XPS (X ray photoemission spectroscopy)/ESCA (Electron Spectroscopy for Chemical Analysis)
XPS
Core-levels: electroniccore-levels are more atomic-like: elementchemical shifts from formal oxidation state of the atom, the local chemical and physical environment : chemicallike Auger electron it has also short : surface sensitive Angular dependence has diffraction effects: structure Magnetic dichroism magnetic
Theoretical consideration
As photoemission is much more simple process than Auger process, conservation of energy then requires that : KE = h - ( E(A+ ) - E(A) ) - The final term in brackets,
representing the difference in energy between the ionized and neutral atoms, is generally called the binding energy (BE) of the electron . ( is the work function of the solid when KE is counted near surface, however, KE detected by analyzer then is the work function of analyzer.)
Z dependenceBE follows the energy levels: BE(1s)>BE(2s)>BE(2p)…BE with same orbital increase with Z: BE(Mg1s)>BE(Na1s)
Elemental specific!
XPS data base
Main features of XPS
Three Step Model
1. Absorption of the photon and excitation of electrons
2.Transport of electrons to the surface
3.The escape of the electrons from surface to the vacuum.
Inelastic scatteringIn the step 2, inelastic scattering let XPS spectra consists of core-level photo-emission peaks imposed by a step-like structure (background) due to the various mechanism to lose kinetic energy. Besides, there are also AES processes visible.
XPS peak fit
XPS peak identification
•Photoelectron lines: core-level, valence bands, spin-orbit splitting•Auger lines•Chemical shifts•X-ray satellites•X-ray “Ghost”•Shake-up satellite•Multiplet satellite•Energy loss lines
Spin-Orbit splitting:
Spin-orbit splitting is an initial state effect. For any electron in orbital with orbital angular momentum, coupling between magnetic fields of spin (s) and angular momentum (l) occurs
Lower binding energy
Higher binding energy
Total angular momentum j = |l ± s|, therefore for s electron there is no degeneracy, and other orbitals have two degeneracy:
- s orbitals are not spin-orbit split - singlet in XPS
- p, d, f… orbitals are spin-orbit split - doublets in XPS
- BE of lower j value in doublet is higher (BE 2p1/2 > BE 2p3/2)
- Magnitude of spin-orbit splitting increases with Z
- Magnitude of spin-orbit splitting decreases with distance from nucleus
(increased nuclear shielding) Intensity ratio?Intensity ratio?
Core Level Chemical ShiftsPosition of orbitals in atom is sensitive to chemical environment of atom. In solid all core levels for that atom shifted by approx. same amount (<10 eV). Chemical shift correlated with overall charge on atom (Reduced charge ®increased BE)For k-shell of an atom in a compound:
EB(k) = EB(k,qA) +Vwhere EB(k,qA) is the binding energy of a free ion, A, qA is the net charge of A, and V is the potential at A due to all other atoms. V can be described as:
V = e2qAqi/riA
(i is any other atoms except A), riA is the distance between i and A, therefore chemical shift is:
EB(k)=EB(k,qA1)-EB(k,qA2 )+V1 –V2
qA1 and qA2 is the difference of charge in two states.
XPS spectra for Si and its compounds with F in a) and chemical shifts vs. the charge in b)
Both S and Si binding energies increase with psitive charge (the loss of negative charge of electron), and the same for C.
FunctionalGroup
Binding Energy(eV)
hydrocarbon C-H, C -C 285.0
amine C-N 286.0
alcohol, ether C-O-H, C -O-C 286.5
Cl bound to C C-Cl 286.5
F bound to C C-F 287.8
carbonyl C=O 288.0
As the samples shown before, binding energies of Al3+ is higher than the metal atom, in the meanwhile, the binding energy of O atom (more positive charge) is higher than the O2- ion.
The chemical shifts due to the variation of the distribution of the charges at the atom site is the main reason for the other name of XPS:
ESCA (Electron Spectroscopy for Chemical Analysis)
Shake-up and shake-off
Photoemission process can leave the ions in the ground state (main peak) and also possibly in an excited sate (shake-up/shake-off satellites), the latter makes the KE of photoelectron less: higher BE.- excitation of electron to bound state shake-up satellite- excitation of electron to unbound (continuum) state shake-off satellite- excitation of hole state shake-down satellite - rare
The shown is XPS
spectra for Cu 2p
photoemission at
different chemical
states. The shake-up
Lines does not exist
in Cu metal, and
is unique for CuO
And CuSO4
Some general rules
Shake-up features especially common in transition metal oxides associated with paramagnetic species. Generally, the shake-up/shake-off satellites have intensities and energy separations from the parent photoelectron line that are unique to each chemical state, which can be used to analyze the chemical state of the elements. Even Some Auger lines also exhibit changes due to these processes. With transition metal, the absence of these lines is the fingerprint for elemental or diamagnetic states. Prominent satellites occurs with paramagnetic states.
Multiplet splitting and shake-up/shake-off lines are generally expected in the paramagnetic states:
MnO XPS spectra
Chemical shifts are too small to distinguish the chemical states of Mn in MnO from a). In b) the satellites are due to Mn2+, while for Mn3+ and Mn4+, although there should be satellites, they are with higher binding energies.
Shake-up/Shake-off satellites are another reason for the chemical sensitivity of XPS
Multiplet satellite Following photoelectron emission, the remaining unpaired electron may couple with other unpaired electrons in the atom, resulting in an ion with several possible final state configurations with as many different energies. This produces a line which is split asymmetrically into several components.
For s-type orbital with other unpaired electrons in the atom there are split lines like in the shown Figure for Mn 3s.
For p or even higher orbital levels, is more complex and subtle
Energy loss lines
eph + esolid e*ph + e**solid
Photoelectrons travelling through the solid can interact with other electrons in the material. These interactions can result in the photoelectron exciting an electronic transition, thus losing some of its energy (inelastic scattering). Most common are due to interband or plasmons (bulk or surface).
Surface plasmon
(bulk plasmon)
The plasmon loss satellites are rarely sharp in insulators but very prominent in the metals. The main peak is normally observed at higher binding energy with several lines with the same energy intervals and reduced intensity, and the interval can be not only single one due to different origins: bulk or surface plasmons, bulk one is more prominent and interval larger (21/2 factor of the surface one).
Energy of LightWavelength() 106m 103m 1 m 10-3m 10-6m
Energy(E)
Bro
ad-
cast
Sh
ort
wav
e ra
dio
Infr
are
d
U VX
-ra
y
Gam
ma
Ray
Vis
ibl
e
1 MeV1 KeV1 eV10-3eV10-6eV
X-ray tubeEarly x-ray source
Standard lab X-ray source is by very high energy e beam hitting the anode.
A common Dual anode X-ray tube
X-ray spectrum from x-ray tube
Characteristic lines from the X ray fluorescence process (XRF) and a broad background (Bremsstrahlung), which is strongly depends on the energy of the electron
Typical X-ray anode material (Mg and Al)
2p3/2 1s and 2p1/2 1s transitions produce soft x-rays
K1,2 radiation (unresolved doublet)
h(eV) FWHM (eV)
Mg 1253.6 0.7
Al 1486.6 0.85
Same transitions in doubly ionized Mg or Al produce K3,4 lines at h~ 9-10 eV higher…
3p 1s transitions produce Kx-rays
Energies and widths of characteristic soft X-ray lines of different materials
Mg K-shell X-ray emission spectrum.
The full line shows the characteristic line emissions after subtraction of a constant background as shown by the dashed line. Note the logarithmic intensity scale.
X-ray satellites
Emission from non-monochromatic x-ray sources produces satellite peaks in XPS spectrum at lower BE.
“ghost peaks”
O Ka at 524.9 eV
Ghost peaks are due to contamination of the x-ray source, which produces x-ray emission at different wavelength and it can also due to contamination of the sample holder etc.
Monochromatic X-ray
1. Narrow peak width
2. Reduced background
3. No satellite & Ghost peaks
Goal to achieve
SampleSample
X-ray AnodeX-ray Anode
Energy Energy AnalyzerAnalyzer
Quartz Quartz Crystal DisperserCrystal Disperser
ee--
Rowland CircleRowland Circle
n=2dsinFor quartz (1010) surface, d=0.42 nm and 78.5 degree for Al K0.93 nm
Synchrotron Radiation
The synchrotron storage ring is a tubular vacuum chamber made to: Hold an electron beam travelling through it at nearly the speed of light. Maintain the high energy of the electron beam. As the accelerating electrons circle the ring at relativistic velocities, they give off intense beams of light including x-rays. By using a monochromator the light will be Monochromatic.
Key properties of synchrotron radiation:high intensitytunability in wide rangenear-coherencepolarized. pulsedwell collimated
NUS has such a source in Singapore!
Sample charging effects
The light for XPS always charges surface positively (shifting of spectrum to higher binding energy) and leads to general instability (spectral noise). For the metal sample, which can be grounded and the charges can be quickly gone. However, for insulator, this effects are serious and need to be treated.
C 1s shifts due to the charging
For XPS For XPS (even AES) (even AES) never forget never forget ground the ground the sample !!!sample !!!
Inhomogeneous Surface ChargingCharging can even change the line shape due to Inhomogeneous Surface Charging, which have different positive voltage on the surface.
In a lot of cases, there is only spectral shift due to charging, which can be determined by comparison with known elemental XPS lines, for example C 1s.
Charge CompensationWhen the spectra is distortedWhen the spectra is distorted
electron flood gun mounted line of sight with sample
electron flood gun mounted in analyzer axis + electromagnet
Which way is B field?Which way is B field?
Other methods including make the sample very thin that is does not insulate, earthed metal mesh and very focused X-ray spot can also help sometimes.
Quantitative analysis
X-ray X-ray penetrate penetrate much deep much deep than the than the escape depth escape depth of electronsof electrons
Can be found inCan be found inhandbookhandbook
How to measure the intensity
Lorentzian or Gaussian functions plus a background (or even more complicated functions) of E can be used to fit the peak to subtract the background. (More complicated Shirely background.)
Instrumentation (analyzers)
resolution
Acceptance angle
Analyzer: most essential part of any electron spectroscopy, its characteristic are: energy range, energy resolution, sensitivity and acceptance angle. Normally its functions involve: retarding of the incoming electron, selection of the electrons with right kinetic energy (pass energy), detecting of the electrons (channeltron)
Hemispherical Analyzer
5 4 . 7
X-rayX-raySourceSource
ElectronElectronOpticsOptics
Outer SphereOuter Sphere
Inner SphereInner Sphere
SampleSample
Analyzer ControlAnalyzer Control
Multi-Channel chanMulti-Channel channeltron Electron Mulneltron Electron Multipliertiplier
Most widely used for XPS
Hemispherical Analyzer
Pass energy: E = e U (b/a - a/b)
E/E = (x1+x2)/2r +2Resolution:
=(a+b)/2
U is the voltage difference between inner and outer sphere; a and b are radii of inner and outer spheres; x1 and x2 are the radii of the entrance and exits apertures, respectively; a is the maximum deviation of the electron trajectories at the entrance with respect to the center line.
Angular resolved XPSPhotoemission is a dipole interaction, its Hamilton can be write as:
Why?Why?
The transition possibility is:
with
Obviously the experimental geometry (the directions of the incident light and electron emission) is crucial to the photoemission process. Moreover, the electronic structure will be influenced by the presence of the surface, its possible influence will be present by the sample normal.
The change of emission angle with respect to the sample normal can also give different surface sensitivity.
The angular dependence of XPS is how the photoelectron diffraction (XPD) is done, which gives the structural information of the surface.
Angular resolved XPS
Various angular dependence
Surface sensitivity change due to angle and photon energy
More Surface More Surface SensitiveSensitive
less Surface less Surface SensitiveSensitive
Same path length but the depth different
Can be done aslo with AES!Can be done aslo with AES!
Sample for surface sensitivity change due to angle
For photoemission, Ekin = h - EB
Change of photon energy can change photoelectron energy that also changes the free path length of the photoelectrons(surface sensitivity).
CannotCannot be done with AES! be done with AES!
AES vs. XPS
1. Common points
2. difference
•Both elemental and chemical sensitive•Both can be used to do quantitative analysis of chemical composition.•Both are electron spectroscopy which have surface sensitivity.
AES: involved two electrons and one hole, due to coulomb interaction (no selection rule), complicated, peak broad, can be excited by many energetic particles including photon, no intrinsic angular dependence, commonly use CMA, AES peak in XPS spectra is with fixed Ekin.
XPS: involved in one electron, due to dipole interaction (selection rule), peak sharp, simple, only excited by photon, sensitive to angular geometry, often use angular resolved analyzer, XPS peak is with fixed binding energy.