a d v a n c e d p a c i n g
DESCRIPTION
PacemakersTRANSCRIPT
ADVANCED PACING
Understanding and Troubleshooting Pacemakers
Sherry L. Knowles, RN, CCRN, CRNI
Objectives
1. Understand the different pacemaker modes according to the NASPE/BPEG (NGB) universal code.
2. Recognize common pacemaker complications in a cardiac tracing.
3. Differentiate between failure to pace, loss of capture, under sensing and over sensing.
4. List appropriate nursing actions for common pacemaker complications.
5. Understand routine pacemaker settings and thresholds.
6. Identify the different types of pacemakers.
7. Understand and recognize pacemaker syndrome.
8. Be familiar with the proper use of a pacemaker magnet.
9. Understand routine nursing management of a patient with a pacemaker.
10. Recognize battery failure in temporary and permanent pacemaker devices.
At the end of this presentation, the view will be able to:
Pacemaker
Nursing Action
Indications for Pacing
Indications for Pacing• Symptomatic Bradycardia
• Heart Blocks– Second Degree AV Block – symptomatic
– Complete AV Block
• Sick Sinus Syndrome
• Overdrive Pacing
• Malfunctioning Pacemaker
• Support During Electrolyte or Acid-Base Disturbances
• To Improve Hemodynamics
• Support During Procedures
• Backup Support With Drug Therapy
Types of Pacing
• Temporary Pacing– Transvenous– Transcutaneous– Epicardial
• Permanent Pacing– High/Low Rate Setting– PR and AV Intervals– Rate Response Pacing– Mode Switching
• Fixed Rate• Controlled• Asynchronous
• Demand• Backup• Synchronous
• AV Sequential• Dual Chamber Pacing• Atrial & Ventricular wires
Pacemaker Types
Epicardial Pacing
Transcutaneous Pacing
Insertion Complications
• Dysrhythmias
• Perforation
• Pneumothorax
• Hematoma
• Hemorrhage
• Tamponade
• Pericardial Friction Rub
• Vein Thrombosis
• Infection
• Inadvertent Arterial Puncture
• Diaphragmatic Stimulation
Modes of Pacing
Chamber PacedO = NoneA = AtriumV = VentricleD = Dual ( A + V)
Chamber SensedO = NoneA = AtriumV = VentricleD = Dual ( A + V)
Response to SensingO = NoneT = Triggered I = InhibitedD = Dual ( T + I)
ProgrammabilityO = NoneP = Simple ProgrammableM= Multiple ProgrammableR = Rate Modulation
Pacemaker ModesAOO: Atrial pacing, no sensing, atrial fixed rate
VOO: Ventricular pacing, no sensing, ventricular fixed rate.
DOO: AV pacing, no sensing, AV fixed rate.
AAI: Atrial Pacing, atrial sensing, inhibited response to sensing. Atrial demand pacing.
VVI: Ventricular pacing, ventricular sensing, inhibited response to sensing. Ventricular demand pacing.
DDI: A pacemaker that senses and stimulates both atrium and ventricle. The stimulation rate is fixed.
AOO
VOO
DOO
AAI
VVI
DDI
Pacemaker ModesSSI: Single chamber pacing (atrial or ventricular), same chamber
sensing, inhibited response to sensing. A pacemaker that senses and stimulates one chamber at one rate.
AAT: Atrial pacing, atrial sensing, triggered response to sensing. Atrial demand pacing.
VVT: Ventricular pacing, ventricular pacing, triggered response to sensing. Ventricular demand pacing.
VAT: Ventricular pacing, atrial sensing, triggered response to sensing. The ventricular stimulus will fire at a set interval after sensing a spontaneous atrial depolarization. AV synchronous pacing.
VDD: A pacemaker that senses the atrium and ventricle and paces the ventricle (atrial synchronous, ventricular inhibited).
DDD: A pacemaker that senses and stimulates both atrium and ventricle. The stimulation rate, tracking rate and minimum
rate are determined.
SSI
AAT
VVT
VAT
VDD
DDD
Pacemaker ModesSSIR: A pacemaker that senses and stimulates one chamber at a rate
determined by a sensor.
AAIR: A pacemaker that senses and stimulates the atrium at a rate determined by a sensor.
VVIR: A pacemaker that senses and stimulates the ventricle at a rate determined by a sensor.
DDIR: A pacemaker that senses and stimulates both atrium and ventricle. The stimulation rate is governed by a maximum rate, the rate determined by a sensor and the minimum rate.
DDDR: A pacemaker that senses and stimulates both atrium and ventricle. The stimulation rate is governed by a maximum rate, the rate determined by a sensor and the minimum rate.
SSIR
AAIR
VVIR
DDIR
DDDR
Single Chamber Pacing
• AAI– Paces and senses only the atria – Must have proper AV conduction
• VVI– Looses AV synchronicity– Used for temporary backup or with loss of
AV conduction (AFib)
Atrial Pacing
Ventricular Pacing
Loss of AV Synchrony
• Loss of atrial contribution
• Decrease in LV preload
• Decrease in stroke volume
• Decrease in cardiac output
• Decrease in cerebral perfusion
• Decrease in coronary perfusion
• Increase risk of pacemaker syndrome
Dual Chamber Pacing
• Maintains / Restores AV Synchrony
• Optimizes heart function
• Reduces pacemaker syndrome
• Useful in CHF
• Capable of mode switching
AV Pacing
Rate Response Pacing
• Uses sensor technology to provide rate modulation with exercise
• AAIR
• VVIR
• DDDR
Rate Response Sensors
• Activity Sensors– sensors that detect body movement. The
more the patient’s body is moving (so the reasoning goes) the faster the heart rate should be.
• Breathing Sensors– measures the patient’s rate of breathing.
The faster the breathing, the faster the heart rate should be.
Rate Response Pacing
Pacemaker Syndrome Often due to loss of AV Synchrony and/or
delayed conduction time
• Shortness of Breath
• Fatigue
• Cough
• Orthopnea
• Headache
• Syncope
• Vertigo
• Choking Sensation
• CHF, Pulmonary Edema
• Dizziness
• Apprehension
• Pulsations in the Neck
• Palpitations
• Chest Pain
• Near Syncope
• Jaw Pain
• Confusion
• Altered Mental Status
Troubleshooting
• Twelve Leads are Better than One.
• Choose leads that show the P wave and QRS morphology best.
• Look for the pacer spikes.
Failure To Pace
Causes of Output Failure
• Lead Fracture
• Lead Dislodgement
• Lead Disconnect
• Electromagnetic Interference
• Battery Failure
• Component Failure
• Over sensing
Nursing Actions
• Assess and Treat Patient
• Temporary Pacemaker– Make Sure Unit Is Turned On
– Check Battery and Connections
– Check Rate, Increase Rate As Needed
– Check mA, Increase Output As Needed
– Decrease Sensitivity (toward asynchronous or 20mV)
• Prepare For Transcutaneous Pacing
• Notify MD As Necessary
Failure To Capture
Failure To Capture
Causes of Loss of Capture
• Lead Dislodgement• Lead Insulation Defect• Battery Failure• Exit Block• Elevated Pacing Thresholds• Acute MI• Electrolyte Imbalance• Medications• Metabolic Disorders• Defibrillation• Long QT Syndrome
Loss of Capture
• Causes:– Catheter Dislodgement
– Catheter Fracture
– Unstable Connection
– Local Myocardial Necrosis/Fibrosis
– Local Myocardial Inflammation
Loss of Capture
• Causes:– Drug Effects
– Metabolic Effects
– Cautery (DC Cardioversion, Ablation)
– Lead Conductor or Insulation Failure
– Generator Malfunction
Nursing Actions
• Assess and Treat Patient
• Temporary Pacemaker– Make Sure Unit Is Turned On
– Check Battery and Connections
– Check mA, Increase Output As Needed
– Check Rate, Increase Rate As Needed
– Decrease Sensitivity (toward asynchronous or 20mV)
• Prepare For Transcutaneous Pacing
• Notify MD As Necessary
Pacemaker Sensitivity
• Low Sensitivity
– Towards Asynchronous
– Toward Higher mV (20mV)
– Decreases pacemaker sensing ability
• High Sensitivity
– Maximal Sensitivity
– Toward Lower mV (1mV)
– Increases pacemaker sensing ability
Failure To Sense
Failure To Sense
Nursing Actions
• Assess and Treat Patient
• Temporary Pacemaker– Make Sure Unit Is Turned On
– Check Battery and Connections
– Increase Sensitivity (toward synchronous or 1mV)
– Check mA, Increase Output As Needed
– Check Rate, Increase Rate As Needed
• Prepare For Transcutaneous Pacing
• Notify MD As Necessary
Pacemaker Interference
• Electromagnetic Interference (EMI)– Electromagnetic fields– Defibrillation– Electrocautery – Ablation – Lithrotripsy– Magnetic Resonance imaging (MRI)– High dose radiotherapy– Cellular telephones – TENS
• Twiddler’s Syndrome
Fusion Beats
Pacemaker Magnet
• A pacemaker magnet placed over a pacemaker generator typically eliminates sensing and returns the pacemaker to an asynchronous pacing mode (AOO, VOO or DOO).
• When using a pacemaker magnet, the rhythm changes to a controlled paced rhythm.
• Use caution when removing a pacemaker magnet because it can cause a reentry tachycardia. The last asynchronous beat is conducted through the AV node to the atrium. In this case, treat this dysrhythmia by reapplying the magnet.
Pacemaker Magnet Tracing
Transvenous Pacing
• Transvenous Pacing
– Start with new battery
– Secure leads and connections
– Set mA to 5 and adjust for capture
– Keep mA 2-3 times the threshold
Transvenous Threshold
• Transvenous Threshold– Gradually decrease mA from 5 (to 0) until
1:1 capture is lost. Then increase mA until 1:1 capture is reestablished.
– Set the mA 2-3 times the threshold.
– Check threshold qd and prn, unless directed otherwise by physician.
Transcutaneous Pacing
• Transcutaneous Pacing
– Keep pads off pacemaker/ICD devices
– Set mA to 40 and adjust for capture
– Remove paste from chest wall
– Avoid arching
– Keep mA 2 X threshold
Transcutaneous Threshold
• Transcutaneous Threshold– Gradually decrease mA from to 40 until 1:1
capture is lost. Then increase mA until 1:1 capture is reestablished.
– Set the mA 2 times the threshold.
– Check threshold qd and prn, unless directed otherwise by physician.
Epicardial Pacing
• Epicardial Pacing
– Start with new battery
– Secure leads and connections
– Set mA to 3-5 and adjust for capture
– Keep mA 3 X threshold
Epicardial Threshold
• Epicardial Threshold– Gradually decrease mA from to 5 until 1:1
capture is lost. Then increase mA until 1:1 capture is reestablished.
– Set the mA 3 times the threshold.
– Check threshold qd and prn, unless directed otherwise by physician.
Pacemaker Sensitivity
• Low Sensitivity
– Towards Asynchronous
– Toward Higher mV (20mV)
– Decreases pacemaker sensing ability
• High Sensitivity
– Maximal Sensitivity
– Toward Lower mV (1mV)
– Increases pacemaker sensing ability
Sensitivity Threshold
• Sensitivity Threshold– Set rate just below patient’s intrinsic rate,
the pacemaker should stop pacing and the sensor light should start flashing.
– Gradually decrease sensitivity until the pacemaker begins to fire. This is the sensitivity threshold.
– Set sensitivity to 2-3 times the sensitivity threshold.
Other Pacemaker Settings
• High/Low Rate– Sets the highest and lowest rates that the
pacemaker can pace.
• AV Interval Control – Set between 150-250 msec to preserve AV
synchrony.
• Refractory Periods – Sets a period of msec in which the pacemaker
will not respond after a sensed or paced beat.
Special Considerations
• Not necessary to disconnect temporary pacemaker during cardioversion or defibrillation
• Changes in pacing threshold may occur with electrolyte imbalance, myocardial ischemia, anemia, CHF, hypoxia, or a change in drug therapy
• Pacing threshold increases over time
Nursing Management• Continuous Cardiac Monitoring
– Watch for pacemaker complications– Always evaluate cardiac tracings
• Electrical Safety Measures– Use rubber gloves when handling external pacemaker
components– Regularly inspect all connections– Replace batteries as needed ( have spares available)– Avoid EMI sources
• Monitor Thresholds• Monitor All Imbalances
• Temperature, Acid-Base, Electrolytes
• Monitor Medication Effects• Watch For Pacemaker Syndrome
Nursing Assessment
• Assess the Patient– Vital Signs
– Cardiac Rhythm (always run strips)
– Patient Complaints
• Assess the Pacemaker– Maintain Electrical Safety
– Inspect and Secure Connections
– Assess the Effectiveness of the Pacemaker
– Check Thresholds
– Change Battery As Needed (keep spares available)
Pacemaker Evaluation
• What is the rhythm?
• Is the pacemaker functioning properly?
• What is the immediate nursing action?
Pacer Rhythm Evaluation What is the pacing Mode?
What is the underlying rhythm?
Is the pacer firing appropriately?
Is the capture appropriate?
Is the sensing appropriate?
What is the low rate limit?
What is the high rate limit?
What is the AV delay?
Nursing Management• Continuous Cardiac Monitoring
– Watch for pacemaker complications– Always evaluate cardiac tracings
• Electrical Safety Measures– Use rubber gloves when handling external pacemaker
components– Regularly inspect all connections– Replace batteries as needed ( have spares available)– Avoid EMI sources
• Monitor Thresholds• Monitor All Imbalances
• Temperature, Acid-Base, Electrolytes
• Monitor Medication Effects• Watch For Pacemaker Syndrome
PRACTICE
AAI
AAI: A pacemaker that stimulates and senses the atrium at one rate
VVI
VVI: A pacemaker that stimulates and senses the ventricle at one rate
DDI
DDI: A pacemaker that senses and stimulates both atrium and ventricle. The stimulation rate is fixed.
DDD
DDD or VVI?
DDDR
DDI
What’s the Mode?
What’s the Mode?
References• Guidelines for Pacemakers and Defibrillators Updated. Journal Watch General Medicine
1998: 2-2.
• Wood, D.A., Fox, K.F., Gibbs, S.R. (2001). Rapid cardiology--for chest pain, breathlessness and palpitations. QJ Med 94: 177-178.
• HAMMILL, S. C., HUBMAYR, R. D. (2000). The Rapidly Changing Management of Cardiac Arrhythmias. Am J Respir Crit Care Med 161: 1070-1073.
• Bauersfeld, U., Nowak, B., Molinari, L., Malm, T., Kampmann, C., Schonbeck, M. H., Schuller, H. (1999). Low-energy epicardial pacing in children: the benefit of autocapture. Ann. Thorac. Surg. 68: 1380-1383.
• KURBAAN, A S, SUTTON, R (1999). Pacing for vasovagal syncope. Heart 82: 649-650
• Braunwall, e. (1992) Heart Disease : a textbook of cardiovascular disease, 4 th edition, WB Saunders
• American Radio Relay League, Inc., Technical Information Service, 225 Main St., Newington, CT 06111 (860) 594-0214.
• Thelan, Lynne A., Davie, Joseph K., Urden, Linda D., Lough, Mary E. (1994) Critical Care Nursing: Diagnosis and Management. Second Edition. Pg 313-322.
• Graver, K (1998) A Practical Guide to EKG Interpretation, 2nd edition
• On-line: http://www.emedicine.com/emerg/topic699.htm