X-ray polarimetry in Xenon gas filled detectors G.S. Botte 1 , L.P.M.M. Carita 1 , A.L. Gouvêa 2 , T.H.V.T. Dias 1 , J.F.C.A. Veloso 2 , P.J.B.M. Rachinhas 1 , R.M. Curado da Silva 1 , F.P. Santos 1 , L.M.N. Távora 1 , C.A.N. Conde 1 and A.D. Stauffer 3 1 Departamento de Física, Universidade de Coimbra, Portugal 2 Departamento de Física, Universidade de Aveiro, Portugal 3 Physics and Astronomy Department, York University, Toronto, Canada MPGD2009 - 12-15 June 2009, Crete, Greece

Upload: stacey

Post on 13-Jan-2016

60 views

Category:

Documents


0 download

DESCRIPTION

X-ray polarimetry in Xenon gas filled detectors. G.S. Botte 1 , L.P.M.M. Carita 1 , A.L. Gouvêa 2 , T.H.V.T. Dias 1 , J.F.C.A. Veloso 2 , P.J.B.M. Rachinhas 1 , R.M. Curado da Silva 1 , F.P. Santos 1 , L.M.N. Távora 1 , C.A.N. Conde 1 and A.D. Stauffer 3 - PowerPoint PPT Presentation

TRANSCRIPT

Page 1: X-ray polarimetry in  Xenon gas filled detectors

X-ray polarimetry in Xenon gas filled detectors

G.S. Botte1, L.P.M.M. Carita1, A.L. Gouvêa2, T.H.V.T. Dias1, J.F.C.A. Veloso2, P.J.B.M. Rachinhas1, R.M. Curado da Silva1, F.P.

Santos1, L.M.N. Távora1, C.A.N. Conde1 and A.D. Stauffer3

1 Departamento de Física, Universidade de Coimbra, Portugal

2 Departamento de Física, Universidade de Aveiro, Portugal

3 Physics and Astronomy Department, York University, Toronto, Canada

MPGD2009 - 12-15 June 2009, Crete, Greece

Page 2: X-ray polarimetry in  Xenon gas filled detectors

Motivation

unpolarized X-rays:

linearly polarized:

cos)sin2

1(cos

2

11

42

2

Pd

d

Photoelectric effect is very sensitive to polarization since photoelectron angular distributions become preferentially aligned with the polarization vector;

including 1st order non-dipole corrections δ and γ to the dipole approximation

(θ is the polar angle relative to the x-ray propagation direction, θ and φ are the polar and azimuthal angles relative to the polarization direction, β is the dipole asymmetry parameter and P2 is the 2nd Legendre polynomial)

cossin)cos(cos 1 4

22

P

d

d

Derevianko et. al., At.Dat.Nucl.Dat. Tables 73 (1999) 153

Trzhaskovskaya et. al., At.Dat.Nucl.Dat. Tables 92 (2006) 245

Page 3: X-ray polarimetry in  Xenon gas filled detectors

Motivation

Why?: great interest for x-ray astronomy & astrophysics

Polarization as a new observational parameter improves the characterization of astronomical x-ray emission sources

Our Goal: Observation of the polarization-induced anisotropy in the profiles of the electron clouds when polarized X-rays (~5-20 keV) are absorbed in xenon

The profiles of the electron clouds produced by photo-ionization events can be used to probe the polarization;

Page 4: X-ray polarimetry in  Xenon gas filled detectors

Summary

Monte Carlo simulation

Experimental measurements

Polarized X-ray source

Gridded-Microstrip Gas Chamber (G-MSGC)

Micro-Hole & Strip Plate (MHSP)

Conclusions

Page 5: X-ray polarimetry in  Xenon gas filled detectors

Monte Carlo simulation: Model

Incident polarized X-rays

Photoelectron absorption:

photoionization & anisotropic emission of photo-electrons

Cascade decay of excited Xe-ions,

involving Auger, fluorescence and shakeoff (isotropic) emissions

Electron energy degradation process through multiple elastic and inelastic (excitation and ionization) scattering collisions in

the gas down to subionization energies

Dias et. al., Phys.Rev. A48 (1993) 2887Botte et. al., IEEE NSS Conf.Rec. (2008) 943 (submitted to IEEE-TNS)

Page 6: X-ray polarimetry in  Xenon gas filled detectors

Monte Carlo simulation: Results

Xe @ 1 atm (Xe E(L-edges) ~ 5 keV )

15 keV unpolarized X-rays 15 keV polarized X-rays

incidence ┴ XY plane, polarization ║ X-axis

Page 7: X-ray polarimetry in  Xenon gas filled detectors

Monte Carlo simulation: ResultsXe @ 1 atm (Xe E(L-edges) ~ 5 keV )

incidence ┴ XY plane, polarization ║ X-axis

15 keV polarized X-rays 25 keV polarized X-rays

Page 8: X-ray polarimetry in  Xenon gas filled detectors

Monte Carlo simulation: Results

40 keV polarized X-rays 50 keV polarized X-rays

incidence ┴ XY plane, polarization ║ X-axis

Xe @ 1 atm (Xe E(K edge)=34.6 keV )

Page 9: X-ray polarimetry in  Xenon gas filled detectors

Polarized X-ray source

Graphite crystal

HPGe detector

X-ray tube

Linearly polarized

X-rays

X-rays

unpolarized

Cu-anode

25kV X-ray tube (K=8.04keV)

Linearly-polarized Bragg-reflection lines characteristic of graphite at

Ex=5.22, 7.84, 10.45, 13.04, 15.66, 18.25, 20.86 keV

Polarized X-rays are obtained by Bragg reflection at 45º from unpolarized radiation generated with an X-ray tube

(HPGe)

Page 10: X-ray polarimetry in  Xenon gas filled detectors

Gridded-MicroStrip Gas Chamber

Absorption/Drift Region (2cm)

X-ray photons Mylar window

Multiplication Region

Xe

1atm

anodegridcathode

Substrate

grid cathode

Physical characteristics

anodes and grids width 5 μm

cathodes width 90 μm

Anode-to-grid gap 12.5 μm

Cathode-to-grid gap 27.5 μm

pitch 185 μm

Page 11: X-ray polarimetry in  Xenon gas filled detectors

G-MSGC: Performance

Gas Microstrip Best R (%) Vanode (V) Vgrid (V) M

XeG-MSGC 13.4 400 100 1150

Standard 15.6 480 - 560

P10G-MSGC 12.6 415 150 590

Standard 13.6 480 - 490

12

16

20

24

28

290 320 350 380 410 440 470 500

Anode voltage (V)

En

erg

y r

es

olu

tio

n,

R (

%)

1

100

10000

Ga

s m

ult

ipli

ca

tio

n f

ac

tor,

M

Dual anode microstrip

Standard microstrip The anode is surrounded by a

close-lying 2nd anode, playing the role of a grid at a potential between anode and cathode, allowing more accurate definition of the multiplication volume.

Carita et. al.,

IEEE TNS-54(5) (2007) 1779

Page 12: X-ray polarimetry in  Xenon gas filled detectors

G-MSGC: Experimental Setup

Graphite crystal

X-ray tube

Linearly polarized X-

rays

i)

ii)

G-MSP active area: 1.5 cm ×1 cm

i) parallel to the MSP length or

ii) parallel to the MSP width

Rectangular G-MSP orientations:

polarization vector is

Polarization direction is //X-axis

Page 13: X-ray polarimetry in  Xenon gas filled detectors

G-MSGC: Experimental Results

Energy spectrum, showing graphite x-ray linesΔ is the observed shift of each peak to lower energies when the rectangular

G-MSP orientation is changed from // to to the polarization vector

Page 14: X-ray polarimetry in  Xenon gas filled detectors

Micro-Hole & Strip Plate

Double-sided microstructure:

•Perforated pattern of a GEM

•Microstrip pattern (bottom)

2 charge-multiplication stages:

inside the holes and

in vicinity of the anodes

Veloso et al., Rev.Sci.Instrum. 71 (2000) 2371

Page 15: X-ray polarimetry in  Xenon gas filled detectors

2D–Micro-Hole & Strip Plate

2D-imaging capability of the detector

Position resolution - x dimension 130 μm (~300 μm FWHM)

Position resolution - y dimension 250 μm (~600 μm FWHM)

Field of view 2.5 cm x 2.5 cm

Physical characteristics and performances of the

2D-MHSP

kapton foil thick 50 μm

holes diameter 50/60 μm

anodes with 20 μm

cathodes with 100 μm

pitch 200 μm

gas filling Xenon

drift region 5 mm

FWHM (@5.9keV) 14%

X-ray imaging using the principle of resistive charge division

Natal da Luz et. al.,

IEEE TNS-55(4) (2008) 2341

Page 16: X-ray polarimetry in  Xenon gas filled detectors

1

10

100

1000

10000

2 4 6 8 10 12 14 16 18 20

X-ray energy Ex (keV)

Co

un

ts

(0 0 4) (0 0 6) (0 0 8) (0 0 10) (0 0 12) (0 0 14)

(18.25 keV graphite line falls under k=19.5 keV flx. peak from the x-ray tube Mo-anode)

2D–MHSP: Experimental Results

Energy spectrum, showing graphite x-ray lines

40 kV x-ray tube

Page 17: X-ray polarimetry in  Xenon gas filled detectors

2D–MHSP: Experimental Results

incidence ┴ XY plane, polarization ║ Y-axis

5.22 keV polarized X-rays 7.84 keV polarized X-rays

Page 18: X-ray polarimetry in  Xenon gas filled detectors

2D–MHSP: Experimental Results

10.45 keV polarized X-rays

polarization ║ Y-axisK fluorescence cloud profile

from the X-ray tube Mo-anode

17.48 keV unpolarized X-rays

Page 19: X-ray polarimetry in  Xenon gas filled detectors

Future work

Conde, X-ray Spectrometry: Recent Technological Advances, John Wiley&Sons, 2004, ch.4

optimization of the absorption region

Photoelectric σP, Rayleigh σR and Compton σC cross-sections for X-rays

and corresponding absorption lengths-L in Xe at 760 Torr.

Absorption depth

Page 20: X-ray polarimetry in  Xenon gas filled detectors

Conclusions

Monte Carlo electron clouds reproduce the anisotropy of the photoelectron emission.

G-MSGC: the shift observed in the peaks of the energy spectrum measured with a Gridded Microstrip Gas Chamber provide clear, albeit indirect, experimental evidence of the alignment of electron cloud profiles in Xe with x-ray polarization direction.

2D-MHSP:The images registered with a 2D position-sensitive detector based on the new Micro-Hole & Strip Plate structure further reinforce that evidence and confirm that a photoelectric polarimeter based on Micro Pattern Gas Chambers is a viable option for X-ray polarimetry.