2 solutions for noninvasive blood pressure measurement
Post on 31-May-2015
1.383 Views
Preview:
DESCRIPTION
TRANSCRIPT
www.emka.frwww.emka.fr
smart tools – swift services – sharp
research
2 solutionsfor noninvasive blood pressuremeasurement
emkaPACK & nibpSNAPSHOT
www.emka.fr
noninvasive blood pressuremeasurement
A
B
C
D
overview
emkaPACK
nibpSNAPSHOT
software analysis
www.emka.fr
overview
A
www.emka.fr
what is NIBP?a noninvasive method for measuring the blood pressure of laboratory animals (dog, primates, etc.)
www.emka.fr
how does it work?
emka TECHNOLOGIES’ noninvasive blood pressure systems are based on the oscillometric technique: pressure is measured from a cuff around a limb or the tail that is inflated to block blood flow then deflated slowly
www.emka.fr
what is analyzed?
time
cuff pressure
www.emka.fr
2 different solutionsemkaPACKnibp is an optional feature of our emkaPACK noninvasive telemetry system - suitable for long term measurements on unrestrained animals
nibpSNAPSHOTself-contained, handheld device designed for single-point measurements
both supplied with dedicated software for data acquisition & analysis
www.emka.fr
common features
low and high pressure range 0-300mmHgadjustable deflation speed5mmHg/s – 15mmHg/slinear deflation speedhigh speed servo-control by instant pressure and dPdtadjustable pulse amplification 5x-100x
www.emka.fr
emkaPACK
B
www.emka.fr
designed for the emkaPACK systemsingle-shot or programmable cyclic operationembedded calibration information
www.emka.fr
setting-up the animal
tail is shaved
cuff is wrapped around the tail and connected to the transmitter
a custom velcro extension protects the cuff, and maintains it in the proper position on the tail
www.emka.fr
nibpSNAPSHOT
C
www.emka.fr
nibpSNAPSHOT
www.emka.fr
features
handheld deviceUSB connection to computerpowered through USBboth a NIBP device and a 7-leads ecg devicedesigned for single-point measurementsmeasurements can be repeated every 10s if necessary
www.emka.fr
specifications
www.emka.fr
software analysis
D
www.emka.fr
real-time & off-line software analysis
real-time analysis performed by iox2 (nibpSNAPSHOT, emkaPACK, emkaPACK 4G)
off-line analysis performed by ecgAUTO (nibpSNAPSHOT, emkaPACK, TOE)
control by iox2
ecgAUTO
www.emka.fr
parameters computed from the deflation curve alone
systolic blood pressurediastolic blood pressuremean blood pressure
ecgAUTO
www.emka.fr
parameters computed with simultaneous ecg
heart ratepressure transit time
ecgAUTO
www.emka.fr
pulse amplitude vs cuff pressure
pressure pulses and ecg
cuff pressure vs time
real-time analysis in iox2
www.emka.fr
pulse amplitude vs cuff pressure
pressure pulses and ecg
cuff pressure vs time
off-line analysis in ecgAUTO
www.emka.fr
how is analysis performed?
detects inflation cyclesapplies hi-pass filterselects valid blood pressure pulsesproduces bell shaped amplitude curvecomputes mean, systolic and diastolic pressure values
ecgAUTO
www.emka.fr
what are the challenges for a successful analysis?
when signal is noisy the algorithm picks pulses that correspond to blood pressure beats and ignores artefacts
for greater efficiency, the software applies a number of criteria and rejects unreliable or spurious results
present experience (around 1000 hours of data from 40 dogs) indicate that from 70 to 95% of inflation cycles produce reliable data
www.emka.fr
examples of good and poor quality signals
pulses from noisy zones are ignored
www.emka.fr
www.emka.fr
facebook :
twitter : @emkaTECH
www.facebook.com/emkaTECHNOLOGIES
www.emka.frwww.emka.fr
solutions for lifesciences research