what is needed for high resolution sem? za small probe size zhigh beam current za mechanically...
TRANSCRIPT
What is needed for High Resolution SEM?
A small probe size High beam currentA mechanically
stable microscope and a quiet lab environment
A skilled operator
Lens performance
The probe size is determined by the aberrations of the lens
The magnitude of the aberrations vary with the focal length of the lens - which is about equal to the working distance
Some lens’ designs are more capable than others at combining both high performance and good sample access
Lens performance cont.
Aberrations including spherical and chromatic are correctable to varying degrees
Corrections depend on a variety of factors including pole piece quality, aperture size, aperture, angle and electron wavelength
Pinhole Lens
The original SEM lens - designed to produce no magnetic field in the sample chamber
Good sample access Long focal length and a
big working distance so high aberrations
Poor EM screening Asymmetric SE
collection due to position of ET
ET
sample
Immersion Lens
Short focal length - so low aberrations
Good EM screening Very stable specimen
mounting in lens Symmetric SE
collection using the through the lens (TTL) detector system
Restricted to small samples (3mm disc)
det
magnetic field
B
Snorkel Lens
Short focal length - so low aberrations and high performance
Good EM screeningThe sample is
outside the lens so there is no limitation on the size of the specimen
TTL
ETMagnetic fieldprojected out of lens
S4700 Snorkel Lens
Up to 45 degrees of sample tilt even at short WD and permits EDS operation at WD of 12mm
Biased deflector plates optimize SE collection for either or both detectors
Improved magnetic screen and stronger stigmators can image magnetic samples at all WD
The lens also acts to filter the SE signal to the TTL
S4700 lens configuration Excitation - 1000 amp.turns
S4700 detector
Snorkel lens permits multiple detectors to be used
In-lens (TTL) detector gives a shadow free image with ultra-high topographical resolution. Super efficient
Lower (ET) detector gives SE images with material contrast information and high efficiency at high tilt angles
These detectors can be used separately or combined as desired for maximum flexibility
Snorkel lenses allow multiple detectors
Detector Flexibility
DRAM with both Upper and Lower detectors MO layer in BSE mode
(DRAM stands for dynamic random access memory, a type of memory)
Multiple imaging modes provide flexibility and problem solving power
What determines spot size? The spot size depends on
the beam energy, WD, and the final aperture convergence angle
Performance improves with higher energies
On the S4700 the aperture size is set automatically
Changing the CL (spot size) does not affect resolution much Variation of probe size with energy
and beam convergence for S4700
Working Distance
Working distance is the most important user controlled parameter
Always use the smallest WD that is possible for a given specimen
Note also that the image resolution is almost independent of the beam energy
Imaging
Microanalysis
Beam current
Typical contrast levels are 3-10% on most samples
Improving contrast lowers required IB , beam current, and improves resolution
Increase IB by raising the tip emission current from 10A to 20 or 30 A if necessary
Resolution
The pixel size is equal to the CRT pixel size divided by the actual magnification e.g a 100µm pixel at 100x gives 1µm resolution
Probe size only limits resolution at high magnifications Image at 1kx magnification
has 0.1µm pixel resolution
Image Content
SE1 - high resolution
SE2 - low (BSE) resolution
SE3 - tertiary signal, interactions of the BSE with the pole piece and chamber walls
ET sees 40% SE3, 45% SE2, 15% SE1
TTL sees 75% SE2 and 25% SE1
SE escape
Lens Detector
ET
TTL
SE1SE2 SE3
SE1
SE2
BSE
SE1/SE2 interaction volumes
The SE1 signal comes from a few nm area at all energies
The SE2 signal comes from an area that can be up to a few microns in diameter at high energies
Pixel size and SE2
At low and medium magnifications the pixel size ( a few µm) is comparable with SE2 interaction volume
So the image is mostly from the BSE generated SE2 component
The SE1 are not a significant contributor
SE2area
pixel
Medium magnification
Medium magnification images have a resolution limited by SE2 interaction volume
SE and BSE images will look similar but not necessarily identical Image at 20kx - 50Å
pixels
High magnification images
Field of view is about size of the SE2 interaction volume so that signal remains about constant as beam scans
The pixel size is about equal with the SE1 area so the SE1 component now provides the image detail
field of view
pixel
Pixels - a summary
High resolution requires the use of a high magnification to keep the pixel size at a small enough value not to limit the resolution
High resolution at high beam energies also requires a high magnification so as to separate the SE1 signal from the lower resolution SE2 signal
High resolution imaging On the S4700
imaging in SE mode with a resolution into the nanometer range is readily possible
What is the ultimate resolution limit?
Optical performance, signal origination, and current to establish sufficient signal quality
Imaging a 10nm thick oxide layer
How good is SE resolution?
The production of SE occurs over a finite volume of space
The initial SE event produces additional SE and so on, leading to a diffusing cloud of SE around the impact point
How far do they travel? Depends on the MFP (mean free path)
SE resolution
The diffusion effect is visible at the edges of a sample as the ‘bright white line' due to extra SE emission
The width of this line is a measure of the SE MFP
The presence of this SE1 edge effect sets an initial limit to the achievable SE image resolution
Molybdenum tri-oxide crystals Hitachi S900 25keVSE diffusion volume
Classical resolution limit
When the object is large its edges are clearly defined by the ‘white lines’
But as the feature reaches a size which is comparable with the edge fringes begin to overlap and the edge contrast falls
20nm
Width =
10 nm
Classical resolution limit
When the feature size is equal to or less than the edge lines overlap and the object is not resolved at all since it has no defined size or shape
This is Gabor’s resolution limit for SE imaging
The resolution in SE mode therefore depends on the value of
Particle contrast
5 nm
width =
High Resolution Imaging
On a high atomic number, very dense , material such as tungsten the SE MFP is only a nanometer or so
So a spatial resolution of about 1nm is likely to be possible
In fact ...
“Lattice” fringes
In this image by Kuroda et al (J.Elect.Micro 34,179, 1985) fringe structures with a spacing of 1.4nm are clearly visible in the SE image
This resolution is consistent with the diffusion model for SE production with =1nm
Image recorded at 20keV on an Hitachi S-900 FEGSEM
The probe size for this image was about 0.9nm
SurfaceSurfaceConfigurationConfiguration
In other samples...
When an object gets small enough to be comparable with then it becomes bright all over and the defining edges disappear.
For low Z, low density materials, this can happen at a scale of 5-10nm
Carbon nanotubes
edge brightness
no edges
The resolution limit
The resolution of the SEM in SE mode is thus seen to be limited by the diffusion range of secondary electrons, especially in low Z materials
In addition the signal to noise ratio is always worse for the smallest detail in the image
Improving the resolution
Improving SEM resolution therefore requires two steps:
minimizing or eliminating the spread of secondary electrons
improving the signal to noise ratio so that detail can be seen
Improving the S/N ratio
Use a metal coat as all metals give more SE than carbon
SE yield tends to rise with Z value
But high Z materials are denser and cause more scatter
Usually consider Cr, or Ti as best choices but W, Pt are also good
Computed SE1 yield at 2keV
Particulate Coatings
Au produces very big particles (30nm)
Au/Pd and W make much smaller (3nm) particles
These have a very high SE yield
Can be deposited in a sputter coater
Coatings are stable Good below 100kx
3nm of Au/Pd at 100kx
Decoration
In some cases the sputtered particles decorate active features on a structure, making them more visible
High Z materials, such as tungsten also permit BSE imaging
Tungsten decorated T4 polyheads 25nm ring diameter 30keV Hitachi S900
Bypassing the limit
Since metals have much lower than carbon, and a higher SE yield, a thin metal film coating on a low Z, low density sample effectively localizes all SE production within itself. The resolution now is a function of the film thickness only and not of
Works even with very thin metal films (few atoms thick)
Can exploit this effect to give interpretable contrast at high resolution
Low SE yield
High SE yield
width film evenwhen <
Mass thickness contrast
The SE1 yield varies with the thickness of the metal film
This effect saturates at a thickness equal to about 3
The conformation of the film to surface topography thus provides contrast 1nm 2nm 3nm
Film thickness
SE
Yield
bulk value
mass thickness variation
Metal builds contrast
The SE localization in the film provides edge resolution
The mass thickness effect gives extra contrast enhancement
The feature is now truly ‘resolved’ since its size and shape are visible
5nm low Z object2nm metal film
Beam positionSE profile with metal film
SE profile without metal
SE
Cr coatings
Cr films are smooth and without structure even at thicknesses as low as 1nm
The mass thickness contrast resolves edges and make the detail visible down to a nanometer scale
The high SE yield of the Cr improves the S/N ratio
However these coatings are not stable - so use Cr coated samples immediately after they have been made
AIDS virus on human cells 500kx 2nm Cr at 20keV Hitachi S900
Coating Summary
Coatings are an essential part of the technique of high resolution SEM because they generate interpretable contrast, improve resolution, and enhance the S/N ratio
Thin coatings are better than thick coatings - do not make your sample a piece of jewelry
Below 100kx particulate coatings are superior because of higher SE yields
Above 100kx use chromium or titanium MRC lab uses Au/Pd coatings on most samples Carbon is a contaminant not a coating
Getting the most from your SEM
Alignment is crucial. Check aperture alignment every time you change areas or imaging conditions and ensure that the stigmators are properly balanced
Minimize vibrations by choice of SEM location. Move pumps away etc.
Keep the room quiet, noise dampening material on the walls.
Check for stray fields. Remove fluorescent lights and dimmer controls.
Keep computer monitors away - use flat screens
Beware of ground loops
Clean Power
Many cases of ‘jaggies’ are due to dirty mains lines not EM pickup
Check waveform at your wall plug
Use clean power from a UPS for critical electronics
Avoids surges
zerocrossings
AC line andEM Field
raster issynchronizedwith field
switching spikes
raster is nownot synchronizedwith field
zerocrossings
Operating tips
Allow the SEM to thermally stabilize and the cold finger to cool down before attempting high resolution - this may take > 1 hour (seldom used at MRC)
Use the stage lock - but don’t forget to turn it off before unloading sample
Use the beam shift rather than stage motion - but remember to recenter the beam before taking a critical image
Look for the scan speed which minimizes ‘jaggies’ when viewing the image live
Getting the best image
Whenever possible take a single slow speed scan rather than accumulating multiple high speed scans
This eliminates blurring due to drift, and distortions in the video amplifier chain and usually produces a higher signal to noise ratio and better contrast32 high speed
framessingle 20
second scan