zeiss neon 40 esb sample insertion/removal
DESCRIPTION
Zeiss Neon 40 EsB Sample Insertion/Removal Load samples (or not if removing) and evacuate the airlock While wearing gloves pick an appropriate stage and load the samples Open the airlock door (swivel the monitor out of the way) and fit the dovetail stage securely onto one of the two teflon holders On the airlock, the red ‘Wait’ light should be lit as well as the yellow ‘Rod Status’ and ‘Gate Closed’ lights While holding the chamber door closed, push the ‘Pump’ button (green light on) followed by the ‘Open’ button (green light on) When the green ‘Proceed’ light turns on and the yellow ‘Gate Closed’ light turns off, wait until the gate valve fully opens (~30 seconds, look for motor rolling to cease) and the sample chamber lights upTRANSCRIPT
Zeiss Neon 40 EsB Sample Insertion/Removal
Close the column chamber valve and make sure sample stage is at the
exchange position Click on Close Column Chamber Valve under the
Airlock panel Using the ChamberScope, make sure sample stage is at
0o tilt and lowered out of the way of the SEM/FIB columns Under the
Stage Points List panel, double click on $exchange to drive the
stage to the exchange position Zeiss Neon 40 EsB Sample
Insertion/Removal
Load samples (or not if removing) and evacuate the airlock While
wearing gloves pick an appropriate stage and load the samples Open
the airlock door (swivel the monitor out of the way) and fit the
dovetail stage securely onto one of the two teflon holders On the
airlock, the red Wait light should be lit as well as the yellow Rod
Status and Gate Closed lights While holding the chamber door
closed, push the Pump button (green light on) followed by the Open
button (green light on) When the green Proceed light turns on and
the yellow Gate Closed light turns off, wait until the gate valve
fully opens (~30 seconds, look for motor rolling to cease) and the
sample chamber lights up Zeiss Neon 40 EsB Sample
Insertion/Removal
Insert or remove the specimen stage from the SEM chamber If
necessary, line up the sample insertion rod with the position of
the sample using the knob underneath the rod Release the lock on
the sample insertion rod and carefully screw the sample insertion
rod into the stage Be careful not to push the stage off the holder
and do not over tighten the rod Insert the sample into the SEM
chamber fitting the dovetail stage onto the holder securely Unscrew
the sample insertion rod from the sample holder (Rod Status light
should be solid yellow, not blinking otherwise system will not
proceed to the next step), retract it all the way and lock it into
the fully withdrawn position Zeiss Neon 40 EsB Sample
Insertion/Removal
Vent the airlock to atmosphere and open the column chamber valve
Release the Open button (green light off) and push the Close button
(green light on) to close the gate valve Wait until the Gate Closed
yellow light comes on Release the Pump button (green light off) and
then the Close button (green light off) Push the Purge button
(green light on) to vent the airlock to atmosphere The red Wait
light will turn on When the airlock is fully vented (~ 10 seconds),
release the Purge button (green light off) Open the column chamber
valve on the Airlock panel