w3051 automatic method development for routine … · mobile phase reservoirs: column salt water...
TRANSCRIPT
TO DOWNLOAD A COPY OF THIS POSTER, VISIT WWW.WATERS.COM/POSTERS ©2014 Waters Corporation
PURPOSE
To develop and validate a routine UPLC based
method for the separation of C-terminal lysine
truncation variants in a chimeric monoclonal
antibody based on IEX gradients generated by
automatic mobile-phase blending
Robert Birdsall*, Thomas E. Wheat, Julann Miller, and Weibin Chen Waters Corporation, 34 Maple Street, Milford, MA 01757
RESULTS
CONCLUSION
Automated method development using
Auto Blend Plus Technology shows highly
reproducible performance.
The integrated platform is suited for robust
and routine method development and can
greatly improve the separation results and
enhance the throughput of the analysis.
W3051 AUTOMATIC METHOD DEVELOPMENT FOR ROUTINE CHARGE VARIANT ANALYSIS OF BIOTHERAPEUTICS BY ION-EXCHANGE CHROMATOGRAPHY
Mobile phase reservoirs:
column
Salt Water
Base Acid
Figure 1. Reservoir Setup. Illustration of a typical Auto Blend
Plus™ Technology reservoir setup.
Day Peak 1 Area
Peak 2 Area
Peak 3 Area
Peak 4 Area
Peak 5 Area
Total
1 252260.0 788195.0 504001.0 296130.0 1052614.0 2893200.0
2 203498.0 660040.0 427519.0 237894.0 940898.0 2469849.0
3 214836.0 686459.0 437254.0 255744.0 974813.0 2569106.0
mean 223531.3 711564.7 456258.0 263256.0 989441.7 2644051.7
SD 25517.4 67665.7 41632.2 29835.9 57276.6 221402.7
% RSD 11.42 9.51 9.12 11.33 5.79 8.37
Test
pH
Buffer
mix 1
Buffer
mix 2
Buffer
mix 3 Mean
Std.
Dev.
%
RSD
5.13 5.10 5.02 5.10 5.07 0.05 0.91
6.12 6.19 6.05 6.19 6.14 0.08 1.32
7.10 7.23 7.06 7.23 7.17 0.10 1.37 Figure 3. Robust Separations with Salt Gradients. The Auto
Blend Plus™ Technology delivers robust separations of C-terminal lysine truncation variants in a chimeric mAb using a
25 minute 25 mM to 65 mM NaCl gradient.
The effects of protein mass load was evaluated
for robustness. The reproducible retention
times and linear response in the Area vs. Mass
Load are indicators of a robust method.
METHODS
An optimized ion exchange method for the
characterization of C-terminal lysine truncation
variants of a chimeric monoclonal antibody was
developed that incorporates a solvent
management system using pure solvents and
concentrated stocks (Figure 1) which will
herein be referred to as Auto Blend Plus™.
Conditions:
LC System: ACQUITY UPLC H-Class System with Auto Blend Plus Technology
Detector: ACQUITY UPLC TUV (Flow cell 5 mm) =280 nm
Column: Protein-Pak Hi Res SP, 4.6x100 mm, 7µm
Temperature: 25 °C Sample Temperature: 4 °C
Injection Volume: 3 µL Flow Rate: 0.500 mL/min
Mobile Phase A: 100 mM MES monohydrate Mobile Phase B: 100 mM MES sodium salt
Mobile Phase C: 1000 mM sodium chloride Mobile Phase D: 18 MΩ H2O
Buffer Conditions: 20 mM MES buffer
Automated algorithms for solvent delivery
Figure 6. Intra-system pH Precision. Separation of C-terminal
lysine truncation variants with three different preparations of MES buffer using Auto Blend Plus™ Technology. Covariance
ranged between 0.91 % – 1.37 % R.S.D.
The optimized method was tested for
robustness using pH, precision, and mass load
as test parameters.
Figure 2. Empirical Table. An example of the empirical
table feature for the Auto Blend Plus™ Technology. Nine
standards were used to calibrate the Auto Blend Plus™
Technology system over a pH range of 5.13 7.09.
The AutoBlend Plus™ Technology delivers
robust separations of C-terminal lysine trunca-
tion variants in a chimeric mAb using salt
gradients (Figure 3) or pH gradients (Figure 4).
The separation method were developed by
systematically varying the salt or pH gradient
generated by The AutoBlend Plus algorithm
(Figure 5).
Figure 4. Auto Blend Plus™ Technology Flexibility. The Auto
Blend Plus™ Technology delivers comparable separations of
C-terminal lysine truncation variants in the same chimeric
mAb using a 15 minute pH gradient from pH 6.50 to pH 7.20 in a 40 mM MES buffer.
Figure 5. The Auto Blend Plus™ Interface. An example of
the Auto Blend Plus™ Technologies intuitive interface for
programming pH or salt gradients. The software algorithm
automatically calculates the percentage of acid and base required for the specified pH or ionic strength.
Intra-system pH precision was tested across
three different MES buffer systems and
determined to be within acceptable range.
The Auto Blend Plus Technology was evaluat-
ed over 40 injections to test reproducibility.
Covariance of individual and total area were
determined to be within acceptable range.
Buffer
mix 1
Buffer
mix 2
Buffer
mix 3
Retention Time [min]
Table 1: Intra-system pH results
+1 Lys
+2 Lys
Retention Time [min]
Chimeric
monoclonal
antibody
+0 Lys
+1 Lys
Retention Time [min]
+2 Lys
Chimeric
monoclonal
antibody
Salt
gradient
+0 Lys
pH
gradient
Day 1
(injection 1)
Day 2
(injection 20)
Day 3
(injection 40)
Table 2: Method precision results
Figure 7. Method Reproducibility. Separation of charge vari-
ants of a chimeric monoclonal antibody at three different time intervals over three days. Integration intervals represent the
five peak areas calculated in Table 2. Individual peak area co-variance was determined to be 11.42 %, 9.51 %, 9.12 %,
11.33 %, and 5.79 %. Total peak area covariance was deter-mined to be 8.37%.
Precision pH Mass load
20 µg
60 µg
100 µg
140 µg
180 µg
Protein Mass Load (µg)
( x1
06
)
To
tal
In
teg
rate
d A
rea R2 = 0.9999
Retention Time [min]
A)
B)
Figure 8. Mass Load Effects. A) Reproducible retention times
were observed over a 9-fold increase in mass load ranging from 20-180 µg of protein. B) The total peak area was
integrated as a measure of precision as shown in the plot of integrated area vs. protein mass load.
100 mM
NaH2PO4
100 mM
Na2HPO4
1000 mM
NaCl H2O
measure pH populate table at
initial setup
flow
low medium
high
measure pH
Protocol: 1) Prepare stock solutions
2) Cross calibrate AutoBlend Plus™
collect
3) Prepare and collect test solutions from instrument
4) Compare
Prepare solutions
The AutoBlend Plus algorithm automatically
calculates the percentage of acid and base
required for the specified pH from the known
pKa of the buffer system or by using an
empirical calibration table (Figure 2).
table populated once at
initial set-up