blood chain - b medical systems · pdf filetest reports of the accredited test laboratory ......
TRANSCRIPT
292.
9706
.78
09/2
015
The Blood Chain line is specially designed to meet the requirements of field con-ditions for military tasks as well as for civil aid organizations for the storage and transport of temperature-sensitive supplies such as blood and blood products, vaccines and pharmaceutical products.
BLOOD CHAINSolutions for safe blood transport
www.bmedicalsystems.com
Your global partner to save lives.
B Medical Systems I Blood Chain
02 | 03 BLOOD CHAIN Your global partner to save lives.
Equipment dedicated to the Armed Forces for the storage and transport of blood, vaccines and medicines
B Medical Systems’s Blood Chain range has been specially dedicated for the supply of blood, vaccines and medicines in war time and in times of natural catastrophes. This range was created after that B Medical Systems had developed the model RCB 42 P for the Pentagon who was searching for a solution for the transport and storage of blood to the front line. This model became a standard and has been adopted by all NATO Armed Forces.
Made of rotomoulded polyethylene, the Blood Chain transport boxes offer excellent mechanical resistance, as proven by drop tests, and are not subject to corrosion. They are perfectly suited to intensive use, with frequent handling under difficult climatic conditions. The polyurethane foam injected between the double walls of these boxes ensures perfect insulation and thus preservation of the quality of blood, blood products and pharmaceutical products even over long periods of transport. Used with frozen icepacks or also wet ice for some versions, they are fully self-suffi-cient with respect to the ambient environment and therefore ideal for all transport stages.
Declaration of Conformity (in accordance with ADR / RID / IMDG / ICAO-TI / IATA-DGR)
> European agreement concerning the international car-riage of dangerous goods by road (ADR) and by railway (RID), directive 2008 / 68 / EC.
> European agreement concerning the international carriage of dangerous goods by sea transport (IMDG), directive 2002 / 84 / EC.
> International agreement for air transport (ICAO-TI / IATA-DGR).
RCB 4 and RCB 12 may contain goods of packing groups I, II and III
RCB 8, RCB 25 and RCB 42 P may contain goods of packing groups II and III
Test reports of the accredited test laboratory IBE-BVI, Belgium.
Dep
endi
ng o
n th
e pr
oduc
t
NSNAPPROVED
ADR RID (2008/68/EC)
IMDG (2002/84/EC)
ICAO-TI / IATA-DGR
SUMMARY
Solar Direct Drive Refrigerators & Freezers for vaccines and medicinesFor rapid deployment under severe conditions
Blood Transport BoxesFor a safe transport of blood under field conditions
AccessoriesFor perfect temperature control
Pages
14 – 19
04 – 13
20 – 21
BLOOD CHAIN Storage and active / passive transport systems dedicated to the military for the supply of blood, vaccines and medicines to the front line.
B Medical SystemsYour global partner to save lives.B Medical Systems S.à r.l. is a pioneer in the medical industry. The company was founded in 1979 when the World Health Organization approached Electrolux in Vianden, Luxembourg, to provide a solution to their problems in safely storing and transporting vaccines around the world. Subsequently, the Electrolux refrigeration division became the Dometic Group.
In March 2015 Navis Capital Partners and the management team acquired the medical division from the Dometic Group under the new name B Medical Systems S.à r.l.. The name change reflects the deep conviction of the team and share-holders, to continuously deploy dedicated solutions in its customers in the healthcare field.
B Medical Systems S.à r.l. specializes in research, develop-ment and the manufacture of professional refrigerators, freezers, storage and solutions for safe transport and manage-ment of blood and vaccines. With our worldwide network of distributors, our own Research and Development capability and flexible manufacturing facilities, we will continually strive for excellence in product development, manufacturing and customer service across all of our product lines.
We are committed to exploring new technologies, increasing the variety of our products and seeking new fields of busi-ness. B Medical Systems S.à r.l. is a flexible, innovative and customer-oriented company, which has always been recog-nized for the excellence and performance of its products. We have always based our reputation on one rule: “never com-promising on standards” to deliver the best performance these programs deserve, for humanity.
Over the past few years, we have been supported extensively and passionately by our regional consultants and country partners who dedicate themselves to our cause and have gained credibility in their own right with the regional agen-cies and country authorities.
BLO
OD
CH
AIN
Solu
tion
s fo
r saf
e bl
ood
tran
spor
tA
CTIV
E SY
STEM
I R
CB /
MT
Blo
od T
rans
port
Box
esA
CCES
SOR
IES
Opt
iona
lSO
LAR
DIR
ECT
DR
IVE
I TCM
Ref
riger
ator
s &
Fre
ezer
sPA
SSIV
E SY
STEM
I R
CBB
lood
Tra
nspo
rt B
oxes
BLO
OD
CH
AIN
Solu
tion
s fo
r saf
e bl
ood
tran
spor
t
Solar Direct Drive
04 | 05 BLOOD CHAIN
Solar Direct Drive solution
Your global partner to save lives.
What is the meaning of “Solar Direct Drive”?
> The Solar Direct Drive solution consists of several refrigerators and freezers for vaccines or medicines working straight from solar panels without batteries and regulatores.
> This solution provides a reliable cold chain for vitally important vaccines or medicines even in the most remoted areas. The Solar Direct Drive solution uses the only energy source that never runs dry : our sun!
Traditional system
What are the regulations put in place? What are the most important characteristics for new products?
WHO has put in place with its PQS a new performance and quality system for the Cold Chain Equipment.
PQS performance specification for :
> Refrigerator or combined refrigerator- icepack freezer : compression cycle Solar direct drive without battery storage.
> Specification reference : E003 / RF05.3
> Product verification protocol : E003 / RF05-VP.3
The development of a new, successful and SAFE product according to PQS requirements has to fulfill a number of defined parameters, the most important of those are :
> Autonomy
Time in days that the device can maintain the vaccine load within a temperature range of +2°C to +8°C under low solar radiation conditions (e.g.: rain ).
> Hold over time
Time in hours that the device can maintain the vaccine load within a temperature range of +2°C to +10°C without any solar power input and at the highest rated ambient temperature.
Solar Direct Drive
NSNAPPROVED
Solar Direct Drive Refrigerators & Freezers for rapid deployment under severe conditions
> You need a quick and reliable response to store vaccines or medicines anywhere in the world!
SOLA
R D
IREC
T D
RIV
E I T
CMR
efrig
erat
ors
& F
reez
ers
SOLA
R D
IREC
T D
RIV
E I T
CMR
efrig
erat
ors
& F
reez
ers
Solar Direct Drive I B Medical Systems TCM SDD
06 | 07 BLOOD CHAIN
Vaccine Refrigerator & Icepack Freezer
Your global partner to save lives.
TCM40SDD
• Net vaccine storage capacity 36 L• Ice-pack storage capacity 8 x 0.6 L• Hold Over Time 93 h 24 at +43°C• Autonomy 81 h 54 at +43°C
+5°C
+43°C
Des
igne
d fo
r tr
opic
al t
empe
ratu
res
Performance
> First and only solar direct drive refrigerator / freezer combination with solo body
> Works at 3,5 KWh / 24 h / m² reference period
> 1 compressor and 1 cooling system allows to work with only 4 solar panels
> Designed for tropical temperatures : +5°C to +43°C
> Simple and user friendly “1 button” operation
> Robust cooling system with Secop (Danfoss) compressor and fan cooled condenser
> Best autonomy times in the market for SDD health center refrigerators
> The quick connector allows for “plug and play” installation
> Rotomoulded body construction : rust free and extremely robust
> 100 mm PU foam insulation guarantees highest possible hold over time
> Triple silicon replaceable gaskets
> Full mice protection
> No freezing risks in the refrigeration compartment
> Newly developed B Medical Systems cold chain digital controller
> Temperature monitoring maintained in power failure periods
> Green technology means : 1. environmentally friendly refrigerant and 2. low power consumption
> Ice pack freezer for the storage of 16 x 0.6 L ice-packs
> Works at 3,5 KWh / 24 h / m² reference period
> The quick connector allows for “plug and play” installation
> Rotomoulded body construction : rust free and extremely robust
> 100 mm PU foam insulation guarantees highest possible hold over time
> 4 heavy duty rollers for easy transport
> 4 hinges for adequate lid support
> Double silicon replaceable gaskets
> Full mice protection
> No freezing risks in the refrigeration compartment
> 2 electronic controllers with integrated digital temperature monitoring
> Temperature monitoring maintained in power failure periods
> Green technology means : 1. environmentally friendly refrigerant and 2. low power consumption
Advantages Advantages
> Requests 8 x 100W / 12V panelsSolar Direct Drive
NSN 4110-13-121-9002 NSN 4110-13-121-9005
TCM 40 SDD TCM 2000 SDDThe new TCM 40 SDD is a combination of a solar direct drive vaccine & medicine refrigerator and ice pack freezer, working straight from solar panels with no batteries and no regulator. [patents pending]
The new TCM 2000 SDD is a double compartment solar direct drive vaccine & medicine refrigerator and Icepack freezer with two cool-ing systems, working straight from solar panels with no batteries and no regulator. [patents pending]
> Requests 4 x 100W / 12V panelsSolar Direct Drive
TCM2000SDD
• Net vaccine storage capacity 99 L• Ice-pack storage capacity 16 x 0.6 L• Hold Over Time 92 h at +32°C• Autonomy 85 h 24 at +32°C
+5°C
+32°C
Performance
SOLA
R D
IREC
T D
RIV
E I T
CMR
efrig
erat
ors
& F
reez
ers
SOLA
R D
IREC
T D
RIV
E I T
CMR
efrig
erat
ors
& F
reez
ers
Icelined Vaccine Refrigerator or Freezer
08 | 09 BLOOD CHAIN
Solar Generator G3 I Solar Direct Drive
Your global partner to save lives.
TCM3000SDD
• Net vaccine storage capacity 156 L• Ice-pack storage capacity -• Hold Over Time 94 h 05 at +32°C• Autonomy 86 h 56 at +32°C
+5°C
+32°C
Performance
TCM 3000 SDDThe new TCM 3000 SDD is a large capacity solar direct drive vaccine & medicine refrigerator or freezer, working straight from solar pan-els with no batteries and no regulator. [patents pending]
> Works at 3,5 KWh / 24 h / m² reference period
> The quick connector allows for “plug and play” installation
> Rotomoulded body construction : rust free and extremely robust
> 100 mm PU foam insulation guarantees highest possible hold over time
> 4 heavy duty rollers for easy transport
> 4 hinges for adequate lid support
> Double silicon replaceable gaskets
> Full mice protection
> No freezing risks in the refrigeration compartment
> 1 electronic controller with integrated digital temperature monitoring
> Temperature monitoring maintained in power failure periods
> Green technology means : 1. environmentally friendly refrigerant and 2. low power consumption
NSN 4110-13-121-9004
> Requests 4 x 100W / 12V panelsSolar Direct Drive
Advantages
The new B Medical Systems Solar Generator is unique. It is a pre-assembled system that only needs to opened and adjusted to work. It requires no professional assistance for the installation. It connects to the TCM 40, 2000 or 3000 SDD by quick connector. Its packing is at the same time its body which is rotomoulded.
G3 - Ground installation, rotomoulded box
G3 Only 3 steps to catch the sun!
Installation within 30 minutes!
2. Adjust the panels
3. Plug & play
> Only 2 persons are necessary to open
the box!
> Incline the panels
until reachingthe desired
angle!
> Connect the connector cable
from the box (length of 25m) to the TCM SDD models,
only one way works!
1. Open the box
fig. T
CM 4
0 SD
D
SOLA
R D
IREC
T D
RIV
E I T
CMR
efrig
erat
ors
& F
reez
ers
SOLA
R D
IREC
T D
RIV
E I T
CMR
efrig
erat
ors
& F
reez
ers
Solar Direct Drive I Installation
10 | 11 BLOOD CHAIN Your global partner to save lives.
Easy setting-up• Quick connectors, only one way works!
Solar Generator - Plug & play Installation
The new B Medical Systems Solar Generators are unique. It connects to the TCM 40 SDD, TCM 2000 SDD or TCM 3000 SDD by quick connectors. The TCM 40 & 3000 SDD needs one generator (= 4 x 100W / 12V) and the TCM 2000 SDD requiere two generators (= 8 x 100W / 12V).
Performance
TCM2000SDD
• Net vaccine storage capacity 99 L• Ice-pack storage capacity 16 x 0.6 L• Hold Over Time 92 h at +32°C• Autonomy 85 h 24 at +32°C
+5°C
+32°C
NSN 4110-13-121-9005
Plug & play T
CM
I S
OLA
R D
IRE
CT D
RIV
E
Re
frig
era
tors
& F
ree
ze
rs
> Requests 8 x 100W / 12V panels
Easy activation• Length of cable : 25 meters each generator• Quick connectors, only one way works!
Options
Contains different basic tools as wrenches, screwdrivers, pencil, tape measure and protractor.
Includes « Basic Tool Kit » and different tools in a special backpack and a roll-up case as well as a digital voltmeter.
Includes « Tool Kit » and a drilling machine and a solar converter.
Basic Tool Kit Tool Kit Premium Tool Kit
SOLA
R D
IREC
T D
RIV
E I T
CMR
efrig
erat
ors
& F
reez
ers
SOLA
R D
IREC
T D
RIV
E I T
CMR
efrig
erat
ors
& F
reez
ers
Solar Direct Drive I B Medical Systems TCM SDD
12 | 13 BLOOD CHAIN
Technical Data
Your global partner to save lives.
Model TCM 40 SDD TCM 2000 SDD TCM 3000 SDDVaccine Refrigerator & Icepack Freezer
Icelined Vaccine Refrigerator & Icepack Freezer
Icelined Vaccine Refrigerator or Freezer
PERFORMANCENet vaccine storage capacity (l) 36 99 156
Ice-pack storage capacity 8 x 0.6 L 16 x 0.6 L -
Freezing Ice-pack capacity 1.89 kg / 24 h at +43°C 4.8 kg / 24 h at +32°C -
Hold Over Time 93 h 24 min. at +43°C 92 h at +32°C 94 h 05 min. at +32°C
Autonomy time (+2°C to +8°C) 81 h 54 min. at +43°C 85 h 24 min. at +32°C 86 h 56 min. at +32°C
Cool down time 36 h 12 h 36 h
Power consumption : Stable running 0.57 KWh / 24 h at +43°C 0.78 KWh / 24 h at +32°C 0.25 KWh / 24 h at +32°C
Cool down 0.67 KWh / 24 h at +43°C 0.74 KWh / 24 h at +32°C 0.34 KWh / 24 h at +32°C
During freezing 0.69 KWh / 24 h at +43°C 0.96 KWh / 24 h at +32°C -
Climate Zone Hot zone (+43°C) Temperate zone (+32°C) Temperate zone (+32°C)
Recommended number and type of solar panels 4 x 100 W / 12 V 8 x 100 W / 12 V (= 2 generators) 4 x 100 W / 12 V
DIMENSIONS / WEIGHTSInner dimensions H x W x D (mm) 410 x 325 x 350 Refrigerator : 585 x 470 x 425
Freezer : 340 x 250 x 505(340 x 330 x 420) + (555 x 600 x 420)
Outer dimensions H x W x D (mm) 900 x 1030 x 780 910 x 1270 x 780 910 x 1270 x 780
Shipping dimensions H x W x D (mm) 1060 x 1040 x 800 1030 x 1300 x 800 1030 x 1300 x 800
Shipping weight (kg) 114 149 144
Gross volume (l) Refrigerator : 46.6 / Freezer : 4.8 Refrigerator : 118 / Freezer : 42 187
TECHNICAL DETAILSInsulation material and thickness PU foam / 100 mm PU foam / 100 mm PU foam / 100 mm
Evaporator type and material Aluminium rollbond Aluminium rollbond Aluminium rollbond
Ice lining quantity of packs 4 x 4.8 L & 4 x 0.6 L PCM+4 4 x 3.8 L & 16 x 0.6 L 6 x 3.8 L & 9 x 0.6 L
Operating voltage range 12 - 20 V 12 - 20 V 12 - 20 V
Minimum starting voltage 12 V 12 V 12 V
Minimum starting power 80 W 80 W 80 W
Minimum power for continious running 70 W Refrig. : 66 W / Freezer : 60 W (à 20 V) 70 W
Minimum solar radiation for continious running 250 W/m² (PQS ref. 3.5 KWh / 24 h / m²) 250 W/m² (PQS ref. 3.5 KWh / 24 h / m²) 250 W/m² (PQS ref. 3.5 KWh / 24 h / m²)
Refrigerant type and quantity R600a 60g R600a Refrig. : 45g / Freezer : 40 g R600a 58g
Compressor type 1 x BD35K 2 x BD35K 1 x BD35K
Electronic type 1 x B Medical Systems / DC solar 2 x B Medical Systems / DC solar 1 x B Medical Systems / DC solar
Sensor type NTC 10kOhm NTC 10kOhm NTC 10kOhm
CERTIFICATIONNSN Code (NATO Stock Number) 4110-13-121-9002 4110-13-121-9005 4110-13-121-9004
WHO test procedure & specifications E003 / RF05-VP.3 & E003 / RF05.3 E003 / RF05-VP.2 & E003 / RF05.2 E003 / RF05-VP.2 & E003 / RF05.2
Low Voltage Directive 2006 / 95 / EC 2006 / 95 / EC 2006 / 95 / EC
ACCESSORIES (standard)Vaccine storage basket 1 4 5
Lockable lid 1 2 1
Spacers 2 2 2
Waterpacks 0.6 L 8 16 -
Waterpacks holders 8 - -
Compass 1 1 1
TCM SDD
+5°C
+32°C
+5°C
+32°C
+5°C
+43°C
Remark : Please note that above figures are for one generator (4 x 100 W / 12 V). The TCM 2000 SDD require 2 generators.
The TCM 40 & 3000 SDD needs one generator.
Model G3 - Solar Generator movable (on ground)
DIMENSIONS / WEIGHTSOuter dimensions H x W x D (mm) 400* x 3690 x 1760
Shipping dimensions H x W x D (mm) 745 x 1885 x 1765
Net weight (kg) 240
Shipping weight (kg) 240
TECHNICAL DETAILSNumber nad type of solar panels 4 x 100 W / 12 V
Minimum solar radiation required 250 W/m²
Maximum power output at STC 320 W
Solar radiation reference period PQS ref. 3.5 KWh / 24 h / m²
Output voltage range nominal 12 V, Voc 21.2 V
Type of connector B Medical Systems / Jaeger 292.9510.22
ACCESSORIES (standard)Anti-theft 1
* with horizontal solar panels
SOLA
R D
IREC
T D
RIV
E I T
CMR
efrig
erat
ors
& F
reez
ers
SOLA
R D
IREC
T D
RIV
E I T
CMR
efrig
erat
ors
& F
reez
ers
Passive & Active Systems I B Medical Systems RCB / MT
14 | 15 BLOOD CHAIN
Passive Transport Boxes for Blood
Your global partner to save lives.
RCB 8
RCB 4
The RCB 8 is perfectly suited for the transport of blood and blood products such as plasma. Its internal polysty-rene compartment protects the blood bags against direct contact with the frozen icepacks. Transparent folders containing identification documents at the front of the RCB 8 allow the contents to be identified without hav-ing to open the lids. The internal temperature can be displayed by an optional temperature recorder located in the container.
The RCB 4 is the smallest model in this range. It is easy to handle, thanks to its shoulder strap. The RCB 4 is fitted with a liquid crystal thermometer to monitor the inside temperature, with a polystyrene interior container and with a synthetic separation wall to prevent direct contact between the temperature-sensitive materials and the icepacks.
Performance RCB 8• Cooling Passve• Gross volume 20 L• Storage capacity (blood bags)
8 x 450 ml14 x 270 ml
Performance RCB 4• Cooling Passive• Gross volume 8 L• Storage capacity (blood bags)
4 x 450 ml6 x 270 ml
Cold Life at +32°C up to 57.02 hCold Life at +43°C up to 16.39 h
Cold Life at +32°C up to 46 hCold Life at +43°C up to 32.5 h
+5°C
+43°C
+5°C
+43°C
Blood Transport Boxes> in accordance with ADR / RID / IMDG / ICAO-TI / IATA-DGR
The boxes are made of rotomoulded polyethylene – a virtually indestructible plastic – and are characterised by an strength for protection against drops and knocks (as proven by drop test), as well as a perfect withstand against cor-rosion.
The polyurethane foam injected between the double walls of these boxes ensures perfect insulation of the contents and thus preservation of vaccine quality even over longer periods of transport. Used with frozen ice packs, they are fully self-sufficient with respect to the ambient environment and therefore ideal for transport stages.
The lid ensures a perfect seal for these containers and is equipped with rotomoulded grooves for facilitating stacking of the boxes, while the hinges are recessed to prevent getting caught on other boxes during handling.
Declaration of Conformity (in accordance with ADR / RID / IMDG / ICAO-TI / IATA-DGR)
> European agreement concerning the international car-riage of dangerous goods by road (ADR) and by railway (RID), directive 2008 / 68 / EC.
> European agreement concerning the international carriage of dangerous goods by sea transport (IMDG), directive 2002 / 84 / EC.
> International agreement for air transport (ICAO-TI / IATA-DGR).
RCB 4 and RCB 12 may contain goods of packing groups I, II and III
RCB 8, RCB 25 and RCB 42 P may contain goods of packing groups II and III
Test reports of the accredited test laboratory IBE-BVI, Belgium.
Blood Transport Boxes for the transport and storage to the front line
RCB passive & active / MT active
PASS
IVE
SYST
EM I
RCB
Blo
od T
rans
port
Box
es
PASS
IVE
SYST
EM I
RCB
Blo
od T
rans
port
Box
es
Passive Transport Boxes for Blood
16 | 17 BLOOD CHAIN
Technical Data
Your global partner to save lives.
Model RCB 4 RCB 8 RCB 12 RCB 25
PERFORMANCECooling Passive Passive Passive Passive
Gross volume (l) 8 20 24 44
Storage capacity (blood bags) 4 at 450 ml 8 at 450 ml 15 at 450 ml 26 at 450 ml
6 at 270 ml 14 at 270 ml 25 at 270 ml 40 at 270 ml
Cold Life at : +32°C Up to 46 h Up to 57.02 h Up to 96.14 h Up to 109.08 h
+43°C Up to 32.5 h Up to 16.39 h Up to 56.5 h Up to 74 h
(without lid opening +2°C to +10°C)with 6 bags (270 ml) and 3 (0.3 l) + 2 (0.6 l) icepacks
with 14 bags (270 ml) and 4 icepacks (0.6 l)
with 25 bags (270 ml) and 6 icepacks (0.6 l)
with 40 bags (270 ml) and 12 icepacks (0.6 l)
DIMENSIONS / WEIGHTSOuter dimensions H x W x D (mm) 299 x 362 x 283 437 x 588 x 288 499 x 550 x 475 499 x 710 x 550
Inner dimensions H x W x D (mm) 186 x 260 x 156 245 x 460 x 180 270 x 340 x 260 264 x 496 x 334
Dimensions of interior container H x W x D (mm) 135 x 245 x 90 167 x 345 x 132 192 x 310 x 237 192 x 390 x 237
Net weight - empty (kg) 3.1 7 11.7 17
Gross weight - fully stocked (kg) 7.6 14.2 25 40
TECHNICAL DETAILSOuter / Interior material Polyethylene Polyethylene Polyethylene Polyethylene
Material of interior container Polystyrene Polystyrene Stainless Steel Stainless Steel
Insulation material and thickness PU foam / 23-27 mm PU foam / 50-60 mm PU foam / 90-105 mm PU foam / 90-105 mm
Color RAL 6031 RAL 6031 RAL 6031 RAL 6031
ACCESSORIES (standard)Cooling element at 0.3 l 3 - - -
Cooling element at 0.6 l 2 4 14 24
Polystyrene interior container 1 1 (with lid) - -
Document compartment (front) 2 1 - -
Document compartment (back) 1 - - -
Stainless steel interior container - - 1 (with lid) 1 (with lid)
Carrying strap (adjustable) 1 1 - -
Securing plastic frame for interior container - - - 1
RCB passive
RCB 12
RCB 25
The RCB 12 is sufficiently large to accommodate blood as well as a substantial amount of plasma. Transport is done on short distances by the RCB 12 type. The one latch of the lid can be lead-sealed to prevent any unauthor-ised handling of the contents during transport. Standard equipment that comes with the RCB 12 includes a remov-able steel compartment as well as 14 icepacks (of 600 ml).
Foreseen for a transport of larger quantities, the RCB 25 is sufficiently large to accommodate blood as well as a large amount of plasma. A plastic reinforcement at the bottom of the RCB 25 allows the removable steel com-partment to be held in place. The latches of the lid can be lead-sealed to prevent any unauthorised handling of the contents during transport. Standard equipment that comes with the RCB 25 includes a removable steel com-partment as well as 24 icepacks (of 600 ml).
Performance RCB 12• Cooling Passive• Gross volume 24 L• Storage capacity (blood bags)
15 x 450 ml25 x 270 ml
Performance RCB 25• Cooling Passive• Gross volume 44 L• Storage capacity (blood bags)
26 x 450 ml40 x 270 ml
Cold Life at +32°C up to 96.14 hCold Life at +43°C up to 56.5 h
Cold Life at +32°C up to 109.08 hCold Life at +43°C up to 74 h
+5°C
+43°C
+5°C
+43°C
PASS
IVE
SYST
EM I
RCB
Blo
od T
rans
port
Box
es
PASS
IVE
SYST
EM I
RCB
Blo
od T
rans
port
Box
es
Active Transport Boxes for Blood
18 | 19 BLOOD CHAIN
Waterpack Freezer
Your global partner to save lives.
MT 900
The model MT 900 has been specifically developed for the freezing of cooling elements. The appliance can ac-commodate up to 187 cooling elements at 0.6 l each and can freeze up to 60 cooling elements within 24 hours at an ambient temperature of +43 °C. Due to its low energy consumption, the use is very economical.
Simple and intuitive operation is ensured by the user-friendly operation and control panel, which is equipped with a red alarm control light for the monitoring of the in-terior temperature. The appliance features a visual alarm signal. The interior is automatically illuminated when the door is opened. An integrated rapid freeze button allows the quick freezing of larger batches.
Performance MT 900• Cooling Active (Compressor)• Gross volume 290 L• Ice-pack storage capacity 187 x 0.6 L• Freezing Ice-pack capacity 60 x 0.6 L at +43°C
+16°C
+43°C
Hold Over Time 14 h at +32°C
Energy consumption 1.84 KWh / 24 h at +43°C
Operating / Alarm temperature -25°C / -20°C
Application range (at +25°C) T (+16°C to +43°C)
DIMENSIONSOuter dimensions H x W x D (mm) 1830 x 595 x 725
Inner dimensions H x W x D (mm) 1498 x 460 x 435
TECHNICAL DETAILSOuter / Interior material Galvanized sheet steel
(STO2Z-AZ150) / Styrol (SAN)
Insulation material and thickness casing / door
PU foam35 - 85 mm / 67.5 - 75 mm
Operating voltage range 220-240 V / 50-60 Hz (10A)
Power 190 W
Relative humidity ≤ 75%
Defrosting Manual
Refrigerant type and quantity R134a 245g
Heat emission 62 Kcal / h
Compressor running time 38%
Noise level (at 1m height / distance) 45 dB(A)
CERTIFICATIONEMV Directive 2004 / 108 / EC
Low-voltage Directive 2006 / 95 / EC
CHARACTERISTICDoor hinge left Optional
Door hinge right Standard
Wooden packaging for ocean transport
Standard
RCB 42 P
The RCB 42 P is a thermostabilizer designed for storing and transporting pre-cooled blood and medicinal prod-ucts over long periods of time. The refrigerating unit op-erates with nearly all types of electricity (AC / DC). The RCB 42 P ensures mobility and provides all the services of a blood bank refrigerator. The RCB 42 P provides huge storage capacity. It can contain up to fifty 270 ml whole blood bags and is standard-equipped with 2 plastic-lami-nated wire baskets that facilitate charging. The forced-air refrigeration system ensures a constant, even tempera-ture. A thermoelectric unit (Peltier block) together with an electronic control unit keeps the temperature of the blood constant. The thermoelectric unit is switched au-tomatically form cooling to heating, depending on the ambient temperature. The perfect insulation ensures the temperature in the cool box to be kept constant for up to 10 hours when transporting or storing items without power supply at an ambient temperature of +32°C. The temperature within the container can be thus maintained at a preset temperature of +1°C to +10°C regardless of external climatic conditions. This is even applicable at ambient temperatures of -32°C to +43°C. The internal temperature can be displayed by an optional tempera-ture recorder located in the container.
Performance RCB 42 P• Cooling Active (Peltier)• Gross volume 43 L• Storage capacity (blood bags)
30 x 500 ml50 x 270 ml
Numerous applications :
> Cars, HGVs, ships, (with conditions) helicopters and planes
> Integrated alarm system (visual / acoustic) with fault indicator port
> Long “hold over time / Cold Life” times (10 h at ambient temperature of +32°C) even without power supply, due to optimum insulation capa-bilities
> Function independent of inclination, cooling free from vibration, impervious to bumps, hardly any moving components, compact aggregate dimen-sions, low weight of cold system, both cooling and heating possible (Thermostabilizer), low-noise operation
-32°C
+43°C
Hold Over Time 10 h at +32°C
Operating temperature +4°C +/-2°C
Application range (at +25°C) -32°C to +43°C
DIMENSIONS / WEIGHTSOuter dimensions H x W x D (mm) 500 x 840 x 550
Inner dimensions H x W x D (mm) 240 x 460 x 335
Net weight - empty (kg) 33
Gross weight - fully stocked (kg) 44
TECHNICAL DETAILSOuter / Interior material Polyethylene
Material of interior container Stainless Steel
Insulation material and thickness PU foam / 105 mm
Color RAL 6031
Operating voltage range AC 90V to 264V (48Hz - 62Hz)115V / 400 Hz
DC 11V to 42V
CERTIFICATIONRadioshielding EMV according to MIL Standard
461 C & 462, 704 D & EC
Directive 2004 / 108 / EC
Low-voltage Directive 2006 / 95 / EC
ADR / RID Group II and III of class 6.1
Vibration test EC Directive DIN / IEC 68-2-6
Shock test EC Directive DIN / IEC 68-2-27
ACCESSORIES (standard)Plastic-coated wire baskets 2
AC connection cable 1
DC connection cable 1
Connecting cable remote alarm 1
* Accumulateurs de froid
ACT
IVE
SYST
EM I
RCB
/ M
TB
lood
Tra
nspo
rt B
oxes
ACT
IVE
SYST
EM I
RCB
/ M
TB
lood
Tra
nspo
rt B
oxes
Accessories (optional)
20 | 21 BLOOD CHAIN Your global partner to save lives.
fig. R
CB 8
38 mm
60 mm
LogTag temperature data logger TRID30-7R
> For continious, tamper-proof temperature recording
> With 30 days statistic display
> Temperature range -30°C to +60°C
LogTag temperature data logger TRID30-7FW
> Vaccine Refrigerator Temperature Recorder with 30 day summary display compliant to WHO PQS specification E006/TR06.3
LogTag temperature data logger TRED30-7R
> For continious, tamper-proof temperature recording
> With 30 days statistic display and connector for remote temperature control
> Temperature range -40°C to +99°C
LogTag Remote probe
> 1.5 m cable, 140 mm tipped sensor
LogTag USB interface cradle
> For reading out the data recordings via PC
Temperature Data LoggersFor perfect temperature control
Testo 174T mini data logger, 1-channel
> incl. wall bracket, battery (2 x CR 2032 lithium) and calibration protocol
> High data integrity, even with empty battery
> Large data memory
> Watertight in accordance with IP65
> Temperature range -30°C to +70°C
> Starter set including logger, interface and software also available
Testo 176 T2, 2-channel temperature logger
> with connections for highly accurate external sensor (Pt100) incl. wall holder, lock, battery and calibration protocol
> Large memory for 2 million measurement values - large, easily legible display up to 8 years battery life
> Standard battery (AA) replaceable by user
> SD card slot
> Temperature range -50°C to +400°C
Eutectic Cooling SystemFor RCB models during long-term, temperature controlled transport without power supply (e.g transport by air or rail)
The phase change material (PCM) allow a safe transport during all seasons. From winter to summer, the PCM have the same „melting point“ which is :
• -32°C : Transport of plasma
• +4°C : Transport of RBC
• +22°C : Transport of unscreened blood
• +37°C : Transport of whole blood
B Medical Systems‘s Eutectic Cooling Elements are available in 2 sizes : 0.3 l and 0.6 l
Please contact us for details.
Cooling Element PCM -32 PCM +4 PCM +22 PCM +37
Color Orange Blue Green Yellow
NominalTemperature -32°C +4°C +22°C +37°C
Application under -30°C +2°C to +8°C +15°C to +25°C over +35°C
Preconditioning Temperature / Time
-40°C / > 24 h +4°C / > 24 h +22°C / > 24 h +37°C / > 24 h
B Medical Systems‘s Eutectic Cooling Elements are heat accumulation elements, containing a so-called phase change material (PCM). The PCM stores latent heat at the required temperature at phase change (liquid / solid). The stored product will therefore remain at a near constant temperature for a specific period of time, without requiring active temperature control. The Eutectic Cooling Elements must be charged for the specified temperature prior to each use.
> The TCM models are also available in sand yellow
RAL 1002
RAL 6031
ACC
ESSO
RIE
SO
ptio
nal
ACC
ESSO
RIE
SO
ptio
nal
All r
ight
s res
erve
d. A
ll in
dica
tions
with
out g
uara
ntee
and
exc
ludi
ng a
ll lia
bilit
y du
e to
faul
ty, i
ncom
plet
e or
out
date
d in
form
ation
. All
tech
nica
l dat
a su
bjec
t to
chan
ge w
ithou
t prio
r noti
ce.
COLD CHAIN Comprehensive range of storage and active / passive transport systems for the storage and distribution of vaccines and in general of all temperature-sensitive preparations under various climatic and technical conditions.
For more information :
BIOMEDICAL REFRIGERATION Special purpose refrigerators and deep freezers, storage and transport concepts for Laboratory, Medicine, Research, Biotechnology and Life Sciences.
BLOOD SAFETY For a reliable cold chain within the scope of safety of blood transfusions.
BLOOD CHAIN Storage and active / passive transport systems dedicated to the military for the supply of blood, vaccines and medicines to the front line.
B Medical Systems S.à r.l 17, Op der Hei
L - 9809 Hosingen, Luxembourg
Tel. : + 352 92 07 31-1
Fax : + 352 92 07 31-300
www.bmedicalsystems.com
B Medical SystemsYour global partner to save lives.
www.bmedicalsystems.com