13 chapter-iii fexofenadine odt method development and method...
TRANSCRIPT
Page 1 of 29
CHAPTER III
FEXOFENADINE ODT METHOD DEVELOPMENT AND
METHOD VALIDATION
This chapter describes about the fexofenadine formulation sample development and
validation of a stability indicating Revere phase-Ultra performace LC procedure fexofenadine
in pharmaceutical formulations. This chapter also describes about materials used and scope of
the present study. It describes the reagents, solvents, glass-ware, standards and test samples
(dosage forms) used in this present study, brief description of equipment and methods
employed in the present study.
3.1 REAGENTS AND CHEMICALS
Sodium hydroxide (NaOH):
Merck supplied AR grade Sodium hydroxide was utilized in this study.
Hydrogen Peroxide (H2O2):
Merck supplied Analytical grade (AR) Hydrogen peroxide (30%) was used for this
study.
Ortho-phosphoric acid (H3PO4):
J.T.Baker, Mumbai supplied analytical grade (AR) of ortho-phosphoric acid was used
for the analysis.
Hydrochloric acid (HCl):
Rankem supplied hydrochloric acid was used for the analysis.
Tri-ethyl amine:
J. T. Baker, Mumbai supplied analytical grade (AR) of Tri-ethyl amine was used for
the analysis.
3.2 SOLVENTS
Acetonitrile (CH3CN):
HPLC grade solvent was used for the analysis, manufactured by Rankem and Merck
India Ltd.
Methanol (CH3OH):
HPLC grade solvent was used for the analysis, manufactured by Merck India Ltd and
Rankem Ltd.
Page 2 of 29
3.3 GLASSWARE:
Brand new of Borosil and Glaxo of class-A, grade glassware like pipette, volumetric
flask, conical flask, glass bottle, test tubes, measuring cylinder and glass beaker were used for
analysis.
3.4 STANDARDS:
Working standards of Fexofenadine hydrochloride and its impurity was used for this
study.
3.5 TEST SAMPLE (DOSAGE FORMS)
Finished products: Fexofenadine orally disintegrating tablet
3.6 EQUIPMENTS AND INSTRUMENTS
Ultra performance liquid chromatography (UPLC):
To develop and validate the chromatographic method Acquity UPLCTM was used. The
LC system has, a auto sampler. The signals recorded by Empower2 software.
Liquid Chromatography-Mass Spectroscopy (LC-MS):
To determine the mass by charge ratio of potential degradant in specificity (forced
degradation) experiment the LC-MS/MS system was utilized. Agilent make LC system which
is connected to Biosystem 4000 Q trap used for study.
Nuclear magnetic resonance (NMR):
The proton and carbon Nuclear magnetic resonance (NMR) experiment were
performed for potential degradant in drug product. Deuterated methanol used as solvent.
Frequency for 1H and 13C was 500 megahertz and 125 megahertz correspondingly applied.
Bruker make INOVA was used for the spectrum.Tetramethylsilane is used as internal
standard which chemical shift considered as Zero ppm.
Analytical balance:
Mettler Toledo make AB265-5/fact model is used for this research work.
Centrifuge:
Remi limited make centrifuge instrument is used for samples preparation.
pH Meter:
Mettler-Toledo, Switzerland make pH meter seven easy model is used to measure pH
of the solution for this work.
Sonicator:
Partech scientific suppliers manufactured 25 Liter sonicator and Life care
manufactured 10 Liter was used for analysis.
Page 3 of 29
Degasser:
Borosil manufactured glass degasser apparatus is used and Millipore 0.45 micron
filter paper is used for degassing.
Milli-Q water purification system:
Milli pore manufactured water purification system which is having 0.22µm membrane
filter is used for analysis.
Water bath:
Water bath, make Cintex, which is manufacture in India, was used for hydrolysis
study.
Photo stability chamber:
Photo-stability experiments were performed in this equipment which is from Sanyo,
Leicestershire, UK.
3.7 SCOPE OF THE PRESENT STUDY:
Keeping in view of the importance of active ingredients and related substances
(process and degradant related impurities) present in finished product to be monitored the
scope of the present study is designed. The present research work focuses on the
identification of an oxidative degradant impurity and the method development and validation
of quantification methods for active ingredients and its impurities present in drug products by
RP-UPLC which are sophisticated analytical techniques. FEX is an antihistamine active
substance utilized in the curing of allergy indication and fever urticaria. Terfenadine is
another choice of fexofenadine which is similar therapeutic effect. Fexofenadine used as an
antagonist to the H1 receptor.
The author selected the following active pharmaceutical drug substances and its
impurity for the present study. The Literature study reveals that no proposed validated test
procedure as available for separation and quantification of the below stated active ingredients
& its impurities present in pharmaceutical dosage forms by the proposed techniques.
Table- 3.7.a: Active Pharmaceutical Ingredients & Its Impurity Structures
Page 4 of 29
3.8 INTRODUCTION:
Fexofenadine, structure is given in
HO
Fig-3.8.a.It shows Structure and Chemical name of fexofenadine hydrochloride
In the literature survey there were few HPLC assay and dissolution methods have
been given. Besides the reported literature for N
method as per our present knowledge no literature specifies about a new potential degradant
of fexofenadine as N-Oxide impurity in fexofenadine dosage form. This degradation product
was formed in formal stability study and forc
elucidated by advanced spectroscopic techniques ESI
indicating chromatographic method is not reported for this impurity. The current research
study explains about isolation and
developmeint and validatiron of selective RP
degradants of fexofenadine and other relate
Page 5 of 29
structure is given in Fig. 3.8.a.,
O
OH
N
O
OH H Cl
3.8.a.It shows Structure and Chemical name of fexofenadine hydrochloride
In the literature survey there were few HPLC assay and dissolution methods have
Besides the reported literature for N-oxide impurity of fexofenadine and selective
method as per our present knowledge no literature specifies about a new potential degradant
Oxide impurity in fexofenadine dosage form. This degradation product
was formed in formal stability study and forced degradation study and its structure was
elucidated by advanced spectroscopic techniques ESI-MS/MS,1H and 13C NMR. The stability
indicating chromatographic method is not reported for this impurity. The current research
study explains about isolation and characterization of this impurity. It is also describes
on of selective RP-LC test procedure for the determination of
degradants of fexofenadine and other relatead substaances (Imp.-A, Imp.-B and Imp.
3.8.a.It shows Structure and Chemical name of fexofenadine hydrochloride
In the literature survey there were few HPLC assay and dissolution methods have
f fexofenadine and selective
method as per our present knowledge no literature specifies about a new potential degradant
Oxide impurity in fexofenadine dosage form. This degradation product
ed degradation study and its structure was
C NMR. The stability
indicating chromatographic method is not reported for this impurity. The current research
characterization of this impurity. It is also describes
LC test procedure for the determination of
B and Imp.-C). The
Page 6 of 29
optimized liquid chromatography procedure method validation was performed for specificity,
linearity, precision, recovery, LOD, LOQ and robustness. Force degradation study in
different stress conditions was performed on placebo and finished dosage forms to ensure the
selectivity of the developed test procedure. These parameters were studied in accordance with
ICH guidelines [17-18].
3.9 OPTIMIZED TEST PROCEDURE
Chromatographic Conditions:
Column: 100 x 2.1 millimeter ,Waters Aquity BEH C18 column, 1.7
micron
Mobile phase:
Solvent-A: pH 7.00±0.05 of water adjusted with OPA with 0.05%TEA
Solvent-B: water: Acetonitrile (10:90)
Gradient program: 0-10min, sol-B: 25-25; 10-15min, sol-B: 25-35; 15-33min, sol-
B: 35-60; 33-35min, sol-B: 60-80; 35-36min, sol-B: 80-25 and
36-40min, sol-B: 25-25.
Flow rate: 0.40 Milliliter per minute
Wavelength: 220nm
Injection volume: 1.5μL
Column Temperature: 30°C
Diluent:
Diluent was 50% Acetonitrile and 50% water mixture with 1 mL of OPA.
System suitability Solution:
Transferred an accurately weighed amount of 60 mg fexofenadine hydrochloride WS
to a 50 milliliter flask and added 35 milliliter of diluent. This sample is sonicated to dissolve
the material totally and then added 1.2 milliliter of impurity solution (0.1 milligram per
milliliter solution of impurity A, B and C) and added diluent to the mark.
Standard solution:
960 μg/mL standard stock solutions were prepared in diluent. About 96 mg of
fexofenadine hydrochloride WS was taken to 100 milliliter flask. It is dissolved in diluent
with vigorous shaking and sonication. Above 5.0mL solution was diluted 20 times in 100mL
volumetric flask with diluent. Further this diluted 20 times in 100mL volumetric flask with
Page 7 of 29
diluent to obtain 2.4 ppm concentration. Five replicate standard injections were made to
determine system suitability.
Test solution preparation:
Weighed separately not less than 20 tablets of fexofenadine ODT and triturated in a
mortar to get fine and homogeneous sample. Separately weighed the powdered sample of
fexofenadine hydrochloride equivalent to 120mg and transferred in to 100 milliliter flasks,
sonicated for 25 minutes with intermediate shivering and sample were diluted volumetric
flask till mark with diluent, centrifuged the sample solutions at 10000 RPM for 10minutes.
Procedure:
Separately injected equal 1.5μL volume of blank, system suitability solution, standard
preparation, test solution in UPLC to record the chromatograms and measured the areas of
peaks in the chromatograms and calculated the results using below formula.
System Suitability:
Fexofenadine peak tailing in standard solution was NMT two, the USP plate count
was not more than 5000 for fexofenadine peak forstd preparation, The Resolution between
Impurity B peak and fexofenadine from system suitability preparation was NLT 3.0 and the
ratio of fexofenadine peak areas from two standard injections was between 0.9 to 1.1.
Calculation:
Percentage (%) of known of impurity =
At*WS*5*5*100*P*Average weight of tablet *100 As*100*100*100*WT*100*L*RRF
Here, “As” is the peak response of fexofenadine in STD. “At” is the peak response of
known impurity in sample preparation.“WT” is fexofenadine content weight taken for sample
preparation in milligram. “WS” is the fexofenadine hydrochloride standard weight taken for
standard preparation in milligram. “P” is fexofenadine potency as fexofenadine
hydrochloride. “L” is the Label claim of fexofenadine hydrochloride in mg per tablet and
''RRF'' is relative response factor of each know impurity.
Percentage (%) of unknown of impurity =
Au * WS * 5 * 5 * 100 * P * Average weight of tablet * 100 As * 100 * 100 * 100 * WT * 100 * L
Where in “As” is the peak response of fexofenadine in STD preparation. “Au” is the
peak response of unknown impurity in sample preparation. “WT” = fexofenadine content
weight taken for sample preparation in milligram. “WS” is the fexofenadine hydrochloride
Page 8 of 29
standard weight taken for STD preparation in milligram. “P” is fexofenadine potency as
fexofenadine hydrochloride and “L” is the Label claim of fexofenadine hydrochloride in mg
per tablet.
3.10 METHOD DEVELOPMENT:
IR Spectra:
Fexofenadine samples were separately scanned in FT-IR range from 4000 to 450cm-1
and the individual IR Spectra’s are represented as Spectrum- 3.10.a
Spectrum- 3.10.a: FTIR spectrum of Fexofenadine
Selection of wavelength:
Fexofenadine hydrochloride and its impurities were scanned across the UV range
from 200-400nm using 2996 PDA Detector. All Fexofenadine hydrochloride and its
impurities have UV maximum absorbance. Spectrum represents the absorption spectra which
are presented in Figure 3.10.a.
Fig 3.10.a:
The above UV spectra indicate that, fexofenadine an
Absorbance at 220 nm. The results were found satisfactory and reproducible at this
wavelength.
Selection of UPLC Column:
Preliminary phase of trial for method development started by isocratic method and
100mm x 2.1 millimeter Acquity BEH C8 which has particle size of 1.7µm column but Imp
B peak was merged with fexofenadine peak and baseline disturbance found. To resolve the
imp-B peak using same column, mobile phase used as per pH 3.0 chlorate buffer as sol
acetonitrile as sol-B with gradient elution experiment was done. To achieve resolution
column stationary phase changed to C18 and Ethylene bridged hybrid (
0.21centimeter) 1.7 micron column. Finally separation was achieved in sol
0.05% TEA, 7.0 pH with OPA and sol
column was proved for Reproducibility, and Ruggedness.
Page 9 of 29
Fig 3.10.a: UV spectrum of Fexofenadine and its impurities
The above UV spectra indicate that, fexofenadine and its impurities have maximum
Absorbance at 220 nm. The results were found satisfactory and reproducible at this
Preliminary phase of trial for method development started by isocratic method and
Acquity BEH C8 which has particle size of 1.7µm column but Imp
B peak was merged with fexofenadine peak and baseline disturbance found. To resolve the
B peak using same column, mobile phase used as per pH 3.0 chlorate buffer as sol
B with gradient elution experiment was done. To achieve resolution
column stationary phase changed to C18 and Ethylene bridged hybrid (
meter) 1.7 micron column. Finally separation was achieved in sol
pH with OPA and sol-B consists of 10 % H2O and 90% ACN mixture. This
column was proved for Reproducibility, and Ruggedness.
UV spectrum of Fexofenadine and its impurities
d its impurities have maximum
Absorbance at 220 nm. The results were found satisfactory and reproducible at this
Preliminary phase of trial for method development started by isocratic method and
Acquity BEH C8 which has particle size of 1.7µm column but Imp-
B peak was merged with fexofenadine peak and baseline disturbance found. To resolve the
B peak using same column, mobile phase used as per pH 3.0 chlorate buffer as sol-A and
B with gradient elution experiment was done. To achieve resolution
column stationary phase changed to C18 and Ethylene bridged hybrid (10centimeter x
meter) 1.7 micron column. Finally separation was achieved in sol-A which has
B consists of 10 % H2O and 90% ACN mixture. This
Page 10 of 29
Selection of Diluent:
Selection of diluent for fexofenadine and its impurity was a major challenge in the
method development as it is unreservedly soluble in CH3CH2OH and CH3OH, insoluble in
hexane & in water its slightly soluble. All of molecules are soluble in Acetonitrile and
Methanol and solubility is very less in water. Initially water and acetonitrile mixture was used
for dilution but because of peak shape of impurity 0.1% orthophosphoric acid added in
diluent. Finally, Acetonitrile and water in the proportion of 1:1, added 1 mL of ortho
phosphoric acid was used as diluents.
Selection of Run Time and Flow Rate:
With this developed method the retention times for fexofenadine and its impurities
were found within 40 min where mobile phase FR is 0.4 milliliter per minute. Above 0.4
milliliter per minute flow rates leads to early retention and merging of impurities it causes
quantification problems for fexofenadine impurities.
Selection of Standard and Sample Solution Concentration:
The standard concentration was based on 0.2% concentration of test concentration.
The sample concentration selected as 1.2 mg/mL to get LOD and LOQ of impurity below
reporting threshold as per ICH guideline.
Selection of Other Chromatographic Parameters:
The test procedure development was initiated with the prime focus on the resolution
of the impurities from main peak. The aim also predefined that less chromatographic time
non-polar Imp-C should properly elute. The solution having 1.2mg/milliliter of fexofenadine,
2.4 microgram/milliliter of all impurities was as used for development. A gradient method
used with sol-A which has 50 Millimolar NaH2PO4 + 10 millimolar NaoCl pH three and
100% of acettonitrile used as Sol-B. UPLC column was used as Ethylene bridged hybrid C8,
100 millimeter x 0.21 centimeter 1.7 micron column particle size. The FR was 0.5 milliliter
per minute on UPLC equipped with PDA detector. Imp-B peak was merged with
fexofenadine peak and at RT of Imp-C baseline disturbance found. To achieve resolution
column stationary phase changed to C18 and Ethylene bridged hybrid (10 centimeter x 0.21
centrimeter) 1.7 micron column. Finally separation was achieved in sol-A which has 0.05%
TEA, 7.0 pH with OPA and sol-B consists of 10 % H2O and 90% ACN mixture. After
altering of gradient, column and MP, all peaks were well resolved, satisfactory results were
obtained also confirm the column lot to lot variation. Based on above study below mentioned
method conditions were further optimized.
Page 11 of 29
Ethylene bridged hybrid C-18 (10 centimeter x 0.21 centrimeter) 1.7 micron column
was as UPLC column. Sol-A which has 0.05% TEA, 7.0 pH with OPA and sol-B consists of
10 % H2O and 90% ACN mixture used as mobile phase. The FR was 0.4 milliliter per min
with a gradient program (Time/%B) was set 0/25, 10/25, 15/35, 33/60, 35/80, 36/25 and
40/25. The column temp was kept at 30 degree centigrade. The output signals were recorded
at 220 nanometer. The vol of inj. was 1.5µL.
3.11 IDENTIFICATION AND CHARACTERIZATION OF OXIDATIVE
IMPURITY:
Semi-Preparative LC Parameter:
Isolation and resolution of degradant was performed on Inertsil octa decyl sillane 3V
(1 centimeter × 25 centimeter, particle size 5 μm) LC column. MP solvent sol-A which has
10 Milli molar ammonium formate solution and sol-B consists of 10 % 10 Milli molar
ammonium formate and 90% ACN mixture. The FR was 4.0 milliliter per minute. The
gradient (Time/%B) programmed 0/20, 10/35, 25/50, 35/90, 40/100, 60/100, 65/20 and 75/20.
The detection wavelength was 220 nanometer. The column oven temp was set at 25°C.
Liquid Chromatography-Mass Spectrometry (LC-MS) Conditions:
To establish the mass by charge ratio of potential degradant in specificity (forced
degradation) experiment the LC-MS/MS system was utilized. Agilent make LC system which
is connected to Biosystem 4000 Q trap used for study. The test procedure was optimized on
Zorbax Stable bond Phenyl, 25centimeter X 0.46 centimeter, five micron column. MP was
having a gradient program. MP solvent sol-A which has 10 Milli molar ammonium formate
solution and Sol-B consists of 10 % 10 Milli molar ammonium formate and 90% ACN
mixture. Both the solvent mixture was degassed before use. The gradient elution program
was set as (Time (min)/% sol-B) 0/30, 7/40, 20/50, 25/90, 27/100, 50/100, 52/30 and 60/30.
The FR was kept as 1 milliliter per min. The sample was injected as 20 micro liter. The
column oven temp was set as 25 degree centigrade. +ve ESI mode was used for analysis.
NMR Conditions
NMR and HRMS Conditions:
The proton and carbon Nuclear magnetic resonance (NMR) experiment were
performed for potential degradant in drug product. Deuterated methanol used as solvent.
Frequency for 1H and 13C was 500 megahertz and 125 megahertz correspondingly applied.
Bruker make INOVA was used for the spectrum.Tetramethylsilane is used as internal
standard which chemical shift considered as Zero ppm.
Identification of N-Oxide Impurity
During the drug product stability evaluation and specificity study, especially in
peroxide degradation experiment 0.71 RRT impurity was found potential. In order to enhance
the impurity, fexofenadine sample oxidation study was performed at 80
eight hours. The impurity has been found to enhance up to 65
impurity was performed on semi
impurity was and evaporated by using rotavapour with 40 d
resultant aqueous layer was lyophilized. The purity of the lyophilized impurity was checked
in HPLC and found highly pure. Further it is utilized to identify the impurity by advanced
technique.
The positive electro-spray ioniza
mass to charge ratio 518 amu [M+H]
fexofenadine. The m/z result reveals the existence of extra ‘O’ oxygen in potential degradant.
Employ HRMS found C32H39NO
Fig. 3.11.a. Mass spectrum of N
Page 12 of 29
xide Impurity
During the drug product stability evaluation and specificity study, especially in
peroxide degradation experiment 0.71 RRT impurity was found potential. In order to enhance
the impurity, fexofenadine sample oxidation study was performed at 80 degree centigrade for
eight hours. The impurity has been found to enhance up to 65-70%. The isolation of the
impurity was performed on semi-prep chromatographic system. Combined isolated portion of
impurity was and evaporated by using rotavapour with 40 degree centigrade vacuum. The
resultant aqueous layer was lyophilized. The purity of the lyophilized impurity was checked
in HPLC and found highly pure. Further it is utilized to identify the impurity by advanced
spray ionization mass spectrometer of potential degradant found
mass to charge ratio 518 amu [M+H]+ (Fig.3.11.a). This m/z is 16 amu additional than
fexofenadine. The m/z result reveals the existence of extra ‘O’ oxygen in potential degradant.
NO5 as elemental composition of the potential degradant.
Fig. 3.11.a. Mass spectrum of N-oxide impurity
During the drug product stability evaluation and specificity study, especially in
peroxide degradation experiment 0.71 RRT impurity was found potential. In order to enhance
degree centigrade for
70%. The isolation of the
prep chromatographic system. Combined isolated portion of
egree centigrade vacuum. The
resultant aqueous layer was lyophilized. The purity of the lyophilized impurity was checked
in HPLC and found highly pure. Further it is utilized to identify the impurity by advanced
tion mass spectrometer of potential degradant found
(Fig.3.11.a). This m/z is 16 amu additional than
fexofenadine. The m/z result reveals the existence of extra ‘O’ oxygen in potential degradant.
as elemental composition of the potential degradant.
(A)
Fig. 3.11.b. Fexofenadine (B)
Tab. 3.11.a.fexofenadine and N
Page 13 of 29
(B)
Fig. 3.11.b. Fexofenadine (B)andN-Oxide imp (A) Structural formula
Tab. 3.11.a.fexofenadine and N-oxide imp.NMR .
Structural formula
Page 14 of 29
3.12 METHOD VALIDATION:
The method validation of optimized test procedure was performed as per ICH
guidelines1. The method was validated by executing the studies for the parameter are,
selectivity, linearity, recovery, precision, and robustness.
System Suitability
This parameter was verified by test procedure characteristic criteria before initiate the
analysis which determination has been done base on specification. The SST was assessed
based on retention time, the resolution between impurity-B and fexofenadine peak, ratio of
fexofenadine peak areas from two standard injections, fexofenadine peak plate count and
asymmetric factor from STD preparation. SST was checked before analysis from two
standard injections which has 2.4 ppm fexofenadine. The acceptance criteria were ratio of
fexofenadine peak aireas should be between 0.9 and 1.1, tailing factor should be LT 2.0 &
theoretical plate should be MT 5000 for fexofenadine from STD. From system suitability
solution, the resolution between impurity B and fexofenadine peak should be not less than
3.0. All parameter are found to be meeting the SST specification. (Table-3.12.a).
Tab. 3.12.a. System Suitability Test Results
Specificity:
The selectivity of the proposed test procedure was verified for its performance and
monitored the resolution of fexofenadine all impurities from each other. In this experiment
also conducted to check the interference from placebo and degradation product at retention
time of fexofenadine and fexofenadine all impurities. Drug product and placebo was stressed
as per bellow mentioned conditions.
Page 15 of 29
The finished product was stressed for base hydrolysis with 2 Normal NaOH for 24 hrs
at 60 degree centigrade, acid hydrolysis with 1N HCl for 3.4 hrs at 60°C, Hydrolytic
degradation performed at 60°C for 24 hours. Oxidation was done with 3 % H2O2 at 60 degree
centigrade for five hours. The photo stability experiment was performed as per ICH Q1B,
exposed to visible light (1.2Million Lux hours) and ultra violet light (200 watt hours per
meter2)[19]. For humidity study sample were kept at 25 degree centigrade, 90 percent relative
humidity for 10 days. Heat stressed experiment done at 105 degree centigrade for one day.
Drug product and placebo sample were treated as same for above mentioned stress condition.
The fexofenadine peak purity was tested with photo diode array detector for degraded test.
The placebo interference was verified with the placebo preparation which is equivalent
amount present in sample. The result shows that there is purity is ok for its all related
substances.
Specificity study for placebo intereference had performed by injecting placebo
preparation as per methodology. There is no intereference found at RT of FEX and its
impurities. (Fig. 3.12.a). 1.2 milligram per milliliter FEX sample was injected in the UPLC
(PDA) to confirm FEX peak purity (Fig. 3.12.b-h). More degradation was found in peroxide
degradation and minor degradation was found in acid hydrolysis, alkali hydrolysis and heat
degradation. Majorly unknown impurity was found in oxidation condition forced degradation
.Oxidative potential degradant, proton and carbon NMR. FEX was observed to be non
degrdaing in water.The optimized test procedure mass balance was calculated for to ensure its
specificity by summation of degradation sample assay and impurities and results are more
than 98.4% (Table- 3.12.b). The result shows that there is found RT of all related
substances.and nature of the optimized test procedure.
Fig.3.12.a.: Representative Chromatogram of Placebo
Page 16 of 29
Tab.3.12.b. Summary of FDstudy
Fig.3.12.a.: Representative Chromatogram of Placebo
Page 17 of 29
Fig.3.12.b.: Representative Chromatogram of Acid Stressed Sample
Fig.3.12.c.: Representative chromatogram of Alkali Stressed Sample
Fig.3.12.d.: Representative Chromatogram of Oxidation Stressed Sample
Page 18 of 29
Fig.3.12.e.: Representative chromatogram of Thermal Stressed Sample
Fig.3.12.f.: Representative chromatogram of Aqueous stressed sample
Fig.3.12.g.: Representative Chromatogram of Photolytic Stressed Sample
Page 19 of 29
Fig.3.12.h.: Representative chromatogram of Humidity stressed sample
Fig.3.12.i.: Representative chromatogram of test spiked with its impurities.
Precision of the method:
This parameter was studied by repeatability and intermediate precision. The 6 each
sample of fexofenadine ODT spiked with its 4 impurities was made and injected in
repeatability study. Sample were prepared by impurity -B and N-oxide at specification limit,
impurity A specification limit and imlimit purity C (decarboxylated impurity) specification
limit (% of impurities respecting 1.2 mg/milliliter FEX) spiking. %RSD of area for eaach
imapurity was calculeted. The test procedure intersmediate precission study were assessed by
differenet analyset, instrument and day.
Based on outcome of the study it is found that proposed test procedure is very precise.
The % RSD for the area of Impurity-A, B, C and N-oxide was maximum 3.0% and 3.6% in
repeatability and intermediate precision correspondingly. The results for each impurity are
tabulated in Table-3.12.a and 3.12.b.
Page 20 of 29
Table 3.12.a : Precision Result of Fexofenadine Impurities
Table 3.12.b: Intermediate Precision Result of Fexofenadine Impurities
Limits of Detection and Quantification of Method
The LOD and LOQ for fexofenadine HCl related substance were established by S/N
ratio which should be 3:1 and 10:1, respectively. At LOQ level six preparation fexofenadine
HCl impurities were prepared to perform the precision study. The results were interpreted in
the %RSD of the peak response. The LOD, LOQ in Table-3.12.c and Table-3.12.d.
Page 21 of 29
Table 3.12.c : LOD and LOQ Value with S/N Ratio of Fexofenadine Impurity
Table 3.12.d : Fexofenadine impurities LOQ precision
Page 22 of 29
Linearity of the Method
To perform this parameter of the proposed test procedure 6 different parts per million
solutions were prepared. The series of level covers the LOQ to 200% of specification level. A
good correlation found from responses of peak Vs standard concentrations. It was achieved in
the range of calibration, i.e. Limit of detection to 200 percent of target level. The R was more
than 0.998 and percent bias at 100 percent concentration was ± 2 percent for all four
Impurities. The results demonstrate the exceptional correlation of area of peak and concen of
Imp- A, Imp- B, Imp- C, and N- Oxide. The R2, c and y-intercept are summarized in
presented in Fig.3.12.j-m.
Fig.3.12.j.: Linearity graph for N-Oxide impurity
Fig.3.12.k.: Linearity graph for impurity-A
y = 5815x + 361.7
R² = 0.999
0
10000
20000
30000
40000
50000
0 2 4 6 8 10
Are
a
Concentration (µg/mL)
N-Oxide
y = 6080.x + 13.66
R² = 0.999
0
10000
20000
30000
40000
50000
0 2 4 6 8
Are
a
Concentration (µg/mL)
Impurity-A
Page 23 of 29
Fig.3.12.l.: Linearity graph for impurity-B
Fig.3.12.m.: Linearity graph for impurity-C
Accuracy
The recovery parameter of impurity was performed in three sample preparation at
Limit of quantification to 50 percent, 75percent, 100percent, 125percent and 150percent of
targeted concen level. All impurities % accuracy was calculated. The % accuracy of
impurities of FEX tests varies from 92.6 to 108.6% at limit of quantification, 50percent,
75percent, 100percent, 125percent and 150percent levels of test concentration. The Liquid
chromatography chromatogram is shown in figure Fig. 3.12.i. with spiked test at target
concentration of all impurities in FEX. Individual % accuracy results for impurities are
tabulated in Table-3.12.e to Table-3.12.i.
y = 5118.x + 249.3
R² = 0.998
0
2000
4000
6000
8000
10000
12000
14000
16000
0 0.5 1 1.5 2 2.5 3
Are
a
Concentration (µg/mL)
Impurity-B
y = 6830.x - 478.9
R² = 0.999
0
5000
10000
15000
20000
25000
30000
0 1 2 3 4 5
Are
a
Concentration (µg/mL)
Impurity-C
Page 24 of 29
Tab.3.12.e. Recovery Data of N-oxide Impurity
Tab.3.12.f. Recovery Data of Impurity-A Impurity
Page 25 of 29
Tab.3.12.g. Recovery Data of Impurity-B Impurity
Page 26 of 29
Tab.3.12.h. Recovery Data of Impurity-C Impurity
Page 27 of 29
Tab.3.12.i. Recovery Data of Impurity at LOQ Level
Robustness
The optimized test procedure robustness was verified by deliberate change in
predefined method parameter. In this experiment response measured as resolution of
fexofenadine and impurity-B and other SST parameter. MP buffer pH was altered to 6.8
and7.2 unit. The FR was changed to 0.36 and 0.44 milliliter per minute to check SST
performance. The column oven temp was altered to 25°C and 35°C in robustness study.ACN
proportion was changed to 90% -110% to check its effect on resolution and SST criteria.
In every the deliberate changes of chromatographic parameter, the results were met
the specification criteria for system suitability. The impurities and main peak are separated
well and all peaks are retained as per RRT. The FEX and impurity-B resolution was found to
be more than 5.4.The FEX tailing was observed 1.0 and theoretical plate count found more
than 11434 in STD injection chromatogram. The similarity factor of peak response was 1.0.
The system suitability parameters evaluated are shown in Table-3.12.j.
Tab. 3.12.j. UPLC Method
Stability in Solution and in the Mobile Phase
To study this parameter FEX sample with spiked all impurities were kept on bench
top for 2 days in closed container. Similar process was followed for STD preparation. Every
day all impurity level and STD preparation similarity factor was evaluated with new
preparation. The same process was followed to established mobile phase stability for 2 days
where every day sample spiked with all impurities were evaluated against old mobile phase.
The result shows that mobile phase is stable up to 2 days. The FEX
with impurities result is not varying more than 10%. Hence, samples are stable up to 2 days.
The similarity factor of new and old prepared standard is within the specification limit.
Therefore, STD is also stable up to 2 days on room temperat
Page 28 of 29
Tab. 3.12.j. UPLC Method - Robustness Data
Stability in Solution and in the Mobile Phase
To study this parameter FEX sample with spiked all impurities were kept on bench
top for 2 days in closed container. Similar process was followed for STD preparation. Every
day all impurity level and STD preparation similarity factor was evaluated with new
preparation. The same process was followed to established mobile phase stability for 2 days
where every day sample spiked with all impurities were evaluated against old mobile phase.
The result shows that mobile phase is stable up to 2 days. The FEX
with impurities result is not varying more than 10%. Hence, samples are stable up to 2 days.
The similarity factor of new and old prepared standard is within the specification limit.
Therefore, STD is also stable up to 2 days on room temperature.
To study this parameter FEX sample with spiked all impurities were kept on bench
top for 2 days in closed container. Similar process was followed for STD preparation. Every
day all impurity level and STD preparation similarity factor was evaluated with new STD
preparation. The same process was followed to established mobile phase stability for 2 days
where every day sample spiked with all impurities were evaluated against old mobile phase.
The result shows that mobile phase is stable up to 2 days. The FEX sample spiked
with impurities result is not varying more than 10%. Hence, samples are stable up to 2 days.
The similarity factor of new and old prepared standard is within the specification limit.
Page 29 of 29
The results (precision, specificity, linearity, accuracy, robustness and ruggedness) of
the present study indicate that the developed and validated methods are reproducible, rugged
and stability indicating.