Page 1: ALLEGRO G050208-00-Z LSC, Livingston 23 March, 2005 1 Calibration for the ALLEGRO resonant detector -- S2 and S4 Martin McHugh, Loyola University New Orleans

G050208-00-Z LSC, Livingston 23 March , 2005

1

ALLEGRO

Calibration for the ALLEGRO resonant detector -- S2 and S4

Martin McHugh, Loyola University New Orleans

Warren Johnson, Louisiana State University

with the ALLEGRO group, Stochastic Sources Subgroup

Page 2: ALLEGRO G050208-00-Z LSC, Livingston 23 March, 2005 1 Calibration for the ALLEGRO resonant detector -- S2 and S4 Martin McHugh, Loyola University New Orleans

G050208-00-Z LSC, Livingston 23 March , 2005

2

ALLEGRO

Outline

• Motivation, background

• Signal flow diagram, transfer function equations

• Discussion of calibration measurements

• Recent mysteries

• summary

Page 3: ALLEGRO G050208-00-Z LSC, Livingston 23 March, 2005 1 Calibration for the ALLEGRO resonant detector -- S2 and S4 Martin McHugh, Loyola University New Orleans

G050208-00-Z LSC, Livingston 23 March , 2005

3

ALLEGRO

Motivation

• Provide input to stochastic background analysis using ALLEGRO and LLO (John Whelan’s talk earlier this week)

• Unlike an event list based search, a coherent search such as this requires a phase consistent response function for the detector signal path

Page 4: ALLEGRO G050208-00-Z LSC, Livingston 23 March, 2005 1 Calibration for the ALLEGRO resonant detector -- S2 and S4 Martin McHugh, Loyola University New Orleans

G050208-00-Z LSC, Livingston 23 March , 2005

4

ALLEGRO

ALLEGRO schematic

Page 5: ALLEGRO G050208-00-Z LSC, Livingston 23 March, 2005 1 Calibration for the ALLEGRO resonant detector -- S2 and S4 Martin McHugh, Loyola University New Orleans

G050208-00-Z LSC, Livingston 23 March , 2005

5

ALLEGRO

Signal path

Page 6: ALLEGRO G050208-00-Z LSC, Livingston 23 March, 2005 1 Calibration for the ALLEGRO resonant detector -- S2 and S4 Martin McHugh, Loyola University New Orleans

G050208-00-Z LSC, Livingston 23 March , 2005

6

ALLEGRO

˜ z v ( f − f r ) = A( f ) ˜ v S ( f ) = aLe i(td 2π ( f − fr ))e−iφ ˜ v S ( f )

˜ F ( f ) = J( f ) ˜ h ( f ) = 4MLf 2 ˜ h ( f )

˜ x ( f ) = G( f ) ˜ F ( f )

= α1

f p2 − f 2 +

if p f

Qp( )−

1

fm2 − f 2 + ifm f

Qm( )

⎜ ⎜

⎟ ⎟˜ F ( f )

˜ I ( f ) = K ⋅ ˜ x ( f ) =IDC

g

⎝ ⎜

⎠ ⎟˜ x ( f )

˜ v S ( f ) = Z ⋅ ˜ I ( f )

˜ z c ( f − f r ) = D ⋅ ˜ z v ( f − f r )

Form of each Transfer Function€

˜ F C ( f ) = H ˜ v e ( f ) =CcVDC

gc

˜ v e ( f )

Page 7: ALLEGRO G050208-00-Z LSC, Livingston 23 March, 2005 1 Calibration for the ALLEGRO resonant detector -- S2 and S4 Martin McHugh, Loyola University New Orleans

G050208-00-Z LSC, Livingston 23 March , 2005

7

ALLEGRO

Need to determine --

• Mode frequencies and Q’s -- fm , fp , Qm , Qp

• overall scale -- in practice we measure is mechanical gain -- includes

‘tuning factor’

• Lock-in amplifier parameters -- gain, filter delay and phase shiftAlso need to know the phase of

the lock-in reference oscillator

⋅K ⋅Z

So in practice the calibration amounts to --

˜ h ( f − fr ) =˜ z C ( f − f r )

J( f )G( f )KZA( f − f r)D

Inverse fft then gives

h H (t) complex heterodyned strain time series

Page 8: ALLEGRO G050208-00-Z LSC, Livingston 23 March, 2005 1 Calibration for the ALLEGRO resonant detector -- S2 and S4 Martin McHugh, Loyola University New Orleans

G050208-00-Z LSC, Livingston 23 March , 2005

8

ALLEGRO

New calibrator installed between S2 and S4 -- known transfer function

• One plate of capacitor is tightly coupled to bar.

• Other plate is weakly coupled to the bar, so acts like a free mass.

• Both plates electrically isolated.

Rest of detector unchanged -Measurements will apply to S2

Page 9: ALLEGRO G050208-00-Z LSC, Livingston 23 March, 2005 1 Calibration for the ALLEGRO resonant detector -- S2 and S4 Martin McHugh, Loyola University New Orleans

G050208-00-Z LSC, Livingston 23 March , 2005

9

ALLEGRO

The calibrator mounted on the bar

Page 10: ALLEGRO G050208-00-Z LSC, Livingston 23 March, 2005 1 Calibration for the ALLEGRO resonant detector -- S2 and S4 Martin McHugh, Loyola University New Orleans

G050208-00-Z LSC, Livingston 23 March , 2005

10

ALLEGRO

Transfer function - white noise excitation to measured output

measurements from 20 March 2004 -- excitation measured through lock-in and A/D plotted here we have

TF = H ⋅G( f ) ⋅K ⋅Z

gives us the overall scale -

Page 11: ALLEGRO G050208-00-Z LSC, Livingston 23 March, 2005 1 Calibration for the ALLEGRO resonant detector -- S2 and S4 Martin McHugh, Loyola University New Orleans

G050208-00-Z LSC, Livingston 23 March , 2005

11

ALLEGRO

Band-limited white noise injection -- recorded directly and through lock-in/anti-aliasing filters

Lock-in/ filter measurements

Page 12: ALLEGRO G050208-00-Z LSC, Livingston 23 March, 2005 1 Calibration for the ALLEGRO resonant detector -- S2 and S4 Martin McHugh, Loyola University New Orleans

G050208-00-Z LSC, Livingston 23 March , 2005

12

ALLEGRO

Compare fourier coefficients

Lock-in/filter introduces an 11ms delay and 18 degree phase shift

Page 13: ALLEGRO G050208-00-Z LSC, Livingston 23 March, 2005 1 Calibration for the ALLEGRO resonant detector -- S2 and S4 Martin McHugh, Loyola University New Orleans

G050208-00-Z LSC, Livingston 23 March , 2005

13

ALLEGRO

Page 14: ALLEGRO G050208-00-Z LSC, Livingston 23 March, 2005 1 Calibration for the ALLEGRO resonant detector -- S2 and S4 Martin McHugh, Loyola University New Orleans

G050208-00-Z LSC, Livingston 23 March , 2005

14

ALLEGRO

Calibration lineCalibration line

Detector resonancesDetector resonances

Extra mechanical resonancesExtra mechanical resonances

Raw data

Page 15: ALLEGRO G050208-00-Z LSC, Livingston 23 March, 2005 1 Calibration for the ALLEGRO resonant detector -- S2 and S4 Martin McHugh, Loyola University New Orleans

G050208-00-Z LSC, Livingston 23 March , 2005

15

ALLEGRO

Calibrated strain spectrum from S4

Page 16: ALLEGRO G050208-00-Z LSC, Livingston 23 March, 2005 1 Calibration for the ALLEGRO resonant detector -- S2 and S4 Martin McHugh, Loyola University New Orleans

G050208-00-Z LSC, Livingston 23 March , 2005

16

ALLEGRO

Stability of calibration -- calibration line tracked through S2

Page 17: ALLEGRO G050208-00-Z LSC, Livingston 23 March, 2005 1 Calibration for the ALLEGRO resonant detector -- S2 and S4 Martin McHugh, Loyola University New Orleans

G050208-00-Z LSC, Livingston 23 March , 2005

17

ALLEGRO

• Discrepancy between white noise injection data sets discovered to be due to change in polarity of DC

bias applied to calibrator (?!)

• Two mysteriesNuisance mode at 885 Hz -- relative phase changes

with DC bias polarityApparent offset in DC bias voltage on calibrator

A Problem

Page 18: ALLEGRO G050208-00-Z LSC, Livingston 23 March, 2005 1 Calibration for the ALLEGRO resonant detector -- S2 and S4 Martin McHugh, Loyola University New Orleans

G050208-00-Z LSC, Livingston 23 March , 2005

18

ALLEGRO

Transfer function magnitude, white noise injection --blue curve +3.99VDC nominal biasred curve -3.99VDC nominal bias

Page 19: ALLEGRO G050208-00-Z LSC, Livingston 23 March, 2005 1 Calibration for the ALLEGRO resonant detector -- S2 and S4 Martin McHugh, Loyola University New Orleans

G050208-00-Z LSC, Livingston 23 March , 2005

19

ALLEGRO

… now flip the plate to which the voltage is applied

Page 20: ALLEGRO G050208-00-Z LSC, Livingston 23 March, 2005 1 Calibration for the ALLEGRO resonant detector -- S2 and S4 Martin McHugh, Loyola University New Orleans

G050208-00-Z LSC, Livingston 23 March , 2005

20

ALLEGRO

we can certainly model extra mechanical modes -- but the phase flip is a mystery

Page 21: ALLEGRO G050208-00-Z LSC, Livingston 23 March, 2005 1 Calibration for the ALLEGRO resonant detector -- S2 and S4 Martin McHugh, Loyola University New Orleans

G050208-00-Z LSC, Livingston 23 March , 2005

21

ALLEGRO

837Hz calibration line strength vs. calibrator bias

Page 22: ALLEGRO G050208-00-Z LSC, Livingston 23 March, 2005 1 Calibration for the ALLEGRO resonant detector -- S2 and S4 Martin McHugh, Loyola University New Orleans

G050208-00-Z LSC, Livingston 23 March , 2005

22

ALLEGRO

Summary

• We have calibrated h(t) for S2 data set • Hardware injections done for E12, S4 -- good news

we are able to extract coherent signal between L1 and A1 (see Sukanta’s earlier talk)

• Analyze burst injections done to determine overall sign of the calibration

• Need to investigate calibrator behavior with post S4 measurements -- be sure there is no problem with overall scale determination