analyzer-204-v.pdf

2
The ability to analytically quantify highly reactive volatile organi c compounds (VOCs) in water is greatly enhanced with the Sample Transfer Stripper TM  (Patented Membrane Technology) a nd the solid state sensor offered by Analytical Systems International Keco (ASI KECO). The analyzer system utilizes many patented and exclusive features only available from ASI KECO. The principle of operation requires a consistent free flowing liquid sample into a Sample Transfer Stripper TM  unit that separates the liquids from the VOCs. A hydrocarbon free, consist ently flowing air carrier then sweeps the VOCs in a gaseous form into the solid state sensor for quantitative analysis in ppb w , ppm w , or up to saturation levels. An optional patented liquid validation/calibr ation by PermTube TM  is available. Thi s unattended on-line analys is is economically superior to inaccurate laboratory analyzers where unstable grab samples result in VOCs deterioration that produce analytical errors. Downtime is reduced to less than 10% of the El Paso, TOC or Sparger methods and the economic pay out is increased because of reduced product loss. In one case, 12,000 pounds of product loss per day was quickly identified by this proven and accurate on-line analytical method. TCEQ letters have designated the Model 204 as  “Best Available Control T echnology” and “Equivalency” for continuous monitoring of cooling tower applications. Applications include analysis of VOCs and oil in water for cooling towe rs, heat exchangers, holding ponds, run off water, and waste water. Model 204-V HYDRO CA RB ON (VOCs ) IN WATER ANALYZER CONTINUOUS ON-LINE PROCESS Continuous On-L ine Ana ly sis • 20+ Years Experience Wide ra ng es: 1 ppb w , ppm w and saturation Liquid V alidation by PermTube 4- 20mA & RS-232/ 485 No Normal Calibrations Required Proven On-Line Reliabilit y TCEQ “Best Available Control T echnology  Pre fer red V endor by Sh ell & Saudi Aramco Features - Patented & Exclusive De scrip tion & Principle of Operation (281) 516-3950  FAX (281) 351-8925  www.asikeco.com  Sales@a sikeco.com Wastewater • Heat Exchanger • Cooli ng Water • Holdin g Ponds • Runof f Water • Plant Efluent  Awarded Shell Global & Saudi Aramco Preferred Vendor TCEQ “Best Available Control Technology”

Upload: luis-cervantes-barradas

Post on 01-Nov-2015

1 views

Category:

Documents


0 download

TRANSCRIPT

  • The ability to analytically quantify highly reactive volatile organic compounds (VOCs) in water is greatly enhancedwith the Sample Transfer StripperTM (Patented Membrane Technology) and the solid state sensor offered by AnalyticalSystems International Keco (ASI KECO). The analyzer system utilizes many patented and exclusive features onlyavailable from ASI KECO.

    The principle of operation requires a consistent free flowing liquid sample into a Sample Transfer StripperTM unit thatseparates the liquids from the VOCs. A hydrocarbon free, consistently flowing air carrier then sweeps the VOCs in agaseous form into the solid state sensor for quantitative analysis in ppb w, ppm w, or up to saturation levels. Anoptional patented liquid validation/calibration by PermTubeTM is available. This unattended on-line analysis iseconomically superior to inaccurate laboratory analyzers where unstable grab samples result in VOCs deteriorationthat produce analytical errors. Downtime is reduced to less than 10% of the El Paso, TOC or Sparger methodsand the economic pay out is increased because of reduced product loss. In one case, 12,000 pounds of productloss per day was quickly identified by this proven and accurate on-line analytical method.

    TCEQ letters have designated the Model 204 as Best Available Control Technology and Equivalency forcontinuous monitoring of cooling tower applications.

    Applications include analysis of VOCs and oil in water for cooling towers, heat exchangers, holding ponds, run offwater, and waste water.

    Model 204-V

    HYDROCARBON (VOCs) IN WATER ANALYZERCONTINUOUS ON-LINE PROCESS

    Continuous On-Line Analysis 20+ Years Experience Wide ranges: 1 ppb w, ppm w and

    saturation Liquid Validation by PermTube 4-20mA & RS-232/485 No Normal Calibrations Required Proven On-Line Reliability TCEQ Best Available Control Technology Preferred Vendor by Shell & Saudi Aramco

    Features - Patented & Exclusive

    Description & Principle of Operation

    (281) 516-3950 FAX (281) 351-8925 www.asikeco.com [email protected]

    Wastewater Heat Exchanger Cooling Water Holding Ponds Runoff Water Plant Efluent

    Awarded Shell Global & Saudi AramcoPreferred Vendor

    TCEQ Best Available Control Technology

  • MODEL 204-V SPECIFICATIONS

    Analytical Systems Intl reserves the right to make revisions and or changes without notice. Analytical Systems Intl shall not be liable for any errorsor omissions made or any damages resulting from the use of the analyzer, information, or the manual. 03095 1995 Printed in the United States ofAmerica.

    Analyzer Quote Request Form:Analytical Systems International provides design and application engineering assistance for the Users analyzerrequirements. For a quotation, please complete the ASI Analyzer Quote Request Form atwww.ASIWebPage.com. For a quote request form, see main web page top right orange button.

    POWER INPUT:110 VAC 50/60 Hz @ 6 Amps.220 VAC 50/60Hz (optional)

    TEMPERATURE:Oven Setting with Automatic ControlOperating: 5C to 50CStorage: 0C to 70CPPB W Range Requires Temp. Control Building

    PERFORMANCE (APPLICATION DEPENDENT):Range: Customer Specified PPB W thru PPM W to SaturationPrimary Chemical Component Customer SpecifiedResolution: 1 PPBAccuracy + 2% of Full ScaleRepeatability: + 2% of Full ScaleLinearity + 1% of Full Scale

    ACCESSORY OPTIONS:Primary Filter, One Liter/min. SweepCabinet Heater & ThermostatSpecial Sampling SystemsPatented Liquid Validation System by Perm TubeTMAir Purifier For Hydrocarbon Free Carrier

    MEMORY / COMPUTER INTERFACE:RS-232/485 Serial Port Communication

    DISPLAY:Alpha Numeric Large LCD

    ANALOG OUTPUT SIGNAL:4-20ma Isolated (optional)4-20ma Linearized (standard)

    SAMPLE INPUT:Pressure Regulator & Gauge Provided 10 PSIG Minimum @ 20 ml/min. Sample FlowTemperature Control Automatic

    UTILITY SUPPLY:Carrier: Zero Air or Breathing Quality (Hydrocarbon Free) @ 200 ml/min.Calibration Supply Chemical

    ALARM CONTACTS: (optional)High and Low Concentration RelayDiagnostic Alarms for Loss of Sample Flow, Air Flow, Purge Flow & Power Loss

    AREA CLASSIFICATIONS:General Purpose (standard)Type Z Purge Class I Div. II, Gr. B,C,D (optional)Type X Purge Class I Div. I, GR. B,C,D (optional)

    CABINET DIMENSIONS & WEIGHT:Stainless Steel NEMA 4X 36" Hx24" Wx10" D(90cm Hx60cm Wx26cm D) 200 lbs. (91 kgs)

    DIGITALDISPLAY

    4-20 MAOUTPUT

    & ALARMS

    FLOWDIAGRAM

    PRINTER

    CARRIER AIRFLOW CONTROLLER

    SAMPLEINPUT

    SECONDARYFILTER

    SAMPLE SWEEP

    INTERFACERS-232/485

    3 4

    1 2

    SAMPLETRANSFERSTRIPPER

    (STS)

    PRE-HEATEXCHANGER

    WASTE SAMPLE20 PSIG MAXIMUM CARRIER/SAMPLE DIFFERENTIAL PRESSURE

    AIR INPUTHYRDROCARBON

    FREE & BREATHINGQUALITY

    FLOW METER

    THERMOSTATEDOVEN

    REGULATOR& GAUGE

    SOLIDSTATE

    SENSOR

    VENTNO BACKPRESSURE

    NEEDLEVALVE

    OPTIONAL LIQUID VALIDATIONSYSTEM NOT SHOWN

    ELECTRONICMODULE