wstf supports an extensive number of test programs

Upload: pritheemadan

Post on 03-Jun-2018

214 views

Category:

Documents


0 download

TRANSCRIPT

  • 8/11/2019 WSTF Supports an Extensive Number of Test Programs

    1/4

    WSTF supports an extensive number of test programs, most of which are instrumented for temperature, pressure,

    load, acceleration, and other measurements. The WSTF Measurement Standards and Calibration Laboratory

    ensures that the instruments used to make these measurements are accurate by calibrating them against recognized

    standards. A two-level tier of standards is used. The first level contains high-accuracy reference standards. These

    standards are traceable to the National Institute of Standards and Technology (NIST), and their accuracy is known in

    terms of national standards. Reference standards are highly stable and repeatable; for example, measurement

    uncertainty of the voltage standard is less than 0.5 ppm. The second level of standards is a group of 80+ working

    standards. These standards are calibrated by the reference standards and used to calibrate customer instruments in

    25 major measurement disciplines.

    The Measurement Standards and Calibration Laboratory has experience in providing support to NASA, other

    government agencies, and private industry. Current customers include NASA, U.S. Air Force, and U.S. Army. The

    Measurement Standards and Calibration Laboratory is certified to meet the ISO 9001 Standard.

    Mechanical Laboratory

    Mechanical calibration is primarily based on mass and length, both of which exist in the form of artifacts. The other

    mechanical measurements are derived from these basic units. Temperature is defined by the ITS-90 (International

    Temperature Scale of 1990) in terms of the melting and freezing points of pure materials as measured by a

    standardized platinum resistance thermometer.

    Physical Standards: WSTF can calibrate mass within the range of 0.5 micrograms to 30 kilograms, and length in a

    range from 50 microinches to 120 inches (1.3 m to 3 m).

    Pressure Calibration Laboratory:WSTF has two deadweight pressure systems: 1000 psi (7 MPa) and 10,000 psig

    (70 MPa). They operate using the basic definition of pressure-force per unit area. In this case, weights are placed on

    a piston giving pounds per square inch (psi). Using very small area pistons, it is possible to generate 10,000 psi (70

    MPa) without placing 10,000 lb (4 Mg) of weight on the piston. The 1000 psi system can be operated in absolute

    (vacuum reference) or gauge (atmosphere reference) mode.

    Straight Pressure Calibration Stations: WSTF has five stations with four different pressure ranges: 0 to 30 in. of

    water (7 kPa), 0 to 1000 psia (7 MPa) (one manually operated and one fully automated), 0 to 10,000 psia (70 MPa),

    and 0 to 30,000 psig (21 MPa).

    Differential Pressure Station: WSTF has the capability of holding a base pressure of up to 250 psig (1.7 MPa) while

    measuring differentials in the range of 0 to 100 psi (700 kPa), or we can do differentials vs. atmospheric pressure up

    to 1,000 psig (7 MPa).

    Flow Calibration Laboratory

  • 8/11/2019 WSTF Supports an Extensive Number of Test Programs

    2/4

    Liquid (Incompressible):The liquid used is water, and WSTF can measure flows from 0.4 to 400 gal/min (1.5 to

    1,500 L/min).

    Gas (Compressible): WSTF has the capability of flowing nitrogen, helium, compressed air, or anything

    nontoxic/nonflammable that can be supplied in a standard DOT pressure bottle in the lower flow ranges: 0.0008 to 12

    SCFM (0.37 to 5,400 cm3/s). In the higher flow ranges up to 200 SCFM (90,000 cm3/s) only nitrogen is available, but

    WSTF can test with any nontoxic/nonflammable gas the customer can supply in sufficient quantity/pressure. One

    customer supplied a helium tube bank semitrailer for high flow calibrations.

    Temperature and Humidity Laboratory

    Temperature: WSTF's temperature lab uses Standard Platinum Resistance Thermometers (SPRT) calibrated to ITS-

    90 as the primary standard, and can calibrate in the range from -197 C (liquid nitrogen) to 500 C.

    Humidity: WSTF's humidity standard and chamber can calibrate in the range of 1 to 98 percent relative humidity,

    within a temperature range of 15 to 30 C (normally 20 to 23 C).

    Force and Tool Calibration: We can calibrate force measuring devices in tension or compression up to 50,000 lbf

    (220 kN). We can also calibrate cable tensiometers vs. weights.

    Torque Wrenches: WSTF can calibrate torque wrenches in the range of 3 in-oz to 1,000 ft-lb (0.02 to 1,400 N-m).

    Crimpers and Strippers: WSTF can gauge, inspect, and conduct pull tests for a wide range of these tools.

    Vacuum Calibration Laboratory: WSTF has a vacuum calibration system, with a range from 1000 torr (100 kPa)

    (760 torr is atmospheric pressure) down to 10-7 torr (10 Pa).

    Software Engineering: Computer-aided calibration systems are used extensively at WSTF. Both off-the-shelf and

    custom designed, written by in-house staff programmers, are utilized. Electronic data can be provided upon customer

    request.

    Electrical Laboratory

    The Electrical Calibration Laboratory provides calibration and repair services for a wide variety of equipment types

    including:

    Resistance and capacitance decade boxes

    Digital and analog multimeters

    Power supplies

  • 8/11/2019 WSTF Supports an Extensive Number of Test Programs

    3/4

    Electronic loads

    Thermocouple simulators and readouts

    Current shunts

    Digitizing and analog oscilloscopes

    Microwave signal generators Spectrum analyzers

    Power sensors

    Equipment outside of WSTF capabilities are outsourced to the manufacturer or other government laboratories,

    allowing customers to use WSTF's laboratory as a single source for all of their calibration support requirements.

    Expense statements are provided quarterly to all non-WSTF customers.

    Microwave Station

    Electrical Laboratory: This station is used to calibrate microwave signal generators, spectrum analyzers,

    attenuators, frequency counters, amplifiers and other associated equipment. The system consists of measuring

    receiver, mixer, local oscillator, microwave signal generator, and spectrum analyzer. The system uses automated

    calibration software for most of the microwave signal generators calibrated by WSTF. Software is also used for most

    attenuator and spectrum analyzer calibrations.

    Scope Station:All of the oscilloscopes calibrated by WSTF are done at this station. The standard is a calibration

    generator for timing and amplitude accuracy, and a signal generator for bandwidth measurements. This system is

    controlled by MET/CALsoftware procedures. Seventy percent of the oscilloscopes are calibrated using

    MET/CAL

    procedures.

    Digital MultiMeter (DMM) Station: There are two calibration standards at the DMM station: a multifunction calibrator

    and a multiproduct calibrator. A custom subroutine was written by WSTF to allow MET/CALto control the

    multifunction calibrator, which is used to calibrate all of the benchtop multimeters of 5-1/2 digit or greater resolution.

    Seventy-five percent of these meters can be calibrated in a fully automated fashion using MET/CALprocedures

    written by WSTF. The multi-product calibrator is used for the rest of our DMM workload, along with some

    thermocouple readouts and simulators. Ninety percent of the handheld DMMs are calibrated using

    MET/CALprocedures.

    Reference Standards: WSTF Measurement Standards and Calibration Laboratory reference standards are

    maintained in the shield room. The volt is represented by a bank of three voltage cells. They are calibrated by the

    NASA Portable Josephson Array Voltage Standard (PJAVS). The PJAVS returns to WSTF approximately every 6 to 9

    months. The ohm is represented by an Thomas-type1-ohm standard resistor. This resistor was originally acquired in

    1964, and since then has been the WSTF ohm standard. The value of this resistor is maintained by annual

    calibrations by higher laboratories, primarily NIST and the Primary Standards Laboratory at Sandia National

  • 8/11/2019 WSTF Supports an Extensive Number of Test Programs

    4/4

    Laboratories. Additionally, WSTF is participating in the NASA Ohm Measurement Assurance Program administered

    by Kennedy Space Center.