Organic Elemental Analysis (OEA) using the dynamic flash combustion technique typically measures Carbon, Nitrogen, Hydrogen, Sulfur and Oxygen (CHNS/O) at <100ppm to 100% levels. However, with the addition of a Flame Photometric Detector (FPD) the Thermo Scientific™ FLASH 2000 analyzer (Figure 1) offers the unique capability of detecting lower sulfur levels of 5–10ppm.
Thermo Scientific FLASH 2000 elemental analyzer with FPD (Flame Photometric Detector) Trace level sulfur determinations beyond the classical OEA limit
Principle of operationThe instrument operation remains the same according to the standard FLASH 2000 sulfur analyzer, with the exception of detection by optional Flame Photometric device (Fig. 2). Samples are weighed in a tin capsule and introduced into the combustion reactor via the Thermo Scientific™ MAS™ 200R autosampler, with the proper amount of pure oxygen.
After combustion – 1800 °C for few seconds – the resultant gases are carried by a helium flow to a layer filled with copper, then to a trap filled with anhydrone (magnesium perchlorate) and swept through a GC column before detection by the Flame Photometric detector (FPD). Total time for this lower level sulfur determination is just 5-6 minutes.
Advantages:• Lowsulfurdetermination(5-10ppm)insolid,
liquid and gas samples
• High(<100ppm-100%)andlow(5-5000ppm) sulfur detectable by the same instrument using TCD and FPD detector respectively
• BothTCDandFPDdetectorsavailableforuseat all times
• Handlesallsampletypes(solid,liquid,viscousand gas)
• Nosamplepre-treatmentrequired
• Inerttubingandconnectorspreventanysulfuradsorption
• DedicatedStandardandReferenceMaterials(solid and liquid) kit for calibrating the analyzer according to the users samples is included in the package included in the OEA/FPD option
• Noadditionalsoftwaretolearnorupgraderequired: the Thermo Scientific Eager Xperience OEA software is ready to use with FPD coupling
• Noneedtowaitforservicecalls–theFPDoption is user installable on any FLASH 2000 model and our previous elemental analyzers (branded Carlo Erba and Fisons)
• Lowerownercostscomparedtootheralternative techniques
Figure 2. FPD / sulfur determination layout
Figure 1. FLASH 2000 analyzer with FPD detector
E L E M E N T A L A N A L Y S I S
Sample Sulfur ppm RSD %
Diesel
25
9.03022
21
22
Gasoline28
6.81533
33
Catalyst73
3.43276
71
Cellulose12
4.68113
12
Sample Sulfur ppm RSD %
Sand
8
10.2069
8
7
Clay59
4.10361
64
Soil272
1.510269
264
Pine Needles331
2.168318
320
FPD SpecificationsEssential for a wide range of applications:
Gas requirements Helium GC grade Air99.995% Hydrogen99.995% Oxygen99.995%(forsamplecombustion)
Power requirements 120/230 V AC, 50/60 Hz
Host PC requirements Pentium processor Windows® 7 and previous operating systems Eager Xperience
Dimensions Gas control box: 177 x 235 x 287 mm (l x h x d) FPD control module: 180 x 85 x 485 mm (l x h x d)
Weight Gas control box: 5 kg FPD control module: 6 kg
FPD package includes Operating manual – includes suggestions on standards, standard weight, sample weight, separation column, FPD analytical range and more to obtain the best results FPD detector FPD control module Gas control device Standard outfit Sulfur Reference Material (solid/liquid samples): –PastaReferenceMaterial(S%:0.131) –SoilReferenceMaterial(S%:0.032) – Ref. Solutions Kit (4 flasks of 10, 50, 100 and 500 ppm of sulfur conc. solutions)
Part Number 432 10145 FPD Option for Flash 2000
Petrochemical Products Lubricants, Gasoline, Diesel, Pitches, Rock oils, Coke, Coal, Gasoil, Rubbers, Tires
Geological material Soils, Sediments, Rocks, Shale, Kerogene, Sand, Clay
Environmental Compost,Bio-mass,WasteWater
Inorganic Compounds Ceramics, Glass fibres, Graphite cathodes , Refractory materials
Organic Compounds Organo-metallic compounds, Pigments, Solvents, Urea, Formaldehyde
Materials Catalysts, Cements, Metals (steel and alloy), Polymers, Polyethylene films, Pulp, Paper and Cellulose
Vegetation Cereals, Leaves, Pine needles, Plants, Tobaccos, Algae, Lichens, Wood and Soil amendments
Food Pasta, Chocolate
Animal Feed Alfalfa
Human samples Nails,Hair,Urine,Blood
More details on FPD applications are available on the OEA CookBook (p/n 317 08256),
a CD which contains a compendium of over 5000 analyses, from more than
3000 sample matrices.
Example performance dataP
rod
uct S
pe
cificatio
ns
PS42193_E 11/12C
Africa-Other +27 11 570 1840Australia +61 3 9757 4300Austria +43 1 333 50 34 0Belgium +32 53 73 42 41Canada +1 800 530 8447China +86 10 8419 3588Denmark +45 70 23 62 60
Europe-Other +43 1 333 50 34 0Finland/Norway/Sweden +46 8 556 468 00France +33 1 60 92 48 00Germany +49 6103 408 1014India +91 22 6742 9434Italy +39 02 950 591
Japan +81 45 453 9100Latin America +1 561 688 8700Middle East +43 1 333 50 34 0Netherlands +31 76 579 55 55New Zealand +64 9 980 6700Russia/CIS +43 1 333 50 34 0South Africa +27 11 570 1840
Spain +34 914 845 965Switzerland +41 61 716 77 00UK +44 1442 233555USA +1 800 532 4752
www.thermoscientific.com©2012 Thermo Fisher Scientific Inc. All rights reserved. ISO is a trademark of the International Standards Organization. Windows is either a registered trademark or trademark of Microsoft Corporation in the United States and/or other countries. Pentium is either a registered trademark or trademark of Intel Corporation in the United States and/or other countries. All other trademarks are the property of Thermo Fisher Scientific Inc. and its subsidiaries. This information is presented as an example of the capabilities of Thermo Fisher Scientific Inc. products. It is not intended to encourage use of these products in any manners that might infringe the intellectual property rights of others. Specifications, terms and pricing are subject to change. Not all products are available in all countries. Please consult your local sales representative for details. Thermo Fisher Scientific S.p.A. Milan, Italy is ISO Certified.
top related