analyzers: monitoring free chlorine with improved accuracy · analyzers: monitoring free chlorine...
TRANSCRIPT
![Page 1: Analyzers: Monitoring Free Chlorine with Improved Accuracy · Analyzers: Monitoring Free Chlorine with Improved Accuracy Gander Safe Drinking Water Seminar - March 2019. Agenda •](https://reader031.vdocument.in/reader031/viewer/2022012003/60ab0a8aa0991a021f733581/html5/thumbnails/1.jpg)
Analyzers: Monitoring Free Chlorine with Improved Accuracy
Gander Safe Drinking Water Seminar - March 2019
![Page 2: Analyzers: Monitoring Free Chlorine with Improved Accuracy · Analyzers: Monitoring Free Chlorine with Improved Accuracy Gander Safe Drinking Water Seminar - March 2019. Agenda •](https://reader031.vdocument.in/reader031/viewer/2022012003/60ab0a8aa0991a021f733581/html5/thumbnails/2.jpg)
Agenda• Purpose of an on‐line free chlorine Analyzer• Typical Installation• Analyzer Models Used in Newfoundland and Labrador• Typical Maintenance and Why you should do it• Questions/Wrap up
![Page 3: Analyzers: Monitoring Free Chlorine with Improved Accuracy · Analyzers: Monitoring Free Chlorine with Improved Accuracy Gander Safe Drinking Water Seminar - March 2019. Agenda •](https://reader031.vdocument.in/reader031/viewer/2022012003/60ab0a8aa0991a021f733581/html5/thumbnails/3.jpg)
Purpose of an on‐line Free Chlorine Analyzer
• Continuously monitor free‐chlorine at the outlet of the chlorine building
• Monitoring only• Monitoring and Control
• Send signal to chlorination system• Metering Pump• Gas Chlorinator
![Page 4: Analyzers: Monitoring Free Chlorine with Improved Accuracy · Analyzers: Monitoring Free Chlorine with Improved Accuracy Gander Safe Drinking Water Seminar - March 2019. Agenda •](https://reader031.vdocument.in/reader031/viewer/2022012003/60ab0a8aa0991a021f733581/html5/thumbnails/4.jpg)
Typical Installation
• Sample line connected downstream from Cl2 injection• Most time outside building, sometime within the building• Sample line: Isolation valve, Y‐Strainer, PRV, Pressure gauge
![Page 5: Analyzers: Monitoring Free Chlorine with Improved Accuracy · Analyzers: Monitoring Free Chlorine with Improved Accuracy Gander Safe Drinking Water Seminar - March 2019. Agenda •](https://reader031.vdocument.in/reader031/viewer/2022012003/60ab0a8aa0991a021f733581/html5/thumbnails/5.jpg)
Typical Installation
![Page 6: Analyzers: Monitoring Free Chlorine with Improved Accuracy · Analyzers: Monitoring Free Chlorine with Improved Accuracy Gander Safe Drinking Water Seminar - March 2019. Agenda •](https://reader031.vdocument.in/reader031/viewer/2022012003/60ab0a8aa0991a021f733581/html5/thumbnails/6.jpg)
Typical InstallationpH probe
Cl2 sensor
Press. gauge
PRV
Y‐Strainer
Shut‐Off Valve
![Page 7: Analyzers: Monitoring Free Chlorine with Improved Accuracy · Analyzers: Monitoring Free Chlorine with Improved Accuracy Gander Safe Drinking Water Seminar - March 2019. Agenda •](https://reader031.vdocument.in/reader031/viewer/2022012003/60ab0a8aa0991a021f733581/html5/thumbnails/7.jpg)
Analyzer Technologies
• 3‐electrode w/pH probe compensation• Membrane 3‐electrode w/pH probe compensation• 3‐electrodeWith pH Reagent• Amperometric
![Page 8: Analyzers: Monitoring Free Chlorine with Improved Accuracy · Analyzers: Monitoring Free Chlorine with Improved Accuracy Gander Safe Drinking Water Seminar - March 2019. Agenda •](https://reader031.vdocument.in/reader031/viewer/2022012003/60ab0a8aa0991a021f733581/html5/thumbnails/8.jpg)
Analyzer Technologies – Pros & ConsTechnology Pros Cons Comments
3‐electrode w/pH compensation
Faster sensingNo reagent – pH probeLittle operational costs
Calibration Clean/Replace pH probe
Replace Electrolyte
Electrolyte Expiry
Membrane 3‐electrodew/pH compensation
No Reagent – NO pH probe required Little operational costs
Slower Sensing pH compensation achieved by electrolyte gel in Chlorine sensor
3‐electrodew/pH Reagent
ReagentOn going operational costs
Reagent expiry
Wears out Requires
freeze protected shipments
Hazardous
![Page 9: Analyzers: Monitoring Free Chlorine with Improved Accuracy · Analyzers: Monitoring Free Chlorine with Improved Accuracy Gander Safe Drinking Water Seminar - March 2019. Agenda •](https://reader031.vdocument.in/reader031/viewer/2022012003/60ab0a8aa0991a021f733581/html5/thumbnails/9.jpg)
Analyzer Models Used in NL
• Evoqua Depolox 3 plus – 3‐electrode + pH Probe• Evoqua Micro 2000 – 3‐electrode + Reagent• Severn Trent – 3‐electrode + Reagent• Superior – Amperometric + Reagent
![Page 10: Analyzers: Monitoring Free Chlorine with Improved Accuracy · Analyzers: Monitoring Free Chlorine with Improved Accuracy Gander Safe Drinking Water Seminar - March 2019. Agenda •](https://reader031.vdocument.in/reader031/viewer/2022012003/60ab0a8aa0991a021f733581/html5/thumbnails/10.jpg)
MaintenanceAs a General Rule:• Most analysers drain sample water – make sure free flowing• Sample flow is critical i.e. 500ml/min
![Page 11: Analyzers: Monitoring Free Chlorine with Improved Accuracy · Analyzers: Monitoring Free Chlorine with Improved Accuracy Gander Safe Drinking Water Seminar - March 2019. Agenda •](https://reader031.vdocument.in/reader031/viewer/2022012003/60ab0a8aa0991a021f733581/html5/thumbnails/11.jpg)
Wallace & Tiernan Depolox 3 plus
• With pH probe • Calibration
• Cl2 – Span ‐ daily• Cl2 – Zero – Bi‐Monthly• pH – no set frequency
• Maintenance• Monthly: check Y‐strainer• Bi‐Monthly:
• Grit – Add half a cap• Top up Electrolyte regularly• Inspect membranes, replace
as required• 6‐month: remove Cl2 sensor,
rince, replace membranes, electrolyte and grit, calibrate, replace membrane if required
![Page 12: Analyzers: Monitoring Free Chlorine with Improved Accuracy · Analyzers: Monitoring Free Chlorine with Improved Accuracy Gander Safe Drinking Water Seminar - March 2019. Agenda •](https://reader031.vdocument.in/reader031/viewer/2022012003/60ab0a8aa0991a021f733581/html5/thumbnails/12.jpg)
Wallace & Tiernan Micro 2000• With Reagent• Calibration
• Chlorine calibration every 6 months• Maintenance
• Reagent (pH 4) use 1 gal every 10 weeks
• Add Grit as required• Check and top up electrolyte level every month
• Replace electrolyte every 12 months• Every 6‐12 months, replace…
• Impeller shaft, and shaft seal• Sample and reagent tubing• Clean electrodes with clean paper towel
![Page 13: Analyzers: Monitoring Free Chlorine with Improved Accuracy · Analyzers: Monitoring Free Chlorine with Improved Accuracy Gander Safe Drinking Water Seminar - March 2019. Agenda •](https://reader031.vdocument.in/reader031/viewer/2022012003/60ab0a8aa0991a021f733581/html5/thumbnails/13.jpg)
Superior WaterGuard
• With Reagent• Calibration
• Chlorine calibration every 6 months• Maintenance
• Clean pre‐filter every month• (2) Reagent 500ml bottles last 4‐8 weeks• Reagent pump Heads and Tubes Replacement every 12 months
![Page 14: Analyzers: Monitoring Free Chlorine with Improved Accuracy · Analyzers: Monitoring Free Chlorine with Improved Accuracy Gander Safe Drinking Water Seminar - March 2019. Agenda •](https://reader031.vdocument.in/reader031/viewer/2022012003/60ab0a8aa0991a021f733581/html5/thumbnails/14.jpg)
Severn Trent MicroChem 2
• Chlorine membrane probe CL4000• Calibration
• Check weekly and calibrated as needed
• Maintenance• Replace electrolyte• Every 12 months
• Replace membrane cap and fill with new electrolyte
• May require to clean the gold electrode with the blue abrasive paper
![Page 15: Analyzers: Monitoring Free Chlorine with Improved Accuracy · Analyzers: Monitoring Free Chlorine with Improved Accuracy Gander Safe Drinking Water Seminar - March 2019. Agenda •](https://reader031.vdocument.in/reader031/viewer/2022012003/60ab0a8aa0991a021f733581/html5/thumbnails/15.jpg)
Maintenance and Why You Should Do It
• Critical if used for control• Follow Manufacturers recommendations
• Regular calibration to ensure accuracy• Chlorine sensor more frequently than pH
• Top‐up/ replace chemicals: Electrolyte, Reagent• Clean sensors as recommended by manufacturers• Less costly to be preventive then to be sorry