1-509-332-2756 [email protected] apor sorption analyzermanuals.decagon.com/quick start...
TRANSCRIPT
236
5 N
E H
op
kin
s C
ou
rt
Pullm
an, W
ash
ing
ton
99
163
USA
1-
509-
332-
275
6in
stru
men
ts@
aqu
alab
.co
m
ww
w.a
qu
alab
.co
m
1451
0 | 1
3-01
-16
Vapor
Sorp
tion
A
naly
zer
Quic
k St
art
Guid
e
VS
A Q
uic
k Sta
rt Gu
ide
• The delete en
try is used to rem
ove an unwanted stage.
• Follow
the m
essage on the screen
to place an em
pty stain
less steel cup in
the ch
amber an
d seal it. Press
to tare th
e cup.
• Follow
the m
essage on the screen
to place the sam
ple in
the cu
p and press
to weigh
t the sam
ple. • P
ress to begin
the test.
5. Running a w
ater activity test— (C
an not be done when
isotherm test is running)
• R
emove th
e weigh
ing pan
. • P
ress until you
are at the M
easurem
ent Tab (2n
d tab). Move latch
left and/or press
to start a reading. T
he
pin
wheel w
ill spin indicatin
g a measu
rement is ru
nning.
• T
he a
w reading for th
e sample an
d the tem
perature w
ill display w
hen finished.
• T
o change betw
een dew
point an
d capacitive readings,
you
will n
eed to toggle the Volatiles Y
es/No option
in
the C
onfiguration
Tab.
Volatiles = Y
es g Capacitive Sen
sor Volatiles = N
o g Dew Poin
t Sensor
Note: A
ccuracy ch
anges to +/- 0.02 w
hen
volatiles “Yes”
is selected
.
VS
A Q
uic
k Sta
rt Gu
ide
1. Getting Started—
• P
lace the A
quaLab V
SA on a fl
at surface. U
se the bu
bble level(Figu
re 1: Adjustm
ent an
d bubble level) an
d use th
e adju
stable feet to level the in
strument. T
he sm
all rubber
feet on
the back
of the low
er plate are to prevent tip-over
and do n
ot need to be tou
ching th
e table. • U
se the A
quaLab V
SA where en
vironmental tem
perature
and humidity rem
ain fairly stable.
Figure 1: A
dju
stmen
t and
bu
bb
le level.
• T
he A
quaLab V
SA needs to be on
a stable platform and
evalu
ated to verify shippin
g did not dam
age the scale
perform
ance. T
o test scale stability: 1. T
urn on the V
SA and allow
instru
ment to w
arm
u
p for 2 hours.
2. N
avigate to the D
iagnostics screen
in the
Configuration
menu and press
. 3. Scroll dow
n to th
e screen that sh
ows th
e weigh
t. Readin
gs should vary n
o more th
an 0.5 m
g.
Adjustm
ent feet:
Two on
front an
d one
on the m
iddle back.
Bu
bble Level
Lower B
lock
VS
A Q
uic
k S
tart
Gu
ide
the up and down arrows. Choosse any value between
0.003 and 1.0
a w. The step setting for DVS determines
th
e a w values for equilibration and consequently the
resolution. If starting
a w is 0.1 and has a step setting of
0.10 aw results in humidity being controlled to 0.1, 0.2,
0.3 … up to the final aw, remaining at each level until
either equilibrium is achieved based on the trigger value
or it reaches the timeout value. Setting a higher
value makes the test faster, but decreases the quantity
of equilibtation points.
• To choose the trigger %dm
/dt value that indicates
equilibration, press enter with “Trigger” highlighted
and then cycle to the desired value using the up and
down arrows. The trigger can be shut off or be any
value between 0.001-1%/hr. Choose the number of
events required to meet the trigger value. The trigger
value represents an acceptable change in mass per
change in time to indicate equilibrium for a given step.
Setting a higher value will make the test faster, but
may not result in com
plete equilibrium. If the trigger
is turned off, the sample will be held at each step for
the time indicated by the timeout setting.
• To choose to include a timeout setting, press with
“Timeout” highlighted and then cycle to the desired
value using the up and down arrows. You can select any
value between 5.0 min to 30 days. The timeout
determines the maximum time allotted for each aw step
NOT the time from
starting
a w to final aw. The default
value is off, which means there is no timeout value
and only the trigger value determines the end of step.
If you set both a trigger value and timeout value,
whichever VSA reaches first determines the end of step.
• To choose to have the isotherm loop, press with
“Loop” highlighted to toggle between on and off.
Turning loop on causes the isotherm to automatically
return to the start
a w (with the same settings) once
the final aw has been achieved. This would create
both an adsorption and desorption curve (or vice versa
depending on the starting direction).
VS
A Q
uic
k S
tart
Gu
ide
4. Lightly tap the table with one finger. If the readings
vary more than 0.5 mg after tapping, the table is
probably too unstable for accurate weight readings
during a test and will need to be made more rigid
or another location found for this equipment.
Contact your AquaLab distributor if the scale
readings are unstable after placing the unit on a
stable structure.
Figu
re 2
: Des
icca
nt
inst
alla
tion
• Desiccant tubes are installed by orienting the tube so that
the black colored valve inserts into the black receiving hole
as shown in Figure 2. Before insertion make sure that the
desiccant tube lid is firmly sealed onto the desiccant tube.
(See Manual for instructions) Lift the tube into the area
above the desiccant receiver and then insert the tube into
the mating holes while shaking the tube slightly. Once in
place, the desiccant tube is sealed for flow purposes and
may move side to side if contacted.
Desiccant Cartridge
Desiccant Receiver
Black Valve
Black Receiving Hole
VS
A Q
uic
k Sta
rt Gu
ide
• To ch
oose to include a tim
eout settin
g, press with
“Timeou
t” highligh
ted, and th
en cycle to th
e desired valu
e using th
e up an
d down arrow
s. Any valu
e between
5.0 m
in to 30 days can
be selected. The tim
eout
determines th
e maxim
um tim
e allotted to move from
th
e starting a
w to the final a
w (DDI on
ly) and is option
al. T
he defau
lt value is off, w
hich mean
s there is n
o tim
eout valu
e.
• T
o choose to h
ave the isoth
erm loop, press
with
“Loop” h
ighligh
ted to toggle between
on and off.
Turning loop on
causes th
e isotherm
to autom
atically retu
rn to th
e start aw (w
ith the sam
e settings) on
ce the final a
w has been
achieved. T
his w
ould create both
an
adsorption and desorption
curve (or vice versa
dependin
g on the startin
g direction).
• For D
VS m
ethod
• T
o choose th
e starting a
w , press with “Start”
highligh
ted, and th
en cycle to th
e desired value using
the up an
d down arrow
s. Any valu
e between
0.03 and 0.95 a
w can be ch
osen includin
g “curren
t”. Choosin
g “curren
t” mean
s that th
e stage will start at
th
e initial a
w of the sam
ple. Press
when finished.
• To ch
oose the final a
w , press with “Fin
al” highligh
ted, and th
en cycle to th
e desired value using
the up an
d down arrow
s. Any valu
e between
0.03 and 0.95 a
w can be ch
osen includin
g curren
t (see“cu
rrent” defi
nition
above). Selecting a fi
nal a
w that is low
er than the startin
g aw w
ill result in
desorption
, while selectin
g a final a
w higher th
an the
startin
g aw w
ill result in
adsorption. Press
when
finished.
• To ch
oose the tem
perature of th
e stage, press with
“Temperatu
re” highligh
ted, and th
en cycle to th
e desired tem
perature u
sing th
e up an
d down arrow
keys. A
ny tem
perature betw
een 15-60°C
can be
chosen
. Each
stage can run at a u
niqu
e temperatu
re.
• T
o choose th
e desired aw step, press
with “Step”
highligh
ted and th
en cycle to th
e desired value using
VS
A Q
uic
k Sta
rt Gu
ide
Figure 3: Tan
k fi
ll extension
• R
emove th
e black plu
g on the clear plastic tan
k fill
exten
sion located on
the back
right of th
e VSA.
(Figu
re 3). Fill the tan
k with deion
ized water u
ntil fu
ll. Fin
ger tighten
the w
ater fill plu
g once fi
nished.
• T
o remove th
e water from
the w
ater tank, th
e water tan
k
drain
plug is u
sed. • Figu
re 3 shows th
e location of a fi
lter that w
ill need to be
rem
oved from tim
e to time w
hen dust bu
ilds up on
it. The
housing arou
nd th
e filter is a sn
ap on housing th
at is easily rem
oved.
Water Fill P
lug
Water T
ank Fill E
xtension
Upper H
orizontal D
rilled Hole
Water T
ank Drain
Plug
Fan Filter
VS
A Q
uic
k S
tart
Gu
ide
•
For
DD
I m
etho
d
• To choose the starting a w, press with “Start”
highlighted and then cycle to the desired value using
the up and down arrows. Any value between 0.03 and
0.95 a w can be chosen including “current”. Choosing
“current” means that the stage will start at the initial a w
of the sample. Press when finished.
• To choose the final aw, press with “Final” highlighted
and then cycle to the desired value using the up and
down arrows. Any value between 0.03 and 0.95 aw c
an
be chosen including current (see “current” definition
above). Selecting a final aw that is lower than the
starting a w will result in desorption, while selecting a
final aw higher than the starting a w results in adsorption.
Press when finished.
• To choose the temperature of the stage, press with
“Temperature” highlighted and then cycle to the
desired temperature using the up and down arrow keys
(hold down arrows for accelerated scrolling). You may
choose any temperature between 15 and 60 °C, 25 °C is
default. Each stage can run at a unique temperature.
• To choose the desired
a w resolution, highlight
“Resolution” press
with “Resolution” highlighted and
then cycle to the desired value using the up and down
arrows. You may choose any value between 0.003
and 1.0
a w. The resolution for DDI determines the target
a
w resolution. Setting a higher resolution value makes
the test faster, but decreases the
a w resolution. W
e recom
mend a DDI resolution of 0.01 aw.
• To choose the air flow rate, press with “Flow”
highlighted, and then cycle to the desired value using
the up and down arrows. Any value between
10-163 ml/min can be selected. Higher flow rates will
speed up the test, but may result in lower data
resolution and non-equilibrium conditions. The
default value of 100 ml/min is suitable for most
products.
VS
A Q
uic
k S
tart
Gu
ide
Figu
re 4
: Sam
ple
cup
and
low
er b
lock
ass
emb
ly
• Figure 4: assembly illustrates the proper positioning of the
parts in the lower block. The air cylinder is placed into
the bore of the base block. It should slide into the block
easily and settle at the bottom
of the bore. The weigh
pan connects to the scale and holds the sample cup. For
best results, verify cleanliness of all parts.
2. A
quaL
ab V
SA P
erfo
rman
ce V
erifi
cati
on o
verv
iew
—
• AquaLab VSA performance and cleanliness should be
checked frequently; before each isotherm test is
recommended.
• Press until you are at the Measurement Tab.
• Use AquaLab Verification Standards to verify performance
or adjust linear offset.
Sam
ple
Cup
Lid
Only used for sample storage and
must be removed during tests.
Sam
ple
Cup
Stainless steel cups are
recommended. Plastic cups
absorb moisture over time.
Wei
ghin
g P
an
Air Cylinder
Base Block Bore
VS
A Q
uic
k Sta
rt Gu
ide
• E
nsure sam
ple temperatu
re is within 4 °C
of chamber
temperatu
re.
4. Starting a Test—
• B
efore starting a test, set th
e Volatiles: Y
es/No option
in
the con
figuration
menu. If you
r sample h
as volatiles, select “Y
es” otherw
ise select “No”.
• O
n the m
ain screen
, select the test w
izard .
• Follow
ing several w
arnings, begin
configuring th
e test by providin
g a name for th
e test using th
e arrow keys
and th
en press
. (The name of a previou
s test will appear
by default.)
• E
nter in
itial moistu
re conten
t if known (option
al). Press
to con
tinue or
to go back.
• Setu
p each stage of th
e isotherm
test. A stage is u
sed to adju
st the isoth
erm settin
gs for an individu
al sample (u
p to 20 stages). E
ach stage is setu
p with the isoth
erm meth
od (DDI or D
VS) an
d the settin
gs for that portion
of the
isoth
erm. Once a stage is created, it can
be modifi
ed or deleted an
ytime before th
e test is started.
i.e. For m
ilk pow
der, Stage 1 cou
ld u
se the D
DI m
ethod
to
ram
p qu
ickly from
0.1 to 0.4 aw an
d Stage 2 cou
ld u
se
th
e DV
S meth
od to step from
0.4 to 0.8 aw slow
ly in
0.05 aw in
cremen
ts.
• T
o setup a n
ew stage, select “n
ew” an
d press . A screen
with several param
eters will appear. If a previou
s test has been
run, th
e stages for that test w
ill be listed by defau
lt. These stages can
be used as-is or u
pdated with
new settin
gs. If you need to delete a stage, go to th
e specifi
c stage and press
. Scroll down to th
e bottom of
the stage settin
gs select “Delete” an
d press .
• Select th
e isotherm
meth
od as either D
ynamic D
ewpoin
t Isoth
erm (DDI) or D
ynamic V
apor Sorption (DVS). D
DI
will create a dyn
amic isoth
erm and D
VS w
ill create a static or equ
ilibrium isoth
erm. The oth
er choices w
ill adju
st based on which meth
od is selected.
VS
A Q
uic
k Sta
rt Gu
ide
• T
o verify, remove w
eigh pan
and th
en place sam
ple cup
con
taining on
e of AquaLab’s verifi
cation stan
dards in
sam
ple chamber. C
lose lid and carefu
lly avoid spillage; See Figu
re 5. • M
ove latch left or press
to start a reading. T
he
pin
wheel spin
s to indicate a m
easurem
ent is ru
nning.
The a
w reading for th
e standard an
d the tem
perature
displays w
hen finished. M
ake su
re the a
w reading
on the stan
dard is within +/-0.005 of th
e correct aw for th
e stan
dard. (+/- 0.02 if cap sensor is u
sed) • If an
offset adjustm
ent is n
eeded, refer to chapter 7 in
the
user’s m
anual.
Figure 5: Sam
ple chamber
3. Sample Preparation—
• Fill sam
ple cup no m
ore than half fu
ll (~7ml).
Do N
ot Overfill!
• Sam
ple size can be betw
een 500 an
d 2,000 mg. W
e recom
mend 1000 m
g. • E
nsure th
e rim and ou
tside of sample cu
p are clean.
Lid
Base