EKG2000 SERVICE MANUAL
Version 1.0
www.ebionet.com
http://www.ebionet.com/
-
C7SM-1.01 2/70
Service Manual of Cardio 2000
Revision History
Edition Date Page Comment
Ver 1.0 2019-10-18 Initial Release
-
C7SM-1.01 3/70
Service Manual of Cardio 2000
- TABLE OF CONTENTS –
Section 1. General Information ............................................................................................... 5
1-1. Introduction ...................................................................................................................................................................... 5
1-2. Features of the Equipment ...................................................................................................................................... 5
1-3. Equipment Description .............................................................................................................................................. 6
1-4. System Installation .....................................................................................................................................................15
1-5. Terms of Warranty .......................................................................................................................................................16
1-6. Specification .................................................................................................................................................................16
Section 2. Troubleshooting .....................................................................................................18
2-1. Power .................................................................................................................................................................................18
2-2. Wave Printing ................................................................................................................................................................19
2-3. Diagnosis .........................................................................................................................................................................22
2-4. Filter ....................................................................................................................................................................................23
2-5. Trouble Shooting Guide ..........................................................................................................................................25
Section 3. Adjustment .............................................................................................................31
3-1. LCD Backlight................................................................................................................................................................31
3-2. Lead Fault ........................................................................................................................................................................32
3-3. Filter Setting ....................................................................................................................................................................33
Section 4. Block Diagram/Description ................................................................................34
4-1. Circuit Block Diagram ..............................................................................................................................................34
4-2. Equipment Overview ................................................................................................................................................35
4-3. Size and Weight ...........................................................................................................................................................38
Section 5. Disassembly and Assembly ................................................................................39
5-1. Module Description ...................................................................................................................................................39
5-2. Internal Block Diagram & Cable ........................................................................................................................46
5-3. Disassembly ...................................................................................................................................................................49
5-4. Assembly .........................................................................................................................................................................59
-
C7SM-1.01 4/70
Service Manual of Cardio 2000
Section 6. Replaceable Parts List and Accessory .............................................................60
6-1. Replaceable Part List ................................................................................................................................................60
Section 7. Connector Pin Assignment .................................................................................63
7-1. Patient Cable Port ......................................................................................................................................................63
Section 8. PCB Part Disposition ..............................................................................................64
8-1. Base (Power+Analog) Board ................................................................................................................................64
8-2. CDUI Board .....................................................................................................................................................................65
8-3. CPU Board .......................................................................................................................................................................66
Section 9. Software Upgrade .................................................................................................67
9-1. Requirement of Upgrading....................................................................................................................................67
9-2 Procedure of Upgrading (Cardio7 UI screen, upgrade method is same) .................................68
-
C7SM-1.01 5/70
Service Manual of Cardio 2000
Section 1. General Information
1-1. Introduction
CardioCare2000 is an electrocardiogram (ECG) device capable of measuring and recording 12
channels of cardiac signals of patients. The device enables operators to record cardiac signals,
review them using various viewing modes, and perform automatic detection of cardiac conditions.
In addition, the device allows operators to enter patient information for it to be printed alongside
cardiac signals, so that operators can sort and manage charts conveniently. It is also possible to
manage the digital files by transferring the stored data to a networked PC.
Furthermore, user convenience has been further enhanced by addition of a functionality that
can initiate ECG recording & saving, filter enhancing, parameter sorting and automatically
detecting cardiac signal all at once by pressing one key.
A battery (optional) can be added to the device so that the device can be operated
conveniently during bedside visits or emergency conditions.
1-2. Features of the Equipment
- Print 12-channel (ch) cardiac signals with reports in: 3ch+1 rhythm, 6ch+1 rhythm, and 12ch
rhythm patterns on A4 or Letter size printing paper.
- Record and print 12 channel rhythms continuously and simultaneously.
- Calculate heart rate, P-R-T axis and PR, QRS, QT and QTc widths from recorded signals
automatically and print them on a report for use in rhythm analysis.
- Get diagnostic reports using automatic detection functionality.
- Able to modify filter setting, signal sensitivity, printing speed, channel view settings and rhythm
settings, and print on previously recorded EKG signals to aid data analysis.
- Able to attach a battery so that the device can become portable.
- More than 200 patient‟s data can be saved and transferred to a PC through LAN.
- It offers various protocols to make it possible to link with hospitals‟ computing networks. Also File
Databases are strengthened.
-
C7SM-1.01 6/70
Service Manual of Cardio 2000
1-3. Equipment Description
The CardioCare2000 system consists of the items below. Unpack the package and check the
items below are included. Also, be sure to check for any damage to the body and accessories.
Basic Components and Accessories
① CardioCare2000 Body (1 EA) – Dimension: 290(W) 300(D) 97.5(H)mm
② Patient Cable (1 EA) – Length: 3,700mm (Max)
③ Limb Electrodes (1 SET)
④ Chest Electrodes (1 SET)
⑤ ECG Paper (1 EA)
⑥ Power Cable (1 EA) – Length: 3,000mm (Max)
-
C7SM-1.01 7/70
Service Manual of Cardio 2000
Optional Components
① Battery (1 EA) – Replaceable and Rechargeable, Lithium ion, 11.1V, 2600mA
② Hanger (1 EA)
③ Cart (1 EA)
CAUTIONS
You may have distortion or signal noise when you use nonstandard or other brand
accessories. We strongly recommend you use only the authorized accessories which we
supply.
WARNING
How to replace battery: Please make sure you use the right battery we show here.
Otherwise we are not liable for any damages and/or explosion/fire caused by using the
wrong battery.
①
②
③
-
C7SM-1.01 8/70
Service Manual of Cardio 2000
Body Configuration
▣ Top View
① Handle
② Printer Cover
③ Printer Cover Switch
④ LCD
⑤ Control Panel
①
② ③
④
⑤
③
-
C7SM-1.01 9/70
Service Manual of Cardio 2000
▣ Front View
① Printer Cover Switch
▣ Rear View
① ②
③
④ ⑤
① Protective Ground Terminal
② Power Switch
③ AC Power Connection Port ( Fuse : 250V , 5A ( High breaking type ))
④ USB Port(Use for equipment upgreades only)
⑤ RJ45 LAN Port
①
-
C7SM-1.01 10/70
Service Manual of Cardio 2000
▣ Left Side View
①
① Handle
▣ Right Side View
①
① Patient cable connection port
WARNING
There is a risk of electric shock if the Rest stand of the equipment is damaged or can
not be fixed to the product. Do not use the product and immediately ask the
manufacturer and the seller for repair
NOTE
To avoid an expected electric shock, do not open the equipment cover or
disassemble the equipment. Refer servicing to Bionet, Inc.
-
C7SM-1.01 11/70
Service Manual of Cardio 2000
LCD Panel
LCD panel displays system setting status after indicating the version of the system and
manufacturer name for 2 seconds when power is turned on.
Displayed items on the LCD are as follows.
① ② ③ ④
⑤ ⑥ ⑦
① Display ECG signal level out of 5, 10, 20, aut (Auto Gain)
② Display printing speed out of 12.5, 25, 50
③ Display base line filter setting on or off
④ Display EMG filter setting on or off
⑤ Display heart rate
⑥ Display channel form of the output report out of 3ch+1rhy, 6ch+1rhy, 12ch rhy, 60s 1rhy.
⑦ Display rhythm channel setting out of I, II, III, aVR, aVL, aVF, V1, V2, V3, V4, V5, V6..
1 0 1 2 . 5 O F F O F F
1 0 0 6 C H + 1 R H Y a V R
① ②
-
C7SM-1.01 12/70
Service Manual of Cardio 2000
Indicator Lamp
I Indicate the lead connection status.
If LED is illuminated green, connection status is good .
If LED is off, connection status is bad.
In this case, check off which one of lead connection is in lead-off status
through monitor mode output.
Caution: Start printing when LED is illuminated green.
Indicate the battery capacity status: Green and Red colors are used
to display 2 different statuses.
The battery indicator is red when the device is operational without having
an A/C power supply. When an A/C power is supplied, the battery
indicator will be either red or green depending on the charge level. No
color will be shown, if there is no battery installed in the device.
Indicate AC power is operated if LED is illuminated green and not in use if
LED is off.
Control panel
Select signal level(mm/mV) out of 5, 10, 20, aut
(I-aVF:10, V1-V6:5 )
Select printing speed(mm/sec) out of 12.5, 25, 50
Select on or off whether or not to activate the filter that eliminates base line
drift
Select on or off whether or not to activate the filter that eliminates EMG.
ECG signal can be distorted by applying this filter.
This key is excepted to use when network function is added ahead.
-
C7SM-1.01 13/70
Service Manual of Cardio 2000
Select channel form of the output report out of 3ch+1rhy, 6ch+1rhy, 12ch
rhy, 60s 1rhy
Select rhythm channel
Call up the menu to enter patient ID, name, age, sex, height, weight
Display real-time ECGs from the patient continuously or monitoring the
patient ECGs for long time,
Record ECGs for 10 seconds by pressing key and prints in A4 sized report
form after calculating the measurement parameter of the recorded ECGs
Print the same report as previously printed one in record mode or prints
after changing system setting such as filter, signal level, output speed,
channel form, rhythm cannel of previously recorded ECGs data in record
mode.
Exit during printing or recording of ECGs.
Except that, call up the menu or select menu.
Move the selected focus to the left in menu mode
Move the selected focus to the right in menu mode
-
C7SM-1.01 14/70
Service Manual of Cardio 2000
Power
When AC power is supplied to the equipment, AC power indicator lamp at the front side is
illuminated green and battery is being charged automatically.
If no battery is installed, battery indicator lamp will be off. Once a battery is plugged in and
the device is powered on, it will change its power mode to supply power to the battery. If the
device is running on the power from the battery, battery charge indicator lamp will light up
red color. If the battery is running low on power, three continuous buzzer alarms will sound. If
this occurs, turn off the device and connect the A/C power before turning the device back
on.
- Required time for charging : more than 4 hours
- Continuous battery operating time : 1 hour
Caution
For the protection of environment, do NOT throw wasted battery away. Please inquire
at the biomedical engineering lab of the hospital and dispose according to the
appropriate procedure to the authorized place(according to national regulations).
Battery Charge
Indicator Lamp
AC Power
Indicator Lamp
-
C7SM-1.01 15/70
Service Manual of Cardio 2000
1-4. System Installation
Precautions for Installation
While installing CardioCare2000, please pay attention to the following items:
- Use the equipment between the ambient temperature 5 to 40℃ and humidity 10 - 90%.
- Check the power cord is properly connected, and the probe carefully handled.
- Do not plug multiple cords in a power outlet.
- Install and operate unit on a flat surface.
- If you experience noise, ground the device.
- Do not use a power cord that may make a connection noise.
- Device settings will be recorded in the internal memory even when it is off.
- Prevent any shock or excessive force that may cause damage to the device.
- Place the device away from any dust or flammable materials.
Power Connection
The equipment needs electrical power to operate. Plug in one end of the power cable to wall
socket and the other to CardioCare2000.
Patient Cable Connection
- Connect the patient cable to the patient cable connection port on the right side of the body.
- Connect limb electrodes to RL (N), LL (F), RA(R), and LA (L) leads of the patient cable and chest
electrodes to V1 (C1), V2 (C2), V3 (C3), V4 (C4), V5 (C5), and V6 (C6), respectively.
Paper Installation
- Push the printer cover release switch to the right to open the printer door of the
CardioCare2000. Install EKG paper with the side to be recorded appearing on top.
Close the cover to finish the paper installation process.
WARNING
No modification of this equipment is allowed.
Do not modify this equipment without authorization of the manufacturer.
If this equipment is modified, appropriate inspection and testing must be conducted to
ensure continued safe use of equipment.
-
C7SM-1.01 16/70
Service Manual of Cardio 2000
1-5. Terms of Warranty
- This product was made through strict quality control and examination. The repair and
compensation standards follow the consumer damage compensation regulations of the Ministry of
Finance and Economy.
- The warranty for this product runs for 1 year. (2 years in EU) But the warranty for accessories runs for
6 months.
- This product is guaranteed against faulty workmanship for a period of 1 year. Under this warranty,
we undertake to repair free of charge at our service centers.
- When having troubles during this period, please provide us with the model name, serial number,
date of purchase and a description of the problem you are experiencing.
CAUTIONS
Federal law restricts this device to sale by or on the order of a physician
1-6. Specification
ECG Leads Simultaneous 12 Leads Resting ECG
Recording Channel 3CH+1RHY, 6CH+1RHY, 12CH,
1CH Long Time (1min)
Gain 5, 10, 20, Auto (I~aVF: 10, V1~V6: 5) mm/mV
Printing Speed 12.5, 25, 50 mm/sec
Sampling Rate Analysis Sampling Rate – 500sample/sec
Digital Sampling Rate - 8,000sample/sec
Filters AC (50/60Hz, -20dB or better),
Muscle (25~35Hz, -3dB or better),
Base Line Drift (off, 0.05, 0.1, 0.2Hz, -3dB or better),
Low Pass Filter(off, 40Hz, 100Hz, 150Hz)
-
C7SM-1.01 17/70
Service Manual of Cardio 2000
Display 2 x 16 Char LCD Display
Sensitivity, Speed, Filter Status, HR, Printing Form, Rhythm Lead
User Interface Short Key
Printer Resolution Thermal Print Head, Roll Paper
Report Paper
- Width: A4/Letter – 215mm (8.5”)
- Length: A4 – 297mm (11.7”)
Letter – 279mm (11”)
Resolution: 8dot/mm (0.125mm pitch)
Patient data ID, Name, Age, Gender, Height, Weight
Basic Measurement Heart Rate: 30~300bpm (±3bpm)
PR, QRS, QT/QTc, P-R-T axis
Electrical Internal Noise: 20uV(p-p)max
Input Impedance: More than 50MΩ
Input Voltage range: ±5mV
CMRR: Greater than 105dB
DC offset Voltage: ≥ ±400mV
Patient Leakage Current: < 10uA
Frequency Response: 0.05~150 with in –3dB
Isolated, Defibrillation and ESU Protected
Line Power Input: 100- 240Vac, 1.5-0.75A, 50-60Hz, Output: 15Vdc, 4.2A
Battery type Replaceable and Rechargeable, Lithium ion, 10.8V, 3350mA
Battery Capacity 360 minutes of continuous operation without recording or
200 ECGs in 12 channel format at 25mm/s and 10mm/mV
with a battery charged during approximately 3 hours from total
discharge (with display off)
Communication LAN
Safety Conformity Class I, Type CF Defibrillation-proof applied Part.
Environmental Operating Temperature: 5~40℃
Operating Humidity: 10~90%RH
Atmospheric Pressure: 70~106KPa
Dimensions 290(W) 300(D) 97.5(H)mm, Approx 3.0kg (Main Body)
Standard Accessory Patient Cable (1 EA), Limbs Electrodes (1 SET),
Chest Electrodes (1 SET), ECG Paper (1 EA),
Power Cord (1 EA)
Options Rechargeable Battery (1 EA),
Cart(1 EA), Hanger (1 EA)
-
C7SM-1.01 18/70
Service Manual of Cardio 2000
Section 2. Troubleshooting
2-1. Power
Problem Solution
Case1.
The EKG2000
open will turn on
and off
frequently.
Case2.
Power will not
turn on.
< General Checking >
1. Check the power cable is connected with AC Power connection at the
back of the system.
2. Check the power source of the multi-tab with connection to the power
cable. (If power is connected, AC power indicator is lighted.)
3. Check the power switch.
4. Check any of two fuses on the Power Module (SMPS) and on fuse on the
Base board is cut.
If it is, replace Power board.
Case1:
Check if the soldering of J1 on the Base board (SMPS + Base board
Connector) is poor. Re-solder the cable.
● When the EKG2000 is turned on, the pin no. 8‟s voltage has to be always
14.6V.
1) Is the pin no. 1/2‟s voltage is 14.6 Yes replace the main board.
2) Is the pin no 1/2‟s voltage is 14.6 No replace the power board.
● When the REC key is pressed,(during printer is working)
1) Pin no. 12‟s voltage has to be changed from 0V to 24V
If not, faulty power board, replace the power board.
2) Pin no. 15‟s voltage has to be changed from 3V to 0V.
If not, faulty CPU, replace the main board.
-
C7SM-1.01 19/70
Service Manual of Cardio 2000
2-2. Wave Printing
Problem Solution
1-1. The wave pattern changes from the flat line
on the left to the wave forms on the right.
1. If the normal wave patterns show up to the
right of chart as shown in problem1, 2,
diagnosis output print too early.
2. It takes some time to show the normal wave
patterns.
3. The normal wave patterns will show if output
is requested after the system is stabilized
by showing the HR variance of ±5 on the LCD
display.
4. Check connection of the limbs electrode
and the chest electrode to the body parts.
5. If used on the dry skin or in the winter
season, connectivity may be degraded. In
that case, used water or ECG gel. The
electrode must be washed after using ECG
gel. Otherwise, the electrode will be rusted
and corroded.
1-2. Some wave patterns are not shown in V1~6.
* Connection location of the limbs electrode
and the chest electrode
2. The diagnosis is not shown in print.
1. Press the menu key, showing the overlapped
circle and triangle, at the bottom of the
Keypad.
2. Press the right arrow button twice to select
PRN and then press the menu key.
3. Press the left arrow button once to select
DIAG and then press the menu key.
4. Press ESC key from the menu to enter the
basic display and then press '0' to printout.
-
C7SM-1.01 20/70
Service Manual of Cardio 2000
3-1. There are too many noises in the wave
pattern.
3-2. The noises periodically occur throughout
the wave patterns.
3-3. The thick noise is made on the wave
patterns.
3-4. The wave output looks like the noise and
are not properly shaped.
1. This problem occurs when the filter value is
not corrected.
2. Press the menu key, showing the overlapped
circle and triangle, at the bottom of the
Keypad.
3. Press the right arrow key once to select FLT
and then press the menu key.
4. Check if the filter value shows BASE: ON,
MUSC: OFF, AC: 60 Hz, LDF: 150Hz and, if not,
correct it.
5. Since the noise occurs as in problems 2 and
3 if the AC filter is turned off or set at
different frequency from the power source,
the system must be set at 50Hz or 60Hz to
correspond to the input power. (It is different
each country) The AC filter status is shown at
the left bottom of the printout as
'CUT OFF (ON) AC'.
6. Problems 3.4 occurs when the BASE is set at
OFF. Change it to ON.
-
C7SM-1.01 21/70
Service Manual of Cardio 2000
4. The wave output looks like the noise and are
not properly shaped.
1. This problem occurs when the chest
electrode is not properly in contact with the
body.
2. Check with simulator there is any cut on the
patient cable.
3. Check there is any damage to the chest
electrode.
4. Test it again after applying alcohol, water or
ECG gel to the electrode.
* If the problem can not be solved by the
above procedures, please contact us.
5. Heavy noise on the wave forms
1. The noise occurs as there is no signal from
the limbs electrode.
2. If the polar plate in the limbs electrode is
blacked or copper colored, or if ECG gel
residue remains, the signal may not be
detected. Replace it.
3. If discoloring is too noticeable in the limbs
electrode, replace it.
6. The wave pattern shows the straight line
among V1~6.
1. If some wave patterns are too small as
shown in the problem 6, the problem is most
likely to be with the patient cable, limbs
electrode or chest electrode.
2. If the problem is not with the accessories,
then the main board is the problem.
3. The main board must be shipped to us for
repair since the condenser of the main
board must be replaced.
7-1. When diagnosis output key is pressed, the
measurement goes on past 10S.
7-2. Operation stops at processing, print after
10S measurement.
1. Replace the main board.
* If the above procedure does not solve the
problem, please contact us
8. The backlight on the screen is on, but there is
no display.
1. Check the Key cable which connects Key
board and the main board and replace the
main board.
-
C7SM-1.01 22/70
Service Manual of Cardio 2000
2-3. Diagnosis
Problem Resolution
1. The diagnosis shows LVH.
1. If disease is diagnosed, abnormal ECG always
appears.
2. LVH occurs when the sum of one of V1 or V2
and one of V5 or V6 is 3.5 or higher.
3. Upgrade to the latest version of the software.
2. Diagnosis shows MI.
1. If MI is diagnosed, it is likely that the print button
was pressed before the wave patterns are
stabilized.
2. Check HR before printout and print it again.
3. Upgrade to the latest version of the software.
-
C7SM-1.01 23/70
Service Manual of Cardio 2000
2-4. Filter
Base line Filter
Base line drift is caused by a patient‟s breathing. ECG signal is overlapped on a large parabola. If
you want to set Base line drift, select „ON‟. Otherwise, select OFF. The set value appears on the
bottom part of the printed form indicating 0.1HZ- when set to ON, and 0Hz- when set to OFF.
AC Filter
In addition to ECG, other signals such as power noise, base line drift by respiration and
electromyogram may be recorded in the output signals.
There are three types of the noise filter; 60Hz, 50Hz and OFF. When set to OFF, noise is not
eliminated. Set to 50Hz, it indicates to filter 50Hz power noise. Set to 60Hz, it means filtering 60Hz
power noise.
When the system is powered by battery, power noise is not generated, so you may have clear
ECG signal if set to OFF. In Europe where the power noise is 50Hz, the system is set to 50Hz. In other
nations including Korea and the U.S., it is set to 60Hz.
The status of the set filter can be checked on the bottom part of the printout, indicating AC 60Hz,
AC 50Hz or AC OFF.
-
C7SM-1.01 24/70
Service Manual of Cardio 2000
Low Pass Filter
When ECG is not clear even after all three filters are set to ON, the use of low frequency filter will
produce clear ECG signal. The low frequency filter can be set to 40Hz, 100Hz, and 150Hz and OFF.
Set to 40Hz, it eliminates all signals over 40Hz. The set value is indicated on the lower part of the
printed form. If you set the filter to OFF, -250Hz is indicated. If set to 40Hz, -40Hz is indicated. Set to
100 Hz, -100Hz is indicated. If set to 150Hz, -150Hz is indicated.
FREQUENCY 40HZ100HZ 150HZ 250HZ
-
C7SM-1.01 25/70
Service Manual of Cardio 2000
2-5. Trouble Shooting Guide
1) Power is off
The power is off
Does the power input cable break
down?
Change the power input cable
Is the fuse blown?
Change the fuseAgain blown
fuse
Does the power switch cable short-
circuit?
Change the power switch cable
Does the main power
connector break down?
Change the main power connector
Normal operation
UseNo
Yes
Yes
Yes
Yes No
Yes
B
A
B
Does the main power module produce the
power properly?
Repair or change the main power
moduleNo
Yes
Ask the manufacturer
about it, then take action
No
Yes
A
A
A
A
A
No
No
No
-
C7SM-1.01 26/70
Service Manual of Cardio 2000
2) LCD is blank although the power is on
LCD is blank although the power is on
Broken 20 pin shield cable?
Change the 20 pin shield cable
Does the LCD cable have a poor condition of connection or short-
circuit?
Check the condition of the connection and then change if it's not good
Is the LCD module
unsatisfactory?
Change the LCD Module
Normal operation
Use
No
Yes
Yes
Yes
No
Yes
B
A
B
Ask the manufacturer
about it, then take action
A
A
A
No
No
-
C7SM-1.01 27/70
Service Manual of Cardio 2000
3) Printer is not working properly
When the printer is not working
properly
Does the paper come out while printing?
Does the motor work properly?
Does the signal come out on the paper while printing?
Normal operation
Use
No
No
Yes
B
A
B
Ask the manufacturer
about it, then take action
Have you connected the motor power
cable correctly?
Change the motorNo
Connet the motor power cable
correctlyNo
Have you connected the
TPH powor cable correctly
Connect the THP power cable
correctlyNo
Does you place on the
printer correctly?
No
Make the paper flat and place it facing up in order to let printer
cover shut?
Is the motor power cable satisfactory?
Change the motor power cable
Is the TPH power cable is satisfactory?
Change the TPH power cable
Yes
Yes
Yes
Yes
Yes
No
No
Yes
Yes
A
A
A
A
A
-
C7SM-1.01 28/70
Service Manual of Cardio 2000
4) EKG wave does not come out or it's very small
EKG wave does not come out or
small
Is the patient cable
connecter correctly?
Reconnect the patient cable
carefully
Is the patient cable short-circuited?
Is the electrode
connected to the patient
cable correctly?
Normal operation
UseNo
No
Yes
Yes
B
A
B
Is the electrode
connected to the patient
cable correctly?
A
Change the patient cable
A
Cleanse the electrode with
alcohol or water, then it use
Reconnected the electrode to the
patient cable properly
Ask the manufacturer
about it, then take action
No A
A
Yes
No
Yes
Control the volume by setting
a gain on the main board
Is the EKG signal adjusted
properly?No
Yes
A
-
C7SM-1.01 29/70
Service Manual of Cardio 2000
5) USB memory does not working properly
EKG files do not write to USB memory
properly
Is USB memory formatted in FAT file system?
Is USB memory space full ?
Does the USB sig cable have a poor condition of connection?
Is USB sig cable broken?
Format USB memory in FAT
file system
Remove all files in USB memory
Check the condition of the connection and Re-
connect the cable.
Exchange the USB sig cable
No
No
No
Yes
Yes
Yes
Yes
No
Ask the manufacturer about it, then take action
-
C7SM-1.01 30/70
Service Manual of Cardio 2000
6) USB barcode reader does not working properly
Does USB Barcode reader not work
properly?
Does the USB sig cable have a poor condition of connection?
Is USB sig cable broken?
Check the condition of the connection and Re-
connect the cable.
Exchange the USB sig cable
No
Yes
Yes
No
Ask the manufacturer about it, then take action
-
C7SM-1.01 31/70
Service Manual of Cardio 2000
Section 3. Adjustment
3-1. LCD Backlight
The LED backlight is powered by two resistors as shown in the following figure.
-
C7SM-1.01 32/70
Service Manual of Cardio 2000
3-2. Lead Fault
NOTE
Ready LED is turned off when the lead fault occurs.
-
C7SM-1.01 33/70
Service Manual of Cardio 2000
3-3. Filter Setting
AC power noise, base line drift caused by breathing , EMG signal can be recorded along with
the output ECG signal. To set the AC noise filter, move to “FLT” menu to call up “AC” menu and
press key. key toggles between off, 50hz, 60hz. Off indicates that filter does not
suppress ac interference. As for 50Hz or 60Hz, filter suppresses 50Hz or 60Hz AC power noise. In case
of battery power, ECG signal is clear only with Off setting because AC noise interference hardly
exist. The selection is 50Hz for most of Europe and 60Hz for USA and other countries. The selected
setting is shown on the bottom line of the output form indicated AC 60Hz, AC 50Hz or AC OFF
Base line drift is caused by patient breath. Base line drift filter can be set by pressing key.
BASE key switches on or off. off, 0.05, 0.1 or 0.2Hz indicates base lien filter is applied or not.
The selected setting is shown on the bottom line of the output form indicated by „0.1hz-„ as for filter
on and „0hz-„ as for filter off.
EMG signal is induced from muscle and internal organs of the patient.
For clear ECG signal output, it is necessary to eliminate EMG signal interference.
EMG filter can be set by pressing MUSC key. MUSC key switches on or off.
On or off indicates EMG filter is applied or not. The selected setting is shown on the bottom line of
the output form indicated „EMG:On‟ or „EMG:Off
Low pass filter can be selected to optimize ECG signal after applying above three filters and if
ECG signal is still interfered. To set the low pass filter, move the menu to “FLT” to call up “LPF” menu
and press key. key toggles between off, 40hz, 100hz, 150hz. As for 40Hz, it indicates
filter suppresses higher frequency than 40Hz. The selected setting is shown on the bottom line of the
output form indicated by „Off„ as for off, by‟-40hz‟ as for 40hz, by‟-100hz‟ as for 100hz, by ‟-150hz‟ as
for 150hz.
To apply filter could result in ECG distortion. So, filter should be appropriately applied to reduce
the fluctuation without affecting the ECG signal when output signal is badly interfered only.
For your reference, base line file and ac filter should be always active and EMG filter can be
applied appropriately. It is recommended for LPF to set 150Hz during diagnosis.
-
C7SM-1.01 34/70
Service Manual of Cardio 2000
Section 4. Block Diagram/Description
4-1. Circuit Block Diagram
Basic Module
CPU Module
Analog Module
Character LCD
User Interface Module
CPU
DDR32G x 2
EMMC4G
USB OTGConnector
PMIC
JTAGConnector
PRINTER ASS’Y
MCU3.3V, 5V Power
Isolator
Shield Driver
ECG AFEPACE Detec-tor
Membrane Keys Isolator
ThermalPrinterHeader
StepMotor
PaperSensor
Analog Module Connector
Analog Module Connector
User Interface Module Connector
Printer Connector
USB 1 Port
LAN PORT(RJ45)
Touchcontroller
BacklightPower
RS232Transceiver
EthernetTransceiver
PaperSensorDriver
5V, 3.3V, 4.2V
Power
MotorDriver
TPH SignalTransceiver
Charger Power MCU
System Power Switch
Coin Battery Buzzer
User Interface Module ConnectorPSU ConnectorBattery Connector
Ready LED
AC / Battery Indicator LED
24VPrinter Power
PrinterMCU
Battery Cable PSU CableUser Interface Signal Cable Analog Board Cable
Printer Cable
Character LCD Pin Header
Character LCD Pin Header
Patient Connector
Power SupplyUnit
Battery
Uartport
-
C7SM-1.01 35/70
Service Manual of Cardio 2000
4-2. Equipment Overview
1) Top
A : Printer Assembly
B : LCD Assembly
C : Control Panel
2) Bottom
A : Battery Cover
A C
B
A
-
C7SM-1.01 36/70
Service Manual of Cardio 2000
3) Front
A : Printer Cover Switch
B : Top Cover Assembly
C : Bottom Cover Assembly
4) Rear
A : Protective Ground Terminal
B : Power Switch
C : AC Power Connection Port
D : USB Port (Use for equipment upgrade only)
E : RJ45 LAN Port
A
C
B
C
A B
D E
-
C7SM-1.01 37/70
Service Manual of Cardio 2000
5) Right Side
A : Patient Cable Connection Port
6) Left Side
A: Handle
A
A
-
C7SM-1.01 38/70
Service Manual of Cardio 2000
4-3. Size and Weight
Size: 290(W) X 300(H) X 89.5(D) mm
Weight: approx. 3 Kg
-
C7SM-1.01 39/70
Service Manual of Cardio 2000
Section 5. Disassembly and Assembly
5-1. Module Description
Base Board
CPU Board
Power Board
(SMPS)
UI Board
-
C7SM-1.01 40/70
Service Manual of Cardio 2000
1) UI Board
A : Main Cable (Input Voltage)
B : Membrane Keys Connector
C : LCD Connector (LCD)
D : FFC Cable (40 pin)
It is linked to the Base Board with a Cable and consists of Membrane keys, Power Indicator LED, Battery
Indicator LED, Ready LED.
- Membrane Keys
- Power Indicator LED : Show the state of connection with AC adapter and a green light indicator
when plugging AC adapter into device.
- Battery Indicator LED : A light indicator lets you know that there is battery and show your current
battery charge strength. A red light indicator lets you know that the battery is charging and a
green light indicator lets you know the batter is fully charged.
- Ready LED : It Indicate the lead connection status.
A
B
C
D
-
C7SM-1.01 41/70
Service Manual of Cardio 2000
2) Base (Power + Analog) Board
A : USB Port
B : LAN Port
C : Switch Connector
D : Input DC Connector (SMPS Output)
E : Voltage Regulator (3 ea)
F : LCD Data Connector
G : Motor Driver
H : Main Connector
I : Transformer
J : Isolation IC (2 ea)
K : Patient Cable Connector
L : MCU (3 ea)
The digital part consists of MCU, DC/DC Converter, DC/AC Controller and chip, which are linked
to other external modules.
A
B
C D E F
G
H
J I K
L
-
C7SM-1.01 42/70
Service Manual of Cardio 2000
3) CPU Board
A : MCU
B : DDR3 SDRAM
C : Power Management
D : eMMC Flash
-
C7SM-1.01 43/70
Service Manual of Cardio 2000
4) Power Module (MDS-065APS15-BA)
It receives 100-240V AC, 50-60Hz signal or battery and supplies 15V to the Base board.
5) Printer
It consists of the TPH (Thermal Printer Head), motor, and paper sensor. And it is linked to the Base
board with 3 pin, 6 pin, and 14 pin cable and approves 24V AC power which will be approved only
if the printer assembly is working.
-
C7SM-1.01 44/70
Service Manual of Cardio 2000
6) Battery
It is linked to the Digital board with a 3 pin cable. It is 10.8V, 3350mA/h and consists of three 4.2V,
3350mA/h Li-ion Rechargeable battery. It is packed in parallel with 2 sets which are connected with
3 cells in series and 1 spacer.
The cables of battery consist of 3 pins connector which are VCC, GND, GND (Battery Check) as
follow picture. GND (Battery Check) cable should be connected to the board, because the cable
is for checking the status of battery.
GND
GND (Battery Check) VCC
-
C7SM-1.01 45/70
Service Manual of Cardio 2000
7) LCD
Character LCD : It connects to CDUI Board LCD with 14 pin Pin Header. It consists of LCD.
-
C7SM-1.01 46/70
Service Manual of Cardio 2000
5-2. Internal Block Diagram & Cable
Power
Managing
MCU
Lithium Ion
Battery
Power
Supply
Unit
System
5V Regulator
System
3.3V Regulator
PMIC
USB
Battery
Green
LED
Battery
Orange
LED
AC
Green
LED
Power
Button
24V
Printer Power
PM
3.3V LDO
Printer
MCU
VSYS(15V, 9V~12.6V)
VLINE(15V)
VBAT(9V~12.6V)
SYS_5V
SYS_3.3V
System PowerControl
PowerState
Signals
BatteryLevel
BatteryTemp.
User interface
PrinterPowerControl
VPM_3.3V
Battery
Charger
LCD
Power
CPU
VCPU
Analog Isolator
Charger State/Control
BatteryConnection
System
Power
Switch
VSYS_SW(15V, 9V~12.6V)
-
C7SM-1.01 47/70
Service Manual of Cardio 2000
2) Cable
LCD Assembly
40Pin FFC Cable
Battery Cable Assembly
-
C7SM-1.01 48/70
Service Manual of Cardio 2000
QN-TPH-Cable
Motor Assembly
Power Switch Cable Assembly
-
C7SM-1.01 49/70
Service Manual of Cardio 2000
5-3. Disassembly
Step1. Removing the Top cover
a. Removing the 5 screws from the bottom cover.
-
C7SM-1.01 50/70
Service Manual of Cardio 2000
b. Removing the Top cover.
1) Open the printer cover by pushing slide knob.
2) Disconnect 40 pin FFC cable and Main Cable.
3) Remove Top cover as shown below.
• Pull 40 pin FFC Cable from the connector of Base board.
-
C7SM-1.01 51/70
Service Manual of Cardio 2000
-
C7SM-1.01 52/70
Service Manual of Cardio 2000
Step2. Replacing the Printer Assembly
a. Remove the 4 screws from the bottom cover.
b. Disconnect the 3 printer cables from the Base Board.
-
C7SM-1.01 53/70
Service Manual of Cardio 2000
Step3. Replacing the Base Board
a. Remove the 9 screws from the bottom cover.
-
C7SM-1.01 54/70
Service Manual of Cardio 2000
b. Disconnect the power connector and the battery cable.
c. Disconnect the power switch cable and ground cable.
-
C7SM-1.01 55/70
Service Manual of Cardio 2000
Step5. Replacing the Power Module (SMPS)
a. Disassemble Ground cable of Power Board (SMPS) and AC Inlet Ground cable from
Protection Grounding Terminal.
b. Remove the 4 screws from SMPS Bracket.
-
C7SM-1.01 56/70
Service Manual of Cardio 2000
Step6. Replacing the Key Board
a. Remove the 4 screws from Inside of Top cover
-
C7SM-1.01 57/70
Service Manual of Cardio 2000
b. Remove the 4 screws and disconnect the 40Pin SMD FFC cable, Main cable, Membrane
Keys cable.
Step7. Replacing the LCD
a. Remove the 4 screws from LCD bracket
-
C7SM-1.01 58/70
Service Manual of Cardio 2000
Step8. Replacing the Battery
a. Remove the 2 screws from Bottom cover
b. Take out Battery from Bottom cover
c. Remove the Battery cable connector
-
C7SM-1.01 59/70
Service Manual of Cardio 2000
5-4. Assembly
Assemble the reverse way from Step.8 to Step.1
* Note
Please check whether two parts which is circled in red (rubber handle & printer lock) are
connected correctly.
-
C7SM-1.01 60/70
Service Manual of Cardio 2000
Section 6. Replaceable Parts List and Accessory
6-1. Replaceable Part List
< Top Cover Ass’y>
-
C7SM-1.01 61/70
Service Manual of Cardio 2000
-
C7SM-1.01 62/70
Service Manual of Cardio 2000
-
C7SM-1.01 63/70
Service Manual of Cardio 2000
Section 7. Connector Pin Assignment
7-1. Patient Cable Port
Pin Description
Pin Name
1 V2
2 V3
3 V4
4 V5
5 V6
6 GND
7 N.C
8 N.C
9 RA
10 LA
11 LL
12 V1
13 N.C
14 RL
15 N.C
-
C7SM-1.01 64/70
Service Manual of Cardio 2000
Section 8. PCB Part Disposition
8-1. Base (Power+Analog) Board
-
C7SM-1.01 65/70
Service Manual of Cardio 2000
8-2. CDUI Board
-
C7SM-1.01 66/70
Service Manual of Cardio 2000
8-3. CPU Board
< Bottom>
-
C7SM-1.01 67/70
Service Manual of Cardio 2000
Section 9. Software Upgrade
9-1. Requirement of Upgrading
[ AdvancePreparation ]
1) USB memory (*Note 1)
2) Unzip the bionet.zip file and copy to USB memory the bionet.zip file.
3) Copy to bionet_ekg2000\packages folder the bup file.
* Note 1: File system of USB memory must be the FAT32 or FAT .
-
C7SM-1.01 68/70
Service Manual of Cardio 2000
9-2 Procedure of Upgrading
(Cardio7 UI screen, upgrade method is same)
1) Connect to the usb port that upgrade software files are copied to a usb memory.
Setup Path
- key → SYS → UPGRADE
2-1) Enter the Password
2-2) Run Upgrade Program that check the password.(*Note 1)
* Note 1 :PASSWORD: Products that display the current date (MMDD) .
- key → SYS → DATE
- ex) Basic → Date : Aug, 25 → Password : 0825
-
C7SM-1.01 69/70
Service Manual of Cardio 2000
3-1) Select the upgrade file.
3-2) Press key.
4-1) Select the APP(version) → Press key.
4-2) Run to upgrade.
5) When the upgrade is complete, turn off and on.
-
C7SM-1.01 70/70
Service Manual of Cardio 2000
International Sales & service
Bionet Co., Ltd. :
5F, Shinsegae I&C Digital Center 61 Digital-ro 31 gil,
Guro-gu, SEOUL 08375, REPUBLIC OF KOREA
Tel : +82-2-6300-6410 / Fax : +82-2-6499-7789 / e-mail: sales@ebionet.com
Website: www.ebionet.com
U.S.A sales & service representative
Bionet America, Inc. :
2691, Dow Ave, Suite B
Tustin, CA92780U.S.A.
Toll Free: 1-877-924-6638 / Fax: 1-714-734-1761 / e-mail: support@bionetus.com
Website: www.bionetUS.com
European sales & service representative
MGB Endoskopische Gerä te GmbH Berlin :
Schwarzschildstraße 6
D-12489 Berlin, Germany
Tel: +49(0)-30-6392-7000 / Fax: +49(0)-30-6392-7011 / e-mail: sales@mgb-berlin.de
Website: www.mgb-berlin.de
Bionet Co., Ltd
Model Name: EKG2000
http://www.ebionet.com/http://www.bionetus.com/http://www.mgb-berlin.de/