ac7114-1 rev h audit criteria for nondestructive testing facility penetrant survey-1

Upload: nayan-vyas

Post on 24-Feb-2018

235 views

Category:

Documents


0 download

TRANSCRIPT

  • 7/25/2019 Ac7114-1 Rev h Audit Criteria for Nondestructive Testing Facility Penetrant Survey-1

    1/33

    PRI operating procedures provide that "This report is published by PRI to advance the state of technical, engineering, and qualitysciences. The use of this report is entirely voluntary, and its applicability and suitability for any particular use, including any pateninfringement arising there from, is the sole responsibility of the user."

    PRI values your input. To provide feedback on this document, please contact the appropriate commodity staff engineer.(Contact information is located athttp://www.eauditnet.comunder Contact Us.)Copyright 2015 Performance Review Institute. All rights reserved.

    t-frm-15 19-Apr-2015

    161Thorn Hill Road

    Warrendale, PA 15086-7527

    AUDIT CRITERIA

    AC7114/1 REV. H

    Issued 1997-07

    Revised 2015-04

    SupersedingAC7114/1 Rev G

    TO BE USED ON AUDITS STARTING ON OR AFTER JANUARY 25, 2015Nadcap

    AUDIT CRITERIA FORNONDESTRUCTIVE TESTING FACILITY PENETRANT SURVEYEditorial Change was made to Section 6.1 on December 03, 2014

    Editorial Change made to reflect the Nadcap Audit Criteria Template Form on 08 January 2015

    Editorial Change made to Section 2.1 on April 19, 2015

    1. SCOPE

    These Audit Criteria define the requirements for suppliers seekingNadcap accreditation in Penetrant and are to be used as supplement toPRI AC7114. These criteria and the applicable User PrimeSupplemental Checklists will ensure that NDT suppliers meet therequirements for Nadcap accreditation in Penetrant.

    2. GENERAL INFORMATION

    In completing this assessment, auditors are instructed to respond witha "Yes or No" to address compliance with each statement ofrequirement. For any negative responses, the auditor must clearlyindicate in the NCR if the No" reflects noncompliance with respect toexistence, adequacy, and/or compliance. Existence relates to evidenceof a documented procedure or policy, adequacy relates to thecompleteness of the procedure or policy, and compliance relates toevidence of effective implementation. This checklist includes the use ofthe word shall which indicates a clear requirement that must be metat all times. The word, may indicates a possible resolution to arequirement or one recommended method to achieve a givenrequirement, but does not represent the only way to meet thatrequirement. In addition the checklist includes the term Compliance

    Assessment Guidance. This term is used to reflect the Task Groupsexpectation when answering the checklist question as YES or NA.

    The auditor must verify the list of procedures provided by the supplierat the time of the audit. Any corrections or updates to the list must beidentified using notes inserted at the applicable criterion.

    http://www.eauditnet.com/http://www.eauditnet.com/http://www.eauditnet.com/http://www.eauditnet.com/
  • 7/25/2019 Ac7114-1 Rev h Audit Criteria for Nondestructive Testing Facility Penetrant Survey-1

    2/33

    PRI AC7114/1 Rev. H - 2 -

    All negative responses require a Nonconformance Report (NCR). NotApplicable (NA) responses do not require an explanation unlessotherwise noted. There is only one plausible reason for an NA, whichis, that a particular operation or issue is not being used at the supplier.There are no NAs simply for a lack of a customer requirement. If a

    system is in use, then all questions pertaining to that system areapplicable. If verification of results require documentation it shall be sonoted in this checklist.

    The audit results shall not include any customer proprietaryinformation. Technical information on parts which have beendesignated Export Controlled License Required (EC-LR) cannot beinput into eAuditNet. If auditors have any questions about this, theyshould contact the Staff Engineer for directions.

    GO FORWARD PLANAny new or revised procedure made after 4December 2006 requires compliance to the following requirements.

    Procedures written prior to this date shall be acceptable; however asystem needs to be in place to ensure that changes reflect the currentNadcap requirements.

    2.1 References:

    SAE Publications: Available from SAE, 400 Commonwealth Drive,Warrendale, PA 15086-0001

    AMS 2644 Inspection Material, Penetrant

    QPL-SAE-AMS-2644 Qualified Products List, InspectionMaterial, Penetrant

    PRI Publications: Available from PRI, 161 Thorn Hill Road,Warrendale, PA 15086-0001

    PD 1100 Nadcap Program Requirements

    ASTM Publications: Available from ASTM, 100 Bar Harbor,Conshohocken, PA 19428 2959

    ASTM-E-1417 Standard Practice for LiquidPenetrant Examination

    ASTM-E-1135 Test Method for Comparing theBrightness of FluorescentPenetrant

    ASTM-D 95 Test Method for Water inPetroleum Products andBituminous Materials byDistillation

    3. MATERIALS AND EQUIPMENT

  • 7/25/2019 Ac7114-1 Rev h Audit Criteria for Nondestructive Testing Facility Penetrant Survey-1

    3/33

    PRI AC7114/1 Rev. H - 3 -

    3.1 Materials, Equipment Listing and Information

    The supplier shall complete and the auditor shall review, the supplier'sprovided materials and equipment listing, for adequacy, content and

    correctness. Attachments are for information only but shall becompleted fully. NCRs shall not be written against attachments.(Attachment AC7114/1-A).

    ATTACHMENT

    3.2 Material Certification

    3.2.1 Is the manufacturers certification available for each batch of penetrant,emulsifier and developer in use, and is there evidence that theyconform to the applicable AMS?

    YES NO

    Compliance Assessment Guidance: The last batch added to the tankdefines the batch for the tank.

    3.3 Material Identification

    3.3.1 Are the contents of all containers (tanks/drums/bulbs) labeled with, oris there traceability to, the material and batch number(s)?

    YES NO

    3.4 Equipment Arrangement

    3.4.1 Is equipment arranged to prevent mixing, contamination or incorrectapplication of materials?

    YES NO

    Compliance assessment guidance: For example tanks should becovered when not in use, sprays and containers shall be labeled etc.

    4. PROCEDURES

    4.1 NDT Facility Written Procedures

    4.1.1 Is there a procedure, general or specific, for processing and inspectionusing the liquid penetrant method?

    YES NO

    4.2 Approval

    4.2.1 Does the procedure(s) indicate approval by the applicable Level 3? YES NO

    Compliance Assessment Guidance: The applicable Level 3 may be theNadcap User, Nadcap User qualified or approved Level 3, internalLevel 3 and/or external Level 3.

    4.3 Procedure / Technique

    Does the written procedure, general or specific, contain the followinginformation as a minimum?

    Compliance Assessment Guidance:The term, procedure may be synonymous with a traveler or a routecard, etc. All non-technical requirements that have been flowed down

  • 7/25/2019 Ac7114-1 Rev h Audit Criteria for Nondestructive Testing Facility Penetrant Survey-1

    4/33

    PRI AC7114/1 Rev. H - 4 -

    solely by the baseline will only be required to be met if that in-usetechnique or process has been revised since the inception of thebaseline on December 4, 2006.

    4.3.1 Traceability to the name and address of the facility performing the

    inspection?

    YES NO

    Compliance Assessment Guidance: In some cases it may not benecessary for the actual address as long as the facility is traceable to aunique identifier.

    4.3.2 Procedure identification number and the date the procedure wasapproved?

    YES NO

    4.3.3 A requirement that all personnel are qualified and certified to therequired level for the activity undertaken and the applicable contractrequirements?

    YES NO

    4.3.4 Part number and/or material to be examined? YES NO

    4.3.5 Complete processing parameters for the penetrant inspection materialsincluding dwell times, application methods, drying times, concentrationof emulsifiers, temperatures and controls to prevent excessive dryingor overheating, as appropriate?

    YES NO

    4.3.5.1 A requirement that components, penetrant, and ambient temperatures

    shall be maintained within the range of 50 F (10 C) to 125 F (52 C)?

    YES NO

    4.3.5.2 A requirement that rinse water temperatures are controlled within the

    range of 50F (10C) to 100 F (38C)?

    NOTE: NA only if Method C is used

    YES NO NA

    4.3.6 Method, Type and Sensitivity Level of Penetrant, Emulsifier, andDeveloper Form in accordance with AMS 2644?

    YES NO

    4.3.7 How hydrophilic emulsifier concentrations are established?

    What is the established concentration range as defined by theprocedure?

    ________________________________________

    YES NO NA

    4.3.8 Identification of the components or areas within a component to beinspected?

    YES NO

    4.3.9 Address or provide traceability to customer acceptance criteria? YES NO

    4.3.10 Special equipment required including type and intensity of light ifdifferent from standard processing?

    YES NO NA

    4.3.11 Complete pre-cleaning and post-cleaning instructions, including YES NO

  • 7/25/2019 Ac7114-1 Rev h Audit Criteria for Nondestructive Testing Facility Penetrant Survey-1

    5/33

    PRI AC7114/1 Rev. H - 5 -

    cleaning materials, sequencing, times, concentrations andtemperatures in accordance with customer requirements or referenceto procedures containing such instructions?

    Compliance Assessment Guidance: Procedure may make reference to

    another procedure or to an outside source.

    4.3.12 Description for the method of identifying inspection status? YES NO

    4.3.13 Method for control of software or programming used for automatedprocessing equipment as applicable?

    YES NO NA

    Compliance Assessment Guidance: This would include software usedto control automated systems.

    4.3.14 All required process control tests and checks called out in section 5 ofthis checklist?

    YES NO

    5. PENETRANT PROCESS CONTROLS

    Compliance Assessment Guidance:

    The following tests marked re-used are required when, and if, thematerials are used from a tank, container or are sprayed andrecovered. If materials are not recovered in any way, tests markedas re-used are not applicable and shall be noted as such.

    If a particular system is not used by the supplier, (i.e., PostEmulsified, Lipophilic) then questions pertaining to that system areNA and no explanation is required. Where equipment is not usedfor a period in excess of the check frequency, the check shall be

    performed immediately prior to use.

    If the following process control checks are performed at thesupplier, adequate procedural coverage is expected to be evidentand is to be reviewed by the auditor. If the tests are performed byan external source, the auditor is expected to review evidence offlow-down (i.e., a Purchase Order) with the appropriate informationto include at a minimum; traceability to the required test and thestandard used. In addition an adequate certification or otherdocument is required to document the results of the tests.

    5.1 Penetrant Brightness Test (reused) SECTION NA

    5.1.1 Is the fluorescent brightness of penetrant being tested at leastquarterly?

    YES NO

    5.1.2 Is ASTM-E-1135 the standard test method used? YES NO

    5.1.3 Is the acceptable limit within 90 -110% of the unused standard? YES NO

    5.1.4 Are records of this test on file and do they indicate acceptable results? YES NO

  • 7/25/2019 Ac7114-1 Rev h Audit Criteria for Nondestructive Testing Facility Penetrant Survey-1

    6/33

    PRI AC7114/1 Rev. H - 6 -

    Compliance Assessment Guidance: The test results may indicate theexact percentage or simply state acceptable or some other term thatdenotes acceptance.

    5.2 Water Content of Water-Washable Penetrant (reused) SECTION NA

    5.2.1 Is the water content of water washable (Method A) penetrant beingtested at least monthly?

    YES NO

    5.2.2 Is ASTM-D-95, Test Method D6304 or Karl Fischer test method used? YES NO

    5.2.3 Is the maximum acceptable limit 5%? YES NO

    5.2.4 Are records of this test on file and do they indicate acceptable results? YES NO

    Compliance Assessment Guidance: The test results may indicate theexact percentage or simply state acceptable or some other term thatdenotes acceptance.

    5.3 Water Content of Water-Based Penetrant (reused) SECTION NA

    5.3.1 Is the water concentration of water based (Method A) penetrant beingtested at least weekly?

    YES NO

    5.3.2 Is a refractometer used to perform this test? YES NO

    5.3.3 Is the concentration in accordance with manufacturersrecommendations?

    What is the current concentration percent? _______________

    YES NO

    5.3.4 Are records of this test on file and do they indicate acceptable results? YES NO

    Compliance Assessment Guidance: The test results may indicate theexact percentage or simply state acceptable or some other term thatdenotes acceptance.

    5.4 Water Content of Lipophilic Emulsifiers (reused) SECTION NA

    5.4.1 Is the water content of in-use lipophilic (Method B) emulsifier beingtested at least monthly?

    YES NO

    5.4.2 Is ASTM-D-95, Test Method D6304 or Karl Fischer test method used? YES NO

    5.4.3 Is the maximum acceptable limit 5%? YES NO

    5.4.4 Are records of this test on file and do they indicate acceptable results? YES NO

    Compliance Assessment Guidance: The test results may indicate theexact percentage or simply state acceptable or some other term thatdenotes acceptance.

    5.5 Concentration of Hydrophilic Emulsifiers SECTION NA

  • 7/25/2019 Ac7114-1 Rev h Audit Criteria for Nondestructive Testing Facility Penetrant Survey-1

    7/33

    PRI AC7114/1 Rev. H - 7 -

    5.5.1 Concentration of in-use (Method D) hydrophilic emulsifiers (used fromcontainers / tanks)

    SECTION NA

    5.5.1.1 Is the concentration of in-use (Method D) hydrophilic emulsifiers being

    checked at least weekly or prior to use?

    YES NO

    Compliance Assessment Guidance: This test is required to beconducted at least prior to use in systems that experience a lowvolume of activity, but if used more often than once a week, weeklywould be the minimum required frequency.

    5.5.1.2 Does this maximum concentration percentage comply with QPL-AMS-2644?

    What is the current concentration percent? _______________

    YES NO

    5.5.1.3 Is a refractometer used to perform this test? YES NO

    5.5.1.4 Has a correction Graph(s) for in use emulsifier been established? YES NO

    Compliance Assessment Guidance: The graph shall have beenprepared for each emulsifier in use and traceable to a specificrefractometer as identified on the graph. A change in batch numberdoes not necessitate a new graph. Generic graphs supplied by themanufacturer are not acceptable.

    5.5.1.5 Is the variation of the concentration within +/- 3% points of theestablished nominal value?

    YES NO

    Compliance Assessment Guidance: The nominal value, and allowedvariation (maximum of +/- 3 % points), must be set so that it does notexceed the maximum concentration permitted by QPL-AMS-2644.

    5.5.1.6 Is the concentration verified upon initial mixing and when adjusted? YES NO

    5.5.1.7 Are records of this test on file and do they indicate acceptable results? YES NO

    Compliance Assessment Guidance: The actual percentage shall berecorded by the supplier or the reading from the refractometer iftraceable to the percentage. Recording, OK or indicating a checkmark

    or other notation, is not acceptable.

    5.5.1.8 Did the facility properly demonstrate this check? YES NO

    Compliance Assessment Guidance:a) The supplier shall be able to demonstrate the correct use of therefractometer in accordance with the manufacturers instructions andshall record the exact concentration or the refractometer readingtraceable to the concentration.b) The refractometer shall be zeroed using water prior to use.

  • 7/25/2019 Ac7114-1 Rev h Audit Criteria for Nondestructive Testing Facility Penetrant Survey-1

    8/33

    PRI AC7114/1 Rev. H - 8 -

    5.5.2 Concentration of spray (Method D) hydrophilic emulsifiers SECTION NA

    5.5.2.1 Are concentrations for spray applications kept at or below 5%? YES NO

    5.5.2.2 Is the concentration of the spray hydrophilic emulsifiers being checkedat least weekly or prior to use? YES NO

    5.5.2.3 Is a refractometer used to perform this test? YES NO

    5.5.2.4 Has a correction graph(s) for in use emulsifier been established? YES NO

    Compliance Assessment Guidance: The graph shall have beenprepared for each emulsifier in use and traceable to a specificrefractometer as identified on the graph. A change in batch numberdoes not necessitate a new graph. Generic graphs supplied by themanufacturer are not acceptable.

    Compliance Assessment Guidance: The graph may be prepared by anexternal source but must meet the requirement defined above.

    5.5.2.5 Is the concentration verified upon initial start up and when maintenanceis performed?

    YES NO

    5.5.2.6 Are records of this test on file and do they indicate acceptable results? YES NO

    Compliance Assessment Guidance: The actual percentage shall berecorded by the supplier or the reading from the refractometer iftraceable to the percentage. Recording, OK or indicating a checkmarkor other notation, is not acceptable.

    5.5.2.7 Did the facility properly demonstrate this check? YES NO

    Compliance Assessment Guidance:a) The supplier shall be able to demonstrate the correct use of therefractometer in accordance with the manufacturers instructions andshall record the exact concentration or the refractometer readingtraceable to the concentration.b) The refractometer shall be zeroed using water prior to use.

    5.6 Dry Developer Characteristics (Form a) SECTION NA

    5.6.1 Is the dry developer being checked at least daily or prior to use? YES NO

    Compliance Assessment Guidance: All systems employing drydeveloper must be checked at least for caking; this includes spray onsystems and portable bulbs or blowers. This may be visually verifiedor demonstrated by system performance.

    5.6.2 Does the check for recycled dry developer ensure the developer hasnot picked up excessive fluorescent material (10 or more fluorescentspecks observed in a 4 inch/10cm circle), when tested utilizing thesame method used for applying developer to hardware?

    YES NO NA

  • 7/25/2019 Ac7114-1 Rev h Audit Criteria for Nondestructive Testing Facility Penetrant Survey-1

    9/33

    PRI AC7114/1 Rev. H - 9 -

    Compliance Assessment Guidance: For the purpose of this check,recycled refersto any material that is drawn from a source wheresome or all of the remaining material is returned to that source. Thetest panel selected shall be nonfluorescent and non-reflective and have

    a working surface color which provides a good contrast with thedeveloper powder. Note: some spray systems actually returnmaterial to the source. The auditor shall verify that a system labeled asnotrecycled is as such in actual performance. The holding tanks fordry developer shall be verified for lack of fluorescent material when

    possible.

    5.6.3 Does the check ensure that the developer is not caked? (applicable torecycled and closed systems)

    YES NO

    5.6.4 Are records of this test on file and do they indicate acceptable results? YES NO

    Compliance Assessment Guidance: Records indicating OK or someother check mark are acceptable for this test.

    5.6.5 Did the facility properly demonstrate this check? YES NO

    5.7 Concentration of Aqueous Developer (Soluble [Form b] & Suspended[Form c])

    SECTION NA

    5.7.1 Is the concentration of in-use Form b and Form c developer beingchecked at least weekly or prior to use?

    YES NO

    5.7.2 Is a hydrometer and a specific gravity chart being used for this test? YES NO

    5.7.3 Are the acceptance limits in accordance with manufacturersinstructions?

    YES NO

    5.7.4 Are records of this test on file and do they indicate acceptable results? YES NO

    Compliance Assessment Guidance: Records indicating OK or someother check mark are acceptable.

    5.7.5 Did the facility properly demonstrate this check? YES NO

    5.8 Characteristics of Aqueous Developer (Soluble [Form b] & Suspended[Form c])

    SECTION NA

    5.8.1 Is the aqueous developer checked at least daily or prior to use forfluorescence by immersing a clean 3x10 inch (7.5x25 cm) aluminumpanel into the developer?

    YES NO

    5.8.2 Does the aqueous developer uniformly wet the test panel? YES NO

    5.8.3 Is the aqueous developer coated panel free from fluorescentcontamination?

    YES NO

    5.8.4 Are records of this test on file and do they indicate acceptable results? YES NO

  • 7/25/2019 Ac7114-1 Rev h Audit Criteria for Nondestructive Testing Facility Penetrant Survey-1

    10/33

    PRI AC7114/1 Rev. H - 10 -

    Compliance Assessment Guidance: Records indicating OK or someother check mark are acceptable.

    5.8.5 Did the facility properly demonstrate this check? YES NO

    5.8.6 If Aqueous soluble developers were used in combination with type I,Method A penetrant, was specific customer documentation or approvalon file?

    YES NO NA

    5.9 Penetrant System Performance

    5.9.1 Has the supplier performed an initial check to establish a baseline foreach known defect standard and material in use?

    YES NO

    Compliance Assessment Guidance: The supplier shall perform an initialcheck with all new materials to establish a baseline for each known

    defect standard and material in use. This check shall be performedutilizing a color photograph approximately 1:1, whose background doesnot interfere with the comparison or by direct comparison of usedversus unused materials; to record the baseline performance. Therequirement is that the performance of the in use materials can not fallbelow that of the unused materials. Therefore a baseline performancemust be established to compare to the daily results. The requirementfor a photo does not apply when directly comparing new and in usematerials utilizing two identical test panels. (example: NiCr Panels)

    NOTE: The establishment of the baseline photo may be performed byan external source providing that the color photo is representative ofthe panel when run at the suppliers facility. The outside source shall

    process the panel utilizing the same penetrant materials and testparameters as the supplier and produce a color photo of approximately1:1 scale.

    5.9.2 Is the penetrant system performance checked at least daily or prior touse?

    YES NO

    Compliance Assessment Guidance: The check is intended to beperformed at least prior to use, that is if the system is not used everyday, the check would not be required to be performed every day. Thischeck includes all in use materials, recycled and not recycled howevera change in the penetrant application method from the same containershall not necessitate another system performance check. Forexample: brush on, electrostatic application, dipping, and pouring fromone container would not require separate checks for the same

    penetrant from the same container.

    NOTE: This includes materials sprayed to waste but is not applicableto Method C penetrant utilized from sealed aerosol containers.

    5.9.3 Are records of this test on file and do they indicate acceptable results? YES NO

    Compliance Assessment Guidance: Results of the system performance

  • 7/25/2019 Ac7114-1 Rev h Audit Criteria for Nondestructive Testing Facility Penetrant Survey-1

    11/33

    PRI AC7114/1 Rev. H - 11 -

    test, utilizing the in-use materials, must indicate the same number andappearance (e.g. size, etc.) of the flaws detected originally when thebaseline was established. Example: If 4 stars of a certain size andappearance are originally detected, then the same number andappearance of the stars is expected each subsequent test and the

    actual number must be recorded. This comparison shall be madeutilizing a color photograph (approximately 1:1, whose back grounddoes not interfere with the comparison) or by direct comparison of usedversus unused materials.

    5.9.4 Did the facility properly demonstrate this check? YES NO

    Compliance Assessment Guidance: The objective of the systemperformance test is that it be used as a comparison tool. Thatcomparison day to day may either be performed utilizing the samesystem parameters as the original (i.e. materials, dwell times,temperatures, etc), or may be processed in the same manner as

    production components, but the method of processing shall beproceduralised. The system performance test shall be witnessed by theauditor with all aspects performed properly.

    5.9.5 Does the procedure address the actions to be taken when thesensitivity or performance of the in-use materials falls below theperformance of the unused materials?

    YES NO

    5.10 Known Defect Standard Maintenance / Degradation

    5.10.1 Are there maintenance procedures in place that assure that cleaning ofthe known defect standards, between usage, is adequate and that

    physical changes in the standard that make it unsuitable for use, canbe detected?

    YES NO

    Compliance Assessment Guidance: The procedure must define what isactually done by the supplier. The auditor shall require that thetechnician demonstrate that the panel or test standard is clean andindication free prior to performance of the test. Unless defined bysupplemental requirements, this does not require application ofdeveloper.

    5.10.2 Is a degradation check performed for the known defect standard atleast annually?

    YES NO NA

    Compliance Assessment Guidance: This check must be performed bycomparing the original results obtained utilizing all new materials to thecurrent results, also using all new materials. This check is a check ofthe defect standard, not of the system. It shall be performed utilizingmeasurements Any suitable method may be used to measure theindications (e.g. inscribed circle, longest length) but the method usedshall be defined and repeated for the degradation measurement.. Thedegradation check is not applicable if the defect standard is replacedbefore the requirement comes due or for known defect standardswhere a direct comparison of new versus in-use materials is made on adaily basis.

  • 7/25/2019 Ac7114-1 Rev h Audit Criteria for Nondestructive Testing Facility Penetrant Survey-1

    12/33

    PRI AC7114/1 Rev. H - 12 -

    5.10.3 Is the stated tolerance for the degradation check +/- 30% of thebaseline measurement?

    YES NO NA

    5.11 Penetrant Contamination Check (Reused) SECTION NA

    5.11.1 Is a visual assessment performed on the penetrant for contamination atleast daily or prior to use?

    YES NO

    Compliance Assessment Guidance: This test may be performed in anyway that allows contamination to be detected in the penetrant material,and the method used shall be specified in the suppliers procedure.Determine if any of the following conditions are evident; precipitates,waxy deposits, white coloration, separation of constituents, surfacescum, or any other evidence of contamination or breakdown. When anyof the above conditions are detected the material shall be discarded ormodified in accordance with the manufacturers instructions.

    5.11.2 Are records of this check on file and do they indicate acceptableresults?

    YES NO

    Compliance Assessment Guidance: Records indicating OK or someother check mark are acceptable.

    5.12 Dryer Controls SECTION NA

    5.12.1 Is the dryer oven controlled to a tolerance of +/-15 F (8.3C) or better? YES NO

    Compliance Assessment Guidance: It is expected that the auditor shallverify the performance is acceptable via the calibration certificate. Thecalibration certificate at a minimum shall indicate the set point, themaximum temperature reached during a cycle, and the minimumtemperature reached during a cycle at each set temperature used forcalibration. In addition the variation shall be recorded from the set pointand shall meet the required tolerance. This is a calibration oftemperature variation with time and does not require a temperatureuniformity survey (TUS). The dryer temperature shall not vary morethan +/-15 F (8.3C) from the set value. The minimum test period usedfor the calibration shall be representative of the duration of a dryingcycle.

    5.12.2 Is the temperature controlling device calibrated at a minimum of one

    point within its range of use?

    YES NO

    Compliance Assessment Guidance: If the temperature controllingdevice is to be calibrated at one point, then that point shall be the pointof use for that dryer. If multiple points are used, then they shallencompass the range of use for that dryer. Example: The dryer shouldnot be calibrated at 130F (54C), 140F (60C) and 150F (66C) andthen used at 160F (71C).

    5.12.3 Is the calibration performed at least quarterly or with adequatedocumentation, extended to no more than every 6 months?

    YES NO

  • 7/25/2019 Ac7114-1 Rev h Audit Criteria for Nondestructive Testing Facility Penetrant Survey-1

    13/33

    PRI AC7114/1 Rev. H - 13 -

    Compliance Assessment Guidance: The calibration shall be performedat least quarterly or extended as allowed by the governing standard butnot to exceed 6 months and only when 4 consecutive acceptablecalibrations and adequate documentation is available to support the

    decreased frequency.

    5.12.4 Is the calibration current? YES NO

    5.12.5 Is there a temperature indicator for the dryer and is it calibrated to atolerance of +/-10 F (5.6C ) or better?

    YES NO

    5.12.6 Is the temperature indicator calibrated at a minimum of one point withinits range of use?

    YES NO

    Compliance Assessment Guidance: If the indicator is to be calibratedat one point, then that point shall be the point of use for that dryer. If

    multiple points are used, then they shall encompass the range of usefor that dryer. Example: The dryer should not be calibrated at 130F(54C), 140F (60C) and 150F (66C) and then used at 160F (71C).

    5.12.7 Is the calibration performed at least quarterly or with adequatedocumentation, extended to no more than every 6 months?

    YES NO

    Compliance Assessment Guidance: The calibration shall be performedat least quarterly or extended as allowed by the governing standard butnot to exceed 6 months and only when 4 consecutive acceptablecalibrations and adequate documentation is available to support thedecreased frequency.

    5.12.8 Is the calibration current? YES NO

    5.13 Black Light Meters and Measurements

    5.13.1 Is a digital meter, with a spectral sensitivity from 320-400 nm, utilized toverify black light intensity?

    YES NO

    5.13.2 Is calibration of the meter being performed at least semi-annually or,with adequate documentation, extended as allowed by the governingstandard but not to exceed 12 months when 4 consecutive acceptablecalibrations and adequate documentation is available to support thedecreased frequency?

    YES NO

    5.13.3 Is calibration performed at a minimum of three points to establishlinearity?

    YES NO

    Compliance Assessment Guidance: No specific values need to beused.

    5.13.4 Is the light meter accurate to within 5% of the standard reading? YES NO

    Compliance Assessment Guidance: Note that the value of 5% is inreference to a comparison of the light meter being calibrated to another

  • 7/25/2019 Ac7114-1 Rev h Audit Criteria for Nondestructive Testing Facility Penetrant Survey-1

    14/33

    PRI AC7114/1 Rev. H - 14 -

    calibrated (traceable to NIST or other National Standard) light meter. Itis not meant to represent a true 5% calibration tolerance.

    5.13.5 Is the calibration current? YES NO

    5.13.6 Is the black light meter traceable to calibration data? YES NO

    Compliance Assessment Guidance: To meet this requirement, themeter shall be serialized, stickered or otherwise identifiable to thecalibration data.

    5.13.7 Is the black light intensity of the inspection light(s) checked at leastdaily, prior to use and after equipment maintenance?

    YES NO

    5.13.8 Is the minimum acceptable limit 1200 W/cm2at 15 inches (38cm)? YES NO

    Compliance Assessment Guidance: Borescope intensity shall be a

    minimum of 1000 W/cm2 at the inspection surface as measured fromthe inspection distance.

    5.13.9 Are records of this check on file and do they indicate acceptableresults?

    YES NO

    Compliance Assessment Guidance: The actual light intensity shall berecorded for this check; OK or some other mark is not acceptable.

    5.13.10 Did the facility properly demonstrate this check? YES NO

    Compliance Assessment Guidance: The auditor shall observe thecheck of the black light intensity being conducted.

    5.13.11 Are the black light reflectors and filters clean and in good condition? YES NO

    5.13.12 Where battery powered lamps are used does the procedure addressconsistency of U/V output and the controls used to ensure adequateoutput during all inspection activity?

    YES NO NA

    Compliance Assessment Guidance: The procedure shall address thecheck requirements for battery powered sources.

    5.14 White Light Meters and Measurements

    5.14.1 Is the calibration of the meter performed at least semi-annually or, withadequate documentation, extended as allowed by the governingstandard but not to exceed 12 months when 4 consecutive acceptablecalibrations and adequate documentation is available to support thedecreased frequency?

    YES NO

    5.14.2 Is calibration performed at a minimum of three points to establishlinearity?

    YES NO

    Compliance Assessment Guidance: No specific values need to beused.

  • 7/25/2019 Ac7114-1 Rev h Audit Criteria for Nondestructive Testing Facility Penetrant Survey-1

    15/33

    PRI AC7114/1 Rev. H - 15 -

    5.14.3 Is the light meter accurate to within 5% of the standard reading? YES NO

    Compliance Assessment Guidance: Note that the value of 5% is inreference to a comparison of the light meter being calibrated to another

    calibrated (traceable to NIST or other National Standard) light meter. Itis not meant to represent a true 5% calibration tolerance.

    5.14.4 Is the calibration current? YES NO

    5.14.5 Is the white light meter traceable to calibration data? YES NO

    Compliance Assessment Guidance: To meet this requirement, themeter shall be serialized, stickered or otherwise identifiable to thecalibration data.

    5.14.6 Is ambient white light level in the inspection area, controlled for Type I

    penetrant, not to exceed 2 ft candles (20 lux) as measured at theexamination surface?

    YES NO

    Compliance Assessment Guidance: This measurement is taken in thedarkened inspection area to include the examination surface with theblack lights operating in the fixed position of 15 inches (38 cm).

    5.14.7 Is this check being performed at least weekly, prior to use and afterequipment maintenance?

    YES NO

    5.14.8 Are records of this check on file and do they indicate acceptableresults?

    YES NO

    Compliance Assessment Guidance: The actual light intensity shall berecorded for this check, OK or some other mark is not acceptable.

    5.14.9 Did the facility properly demonstrate this check? YES NO

    5.14.10 Is 100 ft-candles/1000 lux, or more, available at the part surface forvisual verification of penetrant indications?

    YES NO NA

    Compliance Assessment Guidance: This question shall be applicablewhere accept/reject decisions rely on the use of white light (otherwiseNA)

    5.14.11 Is the white light checked and documented at least weekly or prior touse?

    YES NO NA

    Compliance Assessment Guidance: The actual light intensity shall berecorded for this check; OK or some other mark is not acceptable.

    5.14.12 Did the facility properly demonstrate this check? YES NO NA

    5.15 Timing Devices

    5.15.1 Are timing devices that are used to monitor penetrant, emulsifier anddeveloper dwell times calibrated at a minimum of one point at least

    YES NO

  • 7/25/2019 Ac7114-1 Rev h Audit Criteria for Nondestructive Testing Facility Penetrant Survey-1

    16/33

    PRI AC7114/1 Rev. H - 16 -

    annually using a calibration traceable to NIST or other NationalStandard?

    5.16 Water Temperature Monitor

    5.16.1 Are water temperature indicators calibrated at a minimum of one pointin the range of use at least semi-annually using a calibration traceableto NIST or other National Standard? NA if only Method C is utilized.

    YES NO NA

    Compliance Assessment Guidance: The indicator (gage) utilized tocontrol this process must have a measurement range sufficient to

    provide adequate resolution of the desired readings.

    5.17 Pressure Monitors

    5.17.1 Are water pressure indicators calibrated at a minimum of one point inthe range of use at least semi-annually using a calibration traceable to

    NIST or other National Standard? NA if only Method C is utilized.

    YES NO NA

    Compliance Assessment Guidance: The indicator (gage) utilized tocontrol this process must have a measurement range sufficient to

    provide adequate resolution of the desired readings.

    5.17.2 Are air pressure indicators calibrated at a minimum of one point in therange of use at least semi-annually using a calibration traceable toNIST or other National Standard?

    YES NO NA

    Compliance Assessment Guidance: The indicator (gage) utilized tocontrol this process must have a measurement range sufficient to

    provide adequate resolution of the desired readings.

    6. COMPLIANCE

    Compliance Assessment Guidance: The supplier shall performrepresentative inspections from current production to determinecompliance with these requirements. These parts shall be selected torepresent a variety of Nadcap participating customer requirements andseveral different types of processing equipment if more than one

    penetrant line is in use at this facility. Borrowing hardware from onecompany to another to demonstrate compliance is not acceptable;unless there is objective evidence the design authority has authorizedmovement of the hardware to the company undergoing the audit.Investigate the purchase order specifications to identify any customerunique acceptance, process control, or procedure requirementsapplicable to these parts. Witness the facilitys ability to performinspection of these parts in accordance with the requirements.

    It is expected that the auditor shall review and follow traceability fromthe purchase order to completion of the NDT process or finalcertification by reviewing as a minimum the items listed below. Theauditor shall verify that the process applied complies with the

    procedural system established by the company in terms of flow down,customer specification retrieval (as applicable), inspection

  • 7/25/2019 Ac7114-1 Rev h Audit Criteria for Nondestructive Testing Facility Penetrant Survey-1

    17/33

    PRI AC7114/1 Rev. H - 17 -

    identification, etc.

    a) Purchase Orderb) Drawing (if applicable)c) Work Order

    d) Process Specifications and Revisione) Internal Procedures and Revisionf) Lot / Batch / Serial Number:g) Applicable Procedures / Techniques / Work Instructionsh) Travelers / Route Cardsi) Inspection Log

    6.1 Provide the following documentation for auditor review for each of theparts being tested during the compliance portion of the audit: Note: EC-LR materials shall not leave the suppliers facility.

    a. Copy of completed technique card or supplier procedure with

    evidence of customer approval if required.

    b. Copy of the applicable shop traveler or work order.

    c. If neither a or b above indicate the process and acceptance standardand their revision status, then the document that shows the link to thisdata shall also be reviewed.

    A. Compliance YES NO NA

    Customer End User

    Purchase Order Number:

    Work Order Number:

    Lot / Batch / Serial Number:

    Process Specification(s): Revision(s)

    Internal Procedure(s): Revision(s)

  • 7/25/2019 Ac7114-1 Rev h Audit Criteria for Nondestructive Testing Facility Penetrant Survey-1

    18/33

    PRI AC7114/1 Rev. H - 18 -

    Acceptance Specification(s) / Drawing Notes: Revision(s) Class and/or Grade(if applicable)

    Part Description:

    Part Number: Revision

    Penetrant Type: Method: Sensitivity Level:

    Developer Form:

    Inspectors Level(s):

    B. Compliance YES NO NA

    Customer End User

    Purchase Order Number:

    Work Order Number:

    Lot / Batch / Serial Number:

    Process Specification(s): Revision(s)

  • 7/25/2019 Ac7114-1 Rev h Audit Criteria for Nondestructive Testing Facility Penetrant Survey-1

    19/33

    PRI AC7114/1 Rev. H - 19 -

    Internal Procedure(s): Revision(s)

    Acceptance Specification(s) / Drawing Notes: Revision(s) Class and/or Grade(if applicable)

    Part Description:

    Part Number: Revision

    Penetrant Type: Method: Sensitivity Level:

    Developer Form:

    Inspectors Level(s):

    C. Compliance YES NO NA

    Customer End User

    Purchase Order Number:

    Work Order Number:

  • 7/25/2019 Ac7114-1 Rev h Audit Criteria for Nondestructive Testing Facility Penetrant Survey-1

    20/33

    PRI AC7114/1 Rev. H - 20 -

    Lot / Batch / Serial Number:

    Process Specification(s): Revision(s)

    Internal Procedure(s): Revision(s)

    Acceptance Specification(s) / Drawing Notes: Revision(s) Class and/or Grade(if applicable)

    Part Description:

    Part Number: Revision

    Penetrant Type: Method: Sensitivity Level:

    Developer Form:

    Inspectors Level(s):

  • 7/25/2019 Ac7114-1 Rev h Audit Criteria for Nondestructive Testing Facility Penetrant Survey-1

    21/33

    PRI AC7114/1 Rev. H - 21 -

    6.2 If applicable and in conjunction with the preceding, the supplier shallalso provide an additional three paperwork compliance jobs across theNadcap User base. Provide the following documentation for auditorreview for each of these jobs: Note: The goal is to provide at least one

    set of compliance documentation for each of the suppliers NadcapUser customer base. If all of the suppliers Nadcap customers wererepresented in the actual compliance audits, then this section is NA.Note: EC-LR materials shall not leave the suppliers facility.

    a. Copy of completed technique card or supplier procedure withevidence of customer approval if required.

    b. Copy of completed shop traveler or work order.

    c. If neither a or b above indicate the process and acceptance standardand their revision status, then the document that shows the link to this

    data shall also be reviewed.

    NA

    A. Compliance YES NO

    Customer End User

    Purchase Order Number:

    Work Order Number:

    Lot / Batch / Serial Number:

    Process Specification(s): Revision(s)

    Internal Procedure(s): Revision(s)

    Acceptance Specification(s) / Drawing Notes: Revision(s) Class and/or Grade(if applicable)

  • 7/25/2019 Ac7114-1 Rev h Audit Criteria for Nondestructive Testing Facility Penetrant Survey-1

    22/33

    PRI AC7114/1 Rev. H - 22 -

    Part Description:

    Part Number: Revision

    Penetrant Type: Method: Sensitivity Level:

    Developer Form:

    Inspectors Level(s): Date Performed:

    B. Compliance YES NO NA

    Customer End User

    Purchase Order Number:

    Work Order Number:

    Lot / Batch / Serial Number:

    Process Specification(s): Revision(s)

    Internal Procedure(s): Revision(s)

  • 7/25/2019 Ac7114-1 Rev h Audit Criteria for Nondestructive Testing Facility Penetrant Survey-1

    23/33

    PRI AC7114/1 Rev. H - 23 -

    Acceptance Specification(s) / Drawing Notes: Revision(s) Class and/or Grade(if applicable)

    Part Description:

    Part Number: Revision

    Penetrant Type: Method: Sensitivity Level:

    Developer Form:

    Inspectors Level(s): Date Performed:

    C. Compliance YES NO NA

    Customer End User

    Purchase Order Number:

    Work Order Number:

    Lot / Batch / Serial Number:

    Process Specification(s): Revision(s)

  • 7/25/2019 Ac7114-1 Rev h Audit Criteria for Nondestructive Testing Facility Penetrant Survey-1

    24/33

    PRI AC7114/1 Rev. H - 24 -

    Internal Procedure(s): Revision(s)

    Acceptance Specification(s) / Drawing Notes: Revision(s) Class and/or Grade

    (if applicable)

    Part Description:

    Part Number: Revision

    Penetrant Type: Method: Sensitivity Level:

    Developer Form:

    Inspectors Level(s): Date Performed:

  • 7/25/2019 Ac7114-1 Rev h Audit Criteria for Nondestructive Testing Facility Penetrant Survey-1

    25/33

    PRI AC7114/1 Rev. H - 25 -

    6.3 Customer Specifications

    Compliance Assessment Guidance: The auditor shall verify complianceto each of these requirements by direct observation.

    6.3.1 Is the facility using the appropriate specifications and revisions asrequired by the contract document?

    YES NO

    Compliance Assessment Guidance: This will be confirmed by theauditor by reviewing the customer P.O. and End User drawing ifavailable.

    6.3.2 If the facility is working to any deviations, have they been approved bythe customer?

    YES NO NA

    Compliance Assessment Guidance: The auditor will confirm

    documented customer approval; this may be in the form of a writtencommunication, technique approval or by a formal deviation from theEnd User.

    6.3.3 Were the applicable requirements, procedures and acceptance criteriamade available to the penetrant inspectors?

    YES NO

    Compliance Assessment Guidance: Work instructions/routings/shoptravelers/techniques/procedures, acceptance criteria and uniquecustomer requirements shall be available to the inspectors. The termavailable means that these documents are within easy access by thetechnicians. These documents are available and are used to performthe inspections, not used from memory.

    6.3.4 If written procedures, and or techniques, require review and/orapproval by the Cognizant Engineering Organization, is that approvalevident?

    YES NO NA

    Compliance Assessment Guidance: This shall be investigated andconfirmed by the auditor by reviewing the last ten part numbersshipped.

    6.3.5 If applicable, are part specific drawing inspection and/or acceptancenotes being flowed down and followed by the inspectors?

    YES NO NA

    6.3.6 Was there evidence of compliance to the document control procedure? YES NO

    6.3.7 If applicable, are all specifications, drawings, procedures or techniquechanges authorized? Auditor to confirm approval.

    YES NO NA

    Compliance Assessment Guidance: The auditor shall review thedocuments for any unauthorized changes.

    6.4 Inspection Process

    Compliance Assessment Guidance: The auditor shall verify compliance

  • 7/25/2019 Ac7114-1 Rev h Audit Criteria for Nondestructive Testing Facility Penetrant Survey-1

    26/33

    PRI AC7114/1 Rev. H - 26 -

    to each of these requirements by direct observation.

    6.4.1 Do applicable Work Orders/Shop Travelers and Technique Cardscontain or provide traceability to the following as a minimum?

    6.4.1.1 Part number / Material Identification? YES NO

    6.4.1.2 Specification reference? YES NO

    6.4.1.3 Inspection acceptance requirements? YES NO

    6.4.1.4 Correct sequence of operations? YES NO

    6.5 Pre-Cleaning

    Compliance Assessment Guidance: The auditor shall verify complianceto each of these requirements by direct observation.

    6.5.1 Are parts being cleaned in accordance with applicable workinstructions/procedure?

    YES NO

    Compliance Assessment Guidance: There shall be a procedure thataddresses the cleaning materials, sequence, times, concentrations andtemperatures in accordance with customer requirements.

    6.5.2 Were all parts clean? YES NO

    Compliance Assessment Guidance: Parts shall be cleaned inaccordance with customer requirements. The auditor shall confirm thecleanliness of the parts prior to inspection by whatever means areavailable and reasonable.

    6.6 Pre-Penetrant Etching SECTION NA

    Compliance Assessment Guidance: The auditor shall verify complianceto each of these requirements by direct observation.

    6.6.1 Is there a procedure that addresses etching materials, sequence,times, concentrations and temperatures in accordance with customerrequirements?

    YES NO

    6.6.2 If required, were all parts etched? YES NO NA

    Compliance Assessment Guidance: All parts shall be properly etched ifapplicable. The auditor shall determine applicability based on a reviewof the purchase order, specification, customer instruction, etc.

    6.6.3 If required, is an outside source used? If so, identify the source nameand location:

    ______________________________________________

    YES NO NA

    6.6.4 If required, was the source performing etch (internal or external) YES NO NA

  • 7/25/2019 Ac7114-1 Rev h Audit Criteria for Nondestructive Testing Facility Penetrant Survey-1

    27/33

    PRI AC7114/1 Rev. H - 27 -

    approved by the customer?

    6.7 Penetrant Application

    Compliance Assessment Guidance: The auditor shall verify compliance

    to each of these requirements by direct observation.

    6.7.1 Are components, penetrant, and ambient temperatures maintainedwithin the range of 50F (10C) to 125F (52C)?

    YES NO

    Compliance Assessment Guidance: If the auditor or supplier suspectsthat the temperature is out of this range, the temperature must beverified.

    6.7.2 Is a calibrated timing device used to monitor penetrant dwell time? YES NO

    6.7.3 Are parts dry prior to penetrant application? YES NO

    Compliance Assessment Guidance: It is expected that a dry conditionbe confirmed by the auditor.

    6.7.4 Are penetrant application operations controlled to ensure completecoverage of areas to be inspected?

    YES NO

    Compliance Assessment Guidance: It is expected that adequate andcomplete coverage be confirmed by the auditor by whatever means issuitable.

    6.7.5 Was the correct method, type, and level of penetrant used? YES NO

    Compliance Assessment Guidance: The auditor shall confirm theseissues by researching the appropriate customer specifications anddocumentation.

    6.7.6 Were the parts properly drained to prevent pooling? YES NO

    6.7.7 Was the correct penetrant dwell time utilized as specified by customerrequirements?

    YES NO

    6.7.8 If parts exceeded the maximum dwell time, was penetrant reapplied? YES NO NA

    6.8 Use of Emulsifier SECTION NA

    Compliance Assessment Guidance: The auditor shall verify complianceto each of these requirements by direct observation.

    6.8.1 Is a calibrated timing device used to monitor emulsification time?

    List Emulsification Time .

    YES NO

    Compliance Assessment Guidance: The emulsification time used shallbe recorded by the auditor.

  • 7/25/2019 Ac7114-1 Rev h Audit Criteria for Nondestructive Testing Facility Penetrant Survey-1

    28/33

    PRI AC7114/1 Rev. H - 28 -

    6.8.2 Are methods of application and removal of emulsifiers properlycontrolled?

    YES NO

    Compliance Assessment Guidance: It is expected that the auditor shall

    watch these operations carefully to detect discrepancies.

    6.8.3 Was the correct emulsifier utilized? YES NO

    6.8.4 If hydrophilic emulsifier was used, were the pre rinse parameterscontrolled? (e.g., temperature, pressure, etc.)

    YES NO NA

    Compliance Assessment Guidance: It is expected that parameterssuch as: temperature, pressure, etc, be observed and evaluated bythe auditor.

    6.8.5 Was the entire part flooded with water prior to reaching the maximum

    emulsification time?

    YES NO

    6.9 Penetrant Removal

    Compliance Assessment Guidance: The auditor shall verify complianceto each of these requirements by direct observation. Some questionswill be NA if only method C is utilized.

    6.9.1 Is the rinse water controlled to provide a coarse spray? YES NO NA

    6.9.2 Is the rinse water temperature controlled within the range of 50F

    (10C) to 100 F (38C)?

    YES NO NA

    6.9.3 Is the rinse water pressure controlled to 40 psi (275 kPa) maximum? YES NO NA

    6.9.4 Is the water pressure and temperature monitored at least daily or priorto use?

    YES NO NA

    Compliance Assessment Guidance: This check does not need to berecorded, however the auditor is expected to verify compliance bywitnessing the check being performed without prompting.

    6.9.5 Is penetrant removal, for manual stations, verified under appropriatelighting?

    YES NO NA

    Compliance Assessment Guidance: NA shall only be used where anautomated system is used.

    6.9.6 Is over and under washing/removal controlled? YES NO

    Compliance Assessment Guidance: Removal, using water, removersand solvents shall be controlled. The appearance of the part at finalinspection may also provide evidence of correct removal.For manual systems the auditor shall observe all removal processes.Wash distance of 12 inches (30 cm) where possible, application ofremovers and the use of solvents are included as this issue applies to

  • 7/25/2019 Ac7114-1 Rev h Audit Criteria for Nondestructive Testing Facility Penetrant Survey-1

    29/33

    PRI AC7114/1 Rev. H - 29 -

    all washing and removal methods.Automated lines may not be designed to work at 12 inches or more andverification of distance may not be possible by the auditor. It isexpected that automated lines are substantiated for product or producttypes and thereafter regular controls are carried out to ensure

    continued adequate performance.

    6.9.7 If used, is the air pressure on the hydro air nozzle controlled to 25 psi(170 kPa) maximum and monitored at least daily or prior to use?

    YES NO NA

    Compliance Assessment Guidance: This check does not need to berecorded, however the auditor is expected to verify compliance bywitnessing the check being performed without prompting.

    6.10 Developer Application SECTION NA

    Compliance Assessment Guidance:

    Developer is required unless specific and clear customer allowancehas been given. When developer is not used, the auditor shall providean explanation. Section 6.10 is required to be answered in all caseswhere developer is used. In addition, the following sections 6.11, 6.12and 6.13 will be required when that particular material is used;otherwise those sections not used are NA with no explanation beingrequired.

    The auditor shall verify compliance to each of these requirements bydirect observation.

    6.10.1 Was the developer utilized in accordance with the applicable customerrequirements?

    YES NO

    Compliance Assessment Guidance: The auditor shall confirm this issueby researching the appropriate specifications and documentation.

    6.10.2 Is a calibrated timing device used to monitor developer dwell time? YES NO

    Compliance Assessment Guidance: A timing device shall be calibratedat a minimum of one point at least annually. Any calibration traceableto an acceptable national source is acceptable for these timing devices.

    6.10.3 Was the developing time in accordance with applicable customer

    requirements?

    YES NO

    Compliance Assessment Guidance: The auditor shall confirm this issueby researching the appropriate specifications and documentation.

    6.10.4 When applying developer, are the parts completely covered? YES NO

    Compliance Assessment Guidance: When applying developer, theparts or the areas to be inspected as applicable shall be completelycovered with a thin and uniform developer layer. This issue shall beobserved and confirmed by the auditor.

  • 7/25/2019 Ac7114-1 Rev h Audit Criteria for Nondestructive Testing Facility Penetrant Survey-1

    30/33

    PRI AC7114/1 Rev. H - 30 -

    6.10.5 Are dwell times (in minutes) used by the supplier within the rangesnoted below? Please note all used. An explanation is not required fornot used.

    Compliance Assessment Guidance: 20 to 120 minutes is between 10to 240 minutes. The auditor shall include materials utilized for the wipe-off technique when completing section 6.10.5. If form d is utilized forwipe-off only, the auditor should answer yes with an explanation.

    Form: a Minimum Dwell Time: 10 Maximum Dwell Time: 240 YES NO NA

    Form: b Minimum Dwell Time: 10 Maximum Dwell Time: 120 YES NO NA

    Form: c Minimum Dwell Time: 10 Maximum Dwell Time: 120 YES NO NA

    Form: d Minimum Dwell Time: 10 Maximum Dwell Time: 60 YES NO NA

    6.11 Dry Developer (Form a) SECTION NA

    Compliance Assessment Guidance: The auditor shall verify complianceto each of these requirements by direct observation.

    6.11.1 Are all parts dried in accordance with customer requirements? YES NO

    Compliance Assessment Guidance: The auditor shall confirm this issueby researching the appropriate specifications and documentation. If adryer is used, the set temperature used shall take into considerationthe tolerance of the dryer.

    6.11.2 Are methods for applying developer and removing excess developerproperly controlled?

    YES NO

    6.11.3 If air was used to remove excess dry developer, was it controlled to 5psi (34 kPa) or less?

    YES NO NA

    6.12 Aqueous Developer (Form b & c) SECTION NA

    Compliance Assessment Guidance: The auditor shall verify complianceto each of these requirements by direct observation.

    6.12.1 Was the excess water removed from the parts prior to developer

    application?

    YES NO

    6.12.2 Are methods for applying developer and removing excess developerproperly controlled?

    YES NO

    Compliance Assessment Guidance: The auditor shall observe that theproper techniques are employed.

    6.12.3 Are all parts dried in accordance with customer requirements? YES NO

    Compliance Assessment Guidance: The auditor shall confirm this issueby researching the appropriate specifications and documentation. If a

  • 7/25/2019 Ac7114-1 Rev h Audit Criteria for Nondestructive Testing Facility Penetrant Survey-1

    31/33

    PRI AC7114/1 Rev. H - 31 -

    dryer is used, the set temperature used shall take into considerationthe tolerance of the dryer.

    6.13 Non-Aqueous Developer (Form d) SECTION NA

    Compliance Assessment Guidance: The auditor shall verify complianceto each of these requirements by direct observation. The auditor shallinclude the wipe-off technique when answering section 6.13.

    6.13.1 Are all parts dried in accordance with customer requirements? YES NO

    Compliance Assessment Guidance: The auditor shall confirm this issueby researching the appropriate specifications and documentation.

    6.13.2 Are methods for applying developer properly controlled? YES NO

    Compliance Assessment Guidance: The auditor shall observe that a

    thin even layer of developer is applied and is not in excess.

    6.14 Inspection Area

    Compliance Assessment Guidance: The auditor shall verify complianceto each of these requirements by direct observation.

    6.14.1 Are magnification/viewing aids available for evaluating indications? YES NO

    6.14.2 Are calibrated/verified measuring devices in accordance with customerrequirements utilized?

    YES NO

    Compliance Assessment Guidance: The auditor shall confirm this by

    researching the appropriate specifications and documentation.

    6.14.3 Is the Inspection area free of fluorescent material or othercontamination that may interfere with the inspection process?

    YES NO

    Compliance Assessment Guidance: This would include the area whereparts are evaluated, spills on the floor, tables or other areas wherepenetrant materials may come in contact with parts and baskets, trays,etc. used to hold parts. The auditor is expected to evaluate the area asa whole and determine if any detrimental contamination is present thatwould interfere with the inspection of parts.

    6.15 Evaluation

    Compliance Assessment Guidance: The following evaluation processshall be determined by direct observation, and by researching thespecifications, procedures and other customer documents that areapplicable and required to reach an educated answer to each question.

    6.15.1 For Type I inspection, did inspectors allow at least 1 minute to conditiontheir eyes to the darkness before conducting the inspection?

    YES NO

    Compliance Assessment Guidance: Inspectors shall allow at least 1minute to condition their eyes to the darkness when coming in from

  • 7/25/2019 Ac7114-1 Rev h Audit Criteria for Nondestructive Testing Facility Penetrant Survey-1

    32/33

    PRI AC7114/1 Rev. H - 32 -

    ambient lighting. Adaptation time after evaluation of the indicationunder white light is not required unless otherwise specified by thecustomer. A timing device is required for eye adaptation monitoring, butevidence of calibration is not required.

    6.15.2 Did the inspector adhere to the requirement of not wearingpermanently darkened or photo-chromatic eye glasses?

    YES NO

    Compliance Assessment Guidance: This is not intended to prohibit theuse of eyeglasses with lenses treated to absorb ultraviolet light.

    6.15.3 For Type I inspection, was the light intensity at the inspection surfaceat the time of this survey at least 1200 W/cm2?

    YES NO

    6.15.4 If required, did the inspector properly evaluate every indication andreject parts containing relevant indications that exceeded the

    acceptance limits?

    YES NO NA

    Compliance Assessment Guidance: The auditor is expected to confirmunderstanding by whatever means are required. This may be possibleby observation or, in some cases, may require the auditor to ask theinspector(s) questions regarding the criteria and how it is to be used.

    6.15.5 Were the indications wiped with a solvent dampened swab or brush inaccordance with customer requirements?

    YES NO NA

    Compliance Assessment Guidance: For the purpose of this checklist,the term swab shall include lint-free cloth.

    6.15.6 If redeveloped, was the redevelopment time, for acceptance, adequateto meet customer requirements?

    YES NO NA

    6.15.7 Were all rejectable indications marked on the part and/or appropriatelydocumented?

    YES NO NA

    6.15.8 Was all pertinent paperwork traceable to the part? YES NO

    6.15.9 Were the specific accept/reject criteria properly understood by theinspector?

    YES NO

    Compliance Assessment Guidance: The auditor is expected to confirm

    understanding by whatever means are required. This may be possibleby observation or, in some cases, may require the auditor to ask theinspector(s) questions regarding the criteria and how it is to be used.

    6.15.10 Did the accept/reject criteria properly reflect the customer criteria? YES NO

    Compliance Assessment Guidance: This will be confirmed by theauditor by reviewing the customer P.O and End User drawing ifavailable.

    6.16 Post Cleaning

  • 7/25/2019 Ac7114-1 Rev h Audit Criteria for Nondestructive Testing Facility Penetrant Survey-1

    33/33

    PRI AC7114/1 Rev. H - 33 -

    Compliance Assessment Guidance: The auditor shall verify complianceto this requirement by direct observation.

    6.16.1 If required was an adequate post process cleaning operation

    performed?

    YES NO NA

    6.16.2 After inspection and appropriate cleaning, were parts properlyprotected from corrosion or damage?

    YES NO NA

    6.17 Inspector Qualification

    6.17.1 Were the inspection personnel certified to the correct levels for thework they were performing?

    YES NO

    Compliance Assessment Guidance: This shall be verified by review ofthe certification records.

    6.17.2 Were all non-certified or in-training personnel observed performing PToperations properly supervised?

    YES NO NA

    Compliance Assessment Guidance: This shall be determined by directobservation.

    6.17.3 Were all personnel who require near vision correction to pass thevisual acuity exam wearing vision correction for detection andevaluation purposes during testing/inspections?

    YES NO NA

    Compliance Assessment Guidance: The auditor shall verify thatcorrective lenses were used if required. NA only where none of the

    inspection personnel witnessed working required correction lenses.

    6.18 Inspection Status

    6.18.1 Are symbols and/or marking methods used to denote penetrantinspection status?

    YES NO

    6.18.2 Were all inspection results properly recorded and traceable to theparts?

    YES NO