Anesthesia Machine Pre-use Check Ohmeda Aestiva 3000 Morana Lasic, M.D. BWH Anesthesia Technology Elective Resident 2000 Under the supervision of James H Philip, M.D.

Upload: dangdiep

Post on 20-Aug-2018

232 views

Category:

Documents


3 download

TRANSCRIPT

Page 1: Anesthesia Machine Checkout Ohmeda Aestiva 3000etherweb.bwh.harvard.edu/education/PHILIP/aestivapreuse.pdf · Anesthesia Machine Pre-use Check Ohmeda Aestiva 3000 Morana Lasic, M.D

Anesthesia Machine Pre-use Check

Ohmeda Aestiva 3000

Morana Lasic, M.D.

BWH Anesthesia Technology Elective Resident 2000

Under the supervision of James H Philip, M.D.

Page 2: Anesthesia Machine Checkout Ohmeda Aestiva 3000etherweb.bwh.harvard.edu/education/PHILIP/aestivapreuse.pdf · Anesthesia Machine Pre-use Check Ohmeda Aestiva 3000 Morana Lasic, M.D
Page 3: Anesthesia Machine Checkout Ohmeda Aestiva 3000etherweb.bwh.harvard.edu/education/PHILIP/aestivapreuse.pdf · Anesthesia Machine Pre-use Check Ohmeda Aestiva 3000 Morana Lasic, M.D
Page 4: Anesthesia Machine Checkout Ohmeda Aestiva 3000etherweb.bwh.harvard.edu/education/PHILIP/aestivapreuse.pdf · Anesthesia Machine Pre-use Check Ohmeda Aestiva 3000 Morana Lasic, M.D

Anesthesia Machine Checkout

• The following checkout should be done

every day

Page 5: Anesthesia Machine Checkout Ohmeda Aestiva 3000etherweb.bwh.harvard.edu/education/PHILIP/aestivapreuse.pdf · Anesthesia Machine Pre-use Check Ohmeda Aestiva 3000 Morana Lasic, M.D

Verify Backup Ventilation

Equipment is Available and

Functioning

Page 6: Anesthesia Machine Checkout Ohmeda Aestiva 3000etherweb.bwh.harvard.edu/education/PHILIP/aestivapreuse.pdf · Anesthesia Machine Pre-use Check Ohmeda Aestiva 3000 Morana Lasic, M.D
Page 7: Anesthesia Machine Checkout Ohmeda Aestiva 3000etherweb.bwh.harvard.edu/education/PHILIP/aestivapreuse.pdf · Anesthesia Machine Pre-use Check Ohmeda Aestiva 3000 Morana Lasic, M.D

Check The High Pressure System

• Disconnect the pipeline supplies and close

all cylinder valves--alarms for low O2

supply pressure should sound

Page 8: Anesthesia Machine Checkout Ohmeda Aestiva 3000etherweb.bwh.harvard.edu/education/PHILIP/aestivapreuse.pdf · Anesthesia Machine Pre-use Check Ohmeda Aestiva 3000 Morana Lasic, M.D
Page 9: Anesthesia Machine Checkout Ohmeda Aestiva 3000etherweb.bwh.harvard.edu/education/PHILIP/aestivapreuse.pdf · Anesthesia Machine Pre-use Check Ohmeda Aestiva 3000 Morana Lasic, M.D

• Make sure that the cylinders are full--open

O2 valve and verify that the cylinder is at

least half full (1000 psig)

• Low O2 supply pressure alarm should

disappear

• Close the cylinder and once the alarm again

sounds reconnect the pipeline supplies

Page 10: Anesthesia Machine Checkout Ohmeda Aestiva 3000etherweb.bwh.harvard.edu/education/PHILIP/aestivapreuse.pdf · Anesthesia Machine Pre-use Check Ohmeda Aestiva 3000 Morana Lasic, M.D
Page 11: Anesthesia Machine Checkout Ohmeda Aestiva 3000etherweb.bwh.harvard.edu/education/PHILIP/aestivapreuse.pdf · Anesthesia Machine Pre-use Check Ohmeda Aestiva 3000 Morana Lasic, M.D

• Check that pipeline gauges read about 50

psig

Page 12: Anesthesia Machine Checkout Ohmeda Aestiva 3000etherweb.bwh.harvard.edu/education/PHILIP/aestivapreuse.pdf · Anesthesia Machine Pre-use Check Ohmeda Aestiva 3000 Morana Lasic, M.D
Page 13: Anesthesia Machine Checkout Ohmeda Aestiva 3000etherweb.bwh.harvard.edu/education/PHILIP/aestivapreuse.pdf · Anesthesia Machine Pre-use Check Ohmeda Aestiva 3000 Morana Lasic, M.D

Check The Low Pressure

Systems

• Close flow control valves and turn

vaporizers OFF

• Check fill level and tighten vaporizers’

filler caps

• Make sure that the vaporizers are connected

to an electrical outlet

• Attempt to turn ON more than one

vaporizer at a time

Page 14: Anesthesia Machine Checkout Ohmeda Aestiva 3000etherweb.bwh.harvard.edu/education/PHILIP/aestivapreuse.pdf · Anesthesia Machine Pre-use Check Ohmeda Aestiva 3000 Morana Lasic, M.D
Page 15: Anesthesia Machine Checkout Ohmeda Aestiva 3000etherweb.bwh.harvard.edu/education/PHILIP/aestivapreuse.pdf · Anesthesia Machine Pre-use Check Ohmeda Aestiva 3000 Morana Lasic, M.D
Page 16: Anesthesia Machine Checkout Ohmeda Aestiva 3000etherweb.bwh.harvard.edu/education/PHILIP/aestivapreuse.pdf · Anesthesia Machine Pre-use Check Ohmeda Aestiva 3000 Morana Lasic, M.D
Page 17: Anesthesia Machine Checkout Ohmeda Aestiva 3000etherweb.bwh.harvard.edu/education/PHILIP/aestivapreuse.pdf · Anesthesia Machine Pre-use Check Ohmeda Aestiva 3000 Morana Lasic, M.D

Perform Leak Check of Machine

Low Pressure System

• Make sure that the machine switch and flow

control valves are OFF

• Turn on the auxiliary common gas outlet

• Attach the bulb to it

• Compress and release bulb until it is empty-

-if the bulb inflates in 30 seconds or less--

there is a leak!

Page 18: Anesthesia Machine Checkout Ohmeda Aestiva 3000etherweb.bwh.harvard.edu/education/PHILIP/aestivapreuse.pdf · Anesthesia Machine Pre-use Check Ohmeda Aestiva 3000 Morana Lasic, M.D
Page 19: Anesthesia Machine Checkout Ohmeda Aestiva 3000etherweb.bwh.harvard.edu/education/PHILIP/aestivapreuse.pdf · Anesthesia Machine Pre-use Check Ohmeda Aestiva 3000 Morana Lasic, M.D
Page 20: Anesthesia Machine Checkout Ohmeda Aestiva 3000etherweb.bwh.harvard.edu/education/PHILIP/aestivapreuse.pdf · Anesthesia Machine Pre-use Check Ohmeda Aestiva 3000 Morana Lasic, M.D

Test Each Vaporizer For Low

Pressure Leaks

• Set the vaporizer to 1% and repeat the bulb

test as previously described

• The last vaporizer must be turned off after

the test is completed

Page 21: Anesthesia Machine Checkout Ohmeda Aestiva 3000etherweb.bwh.harvard.edu/education/PHILIP/aestivapreuse.pdf · Anesthesia Machine Pre-use Check Ohmeda Aestiva 3000 Morana Lasic, M.D

• Remember that agent mixtures stay in the

system after the low-pressure test so always

flush the system with O2 (1L/minute for

one minute is sufficient)

• After one minute turn the O2 to minimum

flow

Page 22: Anesthesia Machine Checkout Ohmeda Aestiva 3000etherweb.bwh.harvard.edu/education/PHILIP/aestivapreuse.pdf · Anesthesia Machine Pre-use Check Ohmeda Aestiva 3000 Morana Lasic, M.D

Turn The Machine ON!!

Page 23: Anesthesia Machine Checkout Ohmeda Aestiva 3000etherweb.bwh.harvard.edu/education/PHILIP/aestivapreuse.pdf · Anesthesia Machine Pre-use Check Ohmeda Aestiva 3000 Morana Lasic, M.D
Page 24: Anesthesia Machine Checkout Ohmeda Aestiva 3000etherweb.bwh.harvard.edu/education/PHILIP/aestivapreuse.pdf · Anesthesia Machine Pre-use Check Ohmeda Aestiva 3000 Morana Lasic, M.D

Test Flowmeters

• Test flow of all gases through their full

range making sure that they function

smoothly and that flowtubes are not

damaged

Page 25: Anesthesia Machine Checkout Ohmeda Aestiva 3000etherweb.bwh.harvard.edu/education/PHILIP/aestivapreuse.pdf · Anesthesia Machine Pre-use Check Ohmeda Aestiva 3000 Morana Lasic, M.D
Page 26: Anesthesia Machine Checkout Ohmeda Aestiva 3000etherweb.bwh.harvard.edu/education/PHILIP/aestivapreuse.pdf · Anesthesia Machine Pre-use Check Ohmeda Aestiva 3000 Morana Lasic, M.D

Test The N2O Flow Control

• In the minimum flow as your starting point,

slowly turn N2O flow control

counterclockwise and make sure that the O2

flow increases in 3:1 ratio (N2O:O2)

• Similarly, to test the O2 flow, start with

N2O at 9L and O2 at 3L and make sure that

as you decrease O2 flow, N2O flow

decreases as well

Page 27: Anesthesia Machine Checkout Ohmeda Aestiva 3000etherweb.bwh.harvard.edu/education/PHILIP/aestivapreuse.pdf · Anesthesia Machine Pre-use Check Ohmeda Aestiva 3000 Morana Lasic, M.D
Page 28: Anesthesia Machine Checkout Ohmeda Aestiva 3000etherweb.bwh.harvard.edu/education/PHILIP/aestivapreuse.pdf · Anesthesia Machine Pre-use Check Ohmeda Aestiva 3000 Morana Lasic, M.D

• Stop the O2 supply by disconnecting the

pipeline supply and making sure that the

cylinder O2 valve is closed

• Verify that the low O2 supply alarm sounds

and that N2O, Heliox if available, O2 flow

stop (LAST!), but air flow continues

• Reconnect the O2 supply

Page 29: Anesthesia Machine Checkout Ohmeda Aestiva 3000etherweb.bwh.harvard.edu/education/PHILIP/aestivapreuse.pdf · Anesthesia Machine Pre-use Check Ohmeda Aestiva 3000 Morana Lasic, M.D

Vaporizer Back Pressure Test

• Set O2 flow to 6 L/minute

• Increase the vaporizer concentration from 0

to 1% one click at a time

• The O2 flow must not decrease more than 1

L/minute through the complete range

Page 30: Anesthesia Machine Checkout Ohmeda Aestiva 3000etherweb.bwh.harvard.edu/education/PHILIP/aestivapreuse.pdf · Anesthesia Machine Pre-use Check Ohmeda Aestiva 3000 Morana Lasic, M.D
Page 31: Anesthesia Machine Checkout Ohmeda Aestiva 3000etherweb.bwh.harvard.edu/education/PHILIP/aestivapreuse.pdf · Anesthesia Machine Pre-use Check Ohmeda Aestiva 3000 Morana Lasic, M.D
Page 32: Anesthesia Machine Checkout Ohmeda Aestiva 3000etherweb.bwh.harvard.edu/education/PHILIP/aestivapreuse.pdf · Anesthesia Machine Pre-use Check Ohmeda Aestiva 3000 Morana Lasic, M.D

Power Failure Test

• Unplug the power source while the system

is ON and verify that the power failure

alarm comes ON

• Verify that the alarm cancels once the

power cable is reconnected

Page 33: Anesthesia Machine Checkout Ohmeda Aestiva 3000etherweb.bwh.harvard.edu/education/PHILIP/aestivapreuse.pdf · Anesthesia Machine Pre-use Check Ohmeda Aestiva 3000 Morana Lasic, M.D

Breathing System Test

• Check the breathing circuit thoroughly

• Verify that CO2 absorber is adequate

• Install accessories to be used during the

case (PEEP valve, humidifier, etc.)

• Press the drain button for 10 seconds to

eliminate moisture buildup

Page 34: Anesthesia Machine Checkout Ohmeda Aestiva 3000etherweb.bwh.harvard.edu/education/PHILIP/aestivapreuse.pdf · Anesthesia Machine Pre-use Check Ohmeda Aestiva 3000 Morana Lasic, M.D

Calibrate O2 Monitor

• Ensure monitor reads 21% in room air

• Ensure that low O2 alarm is functioning

• Reinstall O2 sensor and flush breathing

system with O2

• Verify that the monitor displays more than

90%

Page 35: Anesthesia Machine Checkout Ohmeda Aestiva 3000etherweb.bwh.harvard.edu/education/PHILIP/aestivapreuse.pdf · Anesthesia Machine Pre-use Check Ohmeda Aestiva 3000 Morana Lasic, M.D

Perform Leak Check of the

Breathing System

• Set all gas flows to minimum

• Close APL valve

• Occlude Y-piece

• Inflate the bag to 30 cm H2O with the O2

flush

• Verify that the pressure remains for at least

10 seconds and then open the APL valve

Page 36: Anesthesia Machine Checkout Ohmeda Aestiva 3000etherweb.bwh.harvard.edu/education/PHILIP/aestivapreuse.pdf · Anesthesia Machine Pre-use Check Ohmeda Aestiva 3000 Morana Lasic, M.D
Page 37: Anesthesia Machine Checkout Ohmeda Aestiva 3000etherweb.bwh.harvard.edu/education/PHILIP/aestivapreuse.pdf · Anesthesia Machine Pre-use Check Ohmeda Aestiva 3000 Morana Lasic, M.D
Page 38: Anesthesia Machine Checkout Ohmeda Aestiva 3000etherweb.bwh.harvard.edu/education/PHILIP/aestivapreuse.pdf · Anesthesia Machine Pre-use Check Ohmeda Aestiva 3000 Morana Lasic, M.D

Test the APL Valve

• Close the APL valve completely and

occlude the Y-piece

• Press the flush button and verify that the

pressure on the gauge is less than 70 cm

H2O

• Fully open the APL valve and make sure

that the pressure decreases to 0

• Press the flush button again and verify that

the pressure stays around 0

Page 39: Anesthesia Machine Checkout Ohmeda Aestiva 3000etherweb.bwh.harvard.edu/education/PHILIP/aestivapreuse.pdf · Anesthesia Machine Pre-use Check Ohmeda Aestiva 3000 Morana Lasic, M.D

Test Ventilation System

• Test the ventilator circuit for leaks by

setting the bag to Ventilator and decreasing

all flows to minimum

• Set the system switch to standby

• Close the Y-piece and fill the bellows with

O2 flush button

• There is a leak if the bellows fall more than

100ml/min

Page 40: Anesthesia Machine Checkout Ohmeda Aestiva 3000etherweb.bwh.harvard.edu/education/PHILIP/aestivapreuse.pdf · Anesthesia Machine Pre-use Check Ohmeda Aestiva 3000 Morana Lasic, M.D

• Place a test lung (breathing bag) on Y-piece

• Set appropriate Vent settings for the patient

• Fill bellows with O2 flush after turning all

other flows to minimum

• Turn Vent ON

• Verify that the bellows fill appropriately

and that the ventilator displays correct data

Page 41: Anesthesia Machine Checkout Ohmeda Aestiva 3000etherweb.bwh.harvard.edu/education/PHILIP/aestivapreuse.pdf · Anesthesia Machine Pre-use Check Ohmeda Aestiva 3000 Morana Lasic, M.D
Page 42: Anesthesia Machine Checkout Ohmeda Aestiva 3000etherweb.bwh.harvard.edu/education/PHILIP/aestivapreuse.pdf · Anesthesia Machine Pre-use Check Ohmeda Aestiva 3000 Morana Lasic, M.D

• Repeat above with fresh gas flow at 5L/min

making sure that the end expiratory pressure

is 0 cm H2O

Page 43: Anesthesia Machine Checkout Ohmeda Aestiva 3000etherweb.bwh.harvard.edu/education/PHILIP/aestivapreuse.pdf · Anesthesia Machine Pre-use Check Ohmeda Aestiva 3000 Morana Lasic, M.D
Page 44: Anesthesia Machine Checkout Ohmeda Aestiva 3000etherweb.bwh.harvard.edu/education/PHILIP/aestivapreuse.pdf · Anesthesia Machine Pre-use Check Ohmeda Aestiva 3000 Morana Lasic, M.D

• Make sure that unidirectional valves

function properly

• Switch to manual ventilation and verify

inflation and deflation of artificial lungs

• Remove the second bag from the Y-piece

Page 45: Anesthesia Machine Checkout Ohmeda Aestiva 3000etherweb.bwh.harvard.edu/education/PHILIP/aestivapreuse.pdf · Anesthesia Machine Pre-use Check Ohmeda Aestiva 3000 Morana Lasic, M.D
Page 46: Anesthesia Machine Checkout Ohmeda Aestiva 3000etherweb.bwh.harvard.edu/education/PHILIP/aestivapreuse.pdf · Anesthesia Machine Pre-use Check Ohmeda Aestiva 3000 Morana Lasic, M.D

Monitors

• Check and set alarm limits of all monitors--

capnometer, O2 analyzer, pulse oximeter,

spirometer, pressure monitor-high and low

airway pressure alarms

Page 47: Anesthesia Machine Checkout Ohmeda Aestiva 3000etherweb.bwh.harvard.edu/education/PHILIP/aestivapreuse.pdf · Anesthesia Machine Pre-use Check Ohmeda Aestiva 3000 Morana Lasic, M.D

Scavenging System

• Verify that there is a proper connection

between the scavenging system and BOTH

APL valve and ventilator relief valve

• Verify that the scavenging system is

connected to the wall vacuum

Page 48: Anesthesia Machine Checkout Ohmeda Aestiva 3000etherweb.bwh.harvard.edu/education/PHILIP/aestivapreuse.pdf · Anesthesia Machine Pre-use Check Ohmeda Aestiva 3000 Morana Lasic, M.D
Page 49: Anesthesia Machine Checkout Ohmeda Aestiva 3000etherweb.bwh.harvard.edu/education/PHILIP/aestivapreuse.pdf · Anesthesia Machine Pre-use Check Ohmeda Aestiva 3000 Morana Lasic, M.D
Page 50: Anesthesia Machine Checkout Ohmeda Aestiva 3000etherweb.bwh.harvard.edu/education/PHILIP/aestivapreuse.pdf · Anesthesia Machine Pre-use Check Ohmeda Aestiva 3000 Morana Lasic, M.D
Page 51: Anesthesia Machine Checkout Ohmeda Aestiva 3000etherweb.bwh.harvard.edu/education/PHILIP/aestivapreuse.pdf · Anesthesia Machine Pre-use Check Ohmeda Aestiva 3000 Morana Lasic, M.D
Page 52: Anesthesia Machine Checkout Ohmeda Aestiva 3000etherweb.bwh.harvard.edu/education/PHILIP/aestivapreuse.pdf · Anesthesia Machine Pre-use Check Ohmeda Aestiva 3000 Morana Lasic, M.D

• Fully open APL valve and occlude the Y-

piece

• With minimum O2 flow, observe the

scavenger reservoir bag collapsing

completely and the pressure gauge reading

0

• With the O2 flush activated, the scavenger

bag is to distend fully and the gauge should

read less than 10 cm H2O

Page 53: Anesthesia Machine Checkout Ohmeda Aestiva 3000etherweb.bwh.harvard.edu/education/PHILIP/aestivapreuse.pdf · Anesthesia Machine Pre-use Check Ohmeda Aestiva 3000 Morana Lasic, M.D

Check Final Status of Machine

• Vaporizers are OFF

• APL valve open

• Selector switch to Bag

• All flowmeters to minimum

• Verify adequate suction is available

• Breathing system is ready to use

• Equipment ready for intubation, IV

placement, emergency drugs ready

Page 54: Anesthesia Machine Checkout Ohmeda Aestiva 3000etherweb.bwh.harvard.edu/education/PHILIP/aestivapreuse.pdf · Anesthesia Machine Pre-use Check Ohmeda Aestiva 3000 Morana Lasic, M.D