bone plug versus suture-only fixation of medial meniscus … · 2017-07-06 · bone plug versus...
TRANSCRIPT
Bone Plug versus Suture-Only Fixation of Medial Meniscus Allograft Transplants. Biomechanical study
Luiz Felipe Ambra1,2, Amy Phan2, Jack Farr3, Andreas H. Gomoll2.1Universidade Federal de São Paulo, 2Brigham and Women’s Hospital, 3OrthoIndy Hospital
1511283-Ap.jpg1511283-Lateral.jpg
The objective of this study was to evaluate the contact load
characteristics of medial meniscus allograft transplantation
performed by bone-plug versus soft-tissue suture-only fixation
in cadaveric knees.
Our null hypothesis was that there would not be significant
differences between the two fixation groups with regard to load
contact pressures and distribution.
Objetivo
Material and Methods
Meniscal allograft transplantation is a suitable option to treat
meniscal insufficiency. Restoring anatomical reconstruction is
critical to recover the proper function of the meniscus in its
chondroprotective and stability role. In this regard, anatomic
positioning and fixation of the attachment sites are
appreciated to be of critical importance. For medial meniscus
transplant two different techniques has been describe: one two
bone plugs are routinely utilized, one for each meniscal root
is placed within bone tunnels in the anatomical footprint of
the medial meniscus. An alternative technique is to remove
the attachment of the meniscal root from bone graft and use
suture only fixation through drill holes in the tibia. There is
debate on whether the suture only fixation technique
reproduces the same rigidity in fixation as compared to the
bone plug technique.
Background
Table 1. Cadaveric informationKnees (N) 9
Side (left/right) 5 4
Gender (Male/Female) 5 4
Age (mean/±SD) [year] 46.9 2.2
Weight (mean/±SD) [Kg] 76.3 26.3
Height (mean/±SD) [m] 1.57 0.17
Study design: controlled laboratory study.
SpecimenWere used fresh-frozen human cadaveric knee. All the
specimens was evaluated using MRI and XR. Only cadaveric
specimens without evidence of significant degenerative
changes were included in the study.
Meniscal Sizing
Size- and side-matched medial meniscus allografts were
obtained based on XR measurements (Pollard method).
The mismatch between the meniscal allograft size and
radiographic measurement did no exceed 1mm.
Figure 1. Meniscal measurement by pollard method
The mismatch between the meniscal allograft size and
radiographic measurement did no exceed 1mm. Figure 2
demonstrated the a good match size between native meniscus
and the respective graft.
Figure 2. The meniscus native after meniscectomy (right), and respective meniscal allograft (left).
Testing Equipment
Compression force: • INSTRON 8871 servo-hydraulic testing system.
Contact pressure: • A Model 4011 I-Scan sensor (Tekscan Inc, Boston, Mass)
Bone-plug Fixation
Soft tissue only
Meniscal Transplantation preparation technique
Allograft medial meniscus were prepared by fashioning bone
plugs (10x10mm, wide x height) of the anterior and posterior
root attachments. Drill holes in the center of the bone plugs
were made with a 1.2mm drill bit. Heavy suture was run
through the bone plug and secured to the meniscal root in a
locking stitch configuration. The bone plugs were pulled in the
bone tunnels created in the anatomic location of the footprint of
the anterior and posterior horn attachment to the tibia. The
suture were tied over a bone bridge in the anterior tibia. The
meniscal horns and body will be sutured using vertical
mattresses and tied over the capsule
The meniscal horns were detached from the bone plugs,
preserving the suture which were pulled from the bone plug
prior to detachment. The bone plugs were secured in the bone
tunnels with bone cement. The suture attached to the
posterior and anterior horns were pulled through the drill
holes and the suture tied over the anterior tibia. The meniscal
horns and body were sutured using vertical mattresses and
tied over the capsule
The tibia and femur were individually potted in methyl
methacrylate cement. Then the knee was fixed into an
Instron with the tibiofemoral joint surface oriented parallel to
the floor. The knee was constrained but allowing for varus/
valgus angulations.
Contact sensor was inserted underneath the meniscus from
anterior to posterior with minimal disruption of the
meniscocapsular attachments.
Figure 3: Cadaveric knee fixed to the testing apparatus
All the specimens was tested in the following sequence:
Native Meniscec-tomized
Bone-plug
Suture-only
Testing protocol:
• Axial Load: 700N
• Knee Flexion: 0, 30 and 60 degrees.
Results
Data Acquisition
• Mean contact stress on the medial compartment (Mpa)
• Peak contact stress on the medial compartment (Mpa)
• Contact area on the medial compartment (mm2)
The results of mean and peak contact stress and contact area
were compared between intact, meniscectomized, MAT-bone
and MAT-suture using analysis of variance (Dunn test). The
level of significance was set at p ≤ 0.05.
0
100
200
300
400
500
600
0degrees 30degrees 60degrees
[mm
2 ]
ContactArea
Normal Menicectomized Bone-plug Suture-only
**
** *
Meniscectomy significantly increased mean contact pressure
and the contact area was significant reduced in comparison
with the native condition at 0, 30 and 60 degrees (p<0.0001).
Although the peak contact pressure was higher in the
meniscectomy group, it was not statistically significant. Suture
only resulted in higher contact pressure and smaller contact
area than bone plug fixation and native condition. The
difference was statistically significant in 0 and 30 degrees of
flexion. Regardless of technique, no significant differences
were found between transplanted meniscus and native
condition in 60 degrees of flexion.
The following graphs shows the mean contact area, mean
contact pressure and mean peak pressure in each group in 0, 30
and 60 degrees of flexion.
0
0.5
1
1.5
2
2.5
3
3.5
4
0degrees 30degrees 60degrees
[Mpa]
MeanContactPressure
Normal Menicectomized Bone-plug Suture-only
0
1
2
3
4
5
6
7
8
0degrees 30degrees 60degrees
[Mpa]
MeanPeakContactPressure
Normal Menicectomized Bone-plug Suture-only
*
**
* *
*Difference statistically significant (p<0.05) in comparison with native condition.
Native Menisectomized Bone-plug Suture-only
Figure 4: Contact pressure maps on the medial tibia plateau for a typical knee under different meniscal conditions at 0 degrees of flexion.
Conclusion
This study demonstrated that meniscal allograft transplantation
can restore load parameters close to the native condition. Bone-
plug technique has demonstrated improved femorotibial
contact pressure than isolate soft-tissue fixation. This study is
unique in utilizing size-matched meniscal transplants and using
the same transplant for bone plug and suture-only fixation, thus
reducing confounding factors.