Advanced Pulmonary Mechanics during Mechanical Ventilation. Points of Discussion. Basics. Abnormalities. Equation of motion Dynamic Compliance Pressure-volume loop Flow-volume loop Work of breathing Lower and upper inflection points Hysterexis Intrathoracic Pressures. Air-leak - PowerPoint PPT PresentationTRANSCRIPT
![Page 1: Advanced Pulmonary Mechanics during Mechanical Ventilation](https://reader035.vdocument.in/reader035/viewer/2022062304/568143fa550346895db08c4e/html5/thumbnails/1.jpg)
Advanced Pulmonary Mechanics during Mechanical Ventilation
![Page 2: Advanced Pulmonary Mechanics during Mechanical Ventilation](https://reader035.vdocument.in/reader035/viewer/2022062304/568143fa550346895db08c4e/html5/thumbnails/2.jpg)
Points of Discussion
Basics Abnormalities Equation of motion Dynamic Compliance Pressure-volume loop Flow-volume loop Work of breathing Lower and upper
inflection points Hysterexis Intrathoracic
Pressures
Air-leak Air trapping Increased airway
resistance Inadequate flow support Inadequate sensitivity Atelectasis Inadequate PEEP Airway obstruction Over-distension
![Page 3: Advanced Pulmonary Mechanics during Mechanical Ventilation](https://reader035.vdocument.in/reader035/viewer/2022062304/568143fa550346895db08c4e/html5/thumbnails/3.jpg)
Tube + Spring Model
Elastic Forces
Resistive Forces
![Page 4: Advanced Pulmonary Mechanics during Mechanical Ventilation](https://reader035.vdocument.in/reader035/viewer/2022062304/568143fa550346895db08c4e/html5/thumbnails/4.jpg)
Static and Dynamic Pressures
time
Pressure
PEEP
PIP
Pplat
Alveolar Distending (recoil) Pressure difference (Pdis)
Flow-Resistive Pressure difference (Pres)
![Page 5: Advanced Pulmonary Mechanics during Mechanical Ventilation](https://reader035.vdocument.in/reader035/viewer/2022062304/568143fa550346895db08c4e/html5/thumbnails/5.jpg)
Resistive and Elastive Forces
DYNAMIC CHARACTERISTICS:dP = dV / Cdyn
DYNAMIC CHARACTERISTICS:dP = dV / Cdyn
RESISTANCE:dPresistive = R x Flow
RESISTANCE:dPresistive = R x Flow
STATIC COMPLIANCE:dPdistensive = dV / Cst
STATIC COMPLIANCE:dPdistensive = dV / Cst
dP = dPresist. + dP dist.
dP = R x Flow + dV / C st
![Page 6: Advanced Pulmonary Mechanics during Mechanical Ventilation](https://reader035.vdocument.in/reader035/viewer/2022062304/568143fa550346895db08c4e/html5/thumbnails/6.jpg)
Assessment of static P-V curveSuper-syringe method:
Stepwise inflation from a big syringe with multiply occlusions at each volumes to record recoil pressureTime consumingCumbersome to
performDifficult to standardizePatient must be
paralysed, connected to a special equipment
Great risk of oxygen desaturation
Volume
Pressure
![Page 7: Advanced Pulmonary Mechanics during Mechanical Ventilation](https://reader035.vdocument.in/reader035/viewer/2022062304/568143fa550346895db08c4e/html5/thumbnails/7.jpg)
Assessment of static P-V curveSlow Flow Single Inflation Method
Slow (5-10 lpm) inspiratory flow with large Vt and ZEEP
The inspiratory curve of the dynamic P-V loop closely approximates the static curve
The flow-resistive pressure component could be subtracted
Easy to perform, fast and relatively comfortable
Servillo: AJRCCM 1997
Lu: AJRCCM 1999 Pressure
Volume
LPIflex
UPIflex
inspir
ationStat
ic c
urve
![Page 8: Advanced Pulmonary Mechanics during Mechanical Ventilation](https://reader035.vdocument.in/reader035/viewer/2022062304/568143fa550346895db08c4e/html5/thumbnails/8.jpg)
FRC and PV Loop
FRC
VO
LU
ME
TLC
Negative Positive0
DISTENDING PRESSURE
Normal Compliance
FRC
![Page 9: Advanced Pulmonary Mechanics during Mechanical Ventilation](https://reader035.vdocument.in/reader035/viewer/2022062304/568143fa550346895db08c4e/html5/thumbnails/9.jpg)
Components of Pressure-Volume Loop
Components of Pressure-Volume Loop
Volume Volume (mL)(mL)
Insp
irat
ion
Expirat
ion
PIP
VT
PPawaw (cm H (cm H22O)O)
![Page 10: Advanced Pulmonary Mechanics during Mechanical Ventilation](https://reader035.vdocument.in/reader035/viewer/2022062304/568143fa550346895db08c4e/html5/thumbnails/10.jpg)
Pressure-Volume Loop(Type of Breath)
Pressure-Volume Loop(Type of Breath)
Controlled Assisted Spontaneous
Vol (m
l)
Paw
(cm H2O)
I: InspirationE: Expiration
I
E
E
E
II
![Page 11: Advanced Pulmonary Mechanics during Mechanical Ventilation](https://reader035.vdocument.in/reader035/viewer/2022062304/568143fa550346895db08c4e/html5/thumbnails/11.jpg)
PEEP and P-V LoopPEEP and P-V Loop
Volume (mL)
VT
PIPPIPPaw (cm H2O)
PEEPEEP P
![Page 12: Advanced Pulmonary Mechanics during Mechanical Ventilation](https://reader035.vdocument.in/reader035/viewer/2022062304/568143fa550346895db08c4e/html5/thumbnails/12.jpg)
Inflection PointsInflection Points
Pressure (cm H2O)
Volume (mL)
Upper Inflection Point
Lower Inflection Point
Upper Inflection Point: Represents pressure resulting in regional overdistension
Lower Inflection Point: Represents minimal pressure for adequate alveolar recruitment
![Page 13: Advanced Pulmonary Mechanics during Mechanical Ventilation](https://reader035.vdocument.in/reader035/viewer/2022062304/568143fa550346895db08c4e/html5/thumbnails/13.jpg)
Decreased ComplianceDecreased Compliance
Volu
me(m
l)
Pressure (cm HPressure (cm H22O)O)
NormalNormalPatientPatient
![Page 14: Advanced Pulmonary Mechanics during Mechanical Ventilation](https://reader035.vdocument.in/reader035/viewer/2022062304/568143fa550346895db08c4e/html5/thumbnails/14.jpg)
Lung Compliance Changes and the P-V Loop
Lung Compliance Changes and the P-V Loop
Volume (mL)Volume (mL)
PIP levels
Preset VT
PPawaw (cm H (cm H22O)O)
NormalNormal
Volume Targeted Ventilation
DecreasedDecreased
IncreasedIncreased
![Page 15: Advanced Pulmonary Mechanics during Mechanical Ventilation](https://reader035.vdocument.in/reader035/viewer/2022062304/568143fa550346895db08c4e/html5/thumbnails/15.jpg)
Lung Compliance Changes and the P-V Loop
Lung Compliance Changes and the P-V Loop
Volume (mL)Volume (mL)
Preset PIP
VT
levels
PPawaw (cm H (cm H22O)O)
NormalNormal
Pre
ssu
re T
arg
ete
d
Ven
tilatio
n
IncreasedIncreased
DecreasedDecreased
![Page 16: Advanced Pulmonary Mechanics during Mechanical Ventilation](https://reader035.vdocument.in/reader035/viewer/2022062304/568143fa550346895db08c4e/html5/thumbnails/16.jpg)
HysteresisHysteresis
Volume (ml)
Pressure (cm Pressure (cm HH22O)O)
Abnormal Hysteresis
Normal Hysteresis
![Page 17: Advanced Pulmonary Mechanics during Mechanical Ventilation](https://reader035.vdocument.in/reader035/viewer/2022062304/568143fa550346895db08c4e/html5/thumbnails/17.jpg)
Flow-Volume LoopFlow-Volume Loop
Volume (ml)Volume (ml)
Inspiration
Expiration
Flo
w (
L/m
in)
Flo
w (
L/m
in)
PEFR
FRC
PIFR
VT
![Page 18: Advanced Pulmonary Mechanics during Mechanical Ventilation](https://reader035.vdocument.in/reader035/viewer/2022062304/568143fa550346895db08c4e/html5/thumbnails/18.jpg)
Positive Pressure Ventilation: The Equation of Motion
In a passive subject, airway pressure represents the entire pressure (P) applied across the respiratory system.
The work required to deliver a tidal breath (Wb) = tidal volume (VT) x airway pressure
The pressure (P) associated with the delivery of a tidal breath is defined by the simplified equation of motion of the respiratory system (lungs & chest wall):
P = VT/CR+ VT/Ti x RR + PEEP total
Where CR = compliance of the respiratory system, Ti = inspiratory time and VT/Ti = Flow, RR = resistance of the respiratory system and PEEP total = the alveolar pressure at the end of expiration = external PEEP + auto (or intrinsic) PEEP, if any. Auto PEEP = PEEP total – P extrinsic (PEEP dialed in the ventilator) adds to the inspiratory pressure one needs to
generate a tidal breath.
P elastic P resistive P elastic
![Page 19: Advanced Pulmonary Mechanics during Mechanical Ventilation](https://reader035.vdocument.in/reader035/viewer/2022062304/568143fa550346895db08c4e/html5/thumbnails/19.jpg)
Work of BreathingWork of Breathing
A: Resistive Work B: Elastic Work A: Resistive Work B: Elastic Work
Pressure (cm H2O)
Volume (ml)
BBBB
AAAA
![Page 20: Advanced Pulmonary Mechanics during Mechanical Ventilation](https://reader035.vdocument.in/reader035/viewer/2022062304/568143fa550346895db08c4e/html5/thumbnails/20.jpg)
Work of BreathingWOB is a major source of caloric expenditure and
oxygen consumptionAppr. 70% to overcome elastic forces, 30% flow-
resistive workPatient work is a one of the most sensitive indicator of
ventilator dependencyComparison of Ventilator and Patient work is a useful
indicator during weaning processWOB may be altered by changes in compliance,
resistance, patient effort, level of support, PEEP, improper Ti, demand system sensitivity, mode setting
Elevated WOB may contraindicate the weaning process
![Page 21: Advanced Pulmonary Mechanics during Mechanical Ventilation](https://reader035.vdocument.in/reader035/viewer/2022062304/568143fa550346895db08c4e/html5/thumbnails/21.jpg)
WOB Measurements
PA
B C
D
E
VWOB = ∫0
ti P x Vdt
Elasic work: ABCAResistive work
Inspiratory: ADCAExpiratory: ACEA
![Page 22: Advanced Pulmonary Mechanics during Mechanical Ventilation](https://reader035.vdocument.in/reader035/viewer/2022062304/568143fa550346895db08c4e/html5/thumbnails/22.jpg)
Work of Breathing Measurements
WOB = ∫0
ti P x Vdt
Paw: Ventilator Work: The physical force required to move gas into the lung, represents the total work of the resp. system (patient + ventilator)
Peso: Patient Work: done by respiratory muscles, represents the pulmonary work of breathing
Paw-Ptr: Imposed Work by the Endotracheal tube
![Page 23: Advanced Pulmonary Mechanics during Mechanical Ventilation](https://reader035.vdocument.in/reader035/viewer/2022062304/568143fa550346895db08c4e/html5/thumbnails/23.jpg)
P-V Loop and WOB
P
V
P
V
P
VNormal ComplianceNormal Resistance
Normal ComplianceIncreased Resistance
Decreased ComplianceNormal Resistance
![Page 24: Advanced Pulmonary Mechanics during Mechanical Ventilation](https://reader035.vdocument.in/reader035/viewer/2022062304/568143fa550346895db08c4e/html5/thumbnails/24.jpg)
Work per breath is depicted as a pressure-volume area
Work per breath (Wbreath) = P x tidal volume (VT)
Wmin = wbreath x respiratory rate
Pressure Pressure Pressure
Vol
um
e
Vol
um
e
Vol
um
e
VT
WR = resistive work
WEL = elastic work
The total work of breathing can be partitioned between an elastic and resistive work. By analogy, the pressure needed to inflate a balloon through a straw varies; one needs to overcome the resistance of the straw and the elasticity of the balloon.
Work of Breathing
![Page 25: Advanced Pulmonary Mechanics during Mechanical Ventilation](https://reader035.vdocument.in/reader035/viewer/2022062304/568143fa550346895db08c4e/html5/thumbnails/25.jpg)
Intrinsic PEEP and Work of Breathing
Vol
um
e
VT
VT
FRCPressure
PEEPi
Dynamic Hyperinflation
PEEPi = intrinsic or auto PEEP; green triangle = tidal elastic work; red loop = flow resistive work; blue rectangle = work expended in offsetting intrinsic PEEP (an expiratory driver) during inflation
When present, intrinsic PEEP contributes to the work of breaking and can be offset by applying external PEEP.
![Page 26: Advanced Pulmonary Mechanics during Mechanical Ventilation](https://reader035.vdocument.in/reader035/viewer/2022062304/568143fa550346895db08c4e/html5/thumbnails/26.jpg)
++
++
Ventilator
₊ ₊
The Pressure and Work of Breathing can be Entirely Provided by the Ventilator (Passive Patient)
![Page 27: Advanced Pulmonary Mechanics during Mechanical Ventilation](https://reader035.vdocument.in/reader035/viewer/2022062304/568143fa550346895db08c4e/html5/thumbnails/27.jpg)
The Work of Breathing can be Shared Between the Ventilator and the Patient
PAW
PES
patient machine
time
AC mode
The ventilator generates positive pressure within the airway and the patient’s inspiratory muscles generate negative pressure in the pleural space.
Paw = Airway pressure, Pes= esophageal pressure
![Page 28: Advanced Pulmonary Mechanics during Mechanical Ventilation](https://reader035.vdocument.in/reader035/viewer/2022062304/568143fa550346895db08c4e/html5/thumbnails/28.jpg)
Work of breath
Resistive Work
Elastic Work of Lung
Elastic Work of Chest
Paw
Pes
Volume
Work to inflate the chest wall
Inflation Deflation
![Page 29: Advanced Pulmonary Mechanics during Mechanical Ventilation](https://reader035.vdocument.in/reader035/viewer/2022062304/568143fa550346895db08c4e/html5/thumbnails/29.jpg)
Relationship Between the Set Pressure Support Level and the Patient’s Breathing Effort
Carrey et al. Chest. 1990;97:150.
The changes in Pes (esophageal pressure) and in the diaphragmatic activity (EMG) associated with the increase in the level of mask pressure (Pmask = pressure support) indicate transfer of the work of breathing from the patient to the ventilator.
![Page 30: Advanced Pulmonary Mechanics during Mechanical Ventilation](https://reader035.vdocument.in/reader035/viewer/2022062304/568143fa550346895db08c4e/html5/thumbnails/30.jpg)
Partitioning of the Workload Between the Ventilator and the Patient
How the work of breathing partitions between the patient and the ventilatordepends on:
• Mode of ventilation (e.g., in assist control most of the work is usually done by the ventilator)• Patient effort and synchrony with the mode of ventilation• Specific settings of a given mode (e.g., level of pressure in PS and set rate in SIMV)
![Page 31: Advanced Pulmonary Mechanics during Mechanical Ventilation](https://reader035.vdocument.in/reader035/viewer/2022062304/568143fa550346895db08c4e/html5/thumbnails/31.jpg)
Respiratory Mechanics in ARF*
Reduced range of volume
excursion: Low
compliance
Flattering at low and high
volumes: Lower and
upper inflection points
*Bigatello: Br J Anaest 1996
Volume
Pressure
NORMAL
ARDS
![Page 32: Advanced Pulmonary Mechanics during Mechanical Ventilation](https://reader035.vdocument.in/reader035/viewer/2022062304/568143fa550346895db08c4e/html5/thumbnails/32.jpg)
Lung Protective Strategy
1. Set PEEP above the lower Pflex to keep the lung open and avoid alveolar collapse
2. Apply small Vt to minimize stretching forces
3. Set Pplat below the upper Pflex to avoid regional overdistension
Volume
Pressure
![Page 33: Advanced Pulmonary Mechanics during Mechanical Ventilation](https://reader035.vdocument.in/reader035/viewer/2022062304/568143fa550346895db08c4e/html5/thumbnails/33.jpg)
Abnormalities Air-leak Air trapping Increased airway resistance Inadequate flow support Inadequate sensitivity Atelectasis Inadequate PEEP Airway obstruction Over-distension
![Page 34: Advanced Pulmonary Mechanics during Mechanical Ventilation](https://reader035.vdocument.in/reader035/viewer/2022062304/568143fa550346895db08c4e/html5/thumbnails/34.jpg)
Air LeakAir Leak
Volume (ml)
Pressure (cm H2O)
Air Leak
![Page 35: Advanced Pulmonary Mechanics during Mechanical Ventilation](https://reader035.vdocument.in/reader035/viewer/2022062304/568143fa550346895db08c4e/html5/thumbnails/35.jpg)
Air LeakAir LeakInspiration
Expiration
Volume (ml)
Flow Flow (L/min)(L/min)
Air Leak in mL
NormalAbnormal
![Page 36: Advanced Pulmonary Mechanics during Mechanical Ventilation](https://reader035.vdocument.in/reader035/viewer/2022062304/568143fa550346895db08c4e/html5/thumbnails/36.jpg)
Air LeakVolume (mL)
Time (sec)
![Page 37: Advanced Pulmonary Mechanics during Mechanical Ventilation](https://reader035.vdocument.in/reader035/viewer/2022062304/568143fa550346895db08c4e/html5/thumbnails/37.jpg)
Air TrappingAir Trapping
Inspiration
Expiration
NormalNormalPatientPatient
Time (sec)
Flo
w (
L/m
in)
Air TrappingAuto-PEEP
}
![Page 38: Advanced Pulmonary Mechanics during Mechanical Ventilation](https://reader035.vdocument.in/reader035/viewer/2022062304/568143fa550346895db08c4e/html5/thumbnails/38.jpg)
Air TrappingAir TrappingInspiration
Expiration
Volume (ml)Volume (ml)
Flow (L/min)
Does not returnDoes not returnto baselineto baseline
NormalAbnormal
![Page 39: Advanced Pulmonary Mechanics during Mechanical Ventilation](https://reader035.vdocument.in/reader035/viewer/2022062304/568143fa550346895db08c4e/html5/thumbnails/39.jpg)
Response to Bronchodilator
Response to Bronchodilator
Before
Time (sec)
Flo
w (
L/m
in)
PEFR
After
Long TE
Higher PEFR
Shorter TE
![Page 40: Advanced Pulmonary Mechanics during Mechanical Ventilation](https://reader035.vdocument.in/reader035/viewer/2022062304/568143fa550346895db08c4e/html5/thumbnails/40.jpg)
Increased Airway Resistance
Increased Airway Resistance
Inspiration
Expiration
VolumeVolume (ml) (ml)
Flow (L/min)
Decreased PEFR
NormalNormalAbnormal
“Scooped out” pattern
![Page 41: Advanced Pulmonary Mechanics during Mechanical Ventilation](https://reader035.vdocument.in/reader035/viewer/2022062304/568143fa550346895db08c4e/html5/thumbnails/41.jpg)
Increased RawIncreased Raw
Pressure (cm Pressure (cm HH22O)O)
Higher PTA
Norm
al S
lope
Norm
al S
lope
Vol (mL)Vol (mL)
Lower S
lope
Lower S
lope
![Page 42: Advanced Pulmonary Mechanics during Mechanical Ventilation](https://reader035.vdocument.in/reader035/viewer/2022062304/568143fa550346895db08c4e/html5/thumbnails/42.jpg)
Airway Secretions/Water in the Circuit
Airway Secretions/Water in the Circuit
InspirationInspiration
ExpirationExpiration
Volume (ml)Volume (ml)
Flow Flow (L/min)(L/min)
NormalNormalAbnormalAbnormal
![Page 43: Advanced Pulmonary Mechanics during Mechanical Ventilation](https://reader035.vdocument.in/reader035/viewer/2022062304/568143fa550346895db08c4e/html5/thumbnails/43.jpg)
F
VV
F
After SuctionBefore Suction
Airway Obstruction
![Page 44: Advanced Pulmonary Mechanics during Mechanical Ventilation](https://reader035.vdocument.in/reader035/viewer/2022062304/568143fa550346895db08c4e/html5/thumbnails/44.jpg)
Optimising PEEP
V
P
PEEP: 3 cmH2O
V
P
PEEP: 8 cmH2O
![Page 45: Advanced Pulmonary Mechanics during Mechanical Ventilation](https://reader035.vdocument.in/reader035/viewer/2022062304/568143fa550346895db08c4e/html5/thumbnails/45.jpg)
Inadequate SensitivityInadequate Sensitivity
Volume Volume (mL)(mL)
PPawaw (cm H (cm H22O)O)Increased WOB
![Page 46: Advanced Pulmonary Mechanics during Mechanical Ventilation](https://reader035.vdocument.in/reader035/viewer/2022062304/568143fa550346895db08c4e/html5/thumbnails/46.jpg)
Replaced FRC
P
V
Lost FRC
V
P
Atelectasis
![Page 47: Advanced Pulmonary Mechanics during Mechanical Ventilation](https://reader035.vdocument.in/reader035/viewer/2022062304/568143fa550346895db08c4e/html5/thumbnails/47.jpg)
OverdistensionOverdistension
Volu
me (
ml)
Pressure (cm HPressure (cm H22O)O)
With little or no change in VTWith little or no change in VT
Paw risesPaw
rises
NormalAbnormal
![Page 48: Advanced Pulmonary Mechanics during Mechanical Ventilation](https://reader035.vdocument.in/reader035/viewer/2022062304/568143fa550346895db08c4e/html5/thumbnails/48.jpg)
Overdistension
Overdistension occurs when the volume limit of some components of the lung has been exceeded
Abrupt decrease in compliance at the termination of inspiration
Results in a terminal “Beaking” of the P/V Loop
Volume
Pressure
![Page 49: Advanced Pulmonary Mechanics during Mechanical Ventilation](https://reader035.vdocument.in/reader035/viewer/2022062304/568143fa550346895db08c4e/html5/thumbnails/49.jpg)
0.8 Pmax Pmax
Overdistension Index
Volume
Pressure
Cdyn
C20
![Page 50: Advanced Pulmonary Mechanics during Mechanical Ventilation](https://reader035.vdocument.in/reader035/viewer/2022062304/568143fa550346895db08c4e/html5/thumbnails/50.jpg)
THANK YOU