cellatemp pa 40 af 9x
Post on 26-Oct-2014
16 Views
Preview:
DESCRIPTION
TRANSCRIPT
c
CellaTemp PA 40 AF 9x Multifunctional Pyrometer for temperature detection of filaments
Features
§ Dual wavelength technique
§ Temperature range: 700 to 2400 °C
§ High-resolution optics
§ Sighting options: through-the-lens sighting or HD (high dynamic) video camera
§ 2 current output signals and 2 gate inputs/outputs, freely configurable
§ Multifunctional configuration
§ USB digital interface and RS 485 network interface
KELLER HCW GmbH · Division MSR · Carl-Keller-Straße 2-10 · 49479 Ibbenbüren · GermanyTel. +49 (0) 5451 850 · Fax +49 (0) 5451 897392 · www.keller-msr.de · info@keller-msr.de
CellaTemp PA 40 AF 9x was specially designed to measure the temperature of filaments. With its high-resolution optics, this pyrometer can capture target spots as small as 4 mm at a distance of 162 mm.
CellaTemp PA 40 AF 9x detects infra-red energy emitted from the target at two different wavelengths and calcu-lates a temperature reading based on the ratio of these two values. This two-colour or dual wavelength technique produces highly reliable temperature data, even when sight path obstruc-tions such dust or steam or a dirty lens would normally weaken the signal. A two-colour or ratio pyrometer is also far less sensitive to fluctuating emis-sivity than a standard spectral pyrome-ter.
To view the target object, CellaTemp PA 40 AF 9x comes equipped with either through-the-lens sighting or a built-in colour video camera with HD (high dynamic) imaging. The video camera features TBC functionality (target brightness control) which en-ables its light sensitivity to adapt dy-namically to the brightness of the actual target object.
CellaTemp PA AF 9x provides highly versatile configuration options. The pyrometer features two analogue out-puts which the user can custom con-figure. Alternatively, one of the signal outputs can be configured as an input. Two switching outputs also provide a variety of combination possibilities and can be used as gate inputs.
In addition to emissivity, analogue current range and signal smoothing, CellaTemp PA AF 9x offers a variety of other functions and parameter set-tings.
The temperature data shown at the pyrometer’s LED display panel is easy to read from a distance of several meters. Parameters can be configured using the rear keypad. The USB inter-face enables remote configuration and the RS 485 provides network commu-nications capability.
Temperature range: 700 – 2400 °C
Sensor: Photo diode
Spectral sensitivity: 0.95 / 1.05 µm
Sighting options: § through-the-lens § video camera
2 Analogue outputs: 0(4) - 20 mA linear, switchable, scaleable configuration option: 0-10 V voltage input
Load: max. 500 Ω
2 Switching inputs / outputs Open collector 24 V, ≤ 30 mA highly versatile
Data communication: - USB - RS 485 (integrated menu-driven user Interface to set parameters and transmit data to a PC
Video output: PAL, 1Vpp, 75 Ohm Resolution: 722x576 Pixel
Response time t98: ≤ 10 ms (T > 750 °C)
Resolution Current output: 0.2 K + 0.03 % of selected range Temperature reading: 1 K
USB / RS 485: 0.1 K in terminal mode
Measurement uncertainty: 1 % of temp.reading, at least 4 K (at ε = 1 and Ta = 23 °C)
Repeatability: 2 K
Temperature coefficient: ≤ 0.05 %/K of temp. reading [°C] (deviation to 23 °C)
LED Display:
4-digit (digit height 8 mm)
Power requirement: 24 V DC +10% / -20% Current input ≤ 135 mA Ripple ≤ 200 mV
Permissible humidity: 95 % RH max. (non-condensing)
Ambient temperature: 0 – 65 °C (without cooling)
Ambient temperature: -20 – +80 °C
Dimensions: ∅ 65 x 220 mm (incl. connector)
Housing material: stainless steel
Protection: IP65 according to DIN 40050
§ Measuring range § Current range and voltage range
for analogue inputs and outputs § Smoothing function § Simulation of current and tem-
perature for diagnostics § Emissivity § Transmission factor § Data storage mode § Alarm contacts
§ PA AF 9x pyrometer § VK 02/A cable § VK 02/F video cable for model
with video camera § CellaMevis software
§ Air purge § Protective enclosure § Cooling jacket § Mounting bracket § Mounting bracket, adjustable § Calibration certificate according
to DIN ISO 9001
Type Measuring distance
Target size
PA 41 AF 211 225 mm Ø 5,6 mm
PA 41 AF 221 162 mm Ø 4,0 mm
PA 41 AF 222 186 mm Ø 4,6 mm
CellaTemp PA 40 AF 9x
Technical Data
Shipment includes
Subject to change without notice 2012/06