ground fault/leakage detector 2620 - aemc · 10 ground fault/leakage detector model 2620 chapter 4...
TRANSCRIPT
GROUND FAULT/LEAKAGE DETECTOR 2620
with the Artifical Neutral AN-1
CHAUVIN ARNOUX GROUP
®
E N G L I S H User Manual
Statement of Compliance
Chauvin Arnoux®, Inc. d.b.a. AEMC® Instruments certifies that this instrument has been calibrated using standards and instruments traceable to international standards.
We guarantee that at the time of shipping your instrument has met its published specifications.
An NIST traceable certificate may be requested at the time of purchase, or obtained by returning the instrument to our repair and calibration facility, for a nominal charge.
The recommended calibration interval for this instrument is 12 months and begins on the date of receipt by the customer. For recalibration, please use our calibration services. Refer to our repair and calibration section at www.aemc.com.
Serial #: ________________________________
Catalog #: 2125.52
Model #: 2620
Please fill in the appropriate date as indicated:
Date Received: _________________________________
Date Calibration Due: _______________________
Chauvin Arnoux®, Inc.d.b.a AEMC® Instrumentswww.aemc.com
Table of Contents
1. INTRODUCTION ............................................................................... 31.1 International Electrical Symbols ................................................31.2 DefinitionofMeasurementCategories .....................................41.3 ReceivingYourShipment ..........................................................41.4 OrderingInformation .................................................................4
1.4.1 AccessoriesandReplacementParts ............................4
2. PRODUCT FEATURES ...................................................................... 52.1 Description ................................................................................52.2 Connections ..............................................................................6
3. SPECIFICATIONS............................................................................. 73.1 ReferenceConditions ...............................................................73.2 Electrical ...................................................................................73.3 Mechanical ................................................................................83.4 Environmental ...........................................................................93.5 Safety ........................................................................................9
4. OPERATION .................................................................................. 104.1 PrincipleofOperation .............................................................10
4.1.1 MeasuringCurrentLeakage .......................................104.1.2 InterpretingtheMeasurement .....................................10
4.2 EquipmentTroubleshooting ....................................................114.3 TestingVariousCableConfigurations .....................................12
4.3.1 MetalConduits ............................................................124.3.2 GroundedNeutral .......................................................144.3.3 NeutralGroundedThroughanImpedance .................154.3.4 UngroundedSystem ...................................................15
4.4 CommonlyAskedQuestions ...................................................15
5. ARTIFICIAL NEUTRAL .................................................................... 17Warning ...........................................................................................175.1 Description ..............................................................................17
5.2 Specifications ..........................................................................195.2.1 Electrical .....................................................................195.2.2 Mechanical ..................................................................195.2.3 Safety ..........................................................................19
5.3 PrincipleofOperation .............................................................205.4 TestingUngroundedSystems .................................................21
6. MAINTENANCE ............................................................................. 226.1 Cleaning ..................................................................................22
RepairandCalibration ...........................................................................23
TechnicalandSalesAssistance ............................................................23
Limited Warranty ...................................................................................24
WarrantyRepairs ...................................................................................24
GroundFault/LeakageDetectorModel2620 3
CHAPTER 1
INTRODUCTION
Warning Thesesafetywarningsareprovidedtoensurethesafetyofperson-nelandproperoperationoftheinstrument.
•Neverusethisclamponconductorswithavoltagepotentialabove600Vrms.
•Neverclamparoundaconductorunlesstheclampisterminatedtoameasuringinstrumentwithproperinputimpedance.
•Keep the jawmatingsurfacesclean. Ifnecessary,useaslightlyoiledclothtoremoveoxidation.
•Ensurethatthecablesorbusbarareproperlycenteredwithintheclampjaws.
•Stayawayfromotherconductorswhichmaycreateinterference.
•Avoidleavingtheclampindampplacesorexposingittorunningwater.
•Duetoshockorfirehazards,electricalconnectionsoftheinstrumentshouldbeperformedonlybyqualifiedpersonnelandinaccordancewithlocal,state,andfederalelectricalrequirements.
1.1 International Electrical Symbols
Thissymbolsignifiesthattheinstrumentisprotectedbydoubleorreinforced insulation.Useonlyspecified replacementpartswhenservicingtheinstrument.
This symbol on the instrument indicates a WARNING and thattheoperatormustrefertotheusermanualforinstructionsbeforeoperating the instrument. In this manual, the symbol precedinginstructionsindicatesthatiftheinstructionsarenotfollowed,bodilyinjury,installation/sampleandproductdamagemayresult.
Riskof electric shock.The voltageat thepartsmarkedwith thissymbolmaybedangerous.
4 GroundFault/LeakageDetectorModel2620
1.2 Definition of Measurement CategoriesCAT II: Formeasurements performed on circuits directly connected to
theelectricaldistributionsystem.Examplesaremeasurementsonhouseholdappliancesorportabletools.
CAT III: For measurements performed in the building installation atthe distribution level such as on hardwired equipment in fixedinstallationandcircuitbreakers.
CAT IV: For measurements performed at the primary electrical supply(<1000V) such as on primary overcurrent protection devices,ripplecontrolunits,ormeters.
1.3 Receiving Your ShipmentUponreceivingyourshipment,makesurethatthecontentsareconsistentwiththepackinglist.Notifyyourdistributorofanymissingitems.Iftheequip-mentappearstobedamaged,fileaclaimimmediatelywiththecarrierandnotifyyourdistributoratonce,givingadetaileddescriptionofanydamage.Savethedamagedpackingcontainertosubstantiateyourclaim.
1.4 Ordering InformationGround Fault/Leakage Detector Model 2620...................Cat. #2125.52Includes user manual and warranty card.
1.4.1 Accessories and Replacement PartsAdapter–4mmNon-insulatedforSafetyLeads................. Cat. #1017.45(Converts male safety plugs to non-shielded male plugs)
ArtificialNeutralModelAN-1................................................ Cat. #1971.01Includes instrument with soft carrying case; eight 1.5 V “AA” batteries; four 5 ft (1.5 m) color -coded leads (green, red, black, blue) and user manual.
Bag–LargeClassicToolBag .............................................Cat. #2133.71
Bag–SmallClassicToolBag .............................................Cat. #2133.72Lead–Replacement ...........................................................Cat. #2950.29Set of 4 Color-coded 5 ft Safety Leads (4mm straight banana plugs on both ends) for use with the Artificial Neutral Model AN-1. Rated 600V CAT IV
GroundFault/LeakageDetectorModel2620 5
CHAPTER 2
PRODUCT FEATURES
2.1 DescriptionTheModel2620measuresleakagecurrentshuntedtogroundcausedbyinsulationbreakdownor leaks. Itenables theoperator to locate failureswhenorbeforetheyoccur,withoutshuttingdownequipmentorspendinghourstrouble-shooting.It isdesignedspecifically for locating lowcurrent faultsonhighcurrentloads.ThedetectorisasensitiveACcurrenttransformercapableofmea-suring differential or leakage current from500µA, andmay be used tomeasurecurrentupto400Acontinuousduty.TheModel2620provides,ontworanges,1mV/mAACor1mV/AAC.Theoutputleadsareterminatedwithstandard4mmbananaplugscapableofinterfacingwithanystandardmultimeter.Useofadigitalmultimeterwithanalogbargraphisrecommended:Digitaltoprovidethepropervoltageinputimpedance,andananalogbargraphtotracktrends.Whenthe2620isusedasacurrentleakagedetector,itmakesnodifferenceifthesystemissingle-phaseorpoly-phase,orifthecurrentsarein-phaseoroutofphase,balancedorunbalanced.Thenetmagneticfieldatanyinstantintimewillbezeroifalltheconductorssurroundedbythecurrentleakagedetectoraresupplyingallthecurrentdeliveredtoandreceivedfromtheload.Ifanycurrent isdiverted throughanyalternatepath toground,suchasaninsulationbreakdown,thenetlosswillbedetectedproducingaoutputproportionaltotheamplitudeofthefaultcurrent.TheModel2620mayalsobeusedsimplyasahighlyaccurateclamp-oncurrentprobe.Withits4"jawopeningandrangeof500µAto400A,the2620providesaversatileway toanalyzeunbalancedcurrentmeasurements,leakagevaluesongroundingconductorsandgroundloopcurrents.
6 GroundFault/LeakageDetectorModel2620
2.2 ConnectionsThe Model 2620 is compatible with most DMMs, voltmeters, or othervoltagemeasuringinstrumentswiththefollowingfeatures:
• Rangeandresolutioncapableofdisplaying1mVACofinput.• Voltmeteraccuracyof0.75%orbettertotakefulladvantageofthe
accuracyoftheprobe.• Minimuminputimpedanceof1MΩ.
Connecttheprobetothemultimeterorotherinstrument.
SelecttheappropriateACvoltagerangeonyourmultimeter.
TheModel2620hastwoselectableoutputranges.The4Arangewillpro-duceamV/mAwithanoutputof4VACat4A.The400Arangeproduces1mV/Awith400mVAC at 400A.
Warning: User SafetyAlwaysuseaDMM, voltmeter or other displayingdevice, appropriatelyratedfortheworkingvoltageandforsafety.
GroundFault/LeakageDetectorModel2620 7
CHAPTER 3
SPECIFICATIONS
3.1 Reference Conditions
Ambient temperature: 23°C ± 3KRelative humidity: 20to75%RHPosition of the conductor: CenteredinthejawsCurrent frequency and form: Sinusoidal50/60Hz±0.2Hz,distortion<1%Superimposed DC current: NoDCcurrentContinuous magnetic field: Earthfield<40A/mAlternating magnetic field: NoexternalalternatingmagneticfieldProximity of external conductors: NoneMeasuring device’s impedance: ≥10MΩ/100pF
3.2 Electrical
4A Range
Nominal Current: 4AAC
Measurement Range: 0.5mA to 4AAC
Output Signal: 1mVAC/mAAC
4A Range 0.5 to 10mA 10 to 100mA 100mA to 4AAccuracy 3%ofR±1mV 0.5%ofR±0.5mV 0.5%ofR±0.5mVPhaseShift – < 15° < 10°
400A Range
Nominal Current: 400AAC
Measurement Range: 0.5A to 400AAC
Output Signal: 1mVAC/AAC
400A Range 0.5 to 10A 10 to 100A 100 to 400AAccuracy 0.5%ofR±0.5mV 0.35%ofR±0.5mV 0.35%ofR±1mVPhaseShift – < 1° < 0.6°
8 GroundFault/LeakageDetectorModel2620
Overloads• Iplimitcurrent:permanent400ACRMS• Peakcurrent:<1000A• Permissibletransientdi/dt:≤30A/μs• Conductortemperature:≤70°Cwithamaximumpeakof90°C
Load Impedance (DMM):1MΩ min
Frequency:48to1000Hz
Errors caused by external influences:Ambient temperature <0.1%per10KPosition of the gripped conductor(maxwithnotcentredconductor)
0.1%typicofVOUT(non-differentialcurrent);0.2%max
Residual differential(maxwithnotcentredconductor)
0.1%typicofIP(differentialcurrent);0.2%max
External fields, 1V/A (1) <60mVofVsExternal fields, 1mV/mA (1) <100μVofVsCoupled DC current, 1V/A (2) <1mVforcontinuous1ACoupled DC current, 1mVm/A (2) <0.1mVforcontinuous1AFrequency, 1V/A (3) <1.5%from30Hzto1KHzFrequency, 1mV/mA (3) <0.5%from30Hzto1KHz
(1): 400 A/m 50Hz field perpendicular to the clamp opening(2): DC current coupled onto an AC current(3): Limited to 1KHz for 100A
3.3 MechanicalDimensions: 11.2x6.9x1.8"(285x175x45mm)
Weight:2.9lbs(1300g)approx
Jaw opening: 4.4"(112mm)
Maximum Jaw Gap:9.8"(250mm)
Maximum Clamping Capacity: Max4.5"(115mm)Øcable
GroundFault/LeakageDetectorModel2620 9
3.4 EnvironmentalOperating Temperature:-14°to131°F(-10°to55°C);0to85%RH(non-condensing)
Storage Temperature:-40°to158°F(-40°to70°C);0to85%RH(non-condensing)
Operating Altitude: ≤2000monuninsulatedconductors
Transport Altitude: ≤12000m
3.5 SafetyElectrical Safety
EN61010,600VCATIIIPollutionDegree:2
Electromagnetc CompatibilityIndustrialEnvironment:CriterionBEmissivity(EN61326-1)Susceptibility(EN61326-1)
Self-extinguishing CapabilityJawsandCasing:VO(UL94)
10 GroundFault/LeakageDetectorModel2620
CHAPTER 4
OPERATION
4.1 Principle of Operation
4.1.1 Measuring Current Leakage
WhenageneratorsuppliesanACloadthroughapairofwiresinaninsulatedcable,thecurrentgoingoutononewireisequaltothecurrentreturning;their vector sumequalszero.Aground fault changes thisequality,andtheleakagecurrentdetectorpicksitup,measuresit,andprovidesanACvoltageoutputproportionaltotheseverityoffaultcausingtheunbalance.Thegroundconductormustnotpassthroughthedetector.
4.1.2 Interpreting the MeasurementFigure1isaschematicofasingle-phaseinstallationofamotorandanoven,bothgrounded.Thenumbersindicatetheamplitudesofthecurrentsin different conductors.This installation has no faults. If you clamp thedetectoraroundthecablesatpointsA,B,CorD,itwilloutputzero.
115V/230V 15A
15A
10A
10AA B
C D
5A 5A 10A 10A
10A5A
GND
MOTOR
ZEROAMPERE
ZEROAMPERE
OVEN
GND
Figure 1
GroundFault/LeakageDetectorModel2620 11
InFigure2,thereisafaultintheinstallation.AtpointA,theModel2620detects 1A (the difference between 16A and 15A), indicating a faultdownstream.AtpointC, itdetectszero; therefore, thefault isnot in theoven.AtpointsBandD,thedetectorpicksup1A,showingthatthefaultisinthemotor.Ifitgetsworse,theleakagecurrentwillincreaseandthecircuitbreakerwill trip.Thismotorshouldbedisconnected immediatelyandrepaired.
115V/230V 15A
15A
10A
11AA B
C D
5A 5A 10A 11A
10A5A
GND
MOTOR
1A
1A
OVEN
GND
1A
RESIDUAL CURRENT: 1A
Figure 2
In a three-phase installation the principle is the same. But here it isnecessarytopassallactiveconductorsthroughthedetector,includingtheneutralifoneisused.Ifthisisdoneproperly,thedetectoroutputwillbezeroifthereisnocurrentleakingtoground.Ifitdetectssomethingotherthanzero,thereisagroundfaultthatshouldberepaired.
4.2 Equipment TroubleshootingTolocateafaultquickly,beginyourtroubleshootingattheincominglow-voltagedistributionheader.Successivelycheckeachfeeder.IftheModel2620 detects a ground fault, follow the feeder downstream, branch bybranch,totheequipmentwiththeproblem.
It isuseful to takemeasurementsperiodically tocreateahistoryof thequalityoftheinsulation.Thisisgoodpreventivemaintenancewhichcouldprovevaluable later. Ifyoudetectafaultcurrentthat isvery lowinitiallybut increasesfromdaytoday,thismeansthatelectrical failure is immi-nentandshouldbetakencareofduring thenextplannedmaintenanceshutdown.
12 GroundFault/LeakageDetectorModel2620
Forbestresults,centertheconductorswithintheclampjaws.Checkthattheedgesofthejawsareperfectlycleanandthattheyclosetightly.Trytoavoidplacingtheconductorsnearthegaps,andgroupthemasmuchaspossible.
Ifyouarecareless,youcanaffectthemagnitudeofthereading,depend-ingon thecurrentbeingcarried in theconductors. Incertaincases thisrequiresan interpretation.Normally,however, theabsolutevalueof thegroundcurrentisnotimportant.Allyouneedtoknowiswhethertheread-ingiszero.
4.3 Testing Various Cable ConfigurationsOlder installationshaveawidevarietyofelectricaldistributionsystems.Cablesareclampedtowallsandpassthroughmetalconduits,andtherearemanycaseswhereashieldedcablewasinstalledduringrevampingofanareaoradditiontoabuilding.Ifacableisclampedtowalls,detachitandpullitawayfromthesurfaceatseveralpoints,sotheclampcansur-roundalltheconductors.
4.3.1 Metal Conduits
Cablesenclosedinmetalconduitswhicharegroundedarenotaproblem.Simplyapplytheclamparoundtheconduit;themetaltubewillnotblockthemeasurement.Cableswithametalshieldcanbehandledinthesamewayascableinmetalconduit.
Agroundedshieldmaypresentaproblem.Ifitisasimplemetalshield,likeathinmetalwrapping,itcanbeconsideredinthesamewayasthemetalconduit. If thearmor isgroundedbyaconductor, itmaybeaninconve-nience,dependingontheterminalconnections,aswellasonthenumberandtypeofgroundconnectionsalongthecable.
InFigure3 thedetectorsensesthefaultbecausetheshield isnotcon-nected to the motor frame and the leakage current cannot flow backthroughthedetector.
InFigure4,however, the instrumentcannot“see” thefaultbecausetheshield is connected to the motor frame, letting the fault current returnthrough thedetector and causing it to read zero. If possible, take yourmeasurementwhereyoucanavoidclosingtheshieldintheclamp,suchasthenearestcontrolpanelorinsidetheconnectingboxofthemotor.
GroundFault/LeakageDetectorModel2620 13
Figure 3 Figure 4
OnewaytocircumventthisshieldedcableproblemisshowninFigures5and6.Heretheshieldisloopedthroughthedetectortwice,sothatthereadingwillbethatofthefaultcurrent.
Figure 5 Figure 6
14 GroundFault/LeakageDetectorModel2620
4.3.2 Grounded NeutralIt is easy to takemeasurements on this type of network, wheremajorfaults trip thecircuitbreaker.Here, thedetectorcansenseminor faults,sorepairscanbemadebefore theproblemgetsworse.The inspectionmethodisthesameasforasingle-phaseinstallation.Rememberthatonathree-phasenetwork,thecurrentsaddaccordingtothevectorsumlaw.TheModel2620measuresthevectorsumoftheenclosedcurrents.Ifyouhave,forexample,a400mAfaultanda250mAfaultonthesamephase,theModel2620willdetect650mA.But if the two faultsareondifferentphases,theyaddvectoriallyandmayevensumuptozero.Figure7showsanexampleinwhichtherearetwoseparatefaults:400mAonamachine inoneshop,250mAonamachine inanothershop.Thevectorsumisnot650mA,but350mA.Ifyouswitchoffthemachinewiththe250mAfault,thereadingincreases,indicatingfaultsondifferentphases.
Shop 2
Shop 1Fault on Phase 1
400mA
Fault on Phase 2250mA
Figure 7
GroundFault/LeakageDetectorModel2620 15
4.3.3 Neutral Grounded Through an ImpedanceFollowtheprocedurethesamewayasforagroundedneutralinstallation,butwithcare.Whentheneutralisgroundedthroughanimpedance,afault(evenaseriousone)maynottripthecircuitbreaker,andthecircuitwillstay live.
4.3.4 Ungrounded SystemIn order to measure on an ungrounded system, the optional ArtificialNeutralModelAN-1mustbeusedinconjunctionwiththeModel2620.For measurement procedures, see the information in this manual inChapter5.
4.4 Commonly Asked Questions
Q. I have found an indication of 40 to 50mA on one of the main feed-ers with 300A on the line. Is this value a fault?
A. With300Aontheline,itispossibletofindanundesiredcurrentofthatmagnitude.Different cables in the installationmay not have equivalentcapacitanceand resistance.Thesecablesprovide forhighcurrentand,ofcourse,weapplyahighcurrent.Therefore,wemayhaveacapacitiveinfluencewhichcorrespondsto45mA.
Q. Must the conductors be centered within the jaws?
A. Youshouldavoidplacingtheconductorscloseagainstthemagneticcircuit near the gaps, and should group the conductors as much aspossible. Best results are obtained while centering the conductor. Youmustalsoensurethatthejawmatingsurfaceareperfectlycleanandthatnothingpreventsthemfromclosingtightly.
Q. Is it possible when checking cables which run parallel, yet supply the same load, to clamp the detector over each cable separately?
A. Whenthereareparallelcables, themeasurement isvalidonly if thedetectorincludesthecablestogether.Infact,ifyoutakeeachonesepa-rately, it ishighlyprobable that thedistributionofcurrent in theparallelcablesisnoteven.Thesedifferencescanbecausedbythefactthatthecablesmaynothaveexactlythesameimpedance.
16 GroundFault/LeakageDetectorModel2620
Q. I have detected faults of 400mA and 250mA on branch circuits being supplied by the feeder. But clamping onto the incoming feeder resulted in an indication of 350mA. Can you explain why 400mA + 250mA = 350mA?
A. Thedetectormeasuresresidualcurrent;inotherwords,thevectorialsumofthecurrents.Ifyouhavea400mAfaultanda250mAfaultonthesamephase,youhaveeverychanceoffinding650mA.But,ifyouhavetwofaultsondifferentphases,theyaddasvectorsactually,inaverytheo-reticalcasewecouldevenfindzero.
Q. The Model 2620 detects faults caused by insulation failures. How does detecting faults with the detector differ from information result-ing from megohmmeter insulation tests?
A. TheModel2620willdetectelectricalfaultswhiletheequipmentisstillinoperation.Thisofferstwodistinctadvantages.First,itisnotnecessarytoshut theequipmentdown,eliminatingcostlydown time;andsecond,you may detect faults that would not be evident under non-energizedconditions.
Q. Would it be possible to clamp the Model 2620 directly around the grounding detector?
A. Yes,insomeinstancesitmaybepossibletodetectfaultcurrentonaspecificmotorifthereisadedicatedgroundingpath.However,ifseveraldevicesaresharingacommonground, itshouldbenotedthat thefaultmaybeoccurringonanyone(ormultiple)ofthedevices.
GroundFault/LeakageDetectorModel2620 17
CHAPTER 5
ARTIFICIAL NEUTRAL
Warning • Themeasurementofleakagecurrentisdonespecificallywithaclamp
designedforthispurpose.• TheArtificialNeutralmustbeconnected toaknown low resistance
ground.• Iffaultfindingexceeds15minutes,removetheArtificialNeutralfrom
itscarryingcase.• TheArtificialNeutralmustalwaysbeintheOFFpositionwhencon-
nectingtheleads.• Always make connections from the instrument to the circuit under
test.• Alwaysinspecttheinstrumentandaccessoryleadsforserviceability
priortouse.Replacedefectivepartsimmediately.• Neverusetheinstrumentonelectricalconductorsratedabove600V.
5.1 DescriptionTheArtificialNeutral isanaccessoryfortheAEMCModel2620GroundFault/LeakageDetector.Itgeneratesagroundedartificialneutraltoallowthemeasurementoffaultcurrentsonungrounded3phasesystems.Inanungroundedsystem,theimpedancebetweenthetransformer,neu-tral,andgroundisinfinity.Therefore,independentoftheloadcurrent,nogroundfaultcurrentshouldflow.TheArtificialNeutralcreatesanintermittentreturn path for the ground fault current enabling the troubleshooter todistinguishitfromloadcurrent.TheArtificialNeutralisgroundedintermittentlythrougharelaydrivenbyaninternalclock.Thisperiodicgroundingtechniqueperformsthreefunc-tions:
• Limitsthepossibilityofinsulationfaultswithinthecase.• Allowsbetterdiscriminationofactualfaultcurrent.• Limitsinstrumentoverheating
18 GroundFault/LeakageDetectorModel2620
Relayclosuretimeisselectedexternally(FastorSlow)sothatthegroundfaultclampcanbeusedoneitherananalogordigitaldisplay.Whenusingondigitalmultimeter,placetheswitchintheslowposition;whenusingananalogmultimeter,settheswitchtothefastposition.An internal buzzer, also driven by the relay, is activated each time theneutralconnection isgrounded.Three lights,oneforeachphase,showthepresenceofeachofthethreephasesintheArtificialNeutralpriortomeasurement.Afourthgreenlightindicatestheinstrumentisoperating.
WARNINGConnect and disconnectonly in the OFF position
ARTIFICIAL NEUTRALMODEL AN-1
ONSlow
OFF
ONFast
SPECIFICATIONSWorking Voltage: 100 to 600VFrequency: 45 to 65HzBatteries: 8 x 1.5V AA (LR6)EN 61010, 600V, Cat. III, Pol. 2
600V max
ON
ON light(green)
Male-male safety leads,4 mm Ø, length 5 ft (1.5 m) double insulation
CarryingStrap
Ventilation holes3 position switch: • ON, Slow test • OFF • ON, Fast test
[Carrying case not shown]
Safety alligator clips
CV3
AV1
BV2
Red light, showingvoltage present and phase fault
Figure 8
GroundFault/LeakageDetectorModel2620 19
5.2 Specifications
5.2.1 Electrical
Working Voltage:30to600V
Frequency: 45to65Hz
Resistance Per Phase: 3.9kΩ±5%
Relay Duty Time:Slowposition:0.5sFastposition:2.3s
Power Source: Eight1.5V“AA”batteries
Power Consumption:180mA(approx)
Average Battery Life:40hours
5.2.2 Mechanical
Reference Temperature: 23°C ± 3°C
Operating Temperature: 32°to122°F(0°to50°C);10to90%RH
Storage Temperature:-40°to158°F(-40°to70°C);10to90%RH
Case: FireresistantpolycarbonateUL94
Color: Yellow
Dimensions: 8.7x5.4x5.9"(220x136x150mm)
Weight: 2.9lbs(1.3kg)
5.2.3 Safety
Dielectric Strength Test: 6kVbetweenterminalandcase
600VCATIII
20 GroundFault/LeakageDetectorModel2620
5.3 Principle of OperationIfthereisnotaninsulationfaultonthesystem,noleakagecurrentwillflowtoground.Thevectorsumofthecurrentsflowingthroughthethreecon-ductorsenclosedwithintheclampiszero.Thedisplaywill indicatezero(seeFigure9).
A B C N
AN1
M
Figure 9
Ifthereisafault,aleakagecurrentthenflowsbetweenthefaultyphaseandground.Thedetectionofthisfaultcurrentonanungroundedsystemwillbeintermittentasdeterminedbytherateofrelayclosure(Figure10).
Figure 10
Duetothefaultcurrent,thevectorialsumsofthecurrentwillnolongerbezero.Faultcurrentwillbe indicatedbytheselectionontheModel2620clampofeither1mVpermAor1mVperAAC.
GroundFault/LeakageDetectorModel2620 21
5.4 Testing Ungrounded SystemsBefore starting, ensure that theArtificial Neutral batteries are in goodcondition.Switchontheinstrument(FastorSlow).Thegreenlightmustbeilluminated,andthebuzzermustbeactivatedattheselectedrate.Ifthegreenlightisout,orifthebuzzersoundsweak,replacethebatteries.Switch the instrument OFF (green light off). Connect the leads to theappropriatephasesofthesystemundertest.Atthispoint,do not connect the Artificial Neutral to ground.
• PhaseA:RED• PhaseB:BLACK• PhaseC:BLUE
Switch the instrumentON (Fast or Slow).The three phase lightsmustcome on at the same time, then periodically, at the same rate as thebuzzer.Ifoneormoredonotlight,stopthetestandverifyconnectionsortheexistenceofaphasefault.Note:Theremustbeatleast30Vbetweenphases.After verifying that all the lights have illuminated properly, turn theinstrumentOFFandconnecttheGROUNDleadtoaknowngoodground.Tomeasurethevalueofthefault,connecttheModel2620toamultimeterwith an appropriate measurement range. Enclose the three phaseconductorswithinthejaws.SwitchONtheArtificialNeutral(FastorSlow)dependingontheresponsetimeofthemeasuringinstrument.Theactualfaultmeasurementmust bemadewhen the instrument relay is closed(e.g.,when thebuzzersounds).Themeasuredvaluewhen the relay isclosedistheactualvalueofthefaultcurrent,dependingontheselectedoutputsignal.To disconnect theArtificial Neutral, reverse the procedure. Switch theArtificial Neutral OFF, disconnect the GROUND lead, then disconnecteachoneofthephaseleads.
22 GroundFault/LeakageDetectorModel2620
CHAPTER 6
MAINTENANCE
WARNING:
• Formaintenanceuseonlyspecifiedfactoryreplacementparts.
• Toavoidelectricalshock,donotattempttoperformanyservicingunlessyouarequalifiedtodoso.
• Toavoidelectricalshockand/ordamagetotheinstrument,donotgetwaterorotherforeignagentsintothecase.TurntheinstrumentOFFanddisconnecttheunitfromallthecircuitsbeforeopeningthecase.
6.1 Cleaning• Wipethecaseandjawcoverswithalightlymoistenedclothand
milddetergent.
• Donotuseanyabrasivesorsolvents.
• Ifrusted,thejawmatingsurfacesoftheModel2620maybelightlysandedwithveryfinesandpaperandthenverylightlyoiled.Wipeoffexcessoilanddonotletitdripintothecase.
GroundFault/LeakageDetectorModel2620 23
Repair and Calibration
Toensurethatyourinstrumentmeetsfactoryspecifications,werecommendthatitbescheduledbacktoourfactoryServiceCenteratone-yearintervalsforrecalibration,orasrequiredbyotherstandardsorinternalprocedures.
For instrument repair and calibration:Youmust contactourServiceCenter foraCustomerServiceAuthorizationNumber(CSA#).Thiswillensurethatwhenyourinstrumentarrives,itwillbetrackedandprocessedpromptly.PleasewritetheCSA#ontheoutsideoftheshipping container. If the instrument is returned for calibration,weneed toknowifyouwantastandardcalibration,oracalibrationtraceabletoN.I.S.T.(Includescalibrationcertificateplusrecordedcalibrationdata).
Ship To: ChauvinArnoux®,Inc.d.b.a.AEMC®Instruments 15 Faraday Drive Dover,NH03820USA Phone:(800)945-2362(Ext.360) (603)749-6434(Ext.360) Fax: (603)742-2346or(603)749-6309 E-mail:[email protected]
(Orcontactyourauthorizeddistributor)Costsforrepair,standardcalibration,andcalibrationtraceabletoN.I.S.T.areavailable.NOTE: You must obtain a CSA# before returning any instrument.
Technical and Sales Assistance
Ifyouareexperiencinganytechnicalproblems,orrequireanyassistancewiththeproperoperationorapplicationofyourinstrument,pleasecall,mail,faxore-mailourtechnicalsupportteam:
ChauvinArnoux®,Inc.d.b.a.AEMC®Instruments 200FoxboroughBoulevard Foxborough,MA02035USA Phone:(800)343-1391 (508)698-2115 Fax: (508)698-2118 E-mail:[email protected] www.aemc.com
NOTE: Do not ship Instruments to our Foxborough, MA address.
24 GroundFault/LeakageDetectorModel2620
Limited Warranty
TheModel2620iswarrantedtotheownerforaperiodoftwoyearsfromthe date of original purchase against defects inmanufacture.This lim-itedwarrantyisgivenbyAEMC®Instruments,notbythedistributorfromwhomitwaspurchased.Thiswarranty isvoid if theunithasbeentam-peredwith,abusedorifthedefectisrelatedtoservicenotperformedbyAEMC® Instruments.
For full and detailed warranty coverage, please read the Warranty Coverage Information, which is attached to the Warranty Registration Card (if enclosed) or is available at www.aemc.com. Please keep the Warranty Coverage Information with your records.
What AEMC® Instruments will do:Ifamalfunctionoccurswithinthewarrantyperiod,youmayreturntheinstrumenttousforrepair,providedwehaveyourwarrantyregistrationinformationonfileoraproofofpurchase.AEMC®Instrumentswill,atitsoption,repairorreplacethefaultymaterial.
REGISTER ONLINE AT:www.aemc.com
Warranty Repairs
What you must do to return an Instrument for Warranty Repair: First, request aCustomerServiceAuthorizationNumber (CSA#) by phoneorbyfaxfromourServiceDepartment(seeaddressbelow),thenreturntheinstrumentalongwith thesignedCSAForm.Pleasewrite theCSA#on theoutsideoftheshippingcontainer.Returntheinstrument,postageorshipmentpre-paidto:
Ship To: ChauvinArnoux®,Inc.d.b.a.AEMC®Instruments 15FaradayDrive•Dover,NH03820USA Phone:(800)945-2362(Ext.360) (603)749-6434(Ext.360) Fax: (603)742-2346or(603)749-6309 E-mail:[email protected]
Caution:Toprotectyourselfagainstin-transitloss,werecommendyouinsureyourreturnedmaterial.
NOTE: You must obtain a CSA# before returning any instrument.
CHAUVIN ARNOUX GROUP
®
05/17
99-MAN100292v8
Chauvin Arnoux®, Inc. d.b.a. AEMC® Instruments15FaradayDrive•Dover,NH03820USA•Phone:(603)749-6434•Fax:(603)742-2346
www.aemc.com