95°C Lysis Block Workflow Tray Transfer Rack VORTEX 5 seconds 7 HSV/VZV HSV/VZV HSV/VZV HSV/VZV HSV/VZV HSV/VZV Kit Contents Equipment Reaction Tubes Process Buffer Tube Assay Procedure – Heat Lysis VORTEX 5 seconds VORTEX 5 seconds Rehydrate by pipetting up and down a minimum of 5 times. Heat for 5 minutes at 95°C. Close lid tightly and proceed to next step. Transfer 50 μL to each Reaction Tube. Specimen in transport media. PCM302000EN00 (08/16) Amplification and Detection Add 20 μL to Process Buffer Tube. 1 3 4 2 5 Solana Instrument 1 3 4 2 5 Close the lid and select appropriate assay protocol. Run complete in 50 minutes. Review results. 6 Setup the Workflow Tray with patient samples, Process Buffer Tubes and Reaction Tubes. 6 8 7 Use the Transfer Rack to lift Reaction Tubes from the Workflow Tray. Lower the Reaction Tubes into Solana. Release the Reaction Tubes from the Transfer Rack by pulling the Transfer Rack toward your body. Procedure Card HSV 1+2/VZV Assay Study the Package Insert thoroughly before using this Procedure Card. This is not a complete Package Insert.

Upload: tranque

Post on 08-May-2018

213 views

Category:

Documents


1 download

TRANSCRIPT

95°CLysis Block

WorkflowTray

TransferRack

VORTEX5 seconds

7

HSV/VZV

HSV/VZV

HSV/VZV

HSV/VZV

HSV/VZV

HSV/VZV

Kit Contents Equipment

ReactionTubes

ProcessBufferTube

Assay Procedure – Heat Lysis

VORTEX5 seconds

VORTEX5 seconds

Rehydrate by pipetting upand down a minimum of

5 times.

Heat for 5 minutesat 95°C.

Close lid tightly andproceed to next step.

Transfer 50 µL to eachReaction Tube.

Specimen intransport media.

PCM302000EN00 (08/16)

Amplification and Detection

Add 20 µL to Process

Buffer Tube.

1 3 42 5

SolanaInstrument

1 3 42

5

Close the lid and selectappropriate assay protocol.

Run complete in 50 minutes.

Review results.

6

Setup the Workflow Tray withpatient samples, Process Buffer

Tubes and Reaction Tubes.

6 87

Use the Transfer Rack to liftReaction Tubes from the

Workflow Tray.

Lower the Reaction Tubesinto Solana.

Release the Reaction Tubesfrom the Transfer Rack bypulling the Transfer Rack

toward your body.

Procedure Card