Is the key switch in the on position?

TR Detector Fault Analysis

Yes NoExit

Switch the key switch to the ‘ON’ positionDoes the ‘Power’ light come on?

TR Detector Fault Analysis

Yes No

Is the ‘Power’ light on?

TR Detector Fault Analysis

Yes No

Does the monitor power light come on? Switch the monitor switch on and off to check.

TR Detector Fault Analysis

Yes No

Is the mains socket switched on at the wall?

TR Detector Fault Analysis

Yes No

Switch on the mains socket.Is the ‘Power’ light on?

TR Detector Fault Analysis

Yes No

Is the mains switch on the rear of the unit switched on (as shown below)?

TR Detector Fault Analysis

Yes No

Switch on the mains switch at the back of the unit. Does the ‘Power’ light come on?

TR Detector Fault Analysis

Yes No

Check the 13Amp fuse in the mains plug. Is the fuse OK?

TR Detector Fault Analysis

Yes No

Check the 8 Amp fuse in the mains inlet. Is the fuse OK?

TR Detector Fault Analysis

Yes No8A

Contact Todd Research Service – Quote Fault Code 100Telephone 01245 262233

TR Detector Fault Analysis

Exit

Does the computer boot up to show the normal desktop on the monitor?

TR Detector Fault Analysis

Yes No

Switch the mains switch to the ‘ON’ positionDoes the ‘Power’ light come on?

TR Detector Fault Analysis

Yes No

Change the fuse and replace the plug into the wall socket. Is the ‘Power’ light on?

TR Detector Fault Analysis

Yes No

Does the monitor screen boot up to show the normal desktop on the monitor?

TR Detector Fault Analysis

Yes No

Contact Todd Research – Quote Fault Code 200Telephone 01245 262233

TR Detector Fault Analysis

Exit

Contact Todd Research – Quote Fault Code 300Telephone 01245 262233

TR Detector Fault Analysis

Exit

Switch on the mains switch on the rear of the unit. Is the power light on?

TR Detector Fault Analysis

Yes No

Your unit seems to be functioning correctlyCall service on 01245 262233 if this is not the case

TR Detector Fault Analysis

Exit

Replace the 13A fuse and replace the plug into the wall socket. Switch on the socket and switch on the unit. Does the ‘Power’

light now come on?

TR Detector Fault Analysis

Yes No

Replace the 8A fuse and replace the plug into the wall socket. Switch on the socket and switch on the unit. Does the ‘Power’

light now come on?

TR Detector Fault Analysis

Yes No

When the ‘Expose’ switch is pressed do both the ‘Expose’ and the ‘X-rays’ lights come on?

TR Detector Fault Analysis

Yes No

When the ‘Expose’ switch is pressed does the unit produce a clear picture of the object?

TR Detector Fault Analysis

Yes No

Does the unit beep repeatedly or lock up with a message in DOS on the monitor such as ‘The memory x at 0x00000002....’

?

TR Detector Fault Analysis

Yes No

Is the monitor working ?

TR Detector Fault Analysis

Yes No

Is the monitor power light on?

TR Detector Fault Analysis

Yes No

Switch on the monitor. Is the monitor power light on?

TR Detector Fault Analysis

Yes No

Try firmly reconnecting the monitor power lead at the monitor end and at the power outlet from the back of the equipment

(see power outlet photo below). Does the monitor power light come on?

TR Detector Fault Analysis

Yes No

Monitor power outlet

Check the monitor 2.5A fuse shown below. Is the fuse OK?

TR Detector Fault Analysis

Yes No

Monitor 2.5A fuse

Replace the 2.5A fuse and replace the plug into the wall socket. Switch on the socket and switch on the unit. Does the

‘Power’ light on the monitor now come on?

TR Detector Fault Analysis

Yes No

Monitor 2.5A fuse

Contact Todd Research Service – Quote Fault Code 1100Telephone 01245 262233

TR Detector Fault Analysis

Yes No

Contact Todd Research Service – Quote Fault Code 1200Telephone 01245 262233

TR Detector Fault Analysis

Yes No

Is the monitor showing any data?

TR Detector Fault Analysis

Yes No

Is the data or picture sharp and clear?

TR Detector Fault Analysis

Yes No

Does the unit display the message ‘Loading Please Wait’ or revert to first boot mode ?

TR Detector Fault Analysis

Yes No

Contact Todd Research Service – Quote Fault Code 400Telephone 01245 262233

TR Detector Fault Analysis

Exit

Contact Todd Research Service – Quote Fault Code 500Telephone 01245 262233

TR Detector Fault Analysis

Exit

Contact Todd Research Service – Quote Fault Code 550Telephone 01245 262233

TR Detector Fault Analysis

Exit

Contact Todd Research Service – Quote Fault Code 600Telephone 01245 262233

TR Detector Fault Analysis

Exit

Contact Todd Research Service – Quote Fault Code 700Telephone 01245 262233

TR Detector Fault Analysis

Exit

Contact Todd Research Service – Quote Fault Code 800Telephone 01245 262233

TR Detector Fault Analysis

Exit

Contact Todd Research Service – Quote Fault Code 900Telephone 01245 262233

TR Detector Fault Analysis

Exit

Contact Todd Research Service – Quote Fault Code 1000Telephone 01245 262233

TR Detector Fault Analysis

Exit

Does an error message ‘Video Capture Error’ come up on the screen?

TR Detector Fault Analysis

Yes No

Does a message ‘Cabinet Door Open’ come up on the screen?

TR Detector Fault Analysis

Yes No

After checking that the cabinet door is properly closed does the message ‘Cabinet Door Open’ still come up on the screen?

TR Detector Fault Analysis

Yes No

Does the computer produce a totally black image?

TR Detector Fault Analysis

Yes No

Is the image blurred or out of focus?

TR Detector Fault Analysis

Yes No

Does the x-ray tube come on when the ‘Expose’ button is pressed and held on for more than three seconds?

TR Detector Fault Analysis

Yes No