Brain and Spine Treatment Planning

Ryan Foster, PhDAssistant Professor

Director of Clinical Medical PhysicsDepartment of Radiation OncologyUT Southwestern Medical Center

Dallas, TX

Page 2: Brain and Spine Treatment Planning - AAPM: The American ... · Brain and Spine Treatment Planning Ryan Foster, PhD Assistant Professor Director of Clinical Medical Physics Department

Acknowledgements• Zeke Ramirez, CMD, M.S.

Page 3: Brain and Spine Treatment Planning - AAPM: The American ... · Brain and Spine Treatment Planning Ryan Foster, PhD Assistant Professor Director of Clinical Medical Physics Department

Disclosures• I receive research funding from the Cancer

Prevention and Research Institute of Texas.

Page 4: Brain and Spine Treatment Planning - AAPM: The American ... · Brain and Spine Treatment Planning Ryan Foster, PhD Assistant Professor Director of Clinical Medical Physics Department

Overview• Cranial planning and delivery options

– GammaKnife– CyberKnife– Linac

• Spine planning

Page 5: Brain and Spine Treatment Planning - AAPM: The American ... · Brain and Spine Treatment Planning Ryan Foster, PhD Assistant Professor Director of Clinical Medical Physics Department

Learning Objectives• Identify the available delivery techniques for

cranial and spine SRS/SRT/SBRT• Understand the advantages and disadvantages

of the various techniques• Identify the best delivery technique to use for a

given lesion

Page 6: Brain and Spine Treatment Planning - AAPM: The American ... · Brain and Spine Treatment Planning Ryan Foster, PhD Assistant Professor Director of Clinical Medical Physics Department

Options for Cranial Planning• GammaKnife• CyberKnife• Linac

– Circular collimators– Dynamic conformal arcs– Conformal beams– IMRT

Page 7: Brain and Spine Treatment Planning - AAPM: The American ... · Brain and Spine Treatment Planning Ryan Foster, PhD Assistant Professor Director of Clinical Medical Physics Department

Basics of GammaKnife Planning• Small spherical lesions – straightforward

– Pick collimator size that will cover target• Larger or irregular targets – more complicated

– May require several shots– Determine number, size, location and weight of shots– Typically prescribed to 50% isodose line

Page 8: Brain and Spine Treatment Planning - AAPM: The American ... · Brain and Spine Treatment Planning Ryan Foster, PhD Assistant Professor Director of Clinical Medical Physics Department

Pituitary Adenoma on GammaKnife

• 35 Gy at 50%• 5 shots• 4 mm collimator

Page 9: Brain and Spine Treatment Planning - AAPM: The American ... · Brain and Spine Treatment Planning Ryan Foster, PhD Assistant Professor Director of Clinical Medical Physics Department

Meningioma on GammaKnife

• 15 Gy at 50%• 34 shots• 4,8 mm collimators

Page 10: Brain and Spine Treatment Planning - AAPM: The American ... · Brain and Spine Treatment Planning Ryan Foster, PhD Assistant Professor Director of Clinical Medical Physics Department

Brain Mets on GammaKnife

• 18 Gy at 50%• 15 shots• 4,8,16 mm collimators

Page 11: Brain and Spine Treatment Planning - AAPM: The American ... · Brain and Spine Treatment Planning Ryan Foster, PhD Assistant Professor Director of Clinical Medical Physics Department

Multiple Brain Mets on GammaKnife

• 16 mets• 16 or 18 Gy at 50%• 77 shots• ~6.5 hrs• 4,8,16 mm collimators

Page 12: Brain and Spine Treatment Planning - AAPM: The American ... · Brain and Spine Treatment Planning Ryan Foster, PhD Assistant Professor Director of Clinical Medical Physics Department

Acoustic Neuroma on GammaKnife

• 13 Gy at 50%• 10 shots• 4, 8 mm collimators

Page 13: Brain and Spine Treatment Planning - AAPM: The American ... · Brain and Spine Treatment Planning Ryan Foster, PhD Assistant Professor Director of Clinical Medical Physics Department

Trigeminal Neuralgia on GammaKnife

• 85 Gy at 100%• 1 shot• 4 mm collimator

Page 14: Brain and Spine Treatment Planning - AAPM: The American ... · Brain and Spine Treatment Planning Ryan Foster, PhD Assistant Professor Director of Clinical Medical Physics Department

Re-treatment (AVM) on GammaKnife

• AVM – 2011• 18 Gy at 50%• 15 shots• 8 mm collimator• Re-treat 2014• 20 Gy at 50%• 6 shots• 4, 8 mm collimator

Page 15: Brain and Spine Treatment Planning - AAPM: The American ... · Brain and Spine Treatment Planning Ryan Foster, PhD Assistant Professor Director of Clinical Medical Physics Department

Basics of CyberKnife Planning• Fixed cones, IRIS or InCise MLC• Radiation delivered with linac at fixed points in

space called nodes• ~ 100 nodes per fraction• Nodes arranged in spherical (cranial) or

ellipsoidal (extra-cranial) patterns• Forward or inverse planning• Prescribed to 60 - 90% isodose line

Page 16: Brain and Spine Treatment Planning - AAPM: The American ... · Brain and Spine Treatment Planning Ryan Foster, PhD Assistant Professor Director of Clinical Medical Physics Department

SRT on CyberKnife

• Pituitary adenoma• 45 Gy in 25 fx

Page 17: Brain and Spine Treatment Planning - AAPM: The American ... · Brain and Spine Treatment Planning Ryan Foster, PhD Assistant Professor Director of Clinical Medical Physics Department

Staged AVM on CyberKnife

• Large AVM• Too big for single fx

ablative dose• ~ 30 cc• 11 patients• 3-8 stages, 16-20 Gy• Inverse planning better• Patients in follow-up

Ding et al. Radiotherapy and Oncology 109, p. 452.

Page 18: Brain and Spine Treatment Planning - AAPM: The American ... · Brain and Spine Treatment Planning Ryan Foster, PhD Assistant Professor Director of Clinical Medical Physics Department

Trigeminal Neuralgia on CyberKnife

• 66 Gy to 80%• 7.5 mm collimator

Tang et al. J Clin Neurosci 18, 2011.

Page 19: Brain and Spine Treatment Planning - AAPM: The American ... · Brain and Spine Treatment Planning Ryan Foster, PhD Assistant Professor Director of Clinical Medical Physics Department

Linac Cranial SRS Techniques

• Circular arcs• Dynamic arc• Conformal beams• IMRT• How do you decide which technique to use?

Page 20: Brain and Spine Treatment Planning - AAPM: The American ... · Brain and Spine Treatment Planning Ryan Foster, PhD Assistant Professor Director of Clinical Medical Physics Department

Linac Cranial SRS Techniques

• It depends on…• Size of target• Shape of target• Location of target• Do a comparison…

Page 21: Brain and Spine Treatment Planning - AAPM: The American ... · Brain and Spine Treatment Planning Ryan Foster, PhD Assistant Professor Director of Clinical Medical Physics Department

Linac Cranial SRS Techniques

• Plan with circular arcs, conformal beams and dynamic arcs

This image cannot currently be displayed.

Solberg et al. IJROBP 49, No. 5, 2001.

Page 22: Brain and Spine Treatment Planning - AAPM: The American ... · Brain and Spine Treatment Planning Ryan Foster, PhD Assistant Professor Director of Clinical Medical Physics Department

Case 1

Solberg et al. IJROBP 49, No. 5, 2001.

Page 23: Brain and Spine Treatment Planning - AAPM: The American ... · Brain and Spine Treatment Planning Ryan Foster, PhD Assistant Professor Director of Clinical Medical Physics Department

Case 2

Solberg et al. IJROBP 49, No. 5, 2001.

Page 24: Brain and Spine Treatment Planning - AAPM: The American ... · Brain and Spine Treatment Planning Ryan Foster, PhD Assistant Professor Director of Clinical Medical Physics Department

According to Solberg et al, which linac-based SRS technique produces the most

heterogeneous dose in the tumor?

4%3%8%7%78% 1. Multi-iso circular arcs

2. Single iso circular arcs3. Conformal arcs4. Static Conformal beams5. IMRT

Page 25: Brain and Spine Treatment Planning - AAPM: The American ... · Brain and Spine Treatment Planning Ryan Foster, PhD Assistant Professor Director of Clinical Medical Physics Department

Answer

1. Multi-iso circular arcsRef. - Solberg et al.

IJROBP 49, No. 5, 2001

Page 26: Brain and Spine Treatment Planning - AAPM: The American ... · Brain and Spine Treatment Planning Ryan Foster, PhD Assistant Professor Director of Clinical Medical Physics Department

Trigeminal Neuralgia on Linac• 70 – 90 Gy at 100%• 30% IDL touches

brainstem• 5 or 7.5 mm cones• Circular arc delivery

Smith, et al. J Neurosurg 99, 2003.

Figure from BrainLab clinical white paper

Page 27: Brain and Spine Treatment Planning - AAPM: The American ... · Brain and Spine Treatment Planning Ryan Foster, PhD Assistant Professor Director of Clinical Medical Physics Department

Cavernous Sinus Meningioma SRT on Linac

• ~10 conformal noncoplanar beams

• Median 5040 cGy at 180 cGy/fx to ~90%

Selch, et al. IJROBP 59, No. 1, 2004.

Page 28: Brain and Spine Treatment Planning - AAPM: The American ... · Brain and Spine Treatment Planning Ryan Foster, PhD Assistant Professor Director of Clinical Medical Physics Department

IMRT for SRS/SRT on Linac

Benedict, et al. IJROBP 50, No. 3, 2001

Page 29: Brain and Spine Treatment Planning - AAPM: The American ... · Brain and Spine Treatment Planning Ryan Foster, PhD Assistant Professor Director of Clinical Medical Physics Department

Linac Cranial SRS Techniques

• Circular arcs – best for small spherical lesions, TN• Dynamic arc – most frequently used technique• Conformal beams – avoid critical structures• IMRT – necessary to avoid critical structures or for

highly irregularly shaped lesions

Page 30: Brain and Spine Treatment Planning - AAPM: The American ... · Brain and Spine Treatment Planning Ryan Foster, PhD Assistant Professor Director of Clinical Medical Physics Department

Which linac-based SRS delivery technique is best for irregularly shaped

lesions?

2%0%89%9%1% 1. Circular arcs

2. Conformal arcs3. IMRT4. 2D planning5. Conformal beams

Page 31: Brain and Spine Treatment Planning - AAPM: The American ... · Brain and Spine Treatment Planning Ryan Foster, PhD Assistant Professor Director of Clinical Medical Physics Department

Answer3. IMRTRef: Benedict et al., IJROBP 50, No. 3, 2001.

Page 32: Brain and Spine Treatment Planning - AAPM: The American ... · Brain and Spine Treatment Planning Ryan Foster, PhD Assistant Professor Director of Clinical Medical Physics Department

Spine SBRT/SRS Treatment Planning• MR used to delineate spinal cord• Fuse MR and CT carefully• Cord contoured 2.5 cm above and below PTV• Cord constraints must be met!• Typically 10 – 15 non-opposing coplanar beams• Almost always IMRT

Page 33: Brain and Spine Treatment Planning - AAPM: The American ... · Brain and Spine Treatment Planning Ryan Foster, PhD Assistant Professor Director of Clinical Medical Physics Department

SBRT Spine Planning

Page 34: Brain and Spine Treatment Planning - AAPM: The American ... · Brain and Spine Treatment Planning Ryan Foster, PhD Assistant Professor Director of Clinical Medical Physics Department

Image Fusion

Page 35: Brain and Spine Treatment Planning - AAPM: The American ... · Brain and Spine Treatment Planning Ryan Foster, PhD Assistant Professor Director of Clinical Medical Physics Department

Image Fusion

Page 36: Brain and Spine Treatment Planning - AAPM: The American ... · Brain and Spine Treatment Planning Ryan Foster, PhD Assistant Professor Director of Clinical Medical Physics Department

RTOG 0631 Spine SBRT Protocol• Prescription

– 16 or 18 Gy Spine SBRT in 1 fx– 8 Gy EBRT in 1 fx

• Target Coverage– 90% coverage of prescription dose for target

• Target volumes

No margin added to these volumes!

Page 37: Brain and Spine Treatment Planning - AAPM: The American ... · Brain and Spine Treatment Planning Ryan Foster, PhD Assistant Professor Director of Clinical Medical Physics Department

RTOG 0631 Spine SBRT Protocol• Cord constraints

– 10 Gy < 10% of the partial cord– 10 Gy < 0.35 cc of the conventional cord– 14 Gy < 0.03 cc of the conventional cord

Page 38: Brain and Spine Treatment Planning - AAPM: The American ... · Brain and Spine Treatment Planning Ryan Foster, PhD Assistant Professor Director of Clinical Medical Physics Department
Page 39: Brain and Spine Treatment Planning - AAPM: The American ... · Brain and Spine Treatment Planning Ryan Foster, PhD Assistant Professor Director of Clinical Medical Physics Department

According to RTOG 0631, the margin added to the target volume is

0%3%7%8%81% 1. 0 mm

2. 1 mm3. 3 mm4. 5 mm5. 10 mm

Page 40: Brain and Spine Treatment Planning - AAPM: The American ... · Brain and Spine Treatment Planning Ryan Foster, PhD Assistant Professor Director of Clinical Medical Physics Department

Answer1. 0 mm

Page 41: Brain and Spine Treatment Planning - AAPM: The American ... · Brain and Spine Treatment Planning Ryan Foster, PhD Assistant Professor Director of Clinical Medical Physics Department

UTSW Spine SBRT Protocol• Prescription

– CTV 14Gy and GTV 20Gy X 1 fraction– Patients with prior radiation receive 5 fraction

regimen.• Target Coverage

– 95% coverage of prescription dose for CTV.– 90% coverage of prescription dose for GTV.

• Spinal Cord Dose Constraints:<.035 cc 14 Gy<.25 cc 10 Gy<.5 cc 7 Gy

Page 42: Brain and Spine Treatment Planning - AAPM: The American ... · Brain and Spine Treatment Planning Ryan Foster, PhD Assistant Professor Director of Clinical Medical Physics Department

UTSW Spine SBRT Protocol Re-irradiation• Same total dose, but in 4 fractions• 2.8 Gy/fx to CTV, 4 Gy/fx to GTV• Cord constraints same as single fraction

<.035 cc 14 Gy<.25 cc 10 Gy<.5 cc 7 Gy

Page 43: Brain and Spine Treatment Planning - AAPM: The American ... · Brain and Spine Treatment Planning Ryan Foster, PhD Assistant Professor Director of Clinical Medical Physics Department

Spine Treatment Planning• GTV - defines

lesion.• CTV - expansion

to include contiguous bone marrow adjacent to tumor.

• No expansion for PTV.

• Target does notoverlap cord GTV Spinal Cord CTV

Page 44: Brain and Spine Treatment Planning - AAPM: The American ... · Brain and Spine Treatment Planning Ryan Foster, PhD Assistant Professor Director of Clinical Medical Physics Department
Page 45: Brain and Spine Treatment Planning - AAPM: The American ... · Brain and Spine Treatment Planning Ryan Foster, PhD Assistant Professor Director of Clinical Medical Physics Department

Multiple Lesions

Page 46: Brain and Spine Treatment Planning - AAPM: The American ... · Brain and Spine Treatment Planning Ryan Foster, PhD Assistant Professor Director of Clinical Medical Physics Department

Spine SRS/SBRT on CyberKnife

Heron et al. J Neurosurg Spine 17, 2012.

Page 47: Brain and Spine Treatment Planning - AAPM: The American ... · Brain and Spine Treatment Planning Ryan Foster, PhD Assistant Professor Director of Clinical Medical Physics Department

Summary | Conclusion• Several options are available for cranial and

spine SRS/SBRT • Cranial plans and delivery techniques are

determined by the target size, shape and location• Spine SRS/SBRT is almost always IMRT

Page 48: Brain and Spine Treatment Planning - AAPM: The American ... · Brain and Spine Treatment Planning Ryan Foster, PhD Assistant Professor Director of Clinical Medical Physics Department

W.A. “Monty” and Tex Moncrief Radiation Oncology Building

Children’s Hospital

Zale LipshyHospital

Parkland Hospital

St Paul University Hospital

Annette Simmons Stereotactic Treatment

Page 49: Brain and Spine Treatment Planning - AAPM: The American ... · Brain and Spine Treatment Planning Ryan Foster, PhD Assistant Professor Director of Clinical Medical Physics Department

Radiation Oncology at UTSW

Page 50: Brain and Spine Treatment Planning - AAPM: The American ... · Brain and Spine Treatment Planning Ryan Foster, PhD Assistant Professor Director of Clinical Medical Physics Department

SRS/SBRT at UTSW• GammaKnife – cranial SRS, tumors < 4 cm• CyberKnife – cranial SRT, some SBRT (breast)• Linac – All other SBRT

Page 51: Brain and Spine Treatment Planning - AAPM: The American ... · Brain and Spine Treatment Planning Ryan Foster, PhD Assistant Professor Director of Clinical Medical Physics Department

IROC SRS Head Phantom

• CT or MR imaging• Target delineation• TLDs and film

Page 52: Brain and Spine Treatment Planning - AAPM: The American ... · Brain and Spine Treatment Planning Ryan Foster, PhD Assistant Professor Director of Clinical Medical Physics Department

IROC Spine Phantom

Page 53: Brain and Spine Treatment Planning - AAPM: The American ... · Brain and Spine Treatment Planning Ryan Foster, PhD Assistant Professor Director of Clinical Medical Physics Department
Page 54: Brain and Spine Treatment Planning - AAPM: The American ... · Brain and Spine Treatment Planning Ryan Foster, PhD Assistant Professor Director of Clinical Medical Physics Department

Dynamic Arc

Page 55: Brain and Spine Treatment Planning - AAPM: The American ... · Brain and Spine Treatment Planning Ryan Foster, PhD Assistant Professor Director of Clinical Medical Physics Department

Circular Cones

http://www.aktina.com/product/small-field-circular-cones-system-for-siemens/