ardo carum serviceanleitung - mediplus · 2019. 2. 27. · en en 32 ardo medical ag 2. product...
Post on 08-Feb-2021
1 Views
Preview:
TRANSCRIPT
-
Languages: Page:
Serviceanleitung 2–27
Service Manual 29–55
Caru
m –
Ser
vice
anle
itung
/ Se
rvic
e M
anua
l
de
en
-
en en
29 Ardo medical AG
Explanation
a. Carum Breastpumpb. Connection opening for PumpSet c. Control panel with LCDd. Connection for mains cable e. Bottle holder f. EasyFreeze holderg. Mains cable
h. Carum Containeri . Address fieldj. Bottle holderk. EasyFreeze holder l. Carumm. Mains cablen. PumpSet
b.
f.
e.
g.
a.
c.
d.
h.
i.
j.
k.
l.
m.
n.
-
en en
30 Ardo medical AG
taBlE oF ContEntS
1. iMpoRtant SaFEtY DEtailS 31 1.1. Warnings, instructions and advisory symbols 31
2. pRoDUCt DESCRiption 32 2.1. Suckling Curve 32
3. appliCation FoR USE 32 3.1. Indications 32
3.2. Contra- indications 32
4. StaRt-Up / opERation 33 4.1. Icons 33
4.2. Installation 34
4.2.1. Electrical connection 34
4.2.2. Connecting the PumpSet 34
4.2.3. Installing the bottle holder 34
4.2.4. Installing the EasyFreeze holder 34
4.2.5. Installing the Mobile stand (optinal) 35
4.3. Operation 35
4.3.1. Switch on / switch off 35
4.3.2. Regulation of the vacuum and display 35
4.3.3. Cycle regulation and display 35
4.3.4. Switch stimulation mode /
expression mode 36
4.3.5. Using express mode,
skipping stimulation mode 36
4.3.6. Timing / Time Button 36
4.3.7. Backlight 36
5. ERRoR analYSiS / ERRoR DiSplaY / SERViCE inDiCatoR 37 5.1. Further possible errors when using the unit 37
6. MaintEnanCE / CaRE / ClEaninG 37 6.1. Visual inspection before switch-on 37
6.2. Cleaning / disinfection 38
6.3. General Information: Maintenance 38
6.3.1. Checklist: Maintenance inspections
for rentals 38
6.3.2. Checklist: Maintenance inspection
for hospitals 38
7. DEtERMininG tHE CaUSE oF FaUltS 39
8. REpaiR / REplaCEMEnt 40 8.1. General Note: maintenance / repair 40
8.2. Tools Required 40
8.3. Testing Equipment Required 40
8.4. Opening the Pump Assembly 40
8.5. Removal of the Pump Assembly 41
8.6. Installing the Pump Assembly 41
8.7. Closing the Pump Assembly 43
8.8. Replacing the Power supply complete
Art. Nr. 99.03.776 43
8.9. Replacing the Display (with 3 .5 « TFT) PCB
Art. No. 99.03.496 44
8.10. Replacing the Lithium Ion Battery Pack
Art. No. 99.03.498 45
8.11. Replacing the Control PCB
Art. No. 99.03.475 or Art. No. 99.03.770 46
8.12. Replacing the Pump Assembly
Art. No. 99.03.782 47
8.13. Replacing the Left and Right Cylinder
Art. No. 99.03.781 48
8.14. Replacing the Housing Cover Art. No. 99.03.777 49
8.15. Replacing the Housing Base Art. No. 99.03.570 50
9. aCCESSoRiESE / SpaRE paRtS 52 9.1. Spare parts for repairs 52
10. GUaRantEE 53
11. tECHniCal inFoRMation 53 11.1. Version 53
11.2. Technical data 53
11.3. Compliance 54
11.4. EMV Test records 54
12. tRanSpoRt / StoRaGE / DiSpoSal 54 12.1. Transport and storage conditions 54
12.2. Disposal 54
13. annEx 55 13.1. EMC (Electromagnetic compatibility) –
Records of Compliance 55
-
en en
31 Ardo medical AG
1. iMpoRtant SaFEtY DEtailS
Carum is an electrically-operated, safe and reliable breastpump for personal use. Carum is constructed and assembled in accordance with CE guidelines, and in compliance with Security Standards EN 60601-1, EN 60601-1-2, EN ISO 10079-1, EN 60601-1-4 and EN 60601-1-6 and EN 60601-1-11 . For your safety, to avoid damage to the unit and to avoid risk of electric shock, injury or fi re, please observe the following safety regulations.
1.1. Warnings, instructions and advisory symbols
WaRninG
Indicates a possible dangerous situation which, unless it is avoided, may lead to death or severe injury.
CaUtion
Indicates a possible dangerous situation which, unless it is avoided, may lead to slight or moderate injury.
notEIndicates possible non-injury threatening situation,. which may however lead to the risk of, for example, property damage, or have a negative infl uence on functionality or reliability.
Warning: General Hazard Area. Observe Safety Instructions.
Important: general information for the operation or repair of the pump.
Safety Regulations • Afteruse,disconnectCarumfromthemains.• Intheeventofanaccidentalsplash,ortheunitfallingintowater-IMMEDIATELYdisconnectitfromthemains.
Do not touch the unit until it has been disconnected.• Disconnectionfromthemainsisonlyguaranteedifthemainsplugispulledoutofthesocket. • Theunitisdesignedforindooruseonly.Donotuseoutdoors.• UseonlytheoriginalpowercordorcarbatterycablefromARDO.• Donotopentheunit(resultsinvoidingofthewarranty).• Carumisnotwater-proof.Thereforetheunitmustneverbeimmersedinwaterorotherfluids,norusedor
stored in places where it might fall into the water. Do not use while in the bath or while taking a shower.• Keepthemainscableandallaccessoriesawayfromsourcesofheat.• DonotuseCaruminanareawhereoxygengasisinuseorspraysarebeingapplied.• OnlyuseagenuineARDOPumpSetandgenuineARDOparts.SeeSection4.(Start-up/Operation)andSection
8 (Maintenance / Care / Cleaning).• CarumshouldonlybeusedforthepurposesdetailedinSection.2(Application).
Battery option and car battery adapter• Theunitisdesignedforindooruseonly.Donotuseoutdoors.• Iftheoperatingtimeofthebatterydeclinesfromfullychargedtolessthanonehour(ofcharge),thebattery
should be changed as soon as possible, to ensure safe operation of the Carum when on battery power. See Section 8: Maintenance / Care / Cleaning.
• DonotusetheCarumoutintheopenwhenoperatingonthecarcigarette-cable.• Caution:Ensurethatthecarbatterydelivers12volts.• Theunitmustnotbealteredorchangedwithouttheexpresspermissionofthemanufacturer. • AnyalterationorchangetotheCarumisnotpermitted. • IftheCarummustbealteredorchangedtoensuresafecontinueduse,theappropriatestudiesandtesting
must be carried out. • PleasekeepthisUserManualhandyforreferenceatalltimes.
-
en en
32 Ardo medical AG
2. pRoDUCt DESCRiption
The Carum Pump System consists of the breast pump and a PumpSet (single or double pumpset).The Carum breast pump is a quiet, user-friendly double piston pump for hospital use and for rental use.The unit is equipped with a robust plastic housing (Quality: VO) and a simplified, logically organised keyboard with a LCD display readout.The energy-saving electrical equipment is designed for mains- and for car battery-operation, or for (optional) battery use.The sophisticated controls allow each mother to enjoy a comfortable and efficient pumping experience (plus a tested suckling curve) thanks to the flexible system for adjusting vacuum and cycles.
2.1. Suckling Curve The suckling curve is characteristically stable over the entire vacuum and cycle area and corresponds closely to your baby’s suckling behaviour. Therefore the breast tissue is not additionally stressed by pumping.
Stimulation mode Exchange Stimulation-Expression Expression mode
3. appliCation FoR USE
The Carum breastpump system is intended for the expression of breast milk, for the collection and storage of mother’s milk in the event of nursing problems e.g. galactostasis, sore nipples, illness or in the absence of either mother or baby e.g. premature infants. Carum can be used as both a single pump and as a double pump. The unit is intended for indoor use only.
3.1. indicationsCarum • Toalleviateengorgement. • Toincreaseandmaintaintheabilitytobreastfeed. • Ensurethesupplyofbreastmilkshouldmotherandbabybeseparated. • Carumcanbeusedasbothasinglepumpandasadoublepump.
3.2. Contra- indicationsNone known: when the unit is used for the correct purpose.
-
en en
33 Ardo medical AG
4. StaRt-Up / opERation
4.1. Icons
Read the manual
Important information
“ON / OFF“ switch
Suckle strength / Adjust the vacuum (V)
Change the cycles level / number (C)
Switch stimulation mode / express mode
Backlight change (light / dark)
Time measurement (on / stop / off )
Unit with the applied part BF
Cleaning
Single pumpset / Double pumpset (applied part)
Warning: General risk points – check the Instruction Manual
Assembling
0123Compliance with general guidelines: MDD 93 / 42 EWG and MDD 2009
Manufacturer
The unit is compliant with the stipulations of the WEEE 2002 / 96 / EG guidelines
Mains cable
Device of protection category II
ip21 IP 21 drip proof
li-ion Contains a lithium ion battery
Express mode
Stimulation mode
Device is recyclable
-
en en
34 Ardo medical AG
4.2. installation 4.2.1. Electrical connection• Mainsoperation:Insertmainscable(A)intotheCarummains(B)input
socket (Fig.1) and connect the local power plug to the mains power supply.
• Batteryoperation:Beforeusingthebatteryforthefirsttime,connectthe unit to the mains power supply for 6 hours.
4.2.2. Connecting the PumpSetUse of the PumpSet is detailed in the separate user manual: “ARDO PumpSet“. The PumpSet is an applied part of the pump. • TousethesinglePumpSet:
Attach the tube connector (C) to the unit (D). (Fig. 2).
• TousethedoublePumpSet: Attach the tube connector to the unit (Fig 3).
• Thesecondopeningofthetubeconnector(E) remains closed – by single and double pumping.
4.2.3. Installing the bottle holderThe bottle holder is used for short-term storage of the PumpSets. The bottle holder is designed to protect against accidental spilling of milk. The bottle holder can be used free-standing or be plugged into the back of the Carum
4.2.4. Installing the EasyFreeze holderThe easy freeze holder is used for short-term hang up of the PumpSet with bag. The EasyFreeze holder is designed to protect against accidental spilling of milk. The EasyFreeze holder can be plugged onto the back of the Carum instead of the bottle holder.
B
a
D
C
E
-
en en
35 Ardo medical AG
4.2.5. Installing the Mobile stand (optinal)
The installing of the Carum mobile stand is described in separate instruction.
4.3. operationVacuum and cycles can be adjusted independently of each other. Select the adjustment which is the most comfortable for you.
4.3.1. Switch on / switch offSwitch on the unit by pressing the “ON / OFF” button once.A ‘welcome’ page will be displayed.When it is switched on, the breast pump is in stimulation mode (orange display). Medium vacuum and cycle level. To turn the unit off, press the “ON / OFF” button once again.The “ON / OFF” function is the same for battery operation, mains power supply or 12vDC power supply (car battery operation).Display mode:The operation mode is shown on the illuminated LCD display. The display turns off when the unit is switched off.
4.3.2. Regulation of the vacuum and displayThe vacuum (suction strength) can be adjusted, one level at a time, by repeatedly pressing, or by holding down the vacuum button (to the left side of the key pad). The vacuum strength can be individually regulated in both stimulation mode and expression mode. Pressing the ‘up’ arrow increases the vacuum, pressing the ‘down’ arrow reduces the vacuum. The vacuum strength is displayed on the bar chart to the left.
4.3.3. Cycle regulation and display The cycles (frequency) can be adjusted, one level at a time, by repeatedly pressing, or by holding down the cycle button (to the right side of the key pad). The cycle can be individually regulated in both stimulation mode and expression mode. Pressing the ‘up’ arrow increases the cycles level, pressing the ‘down’ arrow reduces the cycles level. The cycles level is displayed on the bar chart to the right.
-
en en
36 Ardo medical AG
4.3.4. Switch stimulation mode / expression modeTo change the stimulation mode display screen (orange display) to the express mode display screen (green display), press the ‘Mode’ button just to the right of the “ON / OFF” button.The change can be made at any time by pressing the ‘Mode’ button.If the Carum is in stimulation mode and the ‘Mode’ button is not pressed, the unit will automatically change to express mode after 2 minutes. Press the ‘Mode’ key to return to stimulation mode.When the mode changes, the background colour of the LCD screen changes accordingly. In addition, the symbol relevant to stimulation or expression mode will be highlighted on the LCD display.Background colour in stimulation mode is orange.Background colour in expression mode is green.In each separate mode of operation, vacuum strength and cycles level can be individually adjusted.When changing from stimulation mode to expression mode, the vacuum strength will automatically be set to 30mbar, and the cycles level to 45 cycles, from the previously set values.When changing from express mode to stimulation mode, the vacuum strength will be set to 60mbar, and the cycles level to 90 cycles. These values also apply when the unit is switched on.
4.3.5. Using express mode, skipping stimulation modeIf the ‘mode’ button is pressed within 2 seconds of switching on the Carum, stimulation mode is skipped. Express mode starts with an especially mild, gentle and low level vacuum of 30mbar and a cycles level of 30 cycles.
4.3.6. Timing / Time ButtonTiming starts automatically when the Carum is switched on. To stop timing, press the ‘time’ button. Timing will remain de-activated until re-started by pressing the ‘time’ button again, or until the unit is turned off. The timer will be reset to 00:00 when the ‘time’ button is pressed for a second time. Press the ‘time’ button again to restart the timer. The timer can be switched on and off in either stimulation or express mode.
4.3.7. BacklightUsing the Carum in a darkened room can be made more comfortable by dimming the intensity of the LCD. To reduce the intensity of the backlighting, press the ‘light’ button (to the left of the LDC) once.The dim function can be turned on and off as required.
-
en en
37 Ardo medical AG
5. ERRoR analYSiS / ERRoR DiSplaY / SERViCE inDiCatoR
Identifiable errors or service / maintenance schedules are represented on the LCD by a warning triangle.
Switch the unit off, and then on again. If the error message comes up again, return the unit to the supplier. The unit must be repaired by trained service personnel, or by the manufacturer.
Error: the unit does not work on battery power i.e. the pump does not operate. Solution: Connect the unit to the mains to recharge the battery. The battery charging time is ca. 4 hrs. Advice: The unit may be used while the battery is being charged, but this will extend the charging time.
5.1. Further possible errors when using the unit• Weakorabsentvacuum Disconnect the PumpSet from the unit. Partly close the connector opening with a finger to test the suction.
If suction is de-tectable, then the pump is working. Test the PumpSet in accordance with the separate instructions, (only use the original ARDO PumpSet)
• Thepumpisnotworking/nodataontheLCDdisplay Check the mains supply; if the unit still does not work, it is defective
Should any of these errors occur, and be confirmed, return the unit to the supplier immediately. The unit must be repaired by trained service personnel or by the manufacturer
6. MaintEnanCE / CaRE / ClEaninG
The silicone pot integrated into the PumpSet constitutes a total barrier, and ensures that neither breast milk nor pathogenic agents can get into the pump unit. The unit is maintenance-free; care of the unit, and its accessories, is limited to inspection and cleaning.
The Carum breast pump can be repaired, but only by trained service personnel or by the manufacturer. For this reason, only the bottle holder, the EasyFreeze holder, the connector cable, and the car charging cable, are available for order as spare parts. See Section: Accessories / Spare Parts.
6.1. Visual inspection before switch-on• Isthepumphousingintact?• Isthemainspowerunitundamaged?• Istheaccessoryclean,completeandintact?• Istheaccessorycorrectlyattached?
Ignoring these checks, or the cleaning instructions, may lead to risk of infection, electric-shock or damage to the unit.
-
en en
38 Ardo medical AG
6.2. Cleaning / disinfection• Carumbreastmilkpump:important: unplug the unit from the mains connection before cleaning. • TheCarumisnotwaterproof.ItconformstoenclosureIP(IngressProtection)ratingIP20.• Cleaningathome:Asneeded,wipewithadampcloth,commercialdetergentmaybeused.• Cleaninginthehospital:Disinfectbywipingwithanon-phenolbaseddisinfectant.• Donotputtheunitintothedish-washerorintoawashing-upbowl.Itisnotpossibletocleanthepumporthe
power unit. Attempting to do so may cause the unit to malfunction or to become contaminated – requiring suitable disposal. (See section 11: Transport / Storage / Disposal).
• AccessoryPumpSet:PleasereadtheseparateUserManual,Section:Cleaning.
6.3. General information: Maintenance• TheCarumbreastpumpismaintenance-freeasmentioned:Stillwerecommend(useintherentingorinthe
hospital application) following controlling/maintenance intervals. They can be carried out by the sales manager (renting) or (renting) or hospital staff. These controls are also prescribed in the instruction manual.
6.3.1. Checklist: Maintenance inspections for rentals
• Visualinspectionofthepumpforcleanliness After every renting
• Visualinspectionofthepumpforcompleteness After every renting
• Visualinspectionofthepumpfordamage After every renting
• Simpleoperationalchecksofthepump(Pumprunsproperly,batterycharged After every renting
• Checkofthevacuumstrength–tobecarriedoutbytrainedservicepersonnel Minimum every 2 years
6.3.2. Checklist: Maintenance inspection for hospitals
• Visualinspectionofthepumpforcleanliness Before every use
• Visualinspectionofthepumpfordamage After every use
• Checkofthevacuumstrength–tobecarriedoutbytrainedservicepersonnel Minimum every 2 years
-
en en
39 Ardo medical AG
7. DEtERMininG tHE CaUSE oF FaUltS
The ‘fault tree’, below, should allow service personnel to diagnose the possible source of faults, and carry out thenecessary repair work.
Star
t
Caru
m fa
ult t
ree
End
Vacu
um to
o lo
w
Det
erm
inat
ion
of fa
ult
Pum
p ru
nsa
pow
er s
ourc
e?
Non
ope
rabl
e
Batt
ery
ok?
Pow
ersu
pply
ok?
Keyb
oard
ok?
Pow
er s
uppl
y re
plac
ed
Cove
r rep
lace
d
Dis
play
repl
aced
Dis
play
ok?
Dis
play
show
s er
ror?
Pum
p un
it re
plac
edPu
mp
unit
ok?
Cont
rol P
rint
repl
aced
?
Pum
pset
right
?
Cylin
der r
epla
ced
Pum
pset
repl
ace
/ tr
ain
Batt
ery
repl
aced
Caru
mfu
nctio
nal?
Def
ectiv
e ho
usin
g
Cove
r rep
lace
d
Oth
er e
rror
s, fi
rst
dete
rmin
e
Repl
ace
case
bac
k
II
O
I
O OO
O O O O
O
OO
Case
bac
kok
?
yes
no
no no no no no no
no no
yes
yes
yes
yes
yes
yes
yes
yes
yes
no
-
en en
40 Ardo medical AG
8. REpaiR / REplaCEMEnt
WaRninG
Repairs may only be carried out by suitably trained personnel.The described checks or controls may not be changed or varied in any respect.Defective Carum parts may only be replaced with parts recommended by the manufacturer.
8.1. General note: maintenance / repair
WaRninG
The Carum must be disconnected from the mains before it is opened for repair or replacement work.
8.2. tools Required
notE
The following tools will be required for all types of repair or replacement work:• TORX®-screwdriverTX20length>=100mm• TORX®-screwdriverTX10length>=60mm• TORX®-screwdriverTX08length>=60mm• Largescrewdriver0or1
8.3. testing Equipment Required
notEThe following testing equipment will be required to check the pump:• Pumpset(Art.No.63.00.40)• Pressuregauge,complete(Art.No.63.00.113)
8.4. opening the pump assembly
WaRninGDisconnect the Carum from the mains.
– Turn the Carum upside down and set it down. Place it on asecure surface.
– Remove the 4 feet (Art. No. 99.03.499).– Unscrew the 6 screws (Art. No. 99.03.751) using the
TORX®-screwdriverTX20.
– Carefully turn the Carum over so that the pump is standing on the base.
– Lift off the housing cover. Disconnect the cable (Art. No. 99.03.774), connecting the control PCB to the display PCB, from the housing cover.
– If the optional Lithium Ion battery pack (Art. No. 99.03.498) has been installed, disconnect this cable from the PCB.
-
en en
41 Ardo medical AG
– Place the housing cover on its side
8.5. Removal of the pump assembly
– Open the Carum pump assembly following the instructions in Section 8.4. opening the pump assembly.
– Disconnect the connecting cable (Art. No. 99.03.773), mains-adapter PCB to the control PCB, from the control PCB (Art. No. 99.03.475 or Art. No. 99.03.770).
– Unplug the connecting cable (Art. No. 99.03.482) between the DC socket and the control PCB (Art. No. 99.03.475 or Art. No. 99.03.770).
– Remove the plugs (Art. No. 99.03.493) from the left and right sides of the housing base.
– Unscrew the 4 pump-anchoring screws (Art. No. 99.03.751) usingtheTORX®TX20screwdriver.
– Carefully remove the small mounting bracket (Art. No. 99.03.574) and the large mounting bracket (Art. No. 99.03.573).
– Remove the mounting plate (Art. No. 99.03.442).– Lift the pump (Art. No. 99.03.782) from the base of the housing.
8.6. installing the pump assembly
– Installing the Pump Assembly.
notE– When inserting, ensure that the protrusions line up with,
and properly engage with the entry holes, and lie flat.
-
en en
42 Ardo medical AG
– Secure the mounting plate (Art. No. 99.03.442) onto the pump by means of the 4 protrusions (Art. No. 99.03.444).
notE– Secure the mounting plate (art. no. 99.03.442) onto the
pump by means of the 4 protrusions (art. no. 99.03.444).
– Carefully line up the small mounting bracket (Art. No. 99.03.574) and the large mounting bracket (Art. No. 99.03.573) with the intended mounting points, and position over the mounting plate (Art. No. 99.03.442).
– Secure the small mounting bracket (Art. No. 99.03.574) and the large mounting bracket (Art. No. 99.03.573) to the housing floorwiththe4screws(Art.No.99.03.751)usingtheTORX®screwdriverTX20.
– Insert the left and right plugs (99.03.493) into the intended socket positions in the housing base.
notE– take care when plugging in. the Vacuum hose must not
be kinked. the hose is fitted with a loop, as shown in the illustration.
– Plug the connecting cable (Art. No. 99.03.773) into the control PCB (Art. No. 99.03.475 or Art. No. 99.03.770).
– Plug the connecting cable (Art. No. 99.03.482), from the DC-socket to the control PCB, into the control PCB (Art. No. 99.03.475 or Art. No. 99.03.770).
– Close the Carum pump assembly following the instructions in: Section 8.7. Closing the pump assembly
-
en en
43 Ardo medical AG
8.7. Closing the pump assembly
– Position the housing cover, and plug the connecting cable, control PCB to display PCB (Art. No. 99.03.774) into the housing cover.
– Fit the housing cover onto the housing base.
– Carefully turn the Carum over and, place it upside down on a secure surface.
– Tightenupthe6screws(99.03.751)usingtheTORX®screwdriverTX20.
– Turn the Carum over so that the pump stands on its feet.– Connect the Carum to the mains and carry out a functionality
test.
8.8. Replacing the power supply complete art. nr. 99.03.776
– Open the Carum pump assembly following the instructions in Section 8.4.: opening the pump assembly.
– Disconnect the connecting cable (Art. No. 99.03.773), mains power supply unit to PCB, from the mains power supply unit
– Disconnect the connecting cable (Art. No. 99.03.481), the power socket to the power supply unit, at the PCB.
-
en en
44 Ardo medical AG
– Unscrew the 2 screws (Art. No. 99.03.751) on the power supply unit(99.03.776),usingtheTORX®screwdriverTX20.
– Lift the power supply unit from the housing.
– Remove the connecting cable (Art. No. 99.03.773), power supply unit to the PCB, at the power supply unit.
– Plug the connecting cable (Art. No. 99.03.773), power supply unit to PCB, into the new power supply unit (Art. No. 99.03.776).
– Slide the power supply unit into the correct position in the base of the housing.
– Screw the power supply unit tightly onto the base of the housing with the 2 screws (Art. No. 99.03.751).
– Plug the connecting cable (Art. No. 99.03.481), power socket to power supply unit, into the PCB.
– Close the Carum pump assembly following the instructions in Section 8.7. Closing the pump assembly.
8.9. Replacing the Display (with 3 .5 « tFt) pCB art. no. 99.03.496
– Open the Carum pump assembly following the instructions in Section 8.4. opening the pump assembly.
-
en en
45 Ardo medical AG
– Disconnect the membrane keyboard ribbon cable (Art. No. 99.03.572) from the display (with 3.5“ TFT) PCB (Art. No. 99.03.496).
– Remove the 4 fixing plates from the PCB board (Art. No. 99.03.495).
– Remove the display (with 3.5“ TFT) PCB (Art. No. 99.03.496).
– Install the new display (with 3.5“ TFT) PCB (Art. No. 99.03.496) into the housing.
– Position the 4 mounting plates onto the PCB (Art. No. 99.03.495)– Secure the LCD PCB with the 4 screws (Art. No. 99.03.752) using
theTORX®screwdriverTX10.– Connect the membrane keyboard ribbon cable
(Art. No. 99.03.496).
– Close the Carum pump assembly following the instructions in Section 8.7. Closing the pump assembly.
8.10. Replacing the lithium ion Battery pack art. no. 99.03.498
– Open the Carum pump assembly following the instructions in Section 8.4. opening the pump assembly.
– Unscrew the 4 screws (Art. No. 99.03.571) on the battery holder (Art.No.99.03.443)usingtheTORX®screwdriverTX20andremove the Lithium Ion battery pack (Art. No. 99.03.498).
– Insert the new Lithium Ion battery pack (Art. No. 99.03.498) into the housing cover.
– Secure the battery pack in place with the battery holder (Art. No. 99.03.443) and the 4 screws (Art. No. 99.03.571) using theTORX®screwdriverTX20.
-
en en
46 Ardo medical AG
– Close the Carum pump assembly following the instructions in Section 8.7. Closing the pump assembly.
8.11. Replacing the Control pCB art. no. 99.03.475 or art. no. 99.03.770
– Remove the pump assembly following the instructions in Section 8.5. Removing the pump assembly.
– Disconnect the connecting cable(Art. No. 99.03.774), control PCB to display PCB, from the control PCB board (Art. No. 99.03.475 or Art. No. 99.03.770).
– Disconnect the motor cable (Art. No. 99.03.473) from the control PCB (Art. No. 99.03.475 or Art. No. 99.03.770).
– Release the 2 screws (Art. No. 99.03.759) on the pump housing usingtheTORX®screwdriverTX8.
– Remove the control PCB (Art. No. 99.03.475 or Art. No. 99.03.770) from the holder.
notE
– Caution: there are two photo sensors on the underside of the pCB which activate the pump assembly. improper removal or installation can damage parts of the pump assembly and / or the pCB. this can lead to functional failures of the Carum.
– Slide the new control PCB (Art. No. 99.03.475 or Art. No. 99.03.770) into the holder and install.
– Attach the control PCB to the unit housing with the 2 screws (Art.No.99.03.759)usingtheTORX®screwdriverTX8.
-
en en
47 Ardo medical AG
– Plug the motor cable (Art. No. 99.03.473) into the control PCB (Art. No. 99.03.475 or Art. No. 99.03.770).
– Plug the connecting cable (Art. No. 99.03.774), control PCB to display PCB, to the control PCB (Art. No. 99.03.475 or Art. No. 99.03.770).
– Install the pump assembly following the instructions in Section 8.6. installing the pump assembly.
8.12. Replacing the pump assembly art. no. 99.03.782
– Remove the pump assembly following the instructions in Section 8.5. Removing the pump assembly.
– Unplug the connecting cable (Art. No. 99.03.774), control PCB to display PCB , from the control PCB (Art. No. 99.03.475 or Art. No. 99.03.770).
– Disconnect the motor cable (Art. No. 99.03.473) from the control PCB (Art. No. 99.03.475 or Art. No. 99.03.770).
– Unscrew the two screws (Art. No. 99.03.759) in the pump unit housingusingtheTORX®screwdriverTX8.
– Release the control PCB (Art. No. 99.03.475 or Art. No. 99.03.770) from the holder and remove.
-
en en
48 Ardo medical AG
notE
– Caution: there are two photo sensors on the underside of the pCB which activate the pump assembly. improper removal or installation can damage parts of the pump assembly and / or the pCB. this can lead to functional failures of the Carum.
– Slide the control PCB (Art. No. 99.03.475 or Art. No. 99.03.770) into the new pump assembly holder and install.
– Secure the control PCB (Art. No. 99.03.475 or Art. No. 99.03.770) with the two screws (Art. No. 99.03.759) to the pump assembly housing,usingtheTORX®screwdriverTX8.
– Connect the motor cable (Art. No. 99.03.473) to the PCB.– Plug the connecting cable (Art. No. 99.03.774), control PCB to
display PCB, into the control PCB (Art. No. 99.03.475 or Art. No. 99.03.770).
– Install the pump assembly following the instructions in Section 8.6. installing the pump assembly.
8.13. Replacing the left and Right Cylinder art. no. 99.03.781
– Remove the pump assembly following the instructions in Section 8.5. Removing the pump assembly.
-
en en
49 Ardo medical AG
– Release the locking washers (Art. No. 99.03.758) from each side using the appropriate tool.
– Remove the complete vacuum cylinder units (Art. No. 99.03.486) from both sides.
– Install the new vacuum cylinder units (Art. No. 99.03.486).– Install the locking washers (Art. No. 99.03.758) on both sides,
using the appropriate tool.
– Install the pump assembly following the instructions in Section 8.6. installing the pump assembly.
8.14. Replacing the Housing Cover art. no. 99.03.777
– Open the Carum pump assembly following the instructions in Section 8.4. opening the pump assembly.
– Disconnect the membrane keyboard ribbon cable (Art. No. 99.03.572) from the display (with 3.5“ TFT) PCB (Art. No. 99.03.496).
– Unscrew the 4 screws on the LCD board using the TORX®screwdriverTX10.
– Remove the 4 fixing plates on the PCB (Art. No. 99.03.495).– Remove the display (with 3.5” TFT) PCB (Art. No. 99.03.496).
– If the optional Lithium Ion battery pack (Art. No. 99.03.498) is installed, unscrew the 4 screws (Art. No. 99.03.752) on the batteryholder(Art.No.99.03.443)usingtheTORX®screwdriverTX20andremovethebatterypack.
-
en en
50 Ardo medical AG
– Replace the housing cover with a new housing cover (art. no. 99.03.777).
– If the optional Lithium Ion battery pack (Art. No. 99.03.498) was installed, reinstall the battery pack in the housingcover.
– Insert the battery pack into the battery pack holder (Art. No. 99.03.443) and re-attach the holder with the 4 screws, (Art.No.99.03.752)usingtheTORX®screwdriverTX20.
– Install the display (with 3.5“ TFT) PCB (Art. No. 99.03.496) in the housing cover.
– Position the 4 fixing plates on the PCB (Art. No. 99.03.495).– Secure the LCD PCB board with the 4 screws (Art. No. 99.03.752).– Connect the membrane keyboard ribbon cable
(Art. No. 99.03.496).
– Close the Carum pump assembly following the instructions in Section 8.7. Closing the pump assembly.
8.15. Replacing the Housing Base art. no. 99.03.570
– Remove the pump assembly following the instructions in Section 8.5. Removing the pump assembly.
– Disconnect the connecting cable (Art. No. 99.03.481), power socket to power supply unit, from the PCB board.
– Unscrew the 2 screws (Art. No. 99.03.751) on the power supply unit(Art.No.99.03.776)usingtheTORX®screwdriverTX20.
-
en en
51 Ardo medical AG
– Remove the power supply unit from the housing.
– Unplug the connecting cable (Art. No. 99.03.482), DC socket to control PCB, from the housing base.
– Remove the connecting cable (Art. No. 99.03.773), power supply unit to PCB, from the housing base.
– Remove the connecting point (Art. No. 99.03.444) from the housing base.
– Replace the old housing base with the new housing base.
– Connect the connecting cable (Art. No. 99.03.773), power supply unit to PCB, to the new housing base.
– Connect the connecting cable (Art. No. 99.03.482), DC socket to control PCB, to the new housing base.
– Install the power supply unit in the intended place in the housing base.
– Firmly screw in the power supply unit with the two screws (Art. No. 99.03.751) to the housing base.
– Plug the connecting cable (Art. No. 99.03.481), mains socket to power unit, into the power supply unit.
– Install the pump unit following the instructions in Section 8.6. installing the pump assembly.
-
en en
52 Ardo medical AG
9. aCCESSoRiESE / SpaRE paRtS
Mains cable Euro Art. No. 63.00.130
MainscableUK Art. No. 63.00.131
Mains cable UL Art. No. 63.00.132
Trolley - optional Art. No. 63.00.129
Car battery adapter 12VDC - optional Art. No. 63.00.133
Bottle holder Art. No. 63.00.127
EasyFreeze holder Art. No. 63.00.128
Various pumpsets www.ardomedical.com
9.1. Spare parts for repairs
Product Name Quantity Article ID No.
Mounting plate / fixing plate 1 99.03.442
Control PCB (Printed Circuit Board) 1 99.03.475
Connecting-cable, power socket to power supply unit 1 99.03.481
Connecting-cable DC socket – PCB 1 99.03.482
Connector-plug 1 99.03.493
PCB mounting plate 1 99.03.495
PCB screen with 3.5” TFT display 1 99.03.496
Lithium Ion battery pack 1 99.03.498
Carum support-stand 1 99.03.499
Lens screw 4 x 12 1 99.03.751
Lens screw 3 x 12 1 99.03.752
Safety washer 1 99.03.758
LensscrewKA25x6 1 99.03.759
PCB with battery switch circuit 1 99.03.770
PCB power connecting cable 1 99.03.773
Connecting-cable Control PCB – display PCB 1 99.03.774
Power supply complete 1 99.03.776
Carum housing cover 1 99.03.777
Carum cylinder set 1 99.03.781
Pump assembly 1 99.03.782
-
en en
53 Ardo medical AG
10. GUaRantEE
The Carum is guaranteed for 3 years from the date of purchase. The purchase invoice serves as proof of guarantee.
General terms and conditionsThis unit is covered by the Ardo medical AG warranty in respect of materials and manufacturing defects.The guarantee starts from the date of purchase. Defective materials will be replaced free-of-charge, during the life of the Guarantee, provided that the Guarantee has not been voided by unauthorised use or access to the unit. To ensure guarantee cover, and the flawless functioning of the unit, the instructions in the User Manual must be followed and only Ardo medical AG accessories used. The guarantee is void if the unit has been opened.
11. tECHniCal inFoRMation
11.1. Version
Carum with Euro mains cable Art. No. 63.03.71 Carum battery with Euro mains cable Art. No. 63.03.75
CarumwithUKmainscable Art.No.63.03.72 CarumbatterywithUKmainscable Art.No.63.03.76
Carum with UL mains cable Art. No. 63.03.73 Carum battery with UL mains cable Art. No. 63.03.77
11.2. technical data
Pump dimensions LxBxH=302x240x180mm Power cable length Length 3 m
Pump weight 3 kg Weight with battery 3.7 kg
Input voltage 100 – 230 VAC + / -10% Frequency 50 – 60Hz
DC Input 12 VDC Lifetime (excluding parts subject to wear and tear)
10 years
Operating temperatures
+5 – +40 °C Relative humidity 10 – 93%
Protection level BF Air pressure 700 – 1060 hPa
Protection Category Double Insulated
Category II Classification in accordance with MDD93 / 42
Category 2a
Cycle and vacuum levels
Vacuum range in stimulation mode
30 – 150 mbar 22 – 114 mmHg 3 – 15 kPa
Cycle range in stimulation mode
72 – 120 / min.
Vacuum range in expression mode
30 – 330 mbar 22 – 250 mmHg 3 – 33 kPa
Cycle range in expression mode
30 – 60 / min.
Permitted tolerance of the vacuum strength
+ / - 5% Permitted tolerance of the cycles
+ / - 3 cycles
The vacuum level generated is sensitive to prevailing atmospheric pressure. The higher the prevailing atmospheric pressure, the more intense the generated vacuum – and vice versa. The vacuum values (including tolerances) specified in the user manual are based on an atmospheric pressure of 880hPa.
-
en en
54 Ardo medical AG
11.3. Compliance
In accordance with guidelines: MDD 93 / 42 CEE of 14.7.1993, and 2010, in respect of medical products and 2010. 2007 / 47 / CE, 2006 / 42 / CE, EN 60601-1, EN60601-1-2, EN ISO 10079-1, EN 60601-1-4, EN 60601-1-6, EN 60601-1-11
0123
11.4. EMV test recordsSee annex 13.1.
12. tRanSpoRt / StoRaGE / DiSpoSal
For the transport and storage of your Carum, we recommend using the original packaging. We further recommend separate storage for the unit is protected from UV (e.g. sunlight / halogen lamps), heat and dust.
12.1. transport and storage conditionsCarum und Carum with optional battery
Temperature -20 – +70°C
Relative humidity 10 – 80%
Air pressure 700 – 1060 hPa
12.2. DisposalThe unit may be returned to the manufacturer, or be dismantled and recycled (as metal or plastic) in an environmentally-friendly manner, through a recycling firm specialising in electrical goods. There are no special or increased risk factors.
Manufacturer: Ardo medical AG Gewerbestrasse 19 6314 Unterägeri Switzerland
Homepage: www.ardomedical.com Email: info@ardo.ch
-
de de
55 Ardo medical AG
13. anlaGEn / annEx
13.1. EMV prüfaufzeichnungen / EMC (Electromagnetic compatibility) – Records of Compliance
-
Ardo medical AG Gewerbestrasse 19 6314 Unterägeri Switzerland
T +41(0)41 754 70 70 F +41(0)41 754 70 71 info@ardo.ch www.ardomedical.com
99.0
3.59
6 /
1.20
12
top related