acclarent se inflation device catheter being used. • do not use air or a gas medium to inflate the...
TRANSCRIPT
Inflation Tubing Luer
Easy-to-read pressure gauge Luminescent Dial
Inflation Tubing
Convenient indicators
Black Fill Line
Black Triangle Fill Markers
Blue Vacuum Lines
Simple-to-operate unlock button
Knob
Unlock Button
®
Warnings and Precautions• Intended for single patient use only. DO NOT REUSE.• After use, the Inflation Device may be a potential biohazard. Handle
and dispose of in accordance with accepted hospital procedures.• Do not use a device where the integrity of the sterile packaging has
been compromised or if the device appears damaged.• Do not exceed the recommended maximum inflation pressure of the
balloon catheter being used.• Do not use air or a gas medium to inflate the balloon.• Not compatible with INSPIRA Air®
Balloon Dilation System.
ACCLARENT® SE Inflation DeviceThis inflation device is intended to inflate, deflate and monitor pressure in sinus balloon catheters. Designed for ease-of-use, this device is a convenient tool to use during balloon sinuplasty procedures.
Product code: SEID
Step 5: Deflate
1
Press and hold , then pull the knob back until the black fill line
aligns with the blue vacuum
lines, then release .
(can be done any time during the procedure)
Note: Rotating the knob counterclockwise will
decrease the pressure.
ACCLARENT® SE Inflation DeviceQuick Start Guide
®
Caution: US law restricts this device to sale by or on the order of a physician.This product can only be used by healthcare professionals. Please read and follow the Instruction for Use (IFU) prior to using for important information including contraindications, warnings, and proper directions. For more product information including warnings and precautions, review the appropriate product User’s Guide.Important information: Prior to use, refer to the instructions for use supplied with this device for indications, contraindications, side effects, warnings and precautions. © Acclarent, Inc. 2016. All rights reserved. 062867-161107
Product code: SEID
Step 1: PreparePlace the end of the inflation tubing into sterile solution.
A. Press and pull back on the knob to draw fluid.
B. Hold the device upright and push the knob in to remove excess air.
C. Pull back on the knob until the black fill line aligns with the black triangle fill markers, then release .
Step 3: Connect
Step 4: InflateRotate the knob clockwise to fill the balloon until target pressure is achieved.
1
1
1
Step 2: Fill
A CB