a tactile resonance sensor system for detection of ...698546/fulltext01.pdf · 4. anatomy,...
TRANSCRIPT
A Tactile Resonance Sensor System for Detection of Prostate Cancer ex vivo – Design and Evaluation on Tissue Models and Human Prostate
Anders P Åstrand
Department of Applied Physics and Electronics
Centre for Biomedical Engineering and Physics
Industrial Doctoral School
Umeå 2014
Department of Applied Physics and Electronics
Centre for Biomedical Engineering and Physics
Industrial Doctoral School
Umeå University
SE-901 87 Umeå, Sweden
This work is protected by Swedish copyright legislation (Act 1960:729)
© 2014 by Anders P Åstrand
ISBN: 978-91-7601-006-8
ISSN: 1653-6789:6
Electronic version http://umu.diva-portal.org/
Printed by: VMC-KBC
Umeå, Sweden 2014
Dedicated to my son Tobias
i
Table of Contents
Table of Contents i Abstract iii 1. Original papers v 2. Abbreviations vii 3. Introduction 1
3.1 General introduction 1 3.2 Aims 2
4. Anatomy, physiology and tactile resonance sensors 3 4.1 Prostate anatomy and pathology 3 4.2 Methods for diagnosis, in brief 4 4.3 Resonance sensors and theory 6 4.4 Tactile sensors in medical applications 8 4.4.1 Piezoelectric tactile resonance sensors 8 4.4.2 Other types of tactile sensors 8 4.5 Soft tissue silicone phantoms and biological tissue 9
5. Materials and methods 11 5.1 The piezoelectric resonance sensor 11 5.2 The measurement system 12 5.3 Technical setup 13 5.4 Preparation of soft tissue phantoms made of silicone 14 5.5 Preparation of biological tissue phantoms 15 5.6 Human prostate samples 16 5.7 Measurement set-up 17 5.7.1 Measurements on soft tissue silicone phantoms 18 5.7.2 Measurements on biological tissue phantoms 19 5.7.3 Measurements on excised human prostate tissue 20 5.8 Maximum estimated errors and overall accuracy 20 5.9 Statistics 21
6. Results and discussion 23 6.1 Instrument performance and accuracy 23 6.2 Angular dependency 24 6.3 Velocity dependency 26 6.4 Depth sensitivity 27 6.5 The clamping force 28 6.6 Measurements on prostate gland ex vivo 29
7. General Summary and Conclusion 33 7.1 Conclusion 33
8. Acknowledgements 35 9. References 37 Summary of papers 45
ii
iii
Abstract
Background
The most common form of cancer among males in Europe and the USA is
prostate cancer, PCa. Surgical removal of the prostate is the most common
form of curative treatment. PCa can be suspected by a blood test for a
specific prostate antigen, a PSA-test, and a digital rectal examination, DRE
where the physician palpates the prostate through the rectum. Stiff nodules
that can be detected during the DRE, and elevated levels of PSA are
indications for PCa, and a reason for further examination. Biopsies are taken
from the prostate by guidance of a transrectal ultrasound. Superficial cancer
tumours can indicate that the cancer has spread to other parts of the body.
Tactile resonance sensors can be used to detect areas of different stiffness in
soft tissue. Healthy prostate tissue is usually of different stiffness compared
to tissue with PCa.
Aim
The general aim of this doctoral thesis was to design and evaluate a flexible
tactile resonance sensor system (TRSS) for detection of cancer in soft human
tissue, specifically prostate cancer. The ability to detect cancer tumours
located under the surface was evaluated through measurements on tissue
phantoms such as silicone and biological tissues. Finally measurements on
resected whole prostate glands were made for the detection of cancer
tumours.
Methods
The sensor principle was based on an oscillating piezoelectric element that
was indented into the soft tissue. The measured parameters were the change
in resonance frequency, Δf, and the contact force F during indentation. From
these, a specific stiffness parameter fF was obtained. The overall
accuracy of the TRSS was obtained and the performance of the TRSS was
also evaluated on tissue models made of silicone, biological tissue and
resected whole human prostates in order to detect presence of PCa. Prostate
glands are generally spherical and a special rotatable sample holder was
included in the TRSS. Spherically shaped objects and uneven surfaces call for
special attention to the contact angle between the sensor-tip and the
measured surface, which has been evaluated. The indentation velocity and
iv
the depth sensitivity of the sensor were evaluated as well as the effect on the
measurements caused by the force with which spherical samples were held in
place in the sample holder. Measurements were made on silicone models
and biological tissue of chicken and pork muscles, with embedded stiff
silicone nodules, both on flat and spherical shaped samples. Finally,
measurements were made on two excised whole human prostates.
Results
A contact angle deviating ≤ 10° from the perpendicular of the surface of the
measured object was acceptable for reliable measurements of the stiffness
parameter. The sensor could detect stiff nodules ≤ 4 mm under the surface
with a small indentation depth of 0.4 to 0.8 mm.
Measurements on the surface of resected human prostate glands showed
that the TRSS could detect stiff areas (p < 0.05), which were confirmed by
histopathological evaluation to be cancer tumours on, and under the surface.
Conclusions
A flexible resonance sensor system was designed and evaluated on soft tissue
models as well as resected whole prostate glands. Evaluations on the tissue
models showed that the TRSS can detect stiffer volumes hidden below the
surface on both flat and spherical samples. The measurements on resected
human prostate glands showed that PCa could be detected both on and
under the surface of the gland. Thus the TRSS provides a promising
instrument aimed for stiffness measurements of soft human tissue that could
contribute to a future quantitative palpation method with the purpose of
diagnosing cancer.
v
1. Original papers
This doctoral thesis is based on the following papers, which are referred to
by Roman numerals. The papers are reprinted with permission from the
publisher when applicable
I. Åstrand AP, Jalkanen V, Andersson BM and Lindahl OA. A flexible
sensor system using resonance sensor technology for soft tissue
stiffness measurements – evaluation on silicone. Presented at the
15th Nordic Baltic Conference on Biomedical Engineering and
Medical Physics, Aalborg, Denmark, 16 June 2011. IFMBE
Proceedings 34 21-24. www.springerlink.com.
II. Åstrand AP, Jalkanen V, Andersson BM and Lindahl OA. Stiffness
measurements on spherical surfaces of prostate models using a
resonance sensor. Presented at the World Congress 2012 on Medical
Physics and Biomedical Engineering, Beijing, China, 30 May 2012.
IFMBE Proceedings 39 1401-1404. www.springerlink.com.
III. Åstrand AP, Jalkanen V, Andersson BM and Lindahl OA. A flexible
resonance sensor instrument for measurements of soft tissue –
evaluation on silicone models. Journal of Medical Engineering &
Technology 2012: 37 (3) 185-196.
DOI: 10.3109/03091902.2013.773097.
IV. Åstrand AP, Jalkanen V, Andersson BM and Lindahl OA. Detection
of stiff nodules embedded in soft tissue phantoms, mimicking cancer
tumours, using a tactile resonance sensor. Journal of Biomedical
Science and Engineering. Accepted 2014.
V. Åstrand AP, Andersson BM, Jalkanen V, Ljungberg B, Bergh A and
Lindahl OA. 2014: The first study on whole human prostate ex vivo
using a tactile resonance sensor for cancer detection. Manuscript.
vi
Peer-reviewed conference publications of relevance, but not
included in the thesis:
Åstrand AP, Jalkanen V, Lindahl OA and Andersson BM. Mätutrustning för
kontrollerad mätning och karatärisering av vävnad. Medicinteknik-dagarna i
Umeå, Sweden , 7 October 2010.
Åstrand AP, Jalkanen V, Lindahl OA and Andersson BM.
Ansättningsvinkelns betydelse vid mätning med piezoelektriska
resonanssensorer på vävnadsmodeller av silikon. Medicinteknikdagarna i
Linköping, Sweden, 12 October 2011.
Other publications:
Åstrand AP. A Flexible Resonans Sensor System for Detection of Cancer
Tissue – Evaluation on Silicone. Licentiate Thesis, 2012. ISBN 978-91-7459-
477-5
vii
2. Abbreviations
BPH Benign prostatic hyperplasia
CT Computer tomography
DRE Digital rectal examination
GS Gleason score
PCa Prostate cancer
MP Measurement position
MRI Magnetic resonance imaging
PLL Phase-locked loop
PSA Prostate specific antigen
PZT Lead zirconate titanate (referring to the piezoelectric element)
SD Standard deviation
TNM Tumour, Nodule, Metastasis
TRSS Tactile Resonance Sensor System
TRUS Transrectal ultrasound TURP Transurethral resection of the prostate
α Contact angle between sensor and object
αr Rotation angle
αsm Angle for sensor movement
d Distance to the surface above the centre of a stiff inclusion
D Diameter
viii
ε Error
Δf Frequency shift
f Loaded resonance frequency
F Contact force
FC Clamping force
Itot Total indentation - indetation depth during measurement
IZ Indentation - indetation depth of interest
n Number of measurements
vi Indentation velocity
fF Stiffness parameter
1
3. Introduction
3.1 General introduction
Research in new and improved methods for early detection of prostate
cancer (PCa) is necessary as PCa is the most common form of cancer among
males in Europe and in the USA. There were about 417000 new cases of PCa
diagnosed in Europe in 2012, and nearly 92000 deaths were reported [1].
The American Cancer Society estimates that more than 238000 new cases
will be diagnosed with PCa in 2013 and that close to 30000 men will die
from PCa [2]. The latest report from the Swedish National Board of Health
(Socialstyrelsen) states that PCa stood for 32.2% of the new cases of males
diagnosed with cancer in 2011, reporting 9663 new cases and 2382 deaths
due to PCa [3].
Common diagnostic methods for detecting PCA are digital rectal
examination (DRE), where the physician palpates the prostate from the
rectum and a blood test for a prostate specific antigen (PSA) [3]. However,
an elevated PSA-level can be due to other causes than PCa and the use of this
alone can lead to over-diagnosis [4]. A transrectal ultrasound (TRUS) guided
systematic needle biopsy is necessary for final diagnosing [5]. The outcomes
of these tests are important for decisions regarding what treatments are
thought to give the best results in each case.
When performing different kinds of open surgery, palpation of areas with
suspected cancer can be used for assessing areas of stiffness and other
abnormalities of the tissue. Earlier studies have reported that tumours in
soft tissue usually are stiffer compared to healthy tissue [6, 7, 8, 9]. An
artificial tactile sensor based on a piezoelectric resonance sensor can be of
aid for the physician when performing the otherwise general palpation of the
tissue [10]. Palpation of soft tissue has been performed during different
thoracotomies, where undetected malignant pulmonary nodules were found
during lobectomies, metastatic tumours has been found on patients with
colorectal cancer, tumours that had not been detected with prior helical
computed tomographies. [11, 12, 13, 14] For PCa diagnosis, the results of the
DRE are subjective and dependant on the physicians’ experience [15]. In a
study where palpation for detecting texture abnormalities in thoracic
paraspinal region were evaluated, they concluded that palpation to assess
2
differences in texture was complex and subjective depending on the
examiners’ experience [16]. For this reason, an objective method and
quantitative parameter related to the prostate tissue might be useful to
physicians.
The tactile resonance sensor system designed and evaluated in this
study, as well as in previous studies [III, IV] provides an instrument aimed
for stiffness measurements on resected whole human prostates.
3.2 Aims
The overall aim was to develop a sensor system for possible detection of
cancer in soft human tissue, specifically prostate cancer.
- To design a flexible resonance sensor system with the possibility to
measure on flat and spherical objects.
- To determine the performance of the system and maximum errors.
- To evaluate how the system parameters are affected by variations in
contact angle and indentation velocity, using flat soft tissue silicone
phantoms.
- To determine the depth sensitivity of the resonance sensor system,
using flat soft tissue silicone phantoms and biological tissue with
embedded harder silicone inclusions.
- To evaluate the functionality of the flexible resonance sensor system
on spherical samples of soft tissue silicone phantoms and biological
tissue with embedded stiff silicone inclusions.
- To perform measurements on whole human prostates ex vivo in
order to find stiffer areas that can indicate the presence of possible
cancer tumours and evaluate the findings against the results from
the histopathological analysis.
3
4. Anatomy, physiology and tactile resonance sensors
4.1 Prostate anatomy and pathology
The prostate is a gland, about 30 to 40 mm in diameter, which is located
below the urinary bladder, figure 1. The prostate encircles a part of the
urethra as it leaves the urinary bladder. The prostate gland consists of
glandular tissue and fibromuscular stroma which is surrounded by and
wrapped in a thick blanket of smooth muscle fibres called the capsule [17,
18].
Figure 1. Illustration of the prostate anatomy. Modified from Wikipedia.
(http://en.wikipedia.org/wiki/prostate).
4
The prostate produces prostatic fluid, a secretion that stands for about 25 %
of the volume of semen and promotes the mobility and viability of the
sperms [18]. The prostate grows until the age of 30, after which it shows a
decrease in the growth rate. After the age of 45-50 it is common that
enlargements start to occur, so called benign prostatic hyperplasia (BPH).
An enlarged prostate can cause pain and problems to urinate, as the swelling
of the prostate constricts the urethra [17]. The condition can be corrected
with a transurethral resection of the prostate (TURP) where some of
constricting parts of the prostate tissue is surgically removed trough the
urethra.
Prostate cancer (PCa) is an adenocarcinoma, a malignant and
metastasizing cancer that often originates from one of the secretory glands.
To metastize, the malignant cancer cells must migrate from the primary
tumour, penetrate and circulate through the blood stream to reach target
tissues such as bone marrow, the lymphatic system and lungs [17, 19]. For
this reason it is important to diagnose PCa at an early stage.
4.2 Methods for diagnosis, in brief
The most common screening methods for PCa are a blood test for a prostate
specific antigen (PSA) and a digital rectal examination (DRE), where the
physician palpates the prostate through the rectum. The aim of the palpation
through the rectal wall is to detect stiff areas or nodules on the prostate, as it
has been shown that cancer tumours in general are stiffer than the
surrounding healthy tissue [7, 8, 9, 20, 21]. In the case that the result from
the DRE is positive for PCa, transrectal ultrasound (TRUS) guided biopsies
are used for the final diagnosis [22]. With TRUS it is possible to determine
the volume of the prostate to determine the PSA density, but the main use
for this method is to guide the needles used for biopsies in different areas of
the prostate. Biopsies are usually obtained from six areas; left and right of
the apex, the midgland and the base, a total of 10 to 12 cores are common
practice [23]. However, for the early stages of PCa when the tumours might
be small, the needles can miss existing nodules which results in false-
negative determinations [24, 25]. The imaging method could be improved by
colour Doppler technology and a contrast enhanced technology with micro
bubbles, which increases the rate of prostate cancer detection [23, 25, 26].
Computer tomography (CT) produces images by the use of X-rays to
create images. CT is not considered an effective method for imaging and
diagnosing PCa as it lacks when it comes to distinguishing the prostate
anatomy from, for example nearby muscle tissue, the bladder wall etc. which
5
makes it difficult to detect cancers within the prostate gland. However, CT
can be useful as an imaging method for high risk patients’ with advanced
local disease [23, 26]. As prostate cancer can metastasize to the bone CT has
been used to monitor bone metastases, but for this purpose Magnetic
resonance imaging (MRI) is a better alternative [23, 27].
MRI can be used for visualizing cancer and determine its volume, extent
and location. MRI uses a strong magnetic field and by analysing the
resonance signal that is generated by the hydrogen nuclei and the protons in
the tissue when the magnetic field is disturbed by pulsing radio waves,
images can be created. MRI have an important role for patients where there
is a discrepancy between PSA-tests and the result from the biopsies taken
with the aid of TRUS, as such discrepancy might indicate missed cancer
tumours [28]. MRI is considered as a good method for imaging the prostate
anatomy and locations of PCa. The MRI has a high resolution for soft tissue
imaging, and can also be used for staging of the PCa [23, 26]. The method
should be avoided earlier than 4 to 8 weeks after biopsies [29, 30] as
haemorrhages, which are common after biopsies, can mimic cancer and
therefore limit the detection of actual PCa [26].
Before any decisions regarding the choice of treatment, the severity of the
PCa is classified by a classification system called the Gleason Score (GS), a
scale ranging from 1 to 5 with under groups. The two most descripting scores
are added which results in a GS with a maximum of 10, which would indicate
a severe condition with a bad prognosis [31, 32]. After surgery or a radical
prostatectomy, the tumour and possible tumour spreading, are classified by
a pathological classification system; tumour, nodule, metastasis (TNM)
staging system [33].
The different options for treatment vary depending on the medical
condition and age of the patient, as well as the grading and the stage of the
cancer. About half of the patients diagnosed with PCa are treated by radical
prostatectomy or radiation therapy which can be administered as external
beam radiation och by implants of radioactive seeds into the prostate
(brachytherapy). Hormonal treatment may be given before or after surgery.
All of these treatments may have an effect on the quality of life for the
patient, as side effects such as urinary and erectile difficulties are common
complications. For some patients with an early diagnose and a slow growing
PCa, an alternative is active monitoring with planned regular PSA-tests and
biopsies. Hormonal treatment is also an alternative for early detected PCa,
which may be given in order to temporarily stop or slow down the growth of
a tumour [2, 3].
6
4.3 Resonance sensors and theory
The theory behind the characteristics of resonance sensors, like the one used
in this study, can approximately be described by a model of vibrations in a
finite rod which has been reported by Omata et al [34] and Jalkanen et al
[35]. The rod is set to vibrate with its resonance frequency, f0, and when the
tip of the sensor comes into contact with an object, a change in the resonance
frequency occurs which can be described with
0
0
2 lZ
Vf x
1
where V0 stands for the wave velocity in the rod, βx is the reactance part of
the impedance, l is the length of the rod and Z0 is the acoustic impedance of
the rod. The impedance of the object with which the rod comes into contact
can be written
xxx iZ 2
where αx is the resistance. The reactance part of the impedance can be
describes as
x
xx
km 3
where mx is the inertia term, ω is the angular frequency and kx is the term for
the hardness (stiffness) of the surface. To calculate mx and kx, for these case
of a spherical tip, also used in this study, the following equations can be used
23
23
11
1
4S
v
amx
4
21
221 1
2S
v
Ek x
5
7
where the Poisson’s ratio is denoted ν (for tissue samples ν = 0.495 is
commonly used [36]) and a11 is a coefficient that depends on Poisson’s ratio
[37]. ρ is the density of the contacted object and S is the surface of the
contact area. E is the elastic modulus (Young’s modulus). The theoretical
model of the finite rod was studied by Jalkanen et al [35] with a Venustron®
resonance sensor system and it was found that as mx >> (kx/ω), the term for
the surface stiffness, can be omitted. Thus with equations 1 to 4 it is possible
to write
23Sf 6
which is valid if Poisson’s ratio is assumed to be constant, and the specific
rod is oscillating at a constant frequency. Since the contact force F between
the sensor and the contacted object can be written as [35].
23ESF 7
Eq / gives that the surface contact area S is depending on F, and by using this
expression in equation 6 the following is valid
E
Ff
8
By introducing a stiffness sensitive parameter, fF , equation 8 can be
written as
E
f
F
9
The stiffness parameter fF was derived from the measured Δf and F in
the work by Jalkanen et al [35], where they verified a theoretical model in
measurements on human prostate tissue. The stiffness parameter was found
valid, as the variations in density were small and mostly non-significant. In
this work, the absolute value, fF , of the stiffness parameter was used.
8
4.4 Tactile sensors in medical applications
4.4.1 Piezoelectric tactile resonance sensors
Resonance measurement systems are based on the principle of a
piezoelectric element oscillating at its resonance frequency, and the change
in the frequency that occurs when the element comes into contact with an
object. Piezoelectric resonance sensors have been used as tactile sensors in
different fields of medical research [6, 34]. Measurements with resonance
sensors have been made to study the differences in stiffness and elasticity of
the skin in order to detect oedema and lesions [38, 39]. In other studies,
measurements have been made to detect lymph nodes containing metastases
[40] and stiffness of the liver that can indicate liver fibrosis [41]. Jalkanen et
al. [9, 42], has reported measurements on slices of resected prostate tissue ex
vivo, which demonstrates the possibility to differentiate between healthy
tissue and malignant tumours. A piezoelectric resonance sensor in an
applanation tonometer has also been used by Hallberg et al. [43] and Eklund
et al. [44], to measure the intraocular pressure of the eye, as an elevated eye
pressure can be an indicator for glaucoma.
4.4.2 Other types of tactile sensors
There are many studies reporting of different experimental devices for
measuring stiffness in soft tissue. Force measurements in order to detect stiff
nodules during large deformations have been reported by Kim et al [45],
showing that they could detect cancer lesions in human prostates at a depth
of 3.4 mm using an 3 mm indentation depth. They have also reported on
measurements for characterization of tissue stiffness using large
deformations on porcine liver [46]. Rolling indenter probes measuring
indentation forces to detect inclusions of stiff nodules have also been used on
silicone phantoms and porcine kidney tissue [47, 48]. Their results showed
that they could detect stiff nodules at a depth of 22 mm with an indentation
depth of 6 mm. One interesting type of tactile sensing system for very soft
human tissue (brain) has been reported by Tanaka et al [49]. The sensor had
a tip made of a latex balloon with 4 mm in diameter that could expand. The
balloon was connected via a surgical line through an endoscope to a syringe
with which the pressure was controlled by a water column and a pressure
sensor.
During minimal invasive surgery, MIS, different solutions to enhance the
tactile feedback to the surgeon have been reported. One common type of
9
technical solution seems to be force sensors added to graspers and other
instruments used during laparoscopic, and robot assisted surgery. Samur et
al [50] developed a robotic indenter to measure soft tissue in the abdominal
region during MIS. They report on successful measurements made on adult
pigs in vivo. Similar instruments were evaluated on bovine liver and porcine
lungs ex vivo with implanted nodules, and it was suggested that such
instruments could give the surgeons the haptic feeling that is often lost
during MIS [51]. A very different approach to the issue of lost haptic
information during MIS was reported by Beccani et al [52]. Their solution
was a wireless tactile sensor to be used to palpate for lumps in soft tissue
during laparoscopic surgery. Their prototype was 60 mm long with a
diameter of 15 mm and was held by a standard laparoscopic grasper,
operated by the surgeon. With this wireless device they could localize
tumours and distribute a stiffness map in real time.
4.5 Soft tissue silicone phantoms and biological tissue
Silicones of different stiffness have been found to be very suitable for
evaluating tactile sensor techniques regarding soft human tissue
characterization as soft tissue phantoms for preliminary research due to the
similarities regarding mechanical properties [42, 53, 55, I]. Silicone is an
elastomer which can be regarded as incompressible with a mechanical
strain-stress relationship that can be described using hyperelastic models
[55, 56]. The silicone can be cast in different geometries and it is possible to
embed small volumes of different stiffness to mimic stiffness variations due
to disease, i.e. cancer nodules. Silicones are also more stable over time,
compared to biological materials which make them appropriate to store for
repeated measurements at a later date.
Biological tissue is more difficult to use as models as its properties can
show changes due to environmental conditions, such as degrading or
dehydrating during the time of measurements. Its normal properties are
different when measurements are made ex vivo, compared to in vivo, due to
the physiological differences and the change in supporting tissue structures.
Indentation measurements on human liver show a difference in elastic
properties between normal and diseased tissue [57]. Mazza et al [58] have
reported of increased stiffness in human liver ex vivo, compared to in vivo.
The fact that ex vivo measurements often are made in laboratory
environment with well-defined boundaries and conditions, can also be an
advantage for the outcome of the measurements.
10
In contrast to silicone, biological tissue does not have a linear stress-strain
relationship [58]. Most biological soft tissues can be described as
hyperelastic, but they also have viscoelastic properties when they are subject
to deformation [36, 56]. Viscoelasticity is a combination of elasticity and
viscosity, and the ratio between these two can vary depending on the type of
biological tissue [59]. Viscosity is the resistance that a certain material has
against changing when subjected to a force as it resists shear and strain
when stress is applied. Elasticity is the property of a material to temporarily
change shape during the time when it is subjected to a force and retract
when the force is unloaded [60]. There are other properties that can
characterise viscoelastic materials, such as hysteresis which is the difference
in behaviour of loading and unloading when subjected to a force. The
mechanical behaviour of healthy and malignant prostate tissue has been
described with viscoelastic models by Phipps el al. [61].
11
5. Materials and methods
5.1 The piezoelectric resonance sensor
The resonance sensor used in this measurement system consists of a
piezoelectric element. A piezoelectric element changes its shape when an
electric field is applied. If this electric field is of a sinusoidal type, the
piezoelectric element will start to oscillate. The resonance sensor used was
divided into two different piezoelectric elements, a driving element and a
pick-up, which is separated by a non-conducting section, figure 2. The signal
from the pick-up section was constantly transferred to a feedback circuit
where a phase-shift circuit ensured that a zero phase condition between the
pick-up and the drive signal was established. This kept the resonance sensor
oscillating at its resonance frequency.
fo f
Healthy tissueCancerous tissue
The frequency shift f
f = f - f0
Pickup element
Driving element
Phase shift
circuit
PEEK tip
Non conducting
section
Force sensor
Sensor movement
Figure 2. A schematic illustration of the piezoelectric resonance sensor in a
measurement situation.
12
When the driving section of the resonance sensor comes into contact with an
object, a change in frequency will occur, f = f - f0 where f0 is the free
(unloaded) resonance frequency of the piezoelectric element, and f is the
loaded resonance frequency when in contact with the object. Contact with a
soft object that can be deformed will result in a f with a negative sign, and a
positive sign will be the result from contact with a hard object.
5.2 The measurement system
The tactile resonance sensor system (TRSS) was developed and used for
these studies. The TRSS was equipped with three motorized translation
stages that were controlled by a computer and a LabView® (National
Instruments, Austin, Texas, USA) program. Two translation stages were used
for horizontal movements, X and Y. A third translation stage for vertical
movements, Z, was mounted on a rotational stage to enable movements at
different contact angles, α, between the sensor and the measured object [38,
III], figure 3.
Stepper
motorHorizontal translation
stages, X and Y
Platform with flat
sample
Stepper motor
Vertical translation
stage, Z
Sensor head
Rotational stage
Electronic circuits, I/O
connector block, etc.
Computer with DAQ
Figure 3. A schematic side view of the basic components of the TRSS in position
for measurements on a flat sample. The movements of the translation stages are
controlled from a desk-top computer.
13
5.3 Technical setup
The three translations stages for the TRSS enabled 3-D movement’s
necessary for the flexibility of the instrument. The translation stages for
horizontal movements (M4424; Parker Hannifin Corporation, Daedal
Division, Irwin, PA, USA) were equipped (in house) with stepper motors
(17HD1008; Moon’s, Shanghai, P.R. China) and connected to the
micrometre screws of the translation stages. The motorized translation stage
(VT-21, Micos, Irvine, CA, USA) for the vertical movements was controlled
by a stage controller (Pollux Box; Micos GmbH, Eschbach, Germany). All
three motorized translation stages had a resolution of 2.5 μm. The rotational
translation stage (NT55-030; Edmund Optics, York UK) with a resolution of
0.5° made it possible to adjust the angle of the movement of the vertical
translation stage ± 90°.
The sensor in the TRSS used for measuring f was a piezoelectric
cylindrical element (Morgan Electro Ceramics, Bedford, Ohio, USA) made of
lead zirconate titanate (PZT). The element was 15 mm long with an outer
diameter of 5 mm and an inner diameter of 3 mm. A hemispherical tip with
the radius of 2.5 mm made of polyether-ether-ketone (PEEK) was glued at
the end of the sensor which came into contact with the measured object. The
PZT-element and a preloaded force sensor (PS-05KC; Kyowa, Tokyo, Japan)
were mounted inside a cylindrical aluminium casing, which will be referred
to as the sensor head, figure 4. The force sensor and the PZT-element were
calibrated before each study to get valid constants for the voltage-to-force,
and voltage-to-frequency conversions. The signals from the force sensor and
the PZT-element were collected by a data acquisition card (NI6036E;
National Instruments, Austin, Texas, USA), and transferred to a computer at
a sampling rate of 1 kHz.
The TRSS can be used for measurements on both flat and spherical
objects, which can be rotated in a special holder. Due to the, often uneven,
and spherical shape of the prostate and the importance of making
measurements perpendicular towards the surface, the sensor was fitted to a
rotational device which allows to be varied. The rotatable holder, figure 4A,
for spherical objects during measurements was equipped with a cantilever
strain gauge (CEA-06-240UZ-120; Measurements Group Inc, Raleigh, NC,
USA) for measuring the clamping force FC and a sensor for recording the
rotation angle, αr [III].
14
1
4
2
3
7
8
9
10A B5 6
Figure 4. The TRSS A) The sensor head B): 1) The vertical and the rotational
stages. 2) Sensor head. 3) Fixture for spherical samples. 4) Rotation angle sensor. 5)
Cantilever strain gauge. 6) Translation stages for horizontal movements. 7) The
platform for flat samples. 8) Force sensor inside. 9) PZT-element. 10) PEEK tip.
5.4 Preparation of soft tissue phantoms made of silicone
For the studies on soft tissue phantoms, different hardness and shapes were
used in [I – V]. The silicone used was a two-component type (Wacker SilGel
612; Wacker-Chemie GmbH, Germany), which has been used in earlier
studies [37, 43, 55, 62, III].
Table 1. The mixing ratios and hardness values for the different silicone mixes used
in papers I – V. The Shore-000 scale was used only in papers IV and V.
Hardness Mixing ratio A:B
Shore-000 value
Cone penetration value (mm x 10-1)
Used in paper
Softest 4:3.75 16 251 I 4:3.50 23,5 221 I 4:3.30 33 188 IV, V 4:3.20 39 163 I, II, III 4:3.00 51 126 I 4: 2.50 82 49 I, II, III
Hardest 4: 2.22 88 33 IV, V
15
All flat silicone samples were cast in standard Petri dishes (diameter 87
mm and height 13 mm). By letting the silicone cure on a horizontal surface
and in room temperature, the surfaces came out even and glossy. For the
purpose of evaluating Δf at different hardness, five homogeneous samples
with the height 13 mm were used.
To study at what depth into soft silicone the TRSS could detect an
embedded stiffer nodule, silicone samples with different height were cast, all
with a stiff hemisphere of silicone (diameter 8 mm, height 4 mm), and a
nodule centred at the bottom of the Petri dish, figure 5A. The same
combination of soft silicone, and hard silicone for the nodules were used in
both paper IV and V, see table 1. The distance from the top of the nodule to
the surface of the silicone, d, varied from 1 to 9 mm between the different
samples.
The TRSS had a rotatable holder to facilitate measurements on spherical
objects. The possibility to perform reliable measurements on spherical
shaped samples needed to be evaluated, as well as how nodules could be
detected under a curved surface. For this purpose soft spherical silicone
samples were cast, using the inside of a table tennis ball (diameter 40 mm)
as a mould. Small spheres of stiff silicone with different diameters (2.5, 4
and 6 mm) were placed inside the mould during the casting procedure,
figure 5B. The embedded small stiff spheres could then mimic cancer
tumours.
d
A B
d
Figure 5. Tissue models made of soft silicone with embedded stiff nodules
mimicking cancer tumours at the distance d from the surface. A) Flat sample in a
Petri dish. B) Spherical sample with nodules.
5.5 Preparation of biological tissue phantoms
The biological tissue phantoms for the flat samples were made of chicken
muscle tissue [IV]. Commercially available chicken breasts were sliced into
different thickness, 5 to 11 mm. Silicone nodules of the same type as for the
16
silicone samples mimicked cancer tumours were used. An incision of the
same size as a nodule was made on the bottom of each slice of chicken
muscle, figure 6A, to make room for the nodule without causing a bulge on
the surface of the slice. Finally the slice was pinned on a polystyrene foam
plate. The nodules were placed with d= 1 to 7 mm under the surface of the
slices.
For the spherical tissue phantoms, pieces of porcine tenderloin were used
instead of chicken muscles because of the volume needed to get the right size
of spherical shaped samples [V]. The pieces were cut approximately into a
spherical shape with a diameter D = 40 mm. All visible membranes were
carefully removed, and pieces with visible tendons were avoided to get as
homogeneous tissue samples as possible. A spherical nodule with the
diameter 6 mm of stiff silicone was inserted under the surface of the sample;
this was done from the side, not to disturb the measurement area, figure 6B.
The nodule was placed at d = 3 ± 0.5 mm under the surface.
d
B
d
A
Figure 6. Cross-section side views of two biological tissue samples with embedded
silicone nodules at the distance d from the surface. A) A flat sample made of
chicken muscle tissue. B) A spherical sample made of porcine muscle tissue.
5.6 Human prostate samples
For the studies in paper V, two prostate glands were obtained with written
consent from the patients, aged 70 and 72 years old. Both were undergoing a
radical prostatectomy at the University Hospital in Umeå. Ethical approval
was given by the Ethics Committee at Umeå University (Dnr 03-423). The
younger man’s prostate gland weighed 60.1 g and had the approximate
diameter D = 50 mm. The other prostate gland weighed 40.6 g and had the
approximate diameter D = 40 mm. Before measurements, the pathologist
dyed the surface of the prostate gland according to their standard, figure 7.
The dorsal aspects (backside) of the prostates were marked with yellow
17
colour, the left anterior (front) red and the right anterior sides were marked
green.
A B C
Figure 7. A resected prostate from a radical prostatectomy. A) Fresh from surgery.
B) The dorsal side of the prostate gland, dyed yellow. The apex is pointing down. C)
The anterior side of the prostate gland seen from the front, dyed green and red.
5.7 Measurement set-up
The setup of the TRSS used in the studies referred to in this thesis has been
presented recently [III - V, 63]. The measured parameters were the applied
force, F, when the sensor was indented into the measured object and the
change in the resonance frequency, f. Jalkanen et al. [35], has shown a
theoretical model that relates the measured stiffness trough F and f to the
Young’s elastic stiffness modulus, as described in 4.3. For evaluating the
stiffness measurements, a parameter called the stiffness parameter fF
was introduced. This is found as an important parameter as it is a function of
the indentation depths, IZ. In the studies presented here, the TRSS has been
evaluated regarding how the measured parameters F, f and fF are
affected by variations in contact angles, and indentation velocities, vi.
Each measurement lasted four seconds, during which the sensor was at a
standstill for two seconds after the pre-set indentation depth, Itot, was
reached, before retracting from the sample. With the pre-set indentation
velocity, i, and the sampling rate of 1 kHz, any indentation depth, IZ, could
be calculated, and hence data of interest could be retrieved from the logged
data at any depth IZ ≤ Itot. The stiffness parameter, fF , was derived by
linear regression for each IZ. All measurements were made at normal room
temperatures, 20 to 22 °C.
For the measurements made on the different spherical samples, the
necessary force, the clamping force FC, used to hold the sample in place
18
between the two concave discs, see figure 4A, were logged as well as the
rotational angle αr. To prevent the biological samples, that were wet and
slippery, from gliding in the rotatable sample holder, pieces of emery cloth
(P80, KK114F, VSM® Abrasives Limited, GB-Milton Keynes, UK ) were
glued to the inside of the concave discs.
5.7.1 Measurements on soft tissue silicone phantoms
The flat silicone samples were mainly used to evaluate the basic functions of
the TRSS, such as the ability to differentiate between silicones of different
hardness. The dependencies of the measured parameters on α and i were
studied using flat silicone samples with two different hardness, see table 1,
and with the height 13 mm. The angle of the sensor movement, α, was
adjusted with the rotationa stage. The deviation in α from the perpendicular
(α = 0°) towards the surface were chosen in steps of 1° from α = 0° to 11°,
and thereafter in steps of 2° up to α = 35°. Five different i were used (i = 1
to 5 mm/s) for that part of the study and Itot was set t0 1 mm [III]. For
papers [IV, V], the indentation velocity was i = 4 mm/s for all
measurements.
The nodules in the spherical silicone models were place under and near
the surface. They were tinted to be visible in the otherwise transparent
silicone. They were aimed as phantoms to mimic a prostate gland with
cancer tumours. The measurements on the spheres were made in different
measurement positions (MPs) with the sensor head movement in different
angles, αsm, but always perpendicular to the surface at the point of contact,
figure 8.
19
0°-10°-20°
10° 20°
O - 360°
αr
αsm
1
2
3
4αr
αsm
A B
Figure 8. A) Illustration of a silicone sample in the rotatable holder, illustrating αsm
and αr. B) A picture of a porcine sample in the rotatable holder. 1) Rotational stage
for αsm. 2) USB-microscope. 3) Cantilever strain gauge. 4) Angle sensor for αr.
The MPs were chosen directly on the nodules, αsm = 0° and at the positions
αsm = 10° and αsm = 20° axially to the left and right of the vertical. The
difference in angle (10°) between each MP was equivalent to a distance of 3.5
mm on the surface of the sphere [V]. For every row of measurement points in
the axial direction, the sphere was rotated approximately αr = 60°, but
making sure to find the correct position for the nodules that was placed at
approximately every αr = 120°. For the measurements on all silicone
samples, there were six measurements made at every MP. The clamping
force, FC, for the silicone spheres were approximately 500 ± 50 mN during
the stiffness measurements.
5.7.2 Measurements on biological tissue phantoms
The biological tissue models of chicken breast muscle were cut in a semi
defrosted state with an electric slicing machine to get even surfaces.
Measurements were made with a vertical sensor movement, αsm = 0°,: one
MP aimed over the nodule wiht another two to the left and two to the right of
the nodule with a distance of 5 mm in between them [IV].
Spherical tissue models were made of porcine muscle from tenderloin.
The measurements on the porcine tissue were made at αsm = 0°, and the
20
tissue was rotated in steps of αr = 5° within an interval of ± 40° from the
nodule. The distance between each MP was approximately 1.75 mm on the
surface of the samples. For all measurements on the biological tissue models,
there was only one measurement made in each point, i.e. several samples of
both chicken and porcine muscles were used [V]. The spherical shaped
porcine samples were subjected to FC = 425 ± 39 mN.The tissue samples
were kept moist by spraying them with saline solution (Sodium Chloride 9
mg/ml, Fresenius Kabi AG, DE-61346 Bad Homburg, Germany) about every
5 minutes, using an atomizer.
5.7.3 Measurements on excised human prostate tissue
For the measurements on the resected prostate glands, the rotatable holder
for spherical objects was used. The two prostates were different in weight
and size, resulting in a higher FC for Prostate 1 (FC = 760 ± 99 mN),
compared to Prostate 2 which was held in place by FC = 460 ± 85 mN. The
measurements were made with a sensor movement in the vertical direction,
αsm = 0°. A total of 36 to 40 measurement points were made as the prostate
specimens were rotated in steps of 10° between each measurement, i.e. only
one measurement in every MP. By doing so, all MP were in a straight line
around the circumference of the prostate gland. Points along this line were
marked with black dye. The reason for this was to guide the pathologist when
cutting the slices and to ensure that the correct slice was used for
comparison with the measurements. The prostate was cut in the horizontal
plane in o.5 cm thick slices and were dehydrated before being embedded in
paraffin. The slices were then cut in thin slices (5 μm) before they were
stained with haematoxylin-eosin and examined by light microscope.
5.8 Maximum estimated errors and overall accuracy
The estimated errors, ε, for the measurement parameters in each study has
been reported [III, IV, V], taking into consideration the specific settings for
each measurement situation. For the indentation depth, the IZ varies
between the different papers, but a general ε for the different IZ was 2 – 4%.
The errors for Δf, F and fF were all < 1.5%. The strain gauge for the
clamping force had an estimated error of 10% due to mechanical friction.
The rotational stage for measuring αsm and the angle sensor for αr both had a
ε = 0.5°. For measurements made on uneven surfaces, for example spherical
21
biological tissue samples and the prostate gland, photographs were taken
during the contact phase which was later studied for incorrect α.
5.9 Statistics
In paper III, the statistical tests were done with a one-tailed t-test to test if
the measured parameters F, Δf and fF deviated from their respective
values at α = 0° with the null hypothesis that the average relative deviation
with respect to the measurements with α = 0° was equal to 10% against the
alternative hypothesis that it was ≤ 10%. Wilcoxon’s signed rank test was
used for the silicone discs in paper IV, and a non-parametric Kruskal-Wallis
method followed by a multiple comparison test for the differences between
the different slices of chicken muscle. In paper V, the statistical tests for the
measurements on silicone were performed with one-way ANOVA and
Tamhane’s post hoc comparison test. The differences between groups of data
from the prostate measurements were done with a Mann – Whitney –
Wilcoxon test.
The software used for the tests were IBM SPSS Statistics 21, and
Microsoft Excel 2010. The measured values were presented as mean ±
standard deviation (SD) and p-values of ≤ 0.05 were considered significant
for all statistical tests.
22
23
6. Results and discussion
The TRSS has been evaluated systematically step-by-step with good results
regarding performance and the different measurement parameters.
However, it must be clarified that the absolute values of the measurements
most likely is unique for this TRSS. The size and geometry of the sensor tip,
the free (unloaded) resonance frequency and the size and type of the PZT-
element are only a few important components that interact uniquely for this
TRSS.
The stiffness measurements on the chicken and porcine muscle tissues,
[IV, V] were only performed once in each MP, as soft biological tissue ex vivo
can be permanently deformed after indentations, and repeated
measurements in the same MP would be affected by this [64]. During the
first measurements on the flat silicone models in Petri dishes, it was noticed
that there was a 0.7 mm protruding rim on the underside. That resulted in
that the bottom could give after during indentation due to the applied force
F. The rim had to be removed to assure full contact between the Petri dish
and the measurement platform. The conditions surrounding the nodules in
the flat silicone models and the spherical samples were different. In fle flat
models, the nodules were supported from underneath, as the y were placed
on the rigid measurement platform. The nodules in the spherical silicone
samples were surrounded by soft silicone in all directions.
6.1 Instrument performance and accuracy
The TRSS was evaluated on five flat silicone samples regarding the ability to
distinguish between different hardness, see table 1 [I]. Figure 9 shows the
linear relationship between Δf and the cone penetration values which is a
method for measuring hardness in soft materials according to DIN ISO
standard 2137.
24
Figure 9. Measurements of Δf (mean ± SD, n = 10) at constant F = 40 mN on five
silicone samples with different hardness. For most Δf, the standard deviations are too
small to be visible as error bars.
The results from figure 8 show a linear correlation between Δf and the cone
penetration values, (p < 0.05, R2 = 0.99). The standard deviations are small,
about 1 to 3 % of the measured values. The higher values for the softer
silicone samples are in agreement with previous studies on the same type of
silicone [42]. The results show that the TRSS could differentiate between
materials with different hardness, which was important for the stiffness
measurements that lay ahead.
6.2 Angular dependency
Two silicone samples with different hardness, see table 1, and two different
indetation depths were used for evaluating the angular dependency of Δf, F
and fF on different α [III]. The dependency of F on α showed a weak
dependency, with a relative deviation due to α, figure 10. It was on average ≤
10% for α up to about 10° (one-tailed t-test, p < 0.05). However, the stiffer
silicone showed stable values of F with respect to α up to about 20° [III]. The
25
dependency of Δf on α showed to be very similar to the dependency of
fF on α, figure 10, with a relatve deviation due to α that on average
was ≤ 10 % for angles up to 10° (one-tailed t-test, p < 0.05). For all other α,
the angular dependency of fF were significant (p > 0.05) [III].
Figure 10. Measurements made on silicone samples with two different hardness at
IZ = 0.20 mm and 0.38 mm for α between 0° and 35°. The figure shows the relative
deviation of fF to fF (α=0°), (mean ± SD, n = 6). The vertical line was
drawn as a guide for the eye.
Measurements on both flat and spherical shaped tissue samples were
especially prone to get the contact angle α ≠ 0° due to the uneven surfaces.
By studying the photographs that were taken at the time of each
measurement, such deviations were possible to identify. Other studies [10,
65] support the findings in paper [III] that the precision in Δf decreases as α
increase, due to the change in contact area between the sensor and the
measured sample. For variations in α ≤ 10°, the measured Δf and F and thus
the fF were not significantly affected [III].
26
6.3 Velocity dependency
The velocity, with which the sensor moved during indentation, could be set
through the LabView® program. In the first papers [I, II], the standard
setting of i, = 5 mm/s was used. In paper III the possible effect that
different i could have on the measured parameters Δf, F and fF were
evaluated. The evaluation of fF as a function of i was made on flat
silicone samples, two different hardness, and α = 0°, figure 11.
Figure 11. Measurements of fF (mean ± SD, n = 10) on flat silicone samples
with two different hardnessfor different i, and at two different IZ.
The results showed that there was a dependency of fF on i, and that it
was increasing with higher i. The relative difference in fF for the
increase in i was more evident for the softer silicone sample compared to
the stiffer. The evaluation also showed results that indicated a lower
dependency for F on i, compared to the dependency for Δf. However, a low
i showed higher accuracy in Itot and Iz since the first sampled data point
showing contact with the surface was more distinct. For the measurements
in [IV, V] the indentation velocity used were chosen to i = 4 mm/s. The
reason was that the silicone samples used were of the softer kind to better
27
mimic biological tissue, and at the same time minimize some of the viscous
effects in the ex vivo biological tissues.
6.4 Depth sensitivity
The ability of the TRSS to detect stiff volumes embedded under a layer of
softer silicone or biological tissue was evaluated on flat silicone samples of
different thickness with embedded stiff hemisphere of silicone [IV]. Figure
12 shows the measurements of fF on ten different silicone samples.
Nine with with embedded hemispheres with the distance from the top of the
hemisphere to the surface of the soft silicone ranging from d = 1 t0 9 mm.
Figure 12. The fF (mean ± SD, n = 6) as a function of d with the sensor
centred above the embedded stiff silicone hemisphere, IZ = 0.4 and 0.8 mm is
shown.
28
The statistical tests of significant differences between fF measured at
the position of the hemisphere, and the average fF of the surrounding
soft silicone were performed using Wilcoxon signed rank test. From figure 11
and the statistical analysis, it was evident that the stiff hemisphere could be
detected at d ≤ 4 mm (p < 0.05). However, the statistical analysis also
showed that the hemisphere, with some exceptions, could be detected with a
weak significance at d = 5 to 9 mm (p < 0.05).
The same type of silicone hemispheres was used for the biological tissue
samples, chicken breast muscles, where a small incision was made on the
underside to make room for the hemisphere. The thicknesses of the
defrosted slices of chicken muscle were, as mentioned earlier, cut in a semi
defrosted state to get an even surface and equal thicknesses. However, the
thawed slices turned out slightly unequal in thickness so for that reason, the
slices were divided into three groups depending on their thickness. This
resulted in six slices for each group with d = 1.5 ± o.5 mm, d = 3.5 ± o.5 mm
and d = 6.5 ± o.5 mm when the nodules were in place. The measured
fF from a point centred over the hemisphere were significantly
different (p < 0.05) compared to the measurements of the surrounding
tissue.
In general, the depth sensitivity for this TRSS agrees with the estimated
sensing depth presented by Jalkanen et al [66]. In that study, the depth
sensitivity was extrapolated from a measured stiffness parameter and the
histological data of the prostate tissue and the results showed an estimated
sensing depth of 3.5 to 5.5 mm for IZ = 1 mm.
6.5 The clamping force
The spherical shaped samples were all subjected to an axially clamping force,
FC when mounted in the rotatable holder [V]. The effect on fF caused
by FC was evaluated with forces up to five or six times higher compared to
what was used later during the other stiffness measurements on spherical
samples. The evaluation on silicone samples showed that fF did not
depend on FC (p > 0.05). The fF for porcine muscle tissue showed a
weak but significant dependence (p < 0.05) on FC. The fF increased
with FC, most noticeable at FC > 800 mN. For this reason, FC was chosen to
be in the range of 500 ± 50 mN for the further studies. For Prostate 1, FC
was higher (FC = 760 ± 99 mN), due to its size and weight.
29
6.6 Measurements on prostate gland ex vivo
In previous works [9, 35, 42, 53, 66, 67], the stiffness measurements were
performed on slices of human prostate tissue with PCa, whereas in this study
the measurements were made on the whole human prostate gland including
the capsule. The previous measurements on other spherical shaped samples
with the rotatable sample holder showed to have been important as the time
frame for measurements on human prostates were very narrow so the
measurements on prostates was quite stressful. The measurements on the
two resected prostate glands took place about 1 hour after that they had been
surgically removed. The prostate glands were mounted in the rotatable
sample holder of the TRSS with the urethra axially. This was done to enable
measurements in a line around the prostate. For the measurements, the
prostate was rotated in steps of approximately 10° between each
measurement, as there was only enough time for a limited number of MPs
due to the short exsess time ( which was 1 hr including transport between the
location of the TRSS and the pathology department).
After the measurements with the TRSS, the prostates were transported
back to the pathologist for routine processing and examination. A tissue slice
of Prostate 1 embedded in paraffin, stained at the periphery in order to locate
the left anterior (red), the right anterior (green) and the dorsal (yellow) parts
of the prostate is shown in figure 13AB. Due to the large size of that
particular prostate gland, the slice had to be cut in a half in this phase, a left
and a right part. Note that they are placed correctly next to each other in
figure 12, even though the mid-section does not look as a straight cut, which
maybe could be expected.
30
A B C D(2) (3)(1)
Figure 13. Prostate 1 in a paraffin block is shown in figure A) and B). At the
periphery, the colours from the staining are visible. Arrow (1) points at punch holes
in the tissue slice where tissue specimens have been taken out after the stiffness
measurements. C) and D) Photomicrographs of the corresponding tissue slice.
Arrows (2) and (3) are aimed at small areas in both the left and right dorsal parts of
the prostate with cancer tumours.
The measurements on Prostate 2 were performed in the same way as for
Prostate 1. A large area of the anterior part of the prostate was marked as
cancer tumours, on the left anterior a Gleason score of 7 (3+4) was noticed,
figure 14.
A B(1) (2)
Figure 14. Prostate 2 is shown in A) and B). A) The tissue slice embedded in
paraffin. B) The photomicrograph of the corresponding tissue slice. Arrow (1) is
pointing at the Gleason staged cancer tumour, and arrow (2) is aimed at the anterior
area containing cancer tumours.
31
Figure 15 shows the results from the stiffness measurements on both
prostates. The areas with higher stiffness parameter values, fF , were
mainly located in the anterior parts of the prostate, except for Prostate 1
where higher values were noticed in the dorsal aspect as well. The anterior
parts consist of mainly smooth muscle and connective tissue, which usually
is stiffer. The dorsal aspect consists more of glandular tissue which is soft.
The dorsal aspect of Prostate 1 had small multiple areas of tumour tissue
encircled in figure 13CD which correspond to the stiffness measurements
and was found ti be significantly different (p < 0.05) to the rest of the
measuremnts on the dorsal side of theprostate [V]. On Prostate 2, figure 14B,
the dorsal aspect showed low stiffness values, but a large area in the anterior
were marked for presence of tumours by the pathologist.
Figure 15. The measurements on prostate 1 (P1) and prostate 2 (P2) are shown.
fF was measured in a line around the prostates,along the great circle,
perpendicular to the urethra. The colours in the figure are the same as how the
pathologist dyed the different areas of the prostate glands.
32
33
7. General Summary and Conclusion
7.1 Conclusion
In this doctoral thesis, the aim was to design and evaluate a flexible tactile
resonance sensor system for the possibility to detect cancer in soft human
tissue with a focus on prostate cancer.
The work started with measurements on flat samples, soft tissue models
made of silicone to evaluate the errors and overall accuracy of the measured
parameters Δf, F and fF which were estimated to be < 1.5% of the
measured values. The effect of a contact angle deviating from a
perpendicular indentation direction and the impact of different indentation
velocity on the measured parameters were evaluated. Contact angles
deviating < 10° from the perpendicular to the surface of the measured object
did not significantly affecte the measurement parameters. Flat silicone and
chicken muscle tissue samples with embedded stiffer volumes (nodules)
were used to investigate the ability to detect stiffer volumes at different
depth under the surface of the tissue phantoms. The results showed that
embedded stiff nodules with certainty could be detected down to a depth of ≤
4 mm under the surface of the measured object, with a small indentation
depth of 0.4 to 0.8 mm into the surface.
Spherical shaped soft tissue phantoms made of silicone and porcine
muscle tissue were used to evaluate the measurement procedures when
using the rotatable holder for spherical objects. Measurements have been
made on resected human prostates and after comparing the measurements
with the histopathological evaluation, it was shown that the TRSS stiffness
parameter could distinguish areas with cancer from normal prostate tissue
(p < 0.05).
In conclusion the work presented here involving the TRSS has shown that
the resonance sensor technique has the potential to contribute to the future
development of new techniques and routines for diagnosis and detection of
prostate cancer tumours, for example during open surgery procedures and
thus become an important clinical application in the future.
34
35
8. Acknowledgements
I would like to express my sincere gratitude to the persons who have
supported me in this work.
First of all, I would like to thank my supervisor Dr. Britt Andersson for
introducing me to the world of a PhD student. Her encouragement and
willingness to discuss problems that I have come across on the way have
been invaluable. Her positive attitude and ability to see things from a
different angle have been of great importance, as has also her help by being a
co-author on my papers.
I also want to thank Professor Olof Lindahl and Dr. Ville Jalkanen in their
role of co-supervisors. They have always been helpful and willing to help me
with obstacles that I have stumbled on. They have also been very important
co-authors on my papers.
Furthermore, I would like to express my gratitude to Sven Elmå, who helped
with the manufacturing of many mechanical components as well as Johan
Pålsson, who helped with the software LabView®, and Lars Karlsson who
helped with the electronics. Morgan Nyberg at Department of Engineering
Sciences and Mathematics, Luleå University of Technology, has assisted with
the measurements and logistics for the measurements on the excised
prostate gland.
I wish to express my gratitude to Kerstin Almroth, Research Assistant at the
Department of Surgical and Perioperative Sciences, Urology and Andrology
at Umeå University, Wanzhong Wang, Pathologist, Pernilla Andersson,
Research Assistant, Birgitta Ekblom, Research Assistant and Susanne
Gidlund, Research Assistant, at the Department of Clinical Pathology at
Umeå University Hospital.
The study was supported by The Industrial Doctoral School at Umeå
University, BioResonator AB, and by grants from Objective 2 North Sweden-
EU Structural Fund.
And last but not least, I would like to thank family and friends who have
supported me through this time.
36
37
9. References
1. Ferlay J, Steliarova-Foucher E, Lortet-Tieulent J, Rosso S, Coebergh
JWW, Comber H, Forman D and Bray F 2013 Cancer incidence and
mortality patterns in Europe: Estimates for 40 countries in 2012
European Journal of Cancer 49 1374-1403
2. American Cancer Society (21/01/2014) at http://www.cancer.org
3. The Swedish National Board of Health and Welfare (Socialstyrelsen)
(21/01/2014) at http://www.socialstyrelsen.se
4. Schröder FH, Hugosson J, Roobol MJ, Tammela TLJ, Ciatto S,
Nelen V, Kviatkowski M, Lujan M, Lilja H, Zappa M, Denis LJ,
Recker F, Berenguer A, Määttänen L, Bangma CH, Aus G, Villers A,
Rebillard X, van der Kwast T, Blijenberg BG, Moss SM, de Koning
HJ, Auvinen A, and E.R.S.P.C. Investigators 2009 Screening and
prostate-cancer mortality in randomized European study N Engl J
Med. vol. 360 no. 13 1320-1328
5. Candefjord S, Ramser K, Lindahl OA 2009 Technologies for
localization and diagnosis of prostate cancer J. Med. Eng. Tech.
33 585-683
6. Lindahl OA, Constantinou CE, Eklund A, Murayama Y, Hallberg P
and Omata S 2009 Tactile resonance sensors in medicine J. Med.
Eng. Tech. 33 263-273
7. Gwilliam JC, Pezzementi Z, Jantho E, Okamura AM, Hsiao S 2010
Human vs. robotic tactile sensing: detecting lumps in soft tissue
IEEE Haptics Symposium 2010 (Waltham, MA) 21-28. ISBN 978-1-
4244-6821-8
8. Tiwana MI, Redmond SJ, Lovell NH 2012 A review of tactile sensing
technologies with applications in biomedical engineering Sensors
and Actuators A179 17-31
38
9. Jalkanen V, Andersson BM, Bergh A, Ljungberg B and Lindahl OA
2006 Resonance sensor measurements of stiffness variations in
prostate tissue in vitro – a weighted tissue proportion model
Physiol. Meas. 27 1373-86
10. Sakai N, Tatsuta M, Yano H, Iishi H, Ishiguro S 2000 Diagnosis of
the extent of gastric cancers by a new endoscopic ultrasonic tactile
sensor Gastrointest Endosc 51 69-73
11. Chung C-C. Hsieh C-C. Lee H-C. Wu M-H. Huang M-H. Hsu W-H
and Hsu H-S. 2010 Accuracy of helical computed tomography in the
detection of pulmonary colorectal metastases. The Journal of
Thoracic and Cardiovascular Surgery 141 no. 5 1207-1212.
12. Cerfolio RJ. and Bryant AS. 2008. Is palpation of the non-resected
pulmonary lobe(s) required for patients with non-small cell lung
cancer? A prospective study. The Journal of Thoracic and
Cardiovascular Surgery 135 no. 2 261-268.
13. Cerfolio RJ. McCarty T. and Bryant AS. 2009. Non-imaged
pulmonary nodules discovered during thoracotomy for
metastasectomy by lung palpation. European Journal of Cardio-
Thoracic Surgery 35 786-791
14. Detterbeck FC. and Egan TM. 2004. Thorascopy using a substernal
handport for palpation. Ann. Thorac. Surg. 78 1031-1036.
15. Ahn B-M, Kim J, Ian L, Rha K-H and Kim H-J 2010 Mechanical
property characterization of prostate cancer using minimaly
motorized indenter in an ex vivo indentation experiment Urology
vol 76 no 4 1007-1011
16. Paulet T. and Fryer G. 2009. Inter-examiner reliability of palpation
for tissue texture abnormality in the thoracic paraspinal region.
International Journal of Osteopathic medicine 12 92-96
17. Martini F 2004 Fundamentals of Anatomy & Physiology. 6th ed.
Pearson Benjamin Cummings, New York
18. Tortora G, Grabowski SR 1996 Principles of Anatomy and
Physilology. 8th ed. Harper Collins College Publishers, New York
39
19. Taichman RS, Cooper C, Keller ET, Pienta KJ, Taichman NS and
McCauley LK 2002 Use of the Stromal Cell-derived Factor-1/CXCR4
Pathway in Prostate cancer Metastasis to Bone Cancer Res 62 1832-
1837
20. Bayar DP, Mostofi FK and the veterans Administration Cooperative
Urological research Group 1972 Carcinoma of the prostate:
prognostig evaluation of certain pathological features in208 radical
prostatectomies Cancer 30 5-13
21. Krouskop TA, Wheeler TM, Kallel F, Garra BS and Hall T 1998
Elastic moduli of breast and prostate tissues under compression
Ultrason imaging 20 260-274
22. Zhang M, Nigwekar P, Castaned B, Hoyt K, Joseph JV, di SAnt
Agnese A, Messing EM, Strang JG, Rubens D and Parker KJ 2008
Quantative characterization of viscoelastic properties of human
prostate correlated with histology Ultrasound in Med & Biol vol 34
no 7 1033-1042
23. Hricak H. Choyke PL. Eberhardt SC. Leibel SA. And Scardino PT.
2007 Imaging prostate cancer: A multidicplinary perspective
Radiology vol. 243 no. 1 28-53
24. Feleppa EJ. Porter CR. Ketterling J. Lee P. Dasgupta S. Urban S and
Kalisz A. 2004 Recent developments in tissue-type imaging (TTI) for
planning and monitoring treatment of prostate cancer Ultrasonic
Imaging 26 163-172
25. Carey BM. 2005 Imaging for prostate cancer Clinical Oncology
17 553-559
26. Ravizzini G. Turkbey B. Kurdziel K and Choyke PL. 2009 New
horizons in prostate cancer imaging European Journal of Radiology
70 212-226
27. Taoka T. Mayr NA. Lee HJ. Yuh WTC. Simonson TM. Rezai K. and
Berbaum KS. 2001 Factors influencing visiulalization of vertebral
metatstases on MR imaging versus bone scintigraphy. AJR 176
1525-1530
40
28. Raz O. haider M. Trachtenberg J. Leibovici D. and Lawrentschuk N.
2010 MRI for men undergoing active surveillance or with rising PSA
and negative biopsies. Nature reviews, Urology 7 543-550
29. Qayyum A. Coakley FV. Lu Y. Olpin JD. Wu L. Yeh BM. Carroll PR.
And Kurhanewicz J. 2004 Organ-confined prostate cancer: Effect of
Prior transrectal biopsy on endorectal MRI and MR spectroscopic
imaging American Journal of Roentgenology 183 1079- 1083
30. Ikonen S. Kivisaari L. Vehmas T. Tervahartiala P. Salo JO. Taari K.
and Rannikko S. 2001 Optimal timing of post-biopsy MR imaging of
the prostate Acta Radiologica 42 70-73
31. Fine SW, Amin MB, Berney DM, Bjartell A, Egevad L, Epstein JI,
Humphrey PA, Mag-Galluzzi C, Montironi R and Steif C 2012 A
contemporary update on pathology reporting for prostate cancer:
Biopsy and radical prostatectomy specimens European Urology 62
20-39
32. Delahunt B, Miller RJ, Srigley JR, Evans AJ and Samaratunga H
2012 Gleason grading: past, present and future Histopathology 60
75-85
33. Srigley, JR, Humphrey PA, Amin MB, Chang SS, Egevad L, Epstein
JI, Grignon DJ, McKiernan JM, Montironi R, Renshaw AA, Reuter
VE and Wheeler TM 2009 Protocol for the examination of
specimens from patients with carcinoma of the prostate gland Arch
pathol Lab Med 133 1568-1576
34. Omata S, Terunuma Y 1992 New tactile sensor like the human hand
and its applications Sens Actuat. 35 9-15
35. Jalkanen V, Andersson BM, Bergh A, Ljungberg B and Lindahl OA
2008 Explanatory models for a tactile resonance sensor system –
elastic and density-related variations of prostate tissue in vitro
Physiol. Meas. 29 729-45
36. Fung YC 1993 Biomechanics – Mechanical properties of living
tissues. New York, Springer-Verlag
37. Gladwell GML and Kleesattel C 1968 The contact – impedance
meter-2 Ultrasonics 6 244-252
41
38. Lindahl OA, Ängquist K-A and Ödman S 1991 Impression technique
for the assessment of oedema – Technical improvement and
methodological evaluation of a new technique Med. & Biol. Eng. &
Comput. 29 591-7
39. Sasai S, Zhen Y-X, Suetake T, Tanita Y, Omata S and Tagami H 1999
Palpation of the skin with a robot finger: an attempt to measure skin
stiffness with a probe loaded with a newly developed tactile vibration
sensor and displacement sensor Skin Res. Tech. 5 237-46
40. Miyaji K, Furuse A, Nakajima J, Kohno T, Ohtsuka, T, Yagyu k, Oka
T and Omata S 1997 The stiffness of lymph nodes containing lung
carcinoma metastases – a new diagnostic parameter measured by a
tactile sensor Cancer 80 1920-25
41. Kusaka K, Harihara Y, Torzilli G, Kubota K, Takayama T, Makuuchi
M, Mori M and Omata S 2000 Objective evaluation of liver
consistency to estimate hepatic fibrosis and functional reserve for
hepatectomy J. Am. Coll. Surg. 191 47-53
42. Jalkanen V, Andersson BM, Bergh A, Ljungberg B and Lindahl OA
2006 Prostate tissue stiffness as measured with a resonance sensor
system: a study on silicone and human prostate tissue in vitro Med.
Biol. Eng. Comput. 44 593-603
43. Hallberg P, Lindén C, Lindahl OA, Bäcklund T and Eklund A 2004
Applanation Resonance Tonometry for Intraocular Pressure in
Humans Physiol. Meas. 25 1053-1065
44. Eklund A, Hallberg P, Lindén C and Lindahl OA 2003 An
Applanation Resonator Sensor for Measuring Intraocular Pressure
Using Combined Continuous Force and Area Measurement Invest.
Ophtalmol. Vis. Sci. 44 no. 7 3017-3024
45. Kim Y. Ahn B. Lee JW. Rha KH and Kim J. 2013 Local property
characterization of prostate glands using inhomogeneous modelling
based on tumor volume and location analysis Med. Biol. Eng.
Comput 51 197-205
42
46. Ahn B. and Kim J. 2010 Measurement and characterization of soft
tissue behaviour with surface deformation and force response under
large deformations Medical Image Analysis 14 138-148
47. Liu H. Sangpradit K. Li M. Dasgupta P. Althoefer K. and Seneviratne
LD. 2013 Inverse finite-element modeling for tissue parameter
identification using a rolling indentation probe Med. Biol. Eng
Comput 52 17-28
48. Sangpradit K. Liu H. Dasgupta P. Althoefer K. and Seneviratne LD.
2011 Finite-element modeling of soft tissue rolling indentation IEEE
Transactions on Biomedical engineering vol. 58 no. 12 3319-3327
49. Tanaka Y. Yu Q. Duomoto K. Sano A. Hayashi Y. Fujii M. Kajita Y.
Mizuno M. Wakabayashi T. and Fujimoto H. 2010 development of a
real time tactile sensing system for brain tumor diagnosis Int J
CARS 5 359-367
50. Samur E. Sedef M. Basdogan C. Avtan L and Duzun O. 2007 A
robotic indenter for minimally invasive measurement and
charaterization of soft tissue response Medical image Analysis
11 361-373
51. Perri MT. Trejos AL. Naish MD. Patel RV. and Malthaner RA. 2010
New tactile sensing sytem for minimally invasive surgical tumour
localization Int J Med Robotics Comput Assist Surg 6 211-220
52. Beccani M. Di Natali C. Sliker LJ. Schoen JA. Rentschler ME.
Valdastri P. 2014 Wireless tissue palpation for inoperative detection
of lumps in the soft tissue IEEE Transactions on Biomedical
engineering vol. 61 no. 2 353-361
53. Lindberg P, Andersson BM, Bergh A, Ljungberg B and Lindahl OA
2006 Prostate cancer detection with an improved resonance sensor
system: parameter evaluation in a silicone model and on human
prostate tissue in vitro Med Bio Eng Comput. 44 1053-59
54. Eklund a, Bergh A, Lindahl OA 1999 A catheter tactile sensor for
measuring hardness of soft tissue: measurement in a silicone modela
and in vitro human prostate model Med. Biol. Eng. Comput. 37 618-
624
43
55. Bergström JS and Boyce MC 1998 Constitutive modelling of the
large strain time-dependent behaviour of elastomers J Mech Phys
Solids 46 931-954
56. Martins PALS, Natal Jorge RM and Ferreira AJM 2006 A
comparative study of several material models for prediction of
hyperelastic properties: Application to silicone-rubber and soft
tissue Strain 42 135-147
57. Carter FJ. Frank TG. Davies PJ. McLean D. and Cuscheri A 2001
Measurements and modelling of compliance of human and porcine
organs. Medical Image Analysis 5 231-236
58. Mazza E. Nava A. Hahnloser D. Jochum W and Bajka M. 2007 The
mechanical response of human liver and its relation to histology: An
in vivo study Medical Inage Analysis 11 663-672
59. McKee CT, Last JA, Russell P, and Murphy CJ 2011 Indentation
versus tensile measurements of Young’s modulus for soft biological
tissues Tissue Engineering, part B vol 17 no 3 155-164
60. Sedef M. Samur E and Basdogan C. 2006 Real-time finite-element
simulation of linear viscoelastic tissue behaviour based on
experimental data Virtual and Augmented Reality Supported
Simulators , IEEE Computer Society nov/dec 58-68
61. Phipps S, Yang THJ, Habib FK, Reuben RL and McNeill SA 2005
Measurements of tissue mechanical characteristics to distinguish
between benign and malignant prostatic tissue Urology 66 447-50
62. Lindahl OA and omata S 1995 impression technique for the
assessment of oedema: comparison with a new tactile sensor that
measures physical properties of tissue Med. Biol. Eng. Comput.
33 27-32
63. Åstrand AP 2012 A Flexible Resonance Sensor Sytem for Detection
of Cancer Tissue – Evaluation on Silicone. Lic Thesis, Chapter 4, 5-8.
Umeå University, Sweden ISBN 978-91-7459-477-5
64. El Sayed T. Mota A. Fraternali F. and Ortiz M. 2008 A variational
constitutive model for soft biological tissues Journal of
Biomechanics 41 1458-1466
44
65. Murayama Y 2005 Development of a micro tectile sensor and its
biomedical applications. PhD thesis, chapter 5. Nihon University,
Japan.
66. Jalkanen V, Andersson BM, Bergh A, Ljungberg B and Lindahl OA
2007 Spatial variations in prostate tissue histology measured by a
tactile resonance sensor. Physiol. Meas. Vol 28 no.10 1267-1281
67. Jakanen V, Andersson BM, Bergh A, Ljungberg B and Lindahl OA
2013 Indentation loading response of a resonance sensor –
discriminating prostate cancer and normal tissue. J Med Eng
Technol vol. 37 no. 7 416-423
45
Summary of papers
Paper I
A flexible measurement system using piezoelectric resonance sensor
technology developed for measurements on both flat and spherical objects is
described. The newly developed measurement system can distinguish
between five flat silicone discs with different stiffness through measurements
of the frequency shift and the applied force. All measurements were made
perpendicular to the surface of the silicone disc and the indentation velocity
of the sensor was 5 mm/s.
The evaluation of the measurement system showed that the measured
frequency shift from the measurements on the different silicone mixtures
was in compliance with the given hardness measurements provided by the
silicone manufacturer and according to the DIN ISO standard 2137. A
linearly correlation with the cone penetration depth used in the standardwas
found (R2 = 0.99, p < 0.05). The paper reports that the measurement system
could distinguish between silicones of different stiffness with repeatability.
The function of the measurement system is further evaluated on spherical
objects and presented in Paper II.
Paper II
The measurement system described in Paper I is further evaluated, but in
this study the measurements were made on spherical objects. Spherical soft
tissue models with a diameter of 40 mm were made of a semi-soft silicone.
Initially, measurements were made with the sensor movement vertical
towards each measuring point along the axis of rotation, disregarding that
the contact angle varied for each point. Secondly, the measurements were
made ensuring that the contact angle was kept perpendicular to the surface
of the same measuring points as in the first measurement series. All
measurements were made with the indentation velocity of 5 mm/s. The two
measurement series were compared. A systematical deviation in the
frequency shift and applied force were noticed between the vertical
measurements compared to the measurements made perpendicular to the
surface of the silicone. The measurements with a contact angle that deviated
from the perpendicular were underestimated. However the applied force
seemed to be less sensitive for the deviating contact angles. The conclusion
was that the measurement system could be used for measurements on
46
spherical objects, and the importance of making measurements with the
sensor movement perpendicular to the surface of the measured object was
pointed out. Further evaluation of the angular dependency of the frequency
shift and applied force is reported in Paper III.
Paper III
Further measurements were made on flat soft tissue silicone models. In this
study, to evaluate the dependence of the frequency shift, applied force and
the stiffness parameter on the contact angle, the indentation velocity for
different impression (indentation) depths. The overall error of the resonance
sensor measurement system was also determined. In this study, two soft
tissue silicone models of different stiffness (hardness) were used.
Measurements were made with the indentation velocities ranging from 1
mm/s to 5 mm/s, and contact angles varying up to 35° from a line
perpendicular towards the surface (0°).
The results showed a velocity dependence for the stiffness parameter, as a
higher contrast between the silicones with different stiffness at higher
indentation velocity was noticed. The three measured parameters: frequency
shift, applied force and the stiffness parameter, all showed to be dependent
on the contact angle. One conclusion from the study was that deviations in
the contact angle from the perpendicular can be accepted up to ± 10°. Larger
deviation in the contact angle resulted in increased errors which lead to an
underestimation of the frequency shift. This could give a wrong indication
for cancer.
The maximum errors for the measured parameters were generally small
compared to the magnitude of the measured values, except for the relative
error in the impression (indentation) depth, which was 12.5% when the
highest indentation velocity was used. Continued work and measurements
on biological tissue are reported in Paper IV.
Paper IV
The focus in this study was to evaluate at what depth a stiff nodule of silicone
could be detected in flat discs of silicone and in biological tissue. A
hemisphere of silicone with a diameter of 8 mm was embedded at different
depths ranging from 1 to 9 mm in a surrounding softer silicone. The same
type of hemispheres were used for measurements on chicken muscle tissue
that were used as phantoms for biological tissue. The measurements were
made perpendicular to the surface of the measured objects and at a
indentation velocity of 4 mm/s. For depths of up to 4 mm of soft silicone
47
covering the stiff silicone hemisphere, the response of fF clearly
indicated the presence of the nodule. For depths exceeding 5 mm the
nodules could be detected using statistical analysis that showed that also
stiffer nodules at depths from 5 to 9 mm could be distinguished with
significance. The measurements on the biological tissue showed that it was
clear that the sensor could detect the stiff silicone hemisphere up to a depth
of 4 mm (p < 0.05, n = 24).
The results are promising for further studies for detection of prostate
cancer. In Paper V, the work is continued on spherical phantoms made of
silicone and biological tissue mimicking a prostate gland.
Paper V
In this paper, the tactile resonance sensor system (TRSS) was evaluated for
measurements on spherical objects, using the rotatable sample holder of the
TRSS. The evaluation was taken in three steps. At first measurements were
made on spherical soft tissue phantoms made of silicone with embedded stiff
nodules at different depths under the surface of the sphere. The next step
was to use spherical shaped biological tissue models of porcine muscle with
embedded stiff nodules before the las phase of the study that included
measurements on resected prostate glands with cancer tumours.
The spherical samples were mounted in a rotatable holder where they were
held in place by two spring loaded concave discs in a way that the sample
could be rotated around its own horizontal axis. The movement of the sensor
was kept perpendicular to the surface of the samples at all times. For the
measurements on the silicone spheres, the measurement positions were
chosen directly over the nodule αsm = 0° and ± 10° and ± 20° to the left and
right. The reason for this was to evaluate if the sensor could detect a nearby
stiff nodule when the sensor movement was angle towards the surface and a
short distance, 3,5 and 7 mm on the side of the curved surface.
In the study, two resected prostate glands from patients undergoing a
radical prostatectomy were obtained for measurements: Ethical approval
was given by the Ethical Committee at Umeå University (Dnr 03-423). From
the measurements it was concluded that the TRSS could detect stiffer areas
on the prostate glands, areas that were confirmed to be prostate cancer
through a histopathological evaluation.
48
Resonance Sensor Lab
1. P. Hallberg, Applanation Resonance Tonometry for
Intraocular Pressure Measurements, Dissertation 2006, ISSN 1653-6789:1, ISBN 91-7264-061-8
2. V. Jalkanen, Resonance Sensor Technology for
Detection of Prostate Cancer, Licentiate thesis 2006, ISSN 1653-6789:2, ISBN 91-7264-153-3
3. P. Lindberg, Development of a Resonance Sensor
System Designed for Prostate cancer Detection, Licentiate thesis 2007, ISSN 1653-6789:3, ISNB 978-91-7264-432-8
4. V. Jalkanen, Tactile Sensing of Prostate Cancer – A Resonance sensor Method Evaluated Using Human Prostate Tissue in Vitro, Dissertation 2007, ISSN 1653-6789:4, ISBN 978-91-7264-441-8
5. A P. Åstrand, A Flexible Resonance Sensor System For Detection of Cancer Tissue – Evaluation on Silicone, Licentiate thesis 2012, ISSN 1653-6789:5 ISBN 978-91-7459-477-5
6. A P. Åstrand, A tactile Resonance Sensor System for Detection of Prostate Cancer ex vivo – Design and Evaluation on Tissue Models and Human Prostate, Dissertation 2014, ISSN 1653-6789:6, ISBN 978-91-7601-006-8
Department of Applied Physics and
Electronics
Centre for Biomedical Engineering and
Physics
Industrial Doctoral School
Umeå University, SE-901 87 Umeå
www.tfe.umu.se
ISSN 1653-6789:6
ISBN 978-91-7601-006-8