Ion Suppression of target ions by matrix interferences in urine were reduced when spotting sub-microliter volumes of sample onto QS-HT mesh screens for DART introduction compared to larger volumes The same low volume pipettor can spot in a 384 sample well format onto QuickStrip-HT screens, allowing DART analysis of samples at 4 seconds per sample, or 384 samples in <25 minutes Quantitation of drugs in urine is possible with low volume spotting and analysis by DART-MS, and sampling is consistent with %RSD 1% Thanks to TTP LabTech for technical support with the mosquito® X1 and for their collaboration Abstract Direct Analysis in Real Time (DART®) coupled to mass spectrometry is an ambient ionization screening technique capable of detecting drugs of abuse in biological matrices, such as urine or oral fluid. However, significant matrix effects are encountered in these samples, leading to substantial ion suppression of the target drugs. To minimize matrix effects, we use a low volume pipettor robot to spot sub-microliter volumes of spiked urine samples. Thermal desorption of the smaller ionization region of the lower volume spiked sample reduces the ion suppression of target drugs from matrix effects. The integration of the low volume pipettor robot to the DART-MS system permitted rapid, reliable, and unattended detection of drugs in up to 384 biological samples consecutively. Implementation of a Novel Parallel Solid Phase Microextraction with DART Technology for High Throughput Screening of Drugs Brittany Laramee, Brian Musselman, Joseph Tice, Steve Shrader, and Frederick Li Conclusion Experimental: Automated Low Volume Pipetting in 384 Well Spacing Format to Facilitate High-Throughput DART-MS Analysis of Target Ions in Matrices Brittany Laramee, Frederick Li, Paul Liang, and Brian Musselman IonSense, Inc., Saugus, MA 01906 Acknowledgement Reduced matrix interference from urine in a high-throughput workflow is demonstrated with the low volume pipettor Cocaine was spiked into urine and analyzed by DART-MS Below is a comparison of target ion intensity and matrix effects between automated 200 nL deposition by Mosquito, and 1 µL deposition Methods Automated Low Volume Pipettor High-throughput analysis was enabled with the low volume liquid handling pipettor, the mosquito® X1 (TTP LabTech). The mosquito® X1 can spot sub-microliter volumes of sample onto mesh screens in a 384 well format, for automated direct introduction to DART-MS for analysis. With Mosquito spotting in a 384 spot format, DART-HT can scan at 4 seconds per sample, analyzing 384 samples in 10 minutes Speed of Analysis of Drugs of Abuse in Urine QuickStrip-96/384™ permits analysis with SBS format samples TTPLabtech Mosquito® Robotics for rapid simultaneous low volume sample pipetting onto the QuickStrip-HT 1 µL 200 nL Cocaine (M+H) = 304.15 m/z With this workflow we are able to show; 1) Reduced matrix interferences and ion suppression of target drugs in biological matrices 2) Speeds of 4 seconds per sample allowing for analysis of 384 samples in less than 10 minutes 3) Accurate quantitation with low %RSD values Reliable and consistent sample deposition and ionization for Quantitation Urine samples spotted by Mosquito and analyzed by DART allow for quantitation of target drugs, with calibration curves showing R 2 values greater than 0.988 Sample deposition and analysis is consistent, with the %RSD for a row of 24 samples 1% with internal standard correction

Upload: others

Post on 28-Jul-2020

3 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: BL- Mosquito DART ASMS 2018 Poster Draft€¦ · Title: Microsoft PowerPoint - BL- Mosquito DART ASMS 2018 Poster Draft Author: laramee Created Date: 5/25/2018 4:27:31 PM

Ion Suppression of target ions by matrix interferences in urine were reduced when spotting sub-microliter volumes of sample onto QS-HT mesh screens for DART introduction compared to larger volumes

The same low volume pipettor can spot in a 384 sample well format onto QuickStrip-HT screens, allowing DART analysis of samples at 4 seconds per sample, or 384 samples in <25 minutes

Quantitation of drugs in urine is possible with low volume spotting and analysis by DART-MS, and sampling is consistent with %RSD 1%

Thanks to TTP LabTech for technical support with the mosquito® X1 and for their collaboration

AbstractDirect Analysis in Real Time (DART®) coupled to mass spectrometry is an ambient ionization screening technique capable of detecting drugs of abuse in biological matrices, such as urine or oral fluid. However, significant matrix effects are encountered in these samples, leading to substantial ion suppression of the target drugs. To minimize matrix effects, we use a low volume pipettor robot to spot sub-microliter volumes of spiked urine samples. Thermal desorption of the smaller ionization region of the lower volume spiked sample reduces the ion suppression of target drugs from matrix effects. The integration of the low volume pipettor robot to the DART-MS system permitted rapid, reliable, and unattended detection of drugs in up to 384 biological samples consecutively.

Implementation of a Novel Parallel Solid Phase Microextraction with DART Technology for High Throughput Screening of Drugs

Brittany Laramee, Brian Musselman, Joseph Tice, Steve Shrader, and Frederick Li

Conclusion

Experimental:

Automated Low Volume Pipetting in 384 Well Spacing Format to Facilitate High-Throughput DART-MS Analysis of Target Ions in Matrices Brittany Laramee, Frederick Li, Paul Liang, and Brian MusselmanIonSense, Inc., Saugus, MA 01906

Acknowledgement

Reduced matrix interference from urine in a high-throughput workflow is demonstrated with the low volume pipettor

Cocaine was spiked into urine and analyzed by DART-MS Below is a comparison of target ion intensity and matrix effects

between automated 200 nL deposition by Mosquito, and 1 µL deposition

MethodsAutomated Low Volume PipettorHigh-throughput analysis was enabled with the low volume liquid handling pipettor, the mosquito® X1 (TTP LabTech). The mosquito® X1 can spot sub-microliter volumes of sample onto mesh screens in a 384 well format, for automated direct introduction to DART-MS for analysis.

With Mosquito spotting in a 384 spot format, DART-HT can scan at 4 seconds per sample, analyzing 384 samples in 10 minutes

Speed of Analysis of Drugs of Abuse in Urine

QuickStrip-96/384™ permits analysis with SBS

format samples

TTPLabtech Mosquito® Robotics for rapid simultaneous low volume sample pipetting onto the QuickStrip-HT

1 µL

200 nL

Cocaine (M+H) = 304.15 m/z

With this workflow we are able to show;1) Reduced matrix interferences and ion suppression of target

drugs in biological matrices 2) Speeds of 4 seconds per sample allowing for analysis of 384

samples in less than 10 minutes3) Accurate quantitation with low %RSD values

Reliable and consistent sample deposition and ionization for Quantitation

Urine samples spotted by Mosquito and analyzed by DART allow for quantitation of target drugs, with calibration curves showing R2 values greater than 0.988

Sample deposition and analysis is consistent, with the %RSD for a row of 24 samples 1% with internal standard correction