6/26/2019

1

SAVe II

Easy Set Up Guide

2

Vent Overview

PEEP

Tidal Volume

Confirmation button

Alarm Silence

Respiratory Rate

Single Manual Breath

Power

Peak Pressure Limit

Height Selection

6/26/2019

2

Circuit Setup

3

4

Start-up

1. Press power

2. Select pt height

3. Press confirm

6/26/2019

3

5

Add PEEP/ Start Ventilator

1. Increase PEEP to 5

2. Press Confirm

*If different Tidal Volume is desired, refer to chart on next slide.

6

6/26/2019

4

7

Add Supplemental FiO2

1. Remove black cap(do not discard cap or white filter)2. Attach green side of large bore tubing to vent3. Fully extend large bore tubing4. Attach small bore tubing to O2 source and turn flow to 6LPM

Monitor Patient

• Maintain end tidal CO2 between 35-45

• Increase RR or VT to lower end tidal CO2

• Decrease RR or VT to raise end tidal CO2

• Maintain SpO2 above 94%

• Increase FiO2 or PEEP (do not raise above 10) to raise SpO2

8

6/26/2019

5

Recommended adjunct placement

9

1. Suction2. Emma3. HME4. MDI5. Neb