analytic techniques sc
TRANSCRIPT
-
7/28/2019 Analytic Techniques Sc
1/95
SPECTROMETRY SPECTOPHOTOMETRY ATOMIC ABSORPTION MASS SPECTROMETRY
LUMINESCENCE FLUORESCENCE CHEMILUMINESCENCE NEPHELOMETRY
ELECTROANALYTIC METHODS ELECTROPHORESIS
POTENTIOMETRY AMPEROMETRY
CHROMATOGRAPHY GAS LIQUID THIN-LAYER
-
7/28/2019 Analytic Techniques Sc
2/95
PHOTOMETRY measure light intensity without consideration of
wavelength
SPECTROPHOTOMETRY Measurement of light intensity in a selected wavelength
ELECTROMAGNETIC RADIATION (EMR) Photons of energy traveling in waves
Includes spectrum of energy from short-wavelength,highly energetic gamma rays and x-rays
-
7/28/2019 Analytic Techniques Sc
3/95
AMPLITUDE Length of electronic vector at maximum peak height.
PERIOD (p) Time in seconds required for the passage of successive maxima or
minima though a fixed point in space. FREQUENCY (v)
Number of oscillation of waveform in second Expressed in Hertz (Hz)
Corresponds to the cycle per seconds
WAVELENGTH Linear distance between any two equivalent points on a successivewave.
UNIT: nanometer (nm) millimicron
angstrom
-
7/28/2019 Analytic Techniques Sc
4/95
STRAY LIGHT
Any wavelength outside the band transmitted by
the monochromator. Common cause: reflection of light from scratches
on optical surfaces or from dust particles and high
order spectra produce by diffraction gratings
Polychromatic light
All visible wavelengths are present
-
7/28/2019 Analytic Techniques Sc
5/95
-
7/28/2019 Analytic Techniques Sc
6/95
-
7/28/2019 Analytic Techniques Sc
7/95
Ultraviolet 180-340 nm
Visible 350 700 nmInfrared 700 - 900 nm
-
7/28/2019 Analytic Techniques Sc
8/95
Wavelength Color absorbed Complementary color
350 430 Violet Yellow-blue
430-475 Blue yellow
475-495 Green-blue orange
495-505 Blue green red
505-555 Green Purple
555-575 Yellow green Violet
575-600 Yellow Blue600-650 Orange Green-blue
650-700 Red Blue-green
-
7/28/2019 Analytic Techniques Sc
9/95
Where h = erg sec(plancks constant)
v= frequencyE= energy
Energy is directly proportional to FrequencyFrequency is inversely proportional to wavelength
Energy is inversely proportional to wavelength
-
7/28/2019 Analytic Techniques Sc
10/95
Concentration of a substance is directly proportionalto the amount of light absorbed or inverselyproportional to the logarithm of transmitted light
A = absorbance
a = absorptivity of a compoundb=light path of the solution
c = concentration of compound
%T= percent transmittance
-
7/28/2019 Analytic Techniques Sc
11/95
Light Source Entrance slit Reduce stray light and prevents scattered light
from entering the monochromator Monochromator Exit slit
Prevent stray light from entering the cuvet. Analytic cell or Cuvet Photodetector Readout device
-
7/28/2019 Analytic Techniques Sc
12/95
-
7/28/2019 Analytic Techniques Sc
13/95
Provides polychromatic light and must generatesufficient radiant power to measure the analyte of
interest.
TYPES
Continuum source
Emits radiation that changes in intensity very slow as a functionof wavelength
EX: tungsten, deuterium, xenon
Line source
Emit a limited number of discrete lines or bands of radiation, eachof which spans a limited range of wavelengths.
-
7/28/2019 Analytic Techniques Sc
14/95
Deuterium-discharge lamp and mercury arc Most commonly used for ultraviolet (UV) work
DEUTERIUM provides continuous emission down to 165 nm
MERCURY LAMP- Low Pressure Mercury Lamps emit sharp line spectrum, with both UV
and visible line.
Medium and High-pressure mercury lamps emit a continuum fromUV to the mid visible region
Incandescent tungsten or tungsten-halogen lamp Most common source of light in the visible and near-infrared
region.
Silicone carbide - IR Merst glower - IR
-
7/28/2019 Analytic Techniques Sc
15/95
Range Spectral distribution within the range
The source of radiant production Stability of the radiant energy Temperature
-
7/28/2019 Analytic Techniques Sc
16/95
A system of isolating a desired wavelengthand excluding others
BANDPASS Total range of wavelength submitted
With at more than half the maximum transmittance.
-
7/28/2019 Analytic Techniques Sc
17/95
Diffraction Gratings Most commonly used Consist of 15000 or 30000 per inch etched onto a polished surface parallel
grooves DIFFRACTION- the separation of light into component wavelength
Prism Short wavelengths are refracted more than long wavelengths, resulting in
dispersion of white light into continuous spectrum. Wedge-shaped pieces of glass , quartz or sodium chloride.
Interference Filter Principle: constructive interference of waves.
Magnesium fluoride with micro-mirror Pass very narrow range of wavelength with good efficiency
Colored-glass filters Pass a relative wide band of radiant energy and have a low transmittance of
the selected wavelength
-
7/28/2019 Analytic Techniques Sc
18/95
Quality of monochromators are described by
Nominal wavelength
Represents the wavelength nanometer at peaktransmittance
Spectral bandwidth
Range of wavelengths above one half peak
transmittance Bandpass
Total range of wavelength transmitted
-
7/28/2019 Analytic Techniques Sc
19/95
Square cuvets have an advantage over roundcuvets in that there is less error from the lens
effect,orientation in the spectrophotometerand refraction. Silicate glasses- 350-2000nm GLASS CUVETS- visible range QUARTZ CUVETS -UV radiation. Polycarbonate plastic both visible and UV
region
-
7/28/2019 Analytic Techniques Sc
20/95
Converts the transmitted radiant energy into an equivalentamount of electrical energy
BARRIER-LAYER CELLS/ PHOTOCELLS/PHOTOVOLTAIC Least expensive
Composed of a film of light sensitive material ~SELENIUM, on a plateof iron Do not require external voltage source, rely on internal electron
transfer to produce a current in an external circuit Output of electrical energy is not easily amplified Used mainly for in filter photometers with wide band pass,
producing a fairly high level of illumination so that there is no need toamplify the signal Inexpensive, durable Temperature sensitive and nonlinear at very low and very high levels
of illumination
-
7/28/2019 Analytic Techniques Sc
21/95
PHOTOTUBE
Also has photosensitive materials that gives off
electrons when light energy strikes it Requires an outside voltage
Has negatively charge cathode and positively
charge anode enclosed in gas case.
CATHODE: (rubidium, lithium) resistor on dark butemits electrons when exposed to light
A vacuum within tubes avoids scattering of thephotoelectrons by collision with gas molecules.
-
7/28/2019 Analytic Techniques Sc
22/95
PHOTOMULTIPLIER TUBE(PMT)
detects and amplifies radiant energy
DYNODES- series of anodes that gives off manysecondary electrons when hit by single electron.
200x more sensitive than phototube.
Extremely sensitive to very low light levels and
light flashes of very short duration.
Analog signalvoltagedigital signal (A/D
converter)absorbance reading
-
7/28/2019 Analytic Techniques Sc
23/95
SILICON DIODE TRANSDUCERS
More sensitive than phototubes but less sensitive
than PMT Spectral ranges from 190-1100 nm.
Positively (p) and negatively (n) charge semi
conductive materials adjoining one another
embedded on a silicon chip
-
7/28/2019 Analytic Techniques Sc
24/95
MULTICHANNEL PHOTON TRANSDUCER
Consist of array of small photoelectric sensitive
elements arranged either linearity or in a twodimensional pattern on a semiconductor chip.
CHIP- usu. SILICON Contains electronic circuitry making it possible to determine
the electric output signal sequentially or simultaneously.
-
7/28/2019 Analytic Techniques Sc
25/95
PHOTODIODE ARRAY (PDA) Produce one dimension array of several hundred photodiodes
set side-by-side on a single integrated circuit (IC) CHARGE-TRANSFER DEVICE
Two dimensional array detectors that operate on a chargetransfer process
CTD- released electron from the bound state to mobile state.
Charge-injection devicesCharge accumulated in each pixel can be measured independently and
nondestructively by using a network of sensing electrodes Charge couple devices
Charged packet are moved in-step along the array row from one pixelto the next as in a bucket chain
-
7/28/2019 Analytic Techniques Sc
26/95
WAVELENGHT ACCURACY
Wavelength indicated on the control dial is the
actual wavelength of light passed by themonochromator
DYDIMIUM or HOLMIUM OXIDE IN GLASS
Frequently used filter to check wavelength accuracy.
MERCURY VAPOR LAMP Verify wavelength accuracy in narrow band pass.
LINEARITY
-
7/28/2019 Analytic Techniques Sc
27/95
Display the magnitude of the electric currentfrom a detector METER READING DEVICE
Displays the analog detector signal by reflecting the needlealong the a scale
DIGITAL READOUT DEVICE Send the detector signal through an analog to digital
converter (A/D) and the microprocessor to display resultsusing a light-emitting diode (LED) or liquid crystal display(LCD)
RECORDER (STRIP CHART OR INTEGRATOR) Maybe connected extremely to give tracing of the detector
output against time or wavelength.
-
7/28/2019 Analytic Techniques Sc
28/95
Splits monochromatic light in 2 components One beam passes through sample and other
through a reference solution or blank Additional beam corrects for variation in light
source intensity The absorbance of the sample can be
recorded as the electrical output of thesample beam
-
7/28/2019 Analytic Techniques Sc
29/95
1. DOUBLE-BEAM IN TIME
Uses one photodetector and alternately passes
the monochromatic light though the samplecuvet and then reference cuvet using a chopper.
2. DOUBLE-BEAM IN SPACE
Uses 2 photodetector, for the sample beam and
reference beam
-
7/28/2019 Analytic Techniques Sc
30/95
-
7/28/2019 Analytic Techniques Sc
31/95
Used to measure concentration by deletingabsorption of electromagnetic radiation by
atoms rather than by molecules The element is not excited but merely
dissociated from its chemical bonds and
placed in a unexcited, ground state.`
-
7/28/2019 Analytic Techniques Sc
32/95
If an element can be excited by externalenergy to emit radiation of a specific
wavelength, the element in the ground statewill absorb radiation of exact wavelength.
-
7/28/2019 Analytic Techniques Sc
33/95
PARTS
1. Light Source
2. Atomizer or burner3. Monochromator
4. Photomultiplier tube
5. Meter or Recorder
-
7/28/2019 Analytic Techniques Sc
34/95
1. Light source:
Hallow-cathode lamp
Consist of an evacuated gas-tight chamber containing ananode, a cylindrical cathode and an inert gas such asHELIUM and ARGON
Electrodeless discharge lamp
Bulb is filled with ARGON and the element to be tested
-
7/28/2019 Analytic Techniques Sc
35/95
Flame serves as a sample cell.2. ATOMIZER or BURNER Produces the flame Two light signals from the flame
1. Alternating signal from hollow cathode lamp2. Direct signal from the flame emission
PREMIX LONG-PATH BURNER Most common burner Sample are aspirated, mixed with air and fuel, large droplets go to
waste.
TOTAL CONSUMPTION BURNER Gasses like hydrogen and air and the sample are not mixed before
entering the flame
-
7/28/2019 Analytic Techniques Sc
36/95
3. MONOCHROMATOR Serves to protect the photodetector from excessive light
emanating from flame emission USES GRATING BECAUSE
Dispersion of spectrum is linear Has higher efficiency at UV region Its spectrum is not distorted by temperature changes.
4. PHOTOMULTIPLIER TUBE Light detector Uses very stable AC-type electronic amplifier to boost
the inherent amplification5. METER OR RECORDER Reads at mEq/L
-
7/28/2019 Analytic Techniques Sc
37/95
Advantage Sensitive and precise For trace metal
Propane or air acetylene flame remain in the groundstate available for light absorption~ accurate, preciseand specific
Disadvantage Inability of the flame to dissociate samples into free
atoms. Ionization of atoms following dissociation by the
flame Matrix interference
-
7/28/2019 Analytic Techniques Sc
38/95
INDUCTIVE COUPLES PLASMA Increase sensitivity for atomic emission.
An argon plasma maintained by the interaction ofradiofrequency field and an ionized argon gas, isreported to have used temperatures between5500K and 8000K
Recommended for determination of zirconium,uranium and boron
ICP with MC is the most sensitive and specificassay technique for all elements in periodic chart.
-
7/28/2019 Analytic Techniques Sc
39/95
Requires electrothermal atomization
Electric furnace to break chemical bonds
Electric current passes through the cylinder walls,evaporated the solvent, ashes the sample andheats the unit to incandescence to atomize the
sample.
Determine amount if light absorbed
-
7/28/2019 Analytic Techniques Sc
40/95
Measures the light emitted by a single atom burned ina flame
Principle: Excitation of electrons from lower to highenergy state
Light source: flame (also serves as cuvette) Method: Indirect internal standard method Internal standard:Lithium/cesium to correct for the variation in flame and atomizer
characteristic Used for the measurement of excited ions (sodium
and potassium) Flickering light indicates changes in fuel reading in
instrument
-
7/28/2019 Analytic Techniques Sc
41/95
Must not normally be present in a biologicfluid
Excitation energies of IS and that of elementbeing analyzed must be close Emission lines of IS and the element analyzed
must be well separated.
-
7/28/2019 Analytic Techniques Sc
42/95
Sodium yellow Potassium Violet
Lithium Red Rubidium Red
Magnesium - Blue
-
7/28/2019 Analytic Techniques Sc
43/95
1. Atomizer or burnerBreaks up the specimen aliquot into fine droplets
2. FlameCauses the evaporation of the solvent from the sample to producedry salt
Hard flame Blue
Soft flame Red Orange
3. Gas and Air supplyPropane gas
4. MonochromatorAllows only emitted line spectrum of an element under study tostrike the PM tube while rejecting the flame spectrum
5. PM tube6. Readout device
Galvanometer or recorder
-
7/28/2019 Analytic Techniques Sc
44/95
Measurement of fluorescence Determine the amount of light emitted by a
molecule after excitation by electromagneticradiation. Fluorescence occurs when electrons give off
light as they drop from the excited state back
to their ground state within a molecule Highly specific because of SIGNATURE or
FINGERPRINT
-
7/28/2019 Analytic Techniques Sc
45/95
-
7/28/2019 Analytic Techniques Sc
46/95
COMPONENTS Light Source (Gas discharge lamps)
Sources of excitation radiant energy
XENON ARC LAMP
MERCURY Commonly use in filter fluorometers
Attenuator Controls light intensity
Primary filter Select wavelength that is best absorb by the solution to be measured
Cuvet Secondary filter
Passes the longer wavelengths of fluorescent lamp
Light detector (PM Tube) Readout device
-
7/28/2019 Analytic Techniques Sc
47/95
Advantage Specificity
selecting the optimal wavelength for both absorption andfluorescence.
Sensitivity 1000x more sensitive than spectrophotometric method
Disadvantage Sensitive to environmental changes
Affects by Quenching pH Temperature Chemical contaminants UV light changes
-
7/28/2019 Analytic Techniques Sc
48/95
Part of the chemical energy generated excitedintermediates that decay .to a ground state withthe emission of photons.
No excitation radiation is required and nomonochromators are needed because thechemiluminescene arises from one species
CHEMILUMINESCENCE reactions are oxidationreactions ofLUMINOL, ACRIDIUM ESTERS and
DIOXETANES characterized by rapid increase inintensity of emitted light followed by a gradualdecay.
-
7/28/2019 Analytic Techniques Sc
49/95
Advantage
Subpicomolar detection limits
Speed With flash type reactions, light is only measured for 10
seconds)
Ease of use (one step procedure)
Disadvantage Impurities can cause background signal that
degrades sensitivity and specificity
-
7/28/2019 Analytic Techniques Sc
50/95
Enhanced chemiluminescence techniquesincrease the chemiluminiescence efficiency
by including an enhancer system in thereaction of a chemiluminescent agent with anenzyme.~60 minutes compared to 30seconds time course for the light intensity of
conventional chemiluminescent reaction
-
7/28/2019 Analytic Techniques Sc
51/95
Determine the amount of scattered light bya particulate matter suspended in a turbid
solution Light scattering depends on wavelength and
particle size For measuring the amount of antigen-
antibody reaction complexes
-
7/28/2019 Analytic Techniques Sc
52/95
Determine the amount oflight blocked by aparticulate matter in a turbid solution.
The amount of light blocked by a suspensionof particles depends not only onconcentration but also on size.
For measuring abundant large particles
proteins and bacterial suspension
-
7/28/2019 Analytic Techniques Sc
53/95
Light amplification by stimulated emission ofradiation is based on the interaction of
radiant energy and suitably excited atoms ormolecules.
-
7/28/2019 Analytic Techniques Sc
54/95
GALVANIC CELLS
Electrodes are connected
There is spontaneous flow of electrons from theelectron of lower affinity(oxidation), theseelectrons pass through the external meter to
cathode (reduction), where OH- ions are liberated.
ELECTROLYTIC CELLS- when current is force tofloe through the dead cells only by applying an
external electromotive force E.
-
7/28/2019 Analytic Techniques Sc
55/95
To rate half cell reaction, a specific electrodereaction is arbitrarily assigned 0.00V.
(standard hydrogen eectrode:H2 gas at 1atm) Every other reaction coupled with this
arbitrary zero reaction is either positive or
negative, depending on the relative affinityfor electrons.
-
7/28/2019 Analytic Techniques Sc
56/95
Measurement of differences in voltage (potential)between two electrodes in a solution for measuringanalyte concentration at a constant current
Electric potentials are produced at the interfacebetween a metal and ions of that metal in solution.Also, when a membrane semipermeable to that ionseparates different concentrations of an ion.
REFERENCE ELECTRODE Electrode with constant voltage
INDICATOR ELECTRODE Measuring electrode
-
7/28/2019 Analytic Techniques Sc
57/95
GLASS ELECTRODE first and mostcommon electrodes for measuring hydrogenion activity (pH or negative log of hydrogenion concentration)
Consist of bulb made of layers of hydratedand nonhydrated glass, which contains a
chloride ion buffer solution
-
7/28/2019 Analytic Techniques Sc
58/95
Electrochemical transducer capable of responding to onespecific ion.
Very sensitive and selective for the ion it measures Consist of a membrane or other barrier separating a
reference solution and a reference electrode from thesolution to be analyzed. Depends on membrane/barrier composition that
determines its ionic selectivity Gas electrode
Liquid membrane electrodes
Precipitate-impregnated membrane electrodes
Solid state electrodes
Gas electrodes
Enzyme electrodes
-
7/28/2019 Analytic Techniques Sc
59/95
BUFFER as known hydrogen ionconcentration
Silver/silver chloride
internal reference electrode
Saturated calomel electrode
External reference electrode
-
7/28/2019 Analytic Techniques Sc
60/95
pH electrode consist of a silver wire coatedwith AgCl, immersed into an internal soln of0.1 mmol/L HCl, and placed into a tube
containing special glass membrane tip The glass membranes are selectively
sensitive to H+ consist of lithium, cesium,lanthanum, barium or aluminum oxide
Potential difference between the internalsolution and test solution is measure as pHand read by voltmeter
-
7/28/2019 Analytic Techniques Sc
61/95
Should be
Reversible and obey the Nernst equation Exhibit a potential that is constant with time.
Return to its original potential after being
subjected to small currents
Exhibit little hysteries (lag with temperaturecycling)
-
7/28/2019 Analytic Techniques Sc
62/95
Generally consist of a metal and its salt in contact witha solution containing the same anion.
CALOMEL ELECTRODE- commonly used
A paste of predominant mercurous chloride, is in directcontact with metallic mercury in an electrolyte solution ofpotassium chloride.
Slow to reach a new stable voltage following temperaturechange and it is unstable above 80C
SILVER/SILVER CHLORIDE Can be used at temperature up to 275C
MERCURY SULFATE AND POTASSIUM SULFATE
-
7/28/2019 Analytic Techniques Sc
63/95
Set up at boundary between two dissimilarsolutions because of positive and negativeions diffusing across the boundary at unequalrates
POTASSIUM CHLORIDE Commonly used filling solution because K+ and Cl- have
nearly the same mobility
-
7/28/2019 Analytic Techniques Sc
64/95
Selection of ISE is determined by the reactionproduct of the immobilized enzyme
Ex:
Urease urea
Glucose oxidase glucose detection
-
7/28/2019 Analytic Techniques Sc
65/95
1. Inert metal electrodes in contact with a redox couple Hydrogen electrode
2. Metal electrodes that participate in redox reaction Ag/AgCl electrode
3. Membrane electrodes1. Solid materials
glass
2. Liquid materials Ion exchange electrodes Calcium ISE
3. Special membrane Gas-sensing Enzyme electrodes
-
7/28/2019 Analytic Techniques Sc
66/95
Measures the quantity of electricity (incoulombs) needed to convert an analyte to adifferent oxidation state
Coulomb is the quantity of electricity orcharge that is transported in 1 second by aconstant current of 1 ampere for t second.
Used to measure chloride ion in serum,plasma , CSF and sweat samples.
-
7/28/2019 Analytic Techniques Sc
67/95
pH electrode contained within a plastic jacket Filled with sodium bicarbonate buffer and has
a gas-permeable membrane (teflon orsilicone) across its opening
-
7/28/2019 Analytic Techniques Sc
68/95
Measurement of the current flow producedby an oxidation- reduction reaction.
For chloride measurement (coloumetry-amperometry)
-
7/28/2019 Analytic Techniques Sc
69/95
Potential is applied to an electrochemical celland the resulting current is measured
ADVANTAGE: sensitivity and capability formulti element measurement
Analytes can be detected in parts per billionrange
ANODIC STRIPPING VOLTAMMETRY Electrochemical technique used to measure heavy
metals such as lead
-
7/28/2019 Analytic Techniques Sc
70/95
CONDUCTANCE
Electrolytic conductivity is a measure of the ability
of a solution to carry an electric current
IMPEDANCE
Based on the change in electrical resistance across
an aperture when a particle in conductive liquid
passes through this aperture Used in hematology laboratory to enumerate
leukocytes, erythrocytes and platelets
-
7/28/2019 Analytic Techniques Sc
71/95
Performed similarly to other electrophoresis methods,except that the separating molecules migrate through pHgradient
deal for separating .proteins of identical sizes but with
different net charges. Protein s move in the electric field until they reach a pHequal to their isoelectric point.
Ph gradient is made by adding acid to anodic side andadding base to cathode area.
ADV: ability to resolve mixture of proteins For measuring serum acid phosphatase isoenzymes,
detection of oligoclonal immunoglobulin bands in CSF andisoenzymes of creatine kinase and alkaline phosphatase inserum.
-
7/28/2019 Analytic Techniques Sc
72/95
ELECTROPHORESIS
Separation of charged compounds based on their
electrical charge
Cations go towards anode, anions go towardscathode
The greater the net charges of a dissolve
compound, the faster it moves through hesolution toward the opposite charged electrode.
-
7/28/2019 Analytic Techniques Sc
73/95
Measures the absorbance of the stain on asupport medium.
COMPONENTS
Light source
Monochromator
Movable carriage
Optical system
Photodetector
-
7/28/2019 Analytic Techniques Sc
74/95
Migration of charged solutes or particles in anelectric field
IONTROPHORESIS
Refers to the migration of small ions ZONE ELECTROPHORESIS Migration of charged macromolecules in a porous
support medium such as paper, cellulose acetate or
agarose gel film ELECTROPHORETOGRAM
Result of zone electrophoresis
Consist of sharply separated zone of macromolecules
-
7/28/2019 Analytic Techniques Sc
75/95
Driving force (electrical power) Support medium Buffer Sample Detecting system
-
7/28/2019 Analytic Techniques Sc
76/95
Charged particles migrate toward the oppositecharged electrode
Velocity of migration depends on: Net charge of particle
Size and shape of the particle Strength of the electric field Chemical and physical properties of the supporting
medium
Electrophoretic temperature
Rate of migration is directly proportional to the netcharge of the particle and inversely proportional to itssize and the viscosity of the buffer
-
7/28/2019 Analytic Techniques Sc
77/95
Cellulose acetate Formed by treating cellulose acetylated with acetic anhydride Dry, brittle, film composed of about 80% air space. Also used in isoelectric focusingjm
Agarose gel
Purified fraction of agar, it is neutral and does not produce electroendosmosis Requires small amount of sample (approx 2mL) Does not bind protein therefore migration is not affected
Polyacrylamide gel Separation of protein on the basis of charge and molecular size Separate serum into 20 or more fractions rather than the usual five fraction.
Widely used to study individual proteins Starch gel
Separates protein on the basis of surface charge and molecular size Not widely use because of technical difficulty in preparing gel.
-
7/28/2019 Analytic Techniques Sc
78/95
Separation is performed in narrow-bore fused silicacapillaries (inner diameter 2575 nm)
Concept: electro-osmotic flow (EOF) EOF is the bulk flow of liquid toward the cathode
upon application of electric field and it issuperimposed on electrophoretic migration. Cation migrate faster because both EOF and
electrophoretic attraction are towards cathode. For separation, quantitation and determination of
molecular weights of proteins and peptides, analysisof PCR products, analysis of organic ions, organicacids, pharmaceuticals, optical isomers and drugs ofabuse in serum and urine.
-
7/28/2019 Analytic Techniques Sc
79/95
The movement of buffer ions an solventrelative to the fixed support
Support media: paper, cellulose acetate andagar gel
Hydroxyl ion remain fixed while free positiveions move toward the cathode.
-
7/28/2019 Analytic Techniques Sc
80/95
Separation method based on differentinteraction of the specimen compounds withthe mobile phase and the stationary phase ascompound travel through a support medium.
RETENTION TIME(tR) Time it takes a compound to elute
-
7/28/2019 Analytic Techniques Sc
81/95
MOBILE PHASE Gas or liquid
Carries the complex mixture (sample)
STATIONARY PHASE Solid or liquid
Through which the mobile phase flows COLUMN Holding the stationary phase
SEPARATED COMPONENTS eluate
-
7/28/2019 Analytic Techniques Sc
82/95
Adsorption AKA liquid-solid chromatography Based on the competition between the sample and the mobile phase
for absorption sites on the solid stationary phase Molecules that are most soluble in the mobile phase , move fastest
Partition AKA liquid-liquid chromatography Separation of solute based on relative solubility in an organic
(nonpolar) solvent and aqueous(polar) solvent
Normal phase mobile solvent is less polar than the stationary solvent Reverse phase mobile solvent is more polar
Applicable to any substance that maybe distributed between twoliquid
Works best in nonionic compound
-
7/28/2019 Analytic Techniques Sc
83/95
Steric exclusion
Variation of liquid-solid chromatography
Separate molecules on the basis of the size and shape
Ion exchange
Solute mixture are separated by virtue of themagnitude and charge of ionic species.
Stationary phase is a resin consisting of largepolymers of substituted benzene, silicates or cellulosederivatives, with charge functional groups
-
7/28/2019 Analytic Techniques Sc
84/95
THINLAYER CHROMATOGRAPHY
Thin layer of sorbent , such as alumina , silica gel,
cellulose or cross linked dextran, is uniformly
coated on a glass or plastic plate
Most commonly used as semiquantitative
screening test
-
7/28/2019 Analytic Techniques Sc
85/95
Uses pressure for fast separations, controlledtemperature, in-line detectors and gradientelution techniques
PUMPS Forces the mobile phase through the column at a
much greater velocity than that accomplished bygravity-flow columns and includes pneumatic,syringe, reciprocating or hydraulic amplifier pumps.
MECHANICAL RECIPROCATING PUMP
Multihead pump with two or more reciprocating pistons.
-
7/28/2019 Analytic Techniques Sc
86/95
COLUMNS
Long stainless steel where the stationary phase is
packed
SILICA GEL most common material used forpacking
Stable and can be used in different ways.
Reverse-phase HPLC Stationary phase is nonpolar molecules(octadecyl C-18
hydrocardbon) bonded to silica gel particles.
-
7/28/2019 Analytic Techniques Sc
87/95
SAMPLE INJECTORS Small syringe Loop injector best and widely used
High reproducibility and used at high temperature.
DETECTORS Monitor the eluate as it leaves the column and produce an
electronic signal proportional to the concentration of eachseparated component
RECORDERS
Used to record detector signal versus the time the mobilephase passed through the instrument, starting from thetime of injection.
CHROMATOGRAM -
-
7/28/2019 Analytic Techniques Sc
88/95
Separate mixture of compounds hat arevolatile or can be made volatile.
GAS-SOLID CHROMATOGRAPHY (GSC)
solid stationary phase
GAS-LIQUID CHROMATOGRAPHY (GLC)
With non volatile liquid stationary phase.
Commonly used in clinical laboratories.
-
7/28/2019 Analytic Techniques Sc
89/95
-
7/28/2019 Analytic Techniques Sc
90/95
ADVANTAGE
Increase the number of tests to be performed in a
given period
Minimizes variation of result from onelaboratorian to another
Eliminates the potential error in manual analyses
such as pipetting, calculation and transcript ofresult
-
7/28/2019 Analytic Techniques Sc
91/95
CONTINUOUS FLOW ANALYZER Liquids are pump through a system of continuous tubing
Samples flow through a common reaction vessel orpathway.
Air bubbles at regular intervals serves as separating andcleaning media
Mixture of sample and reagent takes place using a glasscoil inserted into flow path
A heating bath maintains the required temperature ofthe reaction to allow complete color development-reaction rate is controlled by temperature
Ex Simultaneous multiple analyzer (SMA), technicon
-
7/28/2019 Analytic Techniques Sc
92/95
CENTRIFUGAL ANALYZER Use the force generated by centrifugation to transfer
specimen and reagents
Liquids are placed in separate cuvets for measurement atthe perimeter of a spinning rotor (1000rpm)
It uses acceleration and deceleration of the rotor totransfer the reagents and sample from one chamber toanother
For mixing, centrifugal force or rotor is utilized orbubbling of air
Major advantage: Batch analysis
Ex Cobas- Bio (Roche), IL Monarch
-
7/28/2019 Analytic Techniques Sc
93/95
DISCRETE ANALYZER Most popular and versatile analyzer Employs variety of syringe and pipettes to aspirate and dispense sample and
reagents Positive liquid-displacement pipettes are used for sampling
Capable of running multiple-test-one-sample at-a-time. Each sample-reagent mixture handles separately in its own vessel Has Random access capability that allows STAT samples to be easily
accessed. For mixing, magnetic driven teflon stirring bar located in the bottom of the
reaction chamber is used in Beckman ASTRA For dry slide technology (reflectance photometry, the spreading layer permits
a rapid uniform spreading layer over the reagent layer. REFLECTANCE PHOTOMETRY- measure the light reflected from solid surfaces.
EX: Vitros, dimension Dade, Beckman ASTRA System, Hitachi, Bayer Advia Roche CobasIntegra 800, Roche anlaysis P module, ACA star (Dade)
-
7/28/2019 Analytic Techniques Sc
94/95
BATCH TESTING All samples are loaded at the same time, and singles test is conducted
on each sample PARALLEL TESTING
More than one test is analyzed concurrently on a given clinical
specimen RANDOM ACCESS TESTING
Any test can be performed on any sample in any sequence SEQUENTIAL TESTING
Multiple tests analyzed one after another on a given specimen OPEN REAGENT SYSTEM
A system other than manufacturers reagent can be utilized formeasurement
CLOSE REAGENT SYSTEM A system where the operator can only used the manufacturer
reagents.
-
7/28/2019 Analytic Techniques Sc
95/95
DONT FORGET TO READ YOUR BOOKS!
KEEP CALMANDPRAY
GODBLESS! =)