Z’-LYTE ® Compatible Microplate Reader Documentation Version No.: 08 Jan 15 Page 1 of 18 Setup Guide on the BMG LABTECH CLARIOstar ® Microplate Readers Have a question? Contact our Technical Support Team NA: 800-955-6288 ext. 40266 Email: [email protected] Z’-LYTE ® Assay Setup Guide on the BMG LABTECH CLARIOstar ® Reader The BMG LABTECH CLARIOstar ® Microplate Readers were tested for compatibility with the Life Technologies’ Z’-LYTE ® Assay using the Z'-LYTE ® Ser/Thr 11 Kit (PV3670) and FRAP1 (mTOR) kinase (PV4753). The following document is intended to demonstrate setup of this instrument and provide representative data. For more detailed information and technical support of Life Technologiesassays please call 1-800-955-6288 extension 40266. For more detailed information and technical support of BMG LABTECH instruments or software, please contact BMG LABTECH at 1-877-264-5227 or www.bmglabtech.com. A. Recommended Optics wavelength (nm) Wavelength Selection Excitation 400-12 (or similar) LVF Monochromator Emission 1 445-12 (or similar) LVF Monochromator Emission 2 520-12 (or similar) LVF Monochromator Emission 1 Dichroic 422.5 Linear Variable Dichroic Emission 2 Dichroic 460 Linear Variable Dichroic

Upload: others

Post on 23-Aug-2020

2 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: Z’-LYTE Assay Setup Guide on the BMG LABTECH CLARIOstar … · The BMG LABTECH CLARIOstar® Microplate Readers were tested for compatibility with the Life Technologies’ Z’-LYTE

Z’-LYTE® Compatible Microplate Reader

Documentation

Version No.: 08 Jan 15

Page 1 of 18

Setup Guide on the BMG LABTECH CLARIOstar® Microplate Readers

Have a question? Contact our Technical Support Team

NA: 800-955-6288 ext. 40266 Email: [email protected]

Z’-LYTE® Assay Setup Guide on the

BMG LABTECH CLARIOstar® Reader The BMG LABTECH CLARIOstar® Microplate Readers were tested for compatibility with the Life Technologies’ Z’-LYTE® Assay using the Z'-LYTE® Ser/Thr 11 Kit (PV3670) and FRAP1 (mTOR) kinase (PV4753). The following document is intended to demonstrate setup of this instrument and provide representative data. For more detailed information and technical support of Life Technologies’ assays please call 1-800-955-6288 extension 40266. For more detailed information and technical support of BMG LABTECH instruments or software, please contact BMG LABTECH at 1-877-264-5227 or www.bmglabtech.com. A. Recommended Optics

wavelength (nm)

Wavelength Selection

Excitation

400-12 (or similar)

LVF Monochromator

Emission 1

445-12 (or similar)

LVF Monochromator

Emission 2

520-12 (or similar)

LVF Monochromator

Emission 1 Dichroic

422.5 Linear Variable Dichroic

Emission 2 Dichroic 460 Linear Variable Dichroic

Page 2: Z’-LYTE Assay Setup Guide on the BMG LABTECH CLARIOstar … · The BMG LABTECH CLARIOstar® Microplate Readers were tested for compatibility with the Life Technologies’ Z’-LYTE

Z’-LYTE® Compatible Microplate Reader

Documentation

Version No.: 08 Jan 15

Page 2 of 18

Setup Guide on the BMG LABTECH CLARIOstar® Microplate Readers

Have a question? Contact our Technical Support Team

NA: 800-955-6288 ext. 40266 Email: [email protected]

B. Instrument Setup

1. Make certain plate reader is turned on, and open up CLARIOstar® Control software on

computer. Insert plate into plate reader.

2. Setup protocol for Z'-LYTE®, select "Manage Protocols" from the Protocols tab in the ribbon style menu at the top of the window.

Page 3: Z’-LYTE Assay Setup Guide on the BMG LABTECH CLARIOstar … · The BMG LABTECH CLARIOstar® Microplate Readers were tested for compatibility with the Life Technologies’ Z’-LYTE

Z’-LYTE® Compatible Microplate Reader

Documentation

Version No.: 08 Jan 15

Page 3 of 18

Setup Guide on the BMG LABTECH CLARIOstar® Microplate Readers

Have a question? Contact our Technical Support Team

NA: 800-955-6288 ext. 40266 Email: [email protected]

3. At this point, a new screen will open (below). Click on the “Show all test protocols” or "Fluorescence Intensity" button on the left side of the screen, then select “New” from the tabs at the bottom.

Page 4: Z’-LYTE Assay Setup Guide on the BMG LABTECH CLARIOstar … · The BMG LABTECH CLARIOstar® Microplate Readers were tested for compatibility with the Life Technologies’ Z’-LYTE

Z’-LYTE® Compatible Microplate Reader

Documentation

Version No.: 08 Jan 15

Page 4 of 18

Setup Guide on the BMG LABTECH CLARIOstar® Microplate Readers

Have a question? Contact our Technical Support Team

NA: 800-955-6288 ext. 40266 Email: [email protected]

4. A new window will pop up. Select “Fluorescence Intensity” and “Endpoint” and then select “OK”.

Page 5: Z’-LYTE Assay Setup Guide on the BMG LABTECH CLARIOstar … · The BMG LABTECH CLARIOstar® Microplate Readers were tested for compatibility with the Life Technologies’ Z’-LYTE

Z’-LYTE® Compatible Microplate Reader

Documentation

Version No.: 08 Jan 15

Page 5 of 18

Setup Guide on the BMG LABTECH CLARIOstar® Microplate Readers

Have a question? Contact our Technical Support Team

NA: 800-955-6288 ext. 40266 Email: [email protected]

5. A new Protocol window will open automatically. Enter a test name, select plate type, type “2” into box beside “No. of multichromatics” and in Optic select “Top optic.” In Speed and Precision select “Rapid” or “Precise” or click on ▼.

Page 6: Z’-LYTE Assay Setup Guide on the BMG LABTECH CLARIOstar … · The BMG LABTECH CLARIOstar® Microplate Readers were tested for compatibility with the Life Technologies’ Z’-LYTE

Z’-LYTE® Compatible Microplate Reader

Documentation

Version No.: 08 Jan 15

Page 6 of 18

Setup Guide on the BMG LABTECH CLARIOstar® Microplate Readers

Have a question? Contact our Technical Support Team

NA: 800-955-6288 ext. 40266 Email: [email protected]

6. Additional selections are now made available to change the speed at which the plate is read. When finished, select the "Layout" tab at the top of the Protocol window.

Page 7: Z’-LYTE Assay Setup Guide on the BMG LABTECH CLARIOstar … · The BMG LABTECH CLARIOstar® Microplate Readers were tested for compatibility with the Life Technologies’ Z’-LYTE

Z’-LYTE® Compatible Microplate Reader

Documentation

Version No.: 08 Jan 15

Page 7 of 18

Setup Guide on the BMG LABTECH CLARIOstar® Microplate Readers

Have a question? Contact our Technical Support Team

NA: 800-955-6288 ext. 40266 Email: [email protected]

7. Select the wells you wish to read. Note in this step you can select to designate blanks, positive controls, etc. but for this case we marked all wells "Sample" and calculations were performed manually. When finished, select the “Multichromatic” tab.

Page 8: Z’-LYTE Assay Setup Guide on the BMG LABTECH CLARIOstar … · The BMG LABTECH CLARIOstar® Microplate Readers were tested for compatibility with the Life Technologies’ Z’-LYTE

Z’-LYTE® Compatible Microplate Reader

Documentation

Version No.: 08 Jan 15

Page 8 of 18

Setup Guide on the BMG LABTECH CLARIOstar® Microplate Readers

Have a question? Contact our Technical Support Team

NA: 800-955-6288 ext. 40266 Email: [email protected]

8. Adjust the monochromator settings by typing in the desired wavelength – bandpass. For both chromatics, type in “400-12” for the Excitation. For chromatic 1, type in an Emission of “445-12” and for chromatic 2, type in an Emission “520-12.” The Dichroic settings will be adjusted automatically. At this time you can select “Start measurement” to run this protocol immediately or “OK” to save this protocol.

Page 9: Z’-LYTE Assay Setup Guide on the BMG LABTECH CLARIOstar … · The BMG LABTECH CLARIOstar® Microplate Readers were tested for compatibility with the Life Technologies’ Z’-LYTE

Z’-LYTE® Compatible Microplate Reader

Documentation

Version No.: 08 Jan 15

Page 9 of 18

Setup Guide on the BMG LABTECH CLARIOstar® Microplate Readers

Have a question? Contact our Technical Support Team

NA: 800-955-6288 ext. 40266 Email: [email protected]

9. If you selected “OK” in the previous step you will return to the initial settings window. Select the “Start measurement” button in the menu bar at the top of the screen.

Page 10: Z’-LYTE Assay Setup Guide on the BMG LABTECH CLARIOstar … · The BMG LABTECH CLARIOstar® Microplate Readers were tested for compatibility with the Life Technologies’ Z’-LYTE

Z’-LYTE® Compatible Microplate Reader

Documentation

Version No.: 08 Jan 15

Page 10 of 18

Setup Guide on the BMG LABTECH CLARIOstar® Microplate Readers

Have a question? Contact our Technical Support Team

NA: 800-955-6288 ext. 40266 Email: [email protected]

8. A new window will appear allowing you to select which of your test protocols to be run. Select the protocol created for Z'-LYTE®, and then press “OK.”

Page 11: Z’-LYTE Assay Setup Guide on the BMG LABTECH CLARIOstar … · The BMG LABTECH CLARIOstar® Microplate Readers were tested for compatibility with the Life Technologies’ Z’-LYTE

Z’-LYTE® Compatible Microplate Reader

Documentation

Version No.: 08 Jan 15

Page 11 of 18

Setup Guide on the BMG LABTECH CLARIOstar® Microplate Readers

Have a question? Contact our Technical Support Team

NA: 800-955-6288 ext. 40266 Email: [email protected]

9. A new window will appear. Place the plate into the reader, and select a well to use for adjusting gain and focus by highlighting the well of your choice. The gain or sensitivity can be adjusted for the highlighted channel at this point; in this case a positive control (high FRET) well was used. When finished, click on the "Start Adjustment" tab. This will procedure will need to be followed for both channels.

Page 12: Z’-LYTE Assay Setup Guide on the BMG LABTECH CLARIOstar … · The BMG LABTECH CLARIOstar® Microplate Readers were tested for compatibility with the Life Technologies’ Z’-LYTE

Z’-LYTE® Compatible Microplate Reader

Documentation

Version No.: 08 Jan 15

Page 12 of 18

Setup Guide on the BMG LABTECH CLARIOstar® Microplate Readers

Have a question? Contact our Technical Support Team

NA: 800-955-6288 ext. 40266 Email: [email protected]

10. In a moment, the instrument will have calculated its optimal focal height and the gain adjustments necessary. When finished with adjusting both channels, click on the "Start Measurement" tab to read.

Page 13: Z’-LYTE Assay Setup Guide on the BMG LABTECH CLARIOstar … · The BMG LABTECH CLARIOstar® Microplate Readers were tested for compatibility with the Life Technologies’ Z’-LYTE

Z’-LYTE® Compatible Microplate Reader

Documentation

Version No.: 08 Jan 15

Page 13 of 18

Setup Guide on the BMG LABTECH CLARIOstar® Microplate Readers

Have a question? Contact our Technical Support Team

NA: 800-955-6288 ext. 40266 Email: [email protected]

11. When CLARIOstar® has finished reading, collect the data by clicking the “MARS” icon in the “Results” portion of the ribbon on the toolbar at the top of the window. This will automatically redirect you to the MARS data evaluation file which collects run data. Select the run of interest from the list to open, and view data in a plate layout format.

Page 14: Z’-LYTE Assay Setup Guide on the BMG LABTECH CLARIOstar … · The BMG LABTECH CLARIOstar® Microplate Readers were tested for compatibility with the Life Technologies’ Z’-LYTE

Z’-LYTE® Compatible Microplate Reader

Documentation

Version No.: 08 Jan 15

Page 14 of 18

Setup Guide on the BMG LABTECH CLARIOstar® Microplate Readers

Have a question? Contact our Technical Support Team

NA: 800-955-6288 ext. 40266 Email: [email protected]

C. Z'-LYTE® Kinase Assay using FRAP1 (mTOR) Purpose

The following is a sample assay performed for demonstration purposes. The section below describes how the data was obtained, and is not intended for use as an assay protocol. We recommend all first-time users follow the appropriate protocols and/or validation packets provided with their specific assay kits, and include all proper controls. The instrument settings above would be sufficient for any Z'-LYTE® assay and were optimized for high throughput. The information below is provided as representative data. Assay was run at ATP Km apparent and a kinase concentration producing approximately 30-40% of maximal phosphorylation, as discussed in the Z'-LYTE® protocols. ATP and kinase concentrations should be optimized for each kinase by the actual user. Specific Z’-LYTE® assay protocols and setup information from Life Technologies’ own in-house SelectScreen® Custom Profiling Z’-LYTE®-based kinase assay service can be located at the following link: http://www.lifetechnologies.com/selectscreen and select the SelectScreen® Kinase Profiling Service and select Technology Overview. At a Glance Step 1: Setup the instrument and plate layout. Step 2: Dilution series of inhibitor. Step 3: Prepare assay plate with controls. Step 4: Read. Materials Required

Component Part

Number Storage

FRAP1 (mTPOR) kinase PV4753 -80°C

Z'-LYTE®-LYTE Assay Kit – Ser/Thr 11 containing PV3670

Z'-LYTE® Kinase Assay Kit - Ser/Thr 11 Peptide PV3671 -20°C

Z'-LYTE® Ser/Thr 11 Phospho-Peptide PV3672 -20°C

5X Kinase Buffer A PV3189 20-30°C

ATP PV3227 -20°C

Development Reagent B PV4126 -80°C

Development Buffer P3127 20-30°C

FRAP1 (mTOR) Kinase Buffer – User prepared see

http://www.lifetechnologies.com/selectscreen

Page 15: Z’-LYTE Assay Setup Guide on the BMG LABTECH CLARIOstar … · The BMG LABTECH CLARIOstar® Microplate Readers were tested for compatibility with the Life Technologies’ Z’-LYTE

Z’-LYTE® Compatible Microplate Reader

Documentation

Version No.: 08 Jan 15

Page 15 of 18

Setup Guide on the BMG LABTECH CLARIOstar® Microplate Readers

Have a question? Contact our Technical Support Team

NA: 800-955-6288 ext. 40266 Email: [email protected]

Materials Required But Not Supplied

384-well assay plates. We recommend black Corning® 384-well, low-volume, round-bottom (non-binding surface) assay plates (Corning #4514). Other black-walled, low-binding assay plates, while not tested, may be suitable.

Multi-channel pipettor (12 channels) or any pipetting device that can accurately deliver repeated volumes of 2.5 µl and 5 µl.

96-well microplates that can accommodate 300 µl per well

PI-103 (CAS 371935-74-9), MW 348.4

DMSO Procedure

1. Prepare initial 100X serial dilution curves in Row A of a 96-well plate: Dilute

PI-103 to a 100X initial concentration in 100% DMSO (500 µM). Prepare a set of 3-fold serial dilutions from the initial concentration in a 96-well plate: Add 30 µl DMSO in wells A2 - A10. Add 45 µl of 500 µM of PI-103 in A1. Add 15 µl from well A1 to well A2, and then mix well A2, and take 15 µl from well A2 and add to well A3, mix, and so on.

Figure 1: Schematic of initial compound dilution. PI-103 was titrated from a 500 µM starting concentration in the initial dilution series by preparing a 3-fold dilution in 100% DMSO. A secondary dilution to 4X was then prepared in the rows below the initial dilution curve (lighter gray) using kinase buffer.

2. Prepare a 4X Intermediate Dilution of PI-103 in Row B: Add 48 µl of 1X Kinase Buffer A into all wells B1 – B10. Add 2 µl of diluted inhibitor from the wells in

Page 16: Z’-LYTE Assay Setup Guide on the BMG LABTECH CLARIOstar … · The BMG LABTECH CLARIOstar® Microplate Readers were tested for compatibility with the Life Technologies’ Z’-LYTE

Z’-LYTE® Compatible Microplate Reader

Documentation

Version No.: 08 Jan 15

Page 16 of 18

Setup Guide on the BMG LABTECH CLARIOstar® Microplate Readers

Have a question? Contact our Technical Support Team

NA: 800-955-6288 ext. 40266 Email: [email protected]

Row A to wells in Row B. This will produce a final serial dilution starting at 20 µM (B1), which will then produce a final assay concentration starting at 5 µM.

3. Prepare an Assay Plate: Prepare duplicate wells. Add 2.5 μL of the 4X

compound dilutions per well into a low volume NBS, 384-well plate (Corning Cat. # 4514)., Using row 1, Columns 1 and 2 are replicates of the highest concentration, Columns 3 and 4 are replicates of the second highest concentration, etc. to column 20.

4. Add 2.5 μl of 1X Kinase Buffer A alone to Columns 21 and 22 (0% inhibition, no

compound control), 23 (0% phosphorylation control, no kinase added) and 24 (Phosphopeptide 100% phosphorylation positive control)

5. Add 5 μL of the 2X Peptide/Kinase Mixture (prepared in mTOR kinase buffer) to

Row 1 Columns 1-22. DO NOT ADD TO COLUMN 23 OR 24. 0% phosphorylation control: Add 5 μL of 2 μM substrate alone without kinase to

column 23, rows A-L. 100% phosphorylation control: Add 5 µl of 2 µM phosphopeptide control

substrate to column 24, rows A-L. Buffer-only reference: Add 5 µl Kinase Buffer A alone to the remaining 8 wells

(columns 23 and 24, rows M-P).

6. Add 2.5 μL of 4X ATP Solution (200 μM) per well to all columns to start reaction.

7. Shake assay plate on a plate shaker for 30 seconds.

8. Incubate assay plate for 60 minutes at room temperature.

9. Add 5 μL of the Development Reagent Solution to each well. Use the lot-specific dilutions indicated on your Certificate of Analysis as dilutions may vary based upon Z'-LYTE® peptide and Development Reagent A lot.

10. Shake plate again on a plate shaker for 30 seconds.

11. Incubate for 60 minutes at room temperature.

12. Read and analyze as directed in the protocol.

Page 17: Z’-LYTE Assay Setup Guide on the BMG LABTECH CLARIOstar … · The BMG LABTECH CLARIOstar® Microplate Readers were tested for compatibility with the Life Technologies’ Z’-LYTE

Z’-LYTE® Compatible Microplate Reader

Documentation

Version No.: 08 Jan 15

Page 17 of 18

Setup Guide on the BMG LABTECH CLARIOstar® Microplate Readers

Have a question? Contact our Technical Support Team

NA: 800-955-6288 ext. 40266 Email: [email protected]

1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24

A

B

C

D

E

F

G

H

I

J

K

L

M

N

O

P Figure 2: Assay Plate Schematic. Compound titrations, duplicate 10-point titration shown in row A columns 1-20 with known inhibitor (e.g. PI-103). Columns 21 and 22 prepared without any inhibitor as kinase activity controls (purple), column 23 prepared with no kinase, 0% phosphorylation, (blue) and column 24 prepared using phosphopeptide control, 100% phosphorylation, (green). Note 8 wells in bottom right (yellow M23 – P24), were prepared without any inhibitor or substrates, as buffer controls. White wells can be used for the assay test compounds.

Page 18: Z’-LYTE Assay Setup Guide on the BMG LABTECH CLARIOstar … · The BMG LABTECH CLARIOstar® Microplate Readers were tested for compatibility with the Life Technologies’ Z’-LYTE

Z’-LYTE® Compatible Microplate Reader

Documentation

Version No.: 08 Jan 15

Page 18 of 18

Setup Guide on the BMG LABTECH CLARIOstar® Microplate Readers

Have a question? Contact our Technical Support Team

NA: 800-955-6288 ext. 40266 Email: [email protected]

D. Results

PI-103

0.1 1 10 100 1000 100003.0

3.5

4.0

4.5

5.0

5.5

6.0

6.5

7.0

7.5

IC50 = 49 nM

[PI-103] (nM)

Em

issio

n

Rati

o

Figure 1: Z'-LYTE® Kinase Assay. Z’-LYTE® Assay using Ser/Thr 11 (PV3670) with FRAP1 (mTOR) (PV4753) performed using the BMG LABTECH CLARIOstar®.