analytic techniques sc

Upload: ann-acuzar

Post on 14-Apr-2018

218 views

Category:

Documents


0 download

TRANSCRIPT

  • 7/28/2019 Analytic Techniques Sc

    1/95

    SPECTROMETRY SPECTOPHOTOMETRY ATOMIC ABSORPTION MASS SPECTROMETRY

    LUMINESCENCE FLUORESCENCE CHEMILUMINESCENCE NEPHELOMETRY

    ELECTROANALYTIC METHODS ELECTROPHORESIS

    POTENTIOMETRY AMPEROMETRY

    CHROMATOGRAPHY GAS LIQUID THIN-LAYER

  • 7/28/2019 Analytic Techniques Sc

    2/95

    PHOTOMETRY measure light intensity without consideration of

    wavelength

    SPECTROPHOTOMETRY Measurement of light intensity in a selected wavelength

    ELECTROMAGNETIC RADIATION (EMR) Photons of energy traveling in waves

    Includes spectrum of energy from short-wavelength,highly energetic gamma rays and x-rays

  • 7/28/2019 Analytic Techniques Sc

    3/95

    AMPLITUDE Length of electronic vector at maximum peak height.

    PERIOD (p) Time in seconds required for the passage of successive maxima or

    minima though a fixed point in space. FREQUENCY (v)

    Number of oscillation of waveform in second Expressed in Hertz (Hz)

    Corresponds to the cycle per seconds

    WAVELENGTH Linear distance between any two equivalent points on a successivewave.

    UNIT: nanometer (nm) millimicron

    angstrom

  • 7/28/2019 Analytic Techniques Sc

    4/95

    STRAY LIGHT

    Any wavelength outside the band transmitted by

    the monochromator. Common cause: reflection of light from scratches

    on optical surfaces or from dust particles and high

    order spectra produce by diffraction gratings

    Polychromatic light

    All visible wavelengths are present

  • 7/28/2019 Analytic Techniques Sc

    5/95

  • 7/28/2019 Analytic Techniques Sc

    6/95

  • 7/28/2019 Analytic Techniques Sc

    7/95

    Ultraviolet 180-340 nm

    Visible 350 700 nmInfrared 700 - 900 nm

  • 7/28/2019 Analytic Techniques Sc

    8/95

    Wavelength Color absorbed Complementary color

    350 430 Violet Yellow-blue

    430-475 Blue yellow

    475-495 Green-blue orange

    495-505 Blue green red

    505-555 Green Purple

    555-575 Yellow green Violet

    575-600 Yellow Blue600-650 Orange Green-blue

    650-700 Red Blue-green

  • 7/28/2019 Analytic Techniques Sc

    9/95

    Where h = erg sec(plancks constant)

    v= frequencyE= energy

    Energy is directly proportional to FrequencyFrequency is inversely proportional to wavelength

    Energy is inversely proportional to wavelength

  • 7/28/2019 Analytic Techniques Sc

    10/95

    Concentration of a substance is directly proportionalto the amount of light absorbed or inverselyproportional to the logarithm of transmitted light

    A = absorbance

    a = absorptivity of a compoundb=light path of the solution

    c = concentration of compound

    %T= percent transmittance

  • 7/28/2019 Analytic Techniques Sc

    11/95

    Light Source Entrance slit Reduce stray light and prevents scattered light

    from entering the monochromator Monochromator Exit slit

    Prevent stray light from entering the cuvet. Analytic cell or Cuvet Photodetector Readout device

  • 7/28/2019 Analytic Techniques Sc

    12/95

  • 7/28/2019 Analytic Techniques Sc

    13/95

    Provides polychromatic light and must generatesufficient radiant power to measure the analyte of

    interest.

    TYPES

    Continuum source

    Emits radiation that changes in intensity very slow as a functionof wavelength

    EX: tungsten, deuterium, xenon

    Line source

    Emit a limited number of discrete lines or bands of radiation, eachof which spans a limited range of wavelengths.

  • 7/28/2019 Analytic Techniques Sc

    14/95

    Deuterium-discharge lamp and mercury arc Most commonly used for ultraviolet (UV) work

    DEUTERIUM provides continuous emission down to 165 nm

    MERCURY LAMP- Low Pressure Mercury Lamps emit sharp line spectrum, with both UV

    and visible line.

    Medium and High-pressure mercury lamps emit a continuum fromUV to the mid visible region

    Incandescent tungsten or tungsten-halogen lamp Most common source of light in the visible and near-infrared

    region.

    Silicone carbide - IR Merst glower - IR

  • 7/28/2019 Analytic Techniques Sc

    15/95

    Range Spectral distribution within the range

    The source of radiant production Stability of the radiant energy Temperature

  • 7/28/2019 Analytic Techniques Sc

    16/95

    A system of isolating a desired wavelengthand excluding others

    BANDPASS Total range of wavelength submitted

    With at more than half the maximum transmittance.

  • 7/28/2019 Analytic Techniques Sc

    17/95

    Diffraction Gratings Most commonly used Consist of 15000 or 30000 per inch etched onto a polished surface parallel

    grooves DIFFRACTION- the separation of light into component wavelength

    Prism Short wavelengths are refracted more than long wavelengths, resulting in

    dispersion of white light into continuous spectrum. Wedge-shaped pieces of glass , quartz or sodium chloride.

    Interference Filter Principle: constructive interference of waves.

    Magnesium fluoride with micro-mirror Pass very narrow range of wavelength with good efficiency

    Colored-glass filters Pass a relative wide band of radiant energy and have a low transmittance of

    the selected wavelength

  • 7/28/2019 Analytic Techniques Sc

    18/95

    Quality of monochromators are described by

    Nominal wavelength

    Represents the wavelength nanometer at peaktransmittance

    Spectral bandwidth

    Range of wavelengths above one half peak

    transmittance Bandpass

    Total range of wavelength transmitted

  • 7/28/2019 Analytic Techniques Sc

    19/95

    Square cuvets have an advantage over roundcuvets in that there is less error from the lens

    effect,orientation in the spectrophotometerand refraction. Silicate glasses- 350-2000nm GLASS CUVETS- visible range QUARTZ CUVETS -UV radiation. Polycarbonate plastic both visible and UV

    region

  • 7/28/2019 Analytic Techniques Sc

    20/95

    Converts the transmitted radiant energy into an equivalentamount of electrical energy

    BARRIER-LAYER CELLS/ PHOTOCELLS/PHOTOVOLTAIC Least expensive

    Composed of a film of light sensitive material ~SELENIUM, on a plateof iron Do not require external voltage source, rely on internal electron

    transfer to produce a current in an external circuit Output of electrical energy is not easily amplified Used mainly for in filter photometers with wide band pass,

    producing a fairly high level of illumination so that there is no need toamplify the signal Inexpensive, durable Temperature sensitive and nonlinear at very low and very high levels

    of illumination

  • 7/28/2019 Analytic Techniques Sc

    21/95

    PHOTOTUBE

    Also has photosensitive materials that gives off

    electrons when light energy strikes it Requires an outside voltage

    Has negatively charge cathode and positively

    charge anode enclosed in gas case.

    CATHODE: (rubidium, lithium) resistor on dark butemits electrons when exposed to light

    A vacuum within tubes avoids scattering of thephotoelectrons by collision with gas molecules.

  • 7/28/2019 Analytic Techniques Sc

    22/95

    PHOTOMULTIPLIER TUBE(PMT)

    detects and amplifies radiant energy

    DYNODES- series of anodes that gives off manysecondary electrons when hit by single electron.

    200x more sensitive than phototube.

    Extremely sensitive to very low light levels and

    light flashes of very short duration.

    Analog signalvoltagedigital signal (A/D

    converter)absorbance reading

  • 7/28/2019 Analytic Techniques Sc

    23/95

    SILICON DIODE TRANSDUCERS

    More sensitive than phototubes but less sensitive

    than PMT Spectral ranges from 190-1100 nm.

    Positively (p) and negatively (n) charge semi

    conductive materials adjoining one another

    embedded on a silicon chip

  • 7/28/2019 Analytic Techniques Sc

    24/95

    MULTICHANNEL PHOTON TRANSDUCER

    Consist of array of small photoelectric sensitive

    elements arranged either linearity or in a twodimensional pattern on a semiconductor chip.

    CHIP- usu. SILICON Contains electronic circuitry making it possible to determine

    the electric output signal sequentially or simultaneously.

  • 7/28/2019 Analytic Techniques Sc

    25/95

    PHOTODIODE ARRAY (PDA) Produce one dimension array of several hundred photodiodes

    set side-by-side on a single integrated circuit (IC) CHARGE-TRANSFER DEVICE

    Two dimensional array detectors that operate on a chargetransfer process

    CTD- released electron from the bound state to mobile state.

    Charge-injection devicesCharge accumulated in each pixel can be measured independently and

    nondestructively by using a network of sensing electrodes Charge couple devices

    Charged packet are moved in-step along the array row from one pixelto the next as in a bucket chain

  • 7/28/2019 Analytic Techniques Sc

    26/95

    WAVELENGHT ACCURACY

    Wavelength indicated on the control dial is the

    actual wavelength of light passed by themonochromator

    DYDIMIUM or HOLMIUM OXIDE IN GLASS

    Frequently used filter to check wavelength accuracy.

    MERCURY VAPOR LAMP Verify wavelength accuracy in narrow band pass.

    LINEARITY

  • 7/28/2019 Analytic Techniques Sc

    27/95

    Display the magnitude of the electric currentfrom a detector METER READING DEVICE

    Displays the analog detector signal by reflecting the needlealong the a scale

    DIGITAL READOUT DEVICE Send the detector signal through an analog to digital

    converter (A/D) and the microprocessor to display resultsusing a light-emitting diode (LED) or liquid crystal display(LCD)

    RECORDER (STRIP CHART OR INTEGRATOR) Maybe connected extremely to give tracing of the detector

    output against time or wavelength.

  • 7/28/2019 Analytic Techniques Sc

    28/95

    Splits monochromatic light in 2 components One beam passes through sample and other

    through a reference solution or blank Additional beam corrects for variation in light

    source intensity The absorbance of the sample can be

    recorded as the electrical output of thesample beam

  • 7/28/2019 Analytic Techniques Sc

    29/95

    1. DOUBLE-BEAM IN TIME

    Uses one photodetector and alternately passes

    the monochromatic light though the samplecuvet and then reference cuvet using a chopper.

    2. DOUBLE-BEAM IN SPACE

    Uses 2 photodetector, for the sample beam and

    reference beam

  • 7/28/2019 Analytic Techniques Sc

    30/95

  • 7/28/2019 Analytic Techniques Sc

    31/95

    Used to measure concentration by deletingabsorption of electromagnetic radiation by

    atoms rather than by molecules The element is not excited but merely

    dissociated from its chemical bonds and

    placed in a unexcited, ground state.`

  • 7/28/2019 Analytic Techniques Sc

    32/95

    If an element can be excited by externalenergy to emit radiation of a specific

    wavelength, the element in the ground statewill absorb radiation of exact wavelength.

  • 7/28/2019 Analytic Techniques Sc

    33/95

    PARTS

    1. Light Source

    2. Atomizer or burner3. Monochromator

    4. Photomultiplier tube

    5. Meter or Recorder

  • 7/28/2019 Analytic Techniques Sc

    34/95

    1. Light source:

    Hallow-cathode lamp

    Consist of an evacuated gas-tight chamber containing ananode, a cylindrical cathode and an inert gas such asHELIUM and ARGON

    Electrodeless discharge lamp

    Bulb is filled with ARGON and the element to be tested

  • 7/28/2019 Analytic Techniques Sc

    35/95

    Flame serves as a sample cell.2. ATOMIZER or BURNER Produces the flame Two light signals from the flame

    1. Alternating signal from hollow cathode lamp2. Direct signal from the flame emission

    PREMIX LONG-PATH BURNER Most common burner Sample are aspirated, mixed with air and fuel, large droplets go to

    waste.

    TOTAL CONSUMPTION BURNER Gasses like hydrogen and air and the sample are not mixed before

    entering the flame

  • 7/28/2019 Analytic Techniques Sc

    36/95

    3. MONOCHROMATOR Serves to protect the photodetector from excessive light

    emanating from flame emission USES GRATING BECAUSE

    Dispersion of spectrum is linear Has higher efficiency at UV region Its spectrum is not distorted by temperature changes.

    4. PHOTOMULTIPLIER TUBE Light detector Uses very stable AC-type electronic amplifier to boost

    the inherent amplification5. METER OR RECORDER Reads at mEq/L

  • 7/28/2019 Analytic Techniques Sc

    37/95

    Advantage Sensitive and precise For trace metal

    Propane or air acetylene flame remain in the groundstate available for light absorption~ accurate, preciseand specific

    Disadvantage Inability of the flame to dissociate samples into free

    atoms. Ionization of atoms following dissociation by the

    flame Matrix interference

  • 7/28/2019 Analytic Techniques Sc

    38/95

    INDUCTIVE COUPLES PLASMA Increase sensitivity for atomic emission.

    An argon plasma maintained by the interaction ofradiofrequency field and an ionized argon gas, isreported to have used temperatures between5500K and 8000K

    Recommended for determination of zirconium,uranium and boron

    ICP with MC is the most sensitive and specificassay technique for all elements in periodic chart.

  • 7/28/2019 Analytic Techniques Sc

    39/95

    Requires electrothermal atomization

    Electric furnace to break chemical bonds

    Electric current passes through the cylinder walls,evaporated the solvent, ashes the sample andheats the unit to incandescence to atomize the

    sample.

    Determine amount if light absorbed

  • 7/28/2019 Analytic Techniques Sc

    40/95

    Measures the light emitted by a single atom burned ina flame

    Principle: Excitation of electrons from lower to highenergy state

    Light source: flame (also serves as cuvette) Method: Indirect internal standard method Internal standard:Lithium/cesium to correct for the variation in flame and atomizer

    characteristic Used for the measurement of excited ions (sodium

    and potassium) Flickering light indicates changes in fuel reading in

    instrument

  • 7/28/2019 Analytic Techniques Sc

    41/95

    Must not normally be present in a biologicfluid

    Excitation energies of IS and that of elementbeing analyzed must be close Emission lines of IS and the element analyzed

    must be well separated.

  • 7/28/2019 Analytic Techniques Sc

    42/95

    Sodium yellow Potassium Violet

    Lithium Red Rubidium Red

    Magnesium - Blue

  • 7/28/2019 Analytic Techniques Sc

    43/95

    1. Atomizer or burnerBreaks up the specimen aliquot into fine droplets

    2. FlameCauses the evaporation of the solvent from the sample to producedry salt

    Hard flame Blue

    Soft flame Red Orange

    3. Gas and Air supplyPropane gas

    4. MonochromatorAllows only emitted line spectrum of an element under study tostrike the PM tube while rejecting the flame spectrum

    5. PM tube6. Readout device

    Galvanometer or recorder

  • 7/28/2019 Analytic Techniques Sc

    44/95

    Measurement of fluorescence Determine the amount of light emitted by a

    molecule after excitation by electromagneticradiation. Fluorescence occurs when electrons give off

    light as they drop from the excited state back

    to their ground state within a molecule Highly specific because of SIGNATURE or

    FINGERPRINT

  • 7/28/2019 Analytic Techniques Sc

    45/95

  • 7/28/2019 Analytic Techniques Sc

    46/95

    COMPONENTS Light Source (Gas discharge lamps)

    Sources of excitation radiant energy

    XENON ARC LAMP

    MERCURY Commonly use in filter fluorometers

    Attenuator Controls light intensity

    Primary filter Select wavelength that is best absorb by the solution to be measured

    Cuvet Secondary filter

    Passes the longer wavelengths of fluorescent lamp

    Light detector (PM Tube) Readout device

  • 7/28/2019 Analytic Techniques Sc

    47/95

    Advantage Specificity

    selecting the optimal wavelength for both absorption andfluorescence.

    Sensitivity 1000x more sensitive than spectrophotometric method

    Disadvantage Sensitive to environmental changes

    Affects by Quenching pH Temperature Chemical contaminants UV light changes

  • 7/28/2019 Analytic Techniques Sc

    48/95

    Part of the chemical energy generated excitedintermediates that decay .to a ground state withthe emission of photons.

    No excitation radiation is required and nomonochromators are needed because thechemiluminescene arises from one species

    CHEMILUMINESCENCE reactions are oxidationreactions ofLUMINOL, ACRIDIUM ESTERS and

    DIOXETANES characterized by rapid increase inintensity of emitted light followed by a gradualdecay.

  • 7/28/2019 Analytic Techniques Sc

    49/95

    Advantage

    Subpicomolar detection limits

    Speed With flash type reactions, light is only measured for 10

    seconds)

    Ease of use (one step procedure)

    Disadvantage Impurities can cause background signal that

    degrades sensitivity and specificity

  • 7/28/2019 Analytic Techniques Sc

    50/95

    Enhanced chemiluminescence techniquesincrease the chemiluminiescence efficiency

    by including an enhancer system in thereaction of a chemiluminescent agent with anenzyme.~60 minutes compared to 30seconds time course for the light intensity of

    conventional chemiluminescent reaction

  • 7/28/2019 Analytic Techniques Sc

    51/95

    Determine the amount of scattered light bya particulate matter suspended in a turbid

    solution Light scattering depends on wavelength and

    particle size For measuring the amount of antigen-

    antibody reaction complexes

  • 7/28/2019 Analytic Techniques Sc

    52/95

    Determine the amount oflight blocked by aparticulate matter in a turbid solution.

    The amount of light blocked by a suspensionof particles depends not only onconcentration but also on size.

    For measuring abundant large particles

    proteins and bacterial suspension

  • 7/28/2019 Analytic Techniques Sc

    53/95

    Light amplification by stimulated emission ofradiation is based on the interaction of

    radiant energy and suitably excited atoms ormolecules.

  • 7/28/2019 Analytic Techniques Sc

    54/95

    GALVANIC CELLS

    Electrodes are connected

    There is spontaneous flow of electrons from theelectron of lower affinity(oxidation), theseelectrons pass through the external meter to

    cathode (reduction), where OH- ions are liberated.

    ELECTROLYTIC CELLS- when current is force tofloe through the dead cells only by applying an

    external electromotive force E.

  • 7/28/2019 Analytic Techniques Sc

    55/95

    To rate half cell reaction, a specific electrodereaction is arbitrarily assigned 0.00V.

    (standard hydrogen eectrode:H2 gas at 1atm) Every other reaction coupled with this

    arbitrary zero reaction is either positive or

    negative, depending on the relative affinityfor electrons.

  • 7/28/2019 Analytic Techniques Sc

    56/95

    Measurement of differences in voltage (potential)between two electrodes in a solution for measuringanalyte concentration at a constant current

    Electric potentials are produced at the interfacebetween a metal and ions of that metal in solution.Also, when a membrane semipermeable to that ionseparates different concentrations of an ion.

    REFERENCE ELECTRODE Electrode with constant voltage

    INDICATOR ELECTRODE Measuring electrode

  • 7/28/2019 Analytic Techniques Sc

    57/95

    GLASS ELECTRODE first and mostcommon electrodes for measuring hydrogenion activity (pH or negative log of hydrogenion concentration)

    Consist of bulb made of layers of hydratedand nonhydrated glass, which contains a

    chloride ion buffer solution

  • 7/28/2019 Analytic Techniques Sc

    58/95

    Electrochemical transducer capable of responding to onespecific ion.

    Very sensitive and selective for the ion it measures Consist of a membrane or other barrier separating a

    reference solution and a reference electrode from thesolution to be analyzed. Depends on membrane/barrier composition that

    determines its ionic selectivity Gas electrode

    Liquid membrane electrodes

    Precipitate-impregnated membrane electrodes

    Solid state electrodes

    Gas electrodes

    Enzyme electrodes

  • 7/28/2019 Analytic Techniques Sc

    59/95

    BUFFER as known hydrogen ionconcentration

    Silver/silver chloride

    internal reference electrode

    Saturated calomel electrode

    External reference electrode

  • 7/28/2019 Analytic Techniques Sc

    60/95

    pH electrode consist of a silver wire coatedwith AgCl, immersed into an internal soln of0.1 mmol/L HCl, and placed into a tube

    containing special glass membrane tip The glass membranes are selectively

    sensitive to H+ consist of lithium, cesium,lanthanum, barium or aluminum oxide

    Potential difference between the internalsolution and test solution is measure as pHand read by voltmeter

  • 7/28/2019 Analytic Techniques Sc

    61/95

    Should be

    Reversible and obey the Nernst equation Exhibit a potential that is constant with time.

    Return to its original potential after being

    subjected to small currents

    Exhibit little hysteries (lag with temperaturecycling)

  • 7/28/2019 Analytic Techniques Sc

    62/95

    Generally consist of a metal and its salt in contact witha solution containing the same anion.

    CALOMEL ELECTRODE- commonly used

    A paste of predominant mercurous chloride, is in directcontact with metallic mercury in an electrolyte solution ofpotassium chloride.

    Slow to reach a new stable voltage following temperaturechange and it is unstable above 80C

    SILVER/SILVER CHLORIDE Can be used at temperature up to 275C

    MERCURY SULFATE AND POTASSIUM SULFATE

  • 7/28/2019 Analytic Techniques Sc

    63/95

    Set up at boundary between two dissimilarsolutions because of positive and negativeions diffusing across the boundary at unequalrates

    POTASSIUM CHLORIDE Commonly used filling solution because K+ and Cl- have

    nearly the same mobility

  • 7/28/2019 Analytic Techniques Sc

    64/95

    Selection of ISE is determined by the reactionproduct of the immobilized enzyme

    Ex:

    Urease urea

    Glucose oxidase glucose detection

  • 7/28/2019 Analytic Techniques Sc

    65/95

    1. Inert metal electrodes in contact with a redox couple Hydrogen electrode

    2. Metal electrodes that participate in redox reaction Ag/AgCl electrode

    3. Membrane electrodes1. Solid materials

    glass

    2. Liquid materials Ion exchange electrodes Calcium ISE

    3. Special membrane Gas-sensing Enzyme electrodes

  • 7/28/2019 Analytic Techniques Sc

    66/95

    Measures the quantity of electricity (incoulombs) needed to convert an analyte to adifferent oxidation state

    Coulomb is the quantity of electricity orcharge that is transported in 1 second by aconstant current of 1 ampere for t second.

    Used to measure chloride ion in serum,plasma , CSF and sweat samples.

  • 7/28/2019 Analytic Techniques Sc

    67/95

    pH electrode contained within a plastic jacket Filled with sodium bicarbonate buffer and has

    a gas-permeable membrane (teflon orsilicone) across its opening

  • 7/28/2019 Analytic Techniques Sc

    68/95

    Measurement of the current flow producedby an oxidation- reduction reaction.

    For chloride measurement (coloumetry-amperometry)

  • 7/28/2019 Analytic Techniques Sc

    69/95

    Potential is applied to an electrochemical celland the resulting current is measured

    ADVANTAGE: sensitivity and capability formulti element measurement

    Analytes can be detected in parts per billionrange

    ANODIC STRIPPING VOLTAMMETRY Electrochemical technique used to measure heavy

    metals such as lead

  • 7/28/2019 Analytic Techniques Sc

    70/95

    CONDUCTANCE

    Electrolytic conductivity is a measure of the ability

    of a solution to carry an electric current

    IMPEDANCE

    Based on the change in electrical resistance across

    an aperture when a particle in conductive liquid

    passes through this aperture Used in hematology laboratory to enumerate

    leukocytes, erythrocytes and platelets

  • 7/28/2019 Analytic Techniques Sc

    71/95

    Performed similarly to other electrophoresis methods,except that the separating molecules migrate through pHgradient

    deal for separating .proteins of identical sizes but with

    different net charges. Protein s move in the electric field until they reach a pHequal to their isoelectric point.

    Ph gradient is made by adding acid to anodic side andadding base to cathode area.

    ADV: ability to resolve mixture of proteins For measuring serum acid phosphatase isoenzymes,

    detection of oligoclonal immunoglobulin bands in CSF andisoenzymes of creatine kinase and alkaline phosphatase inserum.

  • 7/28/2019 Analytic Techniques Sc

    72/95

    ELECTROPHORESIS

    Separation of charged compounds based on their

    electrical charge

    Cations go towards anode, anions go towardscathode

    The greater the net charges of a dissolve

    compound, the faster it moves through hesolution toward the opposite charged electrode.

  • 7/28/2019 Analytic Techniques Sc

    73/95

    Measures the absorbance of the stain on asupport medium.

    COMPONENTS

    Light source

    Monochromator

    Movable carriage

    Optical system

    Photodetector

  • 7/28/2019 Analytic Techniques Sc

    74/95

    Migration of charged solutes or particles in anelectric field

    IONTROPHORESIS

    Refers to the migration of small ions ZONE ELECTROPHORESIS Migration of charged macromolecules in a porous

    support medium such as paper, cellulose acetate or

    agarose gel film ELECTROPHORETOGRAM

    Result of zone electrophoresis

    Consist of sharply separated zone of macromolecules

  • 7/28/2019 Analytic Techniques Sc

    75/95

    Driving force (electrical power) Support medium Buffer Sample Detecting system

  • 7/28/2019 Analytic Techniques Sc

    76/95

    Charged particles migrate toward the oppositecharged electrode

    Velocity of migration depends on: Net charge of particle

    Size and shape of the particle Strength of the electric field Chemical and physical properties of the supporting

    medium

    Electrophoretic temperature

    Rate of migration is directly proportional to the netcharge of the particle and inversely proportional to itssize and the viscosity of the buffer

  • 7/28/2019 Analytic Techniques Sc

    77/95

    Cellulose acetate Formed by treating cellulose acetylated with acetic anhydride Dry, brittle, film composed of about 80% air space. Also used in isoelectric focusingjm

    Agarose gel

    Purified fraction of agar, it is neutral and does not produce electroendosmosis Requires small amount of sample (approx 2mL) Does not bind protein therefore migration is not affected

    Polyacrylamide gel Separation of protein on the basis of charge and molecular size Separate serum into 20 or more fractions rather than the usual five fraction.

    Widely used to study individual proteins Starch gel

    Separates protein on the basis of surface charge and molecular size Not widely use because of technical difficulty in preparing gel.

  • 7/28/2019 Analytic Techniques Sc

    78/95

    Separation is performed in narrow-bore fused silicacapillaries (inner diameter 2575 nm)

    Concept: electro-osmotic flow (EOF) EOF is the bulk flow of liquid toward the cathode

    upon application of electric field and it issuperimposed on electrophoretic migration. Cation migrate faster because both EOF and

    electrophoretic attraction are towards cathode. For separation, quantitation and determination of

    molecular weights of proteins and peptides, analysisof PCR products, analysis of organic ions, organicacids, pharmaceuticals, optical isomers and drugs ofabuse in serum and urine.

  • 7/28/2019 Analytic Techniques Sc

    79/95

    The movement of buffer ions an solventrelative to the fixed support

    Support media: paper, cellulose acetate andagar gel

    Hydroxyl ion remain fixed while free positiveions move toward the cathode.

  • 7/28/2019 Analytic Techniques Sc

    80/95

    Separation method based on differentinteraction of the specimen compounds withthe mobile phase and the stationary phase ascompound travel through a support medium.

    RETENTION TIME(tR) Time it takes a compound to elute

  • 7/28/2019 Analytic Techniques Sc

    81/95

    MOBILE PHASE Gas or liquid

    Carries the complex mixture (sample)

    STATIONARY PHASE Solid or liquid

    Through which the mobile phase flows COLUMN Holding the stationary phase

    SEPARATED COMPONENTS eluate

  • 7/28/2019 Analytic Techniques Sc

    82/95

    Adsorption AKA liquid-solid chromatography Based on the competition between the sample and the mobile phase

    for absorption sites on the solid stationary phase Molecules that are most soluble in the mobile phase , move fastest

    Partition AKA liquid-liquid chromatography Separation of solute based on relative solubility in an organic

    (nonpolar) solvent and aqueous(polar) solvent

    Normal phase mobile solvent is less polar than the stationary solvent Reverse phase mobile solvent is more polar

    Applicable to any substance that maybe distributed between twoliquid

    Works best in nonionic compound

  • 7/28/2019 Analytic Techniques Sc

    83/95

    Steric exclusion

    Variation of liquid-solid chromatography

    Separate molecules on the basis of the size and shape

    Ion exchange

    Solute mixture are separated by virtue of themagnitude and charge of ionic species.

    Stationary phase is a resin consisting of largepolymers of substituted benzene, silicates or cellulosederivatives, with charge functional groups

  • 7/28/2019 Analytic Techniques Sc

    84/95

    THINLAYER CHROMATOGRAPHY

    Thin layer of sorbent , such as alumina , silica gel,

    cellulose or cross linked dextran, is uniformly

    coated on a glass or plastic plate

    Most commonly used as semiquantitative

    screening test

  • 7/28/2019 Analytic Techniques Sc

    85/95

    Uses pressure for fast separations, controlledtemperature, in-line detectors and gradientelution techniques

    PUMPS Forces the mobile phase through the column at a

    much greater velocity than that accomplished bygravity-flow columns and includes pneumatic,syringe, reciprocating or hydraulic amplifier pumps.

    MECHANICAL RECIPROCATING PUMP

    Multihead pump with two or more reciprocating pistons.

  • 7/28/2019 Analytic Techniques Sc

    86/95

    COLUMNS

    Long stainless steel where the stationary phase is

    packed

    SILICA GEL most common material used forpacking

    Stable and can be used in different ways.

    Reverse-phase HPLC Stationary phase is nonpolar molecules(octadecyl C-18

    hydrocardbon) bonded to silica gel particles.

  • 7/28/2019 Analytic Techniques Sc

    87/95

    SAMPLE INJECTORS Small syringe Loop injector best and widely used

    High reproducibility and used at high temperature.

    DETECTORS Monitor the eluate as it leaves the column and produce an

    electronic signal proportional to the concentration of eachseparated component

    RECORDERS

    Used to record detector signal versus the time the mobilephase passed through the instrument, starting from thetime of injection.

    CHROMATOGRAM -

  • 7/28/2019 Analytic Techniques Sc

    88/95

    Separate mixture of compounds hat arevolatile or can be made volatile.

    GAS-SOLID CHROMATOGRAPHY (GSC)

    solid stationary phase

    GAS-LIQUID CHROMATOGRAPHY (GLC)

    With non volatile liquid stationary phase.

    Commonly used in clinical laboratories.

  • 7/28/2019 Analytic Techniques Sc

    89/95

  • 7/28/2019 Analytic Techniques Sc

    90/95

    ADVANTAGE

    Increase the number of tests to be performed in a

    given period

    Minimizes variation of result from onelaboratorian to another

    Eliminates the potential error in manual analyses

    such as pipetting, calculation and transcript ofresult

  • 7/28/2019 Analytic Techniques Sc

    91/95

    CONTINUOUS FLOW ANALYZER Liquids are pump through a system of continuous tubing

    Samples flow through a common reaction vessel orpathway.

    Air bubbles at regular intervals serves as separating andcleaning media

    Mixture of sample and reagent takes place using a glasscoil inserted into flow path

    A heating bath maintains the required temperature ofthe reaction to allow complete color development-reaction rate is controlled by temperature

    Ex Simultaneous multiple analyzer (SMA), technicon

  • 7/28/2019 Analytic Techniques Sc

    92/95

    CENTRIFUGAL ANALYZER Use the force generated by centrifugation to transfer

    specimen and reagents

    Liquids are placed in separate cuvets for measurement atthe perimeter of a spinning rotor (1000rpm)

    It uses acceleration and deceleration of the rotor totransfer the reagents and sample from one chamber toanother

    For mixing, centrifugal force or rotor is utilized orbubbling of air

    Major advantage: Batch analysis

    Ex Cobas- Bio (Roche), IL Monarch

  • 7/28/2019 Analytic Techniques Sc

    93/95

    DISCRETE ANALYZER Most popular and versatile analyzer Employs variety of syringe and pipettes to aspirate and dispense sample and

    reagents Positive liquid-displacement pipettes are used for sampling

    Capable of running multiple-test-one-sample at-a-time. Each sample-reagent mixture handles separately in its own vessel Has Random access capability that allows STAT samples to be easily

    accessed. For mixing, magnetic driven teflon stirring bar located in the bottom of the

    reaction chamber is used in Beckman ASTRA For dry slide technology (reflectance photometry, the spreading layer permits

    a rapid uniform spreading layer over the reagent layer. REFLECTANCE PHOTOMETRY- measure the light reflected from solid surfaces.

    EX: Vitros, dimension Dade, Beckman ASTRA System, Hitachi, Bayer Advia Roche CobasIntegra 800, Roche anlaysis P module, ACA star (Dade)

  • 7/28/2019 Analytic Techniques Sc

    94/95

    BATCH TESTING All samples are loaded at the same time, and singles test is conducted

    on each sample PARALLEL TESTING

    More than one test is analyzed concurrently on a given clinical

    specimen RANDOM ACCESS TESTING

    Any test can be performed on any sample in any sequence SEQUENTIAL TESTING

    Multiple tests analyzed one after another on a given specimen OPEN REAGENT SYSTEM

    A system other than manufacturers reagent can be utilized formeasurement

    CLOSE REAGENT SYSTEM A system where the operator can only used the manufacturer

    reagents.

  • 7/28/2019 Analytic Techniques Sc

    95/95

    DONT FORGET TO READ YOUR BOOKS!

    KEEP CALMANDPRAY

    GODBLESS! =)