![Page 1: x Saving data as .csv for use in Excel. x · Florida LogTag Training t Rough outline 1. Preparing the Loggers x Attaching the probes to logger x Matching logger serial number to probe/vial](https://reader034.vdocument.in/reader034/viewer/2022042312/5eda5d3cb3745412b57138be/html5/thumbnails/1.jpg)
STATE OF FLORIDA LOGTAG TRAINING
Florida LogTag Training – Rough outline
1. Preparing the Loggers
Attaching the probes to logger
Matching logger serial number to probe/vial
2. Installing the Loggers
Mounting the Logger
Running the probe cable.
Conditioning the vials
3. Starting the Logger
Sequence to start recording
Button Functionality
Display icons
4. Reviewing Daily Statistics
Daily Max/min values
Interpreting screen data
5. Installing the Software
System requirements
Identify files on disc
Installing the docking station
Initial Set-up
6. Viewing and saving logger data
Pausing/stopping the recording
Detaching the Probe from the Logger – Tracking
Downloading data
Viewing/Interpreting data – viewing options
Saving data as .csv for use in Excel.
Logger Configuration
Emailing data
Attaching the Logger to the Probe - Tracking
7. Advance options
Automation options
8. Questions
![Page 2: x Saving data as .csv for use in Excel. x · Florida LogTag Training t Rough outline 1. Preparing the Loggers x Attaching the probes to logger x Matching logger serial number to probe/vial](https://reader034.vdocument.in/reader034/viewer/2022042312/5eda5d3cb3745412b57138be/html5/thumbnails/2.jpg)
Outline
• Preparing Loggers
• Installation
• Software & Logger Configuration
• Starting Loggers
• Reviewing Daily Statistics
• Viewing/Saving Data
![Page 3: x Saving data as .csv for use in Excel. x · Florida LogTag Training t Rough outline 1. Preparing the Loggers x Attaching the probes to logger x Matching logger serial number to probe/vial](https://reader034.vdocument.in/reader034/viewer/2022042312/5eda5d3cb3745412b57138be/html5/thumbnails/3.jpg)
Preparing The Loggers
![Page 4: x Saving data as .csv for use in Excel. x · Florida LogTag Training t Rough outline 1. Preparing the Loggers x Attaching the probes to logger x Matching logger serial number to probe/vial](https://reader034.vdocument.in/reader034/viewer/2022042312/5eda5d3cb3745412b57138be/html5/thumbnails/4.jpg)
Match the Probe & Logger – Important!! • Each logger and Probe/vial must be matched by
serial number • When unboxing take care to keep Probe/Vials &
matching loggers together
![Page 5: x Saving data as .csv for use in Excel. x · Florida LogTag Training t Rough outline 1. Preparing the Loggers x Attaching the probes to logger x Matching logger serial number to probe/vial](https://reader034.vdocument.in/reader034/viewer/2022042312/5eda5d3cb3745412b57138be/html5/thumbnails/5.jpg)
Install the Probe/Vial • Place vial in center of fridge/freezer unit – secure
with tape or Velcro – Place logger in plastic tray if used with wire rack
• Run probe cable through hinge side of door
• Attach the cable with zip ties or tape.
• Allow Probe/vial to “normalize” in fridge/freezer for 1 ½ to 2 hours before starting the logger.
![Page 6: x Saving data as .csv for use in Excel. x · Florida LogTag Training t Rough outline 1. Preparing the Loggers x Attaching the probes to logger x Matching logger serial number to probe/vial](https://reader034.vdocument.in/reader034/viewer/2022042312/5eda5d3cb3745412b57138be/html5/thumbnails/6.jpg)
![Page 7: x Saving data as .csv for use in Excel. x · Florida LogTag Training t Rough outline 1. Preparing the Loggers x Attaching the probes to logger x Matching logger serial number to probe/vial](https://reader034.vdocument.in/reader034/viewer/2022042312/5eda5d3cb3745412b57138be/html5/thumbnails/7.jpg)
Installing the Logger
• Identify mounting location outside of fridge/freezer
• Be sure it’s within reach of both freezer/fridge probes & is easily accessible
• Attach LTI-MOUNT (clear plastic mount) using the included double sided tape (located in each mount)
• Slide Logger into mount without the probe attached
![Page 8: x Saving data as .csv for use in Excel. x · Florida LogTag Training t Rough outline 1. Preparing the Loggers x Attaching the probes to logger x Matching logger serial number to probe/vial](https://reader034.vdocument.in/reader034/viewer/2022042312/5eda5d3cb3745412b57138be/html5/thumbnails/8.jpg)
Attach Probe to Logger • Important! - Make sure serial number on Logger
matches serial number on Probe/Vial • If not already connected Insert probe end into plug
on logger. You should hear the probe “Click” into place
![Page 9: x Saving data as .csv for use in Excel. x · Florida LogTag Training t Rough outline 1. Preparing the Loggers x Attaching the probes to logger x Matching logger serial number to probe/vial](https://reader034.vdocument.in/reader034/viewer/2022042312/5eda5d3cb3745412b57138be/html5/thumbnails/9.jpg)
• Logger with probe attached in mount
![Page 10: x Saving data as .csv for use in Excel. x · Florida LogTag Training t Rough outline 1. Preparing the Loggers x Attaching the probes to logger x Matching logger serial number to probe/vial](https://reader034.vdocument.in/reader034/viewer/2022042312/5eda5d3cb3745412b57138be/html5/thumbnails/10.jpg)
![Page 11: x Saving data as .csv for use in Excel. x · Florida LogTag Training t Rough outline 1. Preparing the Loggers x Attaching the probes to logger x Matching logger serial number to probe/vial](https://reader034.vdocument.in/reader034/viewer/2022042312/5eda5d3cb3745412b57138be/html5/thumbnails/11.jpg)
Install Software
• Locate the LTA icon on your desktop and double click it to launch the software
![Page 12: x Saving data as .csv for use in Excel. x · Florida LogTag Training t Rough outline 1. Preparing the Loggers x Attaching the probes to logger x Matching logger serial number to probe/vial](https://reader034.vdocument.in/reader034/viewer/2022042312/5eda5d3cb3745412b57138be/html5/thumbnails/12.jpg)
General Settings
• Choose “Edit” > “Options” to access the General Settings screen
![Page 13: x Saving data as .csv for use in Excel. x · Florida LogTag Training t Rough outline 1. Preparing the Loggers x Attaching the probes to logger x Matching logger serial number to probe/vial](https://reader034.vdocument.in/reader034/viewer/2022042312/5eda5d3cb3745412b57138be/html5/thumbnails/13.jpg)
General Settings
• “Show temperatures in” should be set to Fahrenheit
![Page 14: x Saving data as .csv for use in Excel. x · Florida LogTag Training t Rough outline 1. Preparing the Loggers x Attaching the probes to logger x Matching logger serial number to probe/vial](https://reader034.vdocument.in/reader034/viewer/2022042312/5eda5d3cb3745412b57138be/html5/thumbnails/14.jpg)
General Settings
• Information on File Naming Sequence will be provided by VFC Program
![Page 15: x Saving data as .csv for use in Excel. x · Florida LogTag Training t Rough outline 1. Preparing the Loggers x Attaching the probes to logger x Matching logger serial number to probe/vial](https://reader034.vdocument.in/reader034/viewer/2022042312/5eda5d3cb3745412b57138be/html5/thumbnails/15.jpg)
General Settings
• Choose “Export and Reports”
• Check box next to CSV (Comma delimited)
• Click “OK”
![Page 16: x Saving data as .csv for use in Excel. x · Florida LogTag Training t Rough outline 1. Preparing the Loggers x Attaching the probes to logger x Matching logger serial number to probe/vial](https://reader034.vdocument.in/reader034/viewer/2022042312/5eda5d3cb3745412b57138be/html5/thumbnails/16.jpg)
Configuring the Logger
• Carefully unplug probe from Logger
• IMPORTANT – Loggers must be plugged back into the proper serial number matched probe. Take care to return each logger to it’s original location
• Open LogTag Analyzer Software (LTA) (If not already open)
• Drop Logger into docking station – Push down until you hear logger click into place
![Page 17: x Saving data as .csv for use in Excel. x · Florida LogTag Training t Rough outline 1. Preparing the Loggers x Attaching the probes to logger x Matching logger serial number to probe/vial](https://reader034.vdocument.in/reader034/viewer/2022042312/5eda5d3cb3745412b57138be/html5/thumbnails/17.jpg)
Configuring the Logger
• A notification box will pop up telling you that there is no data on the device
• Message will time out or you can click “Cancel” to proceed
![Page 18: x Saving data as .csv for use in Excel. x · Florida LogTag Training t Rough outline 1. Preparing the Loggers x Attaching the probes to logger x Matching logger serial number to probe/vial](https://reader034.vdocument.in/reader034/viewer/2022042312/5eda5d3cb3745412b57138be/html5/thumbnails/18.jpg)
Configuring the Logger
• Next, click the Configuration Wizard Icon
• This will open the LogTag Access Wizard window.
• Click “Next” to launch the Configuration Window
![Page 19: x Saving data as .csv for use in Excel. x · Florida LogTag Training t Rough outline 1. Preparing the Loggers x Attaching the probes to logger x Matching logger serial number to probe/vial](https://reader034.vdocument.in/reader034/viewer/2022042312/5eda5d3cb3745412b57138be/html5/thumbnails/19.jpg)
Configuration Window Overview
• LTA will automatically populate each field with each loggers previous configuration data so most fields will already be completed
• Even though loggers have been pre-configured you will need to give each logger a User ID – Information on Logger ID name and file name will be provided by VFC Program.
![Page 20: x Saving data as .csv for use in Excel. x · Florida LogTag Training t Rough outline 1. Preparing the Loggers x Attaching the probes to logger x Matching logger serial number to probe/vial](https://reader034.vdocument.in/reader034/viewer/2022042312/5eda5d3cb3745412b57138be/html5/thumbnails/20.jpg)
Configuration
• User ID: Give each logger a unique name as directed by VFC Program
• Choose “Push button start”
• Check box next to “Enable pre-start logging”
![Page 21: x Saving data as .csv for use in Excel. x · Florida LogTag Training t Rough outline 1. Preparing the Loggers x Attaching the probes to logger x Matching logger serial number to probe/vial](https://reader034.vdocument.in/reader034/viewer/2022042312/5eda5d3cb3745412b57138be/html5/thumbnails/21.jpg)
Configuration
• Select: “Record readings so that:”
• Set “Record a reading every” to 15 Minutes
• Set “Begin recording after a delay of” to 5 minutes
![Page 22: x Saving data as .csv for use in Excel. x · Florida LogTag Training t Rough outline 1. Preparing the Loggers x Attaching the probes to logger x Matching logger serial number to probe/vial](https://reader034.vdocument.in/reader034/viewer/2022042312/5eda5d3cb3745412b57138be/html5/thumbnails/22.jpg)
Configuration
• Check box next to “Trigger alarm when readings above/equal”
• Enter 46.5 in the field
• Check box next to “Trigger alarm when readings below/equal”
• Enter 35.5 in the field
![Page 23: x Saving data as .csv for use in Excel. x · Florida LogTag Training t Rough outline 1. Preparing the Loggers x Attaching the probes to logger x Matching logger serial number to probe/vial](https://reader034.vdocument.in/reader034/viewer/2022042312/5eda5d3cb3745412b57138be/html5/thumbnails/23.jpg)
Configuration
• We recommend leaving the “Configure requires a password” and
“Download requires a password” options unchecked.
• If enabled and the password is lost the logger will require factory support to unlock
• Click the “Advance Options” button
![Page 24: x Saving data as .csv for use in Excel. x · Florida LogTag Training t Rough outline 1. Preparing the Loggers x Attaching the probes to logger x Matching logger serial number to probe/vial](https://reader034.vdocument.in/reader034/viewer/2022042312/5eda5d3cb3745412b57138be/html5/thumbnails/24.jpg)
Configuration
• Check box for “Alarm remains on even if readings return to non-violation
range”
• Select “Fahrenheit” for the “Temperature display unit”
• Ensure “Allow logging to stop with STOP button” is checked
• Check box for “Show total summary days collected”
• Click the “OK” button to continue
![Page 25: x Saving data as .csv for use in Excel. x · Florida LogTag Training t Rough outline 1. Preparing the Loggers x Attaching the probes to logger x Matching logger serial number to probe/vial](https://reader034.vdocument.in/reader034/viewer/2022042312/5eda5d3cb3745412b57138be/html5/thumbnails/25.jpg)
Configuration
Click “Next” to send configuration to the logger
![Page 26: x Saving data as .csv for use in Excel. x · Florida LogTag Training t Rough outline 1. Preparing the Loggers x Attaching the probes to logger x Matching logger serial number to probe/vial](https://reader034.vdocument.in/reader034/viewer/2022042312/5eda5d3cb3745412b57138be/html5/thumbnails/26.jpg)
Configuration
Configuration is now being transferred to the logger
Progress bar
![Page 27: x Saving data as .csv for use in Excel. x · Florida LogTag Training t Rough outline 1. Preparing the Loggers x Attaching the probes to logger x Matching logger serial number to probe/vial](https://reader034.vdocument.in/reader034/viewer/2022042312/5eda5d3cb3745412b57138be/html5/thumbnails/27.jpg)
Configuration
• Remove Logger from dock and click the “Close” button to continue
![Page 28: x Saving data as .csv for use in Excel. x · Florida LogTag Training t Rough outline 1. Preparing the Loggers x Attaching the probes to logger x Matching logger serial number to probe/vial](https://reader034.vdocument.in/reader034/viewer/2022042312/5eda5d3cb3745412b57138be/html5/thumbnails/28.jpg)
Configuration
• You can now bring the logger back to it’s proper location and reattach the probe
• Important – Loggers must be plugged into serial number matched probe
![Page 29: x Saving data as .csv for use in Excel. x · Florida LogTag Training t Rough outline 1. Preparing the Loggers x Attaching the probes to logger x Matching logger serial number to probe/vial](https://reader034.vdocument.in/reader034/viewer/2022042312/5eda5d3cb3745412b57138be/html5/thumbnails/29.jpg)
![Page 30: x Saving data as .csv for use in Excel. x · Florida LogTag Training t Rough outline 1. Preparing the Loggers x Attaching the probes to logger x Matching logger serial number to probe/vial](https://reader034.vdocument.in/reader034/viewer/2022042312/5eda5d3cb3745412b57138be/html5/thumbnails/30.jpg)
Temperature Value - Temperature value (in °C or °F) Reading Type - CURRENT = The temperature of the last reading taken MAX = The maximum reading in a given 24hr period MIN = The minimum reading in a given 24hr period ALARM Indicator – ALARM appears when an alarm is triggered
![Page 31: x Saving data as .csv for use in Excel. x · Florida LogTag Training t Rough outline 1. Preparing the Loggers x Attaching the probes to logger x Matching logger serial number to probe/vial](https://reader034.vdocument.in/reader034/viewer/2022042312/5eda5d3cb3745412b57138be/html5/thumbnails/31.jpg)
Time value - The time (in HH:MM) can be current time, time remaining to start of delayed start, or duration of a given max/min value above/below the specified limits.
![Page 32: x Saving data as .csv for use in Excel. x · Florida LogTag Training t Rough outline 1. Preparing the Loggers x Attaching the probes to logger x Matching logger serial number to probe/vial](https://reader034.vdocument.in/reader034/viewer/2022042312/5eda5d3cb3745412b57138be/html5/thumbnails/32.jpg)
The value displayed is indicated by the following: TIME = Displaying Current Time (READY & RECORDING modes) TIME DELAY = Displaying remaining time to recording start when a DELAYED START time has been configured (STARTING MODE) DURATION = Displaying duration of a given MAX/MIN statistic above/below the limits in a given day (REVIEW MODE)
![Page 33: x Saving data as .csv for use in Excel. x · Florida LogTag Training t Rough outline 1. Preparing the Loggers x Attaching the probes to logger x Matching logger serial number to probe/vial](https://reader034.vdocument.in/reader034/viewer/2022042312/5eda5d3cb3745412b57138be/html5/thumbnails/33.jpg)
Battery status – Performs battery test hourly
![Page 34: x Saving data as .csv for use in Excel. x · Florida LogTag Training t Rough outline 1. Preparing the Loggers x Attaching the probes to logger x Matching logger serial number to probe/vial](https://reader034.vdocument.in/reader034/viewer/2022042312/5eda5d3cb3745412b57138be/html5/thumbnails/34.jpg)
State Indicator READY = Ready to start recording STARTING = Preparing to start recording. If a delayed start is configured then displays time remaining to recording start RECORDING = Logger is recording (logging data)
![Page 35: x Saving data as .csv for use in Excel. x · Florida LogTag Training t Rough outline 1. Preparing the Loggers x Attaching the probes to logger x Matching logger serial number to probe/vial](https://reader034.vdocument.in/reader034/viewer/2022042312/5eda5d3cb3745412b57138be/html5/thumbnails/35.jpg)
State Indicator PAUSED = Max/Min statistics paused due to button press activity (if so configured) Stopped = Recording has stopped
![Page 36: x Saving data as .csv for use in Excel. x · Florida LogTag Training t Rough outline 1. Preparing the Loggers x Attaching the probes to logger x Matching logger serial number to probe/vial](https://reader034.vdocument.in/reader034/viewer/2022042312/5eda5d3cb3745412b57138be/html5/thumbnails/36.jpg)
Day Alarm Indicators – The display is organized with a table of “day alarm indicator” segments named “Today” to “-29d” which are switched on when a given day has an alarm trigger Day Number – Default is the day number of the current selected day statistic is displayed when in review mode. TODAY is DAY “0”, yesterday is DAY “-1” Your loggers are configured to display the total number of days collected while recording
![Page 37: x Saving data as .csv for use in Excel. x · Florida LogTag Training t Rough outline 1. Preparing the Loggers x Attaching the probes to logger x Matching logger serial number to probe/vial](https://reader034.vdocument.in/reader034/viewer/2022042312/5eda5d3cb3745412b57138be/html5/thumbnails/37.jpg)
Starting the Logger
1. Confirm that “Ready” is displayed on the LCD Screen
2. Press and hold the start button and observe that the “STARTING” icon starts flashing – Hold button down until flashing stops. (Typically 4-6 seconds but it can take longer if button is not continually depressed)
![Page 38: x Saving data as .csv for use in Excel. x · Florida LogTag Training t Rough outline 1. Preparing the Loggers x Attaching the probes to logger x Matching logger serial number to probe/vial](https://reader034.vdocument.in/reader034/viewer/2022042312/5eda5d3cb3745412b57138be/html5/thumbnails/38.jpg)
Starting the Logger
3. Release the Start button when the “STARTING” icon stops flashing
• The logger is now in STARTING mode – “TIME DELAY 00:05” will be displayed
• Once the start delay times out (after 5 min) the logger will enter “RECORDING” mode. At this time the display will show current temperature, current time, battery status, and any alarm triggers. This information is updated each time a log is taken
![Page 39: x Saving data as .csv for use in Excel. x · Florida LogTag Training t Rough outline 1. Preparing the Loggers x Attaching the probes to logger x Matching logger serial number to probe/vial](https://reader034.vdocument.in/reader034/viewer/2022042312/5eda5d3cb3745412b57138be/html5/thumbnails/39.jpg)
Congratulations!
You're Logging!
![Page 40: x Saving data as .csv for use in Excel. x · Florida LogTag Training t Rough outline 1. Preparing the Loggers x Attaching the probes to logger x Matching logger serial number to probe/vial](https://reader034.vdocument.in/reader034/viewer/2022042312/5eda5d3cb3745412b57138be/html5/thumbnails/40.jpg)
Recording Display
• The normal recording mode display shows the current temperature of the last log taken. It’s updated at the same rate as the logging interval.
• The current time, battery status and alarm trigger day summary are also displayed.
![Page 41: x Saving data as .csv for use in Excel. x · Florida LogTag Training t Rough outline 1. Preparing the Loggers x Attaching the probes to logger x Matching logger serial number to probe/vial](https://reader034.vdocument.in/reader034/viewer/2022042312/5eda5d3cb3745412b57138be/html5/thumbnails/41.jpg)
![Page 42: x Saving data as .csv for use in Excel. x · Florida LogTag Training t Rough outline 1. Preparing the Loggers x Attaching the probes to logger x Matching logger serial number to probe/vial](https://reader034.vdocument.in/reader034/viewer/2022042312/5eda5d3cb3745412b57138be/html5/thumbnails/42.jpg)
Reviewing Daily Statistics
• Press the “Review” button to displays the currents day’s max statistic
• The “Today” segment flashes and ‘Day 00’ is shown to indicate the ‘Today’ selection.
![Page 43: x Saving data as .csv for use in Excel. x · Florida LogTag Training t Rough outline 1. Preparing the Loggers x Attaching the probes to logger x Matching logger serial number to probe/vial](https://reader034.vdocument.in/reader034/viewer/2022042312/5eda5d3cb3745412b57138be/html5/thumbnails/43.jpg)
Reviewing Daily Statistics
• Pressing “Review” again displays the day’s minimum statistic
![Page 44: x Saving data as .csv for use in Excel. x · Florida LogTag Training t Rough outline 1. Preparing the Loggers x Attaching the probes to logger x Matching logger serial number to probe/vial](https://reader034.vdocument.in/reader034/viewer/2022042312/5eda5d3cb3745412b57138be/html5/thumbnails/44.jpg)
Reviewing Daily Statistics
• Pressing “Review” again displays the next previous day’s max statistic
• In this example the selected day has recordings that are above the upper limit and the duration value shows the amount of time recorded above the limit.
• The max statistic is above the upper limit so the up arrow is shown and an alarm was triggered by this statistic as indicated by the ALARM text shown.
![Page 45: x Saving data as .csv for use in Excel. x · Florida LogTag Training t Rough outline 1. Preparing the Loggers x Attaching the probes to logger x Matching logger serial number to probe/vial](https://reader034.vdocument.in/reader034/viewer/2022042312/5eda5d3cb3745412b57138be/html5/thumbnails/45.jpg)
Reviewing Daily Statistics
• Pressing “Review” again displays the next previous days minimum statistic
• With each press of the “Review” button the next previous days max and min statistic will be displayed (up to DAY -29)
• Pressing the “START” button at any time will reverts the display back to normal mode
![Page 46: x Saving data as .csv for use in Excel. x · Florida LogTag Training t Rough outline 1. Preparing the Loggers x Attaching the probes to logger x Matching logger serial number to probe/vial](https://reader034.vdocument.in/reader034/viewer/2022042312/5eda5d3cb3745412b57138be/html5/thumbnails/46.jpg)
View/Save Logger Data
• Slide logger out of mount
• Stop the recording process by pressing and holding the “Start/clear/stop” button
• When “STOPPED” is no longer flashing release the button
• You should see “STOPPED” on screen
• Carefully unplug probe from Logger
![Page 47: x Saving data as .csv for use in Excel. x · Florida LogTag Training t Rough outline 1. Preparing the Loggers x Attaching the probes to logger x Matching logger serial number to probe/vial](https://reader034.vdocument.in/reader034/viewer/2022042312/5eda5d3cb3745412b57138be/html5/thumbnails/47.jpg)
View/Save Logger Data
• IMPORTANT – Loggers must be plugged back into the proper serial number matched probe. Take care to return each logger to it’s original location
• Open LogTag Analyzer Software (LTA)
• Drop Logger into docking station
![Page 48: x Saving data as .csv for use in Excel. x · Florida LogTag Training t Rough outline 1. Preparing the Loggers x Attaching the probes to logger x Matching logger serial number to probe/vial](https://reader034.vdocument.in/reader034/viewer/2022042312/5eda5d3cb3745412b57138be/html5/thumbnails/48.jpg)
View/Save Logger Data
• Data is automatically saved to computer and LTA launches data window and displays the “Chart” tab
![Page 49: x Saving data as .csv for use in Excel. x · Florida LogTag Training t Rough outline 1. Preparing the Loggers x Attaching the probes to logger x Matching logger serial number to probe/vial](https://reader034.vdocument.in/reader034/viewer/2022042312/5eda5d3cb3745412b57138be/html5/thumbnails/49.jpg)
View/Save Logger Data
• Additional views and report options can be chosen from the tabs at the bottom of the window
![Page 50: x Saving data as .csv for use in Excel. x · Florida LogTag Training t Rough outline 1. Preparing the Loggers x Attaching the probes to logger x Matching logger serial number to probe/vial](https://reader034.vdocument.in/reader034/viewer/2022042312/5eda5d3cb3745412b57138be/html5/thumbnails/50.jpg)
View/Save Logger Data
• Once graph is visible the data has been saved and the logger can be re-launched.
• To re-launch the logger click on the LogTag Analyzer Wizard button
![Page 51: x Saving data as .csv for use in Excel. x · Florida LogTag Training t Rough outline 1. Preparing the Loggers x Attaching the probes to logger x Matching logger serial number to probe/vial](https://reader034.vdocument.in/reader034/viewer/2022042312/5eda5d3cb3745412b57138be/html5/thumbnails/51.jpg)
View/Save Logger Data
• The logger can now be configured using the same procedure as described earlier
• While LTA will retain the configuration settings it’s always a good idea to confirm that they are correct before sending the configuration to the logger
• Once configuration is complete you can remove the logger and return it to it’s proper location.
• Broken Record - Loggers must be plugged into serial number matched probe
![Page 52: x Saving data as .csv for use in Excel. x · Florida LogTag Training t Rough outline 1. Preparing the Loggers x Attaching the probes to logger x Matching logger serial number to probe/vial](https://reader034.vdocument.in/reader034/viewer/2022042312/5eda5d3cb3745412b57138be/html5/thumbnails/52.jpg)
View/Save Logger Data
• Be sure to start each logger and ensure that probe is correctly attached
• You can now move on to the next logger
• You will be given separate instruction on exporting the data file into the FLSHOTS program.
![Page 53: x Saving data as .csv for use in Excel. x · Florida LogTag Training t Rough outline 1. Preparing the Loggers x Attaching the probes to logger x Matching logger serial number to probe/vial](https://reader034.vdocument.in/reader034/viewer/2022042312/5eda5d3cb3745412b57138be/html5/thumbnails/53.jpg)
Additional Support
• Detailed instructions on using the logger and LTA can be found in the LogTag Analyzer manual
• Manual can be found on software disc or by clicking on the “help” option from inside the LTA software
• Technical support is available from MicroDAQ.com, Ltd. Monday – Friday from 9:30 – 4:30 Est by calling 603-746-5524 or email at [email protected]
![Page 54: x Saving data as .csv for use in Excel. x · Florida LogTag Training t Rough outline 1. Preparing the Loggers x Attaching the probes to logger x Matching logger serial number to probe/vial](https://reader034.vdocument.in/reader034/viewer/2022042312/5eda5d3cb3745412b57138be/html5/thumbnails/54.jpg)