Page 1: Auto Scan/Spot Read Spectrophotometer · 2011-06-23 · DTP45 SPECTROPHOTOMETER 2 Warranty Information X-Rite, Incorporated (“X-Rite”) warrants each instrument manufactured to

DTP45 Auto Scan/Spot Read

Spectrophotometer

Operators Manual

Page 2: Auto Scan/Spot Read Spectrophotometer · 2011-06-23 · DTP45 SPECTROPHOTOMETER 2 Warranty Information X-Rite, Incorporated (“X-Rite”) warrants each instrument manufactured to

D T P 4 5 S P E C T R O P H O T O M E T E R

1

CE Declaration Manufacturer's Name: X-Rite, Incorporated Manufacturer's Address: 4300 44th Street, S.E. • Grand Rapids, Michigan • U.S.A. Model Name: Spectrophotometer Model No.: DTP45 Directive(s) Conformance: EMC 89/336/EEC LVD 73/23/EEC

WEEE As of August 13, 2005, X-Rite products meet the European Union – Waste Electrical and Electronic Equipment (WEEE) directive. Please refer to www.xrite.com for more information on X-Rite’s compliance with the WEEE directive.

Federal Communications Commission Notice NOTE: This equipment has been tested and found to comply with the limits for a Class A digital device, pursuant to Part 15 of the FCC Rules. These limits are designed to provide reasonable protection against harmful interference when the equipment is operated in a commercial environment. This equipment generates, uses, and can radiate radio frequency energy and, if not installed and used in accordance with the instruction manual, may cause harmful interference to radio communications. Operation of this equipment in a residential area is likely to cause harmful interference in which case the user will be required to correct the interference at his own expense.

Industry Canada Compliance Statement This Class A digital apparatus complies with Canadian ICES-003.

Cet appareil numérique de la classe A est conforme à la norme NMB-003 du Canada.

Equipment Information Use of this equipment in a manner other than that specified by X-Rite, Incorporated may compromise design integrity and become unsafe. WARNING: This instrument is not for use in explosive environments. ADVERTENCIA - NO use este aparato en los ambientes explosivos. AVVERTIMENTO - NON usare questo apparecchio in ambienti esplosivi. WARNUNG: Das Gerät darf in einer explosiven Umgebung NICHT verwendet werden. AVERTISSEMENT: Cet instrument ne doit pas être utilisé dans un environnement explosif.

Proprietary Notice The information contained in this manual is derived from patent and proprietary data of X-Rite, Incorporated. The contents of this manual are the property of X-Rite, Incorporated and are copyrighted. Any reproduction in whole or part is strictly prohibited. Publication of this information does not imply any rights to reproduce or use this manual for any purpose other than installing, operating, or maintaining this instrument. No part of this manual may be reproduced, transcribed, transmitted, stored in a retrieval system, or translated into any language or computer language, in any form or by any means, electronic, magnetic, mechanical, optical, manual, or otherwise, without the prior written permission of an officer of X-Rite, Incorporated.

This product may be covered by one or more patents. Refer to the instrument for actual patent numbers.

Copyright © 2007 by X-Rite, Incorporated “ALL RIGHTS RESERVED” X-Rite® is a registered trademark of X-Rite, Incorporated. All other logos, brand names, and product names mentioned are the properties of their respective holders.

Page 3: Auto Scan/Spot Read Spectrophotometer · 2011-06-23 · DTP45 SPECTROPHOTOMETER 2 Warranty Information X-Rite, Incorporated (“X-Rite”) warrants each instrument manufactured to

D T P 4 5 S P E C T R O P H O T O M E T E R

2

Warranty Information X-Rite, Incorporated (“X-Rite”) warrants each instrument manufactured to be free of defects in material and workmanship (excluding battery pack) for a period of 12 months. This warranty shall be fulfilled by the repair or replacement, at the option of X-Rite, of any part or parts, free of charge including labor, F.O.B. its factory or authorized service center.

This warranty shall be voided by any repair, alteration, or modification, by persons other than employees of X-Rite, or those expressly authorized by X-Rite to perform repairs, and by any abuse, misuse, or neglect of the product, or by use not in accordance with X-Rite’s published instructions.

X-Rite reserves the right to make changes in design and /or improvements to its products without any obligation to include these changes in any products previously manufactured. Correction of defects by repair or replacement shall constitute fulfillment of all warranty obligations on the part of X-Rite.

THIS WARRANTY IS EXPLICITLY IN LIEU OF ANY OTHER EXPRESSED OR IMPLIED WARRANTIES, INCLUDING ANY IMPLIED WARRANTY OF MERCHANTABILITY OR FITNESS FOR ANY PARTICULAR PURPOSE. THIS WARRANTY OBLIGATION IS LIMITED TO REPAIR OR REPLACEMENT OF THE UNIT RETURNED TO X-RITE OR AN AUTHORIZED SERVICE CENTER FOR THAT PURPOSE.

This agreement shall be interpreted in accordance with the laws of the State of Michigan and jurisdiction and venue shall lie with the courts of Michigan as selected by X-Rite, Incorporated.

Table of Contents

Overview and Setup 3 Instrument Description 3 Unpacking and Inspection 3

Packaging Content 4 Connecting the Cabling 4 Removing the Instrument from the Base Assembly 5 Attaching the Instrument to the Base Assembly 5 Instrument LED Indicator 6

Calibration 7 Frequency of Calibration 7 Calibration Procedure 7

Sample Measurement 8 Measuring Strips 8

Measurement Techniques 9 Spot Measurement 10

Appendices 11 Troubleshooting 11 Service Information 12 General Maintenance 12

General Cleaning 12 Cleaning the Strip Reading Mechanism 12 Cleaning the Optics 14 Cleaning the Calibration Reference 14

Technical Information 15 Instrument Reset 15 Technical Notes 15

Instrument Specifications 16 General 16

Page 4: Auto Scan/Spot Read Spectrophotometer · 2011-06-23 · DTP45 SPECTROPHOTOMETER 2 Warranty Information X-Rite, Incorporated (“X-Rite”) warrants each instrument manufactured to

D T P 4 5 S P E C T R O P H O T O M E T E R

3

Overview and Setup

Instrument Description The DTP45 Spectrophotometer is a color measurement instrument that reports densitometric, colorimetric and spectral data. The instrument is designed to calibrate color output devices including large format printers, proofers, and digitally connected color copiers. The DTP45 is open ended at the top of the base assembly to read linear groups of patches on large media sheets without cutting. The instrument can also be used for “spot” measurements when detached from the base assembly, providing L*a*b* and illuminant/observer data on its display.

This manual covers the installation, basic operation and maintenance of the instrument. Specific instructions for using the instrument with your software application can be found in the software documentation.

Unpacking and Inspection After removing the instrument from the shipping carton, inspect it for damage. If any damage has occurred during shipping, immediately contact the transportation company. Do not proceed with installation until the carrier’s agent has inspected the damage.

Your instrument is packaged in a specially designed carton to assure against damage. If shipment is necessary, the instrument should be packaged in the original carton along with all the accessories. If the original carton is not available, contact X-Rite to have a replacement shipped to you.

Interface Cable

3-Color LED indicator

Operation button

Instrument release button

LCD display (spot read mode)

Alignment marks

Page 5: Auto Scan/Spot Read Spectrophotometer · 2011-06-23 · DTP45 SPECTROPHOTOMETER 2 Warranty Information X-Rite, Incorporated (“X-Rite”) warrants each instrument manufactured to

D T P 4 5 S P E C T R O P H O T O M E T E R

4

Packaging Content Your instrument packaging should contain all the items listed below. If any of these items are missing, contact X-Rite or an Authorized Representative.

• DTP45 Instrument with Interface Cable and integrated calibration reference • Strip Measurement Base • Switching Power Supply with line cord • Certificate of Calibration • Manuals and Utilities CD • Registration Material

Connecting the Cabling IMPORTANT: You must install the software before connecting the DTP45 to your computer.

1. Locate the DTP45 next to the computer and plug the input connector from the power supply into the connector on the interface cable. CAUTION: Use only the 100-240VAC input, +12VDC output Switching Power Supply provided to power the instrument.

2. Plug the detachable line cord into the power supply and then plug the line cord into the AC wall receptacle.

3. Plug the USB cable into an available port on your computer. The computer should confirm that the USB connector is plugged in and automatically install the device driver for the instrument.

NOTE: If the third-party software application you are using does not contain the DTP45 USB driver, you will have to install it. The driver is located in the Drivers folder on the DTP Series Manuals and Utilities CD.

AC Wall Receptacle

Detachable Line Cord

Power Input

To ComputerUSB Port

Power Supply

Page 6: Auto Scan/Spot Read Spectrophotometer · 2011-06-23 · DTP45 SPECTROPHOTOMETER 2 Warranty Information X-Rite, Incorporated (“X-Rite”) warrants each instrument manufactured to

D T P 4 5 S P E C T R O P H O T O M E T E R

5

Removing the Instrument from the Base Assembly Press the Release button located on the base assembly, behind the instrument. Lift up on the back of the instrument, freeing it from the base assembly.

Attaching the Instrument to the Base Assembly Insert the front of the instrument into the channel at the front of the base assembly. Push down on the back of the instrument until it locks into position.

Lift Back

Insert Front

Page 7: Auto Scan/Spot Read Spectrophotometer · 2011-06-23 · DTP45 SPECTROPHOTOMETER 2 Warranty Information X-Rite, Incorporated (“X-Rite”) warrants each instrument manufactured to

D T P 4 5 S P E C T R O P H O T O M E T E R

6

Instrument LED Indicator The LED indicates a variety of instrument operation conditions, such as calibration mode and operation. Below is a complete list of conditions reported by the LED.

Operation Mode (green LED) • Solid Green—self-test passed and instrument is ready for use.

• Slow Flashing Green—measurement in progress.

• Fast Flashing Green—measurement failed (strip reading was attempted but the wrong strip was read, or strip was skewed during reading).

Calibration Mode (amber LED)

• Solid Amber—instrument calibration mode.

• Slow Flashing Amber—instrument calibration in progress.

• Solid Green after Flashing Amber—instrument calibration was successful and instrument is ready for use.

• Fast Flashing Amber—instrument calibration failed.

Error/Reset Mode (red LED) • Solid Red—instrument is in an error state. Perform an instrument reset.

• Fast Flashing Red—instrument has a system failure. Contact your service representative.

Page 8: Auto Scan/Spot Read Spectrophotometer · 2011-06-23 · DTP45 SPECTROPHOTOMETER 2 Warranty Information X-Rite, Incorporated (“X-Rite”) warrants each instrument manufactured to

D T P 4 5 S P E C T R O P H O T O M E T E R

7

Calibration

Frequency of Calibration Typically, the software application prompts for an instrument calibration when required. The frequency at which this occurs depends on the application. If desired, a calibration can be started manually by pressing the Instrument button—refer below for procedure. At a minimum, the instrument should be calibrated once a day for best accuracy.

Calibration Procedure The calibration reference is concealed behind a cover in the bottom of the instrument.

1. Remove the instrument from the base assembly as explained earlier.

2. Turn the instrument over and lift up on the front of the calibration reference cover. Take care not to touch the white reference.

3. Position the reference over the optics and press down until it is secure.

4. If a calibration is called for by the application, select the appropriate option to initiate the calibration and advance to Step 6. If you are manually activating a calibration, press and hold the Operation button on top of the instrument until the LED indicator changes to solid amber (approx. 3 seconds).

5. Press the Operation button again.

6. The LED indicator slowly flashes amber during the calibration. The LED indicator then turns solid green after a successful calibration. If the calibration fails (fast flashing amber LED), verify reference is clean and re-measure.

7. Position the reference back into its storage location.

Calibration reference cover

Lift Here

Read switch

Page 9: Auto Scan/Spot Read Spectrophotometer · 2011-06-23 · DTP45 SPECTROPHOTOMETER 2 Warranty Information X-Rite, Incorporated (“X-Rite”) warrants each instrument manufactured to

D T P 4 5 S P E C T R O P H O T O M E T E R

8

Sample Measurement

Measuring Strips You should refer to the documentation for the software program that you are using with your instrument. All applications that utilize this instrument must be running during measurements.

• The strip must have at least a 1.2″ (30.5mm) leader before the leading edge of the first step. Contact X-Rite Applications Support if additional information is required. Inspect the strip for spots or defects on the steps. Defects may cause inaccurate measurements.

• The darkest patch of a strip must be read first unless there is at least 1.1″ (27.9mm) unprinted trailer after the last patch.

• Some software packages define strips as “multi-pass” where all color passes must be read before a confirmation is displayed on the monitor. In these cases, a successful reading of a color pass is shown by the instrument slowly flashing green after the reading. This indicates that you may continue to the next color pass. Fast flashing of green indicates a re-read is required on the pass. The instrument’s LED will stay at solid green after all passes have been successfully read.

Strip Positioning and Patch Requirements

0.75"(19mm)

min.

4" (101.6mm) max. to paper stopPaper Stop

Alignment Marks

0.50" (12.7mm) min. patch width

1.2" (30.5mm)min. leader

length0.40" (10.mm)

min. patchheight

Page 10: Auto Scan/Spot Read Spectrophotometer · 2011-06-23 · DTP45 SPECTROPHOTOMETER 2 Warranty Information X-Rite, Incorporated (“X-Rite”) warrants each instrument manufactured to

D T P 4 5 S P E C T R O P H O T O M E T E R

9

Measurement Techniques The following information is provided to familiarize you with reading strips, a series of printed control patches.

1. Select strip type to measure from the software application.

2. Position the strip so that the group of patches to measure is centered between the two alignment marks on the base assembly, and insert it until it stops against the feed rollers.

For large strips (media that extends outside the left side of the base), guide the strip with slight pressure. This prevents skewing of the strip as it is measured. Do not “tug” forwards or backwards on strip while the internal motor is pulling the strip through.

3. Press the Operation button on the instrument. The LED indicator slowly flashes green while the strip is pulled through the instrument and out the back.

A successfully reading is verified by a solid green LED indicator, a short beep on the instrument, and a confirmation message on the computer monitor.

Press the Operation button to activate the measurement

Insert the strip until it is stops against the feed rollers

Alignment Marks

Page 11: Auto Scan/Spot Read Spectrophotometer · 2011-06-23 · DTP45 SPECTROPHOTOMETER 2 Warranty Information X-Rite, Incorporated (“X-Rite”) warrants each instrument manufactured to

D T P 4 5 S P E C T R O P H O T O M E T E R

10

Spot Measurement Your DTP45 instrument is designed to read spot colors and display the results as L*a*b* data using a D50

2 illuminant/observer as the default.

1. Remove the instrument from its base assembly as explained earlier.

2. Locate the targeting shoe at the front of the instrument and rotate it outward. Continue to rotate the shoe until the target window in the shoe is positioned in front of the optics.

3. Slowly release the targeting window, allowing it to extend out of the instrument.

4. Select a desired spot color of at least .5″ (12.7mm) diameter and center the target window over the patch. Make sure the entire instrument is located on the sample to measure.

5. Lower the instrument to the shoe and hold until the reading is displayed.

6. View the L*a*b* and illuminant/observer results on the instrument display. A successfully reading is verified by a beep and solid green LED indicator on the instrument, or a confirmation message on the computer monitor.

Optics

Targeting Shoe

Target Window

Operation button

Page 12: Auto Scan/Spot Read Spectrophotometer · 2011-06-23 · DTP45 SPECTROPHOTOMETER 2 Warranty Information X-Rite, Incorporated (“X-Rite”) warrants each instrument manufactured to

D T P 4 5 S P E C T R O P H O T O M E T E R

11

Appendices

Troubleshooting The DTP45 interfaces with X-Rite’s Monaco software and a number of third-party software applications. To get the most specialized technical assistance, you should first contact the technical service representative for the specific software that you are using.

Prior to contacting X-Rite Customer Service for instrument problems, try the applicable solution(s) described below. If the condition persists, contact a Customer Service Representative toll-free by phone at: 1-888-826-3059; or by fax at: 1-888-826-3061. You can also contact X-Rite’s support staff through our Support page at www.xrite.com.

LED indicator not lit:

• Ensure that the power supply is plugged in and connected to the Interface Cable.

• Reset the instrument (see Instrument Reset).

Fast flashing green LED: • If error persists, clean instrument and calibration reference (see General Cleaning).

• Reset the instrument (see Instrument Reset).

Solid red LED:

• Remove power from the instrument, reapply power and see if the condition is corrected.

• Reset the instrument (see Instrument Reset) and calibrate.

• This could indicate that the unit may not be programmed properly, which could happen if power was removed during reprogramming. Try reprogramming the instrument again (refer to the programming instructions that came with your update).

Instrument and software not communicating: • Verify that the instruments LED indicates normal operation (see Instrument LED

Indicator). If normal, close and restart the software application. If this does not work reboot the computer.

• Remove power from the instrument, reapply power and see if the condition is corrected.

• Check for proper configuration setting from the software provider.

• Reset the instrument (see Instrument Reset) and calibrate.

Repeated strip measurement failures: • Ensure that the strip is being read in accordance with your software’s documentation.

• Close and restart the software application.

• Perform a calibration on the instrument (see Calibration section).

• Reset the instrument (see Instrument Reset).

Repeated calibration failures:

• Clean instrument and calibration reference (see General Maintenance).

• Reset the instrument (see Instrument Reset) and calibrate.

Page 13: Auto Scan/Spot Read Spectrophotometer · 2011-06-23 · DTP45 SPECTROPHOTOMETER 2 Warranty Information X-Rite, Incorporated (“X-Rite”) warrants each instrument manufactured to

D T P 4 5 S P E C T R O P H O T O M E T E R

12

Service Information The DTP45 Spectrophotometer is covered by a one-year limited warranty and should be referred to an authorized service center for repairs within the warranty period.

X-Rite provides repair service to their customers. Because of the complexity of the circuitry, all repairs should be referred to an authorized service center.

X-Rite will repair any instrument past warranty. The customer shall pay shipping and repair cost to the authorized service center, and the instrument shall be submitted in the original carton, as a complete unaltered unit along with all the supplied accessories.

General Maintenance Your instrument requires very little maintenance to achieve years of reliable operation. However, to protect your investment and maintain reading accuracy, a few simple-cleaning procedures should be performed from time to time.

General Cleaning The exterior of the instrument and base assembly may be wiped clean with a cloth dampened in water or mild cleaner. NOTE: DO NOT use any solvents to the clean the instrument, this will cause damage to the cover.

Cleaning the Strip Reading Mechanism 1. Remove the base assembly from the instrument and set the instrument aside.

2. Turn the base assembly upside-down.

3. Compress the two tabs on opposite sides of the upper base towards each other.

4. Push downward, separating the lower base and upper base.

Tabs (2)

Page 14: Auto Scan/Spot Read Spectrophotometer · 2011-06-23 · DTP45 SPECTROPHOTOMETER 2 Warranty Information X-Rite, Incorporated (“X-Rite”) warrants each instrument manufactured to

D T P 4 5 S P E C T R O P H O T O M E T E R

13

5. Clean both base surfaces, idler rollers, drive wheels, and backer block with a lint-free cloth.

NOTE: To thoroughly clean the entire drive wheel, attach the instrument. Press the Operation button—with instrument plugged in—to rotate wheels.

6. Position the lower base over the upper base and insert tabs into openings. The lower base is positioned properly on the upper base when it snaps into place.

Idler Rollers (5)

Idler Rollers (2)

Idler Rollers (2)

Drive Wheels (5)

Upper Base

Lower Base

Backer Block)

Page 15: Auto Scan/Spot Read Spectrophotometer · 2011-06-23 · DTP45 SPECTROPHOTOMETER 2 Warranty Information X-Rite, Incorporated (“X-Rite”) warrants each instrument manufactured to

D T P 4 5 S P E C T R O P H O T O M E T E R

14

Cleaning the Optics 1. Remove the instrument from the base assembly.

2. Carefully blow short burst of clean, dry air into the measurement aperture. This should remove any accumulated dust in the optics area. CAUTION: DO NOT invert cans that use compressed gas as a propellant, doing so could cause damage to the optics assembly.

Cleaning the Calibration Reference 1. Remove the instrument from the base assembly as explained earlier.

2. Turn the instrument over and lift up on the front of the calibration reference cover. Rotate the cover upwards until it stops against the instrument.

3. Clean the calibration reference using a cotton swab or lint-free cloth.

4. Position the reference back into its storage location.

Calibration Reference

Page 16: Auto Scan/Spot Read Spectrophotometer · 2011-06-23 · DTP45 SPECTROPHOTOMETER 2 Warranty Information X-Rite, Incorporated (“X-Rite”) warrants each instrument manufactured to

D T P 4 5 S P E C T R O P H O T O M E T E R

15

Technical Information

Instrument Reset Note: When performing an Instrument Reset, which restores the X-Rite Factory Defaults, you will have to ensure the correct configuration for your software is then restored.

The following procedure resets DTP45 to the factory default settings.

1. Press and hold the Operation button on top of the instrument until the LED indicator changes to solid red (approx. 6 seconds).

2. Press the Operation button again. The factory defaults are now reinstated.

Technical Notes Firmware Update After updating the instrument firmware you must power down the instrument and then power it back up. This ensures that the instrument functions properly and user-configurable settings are restored to the factory defaults.

Page 17: Auto Scan/Spot Read Spectrophotometer · 2011-06-23 · DTP45 SPECTROPHOTOMETER 2 Warranty Information X-Rite, Incorporated (“X-Rite”) warrants each instrument manufactured to

D T P 4 5 S P E C T R O P H O T O M E T E R

16

Instrument Specifications

General Measurement Geometry: ........ Reflection 45°/0° per ANSI ISO 5-4 (IT2.17)

Spot Size on Sample:............. 1.8 mm (.070 in.) in scan direction, 2.5 mm (.097 in.) wide

Light Source: ........................ Gas pressure @ 2850°K

Spectral Sensor: ................... DRS Technology, 24 point engine, 31 point reporting

Spectral Range: .................... 400nm to 700nm

Illumination Types: ................ A, C, D50, D55, D65, D75, F2, F11, & F12

Standard Observer: ............... 2° & 10°

Response Type:..................... Status T, E, I, A, M, & Others

Measurement Time: ............... Approx. 0.25 seconds per patch (7 mm patches)

Inter-instrument Agreement: .. 0.3 ∆E cmc avg. typical (avg., based on 12 BCRA tiles)

Measurement Range: ............. 0.00D to 2.50D; 0 to 160% R

Repeatability on White: .......... 0.2 DE max; ±0.01D max

Linearity: ............................. ±0.01D or 1%

Warm Up Time:..................... None

Dimensions (overall):............. L: 11.87” (30.14 cm), W: 4.66” (11.84 cm), H: 4.10” (10.41 cm)

Weight:................................ Base 1.50 lbs (680 g) Instrument 1.45 lbs (658 g) Combined 2.95 lbs (1.34 kg)

Power Requirements: ............. +12VDC @ 700 mA maximum

Design and specifications subject to change without notice.

Page 18: Auto Scan/Spot Read Spectrophotometer · 2011-06-23 · DTP45 SPECTROPHOTOMETER 2 Warranty Information X-Rite, Incorporated (“X-Rite”) warrants each instrument manufactured to

Corporate Headquarters - USA 4300 44th Street SE Grand Rapids, Michigan 49512 Phone 1 800 248 9748 or 1 616 803 2100 Fax 1 800 292 4437 or 1 616 803 2705 Corporate Headquarters - Europe Althardstrasse 70 8105 Regensdorf Switzerland Phone (+41) 44 842 24 00 Fax (+41) 44 842 22 22 Corporate Headquarters - Asia Room 808-810 Kornhill Metro Tower, 1 Kornhill Road Quarry Bay, Hong Kong Phone (+852) 2 568 6283 Fax (+852) 2 885 8610 Please visit www.xrite.com for a local office near you.

P/N DTP45-500 Rev. C