2
Integrated safety functions: Safety deactivation of the controllers with digital input • Configurable feed delay • Monitoring sample water flow Depending on whether the pH measuring module is to be used on its own or with other sensor measuring modules, there is a choice of flow cell modules (pressu- rised and non-pressurised). For example, the pH mea- suring sensor can be used with the DEPOLOX® 5 mea- suring cell to measure free chlorine, chlorine dioxide, ozone and potassium permanganate. Other sensors are at hand, e.g. to measure conductivity, redox vol- tage and conductivity as well as membrane-covered sensors for free chlorine, total/combined chlorine, chlo- rine dioxide and ozone (see separate data sheets). The “multi-sensor” integrated into the flow cell module monitors the constant sample water flow, measures the sample water temperature and provides wide-area equipotential earthing. Advantages Modular design for the MFC and pH measuring module and up to four additional sensors Four different control types to choose from, also with automatic set-point adjustment Timer-controlled dosing using programmable time switch contacts Easy to configure with selectable, preset process-adjusted applications Easy to connect to process visualisation systems via an OPC server; to web technology via a ChemWeb server Water Technologies Wallace & Tiernan MFC-pH Multi-function controller to measure the pH value The Wallace & Tiernan MFC-pH is a possible system alternative to the MFC multi-function controller. The control, measuring and analysis features of the MFC make it flexible enough for a variety of possible applications in drinking, process and waste water systems. Dosing pumps, electric stroke positioners and continuous actuators can be con- trolled directly according to the measured pH value. Up to four different control types are available for two measured settings. Choose between ratio con- trol, single feedback closed-loop control, combined control and setpoint trim, which is the combination control with adaptive nominal value. Five preset process applications assist in the configuration pro- cess and can also be used to select the control types. The MFC can therefore easily be configured to the onsite requirements. Design and function The MFC-pH multi-function controller is composed of the MFC electronic module and a flow cell module together with the pH sensor measuring module. The MFC electronic module displays the measured data and functions as the control unit. The MFC mo- dule is equipped with different controller outputs, alarm contacts and integrated safety functions. The controller outputs for positioners, pumps, and ana- logue signals (mA) are easily configured in the set- up menu. The alarm contacts (up to a max. of eight contacts) are application-dependant and freely con- figurable. A multiple assignment of safety functions such as a general alarm for monitoring limit values, sample water failure, for example, is possible as well.

Upload: marlon-isaac-cortes

Post on 26-Dec-2015

1 views

Category:

Documents


0 download

DESCRIPTION

n

TRANSCRIPT

Page 1: w_mfc_ph

Integrated safety functions:• Safety deactivation of the controllers with digital input• Configurable feed delay• Monitoring sample water flow

Depending on whether the pH measuring module is tobe used on its own or with other sensor measuringmodules, there is a choice of flow cell modules (pressu-rised and non-pressurised). For example, the pH mea-suring sensor can be used with the DEPOLOX® 5 mea-suring cell to measure free chlorine, chlorine dioxide,ozone and potassium permanganate. Other sensors are at hand, e.g. to measure conductivity, redox vol-tage and conductivity as well as membrane-coveredsensors for free chlorine, total/combined chlorine, chlo-rine dioxide and ozone (see separate data sheets).

The “multi-sensor” integrated into the flow cell modulemonitors the constant sample water flow, measuresthe sample water temperature and provides wide-areaequipotential earthing.

Advantages

• Modular design for the MFC and pH measuringmodule and up to four additional sensors

• Four different control types to choose from, also with automatic set-point adjustment

• Timer-controlled dosing using programmable time switch contacts

• Easy to configure with selectable, preset process-adjusted applications

• Easy to connect to process visualisation systems via an OPC server; to web technology via a ChemWeb server

Water Technologies

Wallace & Tiernan

MFC-pHMulti-function controller

to measure the pH valueThe Wallace & Tiernan MFC-pH is a possible systemalternative to the MFC multi-function controller.The control, measuring and analysis features of the MFC make it flexible enough for a variety of possible applications in drinking, process and waste water systems. Dosing pumps, electric strokepositioners and continuous actuators can be con-trolled directly according to the measured pH value.Up to four different control types are available fortwo measured settings. Choose between ratio con-trol, single feedback closed-loop control, combinedcontrol and setpoint trim, which is the combinationcontrol with adaptive nominal value. Five presetprocess applications assist in the configuration pro-cess and can also be used to select the controltypes. The MFC can therefore easily be configuredto the onsite requirements.

Design and function

The MFC-pH multi-function controller is composed ofthe MFC electronic module and a flow cell moduletogether with the pH sensor measuring module.

The MFC electronic module displays the measured data and functions as the control unit. The MFC mo-dule is equipped with different controller outputs,alarm contacts and integrated safety functions. Thecontroller outputs for positioners, pumps, and ana-logue signals (mA) are easily configured in the set-up menu. The alarm contacts (up to a max. of eightcontacts) are application-dependant and freely con-figurable.

A multiple assignment of safety functions such as ageneral alarm for monitoring limit values, samplewater failure, for example, is possible as well.

Page 2: w_mfc_ph

Due to continual product development and improvement,

certain specifications may change without prior announcement.

SiemensWater TechnologiesWallace & Tiernan GmbHAuf der Weide 1089312 Günzburg GERMANY

[email protected] 50.580-2DE 3.06

© Siemens AG 2006

Sensor measuring module pH

The measuring module consists of a sensor, sensor ca-ble and a pre-calibrated and tested card.

Protection category sensor plug connector: IP 66

Single-rod electrodeMeasuring range: pH 0 to 12, can be freely scaled in 1 pH value steps; resolution 0.01 pH

Flow cell modules

The following flow cell modules (pressurised/non-pressurised) are available:• DEPOLOX®5• VariaSens• Flow-through adapters

The flow-through module DEPOLOX®5 contains themeasuring module DEPOLOX®5 to measure free chlori-ne, chlorine dioxide, ozone and potassium permanga-nate. The VariaSens flow cell module is designed forusing up to two membrane sensors. For each one, there are other similar measuring modules to choosefrom. If the conductivity sensor module is to be used on its own or together with a redox measurement, aflow-through adapter can also be used.

Integrated into the flow cell modules DEPOLOX®5 andVariaSens are:

Flow control valve:• Controlled sample water flow: 33 l /h• Control range: 0.2 … 4.0 bar at valve inlet• Back pressure: max. 1.5 bar with a pressurised system• Permissible sample water temperature: max. 50°C• Minimum conductivity: 200 μS/cm

“Multi-sensor”:• Monitoring the correct sample water flow• Switching point: 21 l /h +/- 3 l /h,

switching hysteresis: 2 l /h• Measuring the sample water temperature with sensor

Pt 1000 for temperature compensation of the pH value and chlorine measuring where necessary)

• Wide-area sample water earthing with stainless steel sleeve (equipotential earthing)

Sample water connections:PVC hose 6 x 3 mm or PE hose 6 x 1 mmHose connector adaptors to 1/2“ threaded hose connection

Weight (incl. packaging): approx. 2 kg

Dimensions (W x H x D): 215 x 375 x 155 mm

MFC electronic module

Display:Back-lit LCD graphic display Resolution 240 x 64 pixels

Measurement inputs:1 x pH, three additional module slots galv. isolated up to 50 V relative to earth1 x dosing quantity display (feedback signal of positioner)

Digital inputs:1 x digital input for monitoring sample water1 x external stop1 x freely definable, e.g. controller stop, operatingmode switch, external set-point

Output contacts:For max. eight freely definable alarm contacts / generalfault messages as well as controller outputs for themeasured parameters

Special outputs:Time-controlled contact, e.g. timed dosing

Each output contact is visualised by a signal LED,max. 1250 VA to 250 V AC, max. 150 W to 220 V DC

Analogue outputs (optional):4 x 0 /4 … 20 mA, freely configurableload ≤ 1000 Ohm, accuracy < 0.5 % FSgalv. isolated up to 50 V relative to earth

Interfaces:RS 232 for direct printer control or firmware updateRS 485 to connect to ChemWeb server,Wallace & Tiernan OPC server, CMS software Infrared interface for remote calibration with the Wallace & Tiernan photometer P42 i-cal

Power supply:200 – 240 V AC ± 10 %, 50 – 60 Hz, 30 VA100 – 120 V AC ± 10 %, 50 – 60 Hz, 30 VA24 V DC ± 20 % 30 W

Ambient temperature: 0 … 50°C

Protection category: IP 66

Tests and marks:Conform to CE (89/336/EEC)EMC tests acc. to EN 61326Electric safety acc. to EN 61010

Weight (incl. packaging): 5.5 kg