aut proceedure

Upload: nandha-kumar

Post on 14-Apr-2018

220 views

Category:

Documents


0 download

TRANSCRIPT

  • 7/27/2019 Aut Proceedure

    1/42

    Stolt Offshore

    SANTOS LTD

    CASINO DEVELOPMENT OFFSHORE

    INSTALLATION

    Contract Number: 651008

    RTD Automated Ultrasonic Testing

    SAC-CTN-00251

    Originator : RTD Sub Contractor

    Checker : Grant Milne Senior Project Engineer

    Approver : Harry Johnson Project Technical Manager

    0.1 Issued For IDC May.31.05

    Version Reason for Issue Issue Date Prepared by Checked by Approved by Customer

    Approval

  • 7/27/2019 Aut Proceedure

    2/42

  • 7/27/2019 Aut Proceedure

    3/42

    Number : UT-05530 (Ref RTD-Steeltest Proc. RTDST-TP-UT-45)

    Revision : A AUTOMATIC ULTRASONIC TESTING PROCEDURE

    Date : 30.05.2005

    Page : 2 of 41

    Rntgen Technische Dienst bvCopyright RTD Delftweg 144, 3046 NC ROTTERDAM

    2003 Rntgen Technische Dienst bv

    This document has been produced by Rntgen Technische Dienst bv (RTD).

    Each copy is controlled and remains property of RTD.

    No part of this publication may be reproduced, stored in a retrieval system, or

    transmitted in any form or by any means without the prior permission of:

    Rntgen Technische Dienst bv

    Inquiries can be directed to;

    Rntgen Technische Dienst bvDelftweg 144, P.O box 10065

    3004 AB Rotterdam

    The Netherlands

    Phone : +31 10 2088208

    Fax : +31 10 4158022

    E-mail : [email protected]

  • 7/27/2019 Aut Proceedure

    4/42

    Number : UT-05530 (Ref RTD-Steeltest Proc. RTDST-TP-UT-45)

    Revision : A AUTOMATIC ULTRASONIC TESTING PROCEDURE

    Date : 30.05.2005

    Page : 3 of 41

    Rntgen Technische Dienst bvCopyright RTD Delftweg 144, 3046 NC ROTTERDAM

    Content

    1.0 Purpose

    2.0 Scope

    3.0 Reference Specifications

    4.0 RotoscanEquipment Description

    4.1 Multi-channel Ultrasonic Equipment

    4.2 Recording

    4.3 Scanner and Umbilical

    4.4 Calibration Set-up

    5.0 Rotoscan Set-Up Configuration

    5.1 Inspection Set-Up Configuration

    5.2 Probe Configuration5.3 Probe Frame Layout

    5.4 Calibration Block

    6.0 Equipment Calibration

    6.1 General

    6.2 Setting of Inspection Gates

    6.3 Sensitivity Settings

    6.4 Scanning Preparation & Surface Conditions

    7.0 Acoustic Coupling Fluid

    8.0 Examination

    8.1 Calibration Check

    8.2 Step by Step description of Rotoscan Examination

    8.3 Verification circumferential position

  • 7/27/2019 Aut Proceedure

    5/42

    Number : UT-05530 (Ref RTD-Steeltest Proc. RTDST-TP-UT-45)

    Revision : A AUTOMATIC ULTRASONIC TESTING PROCEDURE

    Date : 30.05.2005

    Page : 4 of 41

    Rntgen Technische Dienst bvCopyright RTD Delftweg 144, 3046 NC ROTTERDAM

    9.0 Interpretation of Results

    9.1 General

    9.2 Reportable Indications9.3 Sizing

    10.0 Acceptance/Rejection Criteria

    10.1 Repairs

    11.0 Reporting

    12.0 Personnel Qualifications

    12.1 The Rotoscan Operator

    12.2 The Manipulator Operator

    Attachment 1 : Inspection Set-up Configuration

    Attachment 2 : Transducer Selection

    Attachment 3 : Calibration Manufacturing Drawings

    Attachment 4 : Rotoscan Cover sheet Report

    Attachment 5 : Rotoscan Site Report

    Attachment 6 : Reportable indication sheet

    Attachment 7 : Gain settings calibration record

  • 7/27/2019 Aut Proceedure

    6/42

    Number : UT-05530 (Ref RTD-Steeltest Proc. RTDST-TP-UT-45)

    Revision : A AUTOMATIC ULTRASONIC TESTING PROCEDURE

    Date : 30.05.2005

    Page : 5 of 41

    Rntgen Technische Dienst bvCopyright RTD Delftweg 144, 3046 NC ROTTERDAM

    1.0 Purpose

    The purpose of this procedure is to determine Weld Acceptability during the

    pipeline construction of 12" Line-pipe of the "Santos Casino Pipeline Project"

    as required.

    2.0 Scope

    This procedure details methods for Automated Ultrasonic Weld Inspection(via the RTD Rotoscan Unit) of pipeline girth welds. The Rotoscan is an

    automated ultrasonic inspection system for pipeline girth welds, based upon

    the Pulse Echo Method, enhanced with mapping images and Time of

    Flight Diffraction.

    The weld is divided into a number of depth zones which are each inspected

    by one of an array of probes. The probes are arranged such that the entire

    weld is examined from both sides in a single circumferential scan. The

    ultrasonic information is transferred to a computer for data presentation and

    analysis.

    Inspection results are stored on harddisc ( incl. backup removable harddisc )

    and a hardcopy of each weld is printed in a compressed black & white

    image covering the pipe circumference. After completion of pipeline

    inspection, all data will be stored on CDROM or DVD and presented to

    client.

    This procedure is valid for following welds:

    Diameter : 12

    Wall thickness: 18.1 and 22.2 mm

    Weld bevel : J-prepMaterial : SML 450 IPD

  • 7/27/2019 Aut Proceedure

    7/42

    Number : UT-05530 (Ref RTD-Steeltest Proc. RTDST-TP-UT-45)

    Revision : A AUTOMATIC ULTRASONIC TESTING PROCEDURE

    Date : 30.05.2005

    Page : 6 of 41

    Rntgen Technische Dienst bvCopyright RTD Delftweg 144, 3046 NC ROTTERDAM

    3.0 Reference Specifications

    AS 2885.2 2002 Pipelines Gas and Liquid Petroleum Part 2 WeldingDNV-OS-F101- 2000 Submarine Pipeline Systems Offshore StandardSNT TC 1A - 1996 Society of Non destructive Testing Technical CouncilEN 473 European NormAS ISO/IEC 17025 1999 General requirements for the competence of testing and

    calibration laboratoriesRTD CP 31105 Rev 4 Calibration check procedure for ultrasonic multi-channel

    equipmentRTD CP 31108 Rev 1 Control Procedure of calibration blocks for Rotoscan

    systemUT-05531 latest Rev Manual UT procedure for weld repairs

    A-512 Rev 3 RTD Written Practise

    4.0 Rotoscan Equipment Description

    4.1 Multichannel ultrasonic equipment

    A multichannel ultrasonic flaw detector shall be used, type Rotoscan,

    which provides an adequate number of inspection channels

    (referred to as sequences) to ensure a complete volumetric examination

    of the weld through thickness in one circumferential scan.

    Each inspection channel will provide a linear A scan presentation,

    which can individually selected.

    The system meets the following minimum requirements:

    4.1.1 Global parameters

    - Number of Sequences : 32

    - Main Gain : 0 40 dB

    - Digitiser Sampling Frequency : 50 MHz

    - Overall System P.R.F : 4 kHz

  • 7/27/2019 Aut Proceedure

    8/42

    Number : UT-05530 (Ref RTD-Steeltest Proc. RTDST-TP-UT-45)

    Revision : A AUTOMATIC ULTRASONIC TESTING PROCEDURE

    Date : 30.05.2005

    Page : 7 of 41

    Rntgen Technische Dienst bvCopyright RTD Delftweg 144, 3046 NC ROTTERDAM

    4.1.2 Gate Parameters

    In each sequence, the following parameters shall be selectable:

    - Transmitter and Receiver channel 1- 32

    - Gain 0 - 40 dB ( 2dB increments )

    - Gate start and length. ( min increment 0,1mm )

    - Delay ( mm )

    - Threshold level 5% - 100% Full Screen Height ( FSH ).

    - Recordable output selection between; Amplitude and/or Transit

    distance, Mapping, TOFD and Coupling.- Selection between First / Highest amplitude detection within gated

    section

    4.1.3 General

    - Electronic noise shall be lower than acoustical noise in all

    sequences, for the probes and sensitivities to be used during the

    inspection.

    - The signal to noise ratio of each sequence during examination

    shall be at least 20 dB for shear waves probes.- The signal to noise ratio for TOFD probes, measured on the 2mm

    calibration block notch in the root, shall be at least 12 dB

    4.2 Recording

    Inspection results are presented on-line on a 20" colour monitor.

    The screen layout and number of used sequences are configured to present

    the thickness of the weld to be examined. Distance delays are applied to

    the output channels to compensate for variable transducer circumferential

    positions. The presentation shows the following information;

    - Amplitude and relevant transit distance (superimposed )

    - Mapping image sections

    - Go-Nogo image section

    - Coupling check section

    - TOFD image section

    - Circumferential position information- Header section with possibility to present job specific information- Time and date of inspection

    The raw data of each inspected weld will be stored on a hard disc.

    In addition a back-up will be made on an removable hard disc.Inspection files can be reviewed at any time.

    The inspection results will be printed in a compressed black-white image

    of the weld circumference.

  • 7/27/2019 Aut Proceedure

    9/42

    Number : UT-05530 (Ref RTD-Steeltest Proc. RTDST-TP-UT-45)

    Revision : A AUTOMATIC ULTRASONIC TESTING PROCEDURE

    Date : 30.05.2005

    Page : 8 of 41

    Rntgen Technische Dienst bvCopyright RTD Delftweg 144, 3046 NC ROTTERDAM

    4.3 Scanner and Umbilical

    The scanner to be used shall meet the following requirements:

    - Automated operation during examination, however manual

    operation may be permitted as detailed in section 8.2 part F .

    - Easy and quick mounting to the guideband.

    - Capable of taking up a minimum of 10 individually spring loaded

    ultrasonic probes, of which the distances to the weld centreline

    can individually be set.- Coupling controllable per probe, or per set.

    - 15-45 metres of cable length.

    - The scanner is equipped with an encoder for accurate

    measurement of the scanner position within the pipeline

    circumference.

    - Minimum speed of 40 mm/sec

    - Maximum speed of 80 mm/sec

    4.4 Calibration Set-up

    The calibration Set-up consists of a pipe piece, being a part of thin wall 12" pipe.

    On the support a guiding band is mounted at the same distance from the

    weld centreline as the guidance band on the weld to be tested.

    In the pipe piece a cut-out area is present where a calibration block with

    the proper wall thickness as pipe under test, is mounted which is used

    to monitor the ongoing system performance.

    The calibration block(s) to be used shall be made from original piece of pipe,

    without coating and no weld included.

    5.0 Rotoscan Set-up Configuration

    5.1 Inspection Set-up Configuration

    The inspection set-up is visualised in Attachment 1.It shows the selected

    probe angle(s) for the selected pulse echo, tandem and TOFD technique

    within the zones to be examined within the given weldbevel

    configuration.

  • 7/27/2019 Aut Proceedure

    10/42

    Number : UT-05530 (Ref RTD-Steeltest Proc. RTDST-TP-UT-45)

    Revision : A AUTOMATIC ULTRASONIC TESTING PROCEDURE

    Date : 30.05.2005

    Page : 9 of 41

    Rntgen Technische Dienst bvCopyright RTD Delftweg 144, 3046 NC ROTTERDAM

    5.2 Probe Configuration

    In Attachment 2 the selected probes with their parameters are listed which

    are used for the inspection. All probes have an unique number for QA/QC

    purpose relating to the manufacturing process and tests within the RTD

    probe manufacturing department. The numbers of probes used are part of

    the inspection parameter file within the Rotoscan inspection program.

    5.3 Probe Frame Layout

    The probes used for inspection are positioned within a probe frame.

    5.4 Calibration block

    The calibration block shall be of matching wall thickness of the item

    under test and contain artificial defects to install the inspection

    sensitivity. The block shall be designed with sufficient surface area to

    enable the transducer array to scan the target areas in one single pass.

    Reflectors for the pulse echo technique shall be flat-bottom holes and

    notches, which are positioned at the weldbevel.

    TOFD notches are used to identify the system capability to detect surface

    breaking indications. The calibration block shall be manufactured and

    tested according to procedure CP-31108. The calibration block shall be

    identified with a hard stamped unique serial number.

    Attachment 3 shows the calibration manufacturing drawings.

    Notes:

    * Through-holes 2mm in both ends of calibration block at centre line.These reflectors are to locate weld centre position only.

    * Notches, are to indicate the weldbevel position.

    * TOFD notches are used to identify surface breaking indications.

    6.0 Equipment Calibration

    6.1 General

    The ultrasonic equipment shall be electronically calibrated at least once per

    3- month period according calibration procedure to CP-31105 latest revision.

  • 7/27/2019 Aut Proceedure

    11/42

    Number : UT-05530 (Ref RTD-Steeltest Proc. RTDST-TP-UT-45)

    Revision : A AUTOMATIC ULTRASONIC TESTING PROCEDURE

    Date : 30.05.2005

    Page : 10 of 41

    Rntgen Technische Dienst bvCopyright RTD Delftweg 144, 3046 NC ROTTERDAM

    6.2 Setting of Inspection Gates

    - Pulse echo and Tandem channels

    With each transducer positioned for the peak signal response from the

    calibration reflector the detection gates are to be set. The gate shall start

    2-8 mm ( allowance for width of heat affected zone ) before the

    theoretical weld bevel preparation. The gate ends shall be after the

    theoretical weld centreline. All gates will be programmed to record

    amplitude and/or transit distance information. The gate length of thetransit distance in the root channel will be extended to enable root

    penetration registration and for the amplitude sequence the gate shall

    end at the position of the root TOFD notch.

    - Mapping channels

    The mapping gates in the body of the weld starts 2-8 mm

    (allowance for width of heat affected zone) before the theoretical weld

    bevel preparation. The gate length will be extended to enable cap

    reinforcement registration. The mapping gates in the root will be set

    identical to the pulse echo transit distance channels to enable the

    registration of the root penetration.

    - TOFD channel

    The TOFD gate start will be set 3 mm before the arrival of the lateral

    wave and should extend up to and including the first back-wall echo, to

    achieve full cover of wall thickness.

    Note: The gate settings may be altered if geometry indications dictate.If gate settings are changed the client should be informed.

    6.3 Sensitivity Settings

    With each transducer positioned for the peak signal response from the

    calibration reflector ( flat bottom holes ) the detection gates are to be set. In this

    position the probe holder is fixed to the probe frame. The equipment sensitivity

    ( echo amplitude ) for all inspection channels shall be set at 80% Full Screen

    Height.

    Calibration of surface breaking zones to be done using the 1 mm notches, for

    the embedded zones a 3mm FBH.

  • 7/27/2019 Aut Proceedure

    12/42

    Number : UT-05530 (Ref RTD-Steeltest Proc. RTDST-TP-UT-45)

    Revision : A AUTOMATIC ULTRASONIC TESTING PROCEDURE

    Date : 30.05.2005

    Page : 11 of 41

    Rntgen Technische Dienst bvCopyright RTD Delftweg 144, 3046 NC ROTTERDAM

    Mapping channels in the body of the weld will be used to detect the presents of

    porosity , and in addition to identify the position of the weld cap reinforcement

    for pattern recognition purpose. The sensitivity setting is on a 1.5mm FBH at

    80% FSH + 6 dB but should not be so great as to cause interfering electrical or

    geometric noise that could be misinterpreted.

    Mapping channels in the root will be used to identify the position of the root

    penetration for pattern recognition purpose. The sensitivity as a minimum is

    equal to the related pulse echo channels, increased with additional 8dB to

    ensure proper detection but should not be so great as to cause interferingelectrical or geometric noise that could be misinterpreted.

    The lateral wave (RF signal) of the TOFD channel sensitivity is set at 40% FSH.

    The probe distance for TOFD should be calculated at 2/3 of the wall thickness

    (nominal beam angle) 10mm.

    6.4 Scanning Preparation & Surface Condition

    The scanning area shall be free of weld spatters and other irregularities which

    may interfere with the inspection.

    Adjacent to one weld side, prior to welding, over the complete circumference

    at a fixed distance from the nose of the weld preparation, a scribeline shall be

    available for guidance band position adjustment. The distance from this scribe-

    line to the weld centre-line should be 40mm with an accuracy of 1mm.

    Transition welds shall be machined, to match the thinnest wall thickness over a

    width of at least 150mm.

  • 7/27/2019 Aut Proceedure

    13/42

    Number : UT-05530 (Ref RTD-Steeltest Proc. RTDST-TP-UT-45)

    Revision : A AUTOMATIC ULTRASONIC TESTING PROCEDURE

    Date : 30.05.2005

    Page : 12 of 41

    Rntgen Technische Dienst bvCopyright RTD Delftweg 144, 3046 NC ROTTERDAM

    7.0 Acoustic Coupling Fluid

    - Tap water shall be used as acoustic coupling fluid, to be able to test

    welds in the range up to 95C.

    - The coupling is supplied between probe contact surface and the

    pipe via supply channels incorporated in the probe housing.

    - To enable good acoustical contact between the ultrasonic probes

    and the pipe surface, conditions as mentioned under 6.4 apply.

    8.0 Examination

    8.1 Calibration check

    During the first 10 welds a calibration scan shall be made for every inspected

    weld. Thereafter the frequency of calibration is reduced to a minimum of one

    calibration every five welds. The calibration scan shall be performed every

    weld during slow-lay of the pipeline. Slow-lay is defined as one weld every

    two hours

    The Gain settings calibration record (attachment 7) shall be filled out twiceper shift after being done for the first 10 welds.

    A re-calibration shall be carried out if;

    * The calibration of an inspection function differs more than - 3 or + 2 dB

    from the previous calibration.

    * The gate settings need to be adjusted when they deviate more than +/-

    0.25 mm with the previous calibration.

    * After a weld repair. The print-out of the calibration run shall become part

    of the repair.

    After equipment breakdown If there is any reason to assume that a mechanical impact or other event

    occur.

    In case the calibration differs from the initial setting, outside the given

    tolerances, the applicable probe(s) and coupling shall be checked. If the

    calibration has to be changed, the welds before this calibration up to the previous

    calibration will be re-evaluated taking the differences in the gain and/or gate

    adjustment into account.

    The result of each calibration scan shall be printed, tabulated and stored onharddisc.

  • 7/27/2019 Aut Proceedure

    14/42

    Number : UT-05530 (Ref RTD-Steeltest Proc. RTDST-TP-UT-45)

    Revision : A AUTOMATIC ULTRASONIC TESTING PROCEDURE

    Date : 30.05.2005

    Page : 13 of 41

    Rntgen Technische Dienst bvCopyright RTD Delftweg 144, 3046 NC ROTTERDAM

    8.2 Step by Step Description of a Rotoscan Examination

    Before starting the step by step description the UT-system is already calibrated

    and a calibration scan with a record has been made.

    a. Adjust guiding band on the pipe to a distance of 110mm from the marks(scribe-line, which is 40mm from the weld centre-line) and parallel to

    the weld. The accuracy of the band setting should be 1mm.

    b. Scanner Operator will mark zero-point on the 12 o clock position ofthe pipe.

    c. Switch on coupling supply .

    d. Locate the scanner over the top of the pipe and make sure that all

    probes are past the zero point and are in proper contact with the pipe

    surface.

    e. Attach the scanner to the band.

    As long as cooling water is vaporising ( surface temperature >100)

    scanner shall not be attached to the band..

    f. Switch on system and automatically the weld will be scanned in

    clockwise direction. In the event of a drive motor failure, the probe

    frame may be manually driven provided that the weld scan

    integrity is not compromised. This may be proven by an acceptable

    comparison of the last automated calibration scan to the first

    manual calibration scan. This is envisaged to be temporary

    measure and best efforts will be made to repair and re-install the

    automatic drive system.

    g. Systems and coupling supply will switch off automatically after a full

    scan with an overlap of minimum 5 cm.

    h. Check the inspection results on defect indications and proper coupling.

    Use the transit distance recording and mapping and images for

    validation of defect indications within the root area and proper

    alignment of the guiding band. In case of lack of coupling a second scan

    will be necessary.

    i. Remove the scanner from the band and attach it to the calibration block.

  • 7/27/2019 Aut Proceedure

    15/42

    Number : UT-05530 (Ref RTD-Steeltest Proc. RTDST-TP-UT-45)

    Revision : A AUTOMATIC ULTRASONIC TESTING PROCEDURE

    Date : 30.05.2005

    Page : 14 of 41

    Rntgen Technische Dienst bvCopyright RTD Delftweg 144, 3046 NC ROTTERDAM

    8.3 Verification circumferential position

    The accuracy of the circumference, measured by the system, should be checked

    by comparison of indications (defects or geometrical) in the overlap of the

    system. The tolerance is 10mm. This check should be done at least once

    every five welds

    9.0 Interpretation of results

    9.1 General

    With the transit distance measurements and with the information from the

    Mapping and TOFD channels visible on the result presentation, indications

    shall be judged whether they are from the weld geometry or defects.

    The coupling channels shall be checked for couplant loss. Action shall be taken

    if one of the following events occur.

    Couplant loss length is greater than the smallest allowable defect in theacceptance criteria.

    In case an indication is present in the TOFD channel in a position of

    couplant loss a re-scan shall be required. If couplant loss persists in thezone of the indication in the TOFD, the indication shall be validated by

    manual UT.

    In case an indication is present in the PE channel experiencing loss ofcouplant directly adjacent to the length where couplant was lost, a rescan

    shall be required.

    9.2 Reportable indications

    The inspection result should be evaluated and/or reported as follows ;

    To be agreed

    This shall be performed off-line.

    The report format is given in attachment 6

    This is intended to give a measure of the overall weld quality and allow

    feedback to the welding foreman.

  • 7/27/2019 Aut Proceedure

    16/42

    Number : UT-05530 (Ref RTD-Steeltest Proc. RTDST-TP-UT-45)

    Revision : A AUTOMATIC ULTRASONIC TESTING PROCEDURE

    Date : 30.05.2005

    Page : 15 of 41

    Rntgen Technische Dienst bvCopyright RTD Delftweg 144, 3046 NC ROTTERDAM

    9.3 Sizing

    9.3.1 Length Sizing

    The beginning and end of the defect shall be determined from the points

    where the signal falls below 40%FSH either side of the indication

    If a resolvable TOFD indication is seen without an associated pulse echo

    amplitude >40%FSH, then the defect length shall be determined for the

    resolvable part of the image. This is the only instance where length shall be

    measured by TOFD.

    For the length measurement of porosity mapping channel shall be used. The

    length of the porosity is defined as the start of the cluster to the end of the

    cluster in circumferential direction.

    9.3.2 Height Sizing

    The height of an indication shall be measured by PE (pulse/Echo). Sizing

    curves shall be used where 80%FSH = 2,5mm defect height for the

    embedded zones and 80% FSH= 1.2mm for the surface breaking zones.

    If the indication can be resolved by TOFD (clear separation between upper

    and lower diffraction tip) the measured defect height by TOFD shall

    apply.

    The methodology is summarised in the AUT Defect Height Evaluation Flow

    Chart (see page 17).

  • 7/27/2019 Aut Proceedure

    17/42

    Number : UT-05530 (Ref RTD-Steeltest Proc. RTDST-TP-UT-45)

    Revision : A AUTOMATIC ULTRASONIC TESTING PROCEDURE

    Date : 30.05.2005

    Page : 16 of 41

    Rntgen Technische Dienst bvCopyright RTD Delftweg 144, 3046 NC ROTTERDAM

    9.3.3 Sizing of porosity

    When white points (> 40%FSH) are present in the mapping channels.

    Porosity shall be evaluated using the mapping channel(s).

    The height of the porosity is defined as distance between start and end of

    the porosity along the sound axis divided by the cos (), where is the

    probe angle.

    For the height of the porosity see following table:

    Measured cluster size

    along the sound beam

    (mapping)

    in mm

    Reportable

    height

    in mm

    2 1.5

    3 2.3

    4 3.1

    5 3.8

    6 4.6

    7 5.4

    8 6.19 6.9

    10 7.7

    11 8.4

    12 9.2

    13 10

  • 7/27/2019 Aut Proceedure

    18/42

    Number : UT-05530 (Ref RTD-Steeltest Proc. RTDST-TP-UT-45)

    Revision : A AUTOMATIC ULTRASONIC TESTING PROCEDURE

    Date : 30.05.2005

    Page : 17 of 41

    Rntgen Technische Dienst bvCopyright RTD Delftweg 144, 3046 NC ROTTERDAM

    9.3.4 AUT DEFECT HEIGHT EVALUATION FLOW CHART

  • 7/27/2019 Aut Proceedure

    19/42

    Number : UT-05530 (Ref RTD-Steeltest Proc. RTDST-TP-UT-45)

    Revision : A AUTOMATIC ULTRASONIC TESTING PROCEDURE

    Date : 30.05.2005

    Page : 18 of 41

    Rntgen Technische Dienst bvCopyright RTD Delftweg 144, 3046 NC ROTTERDAM

  • 7/27/2019 Aut Proceedure

    20/42

  • 7/27/2019 Aut Proceedure

    21/42

    Number : UT-05530 (Ref RTD-Steeltest Proc. RTDST-TP-UT-45)

    Revision : A AUTOMATIC ULTRASONIC TESTING PROCEDURE

    Date : 30.05.2005

    Page : 20 of 41

    Rntgen Technische Dienst bvCopyright RTD Delftweg 144, 3046 NC ROTTERDAM

    11.0 Reporting

    At commencement of the project we shall issue a covering UT Report

    ( see attachment 4) with all parameters of the system that shall not change

    throughout the project. Then on a daily basis throughout the project, a

    Rotoscan site report will be issued for reporting of daily test results with

    the following information;

    - Project and RTD project number

    - Weld No;- Date;- Diameter;- Wall thickness;- Number of indications, location on circumference, length and

    location in depth;

    - Acceptable ( C-comply) or rejected (DNC- does not comply).

    For a sample of the Rotoscan site report, see attachment 5

    The following set-up for a filename will be used during pipelay.

    Example of filename for weld 1234.

    1234 Scan of mainline or double joint weld

    1234C Calibration scan performed before scanning weld 1234

    1234R1 Scan of weld 1234 after repair

    1234R1C Calibration scan performed before scanning weld 1234R1

    12.0 Personnel qualifications

    All NDE personnel must be able to speak the English language.

    12.1 The Rotoscan operator

    The ultrasonic lead operator performing the examination shall:

    - Be qualified in accordance to RTD written-practice A-512 which

    conforms to SNT-TC-1A 1996 and EN-473 (level II).

    - Has experience with multichannel UT-equipment;

    - Be trained and passed the RTD in-house testing for MechanisedUltrasonic Inspection.

    12.2 The Manipulator operator

    The manipulator operator, handling of scanner mechanism shall:

    - Be trained in using mechanised UT-manipulator to satisfaction of

    Rotoscan operator

  • 7/27/2019 Aut Proceedure

    22/42

    Number : UT-05530 (Ref RTD-Steeltest Proc. RTDST-TP-UT-45)

    Revision : A AUTOMATIC ULTRASONIC TESTING PROCEDURE

    Date : 30.05.2005

    Page : 21 of 41

    Rntgen Technische Dienst bvCopyright RTD Delftweg 144, 3046 NC ROTTERDAM

    Attachment 1 Inspection Set-up Configuration

  • 7/27/2019 Aut Proceedure

    23/42

    Number : UT-05530 (Ref RTD-Steeltest Proc. RTDST-TP-UT-45)

    Revision : A AUTOMATIC ULTRASONIC TESTING PROCEDURE

    Date : 30.05.2005

    Page : 22 of 41

    Rntgen Technische Dienst bvCopyright RTD Delftweg 144, 3046 NC ROTTERDAM

  • 7/27/2019 Aut Proceedure

    24/42

    Number : UT-05530 (Ref RTD-Steeltest Proc. RTDST-TP-UT-45)

    Revision : A AUTOMATIC ULTRASONIC TESTING PROCEDURE

    Date : 30.05.2005

    Page : 23 of 41

    Rntgen Technische Dienst bvCopyright RTD Delftweg 144, 3046 NC ROTTERDAM

    Attachment 2: Transducer Selection

    Example; Will be updated after arrival cal blocks

    Calibration Block 18.1 and 22.2mm (J-prep)

    Zone Reflectors

    Cal. block

    Frequency

    probe

    Angle Inspec.tech. Crystal Ident.

    Z1

    Notch 15x1x2

    Notch 15x2x1

    4 MHz

    6 MHz

    55

    65

    Pulse Echo

    TOFD

    10 x 13 mm

    6 mm

    E

    G

    Z2 FBH 3 mm 4 MHzTx 40

    Rx 46Tandem

    Tx 10 mm

    Rx 12 x 14 mmA

    Z3 FBH 3 mm 4 MhzTx 40

    Rx 46Tandem

    Tx 10 mm

    Rx 12 x 16 mmB

    Z4 FBH 3 mm 4 MHzTx 40

    Rx 46Tandem

    Tx 11 mm

    Rx 12 mmC

    Z5 FBH 2 mm 4 MHz 50 Pulse Echo 13 x 16 mm D

    Z6Notch 15x1x2

    Notch 15x2x1

    4 MHz

    6 Mhz

    60

    65

    Pulse Echo

    TOFD

    10 x 15 mm

    6 mm

    F

    H

  • 7/27/2019 Aut Proceedure

    25/42

    Number : UT-05530 (Ref RTD-Steeltest Proc. RTDST-TP-UT-45)

    Revision : A AUTOMATIC ULTRASONIC TESTING PROCEDURE

    Date : 30.05.2005

    Page : 24 of 41

    Rntgen Technische Dienst bvCopyright RTD Delftweg 144, 3046 NC ROTTERDAM

    Attachment 3 Calibration manufacturing drawings

  • 7/27/2019 Aut Proceedure

    26/42

    Number : UT-05530 (Ref RTD-Steeltest Proc. RTDST-TP-UT-45)

    Revision : A AUTOMATIC ULTRASONIC TESTING PROCEDURE

    Date : 30.05.2005

    Page : 25 of 41

    Rntgen Technische Dienst bvCopyright RTD Delftweg 144, 3046 NC ROTTERDAM

  • 7/27/2019 Aut Proceedure

    27/42

    Number : UT-05530 (Ref RTD-Steeltest Proc. RTDST-TP-UT-45)

    Revision : A AUTOMATIC ULTRASONIC TESTING PROCEDURE

    Date : 30.05.2005

    Page : 26 of 41

    Rntgen Technische Dienst bvCopyright RTD Delftweg 144, 3046 NC ROTTERDAM

  • 7/27/2019 Aut Proceedure

    28/42

    Number : UT-05530 (Ref RTD-Steeltest Proc. RTDST-TP-UT-45)

    Revision : A AUTOMATIC ULTRASONIC TESTING PROCEDURE

    Date : 30.05.2005

    Page : 27 of 41

    Rntgen Technische Dienst bvCopyright RTD Delftweg 144, 3046 NC ROTTERDAM

  • 7/27/2019 Aut Proceedure

    29/42

    Number : UT-05530 (Ref RTD-Steeltest Proc. RTDST-TP-UT-45)

    Revision : A AUTOMATIC ULTRASONIC TESTING PROCEDURE

    Date : 30.05.2005

    Page : 28 of 41

    Rntgen Technische Dienst bvCopyright RTD Delftweg 144, 3046 NC ROTTERDAM

  • 7/27/2019 Aut Proceedure

    30/42

    Number : UT-05530 (Ref RTD-Steeltest Proc. RTDST-TP-UT-45)

    Revision : A AUTOMATIC ULTRASONIC TESTING PROCEDURE

    Date : 30.05.2005

    Page : 29 of 41

    Rntgen Technische Dienst bvCopyright RTD Delftweg 144, 3046 NC ROTTERDAM

  • 7/27/2019 Aut Proceedure

    31/42

    Number : UT-05530 (Ref RTD-Steeltest Proc. RTDST-TP-UT-45)

    Revision : A AUTOMATIC ULTRASONIC TESTING PROCEDURE

    Date : 30.05.2005

    Page : 30 of 41

    Rntgen Technische Dienst bvCopyright RTD Delftweg 144, 3046 NC ROTTERDAM

  • 7/27/2019 Aut Proceedure

    32/42

    Number : UT-05530 (Ref RTD-Steeltest Proc. RTDST-TP-UT-45)

    Revision : A AUTOMATIC ULTRASONIC TESTING PROCEDURE

    Date : 30.05.2005

    Page : 31 of 41

    Rntgen Technische Dienst bvCopyright RTD Delftweg 144, 3046 NC ROTTERDAM

  • 7/27/2019 Aut Proceedure

    33/42

    Number : UT-05530 (Ref RTD-Steeltest Proc. RTDST-TP-UT-45)

    Revision : A AUTOMATIC ULTRASONIC TESTING PROCEDURE

    Date : 30.05.2005

    Page : 32 of 41

    Rntgen Technische Dienst bvCopyright RTD Delftweg 144, 3046 NC ROTTERDAM

  • 7/27/2019 Aut Proceedure

    34/42

    Number : UT-05530 (Ref RTD-Steeltest Proc. RTDST-TP-UT-45)

    Revision : A AUTOMATIC ULTRASONIC TESTING PROCEDURE

    Date : 30.05.2005

    Page : 33 of 41

    Rntgen Technische Dienst bvCopyright RTD Delftweg 144, 3046 NC ROTTERDAM

  • 7/27/2019 Aut Proceedure

    35/42

  • 7/27/2019 Aut Proceedure

    36/42

    Number : UT-05530 (Ref RTD-Steeltest Proc. RTDST-TP-UT-45)

    Revision : A AUTOMATIC ULTRASONIC TESTING PROCEDURE

    Date : 30.05.2005

    Page : 35 of 41

    Rntgen Technische Dienst bvCopyright RTD Delftweg 144, 3046 NC ROTTERDAM

  • 7/27/2019 Aut Proceedure

    37/42

    Number : UT-05530 (Ref RTD-Steeltest Proc. RTDST-TP-UT-45)

    Revision : A AUTOMATIC ULTRASONIC TESTING PROCEDURE

    Date : 30.05.2005

    Page : 36 of 41

    Rntgen Technische Dienst bvCopyright RTD Delftweg 144, 3046 NC ROTTERDAM

  • 7/27/2019 Aut Proceedure

    38/42

    Number : UT-05530 (Ref RTD-Steeltest Proc. RTDST-TP-UT-45)

    Revision : A AUTOMATIC ULTRASONIC TESTING PROCEDURE

    Date : 30.05.2005

    Page : 37 of 41

    Rntgen Technische Dienst bvCopyright RTD Delftweg 144, 3046 NC ROTTERDAM

  • 7/27/2019 Aut Proceedure

    39/42

  • 7/27/2019 Aut Proceedure

    40/42

    Number : UT-05530 (Ref RTD-Steeltest Proc. RTDST-TP-UT-45)

    Revision : A AUTOMATIC ULTRASONIC TESTING PROCEDURE

    Date : 30.05.2005

    Page : 39 of 41

    Rntgen Technische Dienst bvCopyright RTD Delftweg 144, 3046 NC ROTTERDAM

    Attachment 5 Rotoscan report

  • 7/27/2019 Aut Proceedure

    41/42

    Number : UT-05530 (Ref RTD-Steeltest Proc. RTDST-TP-UT-45)

    Revision : A AUTOMATIC ULTRASONIC TESTING PROCEDURE

    Date : 30.05.2005

    Page : 40 of 41

    Rntgen Technische Dienst bvCopyright RTD Delftweg 144, 3046 NC ROTTERDAM

    Attachment 6 Reportable indication sheet

  • 7/27/2019 Aut Proceedure

    42/42

    Number : UT-05530 (Ref RTD-Steeltest Proc. RTDST-TP-UT-45)

    Revision : A AUTOMATIC ULTRASONIC TESTING PROCEDURE

    Date : 30.05.2005

    Page : 41 of 41

    Rntgen Technische Dienst bvCopyright RTD Delftweg 144, 3046 NC ROTTERDAM

    Attachment 7 Gain settings calibration record