a d v a n c e d p a c i n g

69
ADVANCED PACING Understanding and Troubleshooting Pacemakers Sherry L. Knowles, RN, CCRN, CRNI

Upload: sherry-knowles

Post on 07-May-2015

11.773 views

Category:

Education


3 download

DESCRIPTION

Pacemakers

TRANSCRIPT

Page 1: A D V A N C E D  P A C I N G

ADVANCED PACING

Understanding and Troubleshooting Pacemakers

Sherry L. Knowles, RN, CCRN, CRNI

Page 2: A D V A N C E D  P A C I N G

Objectives

1. Understand the different pacemaker modes according to the NASPE/BPEG (NGB) universal code.

2. Recognize common pacemaker complications in a cardiac tracing.

3. Differentiate between failure to pace, loss of capture, under sensing and over sensing.

4. List appropriate nursing actions for common pacemaker complications.

5. Understand routine pacemaker settings and thresholds.

6. Identify the different types of pacemakers.

7. Understand and recognize pacemaker syndrome.

8. Be familiar with the proper use of a pacemaker magnet.

9. Understand routine nursing management of a patient with a pacemaker.

10. Recognize battery failure in temporary and permanent pacemaker devices.

At the end of this presentation, the view will be able to:

Page 3: A D V A N C E D  P A C I N G

Pacemaker

Page 4: A D V A N C E D  P A C I N G

Nursing Action

Page 5: A D V A N C E D  P A C I N G

Indications for Pacing

Page 6: A D V A N C E D  P A C I N G

Indications for Pacing• Symptomatic Bradycardia

• Heart Blocks– Second Degree AV Block – symptomatic

– Complete AV Block

• Sick Sinus Syndrome

• Overdrive Pacing

• Malfunctioning Pacemaker

• Support During Electrolyte or Acid-Base Disturbances

• To Improve Hemodynamics

• Support During Procedures

• Backup Support With Drug Therapy

Page 7: A D V A N C E D  P A C I N G

Types of Pacing

• Temporary Pacing– Transvenous– Transcutaneous– Epicardial

• Permanent Pacing– High/Low Rate Setting– PR and AV Intervals– Rate Response Pacing– Mode Switching

• Fixed Rate• Controlled• Asynchronous

• Demand• Backup• Synchronous

• AV Sequential• Dual Chamber Pacing• Atrial & Ventricular wires

Page 8: A D V A N C E D  P A C I N G

Pacemaker Types

Page 9: A D V A N C E D  P A C I N G

Epicardial Pacing

Page 10: A D V A N C E D  P A C I N G

Transcutaneous Pacing

Page 11: A D V A N C E D  P A C I N G

Insertion Complications

• Dysrhythmias

• Perforation

• Pneumothorax

• Hematoma

• Hemorrhage

• Tamponade

• Pericardial Friction Rub

• Vein Thrombosis

• Infection

• Inadvertent Arterial Puncture

• Diaphragmatic Stimulation

Page 12: A D V A N C E D  P A C I N G

Modes of Pacing

Chamber PacedO = NoneA = AtriumV = VentricleD = Dual ( A + V)

Chamber SensedO = NoneA = AtriumV = VentricleD = Dual ( A + V)

Response to SensingO = NoneT = Triggered I = InhibitedD = Dual ( T + I)

ProgrammabilityO = NoneP = Simple ProgrammableM= Multiple ProgrammableR = Rate Modulation

Page 13: A D V A N C E D  P A C I N G

Pacemaker ModesAOO: Atrial pacing, no sensing, atrial fixed rate

VOO: Ventricular pacing, no sensing, ventricular fixed rate.

DOO: AV pacing, no sensing, AV fixed rate.

AAI: Atrial Pacing, atrial sensing, inhibited response to sensing. Atrial demand pacing.

VVI: Ventricular pacing, ventricular sensing, inhibited response to sensing. Ventricular demand pacing.

DDI: A pacemaker that senses and stimulates both atrium and ventricle. The stimulation rate is fixed.

AOO

VOO

DOO

AAI

VVI

DDI

Page 14: A D V A N C E D  P A C I N G

Pacemaker ModesSSI: Single chamber pacing (atrial or ventricular), same chamber

sensing, inhibited response to sensing. A pacemaker that senses and stimulates one chamber at one rate.

AAT: Atrial pacing, atrial sensing, triggered response to sensing. Atrial demand pacing.

VVT: Ventricular pacing, ventricular pacing, triggered response to sensing. Ventricular demand pacing.

VAT: Ventricular pacing, atrial sensing, triggered response to sensing. The ventricular stimulus will fire at a set interval after sensing a spontaneous atrial depolarization. AV synchronous pacing.

VDD: A pacemaker that senses the atrium and ventricle and paces the ventricle (atrial synchronous, ventricular inhibited).

DDD: A pacemaker that senses and stimulates both atrium and ventricle. The stimulation rate, tracking rate and minimum

rate are determined.

SSI

AAT

VVT

VAT

VDD

DDD

Page 15: A D V A N C E D  P A C I N G

Pacemaker ModesSSIR: A pacemaker that senses and stimulates one chamber at a rate

determined by a sensor.

AAIR: A pacemaker that senses and stimulates the atrium at a rate determined by a sensor.

VVIR: A pacemaker that senses and stimulates the ventricle at a rate determined by a sensor.

DDIR: A pacemaker that senses and stimulates both atrium and ventricle. The stimulation rate is governed by a maximum rate, the rate determined by a sensor and the minimum rate.

DDDR: A pacemaker that senses and stimulates both atrium and ventricle. The stimulation rate is governed by a maximum rate, the rate determined by a sensor and the minimum rate.

SSIR

AAIR

VVIR

DDIR

DDDR

Page 16: A D V A N C E D  P A C I N G

Single Chamber Pacing

• AAI– Paces and senses only the atria – Must have proper AV conduction

• VVI– Looses AV synchronicity– Used for temporary backup or with loss of

AV conduction (AFib)

Page 17: A D V A N C E D  P A C I N G

Atrial Pacing

Page 18: A D V A N C E D  P A C I N G

Ventricular Pacing

Page 19: A D V A N C E D  P A C I N G

Loss of AV Synchrony

• Loss of atrial contribution

• Decrease in LV preload

• Decrease in stroke volume

• Decrease in cardiac output

• Decrease in cerebral perfusion

• Decrease in coronary perfusion

• Increase risk of pacemaker syndrome

Page 20: A D V A N C E D  P A C I N G

Dual Chamber Pacing

• Maintains / Restores AV Synchrony

• Optimizes heart function

• Reduces pacemaker syndrome

• Useful in CHF

• Capable of mode switching

Page 21: A D V A N C E D  P A C I N G

AV Pacing

Page 22: A D V A N C E D  P A C I N G

Rate Response Pacing

• Uses sensor technology to provide rate modulation with exercise

• AAIR

• VVIR

• DDDR

Page 23: A D V A N C E D  P A C I N G

Rate Response Sensors

• Activity Sensors– sensors that detect body movement. The

more the patient’s body is moving (so the reasoning goes) the faster the heart rate should be.

• Breathing Sensors– measures the patient’s rate of breathing.

The faster the breathing, the faster the heart rate should be.

Page 24: A D V A N C E D  P A C I N G

Rate Response Pacing

Page 25: A D V A N C E D  P A C I N G

Pacemaker Syndrome Often due to loss of AV Synchrony and/or

delayed conduction time

• Shortness of Breath

• Fatigue

• Cough

• Orthopnea

• Headache

• Syncope

• Vertigo

• Choking Sensation

• CHF, Pulmonary Edema

• Dizziness

• Apprehension

• Pulsations in the Neck

• Palpitations

• Chest Pain

• Near Syncope

• Jaw Pain

• Confusion

• Altered Mental Status

Page 26: A D V A N C E D  P A C I N G

Troubleshooting

• Twelve Leads are Better than One.

• Choose leads that show the P wave and QRS morphology best.

• Look for the pacer spikes.

Page 27: A D V A N C E D  P A C I N G

Failure To Pace

Page 28: A D V A N C E D  P A C I N G

Causes of Output Failure

• Lead Fracture

• Lead Dislodgement

• Lead Disconnect

• Electromagnetic Interference

• Battery Failure

• Component Failure

• Over sensing

Page 29: A D V A N C E D  P A C I N G

Nursing Actions

• Assess and Treat Patient

• Temporary Pacemaker– Make Sure Unit Is Turned On

– Check Battery and Connections

– Check Rate, Increase Rate As Needed

– Check mA, Increase Output As Needed

– Decrease Sensitivity (toward asynchronous or 20mV)

• Prepare For Transcutaneous Pacing

• Notify MD As Necessary

Page 30: A D V A N C E D  P A C I N G

Failure To Capture

Page 31: A D V A N C E D  P A C I N G

Failure To Capture

Page 32: A D V A N C E D  P A C I N G

Causes of Loss of Capture

• Lead Dislodgement• Lead Insulation Defect• Battery Failure• Exit Block• Elevated Pacing Thresholds• Acute MI• Electrolyte Imbalance• Medications• Metabolic Disorders• Defibrillation• Long QT Syndrome

Page 33: A D V A N C E D  P A C I N G

Loss of Capture

• Causes:– Catheter Dislodgement

– Catheter Fracture

– Unstable Connection

– Local Myocardial Necrosis/Fibrosis

– Local Myocardial Inflammation

Page 34: A D V A N C E D  P A C I N G

Loss of Capture

• Causes:– Drug Effects

– Metabolic Effects

– Cautery (DC Cardioversion, Ablation)

– Lead Conductor or Insulation Failure

– Generator Malfunction

Page 35: A D V A N C E D  P A C I N G

Nursing Actions

• Assess and Treat Patient

• Temporary Pacemaker– Make Sure Unit Is Turned On

– Check Battery and Connections

– Check mA, Increase Output As Needed

– Check Rate, Increase Rate As Needed

– Decrease Sensitivity (toward asynchronous or 20mV)

• Prepare For Transcutaneous Pacing

• Notify MD As Necessary

Page 36: A D V A N C E D  P A C I N G

Pacemaker Sensitivity

• Low Sensitivity

– Towards Asynchronous

– Toward Higher mV (20mV)

– Decreases pacemaker sensing ability

• High Sensitivity

– Maximal Sensitivity

– Toward Lower mV (1mV)

– Increases pacemaker sensing ability

Page 37: A D V A N C E D  P A C I N G

Failure To Sense

Page 38: A D V A N C E D  P A C I N G

Failure To Sense

Page 39: A D V A N C E D  P A C I N G

Nursing Actions

• Assess and Treat Patient

• Temporary Pacemaker– Make Sure Unit Is Turned On

– Check Battery and Connections

– Increase Sensitivity (toward synchronous or 1mV)

– Check mA, Increase Output As Needed

– Check Rate, Increase Rate As Needed

• Prepare For Transcutaneous Pacing

• Notify MD As Necessary

Page 40: A D V A N C E D  P A C I N G

Pacemaker Interference

• Electromagnetic Interference (EMI)– Electromagnetic fields– Defibrillation– Electrocautery – Ablation – Lithrotripsy– Magnetic Resonance imaging (MRI)– High dose radiotherapy– Cellular telephones – TENS

• Twiddler’s Syndrome

Page 41: A D V A N C E D  P A C I N G

Fusion Beats

Page 42: A D V A N C E D  P A C I N G

Pacemaker Magnet

• A pacemaker magnet placed over a pacemaker generator typically eliminates sensing and returns the pacemaker to an asynchronous pacing mode (AOO, VOO or DOO).

• When using a pacemaker magnet, the rhythm changes to a controlled paced rhythm.

• Use caution when removing a pacemaker magnet because it can cause a reentry tachycardia. The last asynchronous beat is conducted through the AV node to the atrium. In this case, treat this dysrhythmia by reapplying the magnet.

Page 43: A D V A N C E D  P A C I N G

Pacemaker Magnet Tracing

Page 44: A D V A N C E D  P A C I N G

Transvenous Pacing

• Transvenous Pacing

– Start with new battery

– Secure leads and connections

– Set mA to 5 and adjust for capture

– Keep mA 2-3 times the threshold

Page 45: A D V A N C E D  P A C I N G

Transvenous Threshold

• Transvenous Threshold– Gradually decrease mA from 5 (to 0) until

1:1 capture is lost. Then increase mA until 1:1 capture is reestablished.

– Set the mA 2-3 times the threshold.

– Check threshold qd and prn, unless directed otherwise by physician.

Page 46: A D V A N C E D  P A C I N G

Transcutaneous Pacing

• Transcutaneous Pacing

– Keep pads off pacemaker/ICD devices

– Set mA to 40 and adjust for capture

– Remove paste from chest wall

– Avoid arching

– Keep mA 2 X threshold

Page 47: A D V A N C E D  P A C I N G

Transcutaneous Threshold

• Transcutaneous Threshold– Gradually decrease mA from to 40 until 1:1

capture is lost. Then increase mA until 1:1 capture is reestablished.

– Set the mA 2 times the threshold.

– Check threshold qd and prn, unless directed otherwise by physician.

Page 48: A D V A N C E D  P A C I N G

Epicardial Pacing

• Epicardial Pacing

– Start with new battery

– Secure leads and connections

– Set mA to 3-5 and adjust for capture

– Keep mA 3 X threshold

Page 49: A D V A N C E D  P A C I N G

Epicardial Threshold

• Epicardial Threshold– Gradually decrease mA from to 5 until 1:1

capture is lost. Then increase mA until 1:1 capture is reestablished.

– Set the mA 3 times the threshold.

– Check threshold qd and prn, unless directed otherwise by physician.

Page 50: A D V A N C E D  P A C I N G

Pacemaker Sensitivity

• Low Sensitivity

– Towards Asynchronous

– Toward Higher mV (20mV)

– Decreases pacemaker sensing ability

• High Sensitivity

– Maximal Sensitivity

– Toward Lower mV (1mV)

– Increases pacemaker sensing ability

Page 51: A D V A N C E D  P A C I N G

Sensitivity Threshold

• Sensitivity Threshold– Set rate just below patient’s intrinsic rate,

the pacemaker should stop pacing and the sensor light should start flashing.

– Gradually decrease sensitivity until the pacemaker begins to fire. This is the sensitivity threshold.

– Set sensitivity to 2-3 times the sensitivity threshold.

Page 52: A D V A N C E D  P A C I N G

Other Pacemaker Settings

• High/Low Rate– Sets the highest and lowest rates that the

pacemaker can pace.

• AV Interval Control – Set between 150-250 msec to preserve AV

synchrony.

• Refractory Periods – Sets a period of msec in which the pacemaker

will not respond after a sensed or paced beat.

Page 53: A D V A N C E D  P A C I N G

Special Considerations

• Not necessary to disconnect temporary pacemaker during cardioversion or defibrillation

• Changes in pacing threshold may occur with electrolyte imbalance, myocardial ischemia, anemia, CHF, hypoxia, or a change in drug therapy

•  Pacing threshold increases over time

Page 54: A D V A N C E D  P A C I N G

Nursing Management• Continuous Cardiac Monitoring

– Watch for pacemaker complications– Always evaluate cardiac tracings

• Electrical Safety Measures– Use rubber gloves when handling external pacemaker

components– Regularly inspect all connections– Replace batteries as needed ( have spares available)– Avoid EMI sources

• Monitor Thresholds• Monitor All Imbalances

• Temperature, Acid-Base, Electrolytes

• Monitor Medication Effects• Watch For Pacemaker Syndrome

Page 55: A D V A N C E D  P A C I N G

Nursing Assessment

• Assess the Patient– Vital Signs

– Cardiac Rhythm (always run strips)

– Patient Complaints

• Assess the Pacemaker– Maintain Electrical Safety

– Inspect and Secure Connections

– Assess the Effectiveness of the Pacemaker

– Check Thresholds

– Change Battery As Needed (keep spares available)

Page 56: A D V A N C E D  P A C I N G

Pacemaker Evaluation

• What is the rhythm?

• Is the pacemaker functioning properly?

• What is the immediate nursing action?

Page 57: A D V A N C E D  P A C I N G

Pacer Rhythm Evaluation What is the pacing Mode?

What is the underlying rhythm?

Is the pacer firing appropriately?

Is the capture appropriate?

Is the sensing appropriate?

What is the low rate limit?

What is the high rate limit?

What is the AV delay?

Page 58: A D V A N C E D  P A C I N G

Nursing Management• Continuous Cardiac Monitoring

– Watch for pacemaker complications– Always evaluate cardiac tracings

• Electrical Safety Measures– Use rubber gloves when handling external pacemaker

components– Regularly inspect all connections– Replace batteries as needed ( have spares available)– Avoid EMI sources

• Monitor Thresholds• Monitor All Imbalances

• Temperature, Acid-Base, Electrolytes

• Monitor Medication Effects• Watch For Pacemaker Syndrome

Page 59: A D V A N C E D  P A C I N G

PRACTICE

Page 60: A D V A N C E D  P A C I N G

AAI

AAI: A pacemaker that stimulates and senses the atrium at one rate

Page 61: A D V A N C E D  P A C I N G

VVI

VVI: A pacemaker that stimulates and senses the ventricle at one rate

Page 62: A D V A N C E D  P A C I N G

DDI

DDI: A pacemaker that senses and stimulates both atrium and ventricle. The stimulation rate is fixed.

Page 63: A D V A N C E D  P A C I N G

DDD

Page 64: A D V A N C E D  P A C I N G

DDD or VVI?

Page 65: A D V A N C E D  P A C I N G

DDDR

Page 66: A D V A N C E D  P A C I N G

DDI

Page 67: A D V A N C E D  P A C I N G

What’s the Mode?

Page 68: A D V A N C E D  P A C I N G

What’s the Mode?

Page 69: A D V A N C E D  P A C I N G

References• Guidelines for Pacemakers and Defibrillators Updated. Journal Watch General Medicine

1998: 2-2.

• Wood, D.A., Fox, K.F., Gibbs, S.R. (2001). Rapid cardiology--for chest pain, breathlessness and palpitations. QJ Med 94: 177-178.  

• HAMMILL, S. C., HUBMAYR, R. D. (2000). The Rapidly Changing Management of Cardiac Arrhythmias. Am J Respir Crit Care Med 161: 1070-1073.

• Bauersfeld, U., Nowak, B., Molinari, L., Malm, T., Kampmann, C., Schonbeck, M. H., Schuller, H. (1999). Low-energy epicardial pacing in children: the benefit of autocapture. Ann. Thorac. Surg. 68: 1380-1383.

• KURBAAN, A S, SUTTON, R (1999). Pacing for vasovagal syncope. Heart 82: 649-650

• Braunwall, e. (1992) Heart Disease : a textbook of cardiovascular disease, 4 th edition, WB Saunders

• American Radio Relay League, Inc., Technical Information Service, 225 Main St., Newington, CT 06111 (860) 594-0214.

• Thelan, Lynne A., Davie, Joseph K., Urden, Linda D., Lough, Mary E. (1994) Critical Care Nursing: Diagnosis and Management. Second Edition. Pg 313-322.

• Graver, K (1998) A Practical Guide to EKG Interpretation, 2nd edition

• On-line: http://www.emedicine.com/emerg/topic699.htm