1 Agilent G5591A Bravo BenchCel Workstation – Site Preparation Checklist Issued: 11 October 2017, Revision: 1.0 Copyright 2017 Agilent Technologies Page 1 of 15 For Research Use Only. Not for use in diagnostic procedures. Thank you for purchasing the Agilent G5591A Bravo BenchCel Workstation. To get you started and to assure a successful and timely installation, please refer to this specification or set of requirements. Correct site preparation is the key first step in ensuring that your instruments and software systems operate reliably over an extended lifetime. This document is an information guide AND checklist prepared for you that outlines the supplies, consumables, space and utility requirements for your equipment. Customer Responsibilities Make sure your site meets the following specifications before the installation date. For details, see specific sections within this checklist, including: The laboratory space requirements to accommodate the size and weight of the instrument (See Other Requirements section for dimensions and weight of the shipping containers.) The laboratory table specifications The environmental conditions for the lab as well as laboratory gases and plumbing The power requirements related to the product, for example, number and location of electrical outlets The required operating supplies necessary for the product and installation Please consult Other Requirements section below for other product-specific information. For more details, please consult the instrument safety and installation guides listed in the Other Requirements section. If Agilent is delivering installation and familiarization services, users of the workstation should be present throughout these services; otherwise, they will miss important operational, maintenance and safety information. Important Customer Information 1. If you have questions or problems in providing anything described as a Customer Responsibility above, please contact your local Agilent or partner support/service organization for assistance prior to delivery. In addition, Agilent and/or its partners reserve the right to reschedule the installation dependent upon the readiness of your laboratory. 2. Should your site not be ready for whatever reasons, please contact Agilent as soon as possible to re-arrange any services that have been purchased. 3. Other optional services such as additional training and consultation for user-specific applications may also be provided at the time of installation when ordered with the system, but should be contracted separately.

Upload: trinhnga

Post on 07-Sep-2018

249 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: Agilent G5591A Bravo BenchCel Workstation Site Preparation ... · Agilent 1 G5591A Bravo BenchCel Workstation – Site Preparation Checklist Issued: 11 October 2017, Revision: 1.0

1

Agilent G5591A Bravo BenchCel Workstation – Site Preparation Checklist

Issued: 11 October 2017, Revision: 1.0 Copyright 2017 Agilent Technologies Page 1 of 15

For Research Use Only. Not for use in diagnostic procedures.

Thank you for purchasing the Agilent G5591A Bravo BenchCel Workstation. To get you started and to assure a successful and timely installation, please refer to this specification or set of requirements.

Correct site preparation is the key first step in ensuring that your instruments and software systems operate reliably over an extended lifetime. This document is an information guide AND checklist prepared for you that outlines the supplies, consumables, space and utility requirements for your equipment.

Customer Responsibilities

Make sure your site meets the following specifications before the installation date. For details, see specific sections within this checklist, including:

The laboratory space requirements to accommodate the size and weight of the instrument (See Other Requirements section for dimensions and weight of the shipping containers.)

The laboratory table specifications

The environmental conditions for the lab as well as laboratory gases and plumbing

The power requirements related to the product, for example, number and location of electrical outlets

The required operating supplies necessary for the product and installation

Please consult Other Requirements section below for other product-specific information.

For more details, please consult the instrument safety and installation guides listed in the Other Requirements section.

If Agilent is delivering installation and familiarization services, users of the workstation should be present throughout these services; otherwise, they will miss important operational, maintenance and safety information.

Important Customer Information

1. If you have questions or problems in providing anything described as a Customer Responsibility above, please contact your local Agilent or partner support/service organization for assistance prior to delivery. In addition, Agilent and/or its partners reserve the right to reschedule the installation dependent upon the readiness of your laboratory.

2. Should your site not be ready for whatever reasons, please contact Agilent as soon as possible to re-arrange any services that have been purchased.

3. Other optional services such as additional training and consultation for user-specific applications may also be provided at the time of installation when ordered with the system, but should be contracted separately.

Page 2: Agilent G5591A Bravo BenchCel Workstation Site Preparation ... · Agilent 1 G5591A Bravo BenchCel Workstation – Site Preparation Checklist Issued: 11 October 2017, Revision: 1.0

2

Agilent G5591A Bravo BenchCel Workstation – Site Preparation Checklist

Issued: 11 October 2017, Revision: 1.0 Copyright 2017 Agilent Technologies Page 2 of 15

1BDimensions and Weight

Identify the laboratory table space before your system arrives based on the following information.

Pay special attention to the total height and total weight requirements for all workstation components you have ordered and avoid bench space with overhanging shelves.

The following tables and figures list the dimensions for the following G5591A Bravo BenchCel Workstations:

Basic Liquid-Handling Workstation includes the following components:

o G5563A Bravo Platform

o G5580A BenchCel Microplate Handler

o Safety equipment: Light Curtain and shields, emergency-stop pendants, Robot Disable Hub

o Ethernet switch

Liquid-Handling and Sealing Workstation includes the following:

o Basic Workstation components

o G5585A PlateLoc Thermal Microplate Sealer

Liquid-Handling and Storage Workstation includes the following:

o Basic Workstation components

o G5584A Labware MiniHub

Liquid-Handling and Labeling Workstation includes the following:

o Basic Workstation components

o G5581A Microplate Labeler

Liquid-Handling, Sealing, and Centrifugation Workstation includes the following:

o Basic Workstation components

o G5585A PlateLoc Thermal Microplate Sealer

o G5582A Microplate Centrifuge with Loader

Liquid-Handling, Sealing, and Labeling Workstation includes the following:

o Basic Workstation components

o G5585A PlateLoc Thermal Microplate Sealer

o G5581A Microplate Labeler

Note: The tables list the overall workstation height based on 660-mm labware racks. The figures show all three rack heights (250-mm, 660-mm, and 860-mm). The figures show the BenchCel 6R, but provide the widths for the BenchCel 2R (two labware racks), BenchCel 4R (four labware racks), and BenchCel 6R (six labware racks).

Special Notes:

See laboratory table specifications under Other Requirements.

Page 3: Agilent G5591A Bravo BenchCel Workstation Site Preparation ... · Agilent 1 G5591A Bravo BenchCel Workstation – Site Preparation Checklist Issued: 11 October 2017, Revision: 1.0

3

Agilent G5591A Bravo BenchCel Workstation – Site Preparation Checklist

Issued: 11 October 2017, Revision: 1.0 Copyright 2017 Agilent Technologies Page 3 of 15

G5591A Bravo BenchCel Workstation (Basic Workstation) dimensions

Description

Weight Height (660 mm rack) Depth Width

Kg lbs. cm in cm in cm in

Bravo + BenchCel with 146-mm risers

- BenchCel 2R

110.8

244.3

126.0

49.6

68.8

27.1

180.9

71.2

- BenchCel 4R 126.9 279.8 126.0 49.6 68.8 27.1 224.1 88.2

- BenchCel 6R 142.9 315.0 126.0 49.6 68.8 27.1 267.2 105.2

Without 146-mm risers

-12.3* -27.1* 112.1 44.1 - - - -

*Subtract this value from the workstation with BenchCel 2R, 4R, or 6R weight, as applicable. Figure: Dimensions (mm) of G5591A Bravo BenchCel Workstation on 146-mm risers (top and front views)

Page 4: Agilent G5591A Bravo BenchCel Workstation Site Preparation ... · Agilent 1 G5591A Bravo BenchCel Workstation – Site Preparation Checklist Issued: 11 October 2017, Revision: 1.0

4

Agilent G5591A Bravo BenchCel Workstation – Site Preparation Checklist

Issued: 11 October 2017, Revision: 1.0 Copyright 2017 Agilent Technologies Page 4 of 15

G5591A Bravo BenchCel Workstation with PlateLoc Sealer dimensions

Description

Weight Height (660 mm rack) Depth Width

Kg lbs. cm in cm in cm in

Bravo + BenchCel + PlateLoc +146-mm risers

- BenchCel 2R

138.0

304.2

126.0

49.6

69.8

27.5

203.5

80.1

- BenchCel 4R 154.1 339.7 126.0 49.6 69.8 27.5 246.7 97.1

- BenchCel 6R 170.1 375.0 126.0 49.6 69.8 27.5 289.8 114.1

Without 146-mm risers -19.6* -43.2* 112.1 44.1 - - - -

*Subtract this value from the workstation with BenchCel 2R, 4R, or 6R weight, as applicable. Figure: Dimensions (mm) of G5591A Bravo BenchCel Workstation with PlateLoc Sealer on 146-mm risers (top and front views)

Page 5: Agilent G5591A Bravo BenchCel Workstation Site Preparation ... · Agilent 1 G5591A Bravo BenchCel Workstation – Site Preparation Checklist Issued: 11 October 2017, Revision: 1.0

5

Agilent G5591A Bravo BenchCel Workstation – Site Preparation Checklist

Issued: 11 October 2017, Revision: 1.0 Copyright 2017 Agilent Technologies Page 5 of 15

G5591A Bravo BenchCel Workstation with Labware MiniHub dimensions

Description

Weight Height (660-mm rack) Depth Width

Kg lbs. cm in cm in cm in

Bravo + BenchCel + MiniHub with risers

- BenchCel 2R

128.3

282.9

126.0

49.6

68.8

27.1 216.4 85.2

- BenchCel 4R 144.4 318.3 126.0 49.6 68.8 27.1 259.6 102.2

- BenchCel 6R 160.4 353.6 126.0 49.6 68.8 27.1 266.0 119.2

Without 146-mm risers -19* -41.9* 112.1 44.1 - - - -

*Subtract this value from the workstation with BenchCel 2R, 4R, or 6R weight, as applicable. Figure: Dimensions (mm) of G5591A Bravo BenchCel Workstation with MiniHub on 146-mm risers (top and front views)

Page 6: Agilent G5591A Bravo BenchCel Workstation Site Preparation ... · Agilent 1 G5591A Bravo BenchCel Workstation – Site Preparation Checklist Issued: 11 October 2017, Revision: 1.0

6

Agilent G5591A Bravo BenchCel Workstation – Site Preparation Checklist

Issued: 11 October 2017, Revision: 1.0 Copyright 2017 Agilent Technologies Page 6 of 15

G5591A Bravo BenchCel Workstation with Microplate Labeler dimensions

Description

Weight Height (660-mm rack) Depth Width

Kg lbs. cm in cm in cm in

Bravo + BenchCel + Labeler with risers

- BenchCel 2R

138.8

306.0

126.0

49.6

68.8

27.1 216.4 85.2

- BenchCel 4R 154.9 341.5 126.0 49.6 68.8 27.1 259.6 102.2

- BenchCel 6R 170.9 376.8 126.0 49.6 68.8 27.1 266.0 119.2

Without 146-mm risers -12.3* -27.1* 112.1 44.1 - - - -

*Subtract this value from the workstation with BenchCel 2R, 4R, or 6R weight, as applicable. Figure: Dimensions (mm) of G5591A Bravo BenchCel Workstation on 146-mm risers with Microplate Labeler (top and front views)

Page 7: Agilent G5591A Bravo BenchCel Workstation Site Preparation ... · Agilent 1 G5591A Bravo BenchCel Workstation – Site Preparation Checklist Issued: 11 October 2017, Revision: 1.0

7

Agilent G5591A Bravo BenchCel Workstation – Site Preparation Checklist

Issued: 11 October 2017, Revision: 1.0 Copyright 2017 Agilent Technologies Page 7 of 15

G5591A Bravo BenchCel Workstation with PlateLoc Sealer over Centrifuge with Loader dimensions

Description

Weight Height (660-mm rack) Depth Width

Kg lbs. cm in cm in cm in

Bravo + BenchCel + PlateLoc + Centrifuge

- BenchCel 2R

183.2

403.9

126.0

49.6

87.8

34.6 210.5 82.9

- BenchCel 4R 199.3 439.4 126.0 49.6 87.8 34.6 253.7 99.9

- BenchCel 6R 215.3 474.7 126.0 49.6 87.8 34.6 296.9 116.9

Figure: Dimensions (mm) of G5591A BenchCel Workstation with PlateLoc Sealer over Centrifuge with Loader (top and front views)

Page 8: Agilent G5591A Bravo BenchCel Workstation Site Preparation ... · Agilent 1 G5591A Bravo BenchCel Workstation – Site Preparation Checklist Issued: 11 October 2017, Revision: 1.0

8

Agilent G5591A Bravo BenchCel Workstation – Site Preparation Checklist

Issued: 11 October 2017, Revision: 1.0 Copyright 2017 Agilent Technologies Page 8 of 15

G5591A Bravo BenchCel Workstation with PlateLoc Sealer and Microplate Labeler dimensions

Description

Weight Height (660-mm racks) Depth Width

Kg lbs. cm in cm in cm in

Bravo + BenchCel + PlateLoc + Labeler

- BenchCel 2R

163.9

361.3

126.0

49.6

69.8

27.5

245.2

96.5

- BenchCel 4R 180.0 396.8 126.0 49.6 69.8 27.5 288.4 113.5

- BenchCel 6R 196.0 432.1 126.0 49.6 69.8 27.5 331.6 130.6

Figure: Dimensions (mm) of G5591A Bravo BenchCel Workstation with PlateLoc Sealer and Microplate Labeler (top and front views)

Page 9: Agilent G5591A Bravo BenchCel Workstation Site Preparation ... · Agilent 1 G5591A Bravo BenchCel Workstation – Site Preparation Checklist Issued: 11 October 2017, Revision: 1.0

9

Agilent G5591A Bravo BenchCel Workstation – Site Preparation Checklist

Issued: 11 October 2017, Revision: 1.0 Copyright 2017 Agilent Technologies Page 9 of 15

2BEnvironmental Conditions

Operating your workstation within the recommended environmental conditions ensures optimum instrument performance and lifetime. The site’s ambient temperature conditions must be stable for optimum performance.

Special Notes

1. Performance can be affected by sources of heat and cold, for example, direct sunlight, heating or cooling from air conditioning outlets, and drafts.

2. The workstation is intended to operate in a low-vibration environment. Excessive vibration may induce pipettor and robot errors.

3. The workstation is for indoor use only. Pollution degree 2. Installation category II.

4. Ensure that your operating temperature and humidity ranges are within the limits of all the instruments in your given workstation.

Instrument Description

Operating temp

range C (F)

Operating humidity range

(%)

Altitude

G5580A BenchCel Microplate Handler 5 to 40 C 20% to 90% RH, non-condensing

Up to 2000 m

G5563A Bravo Platform 4 to 40 C 10% to 90% RH, non-condensing

Up to 2000 m

G5585A PlateLoc Sealer 4 to 40 C 10% to 90% RH, non-condensing

Up to 2000 m

G5581A Microplate Labeler 10 to 35 C 30% to 85% RH, non-condensing

Up to 2000 m

G5582A, G5583A Microplate Centrifuge with Loader

15 to 40 C 10% to 95% RH, non-condensing

Up to 2000 m

Page 10: Agilent G5591A Bravo BenchCel Workstation Site Preparation ... · Agilent 1 G5591A Bravo BenchCel Workstation – Site Preparation Checklist Issued: 11 October 2017, Revision: 1.0

10

Agilent G5591A Bravo BenchCel Workstation – Site Preparation Checklist

Issued: 11 October 2017, Revision: 1.0 Copyright 2017 Agilent Technologies Page 10 of 15

3BPower Consumption

Special Notes

1. In addition to the following, you should consider the power requirements for the computer workstation and any accessories.

Instrument Description Line Voltage &

Frequency (V, Hz) Maximum Power

Consumption (VA)

G5580A BenchCel Microplate Handler

100 - 240 VAC, 50/60 Hz 5 A at 120 V~ 2.5 A at 240 V~

G5563A Bravo Platform 100 - 240 VAC, 50/60 Hz 300 VA

G5584A Labware MiniHub with power supply

100 - 240 VAC, 50/60 Hz 150 VA

G5585A PlateLoc Sealer

North American model:

100 - 120 VAC, 50/60 Hz

European model:

200 - 240 VAC, 50/60 Hz

4 A at 120 V~

2.5 A at 240 V~

G5581A Microplate Labeler 100 - 240 VAC, 50/60 Hz 4 A at 120 V~

2.5 A at 240 V~

G5582A, G5583A Microplate Centrifuge with Loader

100 - 240 VAC, 50/60 Hz 7A/120V,

3.5A/240V

Robot Disable Hub with power supply

100 - 240 VAC, 50/60 Hz -

Ethernet switch 100 - 240 VAC, 50/60 Hz -

Page 11: Agilent G5591A Bravo BenchCel Workstation Site Preparation ... · Agilent 1 G5591A Bravo BenchCel Workstation – Site Preparation Checklist Issued: 11 October 2017, Revision: 1.0

11

Agilent G5591A Bravo BenchCel Workstation – Site Preparation Checklist

Issued: 11 October 2017, Revision: 1.0 Copyright 2017 Agilent Technologies Page 11 of 15

4BRequired Operating Supplies by Customer

Special Notes

1. See the table below for required supplies in order to ensure successful installation.

Item Description

(including dimensions, etc.)

Vendor’s Part Number

(if applicable) Recommended

Quantity

Disposable, 250 µL tips, filtered, sterile (for 96LT Head)

To order, go to: http://www.agilent.com/en/products/automation-solutions/automated-liquid-handling/disposable-pipette-tips

Agilent 19477-022 Case of 50

PlateLoc Sealer (standard), heat seal

To order, go to: http://www.agilent.com/en/products/automation-solutions/microplate-management-robotics/heat-seal

Agilent seal - Various

-

Microplate Labeler Media Kit: Roll of adhesive labels

Ribbon roll to image one label roll

Used ribbon take-up spool for take-up hub

Print head cleaning pen (multiuse)

Media loading document

http://www.agilent.com/en-us/products/automation-solutions/microplate-management-robotics/adhesive-labels

Agilent -

Labware Various -

Page 12: Agilent G5591A Bravo BenchCel Workstation Site Preparation ... · Agilent 1 G5591A Bravo BenchCel Workstation – Site Preparation Checklist Issued: 11 October 2017, Revision: 1.0

12

Agilent G5591A Bravo BenchCel Workstation – Site Preparation Checklist

Issued: 11 October 2017, Revision: 1.0 Copyright 2017 Agilent Technologies Page 12 of 15

Other Requirements

Compressed Air Requirements The workstation requires the use of clean, dry, compressed air to move pneumatic components. The compressed air can be from a centralized source (house), compressed-air cylinders, or portable pumps. The air source must meet the following specifications:

Instrument Quality Flow rate Pressure

G5580A BenchCel Microplate Handler

Clean, dry, compressed

34.0 Lpm (1.2 cfm) 0.65 to 0.69 MPa (95 to 100 psi)

G5585A PlateLoc Sealer

Clean, dry, compressed

70.8 Lpm (2.50 cfm) 0.62 to 0.69 MPa (90 to 100 psi)

G5581A Microplate Labeler

Clean, dry, compressed

72 Lpm (2.54 cfm) 0.55 MPa (80 psi)

G5582A, G5583A Microplate Centrifuge with Loader

Clean, dry, compressed

30 Lpm (1 cfm) 0.55 MPa (80 psi)

CAUTION: Using oil compressors can cause oil to leak into the BenchCel Microplate Handler and void your warranty. CAUTION: Air pressure greater than 0.69 MPa (100 psi) can damage the BenchCel Microplate Handler.

Laboratory Table Specifications 1. The laboratory table must support the weight of the workstation without excessive shaking or

movement. The table should be fixed in place, for example, castors that lock.

2. The table must be level in the direction of the width and the depth of the workstation. Using a traditional bubble level, the table should be leveled such that the bubble is centered between the two limit lines of the level

3. The table surface must meet the following requirements:

a. Be chemical resistant. Agilent Technologies does not recommend using laminated particle board, because the board might be exposed to liquids around the mounting holes.

b. Have a thickness relative to the material that will prevent warping when the workstation and computer are set upon the table.

c. Permit predrilled or tapped holes (M6 thread) for securing integration parts.

4. The table surface must be attached to the table frame.

5. The table frame must have:

a. A leveling mechanism in the feet or castors.

b. Cross members to prevent the table from swaying when the workstation is in operation and to prevent the table surface from bending when the workstation and computer are placed on top.

Page 13: Agilent G5591A Bravo BenchCel Workstation Site Preparation ... · Agilent 1 G5591A Bravo BenchCel Workstation – Site Preparation Checklist Issued: 11 October 2017, Revision: 1.0

13

Agilent G5591A Bravo BenchCel Workstation – Site Preparation Checklist

Issued: 11 October 2017, Revision: 1.0 Copyright 2017 Agilent Technologies Page 13 of 15

c. Table legs that are spaced a maximum of 90 cm (36 n) apart to properly support and distribute the weight.

d. Dimensions that enable support of the table surface without overhang.

e. Have castors that can be unlocked to enable moving the system away from a wall to provide maintenance access to the rear of the workstation.

Computer Requirements If your organization uses a computer other than one configured by Agilent Technologies, make sure the computer meets the minimum requirements:

Windows 10 (64-bit edition)

VWorks Automation Control software v13.1

BenchCel Workstations software v1.0

3.20 GHz, 8 MB cache, processor, 4 cores

8 GB DIMM

500 GB hard drive capacity (100 GB, minimum)

HD Graphics

1280 x 1924 screen resolution

Browser with JavaScript enabled: Microsoft Internet Explorer 8.0, or later, or Mozilla Firefox 3.0, or later (required for viewing the knowledge base)

A PDF viewer, such as Adobe Reader (required for opening the user guide PDF files)

Dedicated 10BaseT or faster Ethernet card for connecting to the workstation LAN. A second network card is required if the controlling computer will be connected to the site LAN. The Agilent service representative will ensure that the workstation LAN is connected and that communication is established with the workstation devices.

Note: Agilent Technologies is not responsible for establishing communication with your site LAN.

USB ports, 2 minimum

Serial port, if available

Page 14: Agilent G5591A Bravo BenchCel Workstation Site Preparation ... · Agilent 1 G5591A Bravo BenchCel Workstation – Site Preparation Checklist Issued: 11 October 2017, Revision: 1.0

14

Agilent G5591A Bravo BenchCel Workstation – Site Preparation Checklist

Issued: 11 October 2017, Revision: 1.0 Copyright 2017 Agilent Technologies Page 14 of 15

Safety Equipment and Precautions WARNING Changing or modifying the safety equipment can prevent the safe operation of the workstation, invalidate its safety compliance, and lead to personal injury or property damage. Any customer who does not use the supplied safety equipment or who modifies the supplied safety equipment assumes full responsibility for providing an appropriate level of safety for its operators and for providing the applicable safety compliance marking and documentation.

All safety equipment supplied with the workstation will be installed for you. The safety equipment includes shields and a Light Curtain to prevent access to moving-parts hazards. The Robot-Disable Hub provides the emergency-stop function for the integrated devices that are equipped with safety interlock circuitry. The Bravo Platform, BenchCel robot, and Centrifuge Loader are connected to the Robot-Disable Hub. Pressing the red emergency-stop button on a pendant or interrupting the Light Curtain will stop the motion of these devices.

Ensure that you understand the potential safety hazards and how to avoid them. For details see:

• Automation Solutions Product General Safety Guide (part number G5500-90015)

• G5562A, G5563A Bravo Platform Safety and Installation Guide (part number G5562-90001)

• BenchCel Microplate Handler User Guide (part number G5580-90000)

• If applicable, Labware MiniHub Safety and Installation Guide (part number G5584-90004)

• If applicable, PlateLoc Thermal Microplate Sealer User Guide (part number G5402-90001).

• If applicable, Microplate Labeler User Guide (part number G5404-90003)

• If applicable, Microplate Centrifuge with Loader User Guide (part number G5405-90002)

You can find these guides and others in the VWorks Knowledge Base at: http://www.agilent.com/chem/askb

Page 15: Agilent G5591A Bravo BenchCel Workstation Site Preparation ... · Agilent 1 G5591A Bravo BenchCel Workstation – Site Preparation Checklist Issued: 11 October 2017, Revision: 1.0

15

Agilent G5591A Bravo BenchCel Workstation – Site Preparation Checklist

Issued: 11 October 2017, Revision: 1.0 Copyright 2017 Agilent Technologies Page 15 of 15

Important Customer Web Links

For additional information about the G5591A Bravo BenchCel Workstations, please visit our product page:

http://www.agilent.com/en/products/automation-solutions/benchcel-workstations/bravo-benchcel-workstation

Need to get information on your product?

o VWorks Knowledge Base - http://www.agilent.com/chem/askb

o Literature Library - http://www.agilent.com/en-us/library/literature

Need to know more? Customer Education – http://www.agilent.com/crosslab/university/

Need Support Services? - www.agilent.com/crosslab

Document part number: G5591-90005