blood flow measurement

35
BLOOD FLOW METER Prepared by Abhijith Prabha , L6A ,ECE Department.

Upload: abhijith-prabha

Post on 15-Aug-2015

217 views

Category:

Engineering


26 download

TRANSCRIPT

Page 1: blood flow measurement

BLOOD FLOW METER

Prepared by Abhijith Prabha , L6A ,ECE Department.

Page 2: blood flow measurement

INTRODUCTION

• Blood flow is the one of the important physiological parameter and the most difficult to measure accurately.

• The average velocities of blood flow vary over a wide range depending on diameter of blood vessel.

• There are many techniques for measuring the blood flow and velocity.

• They are categorized into 1.invasive(surgical). 2.non invasive(through the skin).

Page 3: blood flow measurement

Need for blood flowmeter

• Inspection for block in blood flow.• Testing artificial blood vessels during organ transplantation.• During Fistula creation in dialysis.

Page 4: blood flow measurement

Typical values of blood flowType of blood vessels

Total cross-section area

Blood velocity in cm/s

Aorta 3–5 cm2 40 cm/s

Capillaries 4500–6000 cm2 0.03 cm/s

Vena cava inferior and superior

14 cm2 15 cm/s

Page 5: blood flow measurement

Introduction

• Most widely used blood flow meters are:

Electromagnetic blood flow meterUltrasonic blood flow metersNMR blood flow meterLaser Doppler blood flow meter

Page 6: blood flow measurement

ELECTROMAGNETIC BLOOD FLOWMETER

Page 7: blood flow measurement

ELECTROMAGNETIC BLOOD FLOWMETER

• Measures instantaneous pulsatile flow of blood

• Works based on the principle of electromagnetic induction

• The voltage induced in a conductor moving in a magnetic field is proportional to the velocity of the conductor

• The conductive blood is the moving conductor

Page 8: blood flow measurement

Principle of EM blood flow meter.

Page 9: blood flow measurement

ELECTROMAGNETIC BLOOD FLOWMETER

• A permanent magnet or electromagnet positioned around the blood vessel generates a magnetic field perpendicular to the direction of the flow of the blood.

• Voltage induced in the moving blood column is measured with stationary electrodes located on opposite sides of the blood vessel and perpendicular to the direction of the magnetic field.

• This method requires that the blood vessel be exposed so that the flow head or the measuring probe can be put across it.

Page 10: blood flow measurement

• The Induced emf

• Where• B = magnetic flux density, T• L = length between electrodes, m• u = instantaneous velocity of blood, m/s

Page 11: blood flow measurement

Fig: Typical flow transducer

• The electromagnetic flow-transducer is a tube of non-magnetic material to ensure that the magnetic flux does not bypass the flowing liquid and go into the walls of the tube.

Page 12: blood flow measurement

Types of EM Flowmeters

• basically, all modern flowmeters consist of a generator of AC, a probe assembly , a series of capacitance coupled amplifiers, a demodulator, a DC amplifier and a suitable recording device.

• Basing shape of the energizing current waveform for the electromagnet 2 types of EM Flowmeters are :

A. Sine wave flometer.B. Square wave flowmeter.

Page 13: blood flow measurement

Sine wave Flowmeters

• Probe magnet is energized with a sine wave and the induced voltage will also be sinusoidal.

• Since the flow of blood acts as a secondary terminal of a transformer w.r.t probe magnet, an additional artifact voltage induced is called transformer voltage.

• This voltage is 900 out of phase with the original signal corresponding to flow of blood .

• A method for eliminating transformer voltage by using a gated amplifier(amplify signal only during flow induced voltage is maximum).

• This type of instrument is known as ‘gated sine wave flowmeter’ ,

Page 14: blood flow measurement

Square wave flowmeter

• Probe magnet is energized with a square wave and induced voltage is also square wave.

• It is easier to control magnitude and wave shape of energizing current.

• Separation of transformer voltage is easy .

• For the measurement action square wave is amplitude modulated by variation in blood flow.

Page 15: blood flow measurement

Block diagram of square wave flowmeter

Page 16: blood flow measurement

Block diagram

Transducer• consist of an electromagnet, a pair of electrodes.• Electrodes may be in contact with either flowing blood or

outer surface of the blood vessel

Preamplifier • The induced voltage pick up by the electrodes, is given to a

low noise differential amplifier through a capacitive coupling• Must have a very high CMRR and input impedance

Page 17: blood flow measurement

Block diagram

Gating circuit• It helps to remove spurious voltages generated during magnet

current reversal• The gating action is controlled by the circuit which provides

an excitation current to the electromagnet

Band pass filter• It is an active RC band pass amplifier , which selectively pass

through it the amplified square wave signal• Peak response is kept for 400Hz

Page 18: blood flow measurement

Block diagram

Detector• A phase sensitive detector is used to recover the signal• It also helps in the rejection of interfering voltages at

frequencies below the carrier frequency

Low pass filter and output stage• Demodulated signal is given to an RC LPF , which provides a

uniform frequency response and a linear phase shift • Followed by an integrator provide output corresponding to the

mean flow

Page 19: blood flow measurement

Block diagram

Magnet current drive• The square wave input to the power amplifier stage which

supplies current to the electromagnet is fed from a free running multivibrator

Zero flow reference line• it establish the signal corresponding to zero-flow before

measurement.

Page 20: blood flow measurement

Ultrasonic blood flowmeter

Page 21: blood flow measurement

Ultrasonic blood flowmeter

• An ultrasonic flow meter is a type of flow meter that measures the velocity of a fluid with ultrasound to calculate volume flow.

• Using ultrasonic transducers, the flow meter can measure the average velocity along the path of an emitted beam of ultrasound.

• By averaging the difference in measured transit time between the pulses of ultrasound propagating into and against the direction of the flow or by measuring the frequency shift from the Doppler effect.

Page 22: blood flow measurement

Fig : ultra sonic flow meter

Page 23: blood flow measurement

(a)Doppler-shift flow-velocity meters

• It is an non invasive method.• It is based on the analysis of echo signals from

erythrocytes(RBCs) in blood. • The incident ultrasound is scattered by the blood cells and

scattered wave is received by the second receiver.• The frequency shift of the scattered wave gives idea about

velocity of scatterers. • The Doppler frequency shift is a measure of size and direction

of the flow velocity.

Page 24: blood flow measurement

Doppler-shift flow-velocity meters

5 MHz exciter

Rx Tx

R F amplifierdetector

LPF Audio amplifier

Zero crossing detector

LPF

Blood vessel

Fig : block diagram

output

Page 25: blood flow measurement

(b)Range gated pulsed Doppler flow meter

• Baker(1970) stated that recording blood flow using Doppler shift are sometimes misleading and inaccurate.

• These difficulties can be overcome , if the ultrasonic source is pulsed.

• If the returning signal is range gated ,then the diameter and velocity of blood stream can measure together.

Page 26: blood flow measurement

Receiver limiter Phase detector LPF Sample and amplifier

Power amp Pulse amp Sample pulse gen. LPF

Freq divisionMaster Oscillator

(4.5-5.5 MHz)

Rx elementTx element

Audible Doppler output

Fig : block diagram of pulsed doppler flow meter.

Page 27: blood flow measurement

(c) Flow measurement by Doppler Imaging

• Doppler ultrasound is not only used for measurement of the absolute value of blood velocity and volume, but it also helps to visualize blood flow.

• The probe this imaging equipment is mechanically coupled to position resolver, which gives electrical output.

• Imaging is done by moving the probe through the skin, and developing a 3D image using a computer.

• Thus, it is possible to construct anterior-posterior , lateral and cross-sectional scans of blood vessel.

Page 28: blood flow measurement

Flow measurement by Doppler Imaging

Figs : Flow measurement by Doppler Imaging

Page 29: blood flow measurement

Flow measurement by Doppler Imaging

• Using this recording of mean blood flow, peak flow, reverse peak flow etc. of cardiac cycle are possible.

• This method is also helpful for taking measurements from brain , which is difficult to access.

Page 30: blood flow measurement

NMR Blood Flowmeter

Page 31: blood flow measurement

NMR Blood Flowmeter

• Non invasive method by Nuclear magnetic resonance principle.

• For flow measurement , behaviour of the 2 Hydrogen atoms of water is studied.(Since blood is approximately 83% water).

• Due to magnetic moment of the H atom , the nucleus behaves as a micro miniature magnet which can affected by externally applied magnetic field.

• the hydrogen nuclei orient themselves to produce a net parallel alignment to a steady magnetic field.

Page 32: blood flow measurement

NMR Blood Flowmeter

• Orientation of H atoms will be change according to the status of blood flow.

• This orientation will results in the magnetization of H atoms.• Therefore, magnitude of magnetization can be related to

either velocity of blood or flow rate.• A crossed coil configuration is used to detect the level of

magnetization in body.• The voltage induced in coil(NMR signal) is proportional to

velocity and area of vessel carrying blood.• NMR flow meters are limited in their application to the

measurement in limbs.

Page 33: blood flow measurement

Laser Doppler blood flowmeter

Page 34: blood flow measurement

Fig : schematic diagram of laser doppler blood flow meter

Page 35: blood flow measurement

THE END