of the total amount of hemoglobin in the body is filled
with oxygen molecules. A range of 96% to 100% is generally considered normal.Anything below 90% could quickly lead to life-threatening complications. The margin
between "healthy" saturation levels (95-98%) and respiratory failure (usually 85-90%) is
narrow. If oxyhaemoglobin is low (below 90%) inadequate amounts of oxygen will reach
body cells.
Pulmonary Care Services Oxygen Protocol requires titrating oxygen to maintain an
SPO2 of 94% for most patients, and 92% for those who are CO2 retainers.
As a rule of thumb, respiratory failure usually occurs when saturation (SpO2) falls to
90%, although some patients with chronic respiratory disease may tolerate lowersaturations
Alarm limits should be set at a level that identifies any significant change in saturation.
Setting lower alarm limits of 90% may be appropriate when saturation is 95%, but
inappropriate if saturation is fluctuating at 90-91%. If setting alarm limits below 90%,nurses should be cautious about the very narrow margin remaining before respiratory
failure. Setting a lower alarm limit of 85% or less should always be avoided! Oxygendelivery to tissues, including vital organs, is likely to be inadequate at this level, and such
low saturations usually require urgent medical intervention (intubation and artificial
ventilation
1.4 APPLICATIONSResuscitation and pacing
1. ICU/CCU2. Emergency Room3. Immediate Care facilitiesBedside Patient Monitoring1. Emergency Room2. Post anesthesia3. Special procedures4. Out Patient Care5. Oral/Dental Clinics
1.5 RESPIRATION RATE
Respiratory rate (RR) is the number of breaths a living being, such as a human, takeswithin a certain amount of time (frequently given in breaths per minute).
There is only limited research on monitoring respiratory rate, and these studies have
focused on such issues as the inaccuracy of respiratory rate measurement and respiratory
rate as a marker for respiratory dysfunction.
1.6 PRINCIPLE OF OPERATION
-
8/6/2019 Viren Pc Based
8/21
The human respiration rate is usually measured when a person is at rest and simply
involves counting the number of breathes for one minute by counting how many timesthe chest rises. When you inhale and exhale air during respiration, this leads to movement
of ball left and right. The displacement of the ball is more for a healthy person.
1.7 NORMAL FINDINGS
Respiration rates may increase with fever, illness, OR other medical conditions. When
checking respiration, it is important to also note whether a person has any difficulty
breathing.Inaccuracies in respiratory measurement have been reported in the literature. One
study compared respiratory
rate counted using a 15 second count period, to a full minute, and found significant
differences in the rates.[citation needed]
Respiratory rates measurement in children under five years, for a 30 second or 60second period, suggesting the 60 seconds resulted in the least variability. Another study
found that rapid respiratory rates in babies, counted using a stethoscope, were 2050%higher than those counted from beside the cot without the aid of the stethoscope.[citation
needed] Similar results are seen with animals when they are being handled and not being
handledthe invasiveness of touch apparently is enough to make significant changes inbreathing.
Average respiratory rate reported in a healthy adult at rest is usually given as 12
breaths per minute (1260 Hz)[1][2] but estimates do vary between sources, e.g., 1220breaths per minute, 1014,[3] between 1618,[4]etc. With such a slow rate, more
accurate readings are obtained by counting the number of breaths over a full minute.
Average respiratory rates, by age:Newborns: Average 44 breaths per minute
Infants: 2040 breaths per minute
Preschool children: 2030 breaths per minute
Older children: 1625 breaths per minute
Adults: 1220 breaths per minute
1.8 APPLICATIONS
ICU/ICCU
Emergency Room
Cardiac patients
-
8/6/2019 Viren Pc Based
9/21
CHAPTER 2
BLOCK DIAGRAM
-
8/6/2019 Viren Pc Based
10/21
CHAPTER 3
HARDWARE
SENSOR 1:
CI-6535 is a Low Pressure Sensor.The CI-6535 Respiration Rate Sensor consists of theCI-6534A Low Pressure Sensor. The range of it isbetween 0 and 10 kiloPascals. The
resolution of the sensor is 0.005 kiloPascals (kPa) when used with a computer interface.
The output voltage from the sensor is +1.00 Volts when the pressure is 1 kiloPascal(kPa), and the output voltage is linear. Therefore, the output voltage should be +10.00
Volts at the top of the range (10 kPa).
Sensor 2:
The sensor head consists of a pulse oximeter two light-emitting diodes and aphotodetector. One LED emits in the visible range, the other in the infared. Each LED is
illuminated in turn. The detector may be placed to detect light transmission (through a
finger,for example) or light reflection from skin. The ratio of the detected light signals isproportional to the blood oxygen level. It is best if the same detector can be used for both
visible and infared signal detection.
ARM ARCHITECTURE ARM7TDMI Architecture 32-bit RISC-processor core 37 pieces of 32-bit integer registers 3 stages pipeline Cached (depending on the implementation) Von Neumann-type bus structure 8 / 16 / 32 -bit data types Two instruction set:
ARM instruction setTHUMB instruction set
Revision ARMv4T ARM processors are designed for good speed / power
consumption ratio and maximum code density.
-
8/6/2019 Viren Pc Based
11/21
CONTROLLER 16/32-bit ARM7TDMI-S microcontroller in a tiny LQFP64 package. 512KB of Flash memory 32KB of RAM+ 8KB shared with USB DMA Two 10bit ADC with total of 14 analog input channels One 10bit DAC Two 32bit timers/external event counters (with four CCP channels
each)
In-System/In-Application Programming (ISP/IAP) via on-chip boot-loader software
Watchdog timer Low power RealTime Clock (RTC) Two UARTs
I2C, SPI and SSP with buffering and variable data length Up to 21 external interrupt pins available. Power saving modes include Idle and Powerdown Brownout Detector Circuit 60 MHz maximum CPU clock available from programmable onchip
PLL with settling time of 100 ms
USB 2.0 Fullspeed compliant device controller
UART
-
8/6/2019 Viren Pc Based
12/21
LCD
SERIAL EEPROM(MEMORY)
-
8/6/2019 Viren Pc Based
13/21
SWITCH
POWER CIRCUIT
-
8/6/2019 Viren Pc Based
14/21
-
8/6/2019 Viren Pc Based
15/21
CHAPTER 4
SOFTWARE
-
8/6/2019 Viren Pc Based
16/21
-
8/6/2019 Viren Pc Based
17/21
-
8/6/2019 Viren Pc Based
18/21
-
8/6/2019 Viren Pc Based
19/21
-
8/6/2019 Viren Pc Based
20/21
CHAPTER 5ADVANTAGES,HURDLES &
FUTURE EXTENSIONS
ADVANTAGES One instrument instead of many Integrated logging of data and pre-servable Patients will receive higher quality care with great efficiency Reduced equipment cost Highly portable device(can be even kept in pocket)
HURDLES Motion artifacts Nail polish or artificial finger nails Inadequate blood flow Loosening of pressure belt Oxygen saturation cannot be measured in case of anemia
FUTURE EXTENSIONS
Measuring, diagnosing and displaying most of the parameters likeECG(simulation), Heart rate, temperature, Sugar level etc
Displaying all these parameters on all Mobilephone,PDAs,TABLETS
-
8/6/2019 Viren Pc Based
21/21
CHAPTER 6
REFERENCES
www.maxim-ic.com/an4671www.atmel.com
G J Tortora & N P Anagnostakos
Anatomy and Physiology, 6th edition, New York:
Harper-Collins, 1990, ISBN 006046669
Lauralee Sherwood, Fundamentals of Physiology:The Lancet. May 2,
1998. 351:13081311.
A Human Perspective, Thomson Brooks/Cole, 2006, William D.
McArdle, Frank I. Katch & Exercise Physiology: Energy, Nutrition,
andHuman Physiology, Lippincott Williams & Wilkins,
(http://books.google.com/books)
Brian S. Beckett, Illustrated Human and Social
Dr. David M. Broday, The Inhalation Dosimetry Biology,
Oxford: Oxford University Press, 1995,
Wikipedia: Pulse oximeter
http://en.wikipedia.org/wiki/Oximeter
Principles of Pulse Oximetry Technology
http://www.maxim-ic.com/an4671http://www.maxim-ic.com/an4671