23.01.2018 tender for fixing rate contract … text add to the end of technical ... 6. govt. rajaji...

31
23.01.2018 TENDER FOR FIXING RATE CONTRACT FOR SUPPLY AND INSTALLATION OF LINEAR ACCELERATOR TO MEDICAL INSTITUTIONS IN THE STATE OF TAMIL NADU TENDER NO. 246/LINAC/TNMSC/ENGG/2017, DT.12.12.2017 CORRIGENDUM A. The following amendment is issued: SI. No. Location Instead of Read as 1. Page no. 51 Section V: Schedule of requirements Existing text Revised text at Annexure I 2. Page no. 54 Section VI: Technical Specification Existing text Add to the end of technical specification. 4. Technical Specification for Advanced Dual Energy Linear Accelerator at Annexure II 3. Page no. 218 Section VII : Bid form and Price Schedules Price Schedule Existing text Revised text Annexure III The due date is extended as follows:- Sale of Bid document : up to 08.02.2018 Date & time of submission of bids : 09.02.2018 up to 11.00 AM Date and time of opening of bids : 09.02.2018 at 12.00 Noon All other terms and conditions of the tender remain unaltered. The above corrigendum with annexure forms part of the tender document. The bidder shall attach a copy of this corrigendum duly signed by their authorized signatory in their bid. Sd/- General Manager (Equip)

Upload: phungkien

Post on 12-Mar-2018

216 views

Category:

Documents


4 download

TRANSCRIPT

Page 1: 23.01.2018 TENDER FOR FIXING RATE CONTRACT … text Add to the end of technical ... 6. Govt. Rajaji Hospital, Madurai Note: 1) ... Company should give the undertaking for supply of

23.01.2018

TENDER FOR FIXING RATE CONTRACT FOR SUPPLY AND INSTALLATION OF LINEAR ACCELERATOR TO MEDICAL

INSTITUTIONS IN THE STATE OF TAMIL NADU TENDER NO. 246/LINAC/TNMSC/ENGG/2017, DT.12.12.2017

CORRIGENDUM

A. The following amendment is issued:

SI.

No.

Location Instead of Read as

1. Page no. 51 Section V: Schedule

of requirements

Existing text Revised text at Annexure I

2. Page no. 54

Section VI: Technical Specification

Existing text Add to the end of technical

specification. 4. Technical Specification

for Advanced Dual Energy

Linear Accelerator at Annexure II

3. Page no. 218 Section VII : Bid form

and Price Schedules

Price Schedule

Existing text Revised text Annexure III

The due date is extended as follows:-

Sale of Bid document : up to 08.02.2018

Date & time of submission of bids : 09.02.2018 up to 11.00 AM

Date and time of opening of bids : 09.02.2018 at 12.00 Noon

All other terms and conditions of the tender remain unaltered.

The above corrigendum with annexure forms part of the tender

document. The bidder shall attach a copy of this corrigendum duly signed by

their authorized signatory in their bid.

Sd/-

General Manager (Equip)

Page 2: 23.01.2018 TENDER FOR FIXING RATE CONTRACT … text Add to the end of technical ... 6. Govt. Rajaji Hospital, Madurai Note: 1) ... Company should give the undertaking for supply of

Annexure-I

SECTION – V

SCHEDULE OF REQUIREMENTS -REVISED

Sch.

No.

Brief Description Unit Qty.

Bid security

(Rs.)

1. Dual Energy Linear Accelerator with optional items

Nos. 6 Rs. 45,00,000/- 2. Dual Energy Linear Accelerator

3. Single Energy Linear Accelerator

4. Advanced Dual Energy linear Accelerator

Delivery Schedule: - 90 days from the date of purchase order

Place of Delivery:- The above equipment is delivered at

1. Rajiv Gandhi Government General Hospital, chennai

2. Govt. Royapettah Hospital, Chennai.

3. Govt. Coimbatore Medical College Hospital, Chennai

4. Govt. Thanjavur Medical College Hospital, Chennai

5. Govt. Tirunelveli Medical College Hospital, Tirunelveli

6. Govt. Rajaji Hospital, Madurai

Note:

1) The quantity indicated is tentative and the actual quantity may vary at the time of

placement of orders. No claim on such variation will be entertained.

2) The bidders are permitted to quote any one or all the Schedule.

3) TNMSC reserves the rights to procure any number of Linear accelerator to any of the

listed hospitals in any of the four categories based on the price differential and case

loads and the relative advantage of prices with respect to higher specification. The

decision of TNMSC in this aspect shall be final.

4) The rate contract will be for a period of 3 years from the date of notification of

award.

5) The bidders are also permitted to quote the imported component price in any foreign

currency which will be converted into Indian Currency fixed by SBI on the date of

Page 3: 23.01.2018 TENDER FOR FIXING RATE CONTRACT … text Add to the end of technical ... 6. Govt. Rajaji Hospital, Madurai Note: 1) ... Company should give the undertaking for supply of

opening of technical bid for evaluation of prices and the bidders should quote for the

other component in Indian Rupees.

6) For the rate contract, the imported component price will be frozen at the foreign

currency rates for a period of three years from the date of notification of award and

the payment will be made at the time of actual placement of order and at the

prevailing exchange rates. The customs duty, GST etc., will be at the actual rate at

the time of placement of order. For the local currency quoted items, an annual

increase of 4% per year will be given.

7) The bidder should indicate the following break up in their price bid.

a) CIF value of imported component.

b) Applicable customs duty on the assessable value.

c) Customs clearance charges.

d) Local transport to site.

e) Local accessories.

f) Installation and commissioning including turnkey.

Page 4: 23.01.2018 TENDER FOR FIXING RATE CONTRACT … text Add to the end of technical ... 6. Govt. Rajaji Hospital, Madurai Note: 1) ... Company should give the undertaking for supply of

Annexure-II

Technical specifications of ADVANCED DUAL ENERGY LINEAR

ACCELERATOR (LA)

Advanced Dual Energy Medical Linear Accelerator shall have futuristic advanced platform and

shall have minimum photons of 6 and 15 MV & any five electron beams from 6 to 20 MeV range. The

Linac shall deliver 1.IMRT, 2.VMAT/RAPID ARC, 3.ARC, 4.3D CRT, 5.SRS/SBRT and gated Delivery as

package and shall be upgradable in future. Such options shall be offered in the tender and shall have

validity of at least 3 years for such upgrade / any options, if the hardware upgrade is required, that

shall be costed in and quoted to avoid any hidden charges in executing such options / upgrades. The

main equipment and major features shall have AERB type approval / NOC and shall have FDA or CE

approval.

1. The unit shall meet all the radiation safety standards & Quality Assurance of it mechanical,

electrical and electronic provisions set by regulatory bodies(AERB).Machine shall be typed

approved by A.E.R.B and System shall have all safety interlocks as per AERB guideline.

2. FDA or CE certificate must be provided.

3. It should be the Current model of the machine and software of Treatment Planning system

should be of latest version and should be fully digital .It shall be brand new machine.

Supporting documents to be furnished.

4. Company should provide the certificate of life of machine.

5. Company should give the undertaking for supply of spares for 10 years.

6. Software updates with all added features should be provided free of cost of warranty

period.

7. Company should be responsible for commissioning and handling over the machine to

institute in fully running (proper working) condition.

8. The scope of works for the rate quoted should include installation and commissioning also.

9. Prices quoted for optional items and future upgradation shall be valid for three years after

installation.

10. Company shall make free replacement of damaged parts within 72 hours during warranty at no

additional cost.

11. In case of any replacement of parts during warranty/ guarantee period all formalities such as

import license AMI and DE certificate etc connected with such as import of replacement shall be

compiled with by the bidder. In addition, the payment of custom duty clearing and transportation

charges shall be borne by the bidder.

Page 5: 23.01.2018 TENDER FOR FIXING RATE CONTRACT … text Add to the end of technical ... 6. Govt. Rajaji Hospital, Madurai Note: 1) ... Company should give the undertaking for supply of

12.Bidder must provide list of spare parts.

13. Proof of all & every statement, certificates should be enclosed from the principals if quoted by

their authorized bidder.

14. Networking and configuration with existing CT/MRI & procuring equipment should be done by

the bidders at their own cost. No extra payment will be made to them for this work.

15. Company shall provid ecertificate of trouble free operation of machine for five years for current

year from existing five users.

16.The warranty for the LINAC and its associated systems shall start from the date of AERB’s

commissioning approval and the date of handing over of the machine after Complete Acceptance

procedures by the enduser shall not be counted for the warranty.

17.Please note that the price SHALL INCLUDE ALL EXPENSES including the Customs clearance,

insurance, freight, customs duty, clearance charges and also all expenses towards the maintenance

and repairs of the entire LINAC assembly including spare-parts, electrical and electronic items, computer systems, Chiller systems (including refilling of distilled / industry standard water),

networking, accessories, etc. All necessary licences should be enabled and made to be avilable to the end user atleast till completion of CAMC .

18.List of consumable items like printer cartridges, etc which are not covered by the warranty clauses

must be clearly identified and declared by the vendors in the tender. The cartridges and other

consumable items of the printer must be available locally for a minimum period of 3 years.

19.Posting of a resident engineer on site is highly preferred . A list of certified service engineers

available should be provided.

20.The installation of the LINAC shall be done by the experienced engineer(s), who has installed at

least 5 installations of similar model.

21.The Chiller system shall be provided along with the machine and the warranty and maintenance

costs should be included in this LINAC offer.Chiller room responsibility (installation, maintenance, off

time etc.) should be of company.Chiller room architectural drawing and other details should be provided

by the company.

22.A closed-circuit color TV system with TV monitors and two cameras in the LINAC treatment room

shall be supplied.

23.A patient calling system with 6 channels shall be supplied.

24.Last five purchase prices of the same equipment should be provided along with supporting

documents and complete address and telephone numbers of purchaser is to be provided.

Page 6: 23.01.2018 TENDER FOR FIXING RATE CONTRACT … text Add to the end of technical ... 6. Govt. Rajaji Hospital, Madurai Note: 1) ... Company should give the undertaking for supply of

25.The company should be responsible for installation, commissioning, maintenance and down time

of the machine, software, accessories and networking which will help directly and indirectly

in proper functioning of the machine.

26.In case the Indian representative (agency/company) of company changes it shall be

responsibility of company to run machine till it’s full life. Original company should submit declaration

for this.

2. BASIC EQUIPMENT

a. The accelerator shall be capable of producing two clinically useful photon beams with energies of 6

MV and 15 MV. The minimum characteristics of each energy for a 10x10 cm field at 100 cm TSD should

be as follows:-

For all the energies quoted, specify the above characteristics. A difference of + 2% in the depth dose

data from the IEC published values.

Energy (MV)

Nominal D max (cm)

%

%Depth Dose at 10 cm

depth (10x10cm field)

6 1.5+0.2 67.1+1.5

15 3.0+0.2 76.5+2.0

b. Dose Rate Beam Stability

1. The X-ray dose rate shall be variable in steps and the X-ray dose rate shall be variable 100-600

MU/minute or more for both 6 and 15 MV X-ray energies. The treatment dose rate should be at least

600 MU/minute.

2.Linac Dosimetry Control System The LINAC shall have built in dosimetry chambers with two separate

sealed or unsealed chambers. Parameters of dosimetry system shall be as follows:- Precision + 1% or 1

MU Linearity + 1% or 1 MU Reproducibility + 2% or 1 MU Dose Rate Dependence.

c. Photon Beam Energy Stability:

1.The quality index of a photon beam should not vary with time by more than + 1% or comply AERB

requirements.

Page 7: 23.01.2018 TENDER FOR FIXING RATE CONTRACT … text Add to the end of technical ... 6. Govt. Rajaji Hospital, Madurai Note: 1) ... Company should give the undertaking for supply of

2.The bend magnet system shall be provided with energy defining apertures such that the nominal

energy of the electrons beam existing from the bend magnet shall be within + 3% of the nominal energy

selected at control console for both photons and electrons.

3. RF Source: Magnetron / Klystron, RF Driver, Thyretron (RF) and RF coupler combination with 10 years

full replacement warranty on full RF chain

4. Waveguide Type : Standing / Travelling wave

The Wave guide shall have at least 15 years full replacement warranty.

5. Electron Gun : Sealed / Unsealed and shall have at least 10 years full replacement warranty.

f. Treatment Modes Normal – TSD / TAD Rotation – CW / CCW ARC – CW / CCW Dose rate – MU /

degree

3. Field size specifications

The field size is defined as the distance along the radial and transverse axes between the points of 50%

density on an x-ray film taken at 100 cm TSD with minimum buildup. The display, light field size and

mechanical display should be accurate to within + 1 mm for field sizes _10 x 10 cm2 and _2 mm for field

sizes >10x10 cm2 or comply AERB requirements.

The accelerator shall provide a continuously variable rectangular, unclipped field size from 1 x 1 cm2 to

35 x 35 cm2 at 100 cm SSD. The Maximum clipped field size should be equal or exceed 40x40 cm2 at 100

cm SSD. Clipped corners are unacceptable for field smaller than 35x35 cm2.

4. Radiation Field Penumbra

The width between the 20% and the 80% isodose lines measured for 10x10 cm at depth of 10 cm at 100

cm SSD should not be more than 7 mm.

5. Congruence between optical & Radiation fields:

The congruence between optical and radiation fields for all photon energies for 5x5 cm2, 10x10 cm2,

30x30 cm2 or for a field of maximum dimension for 0 deg, 90 deg, 180 deg and 270 deg deg gantry

angles with SSD=100 cm at the depth of reference plane should be as per FDA / AERB recommendation.

6. Beam Profile

Field Flatness Specifications Variation of x-ray intensity relative to the central axis shall be _ + 3% at 100

cm SSD and 10 cm depth over the central 80% of the field for the longitudinal, transverse and diagonal

axes of all field sizes from 10x10 cm to 40x40 cm. Stability of the flatness with gantry rotation at 0 deg,

90 deg, 180 deg and 270 deg at 10 cm depth on x, y and diagonal axis for all field sizes from 10x10 cm to

maximum field size should not be more than + 3% (As per IEC guideline). The

Page 8: 23.01.2018 TENDER FOR FIXING RATE CONTRACT … text Add to the end of technical ... 6. Govt. Rajaji Hospital, Madurai Note: 1) ... Company should give the undertaking for supply of

flatness criteria applied to beam profile at D max should show peripheral horns not exceeding 105% of

the central axis dose.

7. Field Symmetry Specifications:

The maximum percent difference of average doses shall not exceed + 2% for the longitudinal and

transverse halves of the field at 100 cm TSD and 10 cm depth at gantry angles of 0, 90, 180 and 270

degrees. Field sizes shall be specified as 10x10 and 40x40 cm2. Average dose is defined as the

arithmetic average of minimum and maximum doses within the central 80% of the field for both axes.

8. Beam matching : The successful bidder shall have to do the beam matching of the photon beams of

the system with the photon beams of same energy of existing LA machines of the Institute.

9. X-ray Contamination

The X-ray / electron / Neutrons leakage and contaminations should comply the AERB guidelines. All

safety systems including head leakage should be as per IEC / AERB guidelines.

The x-ray contamination of the electron beam shall be less than 5% of the maximum dose for all

energies specified previously.

a. X-ray absorbed dose due to leakage radiation (excluding neutrons) outside useful beam but inside

a plane circular area of radius 2 m centered around and perpendicular to central axis at normal

treatment distance as per AERB.

b. The electron contamination should not be more than 1%.

c. Radiation leakage limits shall be within appropriate regulatory agency guidelines as follows:-

1. Photon leakage. The photon leakage rate at any point one meter from the target outside the cone

defined by the primary x-ray collimator shall be less than 0.1% of the absorbed dose at the isocenter.

2. Collimator transmission. The movable collimators shall not permit transmission of radiation exceeding

0.5% of the central axis dose at Dmax measured in air for both photon energies.

d. Neutron leakage. The neutron leakage rate should not exceed 0.15% expressed in neutron dose

equivalent (REM) when added to the photon leakage for a 10x10 cm field at the isocenter at any point

one meter from the target when the jaws are closed or to comply AERB requirement.

10. Electron Energy : minimum 4 Beam energies between 4-18 MeV (more energies if available may be

offered without any additional cost).

Page 9: 23.01.2018 TENDER FOR FIXING RATE CONTRACT … text Add to the end of technical ... 6. Govt. Rajaji Hospital, Madurai Note: 1) ... Company should give the undertaking for supply of

a) Dose-Rate for electron energy : Please specify the dose rates and higher dose rates for special

treatments.

b) Field Size

The electron beam size is defined by the inside dimensions of the electron beam applicators projected

geometrically to a plane surface at 100 cm SSD. At least 4 applicators preferably with

Minimum – 5 x 5 cm2, Maximum – 20x20 cm2 or more; there shall be an arc applicator provided for

electron arc treatment.

A method to obtain irregular field shapes shall be provided.

It shall be possible to visualize both the field defining light and the optical distance indicator with an

electron applicator in place.

c) Beam Flatness (Electrons)

Variation of electron intensity relative to the central axis shall not exceed 5% over within the central

80% of radial and transverse axes for photons field sizes 10x10 cm to 20x20 cm at 10 cm depth and

satisfy local and Indian regulatory AERB requirements.

d) Beam Symmetry:- The maximum percent variation in the average electron intensity to the

longitudinal and transverse halves of the electron field at D max for a 10x10 cm and 25 x 25 cm field at

100 cm SSD shall not exceed + 2% at gantry angles of 0, 90, 180 and 270 degrees.

The average electron intensity is the average of the maximum and minimum points within the central

80% of the field for each of the axes.

11. Other Specifications

a) The target to axis distance should be 100 + 0.2 cm or to comply AERB requirement

b) The isocenter shall lie within a sphere of radius 1 mm or to comply AERB requirement

c) The accelerator gantry shall be capable of rotation equal to or greater than 360 degrees with a

variation of the mechanical and radiation isocenter during rotation of less than + 1.0 mm throughout

the entire rotation with variable gantry speed.

d) Digital scales indicating gantry angle position shall be provided both in then treatment room at the

control console. Accuracy of the scales shall be + 0.5 degree.

e) The distance from the end of the lower collimator to the isocenter shall be > 41 cm.

Page 10: 23.01.2018 TENDER FOR FIXING RATE CONTRACT … text Add to the end of technical ... 6. Govt. Rajaji Hospital, Madurai Note: 1) ... Company should give the undertaking for supply of

f) The bottom of the blocking tray should be greater than 30 cm from the isocenter.

g) The height of the isocenter above the finished floor shall be less than 130 cm

h) Digital scales indicating collimator angle position shall be provided both in the treatment room and at

the control console. Accuracy of the scales shall be + 0.5 degree.

i) The Chiller system shall be the integral part of the equipment supplied. Local chillers shall not be

accepted.

j) Imported voltage stabilizer shall be provided for power spike protection.

k) In addition to meeting above specifications for radiation leakage, the LINAC should also meet all the

mandatory safety and radiation leakage regulations or as specified by AERB.

l) Focal Spot size should be 1-2 mm (smallest is preferable)

m) 80 KVA or more UPS along with battery should be provided to give back up at least 30min for the

machine, machine room & control console room in case of power failure.

12. Photon Arc Therapy Bi-directional arc therapy should be included with Automatic calculation of Dose

per Degree based on the Dose Rate selected and the Arc angle set.

13. Beam characteristics: for electrons and Photons shall satisfy local and Indian regulatory authority

which is mandatory.

14. Gantry

a) Rotation + 1800 (3600 total) b) Read out – Digital and Mechanical c) Accuracy dig-readout 0.50

d) Control – Hand pendent and control –console e) Target – Axis Distance. -100 + 2 mm or better f)

ODI Range – 75 cm to 150 cm g) ODI Accuracy + 0.1 cm h) Gantry Rotation Isocentre less than or

equal to 2 mm dia. Sphere please specify.

15. Collimator: i) Rotation - Preferably + 1800 rotation j) Control – Hand pendent and control –

console k) Readout accuracy - + 0.50 l) Collimator Rotation Isocentre 2 mm dia. Sphere

16. Physical / Motorized / Dynamic / Virtual Wedge – Please specify.

17. Asymmetric jaws

a) X & Y both Asymmetrical b) Travel ranges – X & Y : Please specify

18. Multi-leaf collimator(MLC)

a) No of leaves 120 or more.

b) Independent drives for each leaves .

Page 11: 23.01.2018 TENDER FOR FIXING RATE CONTRACT … text Add to the end of technical ... 6. Govt. Rajaji Hospital, Madurai Note: 1) ... Company should give the undertaking for supply of

c) Leaf size/width : atleast 5mm at the isocentre with suitable thickness at the periphery.

d) Work station : Latest version.

e) Integration (full Networking), conventional Simulator, CT scanner, CT Simulator, MRI & RFA should be

done via Planning System) .

f) 3D CRT, IMRT, VMAT/RAPID ARC delivery shall be offered. VMAT / RAPID ARC shall have dynamic

control of MLC, dose rate, diaphragm, gantry, collimator rotation and shall be capable of full field

VMAT / RAPID ARC, SRT & SBRT capability.

g) Leaf retracting position : please specify

h) High over center travel of MLC leaves (>10cm) for all treatments.

i) Leaf height : please specify

j) Leaf material : please specify

k) Coincidence of light & x-ray field + 2mm .

l) Penumbra shall be < 7 mm for( 10x10 field ) Transmission within 0.5% .

n) X-ray leakage within 0.2% .

o) Minimum leaf speed shall be 0cm / second .

p) Positional accuracy of the leaves during treatment 0.5 mm .

q) Inter-digitization of leaves shall be provided.

s) Two nos. of treatment in-room monitors 19” TFT to be provided.

Stereotactic solution:

Detachable micro-multileaf collimator should be fully networked, integrated and compatible

with offered linear accelerator, latest model and technology to perform SRS/ SRT . All the necessary

software and hardware should be quoted and all the licenses required for routine clinical uses should be

permanent ones and should be offered.

1.14.1. Field size should be at least 10 cm X 10 cm.

1.14.2. Number of leaf pairs should be more than 25

1.14.3. Leaf size/width should be atleast ≤ 5 mm at center with suitable thickness at the periphery.

Page 12: 23.01.2018 TENDER FOR FIXING RATE CONTRACT … text Add to the end of technical ... 6. Govt. Rajaji Hospital, Madurai Note: 1) ... Company should give the undertaking for supply of

1.14.4. Through leaf the radiation leakage should be less than 1% of the photon intensity for all photon

energies. The interleaf radiation leakage should be less than or equal to 3% (i.e. ≤ 3%) of the photon

intensity for all photon energies.

1.14.5. Clearance to isocenter ≥ 30 cm.

1.14.6. Leaf speed ≥1.5 cm./sec.

1.14.7. Leaf over travel = 10 cm.

1.14.8. Optical and radiation field congruence should be within ± 2 mm at 100 cm SSD.

1.14.9. The system should have latest version of CPUs and mMLC controller computer and workstation

computer with adaptability for future upgradation.

1.14.10. The mMLC should be integrated to the computer control system of the LINAC.

1.14.11. The mMLC should be fully integrated with Linear accelerator and should be part of the

networking systems (Stand alone systems will not be considered)

1.14.12. Provide details of complete QA kit for SRS & SRT.

1.14.13. The system should have automatic safety interlocks. System should be able to switch the

LINAC beam off incase of any malfunctioning. Leaf positioning control should be monitored by the

secondary feedback system.

1.14.14. The mMLC should allow IMRT techniques. The device should be supplied on separate storage

trolley.

1.14.15. The Linac system should function and continue to treat patients with conventional treatments

even if MLC leaves are in repair.

1.15. Stereotactic radiosurgery accessories

1.15.1. Head Ring assembly for SRS:

Should be supplied with artifact free, autoclaveable pins of different sizes, a torque wrench to adjust

the safe pin pressure, adjustable fixation post in length to treat deep seated lesions down to C1 and

intubation front piece for the emergency intubation.

1.15.2. Mask System for Cranial SRT:

The device should ensure repeat fixation accuracy in lateral & axial directions. The system should

ensure access to oral treatments & intubation without removal of the device. The system should ensure

consistent reposition accuracy.

1.15.3. Head & Neck System for immobilization:

Page 13: 23.01.2018 TENDER FOR FIXING RATE CONTRACT … text Add to the end of technical ... 6. Govt. Rajaji Hospital, Madurai Note: 1) ... Company should give the undertaking for supply of

The device should cover treatment area up to T1 vertebra. The system should ensure consistent

reposition accuracy. The device should offer tilt adjustment in X & Y axis.

1.16 :Planning System for SRS/ SRT .

1.16.2. The planning system should be compatible, networked and demonstrated to the customer

with the Micro multi-leaf collimator and the linear accelerator and perform conformal planning for

Stereotactic radio-surgery/ radiotherapy .

1.16.2.1. Hardware:

1.16.2.1.1. A Latest High Power Intel dual Processor workstation

1.16.2.1.2. 2048 MB Main Memory

1.16.2.1.3. 2 x 70 GB Hard disk (RAID Level 1) min.

1.16.2.1.4. DVD R/RW burner

1.16.2.1.5. Keyboard and Mouse

1.16.2.1.6. MS-Windows XP

1.16.2.1.7. 21" Flatscreen Monitor (1280 x 1024)

1.16.2.1.8. 48 cm visible screen diagonal

1.16.2.1.9. ISO A3 color laserjet printer for documentation of treatment plan

1.16.2.1.10. Prints patient set-up templates for Target Positioning

1.16.2.2. Software

1.16.2.2.1. The Planning System should provide data acquisition, for treatment planning, can be done

using CT/MRI/SPECT/PET and Simulator CT slices and angiographic films via film scanner.

1.16.2.2.3. The software must be capable to import patients’ data from CT scanner, MRI and simulator

(existing in the department) by RT DICOM. Complete DICOM-RT license must be given (import/export

for all DICOM-RT objects and IMRT fluence) to facilitate seamless integration with all other planning

systems, imaging systems and QA systems in the department.

1.16.2.2.4. The software should include all the licenses required (DICOM RT, DICOM-3 etc.) and these

licenses should be permanent ones. The successful bidder should demonstrate its actual applications

with patients during installation. Mere wordings like capable of / compatible etc are not accepted.

1.16.2.3. Treatment Planning

Page 14: 23.01.2018 TENDER FOR FIXING RATE CONTRACT … text Add to the end of technical ... 6. Govt. Rajaji Hospital, Madurai Note: 1) ... Company should give the undertaking for supply of

1.16.2.3.1. The s/w should offer the automatic treatment planning tools

1.16.2.3.2. Automatic Marker detection

1.16.2.3.3. Automatic image fusion

1.16.2.3.4. Automatic isocenter placement

1.16.2.3.5. Beam library

1.16.2.3.6. Automatic field shaping

1.16.2.3.7. Automatic collimator optimization

1.16.2.3.8. Auto-contouring

1.16.2.3.9. Editing of contours without completely reentering of data Possibility to expand/shrink

Objects (e.g. definition of boost volumes) independently in all 3 dimensions.

1.16.2.3.10. Auto Segmentation using atlas or template information.

1.16.2.3.11. Possibility to elastically fuse structures from a brain atlas onto the specific patient data for

auto segmentation.

1.16.2.4. Intensity Modulation

1.16.2.4.1. Software should feature inverse planning capabilities. The system should offer an integrated

leaf-sequencing algorithm, tongue and groove optimization based on leaf sequencing & leakage for the

dose calculation.

1.16.2.4.2. The system should individually adjustable Dose-Volume constraints for PTV, Boost

Volume and OARs. The system should offer importance weighting for OARs and simultaneous multiple

plan optimization.

1.16.3. The offered mMLC based SRS/SRT system should be able to support frameless stereotactic

radiosurgery procedure. Similarly it would be highly preferable to have the latest clinical applications

like dynamic IMRT, and dynamic conformal arc techniques in the offered mMLC system.

1.16.4. The stereotactic ion chamber detectors (Pinpoint & diamond detector) along with electrometer

shall be provided. Provide its characteristics and specifications.

The chamber must be supplied with calibration certificate.

19. Treatment Couch:

Page 15: 23.01.2018 TENDER FOR FIXING RATE CONTRACT … text Add to the end of technical ... 6. Govt. Rajaji Hospital, Madurai Note: 1) ... Company should give the undertaking for supply of

a) Versatile extended range shall be supplied. b) Movement range : Please specify. c) Electrical &

Mechanical Control in case of power failure. d) Control-Local and / or Remote e) Shall have indexed

carbon fibre table top f} shall be provided with laser gaurd or any other anticollision mechanism.

20. Accessories:

a) Front and back pointer – mechanical and / or laser. b) CCTV / Camera Two nos. one wide angle &

one remote control with remote zoom & focus facility . c) Laser alignment system – Green / Red

colour (3 cross + 1 line) . d) A patient communication system with 6 channels shall be supplied. e)

Interface mount to be provided for shadow trays.

21. Portal Imaging(EPID) &Accessories

a) Portal imaging should be fully integrated with Accelerator. b) Should be able to take images at

any Gantry angles with variable X-Y movements , Robotics Arm with remote control. c) Should have

Digital technology with High Resolution 1024 x 768 pixels or more Imaging (Amorphous Silicon Flat Panel

Based /scintillator Based Technology) d)The retractable arm of the detector should be either manual or

robotics type and remote controlled

22. (a) Preferably KV based 3D IGRT and KV/MV based Portal Imaging shall be provided and such

system shall have FDA/CE clearance. The System shall have x-ray source which may be manual or

automatic movement with an automatic flat panel system of 1024 x 1024 pixel or higher and shall have

software for 2D radiography, 2D fluoroscopy and 3D cone beam (volume) CT software’s, with manual /

automated DICOM. kV / MV IGRT QA tools.

23. Oncology Information System complete with Networking

Features shall include –

a) Record & Verify System

b) Transfer of all parameters from available CT Simulator / scanner, MRI, PET-CT, if DICOM is available to

Treatment Planning System and contouring station for Automatic contouring etc.

c) Transfer of all parameters Treatment Planning System to the linear accelerator for Automatic

treatment setup & delivery.

d) Transfer of Fluoroscopy images from CT Simulator to Portal Imaging System for Comparison.

e) Transfer & Execution of MLC Position Parameters for normal treatment & IMRT treatment including

step & shoot and Sliding Window (Dynamic), VMAT/RAPID ARC or SRS/SRT techniques from Treatment

Planning System.

24. Dosimetry and QA

Radiation Therapy Beam Analyzer

Page 16: 23.01.2018 TENDER FOR FIXING RATE CONTRACT … text Add to the end of technical ... 6. Govt. Rajaji Hospital, Madurai Note: 1) ... Company should give the undertaking for supply of

Require a full-fledged three dimensional Water Phantom & Dosimetry System and therapy beam

analyzer for performing Off-axis profiles, PDD, point dose measurement, beam symmetry tuning, Dose

rate constancy check, vector scan and TG51 lead foil measurement for flow and high energy Photon and

electrons. All the measurements should be computer controlled and user friendly. All components

comply with national and international regulations and safety rules. All components of the system and

all available options are controlled by the same software that runs under Microsoft Windows of latest

version of window 2000 and Windows XP. The system should suitable to measure pulsed radiation with

fluctuation dose rate.

Ion Chamber:

Necessary thimble ionization chamber should be there for measurement of field and reference

signal. A thimble chamber of 0.6 cc volume and a plane parallel chamber should be there for photon and

electron absolute dose measurement. The necessary holding devices and extension cables for the above

chambers must be included. The chamber specifications should be quoted. The position accuracy

should be better than + 0.1mm. The chambers should be properly calibrated and given necessary

calibration certificate. The positioning tool should be there to allow easy and exact

positioning of the chamber’s geometrical centre in the central beam and at the water surface. Apart

from this the exact position of the chamber in the radiation beam should be possible via software.

The detector unit should be driven by stepper motor and step length should be adjustable in steps of 0.1

mm. The scanning speed should be adjustable between 5mm/s and 50mm/s in 5mm/s small steps.

Further the delay times for each step should also be adjustable by the user. The acceleration of the step

movement should also be changed as and when required.

The system should allow simultaneous movement in available direction for any vector scan.

The zero point, reference point and limit of the different detector units should be stored separately and

permanently in the control unit.

The control pendant should display the actual position of the chamber position at any given measuring

time.

Water Phantom / Radiation Field Analyzer:

a)The scanning volume should be large enough to scan and should not be less than 48x40x48 cm.

b)To avoid bending of the tank’s walls by water pressure and water absorption of the acrylic material

the wall thickness should be not less than 2.0 cm.

c)The motor of the moving mechanism should not touch not nor dip to the water to avoid mechanical

stress to the acrylic tank.

d) The reproducibility of a position should be + 0.1 mm throughout the whole phantom.

Page 17: 23.01.2018 TENDER FOR FIXING RATE CONTRACT … text Add to the end of technical ... 6. Govt. Rajaji Hospital, Madurai Note: 1) ... Company should give the undertaking for supply of

e) The digitally driven stepper motors should provide hysteresis free movements (stick and slip free).

f) The lift table should be electrically as well as manually operable.

g)The velocity of the vertical motion should be quoted and preferably should have two vertical

velocities.

h)The Water Tank must be rotatable into positions 0 degree, + 45 degree and + 90 degree.

i)A highly accurate Positioning device directly supplied by the principals must be included.

Water reservoir:

The water reservoir should be large enough to store the water and can be pump and drain to the

water phantom as quick as possible. The water Reservoir must be able to hold the entire weight of the

water without any change .The weight of the whole assembly can be puss or pull though the wheel with

polyethylene or equivalent.

The lifting carriage should be electromechanical / elevating screw mechanism that keeps the height

absolutely accurate.

The Lifting carriage and Water Reservoir must be imported and directly from the suppliers and must be

complete with all facilities including TPR and TMR measurements.

Completely Integrated Lifting Carriage and Water Reservoir.

The Water Reservoir must be compatible for TPR measurements and hence for TPR measurements the

pump of the reservoir should drive automatically and electromagnetic valves makes sure that no water

can flow the phantom tank to the reservoir during automatic TPR measurement.

The water reservoir should have a safety circuit that avoids the dry pump running.

Control Unit / Electrometer:

A separate control unit for controlling the movement of the detector in any three directions should be

possible.

A separate electrometer to collect the ions / dose from the chamber / detector should be there. The

voltage to the chamber should be adjusted in the electrometer in steps of 50 V. The polarity of the

chamber should be toggled between +/-. The electrometer should also be able to measure absolute

doses for low and high energy photon and electron.

The gain of the electrometer should be automatic depending upon the signal collected by the field and

reference detector. Further the user should also be given an option to change the gain to field and

reference separately.

Necessary software to use the electrometer for absolute measurements should be provided.

Page 18: 23.01.2018 TENDER FOR FIXING RATE CONTRACT … text Add to the end of technical ... 6. Govt. Rajaji Hospital, Madurai Note: 1) ... Company should give the undertaking for supply of

The time constant should allow 10ms measurement times.

The external dosimeter should also be connecting to the water phantom.

The control unit should permanently store zero point, reference point and limit points for water

phantom, air scanner and mechanical film densitometer separately.

These different sets of limit, zero and reference points can be retrieved independently.

The co-ordinates of the probe should display for all directions, simultaneously on a control pendant.

The control pendant can be attached either to the water tank or to the control unit.

The communication between the control unit and the computer should performed by a standard RS232

interface.

The high voltage for the probe should be switchable independently for each decreased in different

voltage and sign of the measuring signal can be reversed.

A solid, water equivalent phantom made up of slabs of different thicknesses shall be provided by the

vendor for external beam teletherapy dosimetry. It shall be possible to use this phantom for both

photon and electron beam dosimetry. The phantom shall be free of contaminants and air bubbles. The

slab shall be of 30x30 cm or more size totaling a thickness of 30 cm.

Control Computer: The latest version of windows computer should have all the latest feature with color

monitor and with printer / plotter (color) and branded UPS (45 min. back-up).

The software

Measurements can be done against time, against a monitor signal or against reference chamber.

Within the moving range arbitrary points can be measured. An arbitrary vector scan measurement

should be possible.

Point dose measurement, Beam symmetry tuning and TG51 foil measurement should also be possible.

2D planes can be measured at any solid angle.

Isodose can be displayed and plotted that can constructed out of profiles and depth dose curves or

measured matrices. The Isodose level should be freely closable Warning before unsaved date in the

RAM should be overwritten.

The Isodose levels can be chosen after the measurement and without the necessity to have the water

phantom connected.

Multiple closed Isodose lines and hot spots should be detected automatically.

Page 19: 23.01.2018 TENDER FOR FIXING RATE CONTRACT … text Add to the end of technical ... 6. Govt. Rajaji Hospital, Madurai Note: 1) ... Company should give the undertaking for supply of

Single measuring points, complete curves and parts of curves should be remeasured from a user

definable point.

During the measurement the measuring curve should be display graphically and online on the screen.

A special measuring program allows a dose rate constancy check including a statistical evaluation.

Any kind of open, regular shaped, blocked or wedged field can be measured.

Fields from asymmetric collimators can easily be measured.

A special measuring routine for service purposes allows to easily checking the beam with respect to

symmetry, flatness, homogeneity and energy.

Implemented routines allow the measurement, formatting and transferring of basic date to all-

important therapy planning systems.

Ion chamber based survey meters to be provided.

Secondary standard dosimeter with appropriate thimble chamber and parallel plate chambers with

latest calibrations to be provided including pin point chamber for small field dosimetry with phantoms,

barometer and thermometer.

Solid equivalent slab water phantom with adapters for the above mentioned chambers should be

provided.

Film Dosimetric software should be provided for treatment verification.

Administrative Data:

Comprehensive documentation of the measured data by automatic saving of the used measuring

environment should simplify the interpretation of data even a long time.

The used measuring routine data can be reused either unchanged or with some of the parameter

changed.

Data can be printed and plotted in numerical and graphical form on all printers and plotters that are

supported by windows.

The administrative data can be changed after saving the measuring data. All measuring data should

furnished automatically with their administrative information and comprehensive filter function allows

the easily selection of specific data.

The necessary software to network the 3D TBA system with the existing 3D TPS in the department of

Radiotherapy must be offered.

Data Analysis:

Page 20: 23.01.2018 TENDER FOR FIXING RATE CONTRACT … text Add to the end of technical ... 6. Govt. Rajaji Hospital, Madurai Note: 1) ... Company should give the undertaking for supply of

Various normalization should possible viz. normalization to maximum for depth dose curves,

normalization to maximum or center for profiles and normalization to maximum, enter, position and

value for isodose lines.

Homogeneity and symmetry should calculated automatically and various national and international

protocols can be selected.

Depth dose curves can be analyses according to the protocols DIN 6800/2 IAEA TR277, ICRU 35, CRMRI

no.2, AAPM TH21/TG 25 and NACP.

Data transfer and data presentation:

Modules should allow automatic formatting and transferring of measured data to treatment planning

system available in the department.

The measured data can be stored in two different ASCII-formats (with selectable separation characters).

ASCII-data can be sent from external computers and be imported in to the water phantom software.

Image date for film dosimetry can be imported in to water phantom software. Data can be display

graphically on the screen.

Crosshairs should allow the easy manual evaluation of a curve.

1. Plotting / printing of the measured data and correction functions can be printed (alphanumerically)

and plotted (graphically).

2. Accessories for 3-D planning and treatment delivery

2.1. General: The following items are required in developing and implementing of a comprehensive,

ultra-modern 3-D conformal radiotherapy, and intensity modulated radiation therapy program

in the department of Radiotherapy. For all the items the vendor should provide the product

information brochures.

2.2. Hot-wire Cutter / Manual Styrofoam Cutter: One hot-wire cutter for making custom shaped

blocks for variable SSDs should be quoted. Low melting alloy should be supplied. High quality

Flatfoams 40 PSI DOW® Oncology Styrofoams for Photons (6 & 15 MV) and Electron foams with

different thickness for electron energies 6, 9, 12 & 15 MeV, with 100 numbers for each energy

for both photon and electron beams, should be included in the offer.

2.3. Block Casting System: A block casting system should be quoted for filling the Styrofoam milled

using the above system with low melting point alloy. The system should have a solid-state Alloy

Melter with digital readout. It should have precise temperature control and have advanced

alloy dispenser.

Page 21: 23.01.2018 TENDER FOR FIXING RATE CONTRACT … text Add to the end of technical ... 6. Govt. Rajaji Hospital, Madurai Note: 1) ... Company should give the undertaking for supply of

2.4. Electron Foam Cutter (220 volts with Nickel chrome wire assembly) Two electron foam cutters

with large cutting surface for cutting high density foam for making electron cut outs for

radiotherapy patients. Appropriate high density flat foams of 30x30x1.5 size for making

electron cutouts shall be provided (two cartons).

2.5. 300 Kg of Cadmium free Low melting point alloy for making customized shields should be

included in the offer.

25. Immobilization Device:

1) 2 set Universal treatment base plate Made of Carbon Fiber Immobilization devices having a

total solution to treat Pediatric to Adult, Head and Head & Neck Breast, Thorax, Abdomen, Pelvic with

facility to make custom made Supine and prone head rest for Individual Patients to maintain an accuracy

of less than 2 mm. along with appropriate a)thermoplastic Sheets 200 nos. 40 for head, 40 H&N, 40 for

breast, 40 for thorax, 40 for abdomen and pelvic. The same base plate shall be upgraded by adding

localizer box, thorax abdomen bridges, wedges, Upper Arm support, lower arm support, Indexed Couch

stoppers, knee rest, feet fix to adopt for SBRT and SRS/SRT frameless and there shall be 2 set of each to

be provided.

The vendor shall provide 2 set of carbon fiber based Head rest, prone Head rest universal, Pediatric

Supine, Cushion for shoulder, Breast, board Carbon fiber with all required accessories.

Also the vendor shall provide – - Water bath with digital temperature control (1 no.) - Bolus 0.5 cm (3

nos.) - Bolus 1.0 cm (3 nos.) - Body caliper (2 nos.) - Heat Gun (1 no.) - Essential tools set (1 no.) -

Electron Foam cutter (1 no.) - CT markers (1000 nos.) - Vacuum cushions for Head & Neck (6 nos.) -

Vacuum cushions for Thorax (10 nos.) - Vacuum cushions for Abdomen (10 nos.) - Vacuum cushions for

whole body (10 nos.) - Suitable Vacuum pump (1 no.)

The immobilization devices should be compatible with the existing sets of immobilization devices.

26. Training Schedule

a) On-site training should be provided to all staff for at least two weeks.

b) Additional training to be imparted on the equipment as follows, for one Physicist ,one Radiation

Technologist and one Oncologist for two working weeks in a RT facility located in reputed centre in any

developed country and reputed centre in India where the Linear Accelerator with high precision

technology is being extensively used .

27. Software up-gradation, if any, within the period of comprehensive warranty should be supplied free

of cost.

28. General :

Page 22: 23.01.2018 TENDER FOR FIXING RATE CONTRACT … text Add to the end of technical ... 6. Govt. Rajaji Hospital, Madurai Note: 1) ... Company should give the undertaking for supply of

The vendor should specify whether their system will be compatible with existing XiO TPS / Oncentra TPS

for IMRT, IGRT and VMAT / Rapid Arc planning.

The successful tenders will have to liaison with the agency appointed for site preparation.

Pre installation requirement should be specified clearly.

OPTIONAL ITEMS WITH THE HIGHENERGY LINEAR ACCELERATOR

SYSTEM

Note : The technical and price comparison between various firms will be done for essential items only

and NOT the optional items. Depending on the availability of extra funds at the time of placing the

supply order, each option may be purchased in a sequential manner that is one by one starting from

OPTION-1.

OPTION-1

4D IMAGE GUIDANCE:

4DCBCT image Acquisition software:-

In CBCT, the software shall do in line acquisition & reconstruction of 4D volumetric data by sorting

and reviewing the moving anatomy within the projection images and calculating a respiratory trace

directly from the internal anatomy without external surrogates. Each reconstructed phase of the

respiratory cycle is matched to a 3D reference image automatically to review the results quickly and

efficiently and correction vectors calculated automatically to position the tumor in either the average or

the exhale position. Also, with diagnostic images, there shall be visualization and utilization of 4D image

sets to create structures and review treatment options with respiratory-correlated CT images, the

robust toolset to multiple view such as CINE MODE and variable window formats, and automate

processes for creation of Internal Target Volumes.

The software shall help in visualization and navigation of 4D patient data from CT, PET, MR and

Respiratory Correlated CT, to allow clinicians to display any data set in any window and review changes

to the tumor over the respiration cycle and enhances the planning process by importing Maximum

Intensity Projection and other specialty images directly from the imaging device, or create these images

from 4D image data. MIP and MinIP images are formulated from the original data set based on the

needs of the clinic and the viewing preferences of the clinicians. In this the Internal Target Volumes shall

automatically be created from the envelope of structures that constitute the tumor volume on multiple

respiratory correlated image sets with automatic margins for the Planning Target Volume. Segmentation

with rapid structure delineation performed on any view including transverse, sagittal and coronal and

updated in real-tile on other planar views. The procedure is to correct the base line shift that occurs in

lung tumors and for this the 4D CT scanner software will be made available by the buyer institute to

make the system complete.

Page 23: 23.01.2018 TENDER FOR FIXING RATE CONTRACT … text Add to the end of technical ... 6. Govt. Rajaji Hospital, Madurai Note: 1) ... Company should give the undertaking for supply of

Additionally the CBCT shall have option to do Registration Region of Interest generated from any

structure imported from the Treatment Planning System, or created manually using tools in the

software for generation of a 3D registration volume which conforms to anatomical structures. The

registration of two separate areas of anatomy in CBCT, utilizing both the Clipbox and the Shaped

Registration Region of Interest. CBCT software will calculate the relationship of both areas of anatomy to

the proposed correction vectors and alert the user if the target has moved closer to the critical

structures due to anatomical changes which enable user to select a compromise between the two areas,

or send the patient for re-planning.

Software

a) It should be possible to match kV – MV and MV – kV set-up images with DRR.

b) Compare kV radiographic images with reference to compute correction table coordinates and

execute the correction from console and inside the treatment room.

c) Cone beam CT images should be transferrable to the quoted TPS to do the online / offline planning

d)It should be possible to acquire images in gated mode using both kV and MV imagers

e)Digital Fluoroscopy should be available to visualize the gate-on and gate-off fields

f)The image verification tools should include: Blend images, split window, moving window and

complementary colour blending

g)Calibrated Hounsfield units (HU) must be available for CBCT images to be used for RT planning.The

system must be fully DICOM compliant with CT import and export, RT Plan import, RT structures import

and RT image import.Calibration and Flexmap Phantoms should be provided.

h)Phantoms: All necessary QA phantoms for IGRT, Gating, kV imaging and SRS/SRT should be

supplied

i)Shifts calculated during off line image review should be automatically available on Linear Accelerator

during next treatment.

j)System should have collision detector and laser guard and or any anti collision mechanism.

Page 24: 23.01.2018 TENDER FOR FIXING RATE CONTRACT … text Add to the end of technical ... 6. Govt. Rajaji Hospital, Madurai Note: 1) ... Company should give the undertaking for supply of

OPTION-2

High End Patient Positioning System of Linear Accelerator Meant For

Real Time Image Guided Radiotherapy, Frameless Radiosurgery &

Stereotactic Body Radiotherapy Treatment

The specialized & independent patient positioning platform should be able to perform various

specialized treatment techniques such as : Real time Image Guided Radiotherapy, Frameless

Radiotherapy & Stereotactic Body Radiotherapy. Room based Real time IGRT system (Patient Positioning

and Tracking Platform)

User Workstation Intel quad core 2.66 GHz workstation with below configuration as below: - Min. 4 GB

RAM - Min 500 GB HDD - CD / DVD writer - Min 4x ports - Min 2x x-ray frame grabber - Video frame

grabber - High-end graphic card - TFT 19” for workstation with keyboard & monitor Room based IGRT

Hardware should include- - Ceiling mounted IR cameras - Ceiling mounted Flat panel detector - Power &

cabling - Ceiling mounted Monitor system - Floor mounted KV X-ray system

Optical Tracking System: Real-time Infrared optical tracking of the patient’s position and motion /

external breathing patterns for initial set-up and during treatment delivery. It should have Couch

motion to correct patient setups online based on reflective body markers. Interactive control suitable

for 6D robotic couch. Two Stereoscopic infrared cameras for patient tracking in 3 translational and 3

rotational degree of freedom. In-room software interface via ceiling mounted touchscreen monitor

Infrared tracking system should continuously track the external marker in any possible treatment

position (coplanar & non-coplanar).

Room based KV imaging System : Two linac-independent kV X-ray units floor mounted and two

ceiling mounted flat panel detectors combined with IR tracking to monitor patient’s position throughout

treatment delivery and error in both coplanar & non-coplanar couch position by entering the values in

the Imaging system. It should have dual x-ray generator, x-ray tubes & ceiling mounted digital imagers. It

should have a seamless interaction with Optical tracking system to provide a fast & accurate workflow.

The 6D correction should utilize stereoscopic high resolution kV imaging in the treatment room. The KV

imaging should be such that it should be able to provide standalone as well as intratreatment imaging

(while the beam is on) for any couch and gantry angle to accommodate

verification needs for advanced SBRT treatment. It should have the capability of generating monoscopic

or dual snap shot simultaneously allowing quick intra-fraction tumour motion & immediate output of

required corrections (should support all non-coplanar fields). It should perform multiple 6D DRR

calculation from CT data in different planes to identify rotational errors. It should have the capability to

manually & automatically fuse DRR with X-ray images. It should calculate

Page 25: 23.01.2018 TENDER FOR FIXING RATE CONTRACT … text Add to the end of technical ... 6. Govt. Rajaji Hospital, Madurai Note: 1) ... Company should give the undertaking for supply of

positioning errors in 6 degrees of freedom It should allow import of images, isocenter & volumes in

DICOM RT format It should support region of interest definition for patient registration.

It should have seamless interaction with the Record & Verify system It should have a comprehensive QA

checks It should support the following modules: - Frameless SRS - DICOM RT Export to R&V - X-ray snap

monitoring - IGRT review & approval Advanced Real time IGRT planning. It should have automatic

detection of internal markers & urethral stent markers. It should have an advanced review system for

the room based IGRT X-ray treatment. It should have seamless review & approval of IGRT setup with

password protected workflow. It should have target localization, patient positioning, and motion

management utilizing bony anatomy or implanted fiducials for both Cranial & Extra cranial indication

(whole body IGRT). It should have Frameless Radio Surgery setup in x,y,z and all 3 rotational dimensions

in combination with Robotics module.

Intra-fraction imaging for detection of PTV misalignments and movements at any point in time during

the treatment (beam on / off; any gantry rotation, any couch rotation).

Interaction with Robotic Couch top: The system should have the capability to calculate and generate the

shift values, which can then be entered into robotic couch top for 6D corrections to happen based on

these values with the 6D robotic couch system.

Frameless SRS capabilities integrated with robotic couch top: It should have a non-invasive patient

immobilization for image guided set-up and should be integrated with robotic couch top with the

Frameless Extension. It should have imaging support and correction for highly accurate & precise SRS in

coplanar & non-planar position. It should be capable of positioning the patient within all six degrees of

freedom to precisely align the tumour within the beam path throughout the entire course of treatment.

The workflow optimized x-ray acquisition and the therapist from outside the treatment room operates

verification procedure. It should include customized mask for each patient in order to immobilize the

patient while repositioning should be done based on internal structures or body anatomy detected with

x-rays images.

The Frameless mask system should have all relevant accessories, disposable cranial SRS masks &

Infrared Body markers

Quality Assurance : Relevant QA phantoms like – - Pelvic Phantom - Isocenter Phantom - X-ray

calibration phantom

OPTION-3

Please specify the software and hardware for adaptive radiotherapy.

Note : Software up-gradation if any of optional items purchased, within the period of comprehensive

warranty should be supplied.

Page 26: 23.01.2018 TENDER FOR FIXING RATE CONTRACT … text Add to the end of technical ... 6. Govt. Rajaji Hospital, Madurai Note: 1) ... Company should give the undertaking for supply of

Turnkey

Civil Construction and Electrification : Turnkey will include whole as under mentioned :-

Construction of adequate size UPS room with aluminium partitions, TPS contouring

station (10x10m2), Physicst TPS room planning station (10x10m2), Patient review room

(5x5m2), in the reception area, Chiller and Air handling room. Provision of multi-storey rags in

the mould room for keeping the moulds to be given. Provision of tiles in the patient waiting hall

of old cancer, Block and entrance door to access the path for LINAC room. Vendor must visit

the site / department for inspection of area, construction and scope for any modification.

All wires used must be FRLS (Fire Retardant with Low Smoke) type only.

Floor trenches with block board covers will be provided for the cables in the LA and

Equipment rooms.

Furniture should be of high quality (branded/superbrand). False ceiling in all the areas

will comprise of metal suspension system, perforated fireproof aluminium panels with integrated

acoustic lining. Smoke detectors to be accommodated / integrated in the false ceiling.

All the internal wiring including that of telephone, LAN, DICOM & PACS etc., will be of

concealed variety. Modular Cabinets should be provided. Fire Safety measures : A Fire alarm

system of reputed make with smoke / heat detectors, indicators panels, call boxes, electronic

sirens and wiring will be installed. Audio call bell system, with intercom and remote locking/

unlocking facility, to be provided at the main door of the complex. Closed Circuit cameras of

reputed company should be provided in the examination room, console room, Linear

Accelerator and waiting areas to be provided. Sun-film and Venetian blinds will be put up in all

windows. The entire complex will be made rodent/ pest proof.

Electrical Services : The LA and all connected / required equipment for treatment

planning, scanning, post-processing and filming are to be connected to the supplied UPS. All

the equipment/computers along with peripherals, light points, fire alarm system, EPAX and view

boxes are to be supplied power through the common or a separate UPS with suitable capacity

or more with 60 minutes back up. Dimmer controlled incandescent light fixtures are to be

provided in the gantry room, console room. Different parts of the complex will have separate

wiring for light and power circuits through MCBs of suitable capacity. Adequate safety

measures will be incorporated in the electrical power supply system. Dedicated isolated

earthling is to be provided for the complex.

The entire turnkey project (building and the supplied items) should have an underlying

colour scheme so that it gives a homogenous, aesthetically pleasant and patient friendly

Page 27: 23.01.2018 TENDER FOR FIXING RATE CONTRACT … text Add to the end of technical ... 6. Govt. Rajaji Hospital, Madurai Note: 1) ... Company should give the undertaking for supply of

appearance. Any addition and alteration as and when required during turnkey work will be done

by mutual discussion between the supplier and technical committee. The whole turnkey work

has to be done to the fullest satisfaction of the technical committee and End user. In case of

any difference of opinion the decision of the technical committee shall be final and binding on

the supplier.

Furniture Console, Gantry room Modular furniture of branded company for console

equipment and staff. Modular cupboard and cabinet with lock and key. Equipment Room

Modular Cupboard with lock and key to keep spares and user manual Modular Cupboard for

stationary and record.

Page 28: 23.01.2018 TENDER FOR FIXING RATE CONTRACT … text Add to the end of technical ... 6. Govt. Rajaji Hospital, Madurai Note: 1) ... Company should give the undertaking for supply of

Annexure-III

PRICE SCHEDULE

1 2 3 4 5 6 7 8

PRICE FOR EACH UNIT Sch

No.

Item Description Country

of

origin

Quantity

& Unit

Ex-factory Ex-

warehouse Ex-

showroom off-the

shelf for CIF

(a)

Packing

&

forwarding

(b)

Inland transport,

Insurance and

Incidental costs

incidental to

delivery

(c)

Incidental

services as

listed in

clause 7 of

SCC

(d)

Custom

s duty

(e)

Unit price

a+b+c+d+e

Total price per

schedule for

delivery at final

destination

(4 x 6)

GST –

IGST/CGS

T/ SGST

payable if

contract is

awarded

1.

Dual Energy Linear

Accelerator with optional

items

1 no.

Optional items

1a. Option 1- Treatment

planning System

1 no.

1b. Option 2- Array detector 1 no.

1c. Option 3- Stereotactic

Solution

1 no.

1d. Option 4 – 4D image

Guidance

1 no.

1e. Option 5- Biological

Optimization algorithm

1 no.

1f.

Option 6 - 4D Advanced

patient monitoring device

marker free, surface scan

based gating system for

patient monitoring in LA

Room

1 no.

1g. Option 7 – Robotic Couch

top

1 no.

Page 29: 23.01.2018 TENDER FOR FIXING RATE CONTRACT … text Add to the end of technical ... 6. Govt. Rajaji Hospital, Madurai Note: 1) ... Company should give the undertaking for supply of

1h.

Option 8- High End Patient

Positioning System of

Linear Accelerator Meant

For Real Time Image

Guided Radiotherapy,

Frameless Radio surgery

& Stereotactic Body

Radiotherapy Treatment

1 no.

1i.

Option 9 - software and

hardware for adaptive

radiotherapy

1 no.

2. Dual Energy Linear

Accelerator

1 no.

3. Single Energy Linear

Accelerator

1 no.

4. Advanced Dual Energy

Linear Accelerator

1 no.

Optional items

4a 4D Image Guidance 1 no.

4b

High End Patient

Positioning System of

Linear Accelerator Meant

For Real Time Image

Guided Radiotherapy,

Frameless Radio surgery &

Stereotactic Body

Radiotherapy Treatment

1 no.

4c Software and hardware for

adaptive radiotherapy

1 no.

i. Unit price in (6) (Rs. in words)

1. Dual Energy Linear Accelerator with optional items as per specification with HSN Code/rate of GST..........................................................................................................

Page 30: 23.01.2018 TENDER FOR FIXING RATE CONTRACT … text Add to the end of technical ... 6. Govt. Rajaji Hospital, Madurai Note: 1) ... Company should give the undertaking for supply of

2. Dual Energy Linear Accelerator as per specification with HSN Code/rate of GST..........................................................................................................................................

3. Single Energy Linear Accelerator as per specification with HSN Code/rate of GST......................................................................................................................................

4. Advanced Dual Energy Linear Accelerator as per specification with HSN Code/rate of GST ………………………….……………………………………………………………………………………….

Options for Schedule no. 1

1a.Treatment planning System…………………………………………………………………………………………………………………………………………………

1b.Array detector………………………………………………………………………………………………………………………………………………………………

1c.Stereotactic Solution………………………………………………………………………………………………………………………………………………………….

1d.4D image

Guidance…………………………………………………………………………………………………………………………………………………………………………

1e.Biological Optimization algorithm………………………………………………………………………………………………………………………………………….

1f.4D Advanced patient monitoring device marker free, surface scan based gating system for patient monitoring in LA Room

1g.Robotic Couch top………………………………………………………………………………………………………………………………………………………….

1h.High End Patient Positioning System of Linear Accelerator Meant For Real Time Image Guided Radiotherapy, Frameless Radio surgery & Stereotactic Body

Radiotherapy Treatment………………………………………………………………………………………………………………………………………………………..

1i. software and hardware for adaptive radiotherapy…………………………………………………………………………………………………………………………..

Options for Schedule no. 4

4a. 4D Image Guidance………………………………………………………………………………………………………………………………………………………………………………………………………………………………

4b. High End Patient Positioning System of Linear Accelerator Meant For Real Time Image Guided Radiotherapy, Frameless Radiosurgery & Stereotactic Body

Radiotherapy Treatment……………………………………………………………………………………………………………………………………………………...

4c. Software and hardware for adaptive radiotherapy…………………………………………………………………………………………………………………………...

a) Annual Maintenance Charges (labour only) per year /per unit for 7 years after 3 years free warranty maintenance period

1. Dual Energy Linear Accelerator with optional items as per specification with HSN Code/rate of GST............................................................................... ...........................

2. Dual Energy Linear Accelerator as per specification with HSN Code/rate of GST..........................................................................................................................................

3. Single Energy Linear Accelerator as per specification with HSN Code/rate of GST......................................................... .............................................................................

4. Advanced Dual Energy Linear Accelerator as per specification with HSN Code/rate of GST ………………………………..…………………………………………….

b) Annual Maintenance Charges (Comprehensive) for 7 years / per year / per unit after free warranty maintenance period

1. Dual Energy Linear Accelerator with optional items as per specification with HSN Code/rate of GST..........................................................................................................

2. Dual Energy Linear Accelerator as per specification with HSN Code/rate of GST..........................................................................................................................................

3. Single Energy Linear Accelerator as per specification with HSN Code/rate of GST......................................................................................................................................

4. Advanced Dual Energy Linear Accelerator as per specification with HSN Code/rate of GST ………………………….……………………………………………………………………………………….

Page 31: 23.01.2018 TENDER FOR FIXING RATE CONTRACT … text Add to the end of technical ... 6. Govt. Rajaji Hospital, Madurai Note: 1) ... Company should give the undertaking for supply of

Note:

(a) In case of discrepancy between unit price and total price, the unit price shall prevail.

(b) This price schedule should be placed in separate sealed cover “Cover B”

(c) GST applicable for Annual Maintenance Charges shall be indicated separately.

(d) The bidder should indicate the HSN code of the equipment/ service and applicable GST rates.

(e) The bidder should quote for the prices separately from column 5(a) to 5(e) and should not state “as inclusive”.

(f) The bidders are permitted to quote any one or all the Schedule.

(g) TNMSC reserves the rights to procure any number of Linear accelerator to any of the listed hospitals in any of the four categories based on the price

differential and case loads and the relative advantage of prices with respect to higher specification. The decision of TNMSC in this aspect shall be final.

Place : Signature of Bidder......................……………..

Date : Name ……………………………

Business Address ……………………………