broadband phase correction of fourier transform ion cyclotron mass spectra

157
Florida State University Libraries Electronic Theses, Treatises and Dissertations The Graduate School 2012 Broadband Phase Correction of Fourier Transform Ion Cyclotron Resanonce Mass Spectra Feng Xian Follow this and additional works at the FSU Digital Library. For more information, please contact [email protected]

Upload: others

Post on 11-Sep-2021

3 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: Broadband Phase Correction Of Fourier Transform Ion Cyclotron Mass Spectra

Florida State University Libraries

Electronic Theses, Treatises and Dissertations The Graduate School

2012

Broadband Phase Correction of FourierTransform Ion Cyclotron Resanonce MassSpectraFeng Xian

Follow this and additional works at the FSU Digital Library. For more information, please contact [email protected]

Page 2: Broadband Phase Correction Of Fourier Transform Ion Cyclotron Mass Spectra

THE FLORIDA STATE UNIVERSITY

COLLEGE OF ARTS AND SCIENCES

BROADBAND PHASE CORRECTION OF FOURIER TRANSFORM ION CYCLOTRON

RESANONCE MASS SPECTRA

By

FENG XIAN

A Dissertation submitted to the Department of Chemistry and Biochemistry

in partial fulfillment of the requirements for the degree of

Doctor of Philosophy

Degree Awarded: Spring Semester, 2012

Page 3: Broadband Phase Correction Of Fourier Transform Ion Cyclotron Mass Spectra

ii

Feng Xian defended this dissertation on March 30, 2012.

The members of the supervisory committee were:

Alan G. Marshall Professor Directing Dissertation Christopher L. Hendrickson Professor Co-Directing Dissertation Stephen Hill University Representative Naresh S. Dalal Committee Member Michael Roper Committee Member

The Grduate School has verified and approved the above-named committee members, and certifies that the dissertation has been approved in accordance with university requirements.

Page 4: Broadband Phase Correction Of Fourier Transform Ion Cyclotron Mass Spectra

iii

To my parents, Huaiyun Xian and Mingxiang Gao; my wife, Donghong Min; my daughter,

Stephanie Xian, my son, Anthony Xian.

Page 5: Broadband Phase Correction Of Fourier Transform Ion Cyclotron Mass Spectra

iv

ACKNOWLEDGEMENTS

I am sincerely and heartily grateful to my supervisor, Dr. Alan G. Marshall, for his

excellent guidance, caring and support. Dr. Marshall gives me valuable suggestions on how to

become a successful scientist and provides me with a scientific environment for doing research. I

am sure that it would have not been possible without his patient and intelligent advising.

I would also like to thank Dr. Stephen Hill, Dr. Naresh S. Dalal and Dr. Michael Roper

for their inspiring suggestions, their time and effort to serve on my examination committee.

I would also like to thank Dr. Chris Hendrickson for teaching me basically everything

that I know about mass spectrometry. His valuable suggestions and helpful discussions are very

important part in my research. I also thank Dr. Steve Beu for his expertise in the fields of FT-

ICR instrumentation and Dr. Greg Blakney for technical support in software development.

I would like to thank all the former and current members in Dr. Marshall lab for their

help and a great time we spent together. In particular, I want to thank Dr. Huan He, Dr. Amy

McKenna, Dr. Nathan Kaiser, and Dr. Joshua Savory.

I would also like to thank my parents and two elder sisters. They were always supporting

me and encouraging me with their best wishes.

Finally, I would like to thank my wife, sweet daughter and son, for their firm and

continuous love and support.

This work was supported by NSF Division of Materials Research through DMR-0654118

and the State of Florida.

Page 6: Broadband Phase Correction Of Fourier Transform Ion Cyclotron Mass Spectra

v

TABLE OF CONTENTS

List of Figures .............................................................................................................................. viii

Abstract ........................................................................................................................................ xiv

1. HIGH RESOLUTION MASS SPECTROMETRY............................................... 1 What Defines High Resolution and High Mass Accuracy………………………………... 1 Mass Resolution and Accuracy………………………………………………………….... 2 Time-of-Flight Mass Analyzers…………………………………………………………... 3

Orthogonal Acceleration………………………………………………………….. 3 Reflectron/Multipass TOF………………………………………………………... 5 Recent Advances in TOF Mass Analyzer……………………………………….... 6 Selected Applications………………………………………………………….......7

Fourier Transform Mass Analyzers…………………………………………………......... 8 Common Features of Fourier Transform Mass Analyzers………………………...9 Ion Accumulation and Detection………………………………………………....10 Advances in Fourier Transform Mass Analyzers ...……………………………...10

SelectedApplications………………………………………………...............….. 17

2. AUTOMATED BROADBAND PHASE CORRECTION OF FOURIER TRANSFORM ION CYCLOTRON RESONANCE MASS SPECTRA……….... 24 Introduction…………………………………………………………………………....... 24 Problem………………………………………………………………………….. 24 Prior Solutions…………………………………………………………............... 27 Experimental Methods………………………………………………………………….. 29 Sample Description and Preparation…………………………………………….. 29 Instrumentation: 9.4 Tesla FT-ICR MS……………………………………….... 29 Mass Calibration................................................................................................... 30 Computational Method…………………………………………………………. 31 Choosing the Best Phasing Parameters…………………………………………. 32 Baseline Correction……………………………………………………………. 32 Computational Implementation............................................................................ 33 Results and Discussion..................................................................................................... 35 Absorption-Mode vs. Magnitude-Mode Spectral Display.................................... 35 Mass Accuracy..................................................................................................... 35 Filling Gaps Compositional Assignment............................................................... 37 Baseline Roll and Automated Peak Picking.......................................................... 39 Conclusions....................................................................................................................... 40

Page 7: Broadband Phase Correction Of Fourier Transform Ion Cyclotron Mass Spectra

vi

3. BASELINE CORRECTION OF ABSORPTION-MODE FOURIER TRANSFORM ION CYCLOTRON MASS SPECTRA......................................... 41 Introduction........................................................................................................................ 41 Experimental Methods....................................................................................................... 43 Simulation.............................................................................................................. 43 Sample Description and Preparation...................................................................... 45 APPI Source........................................................................................................... 45 9.4 Tesla FT-ICR MS............................................................................................ 46 Mass Calibration.................................................................................................... 46 Baseline Correction Algorithm.............................................................................. 46 Computational Implementation..............................................................................50 Results and Discussion.......................................................................................................50 Mass Accuracy....................................................................................................... 50 Identified Peaks...................................................................................................... 51 Isotopic Distribution...............................................................................................55 4. EFFECTS OF ZERO-FILLING AND APODIZATION ON FOURIER TRANSFORM ION CYCLOTRON RESONANCE MASS SPECTRAL ACCURACY, RESOLUTION, AND SIGNAL-TO-NOISE RATIO.................... 56 Introduction....................................................................................................................... 56 Experimental Methods...................................................................................................... 57 Sample Description and Preparation..................................................................... 57 Instrumentation...................................................................................................... 57 Mass Calibration.................................................................................................... 58 Data Processing..................................................................................................... 58 Computational Implementation............................................................................. 58 Results and Discussion...................................................................................................... 61 Full Apodization vs. Half Apodization.................................................................. 61 Mass Accuracy....................................................................................................... 65 Signal-to-Noise Ratio............................................................................................. 66 Resolving Power.................................................................................................... 66 5. PHASE SPECTRA OF FOURIER TRNSFORM ION CYCLOTRON RESONANCE MASS SPECTROMETRY............................................................. 67 Introduction........................................................................................................................ 67 Method of Stationary phase............................................................................................... 68 Phase Spectrum.................................................................................................................. 69 Chirp/Sweep Excitation Signal.............................................................................. 69 Detected Signal...................................................................................................... 72 Experimental Methods....................................................................................................... 77 Sample Preparation................................................................................................ 77 Instrumentation...................................................................................................... 77 Mass Analysis........................................................................................................ 77 SWIFT Waveform..................................................................................................79 Computational Method...........................................................................................79

Page 8: Broadband Phase Correction Of Fourier Transform Ion Cyclotron Mass Spectra

vii

Results and Discussion...................................................................................................... 79 Phase Spectrum Method vs. Automated Broadband Phase Correction................. 80 Resolving Power and Mass Accuracy........................................................ 80 Unresolved Peaks....................................................................................... 80 SWIFT Phasing by Phase Spectrum Method......................................................... 85 Modified Phase Spectrum for SWIFT....................................................... 85 Performances..............................................................................................85 6. PHASE CORRECTION OF FOURIER TRANSFORM CYCLOTRON RESONANCE MASS SPECTRA BY SIMUTANEOUS EXCITATION AND DETECTION.......................................................................................................... 87 Introduction...................................................................................................................... 87 Phase Spectrum and Phase Correction................................................................. 87 Experimental Methods..................................................................................................... 90 Sample Preparation............................................................................................... 90 Instrumentation..................................................................................................... 90 Mass Analysis....................................................................................................... 92 SWIFT Waveform Design.................................................................................... 93 SED Experiments.................................................................................................. 93 Computational Method......................................................................................... 94 Results and Discussion...................................................................................................... 94 SED vs. Automated Broadband Phase Correction................................................ 94 Absorption Spectra vs. Magnitude Spectrum form Normal SWIFT Excitation... 97 Absorption Spectra from stepped SWIFT Excitation........................................... 97 Advantages of Broadband Phase Correction by SED.......................................... 97 7. ARTIFACTS INDUCED BY SELECTIVE BLANKING OF TIME-DOMAIN DATA IN FOURIER TRANSFORM MASS SPECTROMETRY....................... 102 Introduction..................................................................................................................... 102 Materials and Methods.................................................................................................... 103 Materials.............................................................................................................. 103 Experiments......................................................................................................... 103 Simulations.......................................................................................................... 104 Estimation of Relaxation of Time ..................................................................... 104 Results and Discussion.................................................................................................... 106 Spectral Profile without Blanking....................................................................... 106 Spectral Profile after Blanking............................................................................ 106 Effect of Noise.................................................................................................... 108 Conclusion....................................................................................................................... 109 REFERENCES...................................................................................................... 111 BIOGRAPHICAL SKETCH................................................................................ 137

Page 9: Broadband Phase Correction Of Fourier Transform Ion Cyclotron Mass Spectra

viii

LIST OF FIGURES

Figure 1.1. High resolution time of flight mass analyzers. Top: Dual stage reflectron. Middle: Multi-pass reflectron with linear segments. Bottom: Multi-pass spiral configuration. ...................4 Figure 1.2. Standard orbitrap (left) and compact high field orbitrap (right) mass analyzers ........11 Figure 1.3. Top: Seven segment compensated ICR cell. Bottom: Dynamically harmonized ICR cell (bottom)...................................................................................................................................12

Figure 1.4. Top: Magnitude-mode and absorption-mode Lorentzian spectra, corresponding to Fourier transformation of an infinite duration time-domain signal, f(t) = A cos(ω0t) exp(-t/ ). Bottom: Magnitude (upper) and absorption-mode (lower) electrospray ionization 9.4 T FT-ICR mass spectra from the same bitumen time-domain data. The absorption display clearly resolves a mass doublet (compositions differing by C3 vs. SH4, 0.0034 Da) unresolved in magnitude mode……………………………………………………………………………………………...16 Figure 1.5. Positive electrospray ionization 14.5 T FT-ICR mass spectrum for Nannochloropsis oculata species lipid extract. Inset: Two peaks differing by 2.37 mDa are resolved and assigned…………………………………………………………………………………………..19

Figure 1.6. Positive ion atmospheric pressure photoionization (APPI) 9.4 T Fourier transform ion cyclotron resonance (FT-ICR) mass spectrum of a petroleum crude oil. Upper inset: Mass scale expansion revealing 90 singly-charged ions within a mass range of 0.32 Da. Lower inset: Baseline resolution of ions differing in mass by 1.1 mDa, made possible by ultrahigh mass resolving power of 1,000,000 at m/z 428……………………………………........................... ...22

Figure 2.1. Absorption, dispersion, and magnitude-mode Lorentzian spectra, corresponding to Fourier transformation of an infinite duration time-domain signal, f(t) = A cos(ω0t) exp(-t/ ).....26 Figure 2.2. Plots of time-domain linear frequency-sweep excitation signal (top) and instantaneous excitation frequency (bottom) vs. time, showing the time elapsed following the instant that the excitation frequency matches the ion cyclotron resonance frequency for ions of each of three different m/z values. The corresponding accumulated phase at ωi-1, ωi, or ωi+1 at the onset of detection is ωi-1 (ti-1 + tdelay), ωi (ti + tdelay), or ωi+1 (ti+1 + tdelay) Here the frequency-sweep is from low to high frequency, but a similar argument applies for sweeping from high to low frequency…………………………………………………………………………………….30 Figure 2.3. Schematic baseline flattening procedure. (A) Original absorption spectrum; (B) Fourier transform of (A); (C) Rectangular weight function to remove high-"frequency" components to yield (D); (E) Inverse Fourier transform of (D) to yield the low-"frequency"

Page 10: Broadband Phase Correction Of Fourier Transform Ion Cyclotron Mass Spectra

ix

spectral baseline with true mass spectral peaks removed; (F) Baseline-flattened spectrum produced by subtracting (E) from (A).......................................................................................... 33 Figure 2.4. Electrospray ionization 9.4 T FT-ICR mass spectra. Top: Raw real data following Fourier transform of discrete time-domain signal. Middle: Magnitude-mode spectrum (obtained from Eq. 1.1a). Bottom: Absorption-mode spectrum. The resolving power for the absorption-mode display is equivalent to that for magnitude-mode at 13.6 Tesla. Note also higher mass accuracy for absorption-mode relative to magnitude-mode display.............................................. 34

Figure 2.5. Mass error distribution for magnitude (top) and absorption (bottom) electrospray ionization 9.4 T FT-ICR mass spectra for a vacuum gas oil. Each bar represents the number of assigned masses within a 50 ppb mass error range. The same relative signal abundance threshold (peak height > 5 of baseline noise) was used for peak picking. The magnitude spectrum was produced with one zero fill and Hanning apodization1 and the absorption after one zero fill and half Hanning apodization.............................................................................................................. 36 Figure 2.6. Magnitude and absorption electrospray ionization 9.4 T FT-ICR mass spectra for the same bitumen data. The absorption display clearly resolves a mass doublet (compositions differing by C3 vs. SH4, 0.0034 Da) that appears as a single magnitude mode peak................... 37 Figure 2.7. Isoabundance-contoured plots of double bond equivalents (DBE = rings plus double bonds) vs. carbon number for species containing carbon, hydrogen, one nitrogen and one sulfur derived from magnitude-mode (left) or absorption-mode (right) electrospray ionization 9.4 T FT-ICR mass spectra. Note that absorption-mode identifies many elemental compositions missing from the magnitude-mode assignments......................................................................................... 38 Figure 2.8. Electrospray ionization 9.4 T FT-ICR absorption-mode mass spectral segments for a vacuum gas oil, before (top) and after (bottom) baseline flattening. Baseline flattening enables automated identification of additional signals from low-abundance ions..................................... 39 Figure 3.1. Zoom insets of Crude oil FT-ICR absorption-mode mass spectrum before low-pass filter (top) and after low-pass filter (bottom)................................................................................. 42 Figure 3.2. Simulated absorption-mode peaks with increased signal amplitude (left) and with increased peaks densities (right).................................................................................................... 44 Figure 3.3. Baseline flattening procedure. (A) Baseline identification from Original absorption spectrum; (B) Linear interpolation for empty spots between each two baseline points chosen in (A). (C) Boxcar smoothing of resulting (B) to yield smoothed baseline. (D) Spectrum with flat baseline after subtraction of resulting (C)..................................................................................... 47 Figure 3.4. Zoom insets of Ribonucleases A FT-ICR absorption-mode spectrum after polynomial modeling baseline correction. Note the discontinuity due to cutting whole data to each small piece.............................................................................................................................................. 49

Page 11: Broadband Phase Correction Of Fourier Transform Ion Cyclotron Mass Spectra

x

Figure 3.5. Zoom insets of Crude oil FT-ICR absorption-mode mass spectrum (top) and Ribonucleases A absorption-mode spectrum (bottom) showing the performance of baseline correction algorithm...................................................................................................................... 51 Figure 3.6. Mass error distribution for Crude oil FT-ICR absorption-mode mass spectrum before baseline correction (top) and after baseline correction (bottom)................................................... 52 Figure 3.7. Top: mass scale-expanded of river bitumen APPI FT-ICR absorption-mode mass spectrum showing the resulting flat baseline. Bottom: zoom insects of river bitumen APPI FT-ICR absorption-mode mass spectrum at nominal mass 746 m/z before and after baseline flattening. Note that the absorption spectrum after baseline correction improves discovery of low-abundance peaks..................................................................................................................... 53 Figure 3.8. Top: 9+ charge state isotopic distribution of Ribonucleases A FT-ICR magnitude-mode (black) and absorption-mode spectrum (red) with flat baseline. Middle: 9+ charge state isotopic distribution of Ribonucleases A FT-ICR absorption-mode spectrum before baseline correction (red) with calculated isotopic distribution profile (black cross). Bottom: 9+ charge state isotopic distribution of Ribonucleases A FT-ICR absorption-mode spectrum after baseline correction (red) with calculated isotopic distribution profile (black cross).................................. 54 Figure 4.1. Twelve full window (left) and half window (right) functions. N is the time-domain data size before zero-filling and n is the index for each data 0 ≤ n ≤ N-1.................................... 59 Figure 4.2. Data processing steps for FT-ICR magnitude-mode and absorption-mode FT-ICR spectra........................................................................................................................................... 60 Figure 4.3. Mass scale-expanded segments of electrospray ionization (ESI) FT-ICR magnitude mass spectra of distillated fraction from crude oil for half Hanning (top) and full Hanning (bottom) apodization functions. The magnitude spectrum with half hanning apodization exhibits greater width at each peak base..................................................................................................... 61 Figure 4.4. Mass error distribution for broadband magnitude-mode FT-ICR mass spectra for a distillate fraction of crude oil for half Hanning (black) and full Hanning (red) apodization functions......................................................................................................................................... 62 Figure 4.5. Mass scale-expanded segments of FT-ICR absorption spectra for a distillated fraction from crude oil for half Hanning (top) and full Hanning (bottom) apodization functions............. 63 Figure 4.6. Mass scale-expanded segments of FT-ICR absorption spectra for bitumen for half Hanning (top) and full Hanning (bottom) apodization functions.................................................. 63 Figure 4.7. Root-mean-square mass error for broadband magnitude-mode (top) and absorption-mode (bottom) FT-ICR mass spectra for a distillate fraction of crude oil, following 0, 1, and 2 zero-fills, for each of several apodization functions. The six apodization functions resulting in the least rms error are shown........................................................................................................ 64

Page 12: Broadband Phase Correction Of Fourier Transform Ion Cyclotron Mass Spectra

xi

Figure 4.8. Peak height-to-noise ratio for peak at m/z 450.4094 (C32H52N1) for a distillate fraction of crude oil, following 0, 1, and 2 zero-fills, for each of several apodization functions.. 64 Figure 4.9. Average mass resolving power for 6 peaks above 6σ of baseline noise at nominal m/z 450 for a distillate fraction of crude oil, following 0, 1, and 2 zero-fills, for each of several apodization functions.................................................................................................................... 65 Figure 5.1. Schematic ICR cell showing the excitation, detection and magnitude field directions....................................................................................................................................... 73 Figure 5.2. A) Plots of time-domain linear frequency-sweep excitation signal (upper) and detected signal (bottom) vs. time. B) Phase spectrum profiles (phase vs. frequency) of linear-sweep excitation and detected signal. Here we only plot the phase spectrum for low-to high frequency-sweep for demonstration purpose................................................................................ 76 Figure 5.3. Distillate of crude oil 9.4T FT-ICR mass spectra. Top: Raw real data following Fourier transform of discrete time-domain signal. Middle: absorption-mode spectrum (obtained after applying analytical phase spectrum). Bottom: Optimized absorption-mode spectrum after fine tune of constant term. Note: additional 0.136 radians in constant term................................. 78

Figure 5.4. Top: Distillate of crude oil of FT-ICR absorption-mode spectra produced by broadband phase correction algorithm (upper) and phase spectrum method (lower). Middle: Plot of calculated phase value vs. frequency from broadband phase correction algorithm (blue) and phase spectrum method (red). Bottom: Plot of phase difference, ε, vs. frequency. Note that every ε value is less than 1 degree (0.0175 radians)............................................................................... 81 Figure 5.5. TOP: Mass scale-expanded segments of distillate of crude oil of FT-ICR absorption-mode spectra from broadband phase correction algorithm (upper) and phase spectrum method (lower). Bottom: Mass error distribution of absorption-mode spectra from phase correction algorithm (upper) and phase spectrum method (lower)................................................................ 82 Figure 5.6. Magnitude and absorption electrospray ionization 9.4 T FT-ICR mass spectra for the same crude oil data from phase spectrum method. The absorption display clearly resolves several mass doublets (compositions differing by C3 vs. SH4, 0.0034 Da) that appears as a single magnitude mode peak................................................................................................................... 83 Figure 5.7. A) Schematic procedure for building SWIFT waveform. B) Detailed SWIFT waveform configuration in real experiment. Note that effective waveform region is located in the middle of whole waveform and is exactly 1/2 of T1..................................................................... 84 Figure 5.8. Top: Crude oil 9.4T absorption-mode (lower) and magnitude-mode (lower) FT-ICR mass spectra excited by SWIFT waveform. Bottom: Mass error distribution for crude oil absorption-mode spectrum (lower) by modified phase spectrum method and magnitude-mode spectrum (upper). Note that much better mass accuracy for absorption-mode spectrum relative to magnitude-mode spectrum excited by SWIFT waveform........................................................... 86

Page 13: Broadband Phase Correction Of Fourier Transform Ion Cyclotron Mass Spectra

xii

Figure 6.1. Top: Plot of time-domain linear frequency-sweep excitation signal. Middle: Plot of detection signal in SED experiments. Bottom: Plot of excitation signal in SED experiment........91

Figure 6.2. Data processing steps for SED experiments. A) Zeroing of containment signal (front saturated signal induced by excitation signal). B) Complex division to produce absorption-mode spectrum. Note that exact same data processing (half apodization and zero-filling) for both zeroed detected signal and excitation signal...................................................................................92 Figure 6.3. Top: Absorption-mode from automated broadband phasing (upper) and SED phasing (lower) for linear-sweep excitation. Middle: Mass error distributions for automated broadband phasing (upper) and SED phasing (lower). Bottom: Isoabundance-contoured plots of double bond equivalents (DBE = rings plus double bonds) vs. carbon number for species containing carbon, hydrogen, one nitrogen for automated broadband phasing (left) and SED phasing (right)............................................................................................................................................. 95 Figure 6.4. Top: Magnitude-mode (upper) and absorption-mode (lower) from normal SWIFT excitation. Middle: Mass scale-expanded segments of crude oil FT-ICR magnitude-mode (upper) and absorption-mode (lower) mass spectra from normal SWIFT excitation. Bottom: Mass error distribution for magnitude (upper) and absorption (lower) electrospray ionization 9.4 T FT-ICR mass spectra for a crude oil........................................................................................................... 96 Figure 6.5. Top: Schematic design of stepped SWIFT waveform. Bottom: Magnitude-mode (upper) and absorption-mode (lower) from stepped SWIFT excitation........................................ 98 Figure 6.6. Electrospray ionization crude oil 9.4 T FT-ICR mass spectra. Top: Raw real data following Fourier transform of discrete time-domain signal. Middle: Result after automatic broadband phasing. Bottom: Absorption-mode spectrum from SED phasing...............................99 Figure 6.7. Top: Schematic automatic peak height correction in SED algorithm. Bottom: Isoabundance-contoured plots of double bond equivalents (DBE = rings plus double bonds) vs. carbon number for species containing carbon, hydrogen, one nitrogen for magnitude-mode (upper) and absorption-mode (lower).......................................................................................... 100 Figure 7.1. Generation of frequency-domain FT-ICR mass spectra from a simulated isotopic distribution with 57 frequency components................................................................................. 105 Figure 7.2. a) Experimental time-domain ICR signal from the isolated 57+ charge state isotopic distribution from electrosprayed humanized IgG1k therapeutic antibody. b) Simulated noiseless time-domain transient with 57 frequency components. c) Absorption-mode frequency spectrum for b). d) Magnitude-mode frequency spectrum for b)............................................................... 107 Figure 7.3. Top: Simulated noiseless time-domain signal after blanking of the data between the "beats". Middle: Absorption-mode frequency spectrum. Bottom: magnitude-mode frequency spectrum....................................................................................................................................... 108

Page 14: Broadband Phase Correction Of Fourier Transform Ion Cyclotron Mass Spectra

xiii

Figure 7.4. Top: Simulated time-domain ICR signal with noise. Middle: Absorption-mode frequency spectrum from the above time-domain signal. Bottom: Absorption-mode frequency spectrum after blanking was performed on the above time-domain signal................................. 109 Figure 7.5. Top: FT-ICR average peak height-to-noise ratio (S/N) for each of the 10 highest absorption-mode peaks (from the time-domain data of Figure 7.4 (top)) with (blue) and without (red) blanking. Bottom: Ion cyclotron frequency errors for the 10 highest absorption-mode peaks with (blue) and without blanking (red)....................................................................................... 110

Page 15: Broadband Phase Correction Of Fourier Transform Ion Cyclotron Mass Spectra

xiv

ABSTRACT

It has been known for 35 years that phase correction of Fourier transform ion cyclotron

resonance (FT-ICR) mass spectral data can in principle produce an absorption-mode spectrum

with mass resolving power as much as a factor of 2 higher than conventional magnitude-mode

display, an improvement otherwise requiring a (much more expensive) increase in magnetic field

strength. However, temporally dispersed excitation followed by time-delayed detection results

in steep quadratic variation of signal phase with frequency. We developed a robust, rapid,

automated method to enable accurate broadband phase correction for all peaks in the mass

spectrum. Low-pass digital filtering effectively eliminates the accompanying baseline roll.

Experimental FT-ICR absorption-mode mass spectra exhibit at least 40% higher resolving power

(and thus an increased number of resolved peaks) as well as higher mass accuracy relative to

magnitude mode spectra, for more complete and more reliable elemental composition

assignments for mixtures as complex as petroleum.

Absorption-mode FT-ICR mass spectrum, which is produced by automatic broadband

phase correction algorithm, demonstrates baseline distortion even with low-pass filter baseline

correction. Significant baseline roll affects peaks picking algorithm and results in incorrect peak

height measurement. Isotopic distribution in spectra presenting large baseline roll couldn’t

display correct information. Thus, identification and characterization of biomolecule become

much more difficult. In Chapter 2, we designed a fast, robust and automated baseline correction

process. Each minimum data point of reversed peak in absorption-mode spectrum has been

collected as bases of baseline model, and then further linear interpolation and boxcar smoothing

technique help to complete the baseline model. Finally, the baseline model is subtracted from

original spectrum to produce a flat baseline. This algorithm has been experimentally proven to

automatically flatten baseline of crude oil, environmental sample and biomolecule FT-ICR mass

spectra. More peaks have been identified from absorption-mode spectrum with flat baseline

without loss of mass accuracy. Isotopic distribution also demonstrates very accurate profile.

Apodization function and zero-filling are two basic steps in data processing of Fourier

transform ion cyclotron resonance (FT-ICR) mass spectrometry and their effect on the

conventional Fourier transform ion cyclotron resonance (FT-ICR) experimental and simulated

Page 16: Broadband Phase Correction Of Fourier Transform Ion Cyclotron Mass Spectra

xv

magnitude-mode mass spectra and single-peak absorption-mode spectra are well known. In

Chapter 4, we examine the effects of each of twelve apodization (window) functions and 0, 1,

and 2 zero-fills for absorption-mode Fourier transform mass spectra peak height-to-noise ratio,

mass measurement accuracy, and mass resolving power for dense FT-ICR mass spectra of

petroleum. Half function windowing is best for resolving close absorption-mode doublets,

whereas full function windowing is best for resolving magnitude-mode doublets. Absorption-

mode offers significantly higher mass resolving power than magnitude-mode for any given

windowing function. Half apodization increases absorption-mode mass accuracy, irrespective of

the choice of window function. One (but not more than one) zero-fill improves mass accuracy for

absorption-mode mass accuracy but not for magnitude-mode. Peak height-to-noise ratio for both

absorption and magnitude spectra is improved by zero-filling.

Although we have successfully demonstrated the automated phase correction method for

complex Fourier transform ion cyclotron resonance (FT-ICR) mass spectrum, we can’t express

the exact quadratic phase function of frequency from calculated phase for discrete data point. In

Chapter 5, we applied stationary phase method to excitation and detection signal and derived the

accurate phase spectra for both the linear chirp excitation and detected FT-ICR signals

analytically. Because phase spectrum of detected signal represents correct variation of

accumulated phase with frequency, it could be directly used to recover the absorption-mode FT-

ICR mass spectra. Also, the phase correction of FT-ICR mass spectra from stored waveform

inverse Fourier transform (SWIFT) by phase spectrum has been experimentally described. The

analytically phase correction results are compared to the previous results produced by automated

phase correction method in terms of resolving power and mass measurement accuracy

Except for phase correction based on mathematical calculation of accurate phase for

different frequencies. Scientists have demonstrated that simultaneous excitation and detection

(SED) enable Fourier deconvolution to provide broadband phase correction with no user

interaction. However, the capacitive nulling technique which is applied in SED method for

removing the saturated excitation signal in front of detected signal is not practical due to unstable

capacitors. In Chapter 6, we describe a new data processing procedure to enable broadband phase

correction of FT-ICR mass spectra by SED without any hardware modification. The resulting

absorption-mode spectra yield improvement in resolving power as well as reduction in

assignment errors relative to conventional magnitude-mode spectra. The Fourier deconvolution

Page 17: Broadband Phase Correction Of Fourier Transform Ion Cyclotron Mass Spectra

xvi

procedure has the additional benefit of correcting for spectral variation resulting from

nonuniform power distribution over the excitation bandwidth and phasing spectra from different

excitation waveforms (e.g., SWIFT with different magnitude modulations).

Fourier transform mass spectrometry (FTMS) of the isolated isotopic distribution for a

highly charged biomolecule produces time-domain signal containing large amplitude signal

"beats" separated by extended periods of much lower signal magnitude. Signal-to-noise ratio for

data sampled between beats is low, due to destructive interference of the signals induced by

members of the isotopic distribution. Selective blanking of the data between beats has been used

to increase spectral signal-to-noise ratio. However, blanking also eliminates signal components,

and thus can potentially distort the resulting FT spectrum. In Chapter 7, we simulate the time-

domain signal from a truncated isotopic distribution for a single charge state of an antibody.

Comparison of the FT spectra produced with or without blanking and with or without added

noise clearly show that blanking does not improve mass accuracy and introduces spurious peaks

at both ends of the isotopic distribution (thereby making it more difficult to identify

posttranslational modifications and/or adducts). Ergo, blanking should never be employed: it has

no advantages and major disadvantages.

Page 18: Broadband Phase Correction Of Fourier Transform Ion Cyclotron Mass Spectra

1

CHAPTER ONE

HIGH RESOLUTION MASS SPECTROMETRY

Mass spectrometry is a powerful analytical technique for analysis of charged particles. Dr. John

B. Fenn had showed us a clear picture of mass spectrometry. “Mass spectrometry is the art of

measuring atoms and molecules to determine their molecular weight. Such mass or weight

information is sometimes sufficient, frequently necessary, and always useful in determining the

identity of a species.” Different mass analyzers coupled with different chromatography have

been powerful tools to routinely do qualitative and quantitative analysis, like, identification of

trace amount contamination in environments and food. Modern mass spectrometry application

mostly concentrated on different bio-analysis field, e.g. proteomics, lipidomics, metablomics and

drug discovery. Among characteristics of mass spectrometry, high resolution and high mass

accuracy are most interested aspects for scientists working in variety research fields in recent

decade years. Also these two characteristics are motivation for improvement of different mass

analyzers (e.g. orthogonal and multi-turn time-of-flight) and invention of new mass analyzer

(orbitrap) in recent years.

What Defines High Resolution and High Mass Accuracy?

Mass resolution determines the ability to distinguish ions with different molecular weight and the

definition of mass resolution is the minimum mass difference, m2-m1, between two mass

spectral peaks such that the valley between their sum is a specified fraction of the height of the

smaller individual peak. The most used value of the valley is the half-maximum height (∆m50%)

of either peak. So for two peaks with equal heights, the mass resolution m2-m1= ∆m50%. Mass

resolving power is another term closely related to the mass resolution. The definition of mass

resolving power could be either for a isolated peak of mass, m, as m/∆m50% or for two equal

amplitude peaks as m2/(m2-m1). For multiple charged compound, the m/z will replace m in

definition of resolving power. 2 Therefore, a greater resolving power definitely improves the

Page 19: Broadband Phase Correction Of Fourier Transform Ion Cyclotron Mass Spectra

2

ability to see smaller mass difference between two ions. Although no exact boundary for

different resolution level, high resolution is usually used to described for mass analyzer with the

resolving power (m/∆m50%) larger than 10,000. Some mass analyzers, such as ion trap,

quadrupole mass filter and triple quadrupole, will not fall into this level even though they are

really useful in different application fields. Mass accuracy is a precision evaluation of mass

measurement provided by mass analyzer. It is often expressed in parts per million (ppm) and

largely depend on the stability and the resolution of different mass analyzers. High mass

accuracy (<5ppm) could provide better mass measurement and reduce the possibility of

elemental composition for each peak.

Mass Resolution and Accuracy

In most current mass spectrometer instruments, the detected signal, e.g., voltage, current or time,

will be collected as discrete data points after digital converter. Each peak has been represented

by discontinuous data points, so peak height and position usually are determined by some sort of

interpolation or fitting procedures. 3-5 The mass measurement precision of peak parameters

depends upon the spectral signal-to-noise ratio, the number of data points per peak width 6 and

different noise distribution 7. Under same signal-to-noise ratio, the peak with more data points

definitely has better mass measurement precision than the peak with less data points. Thus, the

faster data collection rate for recent time-of-flight mass spectrometer is main contribution of

improvement of mass accuracy. After determination of peak position, the whole spectrum will be

calibrated by the accurate masses of two or more different ions. Compare to the external

calibration, the internal calibration could typically gain the better mass accuracy. Because of

unknown sources of m/z-dependent systematic error, some special cared internal Calibrations are

needed, e.g., calibration of spectra in discrete m/z ranges has previously been proposed and

implemented. 8 Considering the ion abundance effect, the small segments internal calibration can

largely reduce the systematic error and improve the rms mass error. 9 Then, the accurate mass

measurement after calibration can be used to assign the elemental composition. This is not a

problem for pure simple compound, and even low resolution mass spectrometer can easily assign

the elemental composition for known chemicals.

Page 20: Broadband Phase Correction Of Fourier Transform Ion Cyclotron Mass Spectra

3

However, as the number of atoms increases, the number of possible combinations also

increases. For unknown complex mixture, mass differences of different analytes become

significantly small and spectral interferences will appear. Such interferences may occur from

different possibilities, e.g., similar elemental combination, oxide chemical formation or multiple

charged ions. Peaks’ overlapping could skew the peak shape of compound and obscure the

analyte of interest, the peak centroid will no longer correspond to accurate mass. Any peak

assignment and database searching based on these peak centroids will produce erroneous results.

The straightforward method for solving this problem is improving resolving power. For

example, molecules of the same nominal mass differing in elemental composition at around m/z

700, e.g. C3 vs. SH4 (~3.4 mDa), could be separated at a resolving power m/∆m50% better than

200,000. High resolution or resolving definitely increases the confidence for assigned elemental

confirmation. In other word, the real mass measurement accuracy of complex mixture totally

relies on the high resolution or resolving power.

Time-of-Flight Mass Analyzers

The time-of-flight mass analyzer, initially proposed by Stephan in 1946,10 and reduced to

practice in 1948, 11 has evolved through several stages. Delayed ion extraction addressed the

problem of kinetic energy spread among ions of the same m/z.12 The advent of fast digitizers,

MALDI (and MALDI imaging), orthogonal ion introduction, and kinetic energy focusing by use

of a reflectron brought TOF MS to its present level of popularity.

Orthogonal Acceleration

Ideally, a TOF MS experiment should begin with all ions with the same initial position and

velocity. Orthogonal ion introduction, 13 in which ions are accelerated in a direction

perpendicular to a collimated ion beam, effectively achieves that condition, and thus improves

mass resolving power. Another advantage of orthogonal introduction is that ions can be

accumulated in the acceleration region while previously accumulated ions fly toward the

detector, thereby more efficiently enabling coupling of continuous ion sources with TOF mass

Page 21: Broadband Phase Correction Of Fourier Transform Ion Cyclotron Mass Spectra

4

analysis. Further optimization of ion guides and use of an ion funnel14 with automatic gain

control (AGC) improves the ion beam quality, for higher sensitivity and mass accuracy.15

Dual Stage Reflectron

IonMirror

High FieldPusher

Ion DetectionSystem

From Ion Source

- Ion Trajectory

- One Figure-Eight Ion Trajectory

To Detector

Detector

Ion Source

Gridless Mirror

Periodic Ion Lenses

Gridless Mirror

Figure 1.1. High resolution time of flight mass analyzers. Top: Dual stage reflectron.16 Middle: Multi-pass reflectron with linear segments.19 Bottom: Multi-pass spiral configuration.17-18

Page 22: Broadband Phase Correction Of Fourier Transform Ion Cyclotron Mass Spectra

5

Reflectron/Multipass TOF

The resolving power for a TOF mass analyzer is m/∆m50% = (T/2∆t), in which T is the ion total

flight time, and ∆t is the mass spectral peak width. Increasing the flight time for improved

resolving power is made possible by the reflectron, introduced by Mamyrin in 1973.20 After an

initial drift region, ions are subjected to a spatially quadratic potential that acts as an ion mirror.

When ions of a given m/z but different kinetic energy fly through the reflectron region, the faster

ions penetrate farther into the reflectron (and therefore travel a greater distance to reach the

detector) than slow ions, and ions of all speeds arrive at the detector at same time. The reflectron

thus provides increased ion flight path, without defocusing due to spread in initial velocity,

thereby increasing mass resolving power. Broadband mass resolving power of 10,000 and rms

mass error of 5-10 ppm may now be routinely attained. With high field pusher and dual stage

reflectrons, commercial TOF mass analyzer can now reach mass resolving power of 40,000. 16

(Figure 1.1, top) If the ion spatial focusing can be maintained with minimal ion loss at each

reflection, the ion flight path may be extended by multiple reflections. Multi-pass 21 and spiral

(Figure 1.1, bottom)17 TOF mass analyzers now attain mass resolving power of 50,000 or higher.

Recently, an optimized multi-pass time-of-flight mass analyzer has been combined with

matrix-assisted laser desorption/ionization (MALDI) for imaging mass spectrometry. The multi-

pass time-of-flight mass analyzer consists of four electric sectors, with cylindrical side electrodes

and Matsuda configuration to generate a toroidal electric field, in which focusing is maintained

by quadrupole triplet lenses at the entrance and exit of each toroidal segment. Mass resolving

power of 130,000 has been achieved for angiotensin II [M+H]+ ions (m/z 1046.542) after 500

turns and 654.8 m total flight path.22 In a typical multi-pass time-of-flight mass analyzer, the

power supply for ion injection/ejection is very large and complicated in order to reduce

instability of voltage switching between ion injection and ejection events. A more recent multi-

pass TOF design adds two more sectors for ion injection and ejection only, thereby minimizing

the size of the power supply.23

Although a cyclic multi-pass TOF mass analyzer offers potentially high mass resolution,

the observable mass range is limited, because low-m/z ions eventually overtake higher-m/z ions,

so that the effective m/z range is reduced by a factor of N for N passes. An elliptical flight path

composed of four toroidal electric sectors works with a new segmentation method to identify the

Page 23: Broadband Phase Correction Of Fourier Transform Ion Cyclotron Mass Spectra

6

number of laps transited by ions of a given m/z.24 The most recent multi-pass designs include

doughnut-shaped ion optics25, a spiral path electric reflector 26 and multiple angle reflecting

electrostatic ion mirrors. 19 (Figure 1.1, middle) The latter two designs have achieved

broadband mass resolving power of 60,000.

Recent Advances in TOF Mass Analyzer

Detection. The most commonly used TOF ion detector is the microchannel plate (MCP)

detector, which converts ions to secondary electrons, with large active area and rapid response

time. However, secondary electron generation efficiency varies directly with incident ion

velocity, and ion velocity varies inversely with the square root of accelerated ion mass, so that

MCP detection efficiency for high-mass ions is low. The mechanical nanomembrane detector

can detect the time-varying field emission of electrons from mechanical oscillation without mass

discrimination and thus improve high-mass detection sensitivity. For example, singly charged

immunoglobulin G (150 kDa) has been experimentally observed by TOF MS with

nanomembrane detection.27 A different nanomechanical resonator,28, 29 whose resonance

frequency is perturbed by surface adsorption of ions, was also recently reported as a sensitive

mass sensor, but further investigation for its practical use in TOF MS is needed. Another

potential useful detector for TOF mass analyzer is the cryodetector. When applied to TOF MS,

cryogenic detectors measure low-energy solid-state excitations, called phonons, created by a

particle impact. The energy of these phonons is less than a few meV which is much smaller than

the energy in the electronvolt range needed to produce secondary electrons in conventional

ionization detectors. The cryodetector is sensitive and good for detection of large, slow-moving

ions. 30 A strong signal at 510 kDa from disulfide liked dimmer has been experimentally

observed in MALDI TOF mass spectrometer coupled with cryodetector. 31

TOF/TOF. Two time-of-flight (TOF) mass analyzers in succession enable MS2

experiments. 32, 33 Usually, The first TOF mass analyzer selects the precursor ion for injection

into a high energy collision cell for collision induced dissociation (CID). The second TOF mass

analyzer then records the fragment ions. The most common TOF/TOF configuration is a linear

TOF mass analyzer followed by a reflectron TOF mass analyzer.34, 35 However, low resolution in

MS1 can render product ion identification difficult.

Page 24: Broadband Phase Correction Of Fourier Transform Ion Cyclotron Mass Spectra

7

Multi-pass TOF/TOF has been achieved with a MALDI ion source, a multi-turn TOF

mass analyzer (TOF1), a collision cell, and a quadratic-field ion mirror,36, 18 providing resolving

power of 5000 for precursor ion selection and 1000 resolving power for product ions throughout

the mass range. The identification of lyso-phosphatidylcholine (LPC) and phosphorylation37 has

been demonstrated. Recently, a spiral ion optical TOF mass analyzer for MS1 was combined

with an offset parabolic ion reflectron TOF as MS2, providing high resolution MS1 precursor

monoisotopic ions up to m/z 2500.38

Selected Applications

Although TOF mass analyzers have lower mass resolution than FT mass analyzers (FT-ICR and

orbitrap), they have no upper m/z limit in principle and are thus particularly useful for

identifying singly charged ions of high molecular weight (as from MALDI). Fast response/scan

rate is also advantageous for applications requiring short acquisition period, e.g., analysis by

liquid or gas chromatography mass spectrometry.

A method for screening doping agents in human urine consists of solid phase extraction

followed by HPLC-TOFMS. Coupled with orthogonal electrospray ionization, 124 substances

including stimulants, narcotics, agonists, diuretics, etc., are identified in less than 30 min.39

Further development of the extraction method enabled identification of 40 more doping

compounds.40 The utility of the LC-TOF method was assessed by parallel analysis of 30

authentic urine samples by use of rapid emergency drug identification (REMEDI) and led to

identification of twice as many drugs.41 Compared to prior methods, the author concluded that

UPLC-TOFMS offers an attractive alternative toxicological screening technique. Because the

mass accuracy tolerance for broad toxicology screening has historically been quite wide (20 ppm

or more),42 TOF mass accuracy less than 5 ppm provided substantial improvement in the number

and confidence of identification.43, 44

Mass spectrometry imaging (MSI) provides rapid detection, localization, and

identification of organic components of complex biological mixtures,45 to establish chemical

correlations with biological function or morphology. MSI experiments are typically performed

with a TOF mass analyzer, due to its good sensitivity and speed. Localization of peptides by

MALDI TOF imaging reveals the distribution of their respective proteins.46 Comparison of

protein profiles from normal and tumor tissue can lead to biomarker candidates.47, 48 The

Page 25: Broadband Phase Correction Of Fourier Transform Ion Cyclotron Mass Spectra

8

profiling and imaging of proteins involved in proliferation, differentiation, and apoptosis by

MALDI TOF MS generates unique and differential proteomic blueprints for implantation and

interimplantation sites.49 Lipid imaging in rat brain tissues during focal cerebral ischemia

reveals dynamic conversion from phosphatidylcholine (PC) to lyso-phosphatidylcholine (LPC)

in brain areas with ischemic injury.50 As for protein profiling, a different lipid distribution

between normal and cancer cells could diagnose disease.51, 52

In aerosol analysis, it is necessary to track changes in aerosol size and chemistry with

sub-second time resolution. A TOF mass analyzer collects high resolution aerosol mass spectra

at rates exceeding 1 kHz for determination of both aerosol size and mass.53 Alternatively, two

dimensional gas chromatography (GCxGC) mass spectrometry requires fast scanning because

the final GC peak width is typically 20-200 ms,54 making TOF the MS method of choice.

GCxGC TOF MS applications include identification of naphthentic acids in oil samples,55, 56

high throughput metabolic profiling,57, 58 characterization of pathogen bacteria,59 and analysis of

organic nitrogen compounds in aerosol samples.60 Moreover, fast transient temporal analysis of

catalyst kinetics targets TOF mass analysis for its sensitivity, detector response, and time

resolution.61

The TOF mass analyzer has been also been applied to polymer analysis by virtue of its

high m/z range and MS/MS capability. Evaluation of top-down TOF MS/MS for poly(α-

peptoid)s synthesized by N-heterocyclic carbine (NHC)-mediated zwitterionic ring-opening

polymerization62 showed that electrospray ionization (ESI) enabled detection of the intact

molecular species, whereas MALDI resulted in elimination of the NHC initiator in the presence

of cationizing salts. Coupled with MALDI, TOF MS can address the mechanism of polymer

backbone degradation via free radical chemistry63 and quantify residual polyethylene glycol

(PEG) in ethoxylated surfactants.64 Cation adducts generated more informative fragment ions

under CID,65 and TOF MS has characterized various polymers with ammonium adducts

generated by ESI.66 Miscellaneous other applications include characterization of C4 explosives

by TOF secondary ion mass spectrometry (SIMS),67 qualitative and quantitative analysis of

herbal medicines (HMs) LC/TOF MS68 and identification of different contamination in human

body by GC/TOF MS.69

Fourier Transform Mass Analyzers

Page 26: Broadband Phase Correction Of Fourier Transform Ion Cyclotron Mass Spectra

9

The FT-ICR mass analyzer, introduced in 1974, 70 has the highest mass resolving power71, 9 and

best mass measurement accuracy9 among current mass analyzers. The orbitrap, another Fourier

transform mass analyzer, invented in 1999,72 has been widely distributed since its commercial

introduction in 2004.73

Common Features of Fourier Transform Mass Analyzers

The ion cyclotron and orbitrap each produces a spatially coherent packet of ions of a given m/z.

Coherence in FT-ICR is achieved by rf excitation at the ion cyclotron frequency, whereas

coherence in the orbitrap is achieved by injecting ions of a given m/z into the mass analyzer in a

time short compared to the ion oscillation frequency. In both devices, the periodic motion

(rotation in ICR, oscillation in orbitrap) is detected from the oscillating current induced in

opposed detection electrodes, as ions pass near each electrode. The signal from ions of a given

m/z is linearly proportional to the number of those ions, and to the radius 1, 74 or amplitude of the

ion motion. Linearity is especially important for FT-ICR, because ions of different m/z may be

selected/excited in any desired combination by a time-domain excitation waveform produced by

inverse Fourier transformation of the desired frequency-domain excitation profile ("stored

waveform inverse Fourier transform" (SWIFT) excitation). 75-77Both devices inherently detect

ions of a wide m/z range simultaneously. The time-domain signal is then digitized, apodized,

zero-filled 78, 79 (to improve digital resolution), and subjected to discrete Fourier transformation

to yield mathematically "real" and "imaginary" frequency-domain spectra, Re(ω) and Im(ω),

which are typically combined to yield a "magnitude" spectrum, M(ω) (see below).1 Spectral

frequency is then converted to m/z by an appropriate calibration equation (see below).

FTMS resolving power is ultimately limited by the acquisition period, T, for the time-

domain transient. However, the time-domain transient duration is in turn limited by magnetic and

electrical field imperfection, ion-neutral collisions, and ion-ion interactions. 79-82 Apodization

functions 83 help to detect small peaks near large peaks by reducing the magnitude of auxiliary

maxima and by narrowing the peak width near the peak base (at the cost of broadening the peak

width at half-maximum peak height).

Ion Accumulation and Detection

Page 27: Broadband Phase Correction Of Fourier Transform Ion Cyclotron Mass Spectra

10

Both ICR and orbitrap mass analyzers are pulsed detectors. Because ion introduction is often

temporally continuous (e.g., ESI and APPI), ions are typically accumulated externally during

detection of ions from the preceding accumulation period. 84 In FT-ICR, ions are externally

accumulated in a multipole electric ion trap and simultaneously ejected toward the ICR cell,

whereas in the orbitrap, ions are collected in a "C" trap,85 and injected simultaneously toward the

orbitrap. In ICR, ion spatial coherence is maintained by the confining magnetic and electrostatic

fields, and by ion-ion interactions, 82 whereas in the orbitrap, ion axial coherence is achieved by

short pulsed injection from the C trap and radial coherence is lost as ions spread out to form a

rotating ring (actually, an advantage, by reducing space charge repulsions and enabling increased

dynamic range), and the ring for ions of a given m/z oscillates axially to produce an image

current on the detection electrodes. The cyclotron frequency in ICR signal varies as (m/z)-1,74

whereas the orbitrap axial frequency (like the ICR "trapping" oscillation) varies as (m/z)-1/2.85 As

a result, ICR mass resolving power varies as (m/z)-1, vs. (m/z)-1/2 for the orbitrap.

Advances in Fourier Transform Mass Analyzers

High field strength. The simplest way to improve FTMS performance is to operate at

higher magnetic field, B (ICR) or higher electric field (orbitrap). ICR mass resolving power or

scan rate increase proportional to B, whereas mass accuracy, dynamic range, and upper m/z limit

increases as B2.86, 87 The highest field FT-ICR instrument is currently 15 T, and 21 T systems are

under construction at the U.S.A. National High Magnetic Field Laboratory and Pacific

Northwest National Laboratory.

The orbitrap axial oscillation frequency can be expressed as: 88

][2

1)ln(

2

)/( 2

1

2

2

1

22 RRR

RR

U

zm

e

m

r

−−⋅=ω

[1.1]

in which e is the elementary charge (1.602 × 10-19 C), R2 is the maximum radius of the outer

barrel-like electrode, R1 is the radius of the central spindle-like electrode, Ur is the voltage

applied to the central electrode, and Rm is the "characteristic" radius at which dU(r,z)/drz=0 = 0.

From Eq. 1.1, it is clear that orbitrap resolving power may be increased by increasing the central

Page 28: Broadband Phase Correction Of Fourier Transform Ion Cyclotron Mass Spectra

11

electrode voltage or decreasing the R2/R1 ratio. The newest orbitrap exhibits increased Ur and

optimized R2/R1 ratio to provide ~2-fold improvement in resolving power (for a given data

acquisition period) or similar reduction in data acquisition period (for faster on-line LC

sampling, for a given mass resolving power). With careful balancing of construction tolerances

and experimental parameters, the new orbitrap resolved an isotopic distribution at a resolving

power in excess of 600,000 at m/z 196. 88

12

mm

30

mm

10

mm

20

mm

Standard Orbitrap High Field Orbitrap

Figure 1.2. Standard orbitrap (left) and compact high field orbitrap (right) mass analyzers.88

Page 29: Broadband Phase Correction Of Fourier Transform Ion Cyclotron Mass Spectra

12

Electric field optimization. The orbitrap mass analyzer is composed of a spindle-like

central electrode and a barrel-like outer electrode (Figure 1.2). Application of a dc voltage to the

two axial electrodes produces an electrostatic potential that ideally is the sum of a quadrupolar

ion trap potential and a logarithmic potential of a cylindrical capacitor. For ICR, a strong

homogeneous static magnetic field confines ion motion transverse to the magnetic field, and an

ideally three dimensional quadrupolar electrostatic potential confines ion motion parallel to the

magnetic field.

Compensation Segments

Trapping Segments

Central S

egment

Figure 1.3 Top: Seven segment compensated ICR cell. 89 Bottom: Dynamically harmonized ICR cell (bottom). 90

Page 30: Broadband Phase Correction Of Fourier Transform Ion Cyclotron Mass Spectra

13

In practice, the potentials in both mass analyzers deviate from ideal, because the

electrodes are truncated, imperfect in shape and alignment, and must have apertures to admit ions

(both) and electrons or photons (ICR). In ICR, an imperfect (anharmonic) quadrupolar

electrostatic potential produces an axially dependent, nonlinear radial electric field (Er) that

makes the observed cyclotron frequencies depend on ion radius and axial amplitude. Different

approaches to the cell anharmonicity problem have been investigated since the 1980s. 91-98 One

approach is to insert compensation rings into an open cylindrical cell. 99, 89, 100-102 (e.g., Figure

1.3, top). Another idea relies on ion cyclotron motion to average the electric potential from

curved segmented electrodes90 (Figure 1.3, bottom). Both cell designs in Figure 1.3 have

achieved baseline unit mass resolution for proteins up to 150 kDa.103 Moreover, a coaxial multi-

electrode cell (O-trap),104 with separate compartments for ICR excitation and detection, has been

experimentally demonstrated. The potential benefit from that design is enhancement of

resolving power equal to detected frequency multiple order that could alternatively yield shorter

detection period for correspondingly higher throughput. (e.g., for LC/MS)

Compared to a standard orbitrap geometry, the outer electrode dimensions of the compact

high-field orbitrap analyzer are scaled down by a factor of 1.5, whereas the central electrode

dimensions are scaled down by a factor of 1.2 (Figure 1.2). The entrance aperture cross-section

is reduced by more than a factor of 2; thus, to avoid a corresponding loss in sensitivity, a new

miniature lens system focuses ions to a much smaller spot. The capacitance of the orbitrap drops

in proportion to size, which results in improved image current detection sensitivity. New

transistors in the preamplifier further increase sensitivity. In spite of higher ion density, the

relatively thicker central electrode results in smaller space charge shifts than for the standard

orbitrap.

Software Development. Calibration. FTMS mass accuracy is determined by signal-to-

noise ratio and number of data points per peak width,6 and by "calibration", namely the

conversion of a frequency-domain spectrum to an m/z spectrum. For FT-ICR, calibration

typically relies on a two-term equation based on a perfectly homogeneous magnetic field and a

purely quadrupolar electrostatic field.105, 106 Other equations incorporating variation in ion

abundance107 and space-charge effects108, 109 have also been evaluated. External calibration

(performed for a different sample than for analyte) is typically 2-3x less accurate than internal

Page 31: Broadband Phase Correction Of Fourier Transform Ion Cyclotron Mass Spectra

14

calibration (based on multiple ions of known m/z values in the analyte sample). Systematic error

has been reduced by a factor of up to 3 (and the number of assigned peaks increased by ~25%)

by separate calibration for each of up to ~30 individual mass spectral segments.8,9 Addition of an

ion abundance-dependent term to the calibration equation can reduce systematic error by another

factor of ~2-3. 89, 9 For a high vacuum gas oil sample with ~5,200 peaks, FT-ICR rms mass error

for an absorption-mode spectrum (see below) drops from 107 ppb to 27 ppb by essentially

eliminating systematic error. For the orbitrap, m/z is to first order proportional to (ω)-1/2, leading

to a one term calibration equation. As for ICR, orbitrap mass calibration may be improved by

considering the effect of space charge, leading to a mass measurement error as low as ~80 ppb

over an m/z range of 100-2,000. 110

Phase Correction. Fourier transformation of a time-domain signal produces

mathematically real and imaginary frequency-domain spectra, Re(ω) and Im(ω), which may be

combined to yield magnitude-mode (also known as absolute-value) and phase spectra, M(ω) and

φ(ω), related according to Eq. 1.2.1

M(ω) = [[Re(ω)]2 + [Im(ω)]2]1/2 [1.2a]

φ(ω) = arctan[Im(ω)/Re(ω)] [1.2b]

If the "phase spectrum", φ(ω), is known, absorption- and dispersion-mode spectra, A(ω)

and D(ω), may in turn be obtained as linear combinations of Re(ω) and Im(ω).

A(ω) = cos[φ(ω)] Re(ω) - sin[φ(ω)] Im(ω) [1.3a]

D(ω) = sin[φ(ω)] Re(ω) + cos[φ(ω)] Im(ω) [1.3b]

An absorption-mode spectral peak is inherently narrower than its corresponding magnitude-mode

spectral peak (Figure 1.4, top) by a factor that depends on the presence and mechanism of signal

damping, 79-82 for an improvement in mass resolving power by a factor of up to 2. In FT-ICR,

temporally dispersed excitation and the delay between excitation and detection result in complex

variation of phase with frequency. For that reason, magnitude-mode has been the conventional

Page 32: Broadband Phase Correction Of Fourier Transform Ion Cyclotron Mass Spectra

15

display for FT-ICR mass spectra. The advantage of absorption-mode display was recognized

from the outset.111 Efforts to achieve broadband phase correction112, 113 culminated in a phase

correction algorithm that can automatically account for first- and second-order phase

accumulation from excitation parameters and iterates the zero-order term.114 An experimental

FT-ICR absorption-mode mass spectrum 40% higher resolving power (and thus an increased

number of resolved peaks) as well as higher mass accuracy relative to its corresponding

magnitude mode spectrum is shown in Figure 1.4, bottom. Another approach for phase

correction is to assume a quadratic relation between phase and frequency, and iteratively solve

for zero-, first-, and second-order coefficients.115

Compared to ICR, the phase "wrapping" (i.e., phase variation by more than 2π across the

frequency spectrum) for the orbitrap is less severe due to the short ion injection period from the

C trap but complicated by the change in central electrode voltage (and therefore resonant

frequency) during ion injection. A "pseudo" phase correction based for the orbitrap combines

absorption-mode display for the top part of the peak with an increasing magnitude-mode

component for the bottom part. Although the peak width at half peak height is thus narrower,116

improvement vanishes for low abundance peaks that are near the base of a high abundance peak,

and the mass measurement accuracy does not improve.

Improved Data Station. Continued improvement in FT mass analyzers requires optimized

control of multiple experimental events, precisely timed data acquisition, and efficient data

analysis to accommodate new hardware modifications and ion optics configurations. A recent

Predator data station117 exploits improved PC bus speed to transfer data to and from the control

PC in real time. Conditional data acquisition can evaluate data in real time and apply user-

defined conditions to average and store data. This capability is primarily used to eliminate low

quality data so as to improve the data S/N ratio and minimize the computer memory required to

store data. Tool command language (Tcl) scripting has been applied for 17 voltages and 18

triggers, and provides a fast and easy way to control the Predator data station and associated

peripherals, e.g., LC, auto-sampler, laser source, and electron emitter.117 Another improved data

control system combines PXI hardware with a workflow-based acquisition and control software

to allow unique types of data-dependent experiments.118 Dynamic switching between different

dissociation technologies may be achieved due to fast and precise hardware response. 119

Page 33: Broadband Phase Correction Of Fourier Transform Ion Cyclotron Mass Spectra

16

ω0

2/τ

A(ω)

M(ω)

Lorentzian Peaks

m/z883.92883.88883.84883.80

883.92883.88883.84883.80

C61H103O3

C58H107O3S1

Bitumen ESI 9.4T FT-ICR Mass Spectra

C61H103O3

Magnitude Mode

Absorption Mode

3.4mDa

2√3/τ

m/Δm50%= 450,000

m/Δm50%= 250,000

Figure 1.4. Top: Magnitude-mode and absorption-mode Lorentzian spectra, corresponding to Fourier transformation of an infinite duration time-domain signal, f(t) = A cos(ω0t) exp(-t/ ). Btootm: Magnitude (upper) and absorption-mode (lower) electrospray ionization 9.4 T FT-ICR mass spectra from the same bitumen time-domain data. The absorption display clearly resolves a mass doublet (compositions differing by C3 vs. SH4, 0.0034 Da) unresolved in magnitude mode.

Page 34: Broadband Phase Correction Of Fourier Transform Ion Cyclotron Mass Spectra

17

Selected Applications

Tandem mass spectrometry. The main difference between ICR and orbitrap for

MS/MS is that precursor ion dissociation can be performed in a Penning trap but not in an

orbitrap. In principle, fragment ions can be brought to coherence in an orbitrap for subsequent

excitation and detection, 120 but MS/MS is typically performed externally (CID or ETD) for

orbitrap detection. 121, 122.MS/MS can be performed in an ICR cell by collision-induced

dissociation (CID), infrared multiphoton dissociation (IRMPD), and electron capture dissociation

(ECD). However, high resolution requires low pressure in the ICR cell, so introduction of

collision gas into the cell requires subsequent pump down before excitation/detection; ergo, CID

FT-ICR experiments are now typically conducted with an external electric multipole ion trap.

Because ECD (or ETD) can fragment a peptide but not dissociate the non-covalently bound

fragments, infrared irradiation is commonly used to release the product ion. CID and IRMPD

heat an ion and typically break the weakest bond to produce predominantly b and y type peptide

fragment ions. However, post-translational linkages (e.g., phosphorylation, glycosylation) are

often lost before peptide backbone cleavage, thereby eliminating knowledge of their location. In

contrast, ECD (or ETD) produces mainly c- and y- type backbone cleavage fragment ions by

breaking the backbone N-C� bond without loss of phosphate or glycan. Thus, ECD or ETD is

much preferred for locating posttranslational modifications. 123, 124.

Proteomics encompasses identification and structural characterization of proteins and

their complexes, determination of post-translational modifications, and quantitation of protein

expression under different treatments. In "top-down" proteomics, a single gas-phase protein is

isolated, and subjected to fragmentation (e.g., CID, IRMPD, ECD, ETD, etc.) to generate

fragment masses for comparison to an appropriate database for protein identification, as well as

identity and location of each posttranslational modification according to its characteristic

additional mass relative to an unmodified amino acid residue. The top-down approach can in

principle locate all posttranslational modifications, even when several are present at the same

time. High resolution and high mass accuracy are essential to resolve isotopic distributions from

peptides of different charge state as well as to assign isobaric peptide compositions (e.g., lysine

vs. glutamine, differing in mass by 0.0364 Da).125 Examples of top-down posttranslational

identifications include: rhesus monkey cardiac troponin,126 integral membrane proteins 127 and

Page 35: Broadband Phase Correction Of Fourier Transform Ion Cyclotron Mass Spectra

18

human histones.128, 129 Further, 99 proteins have been identified by top-down tandem mass

spectrometry in Methanosarcina acetivorans,130 and capillary liquid chromatography coupled

with FT-ICR identified with high confidence 102 endogenous peptides in the suprachiasmatic

nucleus. Robust two dimensional liquid chromatography combined with top down mass

spectrometry increases efficiency for improved quantification and characterization.131, 132 Further

automated data processing can increase top-down throughput.133 In addition to conventional ESI,

IR-MALDESI has been coupled with an FT-ICR mass analyzer to perform top-down analysis of

equine myoglobin (17kDa).134 Discovery and characterization of endogenous peptide135 and

candidate pharmacodynamic markers136 has been demonstrated. Recently, FT-ICR MS achieved

unit mass baseline resolution for an intact 148 kDa therapeutic monoclonal antibody, the highest

mass protein isotopically resolved to date.103 132

In the "bottom-up" approach, proteins are typically separated by HPLC and/or

electrophoresis, and proteolytically digested into peptides, then subjected to on-line LC/MS, and

the most abundant peptides are subjected to MS/MS.137, 138 Any of several algorithms (cite

Sequest, Mascot, etc.) then compare precursor ion mass and MS/MS peak spacing to an

appropriate database for protein identification. 139-141 Identification reliability may be improved

by incorporating LC elution time.142 Alternatively, "shotgun" proteomics also relies on on-line

LC/MS and MS/MS, but without prior gel separation of the proteins.143 Shotgun proteomics has

been applied to microorganisms,144 histones (especially post-translational modifications),145, 146

and discovery of biosynthetic pathways, 147 etc. For proteins too large for efficient gas-phase

fragmentation, "middle-down" proteomics denotes LC/MS and MS/MS for large proteolyic

fragments of the original protein, and has identified 7,454 peptides from 2-20 kDa after 23 LC

MS/MS injections of Lys-C digests of HeLa- S3 nuclear proteins.148 Post-translationally

modified histone H3 variants have also been characterized by that approach. 149

Quantitative proteomics typically requires isotopic labeling, either before (SILAC) 150 or

after peptide isolation.137 Quantification of human monkeypox virus and vaccinia virus

confirmed the level required for pathogenesis with Accurate Mass and Time tag (AMT) method

after peptide isolation.151 With stable isotope labeling of amino acids in cell culture (SILAC), it

is possible to quantitatively evaluate phosphorylation changes in stable cell lines.152 Quantitative

proteomics depends on fragmentation energy. 153, 154 When isotopic labeling is not applicable,

label-free methods can provide a measure of quantitation. 155.

Page 36: Broadband Phase Correction Of Fourier Transform Ion Cyclotron Mass Spectra

19

Metabolomics. High throughput screening of chemical toxicity in Daphnia magna has

been reported.156 Metabolic profiling facilitates biochemical phenotyping of normal and

neoplastic colon tissue at high significance levels 157 and annotates the ions originating from

identical metabolites by fragmentation pattern and database searching analysis.158 Coupled with

different chromatography platforms,159-161 higher dynamic range and identification of more

metabolites can be achieved. Autocorrelation of retention and ionization leads to more

exhaustive metabolic fingerprints.162

500 700 900 1100 1300 1500 1700 1900m/z

730.535 730.545 730.555 730.565m/z

730.53805

PC (32:2), ‐0.11 ppmC40H76N1O

8P1

730.55919

DGTS (32:2) + Na+

‐0.08 ppmC42H76N1O

7Na

1

730.56156

DGTS (34:5)‐0.08 ppmC44H75N1O

7

2.37 mDa

(+) ESI 14.5 T FT-ICR MS of Algae (Nannochloropsis oculata) Lipid Extract

Figure 1.5. Positive electrospray ionization 14.5 T FT-ICR mass spectrum for Nannochloropsis

oculata species lipid extract. Inset: Two peaks differing by 2.37 mDa are resolved and assigned. (Figure kindly provided by Huan He)

Page 37: Broadband Phase Correction Of Fourier Transform Ion Cyclotron Mass Spectra

20

Lipidomics. Characterization of nonpolar lipids and selected steroids by high resolution

mass analysis has emerged as a powerful tool for lipidomics. 163, 164 On-line LC separation can

quickly narrow down the possible phospholipid and glycosphingolipid compositions and

facilitate their identification.165, 166 Unequivocal lipid elemental composition from isotopic fine

structure has been used to validate the identification of sulfur-containing lipids in algae extracts

(Figure 1.5). 167 Direct infusion high resolution mass analysis with computer-assisted

assignment is able to profile the 13C-isotopologues of glycerophospholipids (GPL) directly in

crude cell extracts.168 High resolution MS/MS enabled identification of triacylglycerols

archaeological extracts,169 and neutral lipids.170 Matrix-assisted laser desorption ionization

combined with high-resolution MS enables confident identification and spatial localization of

lipids in tissue sections.171, 172

RNA/DNA Analysis. Most lower resolution MS analyses of oligonucleotides have focused on

synthetic DNA, RNA, aptamers, etc. Double-stranded small interfering RNA (siRNA) presents

numerous fragment ions and multiple chemical modifications. The different metabolite patterns

for synthetic siRNA in different biomatrices (rat and human serum) provide insight into the

stability and pharmacokinetic properties of therapeutic siRNA compounds.173 A method for

sequencing single and double stranded RNA oligonucleotides by acid hydrolysis has been

presented. From the mass differences between adjacent members of a mass ladder, the complete

sequences of different siRNA 21-mer single and double strands could be verified. This simple

and fast method can be applied to controlling sequences of synthetic oligomers, as well as for de-

novo sequencing.174 A high resolution mass spectrometry-based analytical platform for RNA,

employing direct nano-flow reversed-phase liquid chromatography with a spray tip column

predicted the nucleotide composition of a 21-nucleotide small interfering RNA, detected yeast

tRNA post-transcriptional modifications, and analyzed the nucleolytic fragments of RNA at a

sub-femtomole level.175 Various other approaches, e.g., combing chemical modification of RNA

with mass spectrometry, 176 based on neomycin B+ 177 or classical nucleic acid ligands, 178

enable characterization of polypurine tract-containing RNA:DNA hybrids, and show that the

common structural features of lentiviral and retrotransposon PPTs facilitate the interaction with

their cognate reverse transcriptase (RTs). Structural elucidation of the HIV-1 virus has also been

achieved by the combination of footprinting and cross-linking techniques with high resolution

Page 38: Broadband Phase Correction Of Fourier Transform Ion Cyclotron Mass Spectra

21

MS detection.178, 179

Synthetic Polymer Analysis. The most obvious advantage of mass spectrometry for

synthetic polymer analysis is the generation of the full molecular weight distribution, rather than

simply number-average or weight-average molecular weight.180 181 High resolution MS can also

reveal pathways of polymer synthesis, as for a sterically hindered alkoxyamine initiator for the

nitroxide-mediated copolymerization (NMP) of methyl methacrylate (MMA);182 the styrene

monomer was added as polymerization terminator, and the progress monitored from the product

oligomer distributions. Synthetic polymer structure has been probed by various fragmentation

techniques, including Electron Capture Dissociation (ECD), Collision Induced Dissociation

(CID), and Electron Detachment Dissociation (EDD),183-188 e.g., polyamidoamine dendrimer

characterization by ECD and EDD.

Petroleomics. From sufficiently complete characterization of the organic composition of

petroleum and its products, it is becoming possible to correlate (and ultimately predict) their

properties and behavior.189, 190FT-ICR MS is the preferred identification technique for petroleum

analysis because of the need to resolve mass splits down to 3.4 mDa (32S1H4 vs. 12C3) or ~0.5

mDa (for nickel-containing porphyrins) (Figure 1.6). Recently, ion mobility (IM) mass

spectrometry (MS) and FT-ICR MS were combined to confirm the presence of asphaltene

aggregation.191 Removal of contaminants in asphaltenes during the crude oil processing is

necessary to avoid catalyst fouling and hence lower liquid yield. Also, asphaltenes can

precipitate during production and refining. Thus asphaltenes are one of the most discussed

topics in petroleomics. 192-196 Ion mobility mass spectrometry and FT-ICR MS have been

combined to confirm the presence of asphaltene nanoaggregates in toluene at concentrations (10-

4 mass fraction) below those in crude oils. 191 Subsequently, in situ analysis of a 3000 ft vertical

column of crude oil by downhole fluid analysis indicated that the asphaltenes in a black crude oil

exhibit gravitational sedimentation according to the Boltzmann distribution and that the

asphaltene colloidal size is about 2 nm. 197 The detailed characterization of a Middle Eastern

heavy crude oil distillation series fully validated the Boduszynski model, namely that petroleum

is a continuum with regard to composition, molecular weight, aromaticity, and heteroatom

content, and that boiling point can be predicted from aromaticity and heteroatom content. 198, 199

Naphthenic acids (i.e., species containing non-aromatic hydrocarbon rings) can form harmful

Page 39: Broadband Phase Correction Of Fourier Transform Ion Cyclotron Mass Spectra

22

deposits, and may be identified in crude oil before the deposits form200-202 as well as in oil sands

processed water.203-205 Ultrahigh mass resolution also enables identification of biomarkers, such

as nickel and vanadyl petroporphyrins.206, 207

m/z1,000900800700600500400300

m/z636.59636.51636.43636.35636.27

m/z428.311428.308428.306428.303

Positive Ion APPI 9.4 T FT-ICR MS of Heavy Crude Oil

1.1 mDa (13CH332S vs C4)

m/∆m50%=1,000,000

m/∆m50%=1,100,000

54,004 Peaks > 6σ between

m/z 200 and 1000

90 Peaks > 6σ Baseline Noise

Figure 1.6. Positive ion atmospheric pressure photoionization (APPI) 9.4 T Fourier transform ion cyclotron resonance (FT-ICR) mass spectrum of a petroleum crude oil. Upper inset: Mass scale expansion revealing 90 singly-charged ions within a mass range of 0.32 Da. Lower inset: Baseline resolution of ions differing in mass by 1.1 mDa, made possible by ultrahigh mass resolving power of 1,000,000 at m/z 428. (Data kindly provided by Amy McKenna)

Page 40: Broadband Phase Correction Of Fourier Transform Ion Cyclotron Mass Spectra

23

Environmental Analysis. Applications of high resolution MS technology in the fields of

food safety208 and environmental analysis have recently been reviewed.209 High resolution helps

to trace contaminants, as well as to identify analogs for which standards are not available. High-

resolution full scan MS analysis with simultaneous targeted CID MS/MS has been a applied to

analysis of polyether toxin azaspiracid food contaminants in shellfish.208 High resolution is also

useful in the high-throughput screening of micropollutants in wastewater treatment. Both

targeted and non-targeted screening have identified a variety of unreported and previously

reported pesticide transformation products.210 Characteristic polar oil sands components

(naphthenic acids and other related acid fraction components) spiked into plant tissue prior to

extraction can be differentiated from co-extracted endogenous plant components by high

resolution MS.205 LC/high resolution MS showed that the industrial chemical, perfluorooctanoic

acid, is microbiologically inert, and hence environmentally persistent.211 The high production

volume chemicals, benzotriazoles and benzothiazoles, have been extracted from environmental

water by solid-phase extraction and identified by LC/high resolution MS. 212 Induced metabolite

changes in myriophyllum spicatum, an aquatic vascular plant, have been monitored by high

resolution MS.213 The method has also provided molecular characterization as the level of

elemental composition, for dissolved organic matter (DOM), e.g., from pore water,214 secondary-

treated waster water,215 and the Greenland ice sheet, 216 because it is capable of resolving

complex molecular mixtures and providing information about the exact elemental composition.

Page 41: Broadband Phase Correction Of Fourier Transform Ion Cyclotron Mass Spectra

24

CHAPTER TWO

AUTOMATED BROADBAND PHASE CORRECTION OF

FOURIER TRANSFORM ION CYCLOTRON RESONANCE

MASS SPECTRA

Introduction

Fourier transform ion cyclotron resonance mass spectrometry (FT-ICR MS)74 offers the highest

broadband mass spectral resolving power and mass accuracy. For example, isobaric "fine

structure" (i.e. ions of the same nominal (nearest integer) mass but different elemental

composition, CcHhNnOoSs) is routinely resolved to yield thousands of unique elemental

compositions in a single mass spectrum (300-1000 Da) for petroleum crude oil. With great

effort, isotopic fine structure (i.e., ions of the same elemental composition but different

composition of isotopes such as 13C, 15N, 18O, and 34S) has been resolved for a 16 kDa protein,71

at a mass resolving power (m/z)/Δ(m/z)50% ≈ 8,000,000, in which m/z is ion mass-to-charge ratio

and Δ(m/z)50% is the full mass spectral peak width at half maximum peak height.217 Although

mass resolving power increases linearly with increasing magnetic field,2 magnet cost increases at

a high power of magnetic field. Thus, one is led to seek other ways to improve FT-ICR mass

resolving power at a given magnetic field strength. To that end, it is necessary to review the

origin and nature of FT-ICR mass spectral peak shape.

Problem

Fourier transformation of a time-domain ICR signal produces mathematically real and imaginary

frequency-domain spectra, Re(ω) and Im(ω), or alternatively magnitude-mode (also known as

absolute-value) and phase spectra, M(ω) and φ(ω), related according to Eq. 2.1.1

M(ω) = [[Re(ω)]2 + [Im(ω)]2]1/2 [2.1a]

Page 42: Broadband Phase Correction Of Fourier Transform Ion Cyclotron Mass Spectra

25

φ(ω) = arctan[Im(ω)/Re(ω)] [2.1b]

(A frequency-domain spectrum may be converted to a mass-to-charge (m/z) ratio spectrum, at

sub-ppm mass accuracy, by a two-parameter "calibration" equation based on a quadrupolar

electrostatic trapping potential approximation. 105, 106) Absorption- and dispersion-mode spectra,

A(ω) and D(ω), may in turn be obtained as linear combinations of Re(ω) and Im(ω).

A(ω) = cos[φ(ω)] Re(ω) - sin[φ(ω)] Im(ω) [2.2a]

D(ω) = sin[φ(ω)] Re(ω) + cos[φ(ω)] Im(ω) [2.2b]

An absorption-mode spectral peak is inherently narrower than its corresponding

magnitude-mode spectral peak by a factor that depends on the mechanism of signal damping:

ion-molecule reactions, hard-sphere collisions, 80 Langevin collisions to yield the Lorentzian

peak shape, shown in Figure 2.1,217 ion-ion interactions,81, 82 magnetic and electric field

imperfections, and undamped time-domain signal to yield sinc peak shape.79 In any case, our

phasing process is independent of decay model because we don’t fit our data to an assumed

lineshape. Moreover, magnitude-mode doublets that are barely resolved suffer from phase-

dependent frequency shifts that reduce mass accuracy;218 absorption-mode doublets are not

shifted. Dating back to the very first FT-ICR mass spectra,70, 219 it was recognized that

absorption-mode affords higher resolving power than magnitude-mode display.111 However,

recovery of A(ω) from Re(ω) and Im(ω) requires knowledge of the correct "phase" angle, φ(ω).

In Fourier transform nuclear magnetic resonance (FT-NMR) spectroscopy, signals at all

frequencies are excited nearly simultaneously by a short rf pulse, so that the time delay between

excitation and detection for different specific frequencies is constant. The "phase" angle, φi(ωi)

varies linearly with frequency,

φi (ωi) = φ0+ ωi tdelay [2.3]

by less than 2π radians across a typical FT-NMR spectrum, relative to a fixed phase, φ0, that is

Page 43: Broadband Phase Correction Of Fourier Transform Ion Cyclotron Mass Spectra

26

independent of frequency. Thus, one need simply vary φ0 until Re(ω) = A(ω) for a peak at one

end of the spectrum, and then vary tdelay until Re(ω) = A(ω) for a peak at the other end of the

spectrum. Peaks at intermediate frequencies will then all exhibit absorption-mode shape.

ω0

D(ω)

ω0

A(ω)

M(ω)

ω0

2/τ

2/τ

2√ 3 /τ

Figure 2.1. Absorption, dispersion, and magnitude-mode Lorentzian spectra, corresponding to Fourier transformation of an infinite duration time-domain signal, f(t) = A cos(ω0t) exp(-t/ ).

Page 44: Broadband Phase Correction Of Fourier Transform Ion Cyclotron Mass Spectra

27

However, for FT-ICR MS, the frequency bandwidth is much larger than for FT-NMR, and

signals are typically excited by frequency-sweep excitation219, 220, so that the time delay

((ti+tdelay)in Figure 2.2) between excitation and detection also varies linearly with frequency:

Each ti can be calculated as:

ti = [(ωe – ωi)/(sweep rate)] [2.4]

in which ωe is the end frequency of the frequency-sweep excitation, and sweep rate is time rate of

change of excitation frequency. ωe, tdelay, and sweep rate are known in each experiment.

Substituting into Eq.2.3. We see that φi(ωi) varies quadratically with frequency, ωi:

φi (ωi) = φ0+ ωi (ti +tdelay) = φ0+ ωi [((ωe – ωi)/(sweep rate))+tdelay] [2.5]

Moreover, φi(ωi) can vary by 2π radians or more over even a single nominal m/z. Thus,

because sin[φi + n(2π)] = sin(φi), it is not easy to recognize the variation of φi(ωi) with frequency

across a broadband spectrum. As a result, virtually all FT-ICR mass spectra have been limited to

magnitude-mode for the past 35 years.

Prior Solutions

FT-ICR mass spectral phase correction to yield absorption-mode display has been demonstrated

previously for a single peak70 and over a narrow m/z range.112 In another approach, a plot of

dispersion vs. absorption (DISPA) for an isolated Lorentzian peak yields a circle whose center is

rotated by φ radians,221 allowing for determination of φ for each spectral peak. The DISPA plot

has been applied successfully to user-interactive phase correction of FT-NMR spectra 222 and to

FT-ICR mass spectra with widely separated (by ~20 Δ(m/z)50%) peaks,223 but is not suitable for

closely spaced peaks (i.e., the situation for which phase correction is most desirable), because the

"tails" of the dispersion spectrum extend many line widths away from the peak center.

If excitation and detection can be performed simultaneously, then the convolution theorem

of Fourier analysis can be used to recover a broadband absorption spectrum as follows. For a

linear response system (i.e., the response magnitude at a given frequency is linearly proportional

to the excitation magnitude at that frequency), the response, f(t), to a particular time-domain

Page 45: Broadband Phase Correction Of Fourier Transform Ion Cyclotron Mass Spectra

28

excitation waveform, e(t), is the convolution of e(t) and the response, a(t), to an "impulse" (i.e.,

instantaneous) excitation.

f(t) = e(t) * a(t) [2.6]

The convolution theorem states that the corresponding Fourier transforms, F(ω), E(ω), and A(ω)

of the time-domain waveforms, f(t), e(t), and a(t) are related by

F(ω) = E(ω) � A(ω) [2.7a]

so that the absorption-mode spectrum can in principle be "deconvolved" according to:

A(ω) = F(ω)/E(ω) [2.7b]

Eq. 2.7b has been demonstrated for the in silico noiseless ICR response to a linear frequency-

sweep excitation.224 Some 25 years later, deconvolution to yield a broadband absorption-mode

FT-ICR mass spectrum was demonstrated experimentally, based on careful nulling of the signal

induced on the detection electrodes by the excitation electrodes of an ICR trapped-ion cell.113

However, we found that nulling was difficult, incomplete, and unstable, so we seek a simpler

method that can be easily extended to all FT-ICR mass spectra.

Peak height, phase, frequency, and decay constant for each ion signal may estimated by a

maximum likelihood method based on an assumed mathematical ion signal model. 225 Mass

accuracy was shown to improve for a spectrum with 30 non-overlapped peptide peaks.

However, the parameter estimates do not explicitly account for peak overlap, and a large error in

one peak frequency estimate can corrupt the mass estimates for ions of other frequencies.

Here, we present a broadband phase correction method to produce an absorption-mode

spectrum with no user interaction. Following fast Fourier transformation of the discrete time-

domain response to a frequency-sweep excitation to yield discrete real and imaginary spectra, the

algorithm calculates the optimum value of φ(ω) for each spectral data point, based on the initial

and final frequencies of the swept excitation, the sweep rate, and the "ringdown" delay between

Page 46: Broadband Phase Correction Of Fourier Transform Ion Cyclotron Mass Spectra

29

excitation and detection. The absolute phase at each spectral point is then determined iteratively

by optimizing the positive magnitude of the absorption spectrum while varying a frequency-

independent phase correction factor from 0 to 2π (φ0 in Eq. 2.5). This method has been tested for

electrosprayed petroleum mixtures, and yields a larger number of uniquely identified elemental

compositions as well as higher mass accuracy. A time-domain FT-ICR signal with 8 Mwords can

be processed automatically to generate a phase-corrected absorption spectrum in a few minutes.

Experimental Methods

Sample Description and Preparation

Athabasca bitumen and distillate fraction (500 – 538 ○C) from Athabasca bitumen heavy vacuum

gas oil (HVGO) were obtained from the National Centre for Upgrading Technology (Devon,

Alberta, Canada). The California crude oil was obtained from Chevron. The California curde oil

was adsorbed onto a 60g silicic acid chromatographic column and fractions of increasing polarity

were selectively eluted with increasing solvent strength. Bitumen, HVGO and each California

crude oil fraction sample were each diluted to 500 μg/mL in a 50:50 toluene methanol mixture

and used without additional purification. All solvents were HPLC-grade, obtained from Sigma-

Aldrich Chemical Co. (St. Louis, MO).

Instrumentation: 9.4 Tesla FT-ICR MS

Each oil sample was analyzed with a custom-built FT-ICR mass spectrometer equipped with a

9.4 Tesla horizontal 220 mm bore diameter superconducting solenoid magnet operated at room

temperature (Oxford Corp., Oxney Mead, U.K.) and a modular ICR data station (PREDATOR)

facilitated instrument control, data acquisition and data analysis. 226, 227 Positive ions generated

at atmospheric pressure were accumulated in an external linear octopole ion trap84 for 250-1000

ms and transferred by rf-only octopoles to a 10 cm diameter, 30 cm long open cylindrical

Penning ion trap. Octopoles were operated at 2.0 MHz and 240 Vp-p amplitude. Broadband

frequency sweep (chirp) dipolar excitation (70 – 700 kHz at 50 Hz/μs sweep rate and 350 Vp-p

amplitude) was followed by direct-mode image current detection to yield time-domain data sets

Page 47: Broadband Phase Correction Of Fourier Transform Ion Cyclotron Mass Spectra

30

with 8 Mwords. Time-domain data sets were co-added (200-300 acquisitions), Hanning

apodized,1 and zero-filled once before fast Fourier transform and magnitude calculation.

Detect

Time

Frequency-Sweep Excitation

tdelay(m/z)i-1 = 2000

(m/z)i = 500

(m/z)i+1 = 50

E(t)

Fre

qu

en

cy

ti-1

ti

ωi-1

ωi ωi+1

ti+1

(Shift Theorem)

ωs

ωe

Figure 2.2. Plots of time-domain linear frequency-sweep excitation signal (top) and instantaneous excitation frequency (bottom) vs. time, showing the time elapsed following the instant that the excitation frequency matches the ion cyclotron resonance frequency for ions of each of three different m/z values. The corresponding accumulated phase at ωi-1, ωi, or ωi+1 at the onset of detection is ωi-1 (ti-1 + tdelay), ωi (ti + tdelay), or ωi+1 (ti+1 + tdelay) Here the frequency-sweep is from low to high frequency, but a similar argument applies for sweeping from high to low frequency.

Mass Calibration

Each FT-ICR mass spectrum was first calibrated with respect to a prior sample containing

Ultramark®, Met-Arg-Phe-Ala (MRFA) peptide, and caffeine (external calibration), and then

with respect to a homologous series of ions of high abundance common to all three samples.

Page 48: Broadband Phase Correction Of Fourier Transform Ion Cyclotron Mass Spectra

31

Peaks with relative peak abundance greater than 6 times the standard deviation of the baseline

noise (the best empirical compromise between maximizing the number of peaks and maximizing

confidence in identification) were exported to a spreadsheet. All masses were then converted to

the Kendrick mass scale.228 A unique molecular formula could then be assigned to each peak

based on the Kendrick Mass Defect (KMD) series.229 This procedure has been extensively

validated for petroleum and its products.189

Computational Method

Figure 2.2 shows why FT-ICR spectral phase varies linearly and quadratically with frequency.

First suppose that ions of different cyclotron frequency are all excited instantly and

simultaneously, such that the signal at each cyclotron frequency, ωi, varies as cos(ωit). Because

the time-domain excitation rf signal magnitude is typically ~200 Vp-p, whereas the detected ICR

time-domain rf signal is of the order of microvolts, it is necessary to delay the onset of detection

by tdelay ≈ 1 ms to allow the receiver to "ring down" before detection begins. The "shift theorem"

of Fourier analysis230 dictates that the accumulated phase angle (i.e., phase "shift") at the onset of

detection will be ωitdelay. Thus, the phase correction, φ(ω), in Eq. 2.2a will vary linearly with

spectral frequency, as in Eq. 2.4. Next, suppose that ions are excited by a sinusoidal waveform

whose frequency changes linearly with time (Fig. 2.2, left), and whose phase increases

quadratically with time (just as the linear distance traveled by an object subjected to constant

linear acceleration (i.e., linearly increasing velocity) varies quadratically with time). Ions of a

given m/z exhibit initially spatially incoherent cyclotron motion, and coherent cyclotron motion

begins immediately at the onset of resonant excitation.231 The key step is to realize that, to a

good approximation, ions of a given cyclotron frequency are excited to their final cyclotron

radius essentially at the instant that the excitation frequency equals the cyclotron frequency.

Coherent ion cyclotron motion is initiated with a phase shifted by π/2 radians with respect to the

excitation frequency231 and all excited ions will thenceforth exhibit a quadratic variation of initial

cyclotron phase with frequency that corresponds to that of the excitation waveform. After

excitation, the ion cyclotron phase will accumulate linearly (e.g., ti-1 for ions of m/z 2,000 in

Figure 2.2, plus tdelay (for ions of all m/z). Thus, the accumulated cyclotron phase will be the

sum of the temporally quadratic phase accumulation during the frequency sweep plus the

temporally linear increase between the instant of excitation and the onset of detection.

Page 49: Broadband Phase Correction Of Fourier Transform Ion Cyclotron Mass Spectra

32

Moreover, ωe, sweep rate, and tdelay in Eq. 2.5 may be well approximated by knowledge of the

frequency-sweep excitation range and rate and the subsequent time delay between the end of

excitation and the onset of detection. It remains only to determine initial phase angle φ0 in Eq.

2.5. However, because φ0 is independent of frequency, we can simply vary it from 0 to 2π in 1-

degree increments (i.e., 2π/360 radians) until the entire spectrum is optimally "phased".

Choosing the Best Phasing Parameters

We tested several criteria for selecting the best value of phase angle in Eq. 2.5, and settled on the

following. We begin by finding the magnitude-mode spectral data point of greatest magnitude.

We then set φi = φ0 - ω (ti + tdelay) at that data point (ω = ωi). For each successive frequency

increment above or below ωi, we then calculate (ti + tdelay) (given that (ti + tdelay) in Figure 2.2 is

known for each discrete spectral frequency) so that the relative phase for each point in the whole

spectrum is known. We then iteratively calculate the best initial phase (φ0) by variation as

described above. We find it best to judge the success of phase correction based on the full

spectrum rather than just one peak. Because each absorption peak is symmetric about its

midpoint and each absorption spectral data point is positive-valued (in the absence of noise), we

therefore recalculate the best zero-order phase, φ0, as that for which the sum of all of the

absorption-mode spectral data points is maximum positive.

Baseline Correction

The baseline of an absorption-mode spectrum from a zero-filled time-domain data set exhibits

wiggles that are manifested as a periodic "roll" in the frequency-domain spectrum.1 Baseline roll

is significant for frequency-sweep excitation and increases with decreased sweep rate and

increased excitation bandwidth. Fortunately, baseline roll varies much more slowly with

frequency than do ion cyclotron resonance peaks, and thus may be greatly reduced by digital

low-pass filtering (see Figure 2.3).1 The frequency-domain absorption-mode spectrum is

Fourier transformed and the data corresponding to the highest 99% of the transformed abscissa

are removed. The remaining transformed data set is then inverse Fourier transformed to

represent the original (rolling) baseline with all spectral peaks removed. The filtered baseline is

then subtracted from the original absorption-mode data to produce an absorption-mode spectrum

Page 50: Broadband Phase Correction Of Fourier Transform Ion Cyclotron Mass Spectra

33

with greatly reduced baseline roll. Removal of the roll makes it easier to find and quantitate low

abundance signals automatically, because peak-picking algorithms typically select signals higher

than a specified baseline magnitude.

FT

1

Inverse

FT

(A)(B)

(C)

(D)

(F)

(E)

m/z

m/z

"Frequency"’

"Frequency"’

"Frequency"’

Figure 2.3. Schematic baseline flattening procedure. (A) Original absorption spectrum; (B) Fourier transform of (A); (C) Rectangular weight function to remove high-"frequency" components to yield (D); (E) Inverse Fourier transform of (D) to yield the low-"frequency" spectral baseline with true mass spectral peaks removed; (F) Baseline-flattened spectrum produced by subtracting (E) from (A). Computational Implementation

The phase correction software has been implemented as an extension of our modular Predator

ICR data acquisition data system,227 in C/C++ computer language operating under

LabWindows/CVI (National Instruments, Austin, TX). The algorithm achieves broadband phase

correction of 8 Mword discrete time-domain ICR data in two minutes.

Page 51: Broadband Phase Correction Of Fourier Transform Ion Cyclotron Mass Spectra

34

Vacuum Gas Oil

ESI 9.4T FT-ICR MSFFT Real Spectrum

Before Phasing

Figure 4

Magnitude Modem/Δm50% 490,000

RMS Error 140ppb

300 400 500 600 700m/z

700600500400300

Absorption Mode

After Phasing

m/z

m/Δm50% 710,000

RMS Error 110ppb

(13.6 T)

9.4 T

Figure 2.4. Electrospray ionization 9.4 T FT-ICR mass spectra. Top: Raw real data following Fourier transform of discrete time-domain signal. Middle: Magnitude-mode spectrum (obtained from Eq. 1.1a). Bottom: Absorption-mode spectrum. The resolving power for the absorption-mode display is equivalent to that for magnitude-mode at 13.6 Tesla. Note also higher mass accuracy for absorption-mode relative to magnitude-mode display.

Page 52: Broadband Phase Correction Of Fourier Transform Ion Cyclotron Mass Spectra

35

Results and Discussion

Absorption-Mode vs. Magnitude-Mode Spectral Display

Figures of merit for mass spectrometric performance should never be based on one or a few mass

spectral peaks, but rather on many peaks spanning a wide range of signal-to-noise ratio and m/z.

Petroleum crude oil and its distillates thus provide a particularly stringent test for the present

approach, because of their daunting compositional complexity: e.g., up to 50,000 resolved peaks

across a 340 < m/z < 1,500 range.87 Broadband FFT real, magnitude-mode, and phase-corrected

absorption-mode electrospray ionization FT-ICR mass spectra of vacuum gas oil at 9.4 T are

shown in Figure 2.4. Note the ~45% higher mass resolving power (equivalent in that respect to

magnitude-mode display at 13.6 tesla!) and ~30% improvement in mass accuracy for absorption-

mode relative to magnitude-mode display.

Mass Accuracy

Figure 2.5 shows mass error distributions for the broadband ESI FT-ICR magnitude-mode and

absorption-mode mass spectra of Figure 2.4. (The two data sets were apodized differently to

achieve optimal performance—see Future Extensions.) The reduction in rms error (averaged

over all assigned peaks) is achieved mainly by reassignment of those peaks with the highest

magnitude-mode mass errors. Moreover, close mass doublets partially or unresolved in

magnitude mode become resolved in absorption mode (see Figure 2.6), thereby increasing the

number of correctly assigned elemental compositions in Figure 2.5. Thus, the improvement in

resolution is not only quantitative; it can be qualitative (i.e., being able to assign a peak or not).

Finally, the improvement is much greater near the baseline of the peak, where a Lorentzian

magnitude-mode peak can be ten-fold wider than for absorption mode (see Figure 2.1). In

theory, zero-filling increases the peak position precision for an absorption-mode spectrum,

because the first zero-fill effectively captures into the absorption spectrum information residing

in the non-zero-filled dispersion data obtained by Fourier transformation of the time domain

data. 78 However, zero-filling exposes Gibbs oscillations ("wiggles" on the sides of each peak

that can compromise our peak-picking procedure) and apodization removes those wiggles. The

Page 53: Broadband Phase Correction Of Fourier Transform Ion Cyclotron Mass Spectra

36

effects of apodization and zero-filling on phase correction will be discussed at length in a future

paper.

1,692 Assigned Peaks

RMS Error: 140 ppb

Magnitude Mode:

One Zero-Fill &

Hanning Apodization

Vacuum Gas Oil ESI 9.4T FT-ICR MS

Num

ber of P

eaks p

er

Bin

0

100

200

300

400

1 0 -1Mass Error (ppm)

1,732 Assigned Peaks

RMS Error: 110 ppb

Absorption Mode:

One Zero-Fill & Half-

Hanning Apodization

0.8 0.6 0.4 0.2 -0.2 -0.4 -0.6 -0.8

0.8 0.6 0.4 0.2 -0.2 -0.4 -0.6 -0.810

100

200

300

400

0 -1

Mass Error (ppm)

Num

ber of P

eaks p

er

Bin

Figure 2.5. Mass error distribution for magnitude (top) and absorption (bottom) electrospray ionization 9.4 T FT-ICR mass spectra for a vacuum gas oil. Each bar represents the number of assigned masses within a 50 ppb mass error range. The same relative signal abundance threshold (peak height > 5 of baseline noise) was used for peak picking. The magnitude spectrum was produced with one zero fill and Hanning apodization1 and the absorption after one zero fill and half Hanning apodization1 (see text).

Page 54: Broadband Phase Correction Of Fourier Transform Ion Cyclotron Mass Spectra

37

m/Δm50% = 230,000

Magnitude Mode

C58H99O3S1

Bitumen

ESI 9.4T FT-ICR MS

Absorption Mode

3.4mDa

C58H99O3S1

C61H95O3

m/Δm50% = 350,000

m/z875.70 875.75 875.80

Figure 2.6. Magnitude and absorption electrospray ionization 9.4 T FT-ICR mass spectra for the same bitumen data. The absorption display clearly resolves a mass doublet (compositions differing by C3 vs. SH4, 0.0034 Da) that appears as a single magnitude mode peak.

Filling Gaps in Compositional Assignment

A particularly informative way to organize the thousands of assigned neutral elemental

compositions, CcHhNnOoSs, from mass spectra from complex organic mixtures (e.g., petroleum,

dissolved organic matter) is with an isoabundance contoured plot of double bond equivalents

(DBE = number of rings plus double bonds to carbon) vs. number of carbon atoms for each

heteroatom class, NnOoSs. DBE vs. carbon number images for the two most abundant

heteroatom classes from electrospray ionization FT-ICR mass spectra of California crude oil

fractions are shown in Figure 2.7. Note the large gaps in the magnitude-mode image,

Page 55: Broadband Phase Correction Of Fourier Transform Ion Cyclotron Mass Spectra

38

corresponding to high-mass species of relatively low abundance (i.e., lower mass measurement

precision) and non-unique assignment due to increased number of possible compositions for a

given mass tolerance. The corresponding absorption-mode mass spectrum resolves many of

those species, so as to fill in the gaps in the DBE vs. carbon number image.

California Crude Oil

(Fraction 1)

N1S1

0

10

20

30

40

20 60 100

DB

E

N1S1

20 60 1000

10

20

30

40

DB

E

(Fraction 2)

Magnitude Data Absorption Data

Carbon Number Carbon Number

Rela

tive A

bundance

30

20

10

0

30

20

10

020 40 60 80 20 40 60 80

DB

E

DB

E

Figure 2.7. Isoabundance-contoured plots of double bond equivalents (DBE = rings plus double bonds) vs. carbon number for species containing carbon, hydrogen, one nitrogen and one sulfur derived from magnitude-mode (left) or absorption-mode (right) electrospray ionization 9.4 T FT-ICR mass spectra. Note that absorption-mode identifies many elemental compositions missing from the magnitude-mode assignments.

Page 56: Broadband Phase Correction Of Fourier Transform Ion Cyclotron Mass Spectra

39

Baseline Roll and Automated Peak Picking

Automated peak-picking algorithms typically limit consideration to peaks whose height exceeds

a specified multiple of the standard deviation of baseline rms random noise. If the maximum

excursions of that baseline "roll" exceed a few standard deviations of the baseline random noise,

then a peak-picking algorithm can fail to identify true low-magnitude spectral signals whose

frequencies happen to fall near the minima of the "roll". The importance of eliminating baseline

"roll" is evident from Figure 2.8, showing an increase from 13 to 18 in the number of peaks

identified by automated peak picking.

m/z 336330

Vacuum Gas Oil

ESI 9.4 T FT-ICR MS

Absorption Mode Before Baseline Correction:

13 Peaks > 3σ of Baseline Noise

After Baseline Correction:

18 Peaks > 3σ of Baseline Noise

328326 332 338334

Figure 2.8. Electrospray ionization 9.4 T FT-ICR absorption-mode mass spectral segments for a vacuum gas oil, before (top) and after (bottom) baseline flattening. Baseline flattening enables automated identification of additional signals from low-abundance ions.

Page 57: Broadband Phase Correction Of Fourier Transform Ion Cyclotron Mass Spectra

40

Conclusion

Here, we have demonstrated that broadband absorption-mode FT-ICR mass spectra can provide

higher mass resolution, mass resolving power, and mass accuracy than conventional magnitude-

mode spectra. The method has already been successfully applied to FT-ICR MS of petroleum. 198 In future work, we shall address the effects of apodization, zero-filling, systematic errors,

signal-to-noise ratio, and peak separation relative to peak width on mass and amplitude accuracy,

as well as optimal recovery of the information distributed among absorption-mode, dispersion-

mode, and magnitude-mode spectra. For example, optimal magnitude-mode and absorption-

mode resolution can require different apodization ("windowing") functions. 232 And higher mass

resolution can actually result in lower mass accuracy for a close magnitude-mode doublet. Also,

it is not necessarily optimal to begin phasing from the highest magnitude peak: rather, one could

begin phasing at the midpoint of the frequency-sweep (which is not the midpoint of the m/z

range) to reduce the first- and second-order correction errors.

Absorption-mode peaks for ions of different m/z add linearly, whereas magnitude-mode

peaks do not (magnitude is a non-linear operation). Thus, absorption-mode display is an

inherently more accurate representation of ion relative abundances for partially overlapping

resonances. We have observed experimentally (not shown) that absorption-mode display can

exhibit greater peak asymmetry and distortion than magnitude-mode display. Thus, absorption-

mode appears to be more sensitive to the detailed shape and motion of excited ion packets, and

could thus possibly aid in diagnosis (and ultimately reduction) of response nonlinearity as well as

identification (and cure) of various sources of systematic phase and amplitude distortion.

From an application standpoint, the largest resolution enhancement (factor of 2 for an

undamped time-domain ICR signal) will be for on-line LC FT-ICR MS, for which the time-

domain acquisition period is typically 1 s or less, during which the time-domain signal does not

decay significantly. Moreover, the advantages of absorption-mode display for crowded mass

spectra, demonstrated here for petroleum, will also apply to biological samples (proteomics,

glycomics, lipidomics, metabolomics, etc.)

Page 58: Broadband Phase Correction Of Fourier Transform Ion Cyclotron Mass Spectra

41

CHAPTER THREE

BASELINE CORRECTION OF ABSORPTION-MODE FOURIER

TRANSFORM ION CYCLOTRON MASS SPECTRA

Introduction

Previously, broadband Fourier transform ion cyclotron resonance (FT-ICR) absorption-mode

spectrum has been produced by automated phase correction algorithm.114 Crude oil absorption-

mode spectrum demonstrates not only higher resolving power and also better mass accuracy than

conventional magnitude-mode spectrum. However, the baseline of absorption-mode spectra

always exhibits periodic “roll” in the frequency domain spectra (Figure 3.1, top). Significant

baseline roll affects peaks picking algorithm and results in incorrect peak height measurement.

Weak signals in spectra presenting large baseline roll may not be recognized because the

magnitude of baseline roll could be larger than the peak height. Thus, number of peaks with

height greater than 6 of baseline rms noise in absorption-mode spectrum is less than magnitude-

mode spectrum in some cases. In previous method, 1% of frequency data, which contains most

baseline roll information has been taken out by digital low-pass filtering, 1 and subtracted from

original frequency spectrum to produce absorption-mode spectrum with much improved baseline

(Figure 3.1, bottom). Corrected baseline facilitates the peak finding, peak height measurement

and identification of more chemical component in absorption-mode spectrum, anyhow, the

baseline under denser peak area still display certain negative distortion/dip as it is flat within

peak-free area. We notice that negative dip of baseline is significant for frequency sweep

excitation and increases with peak height and peak densities. Simulated results of absorption

frequency peaks further confirm our observations (see below in detail). These residues of

baseline distortion (negative dip) hinder to acquire the accurate peak magnitude in absorption-

mode spectrum and make it unusable in quantification cases. Especially, baseline roll residues

will largely distort the isotopic distribution because big biomolecular FT-ICR mass spectrometry

has really high peak density due to isotopic distribution with highly charged states. Identification

Page 59: Broadband Phase Correction Of Fourier Transform Ion Cyclotron Mass Spectra

42

and structural characterization of different biomolecules require high quality spectra prior to

statistical analysis. 125, 2 Inaccurate isotopic distribution information makes it more difficult to

accurately measure biomolecule mass and causes the failure of database searching algorithm.

m/z

793792791790789788

793792791790789788

Crude Oil 9.4T FT-ICR Absorption-Mode Mass Spectra

After Low-Pass Filter

Before Low-Pass Filter

Figure 3.1. Zoom insets of Crude oil FT-ICR absorption-mode mass spectrum before low-pass filter (top) and after low-pass filter (bottom).

Baseline roll is a major problem in FT NMR and has been extensively investigated by

scientists.233-236 The reasons causing Baseline roll in NMR are various, e.g., dispersive wings

fold back by phase correction,234 nonlinearity response of filters,237 improper weighting of the

initial part of time domain data,238 the discrete nature of Fourier transform and instrumental

instabilities.239, 240 Baseline roll can be reduced by experimental methods, like, justification of

acquisition parameters,236 oversampling241 and digital signal processing,242 or corrected by post

Page 60: Broadband Phase Correction Of Fourier Transform Ion Cyclotron Mass Spectra

43

data processing (e.g., reconstruction of initial part of time-domain data240, 243 and baseline roll

modeling by Fourier series, 244, 245 polynomials246-248 and other functions249-251). Compared to

experimental method, post data processing offers a general and efficient way to correct the

baseline roll, so it has been a most popular approach in FT NMR field and common feature in

NMR analysis software. We believe that the reasons causing baseline in FT ICR is similar to in

FT NMR and baseline correction methods in FT NMR could be applied to FT ICR mass spectra

due to sharing basic Fourier transformation principle and technique. Since baseline correction

method based on post data processing is no dependent on specific experiment process and

parameters, it could be the best choice for FT-ICR mass spectra without further investigation of

origin of baseline roll.

Here, we present a fast and robust algorithm for automatic baseline correction of

frequency spectrum of FT-ICR. The whole procedure, including baseline identification, linear

interpolation and baseline smoothing, has been implemented with phase correction algorithm to

efficiently produce an absorption-mode spectrum with flat baseline. This method is also tested

for crude oil, river bitumen and Ribonucleases A, and yields a larger number of identified peaks

(crude oil) as well as correct isotopic distribution (biomolecule) without loss of mass accuracy.

Compared with calculation time of phase correction, baseline correction only takes less than 10

seconds to flat a 8 Mwords absorption-mode FT-ICR mass spectrum.

Experimental Methods

Simulation

In FT-ICR MS, each time-domain ICR signal, f(t), could be modeled as exponentially damped

sinusoids: 252, 80

f(t) = Acos(ωt+φ)exp(-t/τ) 0<t<T [3.1]

in which i represents each ion signal, is a characteristic damping time constant, or relaxation

time, T is the time-domain data acquisition period, φ is the initial phase, ω is the natural angular

frequency, and A is the amplitude (which is directly proportional to the number of ions of that

Page 61: Broadband Phase Correction Of Fourier Transform Ion Cyclotron Mass Spectra

44

cyclotron frequency). The simulated time-domain signal was apodized, zero-filled and Fourier

transformed to produce absorption and dispersion (A(ω) and D(ω) frequency spectra after phase

correction. The peak height in simulated absorption frequency spectrum is proportional to the A

in time-domain signal. Based on this idea, we simulate single absorption frequency peak with

four different amplitudes (A). Those resulting absorption frequency spectra clearly show

negative dip increasing with increased peak height (Figure 3.2, left).

-32000

0

150000

Frequency (kHz)151,387 151,413

-21000

0

150000

-41000

0

150000

-37000

0

150000

Simulated Absorption-Mode Peaks

-21000

0

150000

-16000

0

150000

-90000

150000

-110000

150000

Peak H

eig

ht

A=0.75

A=1

A=1.5

A=2

Figure 3.2. Simulated absorption-mode peaks with increased signal amplitude (left) and with increased peaks densities (right).

Page 62: Broadband Phase Correction Of Fourier Transform Ion Cyclotron Mass Spectra

45

For investigation of baseline distortion of denser peak area, we simulated the time

domain signal as a sum of exponentially damped sinusoids;

f(t) = ΣAicos(ωit+φi)exp(-t/τ) 0<t<T [3.2]

in which i represents each ion signal. As we add more frequency component into time domain

signal (corresponding to more peaks in absorption frequency spectrum), increased negative dip

in simulated frequency spectrum is observed (Figure 3.2, right). All parameters used in

simulation include T=3.069 s, τ = 1 s, points = 4194304 and bandwidth=568181 Hz.

Sample Description and Preparation

Stock solution of North American crude oil was dissolved in 50:50 (v/v) toluene/methanol to a

final concentration of 1 mg/mL. Each sample was further diluted to 0.5 mg/mL with 50:50 (v:v)

toluene/methanol and 0.1% formic acid, to facilitate protonation during electrospray ionization

(ESI). Peace River bitumen was supplied by National Centre for Upgrading Technology (Devon,

Alberta, Canada) and further diluted to 250 μg/mL in toluene prior to APPI FT-ICR MS analysis

with no additional modification. Rebionuclease A purchased from Sigma-Aldrich Chemical Co.

(St. Louis, MO) was diluted to 1µM in conventional ESI solution: H2O/acetonitrile/formic acid,

50/49/0.5 (v/v/v). All solvents were HPLC grade and purchased from Sigma-Aldrich Chemical

Co. (St Louis, MO).

APPI Source

A custom-built adapter interfaced the APPI source (ThermoFisher Scientific, San Jose, CA) to

the front stage of a custom-built 9.4 T FT-ICR mass spectrometer (see below).253 The sample

flows through a fused silica capillary at a rate of 50 μL/min and is mixed with nebulization gas

(N2 at 50 kPa) inside a heated vaporizer operated at 300 °C for parent crude and maltenes, and

350 °C for asphaltenes.193 After nebulization, gas-phase neutral analytes exit the heated

vaporizer region as a confined jet and a krypton vacuum ultraviolet gas discharge lamp (Syagen

Technology Inc., Tustin, CA) produces 10 eV photons (120 nm). Toluene serves as both solvent

and dopant and increases analyte ionization.254 Charge exchange and proton transfer reactions

Page 63: Broadband Phase Correction Of Fourier Transform Ion Cyclotron Mass Spectra

46

occur between ionized toluene and neutral analytes through collisions with toluene at

atmospheric pressure inside the APPI source.254, 253

9.4 Tesla FT-ICR MS

All samples were analyzed with a custom-built FT-ICR mass spectrometer is equipped with a 22

cm horizontal room-temperature bore 9.4 T magnet (Oxford Corp., Oxney Mead, U. K.) and a

mudular ICR data station (Predator) facilitated instrument control, data acquisition, and data

analysis.117 Ions generated at atmospheric pressure in the external APPI or ESI source enter the

skimmer region at ~ 2 Torr through a heated metal capillary into the first rf-only octopole. Ions

pass through a quadrupoe to a second octopole, 84 where they accumulate for 250-1000 ms.

Helium gas was introduced during accumulation to collisionally cool the ions before transfer

through a 200 cm rf-only octopole into an open cylindrical Penning ion trap. Octopole ion guides

were operated at 2.0 MHz and 240 Vp-p rf amplitude. Broadband frequency chirp excitation

accelerated the ions to a cyclotron orbital radius that was subsequently detected by two opposed

electrodes of the ICR cell to yield 8 Mword time-domain data sets. Multiple (100-300) time

domain acquisitions were summed for each sample, Hanning-apodized (magnitude-mdoe) or half

Hanning-apodized (absorption-mode), and zero-filled once prior to fast Fourier transform and

magnitude calculation.

Mass Calibration

Each FT-ICR mass spectrum was first calibrated with respect to a prior sample containing

Ultramark®, Met-Arg-Phe-Ala (MRFA) peptide, and caffeine (external calibration), and then

with respect to a homologous series of ions of high abundance common to all petroleum samples.

Masses for singly charged ions with relative abundance of > 6 of baseline rms noise were

extracted, converted to the Kendrick mass scale228 and exported to a spreadsheet for easier

identification of homologous series. Peak assignments were performed by Kendrick mass defect

analysis, 229 as previously described.

Baseline Correction Algorithm

How to accurately describe a baseline point is an important step in this algorithm. Most baseline

Page 64: Broadband Phase Correction Of Fourier Transform Ion Cyclotron Mass Spectra

47

identification methods255, 256, 250, 251 will initially differentiate the baseline regions that are

considered to be baseline noise from those that are considered to be signal according to different

noise criteria. These methods only work for sparse spectra where baseline noises can be easily

found between signals.

m/z171317121711171017091708

Minimum Points of Reversed Peak and magnitude<3σ

(A) Identification of Baseline Points

(B) Linear Interpolation

(C) Boxcar SmoothingBi-10, Bi-9, …..Bi-2, Bi-1, Bi, Bi+1, Bi+2, …,Bi+9, Bi+10

m/z171317121711171017091708

After Baseline Correction

Smoothed Baseline

(D) Subtraction

Figure 3.3. Schematic baseline flattening procedure. (A) Baseline identification from Original absorption spectrum; (B) Linear interpolation for empty spots between each two baseline points chosen in (A). (C) Boxcar smoothing of resulting (B) to yield smoothed baseline. (D) Spectrum with flat baseline after subtraction of resulting (C).

Page 65: Broadband Phase Correction Of Fourier Transform Ion Cyclotron Mass Spectra

48

However, for complex analysis, (e.g., petroleum and top-down proteomics’ spectra) it is

difficult to separate baseline noises from signals due to more doublets, triplets and isotopic

distributions (with more closed peaks). Our algorithm (Figure 3.3), which include baseline

identification, linear interpolation and boxcar smoothing, efficiently extracts the baseline from

spectrum without more considering of recognition of baseline.

We start to find base points of baseline by finding the reverse peak point with minimum

magnitude. To decide whether the ith point belongs to the baseline, the magnitude of ith point

was compared with the magnitude of front (i-1) and back (i+1) point and value of 3 of baseline

noise stand deviation (Figure 3.3, (A)). If the magnitude of ith point satisfies both Eq. 3.3 and

Eq. 3.4, it is considered as baseline:

Mi-1>Mi<Mi+1 [3.3]

Mi<3 [3.4]

In which Mi represents the magnitude of ith point, is the noise standard deviation. In this step,

we experimentally observed that one more zero-filling produces more minimum points of reverse

peaks and further facilitates the identification of baseline points. We suggested that two zero-

fillings during data processing should be used to produce absorption-mode spectrum. All

identified point is then stored in a baseline spectrum. The baseline spectrum has the same size of

data count as original absorption spectrum. The index of identified points in baseline spectrum is

also same as in original absorption spectrum. Except for identified points, all other points in

baseline spectrum are zeros. Next, we linearly interpolate values between each two nonzero

points to fill all zero spots in baseline spectrum (Figure 3.3, (B)). Further “boxcar smoothing’’

will remove short term variations, or “noise” and reduced the error in baseline spectrum (Figure

3.3, (C)). The whole “boxcar smoothing” 257 idea can be written as:

∑ ++

=−=

k

kj

jibk

ib ][)12(

1][ [3.5]

In which ][ib represents each baseline point after smoothing, ][ib is each baseline point before

smoothing, 2k +1 is the width of the window used for smoothing data. The value of k is set to a

constant value of 10 in order to get best compromise between calculation speed and final quality

of smoothed spectrum. Finally, the smoothed baseline spectrum is subtracted from original

absorption-mode spectrum to flatten baseline (Figure 3.3, (D)). Although only boxcar smoothing

method has been used to average our data, the choice of smoothing or windowing methods could

Page 66: Broadband Phase Correction Of Fourier Transform Ion Cyclotron Mass Spectra

49

be arbitrary in same case. The 3 in identification step of baseline points is more sensitive to

scale of baseline roll. For large baseline roll, the low frequency pass filter should be considered

to lower/flatten part of distortion before our baseline correction algorithm.

m/z172517201715

m/z17201715

Ribonucleases A 9.4T FT-ICR Absorption-Mode Mass Spectrum

After Cubic Modeled Baseline Correction

Figure 3.4. Zoom insets of Ribonucleases A FT-ICR absorption-mode spectrum after polynomial modeling baseline correction. Note the discontinuity due to cutting whole data to each small piece.

We have to mention that one advantage of our baseline correction algorithm is no need to

cut data as small segments (e.g., baseline modeling by different functions). For Polynomial

baseline modeling, we notice that the discontinuity appears in the edge of each segment (Figure

3.4). Also, we observed that low order polynomial function with small segment (less fitting data)

Page 67: Broadband Phase Correction Of Fourier Transform Ion Cyclotron Mass Spectra

50

performs is better than high order polynomial function with large segment. However, it is not

easy to come up a general polynomial model for different spectra due to experimental data

dependence of optimized parameters.

Computational Implementation

The baseline correction algorithm has been combined with automatic phase correction procedure

and implemented in our modular Predator ICR data acquisition system,117 in C program

operating under LabWindows/CVI (National Instruments, Austin, TX).

Results and Discussion

The baseline correction algorithm is tested experimentally by crude oil, bitumen as well as

Ribonucleases A. Figure 3.5 demonstrates the resulting flat baseline of crude oil and

Ribonucleases A absorption-model FT-ICR mass spectra. We can see that the periodic roll of

baseline in crude oil spectrum is romoved after applying baseline correction algorithm (Figure

3.5, top). For denser spectra, (e.g., Figure 3.5, bottom, for an absorption-mode FT-ICR mass

spectral segment from Rigonucleases A) this algorithm is working pretty well even the negative

dip is large in original spectrum. Note the increaing of peak height of isotopic distribution after

baseline correction.

Mass accuracy

Most important figure of merit for mass spectrometric performance should be measured peak

position but not peak height. For the present experimental FT-ICR mass spectra, elemental

compositions could be determined for all singly charged ions between 300 and 1000 Da with

peak heights greater than 6 of baseline rms noise. For comparison of absorption-mode spectra

with and without baseline correction, we chose the 8000 highest peaks in an ESI FT-ICR mass

spectrum of a crude oil. Figure 3.6 shows mass error distributions for the crude oil ESI FT-ICR

absorption-mode mass spectra of Figure 3.5 top. Both mass error distribution and rms error of

absorption-mode spectrum after baseline correction are similar to them of absorption spectrum

Page 68: Broadband Phase Correction Of Fourier Transform Ion Cyclotron Mass Spectra

51

without baseline correction. This result confirms that our baseline correction algorithm remains

the correct peak position with improved peak height measurement.

Identified Peaks

Because automated peak-picking algorithms typically limit consideration to peaks whose height

exceeds a specified multiple of the standard deviation of baseline rms random noise. As the

876875874873872871870869868867

Abundance 20

15

10

5

0

m/z876875874873872871870869868867

Abundance 20

15

10

5

0

Ribonuclease A 9.4T FT-ICR Absorption-Mode Mass Spectra

Original Absorption-Mode Spectrum

Absorption-Mode Spectrum

After baseline correction

793792791790789788

Crude Oil 9.4T FT-ICR Absorption-Mode Mass Spectra

Before Baseline Correction

After Baseline Correction

m/z793792791790789788

Figure 3.5. Zoom insets of Crude oil FT-ICR absorption-mode mass spectrum (top) and Ribonucleases A absorption-mode spectrum (bottom) showing the performance of baseline correction algorithm.

Page 69: Broadband Phase Correction Of Fourier Transform Ion Cyclotron Mass Spectra

52

Most abundant

8000 peaks

RMS error: 227ppb

200 600 1000

0.5

0.0

-0.5

Ma

ss

E

rro

r (p

pm

)

Before Baseline Smoothing

200 600 1000

0.5

0.0

-0.5

Ma

ss

E

rro

r (p

pm

)

After Baseline Smoothing

Crude Oil 9.4T FT-ICR Absorption-Mode MS (Mass Error)

Most abundant

8000 peaks

RMS error: 223ppb

Figure 3.6. Mass error distribution for Crude oil FT-ICR absorption-mode mass spectrum before baseline correction (top) and after baseline correction (bottom).

baseline exhibits large distortion, peak height will be significantly reduced by negative value of

baseline roll. Then a peak-picking algorithm can fail to identify true low-magnitude spectral

signals whose frequencies happen to fall near the minima of the "roll". Figure 3.7, top, shows

that the baseline correction algorithm successfully removes baseline for river bitumen APPI FT-

Page 70: Broadband Phase Correction Of Fourier Transform Ion Cyclotron Mass Spectra

53

ICR absorption-mode mass spectra. Identified peaks by automated peak picking algorithm (peak

height above 6 rms random noise) increase from 33 to 45 (Figure 3.7, bottom).

River Bitumen APPI 9.4T FT-ICR Absorption-Mode MS

748747

Abundance

4035302520151050

m/z748747

Abundance

4035302520151050

Before Baseline Smoothing

After Baseline Smoothing

m/z746.69746.61746.53746.45746.37746.29

Abundance

40

30

20

10

0

m/z

Abundance

40

30

20

10

0746.69746.61746.53746.45746.37746.29

W/O Baseline Smoothing

33 Peaks

With Baseline Smoothing

45 Peaks

Figure 3.7. Top: mass scale-expanded of river bitumen APPI FT-ICR absorption-mode mass spectrum showing the resulting flat baseline. Bottom: zoom insects of river bitumen APPI FT-ICR absorption-mode mass spectrum at nominal mass 746 m/z before and after baseline flattening. Note that the absorption spectrum after baseline correction improves discovery of low-abundance peaks.

Page 71: Broadband Phase Correction Of Fourier Transform Ion Cyclotron Mass Spectra

54

m/z15331532

Ribonucleases A 9.4T FT-ICR Mass Spectra

Magnitude-Mode

Absorption-Mode

9+

m/z153315321531

Ab

un

dan

ce150

100

50

0

m/z153315321531

Ab

un

dan

ce

100

50

0

150 Before Baseline Smoothing

After Baseline Smoothing

Figure 3.8. Top: 9+ charge state isotopic distribution of Ribonucleases A FT-ICR magnitude-mode (black) and absorption-mode spectrum (red) with flat baseline. Middle: 9+ charge state isotopic distribution of Ribonucleases A FT-ICR absorption-mode spectrum before baseline correction (red) with calculated isotopic distribution profile (black cross). Bottom: 9+ charge state isotopic distribution of Ribonucleases A FT-ICR absorption-mode spectrum after baseline correction (red) with calculated isotopic distribution profile (black cross).

Page 72: Broadband Phase Correction Of Fourier Transform Ion Cyclotron Mass Spectra

55

Isotopic Distribution

The 9+ charge state isotopic distribution of Ribonucleases A absorption-mode spectrum after

baseline correction displays very similar profile to that of normal magnitude-mode spectrum

(Figure 3.8, top). Note that absorption-mode spectrum naturally has narrower peaks width than

in magnitude-mode spectrum. IsoPro software was used to generate the calculated isotopic

distribution based on the amino acid sequences of Ribonucleases A. Compared the calculated 9+

charge state isotopic distribution, the isotopic distribution of absorption-mode spectrum after

baseline correction (Figure 3.8, bottom) demonstrates more accurate profile than before baseline

correction (Figure 3.8, middle).

Page 73: Broadband Phase Correction Of Fourier Transform Ion Cyclotron Mass Spectra

56

CHAPTER FOUR

EFFECTS OF ZERO-FILLING AND APODIZATION ON

FOURIER TRANSFORM ION CYCLOTRON RESONANCE

MASS SPECTRAL ACCURACY, RESOLUTION, AND SIGNAL-

TO-NOISE RATIO

Introduction

All forms of Fourier transform spectroscopy share common features: acquisition and digitization

of an interferogram or time-domain response to pulsed excitation, time-dependent weighting

(windowing; apodization), padding by addition of zeroes, and discrete Fourier transformation to

yield absorption and dispersion (or magnitude) spectra.258 In particular, Fourier transform ion

cyclotron resonance (FT-ICR) broadband mass spectra have until recently been reported in

magnitude mode, for which the effects of apodization and windowing are well known (see

below). With the recently achieved broadband phase correction to yield absorption-mode FT-

ICR mass spectra, 113, 114 it becomes necessary to understand the effects of apodization and zero-

filling on absorption-mode FT-ICR mass spectra, especially for dense spectra from complex

organic mixtures.

An apodization weighting function applied to a time-domain response to a pulsed

excitation is typically designed to reduce sharp edges at one or both ends of the signal to

eliminate Gibbs oscillations (literally, removal of "feet") on either side of the resulting FT

spectral peak. 259, 260 The most common use of apodization is to increase FT spectral peak height-

to-noise ratio, by weighting the earlier time-domain data points higher than the later time-domain

data. 261, 83, 262 A disadvantage of such apodization is that the peak width at half peak height is

broadened, with correspondingly lower resolving power. Zero-filling (see below) effectively

captures into the absorption-mode spectrum information originally residing in the (non-zero

filled) dispersion data, and thus increases absorption-mode peak height-to-noise ratio by a factor

of √2,78 with concomitant improvement in mass accuracy.261, 1

Page 74: Broadband Phase Correction Of Fourier Transform Ion Cyclotron Mass Spectra

57

Optimized apodization for improvement in resolution and peak height-to-noise ratio has

been extensively investigated,78, 263-265 and both apodization and zero-filling are in wide use for

FT-NMR data processing.261, 266 For FT-ICR MS, Chow218 showed that ICR frequency

measurement error depends on the separation between two peaks. FT-ICR magnitude-mode

spectral frequency/mass interpolation error has been evaluated for different various apodization

functions and zero-fills,267, 268, 262, 218, 5 and apodization optimization has also been discussed.

Aarstol, 1987 #82} Investigation of apodized absorption spectra has been limited to simulated or

sparse experimental cases.4, 269, 258. Finally, for FT-ICR MS, it is essential to eliminate

systematic errors (e.g., by dividing a spectrum into 30 separately calibrated segments). 9 in order

to take advantage of the effects of apodization on random noise. Here, we consider the effects of

twelve different apodization functions as well as multiple zero-fills on FT-ICR magnitude and

absorption spectra from bitumen and petroleum crude oil, with thousands of resolved peaks.

Experimental Methods

Sample Description and Preparation

Athabasca bitumen and distillate fraction (500 – 538 ○C) from Athabasca bitumen heavy vacuum

gas oil (HVGO) were obtained from the National Centre for Upgrading Technology (Devon,

Alberta, Canada). Bitumen and HVGO samples were diluted to yield a final concentration of

500 μg/mL by a 50:50 toluene methanol mixture spiked with 0.1% (by volume) formic acid

(positive ESI) and used without additional purification. All solvents were HPLC-grade, obtained

from Sigma-Aldrich Chemical Co. (St. Louis, MO).

Instrumentation

Middle Eastern crude oil distillate fractions were analyzed with a custom-built FT-ICR mass

spectrometer equipped with a 9.4 T horizontal 220 mm bore diameter superconducting solenoid

magnet operated at room temperature (Oxford Instruments, Abingdon, Oxfordshire, U.K.) and a

modular ICR data station (PREDATOR) facilitated instrument control, data acquisition and data

Page 75: Broadband Phase Correction Of Fourier Transform Ion Cyclotron Mass Spectra

58

analysis.270, 271, 227 Positive ions generated at atmospheric pressure were accumulated in an

external linear octopole ion trap84 for 250-1000 ms and transferred by rf-only octopoles to a 10

cm diameter, 30 cm long open cylindrical Penning ion trap. Octopoles were operated at 2.0

MHz and 240 Vp-p amplitude. Broadband frequency sweep (chirp) dipolar excitation (70 - 700

kHz at 50 Hz/μs sweep rate and 350 Vp-p amplitude) was followed by direct-mode image current

detection to yield 8 Mword time-domain data sets. Time-domain data sets were co-added (100

acquisitions), Hanning apodized, and zero-filled once before fast Fourier transform and

magnitude calculation.

Mass Calibration

ICR frequencies were converted to ion masses based on the quadrupolar trapping potential

approximation105, 106and internally calibrated with respect to the most abundant homologous

alkylation series differing in mass by integer multiples of 14.01565 Da (CH2) for each sample.

Peaks with relative peak abundance greater than 6 times the standard deviation of the baseline

noise were exported to a spreadsheet.

Data Processing

A discrete time domain ICR signal is multiplied by one of the twelve window functions listed in

Figure 4.1 padded by n zero-fills (each of which doubles the number of time-domain data Fast

Fourier transformation produces real (Re(ω)) and imaginary (Im(ω)) frequency-domain spectra.

Broadband phase correction 114 performs a linear combination of real and imaginary data to

generate an absorption frequency spectrum (A(ω)). Alternatively, a magnitude-mode frequency

spectrum (M(ω)) is calculated from √ ([Re(ω)]2 + [Im(ω)]2). Both absorption and magnitude

frequency spectra are converted to m/z with a two term calibration equation. 105, 106.

Computational Implementation

Each of various apodizations and zero-fills was programmed in C. Each function was

implemented as an extension of our ICR data system, 117 in C operating under LabWindows/CVI

(National Instruments, Austin, TX). Data processing steps are shown in Figure 4.2. Half

apodization and full apodization have been tested for both magnitude and absorption spectra.

Page 76: Broadband Phase Correction Of Fourier Transform Ion Cyclotron Mass Spectra

59

Cosine

Full Window Half Window

0 1n/(N-1)

Mag

nit

ud

e o

f W

ind

ow

Fu

nc

tio

ns

0

1

Hamming

Hanning

Gaussian

=0.2

Blackman

Blackman-Harris

Bartlett-Hanning

Triangle

Four-Term Blackman-

Harris

Kaiser

α=1.2

Blackman-Nuttall

Kaiser-Bessel

Figure 4.1. Twelve full window (left) and half window (right) functions. N is the time-domain data size before zero-filling and n is the index for each data 0 ≤ n ≤ N-1.

Page 77: Broadband Phase Correction Of Fourier Transform Ion Cyclotron Mass Spectra

60

Time (s)0 3.5

x x

m/z700650600550500450400350

Fourier Transform

Imaginary SpectrumReal Spectrum

Magnitude Spectrum

Phase Correction

Absorption Spectrum

Magnitude Spectrum Absorption Spectrum

Zero-Fill

Time (s) 70

Zero-Fill

Time (s) 70

Full Half

m/z700650600550500450400350

Figure 4.2. Data processing steps for FT-ICR magnitude-mode and absorption-mode FT-ICR spectra.

Page 78: Broadband Phase Correction Of Fourier Transform Ion Cyclotron Mass Spectra

61

448.41448.39448.37448.35448.33448.31

m/z448.41448.39448.37448.35448.33448.31

Crude Oil Distillate 9.4 T ESI FT-ICR MS

(Magnitude-Mode Spectra)

Half Hanning

Apodization

Full Hanning

Apodization

Figure 4.3. Mass scale-expanded segments of electrospray ionization (ESI) FT-ICR magnitude mass spectra of distillated fraction from crude oil for half Hanning (top) and full Hanning (bottom) apodization functions. The magnitude spectrum with half hanning apodization exhibits greater width at each peak base.

Results and Discussion

Full Apodization vs. Half Apodization

Although magnitude-mode peaks with half apodization exhibit better resolution at half-

maximum peak height than with full apodization (Figure 4.3), half apodized peaks are much

broader at the base. Thus, mass accuracy for an FT-ICR magnitude spectrum of crude oil

distillate with half apodization is a factor of 2 worse than for full apodization (Figure 4.4). For

sparse spectra (e.g., Figure 4.5, for an absorption-mode FT-ICR mass spectral segment from a

Page 79: Broadband Phase Correction Of Fourier Transform Ion Cyclotron Mass Spectra

62

crude oil distillate) both half Hanning and full Hanning apodization yield symmetric peaks, but

full apodization introduces negative lobes on both sides of each peak. For denser spectra (e.g.,

Figure 4.6, for an absorption-mode FT-ICR mass spectral segment from bitumen), full Hanning

apodization severely distorts the shapes of closely space peaks. In summary, half apodization is

best suited for absorption-mode display, whereas full apodization is best for magnitude-mode

spectra.

Ma

ss

Err

or

(pp

m)

2

1

-1

-2

0

300 400 500 600 700

m/z

Crude Oil Distillate 9.4 T ESI FT-ICR MS

Magnitude-Mode

With full Hanning window 1700 peaks 130ppb

With half Hanning window 1700 peaks 270ppb

Figure 4.4. Mass error distribution for broadband magnitude-mode FT-ICR mass spectra for a distillate fraction of crude oil for half Hanning (black) and full Hanning (red) apodization functions.

Page 80: Broadband Phase Correction Of Fourier Transform Ion Cyclotron Mass Spectra

63

Half Hanning

Apodization

Full Hanning

Apodization

448.41448.39448.37448.35448.33448.31

m/z448.41448.39448.37448.35448.33448.31

Crude Oil Distillate 9.4 T ESI FT-ICR MS

(Absorption-Mode Spectra)

Figure 4.5. Mass scale-expanded segments of FT-ICR absorption spectra for a distillated fraction from crude oil for half Hanning (top) and full Hanning (bottom) apodization functions.

m/z730.72730.68730.64730.60730.56

Half Hanning

Apodization

Full Hanning

Apodization

Bitumen 9.4 T ESI FT-ICR MS

(Absorption-Mode Spectra)

Figure 4.6. Mass scale-expanded segments of FT-ICR absorption spectra for bitumen for half Hanning (top) and full Hanning (bottom) apodization functions.

Page 81: Broadband Phase Correction Of Fourier Transform Ion Cyclotron Mass Spectra

64

Half C osine

Half Hamming

Half Hanning

Half B lackman

Half T r iang le

Half B art let t - Hann

W / O A p od izat ion

Cosine

Hamming

Hanning

Blackman

Triangle

Bartlett-Hann

W/O Apodization

Absorption-Mode

Magnitude-Mode

Zero-Fills

RM

S M

ass E

rror (p

pm

)R

MS M

ass E

rror (p

pm

) 0.30

0.10

0.25

0.20

0.15

0.05

0.30

0.05

0.25

0.15

0.10

0 1 2 3

0 1 2 3

0.20

Figure 4.7. Root-mean-square mass error for broadband magnitude-mode (top) and absorption-mode (bottom) FT-ICR mass spectra for a distillate fraction of crude oil, following 0, 1, and 2 zero-fills, for each of several apodization functions. The six apodization functions resulting in the least rms error are shown.

Cosine

Hamming

Hanning

Blackman

Triangle

Bartlett-Hann

W/O Apodization

Half Cosine

Half Hamming

Half Hanning

Half Blackman

Half Triangle

Half Bartlett-Hann

W/O Apodization

Peak H

eig

ht-to

-Nois

e R

atio

Zero-Fills

Absorption-Mode

Magnitude-Mode

900

500

800

700

600

400

900

500

800

700

600

400

0 1 2 3

0 1 2 3

Figure 4.8. Peak height-to-noise ratio for peak at m/z 450.4094 (C32H52N1) for a distillate fraction of crude oil, following 0, 1, and 2 zero-fills, for each of several apodization functions.

Page 82: Broadband Phase Correction Of Fourier Transform Ion Cyclotron Mass Spectra

65

Half Cosine

Half Hamming

Half Hanning

Half Blackman

Half Triangle

Half Bartlett-Hann

W/O Apodization

Cosine

Hamming

Hanning

Blackman

Triangle

Bartlett-Hann

W/O Apodization

Absorption-Mode

Magnitude-Mode

Zero-Fills

Resolv

ing P

ow

er

(m/∆

m50%)

Resolv

ing P

ow

er

(m/∆

m50%)

600,000

1,200,000

1,000,000

800,000

200,000

800,000

600,000

400,000

0 1 2 3

0 1 2 3

Figure 4.9. Average mass resolving power for 6 peaks above 6σ of baseline noise at nominal m/z 450 for a distillate fraction of crude oil, following 0, 1, and 2 zero-fills, for each of several apodization functions.

Mass Accuracy

For the present experimental FT-ICR mass spectra, elemental compositions could be determined

for all singly charged ions between 300 and 700 Da with peak heights greater than 6 of baseline

rms noise. For comparison of magnitude and absorption spectra, we chose the 1700 highest

peaks in an ESI FT-ICR mass spectrum of a crude oil distillate. The rms mass errors for

absorption-mode spectra following each of six different apodizations and 0, 1 or 2 zero-fills is

plotted in Figure 4.7. Half apodization produces consistently better mass accuracy for

absorption-mode than for full-apodized magnitude-mode spectra. Moreover, mass accuracy for

absorption-mode (but not magnitude-mode) improves significantly after one time-domain zero-

fill, because the first zero-fill effectively captures the information contained in the non-zero-

filled dispersion spectrum.78 For magnitude mode, full apodization yields higher mass accuracy

than half apodization, whereas the converse applies to absorption mode. A second zero-fill yields

no improvement in mass accuracy for either magnitude-mode or absorption-mode spectra.

Page 83: Broadband Phase Correction Of Fourier Transform Ion Cyclotron Mass Spectra

66

(Moreover, in the absence of noise, the first zero-fill correctly interpolates each additional

spectral data between two adjacent original data, whereas second or higher zero-fills yield a

convolved peak shape that no longer exactly interpolates the additional data values. 1

Signal-to-Noise Ratio

For the same experimental time-domain raw data as for Figure 4.7, Figure 4.8 shows the peak

height-to-noise ratio for peak at m/z 450.4094 (C32H52N1), following each of six different

apodizations and 0, 1 and 2 zero-fills. Each of the full apodization functions reduces magnitude-

mode peak height-to-noise ratio, because the first half of the apodization function reduces the

time-domain signal amplitude. Although the unapodized absorption-mode spectral peak height -

to-noise ratio is lower than that for an unapodized magnitude-mode spectrum, absorption-mode

signal-to-noise ratio improves dramatically after half apodization, because half-apodization

lowers the time-domain signal amplitude most at the end of the acquisition period, where the

noise-to-signal ratio is highest. In fact half apodized absorption-mode spectra typically exhibit

higher signal-to-noise ratio than full apodized magnitude-mode spectra. Absorption-mode peak

height-to-noise ratio is virtually independent of the choice of half apodization function, whereas

magnitude-mode peak height-to-noise ratio varies significantly among different full apodization

choices. Finally, peak height-to-noise ratio for both absorption-mode and magnitude-mode

spectra increases slightly after zero-filling, presumably because interpolation gives higher digital

resolution to help locate the peak center.

Resolving Power

For the same experimental time-domain raw data as for Figure 4.7, six peaks greater in

magnitude than 6s of baseline noise at nominal m/z 450 are chosen for test purpose. Figure 4.9

shows that the average resolving power is consistently as much as 50% higher for absorption

than for magnitude spectra, and depends strongly on the choice of apodization function for both

spectra. As for multiple zero-fillings, the resolving power improves slightly on zero-fillings for

both absorption and magnitude-mode spectra, presumably due to much better digital resolution

of peak height and peak width. Half cosine (for absorption-mode) and full cosine (for magnitude-

mode) apodization provide the highest resolving power.

Page 84: Broadband Phase Correction Of Fourier Transform Ion Cyclotron Mass Spectra

67

CHAPTER FIVE

PHASE SPECTRA OF FOURIER TRNSFORM ION

CYCLOTRON RESONANCE MASS SPECTROMETRY

Introduction

In Fourier transform analysis, an ideal signal f(t) without any time delay could be complex

Fourier transformed to F(ω)=A(ω)+iD(ω). We call real part of F(ω) as absorption-mode

spectrum and imaginary part as dispersion-mode spectrum. For a signal with time delay, f(t-T),

The resulting Fourier transformation is F(ω)eiφ(ω). 1 In which ejφ(ω) =cos(φ(ω)) + isin(φ(ω)), so

F(ω) eiφ(ω) =[A(ω) cos(φ(ω)) + D(ω) sin(φ(ω))] +i[D(ω) sin(φ(ω)) - A(ω)cos(φ(ω)),]. The real

part of F(ω) eiφ(ω) is linear combination of A(ω) and D(ω). We should note that both cases the

sqrt(real2+imaginary2) is equal to sqrt[A(ω)2+D(ω)2] ,which we call magnitude spectrum, it is

independent any phase shift. Compared to conventional magnitude-mode display, absorption-

mode FT-ICR mass spectrum offers a gain up to a factor of 2 in mass resolving power80-82 and

increased mass measurement accuracy. 114, 9 If we can get the phase value for signal with time

delay, we can recover the F(ω) by F(ω) eiφ(ω) * e-iφ(ω) , which is Fourier transformation of

unshifted (ideal) version of f(t), and get absorption-mode spectrum, A(ω), from real part of F(ω).

So the key to this phase correction method is the explicit knowledge of a function of frequency

φ(ω), which we call the phase spectrum of FT-ICR.

Most recent developments of phase correction of FT-ICR mass spectrum are based on

calculation of phase spectrum. Except for linear phase spectrum used in phase correction for

collision mode analysis, 112 broadband phase correction algorithm114 can automatically account

for first- and second-order phase accumulation from excitation parameters and iterates the zero-

order term. Another approach for phase correction is to assume a quadratic relation between

phase and frequency, and iteratively solve for zero-, first-, and second-order coefficients. 115 All

these methods can yield an absorption-mode spectrum with higher resolving power than

conventional magnitude-mode spectrum. However, the physical explanation and explicit form of

Page 85: Broadband Phase Correction Of Fourier Transform Ion Cyclotron Mass Spectra

68

the phase relation/phase spectrum is not known in all prior research. The discovery of physically

justified, experimentally tested, closed-form or or quasi-closed-form expressions would result in

a new and more efficient broadband phase correction algorithm. Here, we derive analytical phase

spectrum for linear-sweep excitation and detection signal of FT-ICR by using method of

stationary phase method. Phase correction of FT-ICR mass spectrometry directly by using phase

spectrum is demonstrated experimentally. Compared to automated broadband phase correction

algorithm, analytical phase spectrum accurately calculate phase value for each frequency and

provide fast and robust phase correction for FT-ICR mass spectra from linear-sweep excitation.

Moreover, this phase spectrum could be modified to phase swift excited signal.

Method of Stationary Phase

The method of stationary phase, which was developed by Lord Kelvin in the 1800s to solve

integrals encountered in the study of hydrodynamics, is useful in evaluating integrals of the

following type:

dtetAE tif∫=+∞

∞−

)()( [5.1]

If we put t = tω+ζ, where f(t) is stationary for t = tω, ζ is very small number, we can develop

f(t) in a Taylor expansion:

....)('''6

)(''2

)(')()()(32

++++=+= ωωωωω

ξξξξ tftftftftftf [5.2]

The definition of stationary point is tω, where f′(t)=0 at t = tω. While A(t) changes comparatively

slowly, f(t) changes by 2π and A(t) is supposed to vary by only a small fraction of itself. That is,

the integrand approximates a constant A(t) multiplied by a rapidly varying function eif(t), which

varies between +1 and -1. Therefore, destructive interference between the various contributions

to the integral will make it vanish, except for those values of t for which f(t) is stationary.

Stationary values do not cancel each other in general and they are not symmetrically distributed

Page 86: Broadband Phase Correction Of Fourier Transform Ion Cyclotron Mass Spectra

69

around zero, for any function f(t). At the stationary value of t, the integrand is approximately one

constant, A(t), multiplied by another, eif(t).

dtetAdtetAEtftfi

tif ∫=∫=∞+

∞−

+∞+

∞−

))(()())(''

2)((

)(

2

ωωξ

ω

ξωω

ωω

ξ

ω

ξ

ω deetAdteetAtfi

tiftfi

tif

∫=∫=∞+

∞−

∞+

∞−

)(''2)(

)(''2)(

22

)(*)(

ξξξ ωωωω dtfitfetA tif )])(''

2

1sin())(''

2

1[cos()( 22)( +∫=

∞+

∞−

)4

)((4)( )(

)(''

2

)(''

2*)(

π

ω

ω

π

ω

ω

ωω

ππ +

==tfii

tif etAtf

etf

etA [5.3]

Phase Spectrum

Chirp/Sweep Excitation Signal

The chirp/sweep excitation signal (Figure 5.2, (A)) for Fourier Transform ion cyclotron

resonance could be represented as

)]2

1cos[()( 2

0tSRtEtE

I∗∗+= ω [5.4]

In which E0 is the amplitude of excitation signal, ωI=2πfI is the initial frequency and SR

represents the sweeprate. The Fourier transform, E(ω), of the excitation signal E(t) is

dtetEET

ti∫= −

0

*)()( ωω [5.5]

Because of Euler’s formula

2

)cos(ixix ee

x−+

= [5.6]

Page 87: Broadband Phase Correction Of Fourier Transform Ion Cyclotron Mass Spectra

70

dteeeE

dteteET

titSRtitSRtiT

ti II

∫ +=∫= −∗∗+−∗∗+

0

)2

1()

2

1(

0

0

*][2

*)()(22

ωωωωω

dteeET ttSRtittSRti II

∫ +=+∗∗+−−∗∗+

0

)2

1()

2

1(

0 ][2

22 ωωωω

∫∫ +=+∗∗+−−∗∗+ T ttSRtiT ttSRti

dteE

dteE II

0

)2

1(

0

0

)2

1(

0

22

[2

[2

ωωωω

[5.7]

We can apply the stationary method here in order to solve the Eq. 5.7. For abbreviation,

we introduce two notations.

ttSRttfI

ωωω −∗∗+= 2

1

2

1),( [5.8]

)2

1(),( 2

2ttSRttf

Iωωω +∗∗+−= [5.9]

According to the stationary point definition (The definition of stationary point is

tω, where f′(t)=0 at t = tω).

0),('11 1

=−∗+==

ωωω ωωtSRtf

Itt

So SR

t Iωω

ω

−=

1 [5.10]

0),('22 2

=−∗−−==

ωωω ωωtSRtf

Itt

So SR

t Iωω

ω

−−=

2 [5.11]

We can see that tω1 is always positive value and tω2 is always negative value. The second

exponential in the Eq.5.7 has no stationary point in the time range [0,T]. If the amplitude e0 is

slowly varying or constant, the contribution to E(ω) from second exponential of Eq.5.7 can be

Page 88: Broadband Phase Correction Of Fourier Transform Ion Cyclotron Mass Spectra

71

neglected, leaving

dteE

dteE

ET ttSRtiT

tfi I

∫=∫=−∗∗+

0

)2

1(

0

0

),((0

2

1

2][

2)(

ωωωω [5.12]

Taylor expansion of f1(ω,t) is developed as

),(''2

),('),(),(11

2

11111 ωωω ωξωξωω tftftftf ++= [5.13]

So ∫=T

tfi dteE

E0

),((0 ][2

)( 1 ωω (applying the stationary point)

∫=−T tttfi

tif dteeE

0

))(,(''2

1

),(0

2111

11

2

ωωω

ωω [5.14]

For tω1 not too close to 0 or T the finite integral in above equation can be replaced by one

extending from -∞ to ∞ without loss of accuracy. Then

∫=−T tttfi

tif dteeE

E0

))(,(''2

1

),(0

2111

11

2)(

ωωω

ωωω

)

4),((

11

0

11

),(''2

πω

ω

ω

ωπ +

=tfi

etf

E

)

4),((

0

11

2

πω ωπ +

=tfi

eSR

E [5.15]

Phase spectrum for chirp/sweep excitation signal can be represented as

4

),()(11

πωωϕ ω += tfE

[5.16]

Page 89: Broadband Phase Correction Of Fourier Transform Ion Cyclotron Mass Spectra

72

Submit Eq.5.8 and Eq.5.10 into Eq.5.16

4

)(2

1)( 2 πωωωωωωωωωϕ +

−−

−∗∗+

−=

SRSRSR

SRIII

IE

42

2 2222 πωωωωωωωωωω+

−−

+−+

−=

SRSRSRIIIII

42

22222 2222 πωωωωωωωωωω+

+−+−+−=

SRIIIII

42

22222 2222 πωωωωωωωωωω+

+−+−+−=

SRIIIII

42

)( 2 πωω+

−−=

SRI [5.17]

Depend on the sweep direction, the final phase spectrum for chirp/sweep excitation signal

for each specific ωi is

42

)()(

2 πωωωϕ ±−−

=SR

Ii

iE [5.18]

The option + means that the sweep of frequency is low to high and the option – means that the

sweep of frequency is high to low.

After we solve the phase spectrum for excitation signal, we can see that the excitation

signal after inverse Fourier Transformation can be represented as

]42

)(cos[()(

2

0

πωωω mSR

tEtE Ii

i

−−= [5.19]

Detected Signal

We start to discuss the detected signal from the Lorentz force. When each ion cloud is

Page 90: Broadband Phase Correction Of Fourier Transform Ion Cyclotron Mass Spectra

73

transferred into ICR cell with a spatially uniform magnetic field, B (Figure 5.1), motion of ion

cloud will subject to a force given by Lorentz force which is described in Eq.5.20.

Excita

tion

Detection

B

x

y

Figure 5.1. Schematic ICR cell showing the excitation, detection and magnitude field directions

BvqEqFvvvv

×+= [5.20]

dt

vdmamF

vvv

== [5.21]

In which m, q, and v are ionic mass, charge, and velocity, and the vector cross product term

means that the direction of the magnetic component of the Lorentz force is perpendicular to the

plane determined by v and B. If the ion maintains constant speed, then the Lorentz force bends

the ion path into a circular motion. If we substitute Eq. 5.21 into Eq. 5.20 and consider x and y

directions, we can get more detailed ionic motion equation.

BjvivqqEidt

vdm

yx×++= )ˆˆ(ˆ

v

BivjvqqEidt

jvivdm

yx

yx )ˆˆ(ˆ)ˆˆ(−+=

+ [5.22]

Page 91: Broadband Phase Correction Of Fourier Transform Ion Cyclotron Mass Spectra

74

Eq. 5.22 can be further separated as two ion motion equations along with x and y directions

iqBviqEdt

idvm

y

x ˆˆˆ

−= [5.23]

jqBvdt

jdvm

x

y ˆˆ

= [5.24]

Second derivative of vy can be acquired from first derivative of Eq. 5.24.

dt

dv

m

qB

dt

vm

qBd

dtdt

dvd

v x

x

y

y===″ )()(

[5.25]

Substitute vx from Eq. 5.24 into Eq. 5.25 to yield

yyv

m

Bq

m

qEqBv

2

22

2−=″

[5.26]

Because of ωi = (qi B)/mi, we can simplify Eq. 5.26 to Eq. 5.27.

2

2

2

i

i

yiy

m

BEqvv =+″ ω [5.27]

Write the Fourier Transform of yv as

yv̂ and substitute excitation signal E(t) (Eq. 5.4) into Eq.

5.27 to yield

∫ ∗∗+=+−∞+

∞−

− dttSRtEem

Bqvv

I

ti

i

i

yiy)]

2

1cos[(ˆˆ 2

02

2

22 ωωω ω

Page 92: Broadband Phase Correction Of Fourier Transform Ion Cyclotron Mass Spectra

75

∫ ∗∗+−

=∞+

∞−

− dttSRtEem

Bqv

I

ti

ii

i

y)]

2

1cos[(

)(ˆ 2

0222

2

ωωω

ω [5.28]

Similar to excitation signal, we can derive the phase spectrum for yv̂ by the stationary phase

method. The resulting yv̂ is

42

)(

02

2 2

2)(8ˆ

πωωπωωω

±−−

−= SR

iii

i

y

I

eSR

Em

Bqv [5.29]

From Eq. 5.29, we can see that the phase spectrum for ion motion is exact same as

excitation signal.

42

)()(

2 πωωωϕ ±−−

=SR

Ii

i

Inverse Fourier transformation of yv̂ will produce the asymptoti solution for

yv . Since the

phase spectrum will not change during Fourier transform, the velocity of ion motion, yv , along

with detection direction will have same phase spectrum as yv̂ . The image current I[t] (detected

signal) is proportional to yv [t], the phase spectrum for detected signal has same expression as

excitation signal at t=0. However, we only start to observe detection signal at tobs due to time

dispersion for linear frequency-sweep and small time delay (tdelay) for avoiding induced

excitation signal (Figure 5.2, (A)). The final phase spectrum for detected signal at t=tobs is

42

)()(

2 πωωωωϕ ±+−−

=obsi

Ii

iDt

SR [5.30]

Figure 5.2, (B) shows the phase spectrum profiles of linear frequency-sweep excitation (blue)

Page 93: Broadband Phase Correction Of Fourier Transform Ion Cyclotron Mass Spectra

76

and detected signal(red) for low to high frequency sweep.

Fre

quency

Detecttdelay

E(t)

fI

0

0

ff

tobs

Frequency-Sweep Excitation Signal

Detected Signal

D(t)

Time (s)

- φD (ω)

- φE (ω)

φ(ω

) (r

ads)

0

fI ffFrequency (Hz)

A)

B)

ω=2πf

Figure 5.2. A) Plots of time-domain linear frequency-sweep excitation signal (upper) and detected signal (bottom) vs. time. B) Phase spectrum profiles (phase vs. frequency) of linear-sweep excitation and detected signal. Here we only plot the phase spectrum for low-to high frequency-sweep for demonstration purpose.

Page 94: Broadband Phase Correction Of Fourier Transform Ion Cyclotron Mass Spectra

77

Experimental Methods

Sample Preparation

Stock solutions of either a distillate fraction (500-538 ○C) from an Athabasca bitumen or a North

American crude oil were dissolved in 50:50 (v/v) toluene/methanol to a final concentration of 1

mg/mL. Each sample was further diluted to 0.25 mg/mL with 50:50 (v:v) toluene/methanol and

0.1% Formic acid to facilitate protonation during electrospray ionization.

Instrumentation

All sample was analyzed with a custom-built FT-ICR mass spectrometer equipped with a 9.4

Tesla horizontal 220 mm bore diameter superconducting solenoid magnet operated at room

temperature (Oxford Instruments, Abingdon, Oxford shire OX13 5QX, UK) and a modular ICR

data station (PREDATOR) facilitated instrument control, data acquisition and data analysis. 226,

227 Positive ions generated at atmospheric pressure were accumulated in an external linear

octopole ion trap84 for 250-1000 ms and transferred into ICR cell. Linear broadband frequency

sweep (chirp) dipolar excitation (10 – 900 kHz at 50 Hz/μs sweep rate and 350 Vp-p amplitude)

or SWIFT excitation was followed by direct-mode image current detection to yield time-domain

data sets with 16 Mwords. The 200 collected individual transients of 2.8s for distillate of crude

oil) or 5.6 s for crude oil were averaged, apodized with a full-Hanning (magnitude spectrum) or

half-Hanning (absorption spectrum 114) weight function, and zero-filled once prior to fast Fourier

transformation. Each m/z spectrum was internally calibrated with respect to an abundant

homologous N1 series, and then walking calibration was applied to the spectrum was dividend

into equal consecutive segments which contained at least two members of a homologous

alkylation series.

Mass Analysis

Each FT-ICR mass spectrum was first calibrated with respect to a prior sample containing

Ultramark®, Met-Arg-Phe-Ala (MRFA) peptide, and caffeine (external calibration), and then was

internally calibrated with respect to an abundant homologous N1 series, and then walking

Page 95: Broadband Phase Correction Of Fourier Transform Ion Cyclotron Mass Spectra

78

calibration was applied and the spectrum was dividend into 56 (for crude oil) equal consecutive

segments which contained at least two members of a homologous alkylation series. Masses for

singly charged ions from m/z 300 to m/z 700 (distillate of crude oil) or from m/z 200 to m/z

1000 (crude oil) with relative abundance of > 6 of baseline rms noise were extracted, converted

to the Kendrick mass scale228 and exported to a spreadsheet for easier identification of

homologous series. Peak assignments were performed by Kendrick mass defect analysis, 229 as

previously described.

4*2

)( 2 πππϕ +∗+−

−= ftSR

ffobs

s

D

After Applying Phase Spectrum

800700600500400300m/z

Distillate of Crude Oil 9.4T FT-ICR Mass Spectra

Before Phasing

After Extra Constant Phase

136.04

*2)( 2

++∗+−

−=πππϕ ft

SR

ffobs

s

D

800700600500400300

Figure 5.3. Distillate of crude oil 9.4T FT-ICR mass spectra. Top: Raw real data following Fourier transform of discrete time-domain signal. Middle: absorption-mode spectrum (obtained after applying analytical phase spectrum). Bottom: Optimized absorption-mode spectrum after fine tune of constant term. Note: additional 0.136 radians in constant term.

Page 96: Broadband Phase Correction Of Fourier Transform Ion Cyclotron Mass Spectra

79

SWIFT Waveform

The procedure of building SWIFT waveform was described in several literatures. 75-77 The

optimized algorithm has been implemented in our modular ICR data station (PREDATOR).

Basically we build the magnitude modulation with constant amplitude and smooth it twice by

window function. The corresponding phase modulation will be automatically calculated based on

the information of magnitude modulation. 272, 273 Both magnitude and phase modulation are then

subjected to inverse Fourier transform to give the time-domain excitation waveform. The

schematic procedure of SWIFT is shown in Figure 5.7, top. Finally, converted analog signal

from the SWIFT waveform is amplified and applied to excitation plate.

Computational Method

Phase correction by phase spectrum can be simply implemented in software. Figure 5.3 shows

the three major steps in phase spectrum method. First, the time-domain will be subjected to half

apodization, one zero-filling and Fourier transformation to yield real spectrum (Figure 5.3, top)

and imaginary spectrum. Second, derived phase spectrum (Eq. 5.30) is applied to both real and

imaginary spectrum according to )Im()](sin[)Re()](cos[)( ωωϕωωϕωDD

A −= and

)Im()](cos[)Re()](sin[)( ωωϕωωϕωDD

D += (Figure 5.3, middle). Although we have

derived the analytical phase spectrum for detected signal from linear frequency-sweep excitation,

we have noticed that there is small phase variation ±150 in constant term π/4 for different

samples. These phase differences between analytical phase spectrum and real phase most likely

come from imperfect experimental conditions, like, electronic circuit, non-chemical noise or

nonzero initial phase value (ion clouds are not perfectly aligned with excitation plate). For best

result, we add a third step to fine tune the constant term within ±150. We simply vary additional

constant term from -150 to 150 in 1-degree increments until the entire spectrum is optimally

"phased". Figure 5.3, bottom shows the optimized absorption-mode spectrum after fine tune of

constant term.

Results and Discussion

Page 97: Broadband Phase Correction Of Fourier Transform Ion Cyclotron Mass Spectra

80

Phase Spectrum Method vs. Automated Broadband Phase Correction

Calculated Phase. Although previous automated broadband phase correction algorithm has no

explicit form for phase function, it works pretty well in term of improvement of resolving power

and mass accuracy. We compared the phasing results from phase spectrum method and

automated broadband phase correction algorithm for distillate of crude oil FT-ICR mass spectra

so as to fully understand both phase correction methods. Figure 5.4, (A) shows the distillate of

crude oil of FT-ICR absorption-mode spectrum from broadband phase spectrum method (lower)

and phase correction algorithm (upper). Phase value vs. frequency for phase spectrum method

(red) and phase correction algorithm (blue) are plotted in Figure 5.4, (B). Although absolute

phase value from two methods is totally different (Figure 5.4, (B)), two curves looks like parallel

to each other. Calculated phase difference between these two methods is very close to 2*π*1033.

It does make sense because two phase value different at 2nπ will have exact position in ICR cell

and both phase value will recover absorption-mode spectrum perfectly. Further quantitative

phase differences are evaluated by ε. Figure 5.4, (C) shows the plot of ε vs. frequency and clearly

exhibit the phase error associated with broadband phase correction algorithm. In broadband

phase correction, we pick the frequency point with highest amplitude within magnitude-mode

spectrum as initial point and add same amount of phase difference for increasing each frequency

data point or reduce same amount of phase difference for decreasing each frequency data point.

This is reason that the highest value of ε (at initial point) is zero and other value of are linearly

decreased toward both ends of frequency spectrum.

Resolving Power and Mass Accuracy. For comparison of resolving power and mass accuracy,

we chose the exact same 1800 highest peaks in an ESI FT-ICR mass spectrum of a distillate of

crude oil for both phase spectrum method and broadband phase correction algorithm. Although

phase error associated with broadband phase correction algorithm have observed, the error is

relative small (Figure 5.4, (C)). Both resolving power (Figure 5.5, top) and mass error

distribution (Figure 5.5, bottom) from phase correction algorithm (upper) and phase spectrum

method (lower) demonstrate very similar result.

Unresolved Peaks. Crude oil of FT-ICR absorption-mode spectrum by phase spectrum method

is not only getting better resolving power than magnitude-mode spectrum, but also getting more

Page 98: Broadband Phase Correction Of Fourier Transform Ion Cyclotron Mass Spectra

81

Phase spectrum method

Phase correction algorithm

200 250 300 350 400 450

Frequency (KHz)

(rad

s)

-0.006

-0.004

-0.002

0

0.002

0.004

200 250 300 350 400 450

0

-4000

2000

4000

6000

8000

10000

-2000

- φ(ω)Phase spectrum

- φ(ω)Phase Correction

φ(ω

) (r

ad

s)

A)

B)

C)

m/z600400300 700500

600400300 700500

Distillate of Crude Oil 9.4T FT-ICR Mass Spectra (Absorption-Mode)

2*1033*π

= φ(ω)phase spectrum- φ(ω)phase correction-2*1033*π

Figure 5.4. Top: Distillate of crude oil of FT-ICR absorption-mode spectra produced by broadband phase correction algorithm (upper) and phase spectrum method (lower). Middle: Plot of calculated phase value vs. frequency from broadband phase correction algorithm (blue) and phase spectrum method (red). Bottom: Plot of phase difference, ε, vs. frequency. Note that every ε value is less than 1 degree (0.0175 radians)

Page 99: Broadband Phase Correction Of Fourier Transform Ion Cyclotron Mass Spectra

82

m/z448.41448.39448.37448.35448.33448.31448.29

765000

758000

748000 767000

760000

448.41448.39448.37448.35448.33448.31

758000

754000

798000 797000

758000

448.29

-0.5

0

0.5

200 300 400 500 700600

Mass E

rror

(ppm

)

m/z

-0.5

0

0.5

200 300 400 500 700600

Mass E

rror

(ppm

)

Phase correction algorithm

Phase spectrum method

1800 peaks

110 ppb

1800 peaks

110 ppb

A)

B)

Distillate of Crude Oil 9.4T FT-ICR Mass Spectra

Phase correction algorithm

Phase spectrum method

Figure 5.5. TOP: Mass scale-expanded segments of distillate of crude oil of FT-ICR absorption-mode spectra from broadband phase correction algorithm (upper) and phase spectrum method (lower). Bottom: Mass error distribution of absorption-mode spectra from phase correction algorithm (upper) and phase spectrum method (lower).

Page 100: Broadband Phase Correction Of Fourier Transform Ion Cyclotron Mass Spectra

83

resolved peaks. Closed mass doublets partially or unresolved in magnitude mode become

resolved in absorption-mode (Figure 5. 6), thereby increasing the number of correctly assigned

elemental compositions.

915.86915.82915.78915.74915.7915.66

m/z915.86915.82915.78915.74915.7915.66

Crude Oil ESI 9.4T FT-ICR MS

Absorption-Mode

Magnitude-Mode

C66H96N113C

C61H92N1S213C

3.4mDa3.4mDa

C63H111O3

C59H115N2S2

C63H100N1S113C

C64H98N1S113C

C64H98N1S113C

C63H100N1S113C

C59H115N2S2

Figure 5.6. Magnitude and absorption electrospray ionization 9.4 T FT-ICR mass spectra for the same crude oil data from phase spectrum method. The absorption display clearly resolves several mass doublets (compositions differing by C3 vs. SH4, 0.0034 Da) that appears as a single magnitude mode peak.

Page 101: Broadband Phase Correction Of Fourier Transform Ion Cyclotron Mass Spectra

84

Phase

Frequencyfs ff

Smooth x2

T1

T

IFT

Magnitude Modulation Phase Modulation

Phase

Frequencyfs ff

10 Tt ≤≤

T1

Tt0t2

10ttt ≤≤

Phase Modulation

Swift Excitation Event

Effective waveform

tdelay

Detection

Excitation 2

)(1

0

TTt

−=

0

t0 t1

A)

B)

X

SWIFT Waveform

Figure 5.7. A) Schematic procedure for building SWIFT waveform. B) Detailed SWIFT waveform configuration in real experiment. Note that effective waveform region is located in the middle of whole waveform and is exactly 1/2 of T1

Page 102: Broadband Phase Correction Of Fourier Transform Ion Cyclotron Mass Spectra

85

SWIFT Phasing by Phase Spectrum Method

Modified Phase Spectrum for SWIFT. Previous scientist has proved that the frequency-

domain excitation waveform from SWIFT has the same structure as frequency domain of a

frequency sweeping waveform.273 Since phase spectrum method is working well for linear

frequency-sweep excitation signal, it should also work for SWIFT. Figure 5.7, top shows the

basic steps for building SWIFT waveform. Magnitude profile with constant amplitude will be

smoothed twice by specific filter in order to avoid the Gibbs’s oscillation caused by discontinuity

points (near both edges of magnitude profile). In the real case scenario, finite time period causes

the power leakage to outside of time window. Although smoothing filterer also help to decrease

the power leakage, expanded window confining most power leakage has to be used in the real

case. Figure 5.7, bottom shows the detailed expanded time window. For convenient purpose, the

whole window length, T1, is expanded to double length of original waveform window. Also, the

original waveform window is located at the middle of large window. The middle part confines

most power inside and represents the most effective part of whole waveform. In other words, the

phase modulation only effectively works in this window area. According to all these information,

we can modify the phase spectrum of detected signal as:

4

)(2

)()(

0

2 πωωωωϕ ±−+−−

= ttSR

obsi

i

iD [5.31]

In which ωI=2πfI is the initial frequency, SR represents the sweeprate and should be calculated

by (ωF- ωI)/T, tobs is the beginning time of detection event (t1+tdelay). The SWIFT only effectively

start at t0, so the effective observation time is tobs-t0.

Performances. Modified phase spectrum method is tested by the SWIFT waveform built for

Figure 5.7. Peaks between m/z 200 to m/z 1000 (crude oil) with relative abundance of > 6 of

baseline rms noise are chosen for test purpose. Figure 5.8, top shows that the average resolving

power is consistently as much as 40% higher for absorption-mode (lower) than for magnitude-

mode spectrum (upper). Crude oil of absorption-mode FT-ICR mass spectrum (lower) from

modified phase spectrum method demonstrates not only much smoother mass error distribution

than magnitude-mode spectrum (upper), but also 40% better mass accuracy than magnitude-

mode spectrum for exact same 11500 with highest peak heights (Figure 5.8, bottom).

Page 103: Broadband Phase Correction Of Fourier Transform Ion Cyclotron Mass Spectra

86

1000900800700600500400300

m/z

1000900800700600500400300

1.0

0.0

-1.01000800200 600

Mass E

rro

r (p

pm

)

400

0.5

-0.5

1.0

0.0

-1.01000800200 600

Mass E

rro

r (p

pm

)

400

0.5

-0.5

11500 peaks

RMS error:

70ppb

11500 peaks

RMS error:

130ppb

Absorption‐Mode

Magnitude‐Mode

Absorption‐Mode

Magnitude‐Mode

m/z

Crude Oil 9.4T FT-ICR Mass Spectra (SWIFT)

m/∆m50% = 600000

m/∆m50% = 850000

Figure 5.8. Top: Crude oil 9.4T absorption-mode (lower) and magnitude-mode (lower) FT-ICR mass spectra excited by SWIFT waveform. Bottom: Mass error distribution for crude oil absorption-mode spectrum (lower) by modified phase spectrum method and magnitude-mode spectrum (upper). Note that much better mass accuracy for absorption-mode spectrum relative to magnitude-mode spectrum excited by SWIFT waveform.

Page 104: Broadband Phase Correction Of Fourier Transform Ion Cyclotron Mass Spectra

87

CHAPTER SIX

PHASE CORRECTION OF FOURIER TRANSFORM ION

CYCLOTRON RESONANCE MASS SPECTRA BY

SIMULTANEOUS EXCITATION AND DETECTION

Introduction

Because mass spectrometers typically measure the mass-to-charge ratio (m/z) of an ion, it is a

powerful analytical technique for analysis of different charged particles. Modern mass

spectrometry application mostly concentrated on different bio-analysis field, e.g., proteomics, 125,

126 lipidomics, 165, 164 metablomics157, 156 and drug discovery. 274 Among characteristics of mass

spectrometry, high resolution/resolving power (m/∆m50%) is one of most interested aspects for

scientists working in variety research fields in recent decade years. As mass-resolving power

increases, more plateaus of chemical information become accessible. For example, m/∆m50%

>200,000, there is a separation of peaks at m/z 700 but differing in elemental composition by C3

vs. SH4 (separated by ~3.4 mDa). 2, 275 Among current mass spectrometers, the FT-ICR mass

analyzer, introduced in 1974,70 has the highest mass resolving power71, 9 and best mass

measurement accuracy9. Because ICR mass resolving power increase proportional to B, The

simplest way to improve FTMS performance is to operate at higher magnetic field, B. 86, 87

However, higher magnetic field is not economic way because magnet cost increases at a high

power of magnetic field. Compared to magnitude-mode spectra, an absorption-mode spectral

peak recovered from phase correction is inherently narrower than its corresponding magnitude-

mode spectral peak up to a factor of 2 that depends on the mechanism of signal damping. 79-82

Recently, phase correction gains scientists’ interest because it needs less hardware modification

(simultaneous excitation and detection113) or only software improvement112, 114, 115 (post-

processing of data) without additional cost.

Phase Spectrum and Phase Correction

Page 105: Broadband Phase Correction Of Fourier Transform Ion Cyclotron Mass Spectra

88

For a time-domain signal f(t) with any phase φ(ω), Fourier transformation of this time-domain

signal produces mathematically real and imaginary frequency-domain spectra, Re(ω) and Im(ω).

The related mathematical equation can be written as Eq.6.1:

∫ +== −T

ti idtetfF0

)Im()Re(*)()( ωωω ω [6.1]

Considering the phase φ(ω), above equation could be expanded as Eq.6.2:

)]()sin()cos()([)]()sin()cos()([)Im()Re( ωϕϕωωϕϕωωω ADiDAi −++=+ [6.2]

)(*)]()([ ωϕωω ieiDA += [6.3]

In which )](sin[)](cos[)( ωϕωϕωϕ ie i += and )]Re(/)arctan[Im()( ωωωϕ =

We can see that Absorption- and dispersion-mode spectra, A(ω) and D(ω), may be obtained

directly as eiφ(ω)=1 (φ(ω) = 2nπ for any integer n). In most general cases (φ(ω) ≠ 2nπ), the real

and imaginary part of Fourier transformation, Re(ω) and Im(ω), are the linear combination of

A(ω) and D(ω) (Eq.6.2)

If the "phase spectrum", φ(ω), is known, absorption- and dispersion-mode spectra, A(ω)

and D(ω), could be obtained as Eq.6.4:

)()(*)]()([*)]Im()[Re( )()()( ωωωωωω ωϕωϕωϕ iDAeeiDAei iii +=+=+ −− [6.4]

or

)Im()](sin[)Re()](cos[)( ωωϕωωϕω −=A [6.5]

)Im()](cos[)Re()](sin[)( ωωϕωωϕω +=D [6.6]

Page 106: Broadband Phase Correction Of Fourier Transform Ion Cyclotron Mass Spectra

89

Eq.6.5 and Eq.6.6 are simplified expression of Eq.6.4 and could be easily used in

different algorithms. Most recently developed phase correction methods 112, 114, 115 are based on

the calculation of the "phase spectrum", φ(ω), and then recover absorption- and dispersion-mode

spectra, A(ω) and D(ω), from Eq.6.5 and Eq.6.6.

Another phase correction method based on the Fourier deconvolution theory 1 is called

simultaneous excitation and detection (SED), 113 which the excitation and detection spectra must

be temporally synchronized. For ion cyclotron radii less than about half of the trapped-ion cell

radius, ICR excitation and detection processes are both highly linear. 276 Therefore, the radius of

the ion cyclotron motion after excitation is a linear function of the excitation signal amplitude,

and the amplitude of the detected image current is a linear function of the ion cyclotron radius.

Especially, the phase spectrum of detected signal will be exact same as excitation signal because

of no time difference between excitation and detection (simultaneous excitation and detection).

Under these circumstances, the detected time domain ion signal, f(t)real, is simply the convolution

of the applied excitation waveform, e(t), and the desired ideal ion response, h(t). Because the

convolution of two time domain functions is equivalent to the product of their respective Fourier

transforms, the absorption mode spectrum that corresponds to the desired ideas response can be

recovered via complex division of the spectrum of the observed response by the spectrum of the

excitation. Unlike phase correction methods based on calculation of phase spectrum, SED

method should theoretically work for any linear FT-ICR system with no need of any phase

information.

In real experiments, this simultaneous excitation and detection (SED) is made difficult

due to capacitive coupling that exists between the excitation and detection circuits. The coupling

causes excitation-induced preamplifier saturation, e(t)add and the detection signal coupled with

the excitation-induced signal. In previous work, capacitive nulling technique has been used to

reduce excitation-induced signal from detected signal. Broadband phase correction of FT-ICR

mass spectra has been demonstrated by SED method. However, the SED method is not

practicable because robust capacitive nulling is not possible.

Here, we describe a new data processing procedure to enable broadband phase correction

of FT-ICR mass spectra by SED without any hardware modification. All signal values in that

portion of the resulting detection signal that were acquired during excitation are replaced with

zero values via computer processing. The resulting absorption-mode spectra yield improvement

Page 107: Broadband Phase Correction Of Fourier Transform Ion Cyclotron Mass Spectra

90

in resolving power as well as reduction in frequency assignment errors relative to conventional

magnitude-mode spectra. Also, the SED method can phase spectra produced from different

excitation waveforms (e.g., SWIFT with different magnitude modulations), and automatically

correct the peak height variation caused by non-uniform power distribution over the excitation

bandwidth.

Experimental Methods

Sample Preparation

Stock solutions of either a distillate fraction (500-538 ○C) from an Athabasca bitumen heavy

vacuum gas oil (HVGO) or a North American crude oil were dissolved in 50:50 (v/v)

toluene/methanol to a final concentration of 1 mg/mL. Each sample was further diluted to 0.25

mg/mL with 50:50 (v:v) toluene/methanol and 0.1% Formic acid to facilitate protonation during

electrospray ionization.

Instrumentation

Crude oil sample was analyzed with a custom-built FT-ICR mass spectrometer equipped with a

9.4 Tesla horizontal 220 mm bore diameter superconducting solenoid magnet operated at room

temperature (Oxford Corp., Oxney Mead, U.K.) and a modular ICR data station (PREDATOR)

facilitated instrument control, data acquisition and data analysis. 226, 227 Positive ions generated

at atmospheric pressure were accumulated in an external linear octopole ion trap84 for 250-1000

ms and transferred by rf-only octopoles to a 10 cm diameter, 30 cm long open cylindrical

Penning ion trap. Octopoles were operated at 2.0 MHz and 240 Vp-p amplitude. Broadband

frequency sweep (chirp) dipolar excitation (70 – 700 kHz at 50 Hz/μs sweep rate and 350 Vp-p

amplitude) or SWIFT excitation was followed by direct-mode image current detection to yield

time-domain data sets with 16 Mwords. The 25 individual transients of 5.6 s collected for the

crude oil were averaged, apodized with a full-Hanning (magnitude spectrum) or half-Hanning

(absorption spectrum 114) weight function, and zero-filled once prior to fast Fourier

transformation. Each m/z spectrum was internally calibrated with respect to an abundant

Page 108: Broadband Phase Correction Of Fourier Transform Ion Cyclotron Mass Spectra

91

homologous N1 series, and then walking calibration was applied to the spectrum dividend into 56

equal consecutive segments which contained at least two members of a homologous

alkylation series.

Fre

quency

E(t)

fs

0

0

ff

Ttotal

Frequency-Sweep Excitation Signal

Detected Signal

D(t)

Time (s)

Time (ms)

0

Ttotal

Time (s)

Ttotal

0

RDT Signal

Figure 6.1. Top: Plot of time-domain linear frequency-sweep excitation signal. Middle: Plot of detection signal in SED experiments. Bottom: Plot of excitation signal in SED experiment.

Page 109: Broadband Phase Correction Of Fourier Transform Ion Cyclotron Mass Spectra

92

Mass Analysis

Each FT-ICR mass spectrum was first calibrated with respect to a prior sample containing

Ultramark®, Met-Arg-Phe-Ala (MRFA) peptide, and caffeine (external calibration), and then was

internally calibrated with respect to an abundant homologous N1 series, and then walking

calibration was applied to the spectrum was dividend into 56 equal consecutive segments which

contained at least two members of a homologous alkylation series. Masses for singly charged

ions from m/z 200 to m/z 1000 with relative abundance of > 6 of baseline rms noise were

extracted, converted to the Kendrick mass scale228 and exported to a spreadsheet for easier

identification of homologous series. Peak assignments were performed by Kendrick mass defect

analysis, 229 as previously described.

)()()()( thtetetfaddreal

∗+=

)()()( thtetf ∗=)()()( ωωω HEF =

)(ωE

)(/)()( ωωω EFH =

)(te

)()()( thtetf ∗=

÷ =Half

Apodization

Zero-filling

FFT

1000900800700600500400300

Data Processing for SED Experiment

xA

B

Figure 6.2. Data processing steps for SED experiments. A) Zeroing of containment signal (front saturated signal induced by excitation signal). B) Complex division to produce absorption-mode spectrum. Note that exact same data processing (half apodization and zero-filling) for both zeroed detected signal and excitation signal.

Page 110: Broadband Phase Correction Of Fourier Transform Ion Cyclotron Mass Spectra

93

SWIFT Waveform Design

Fourier transform (SWIFT) excitation method introduced in 1985 76 can produce specific

excitation magnitude spectra with high mass selectivity. In the SWIFT method, a desired

excitation magnitude spectral profile and the corresponding phase function are specified. 77 They

are then subjected to inverse Fourier transform to give the time-domain excitation waveform.

Converted analog signal from the waveform is amplified and applied to excitation plate. In our

experiments, two specific magnitude spectrum profiles have been designed. First one has

uniform magnitude profile and second one has different magnitude profiles for four frequency

segments (100-250 kHz, 250-500 kHz, 500-750 kHz, 750-1000 kHz). The detailed profile is

shown in Figure 6.5, top. Optimized phase modulation 273 for each magnitude profile has been

used to build each SWIFT waveform.

SED Experiments

Because the concurrent acquisition of both time-domain transient and excitation signal needs the

both hardware modification (duplicate parallel sets of acquisition circuitry) and software

modification (trigger excitation and detection at same time), it increases the instrumental cost

and experimental complexity. Here, we collect time-domain data for the excitation waveform

and detected ion signal with identical instrument parameters and timing sequences separately.

For our experiments, spectra of the excitation waveforms were obtained by directly coupling the

excitation and detection circuits with appropriate attenuation to avoid saturation of the detection

preamplifier. The coupling was accomplished as close to the cell as possible, and included the

preamplifier, so that the excitation waveform was acquired with an excitation and detection

signal path as similar as possible to that used during ion detection. 113 This procedure helps to

ensure that the detected excitation and ion signals are both subject to the same signal path

induced phase shifts. Figure 6.1, top shows time-domain linear frequency-sweep excitation

signal used in SED experiments. Detection signal (Figure 6.1, middle) demonstrates containment

signal with really high amplitude at front end. Figure 6.1, bottom is the profile of excitation

signal. RDT stands for response during transient. Note that total time of containment signal in

detection signal is exact same as total time of excitation waveform (Ttotal).

Page 111: Broadband Phase Correction Of Fourier Transform Ion Cyclotron Mass Spectra

94

Computational Method

In our experiments, prior to phase correction via Fourier deconvolution, all saturated signal

values in that front of the resulting detection signal were replaced with zero values via computer

processing. Both detection signal after zeroing and excitation signal were subjected to half

apodization, zero-filling, Fourier transformation and complex division to produce absorption and

dispersion (A(ω) and D(ω), as shown schematically in Figure 6.2. Note that exact post-

processing steps have been applied to both signals to avoid any error associated with data

processing.

Results and Discussion

SED vs. Automated Broadband Phase Correction

In first experiment, we use the linear-sweep excitation waveform (Figure 6.1, top) and collect

time-domain data for the SED and automated broadband phase correction mehtods with identical

instrument parameters. For the present experimental FT-ICR mass spectra, elemental

compositions could be determined for all singly charged ions between 200 and 1000 Da with

peak heights greater than 6 of baseline rms noise. Figure 6.3, top shows the phasing results

from automated broadband phase correction (upper) and SED method (lower). Negative peak

with large magnitude appears in absorption-mode spectrum from automated broadband phase

correction but not in SED method because the noise peak, which can’t phased properly, could be

reduced by complex division step in SED method. For comparison of phasing result from both

methods, we chose the 5800 highest peaks in an ESI FT-ICR mass spectrum of a crude oil. The

mass errors distribution for absorption-mode spectra from both methods are plotted in Figure 6.3,

middle. SED method produces a little bit better mass accuracy than automated broadband phase

correction method. Moreover, DBE vs. carbon number images for the most abundant heteroatom

classes (N1) from electrospray ionization FT-ICR mass spectra of crude oil are shown in Figure

6.3, bottom. Note the different color distribution in SED phasing result, corresponding to

different abundance for same heteroatom class. The corresponding absorption-mode mass

spectrum from SED method get higher magnitude for low m/z peaks, so as to shift the deep color

Page 112: Broadband Phase Correction Of Fourier Transform Ion Cyclotron Mass Spectra

95

to low m/z of in the DBE vs. carbon number image.

Crude Oil ESI 9.4T FT-ICR Absorption-Mode Mass Spectra

SED Phasing

Broadband Phasing

15 30 45 60 750

10

20

30

40

15 30 45 60 750

10

20

30

40

900800700600500400300

DB

E

DB

E

Carbon Number Carbon Number

200 400 600 800 1000

Broadband Phasing

5800 peaks

140 ppb

SED Phasing

5800 peaks

110 ppb

m/z

m/z

Mass E

rro

r(p

pm

)

1.0

0.5

0

-0.5

-1.0

1.0

0.5

0

-0.5

-1.0

SED PhasingBroadband Phasing

Rela

tive A

bu

ndance

Figure 6.3. Top: Absorption-mode from automated broadband phasing (upper) and SED phasing (lower) for linear-sweep excitation. Middle: Mass error distributions for automated broadband phasing (upper) and SED phasing (lower). Bottom: Isoabundance-contoured plots of double bond equivalents (DBE = rings plus double bonds) vs. carbon number for species containing carbon, hydrogen, one nitrogen for automated broadband phasing (left) and SED phasing (right).

N1 N1

Page 113: Broadband Phase Correction Of Fourier Transform Ion Cyclotron Mass Spectra

96

Crude Oil ESI 9.4T FT-ICR MS (Normal Swift SED)

1000800600400

Absorption-Mode

m/z

1000800600400

Magnitude-Mode

926.86926.84926.82926.80926.78926.76

370000 380000450000

380000

Absorption-Mode

m/z926.86926.84926.82926.80926.78926.76

700000 710000650000

670000

600000Magnitude-Mode

1.0

0.5

0

-0.5

-1.0

Magnitude-Mode

4105 peaks

140 ppb

1.0

0.5

0

-0.5

-1.0200 400 600 800 1000

Absorption-Mode

4100 peaks

100 ppb

m/z

Mass E

rro

r (p

pm

)

Figure 6.4. Top: Magnitude-mode (upper) and absorption-mode (lower) from normal SWIFT excitation. Middle: Mass scale-expanded segments of crude oil FT-ICR magnitude-mode (upper) and absorption-mode (lower) mass spectra from normal SWIFT excitation. Bottom: Mass error distribution for magnitude (upper) and absorption (lower) electrospray ionization 9.4 T FT-ICR mass spectra for a crude oil.

Page 114: Broadband Phase Correction Of Fourier Transform Ion Cyclotron Mass Spectra

97

Absorption Spectra vs. Magnitude Spectra from Normal SWIFT Excitation

In second experiment, we built a SWIFT waveform with constant magnitude profile as the

excitation signal for SED. In this case, we also pick peaks with heights greater than 6 of

baseline rms noise for comparison purpose. Figure 6.4, top shows broadband crude oil ESI FT-

ICR magnitude-mode (upper) and absorption-mode (lower) mass spectrum. Clearly, absorption-

mode mass spectrum shows higher peak heights at low m/z area than magnitude-mode mass

spectrum due to the advantage of SED data processing - automatically correcting the peak height

difference caused by non-uniform excitation waveform (see below in detail). Compared to

magnitude-mode spectrum, absorption-mode spectrum consistently gets a factor of 1.5 better

resolving power than magnitude-mode spectrum (Figure 6.4, middle). Mass error distributions

for the broadband ESI FT-ICR magnitude-mode and absorption-mode mass spectra are

demonstrated in Figure 6.4, bottom. The reduction in rms error (averaged over all assigned

peaks) is achieved mainly by reassignment of those peaks with the highest magnitude-mode

mass errors.

Absorption Spectra from Stepped SWIFT Excitation

The biggest advantage of SWIFT excitation is that one can design different excitation waveform

by specifying the magnitude profile and phase modulation. In this experiment, we build a SWIFT

excitation waveform with four segments magnitude profile for purpose of testing. The detailed

magnitude profile (M(ω)) and stepped SWIFT excitation waveform are shown in Figure 6.5, top.

The resulting absorption-mode and magnitude-mode mass spectra are shown in Figure 6.5,

bottom. Because the radius of the ion cyclotron motion after excitation is a linear function of the

excitation signal amplitude, and the amplitude of the detected image current is a linear function

of the ion cyclotron radius, stepped distribution of peak heights in broadband crude oil

magnitude-mode spectrum (upper) is observed in Figure 6.5, bottom. Absorption-mode

spectrum (lower) from SED shows much better and smoother peak heights distribution than

magnitude-mode spectrum. Automatic peak height correction of SED is contributing to this

difference of peak heights distribution between magnitude-mode and absorption-mode spectra.

Advantages of Broadband Phase Correction by SED

Page 115: Broadband Phase Correction Of Fourier Transform Ion Cyclotron Mass Spectra

98

x

IFT

Time(msec)50403020100

Swift wave form

Crude Oil ESI 9.4T FT-ICR MS (Stepped Swift SED)

Frequency (kHz)1000800600400200

)(ωM

100

6575

85

ωϕ

Frequency (kHz)1000800600400200

Unknown

m/z

1000800600400

Magnitude-Mode

1000800600400

Absorption-Mode

Figure 6.5. Top: Schematic design of stepped SWIFT waveform. Bottom: Magnitude-mode (upper) and absorption-mode (lower) from stepped SWIFT excitation.

Page 116: Broadband Phase Correction Of Fourier Transform Ion Cyclotron Mass Spectra

99

Crude Oil ESI 9.4T FT-ICR MS (Stepped Swift SED)

SED Phasing

Broadband Phasing

1000800600400m/z

Real Spectrum Before

Phasing

1000800600400

1000800600400

Figure 6.6. Electrospray ionization crude oil 9.4 T FT-ICR mass spectra. Top: Raw real data following Fourier transform of discrete time-domain signal. Middle: Result after automatic broadband phasing. Bottom: Absorption-mode spectrum from SED phasing.

Page 117: Broadband Phase Correction Of Fourier Transform Ion Cyclotron Mass Spectra

100

)(

)(

)exp()(

)exp()(

)(

)()(

ωω

ϕωϕω

ωωω

ω

ω

M

M

M

M

E

HE ∗∗

==

)(

)(

ωω

M

M ∗

Frequency (kHz)

)exp()( ωϕω ∗M100

65 75

Rela

tive A

bu

nd

an

ce

15 30 45 60 750

10

20

30

40

15 30 45 60 750

10

20

30

40

DB

E

DB

E

Stepped SwiftNormal Swift

15 30 45 60 750

10

20

30

40

15 30 45 60 750

10

20

30

40

DB

E

DB

E

Carbon Number Carbon Number

Absorption-Mode

Magnitude-ModeMagnitude-Mode

Absorption-Mode

600500400300200

Absorption-Mode

Real Spectrum

Before Phasing

Figure 6.7. Top: Schematic automatic peak height correction in SED algorithm. Bottom: Isoabundance-contoured plots of double bond equivalents (DBE = rings plus double bonds) vs. carbon number for species containing carbon, hydrogen, one nitrogen for magnitude-mode (upper) and absorption-mode (lower).

Page 118: Broadband Phase Correction Of Fourier Transform Ion Cyclotron Mass Spectra

101

Although we don’t see much performance difference between automated broadband phase

correction and phase correction by SED for linear-sweep excitation (Figure 6.3), we do observe

that the SED phasing method is superior to automated broadband phase correction for stepped

SWIFT waveform. Figure 6.6, top is broadband crude oil FFT real mass spectrum. Automated

broadband phase correction algorithm failed to recover the absorption-mode spectrum from real

and imaginary spectrum because phase modulation in stepped SWIFT waveform is totally

different from the quadratic phase spectrum for linear-sweep excitation waveform (Figure 6.6,

middle). SED phasing algorithm is not dependent on any phase spectrum/phase modulation, it

could successfully produce the absorption-mode spectrum for stepped SWIFT waveform (Figure

6.6, bottom). Another advantage of SED phasing is automatic peak height correction in

absorption-mode spectrum. The detailed explanation is demonstrated in Figure 6.7, top.

According to the Fourier deconvolution theory, complex division of detected signal to excitation

signal will directly produce absorption-mode spectrum (the resulting real spectrum of complex

division). The Fourier transformation of detected signal and excitation signal can be represented

as M(ω)exp(φ(ω)) (polar formation of Fourier transformation). Because the phase spectrum φ(ω)

in both detected and excitation signal is same, the exponential part (exp(φ(ω))) will be canceled

during complex division. The resulting complex division (lower) is the ratio of magnitude profile

of detected signal (M*(ω)) to excitation signal (M(ω)) . Since the FT-ICR mass spectrometer is

highly linear system, magnitude profile of detected signal (M*(ω)) will be directly proportional

to excitation signal (M(ω)). The final peak height profile of absorption-mode mass spectrum

form SED is independent on different excitation waveforms. In other words, SED phasing will

automatically correct the peak height caused by non-uniform of excitation waveform amplitude.

Figure 6.7, bottom shows DBE vs. carbon number images for the most abundant heteroatom

classes (N1) from FT-ICR mass spectra of crude oil (SWIFT excited). For both normal SWIFT

(left) and stepped SWIFT (right) cases, absorption-mode spectra by SED (lower) achieve very

similar images (color coding represents peak height). However, magnitude-mode spectra

demonstrate the excitation waveform amplitude dependences.

Page 119: Broadband Phase Correction Of Fourier Transform Ion Cyclotron Mass Spectra

102

CHAPTER SEVEN

ARTIFACTS INDUCED BY SELECTIVE BLANKING OF TIME-

DOMAIN DATA IN FOURIER TRANSFORM MASS

SPECTROMETRY

Introduction

In Fourier transform mass spectrometry, periodic coherent ion motion is detected from the

oscillating current induced in opposed detection electrodes. Fourier transformation of the

digitized time-domain signal produces mathematically real and imaginary frequency-domain

spectra, Re(ω) and Im(ω), or alternatively a magnitude-mode (also known as absolute-value)

spectrum, M(ω) = [[Re(ω)]2 + [Im(ω)]2]1/2. Absorption- and dispersion-mode spectra, A(ω) and

D(ω), may in turn be obtained as linear combinations of Re(ω) and Im(ω).1 The magnitude-mode

spectral peak width at half-maximum peak height is higher by a factor of up to 2 relative to

absorption-mode spectra. 78, 80, 112, 113 With appropriate phase correction,114, 116, 115 apodization,277,

278 and zero-filling, 78,279 an absorption-mode spectrum may be generated from linear

combination of the real and imaginary spectra, to yield higher resolving power and better mass

accuracy than magnitude-mode spectrum. Apodization277, 278 can help to detect a small peak near

a large peak by reducing the amplitudes of auxiliary maxima and narrowing the peak width near

its base. Zero-filling effectively recovers into the absorption spectrum the information residing in

the non-zero-filled dispersion data obtained by Fourier transformation of the time domain

data.78,1

Mass resolution to better than 1 Da enables assignment of charge for a single charge

state,280 and is especially important for identifying post-translational modification(s) and

adduct(s) in intact proteins.281-283, 132, 284, 103 The time-domain signal for highly charged high-mass

ions displays large amplitude signal "beats" separated by extended periods of much lower signal

amplitude due to destructive interference of the signals induced by ions of many different m/z

values (and thus many frequencies).285 The highest mass resolution is achieved by isolating the

Page 120: Broadband Phase Correction Of Fourier Transform Ion Cyclotron Mass Spectra

103

isotopic distribution from ions of a single charge state, to yield a time-domain signal

characterized by multiple high-amplitude "beats", separated in time by the period of the

(approximately constant) frequency separation between ions differing in mass by ~1 Da.

Because the signal-to-noise ratio is typically low between "beats", selective blanking (zeroing) of

the data between beats eliminates much of the noise, and has been used to increase spectral

signal-to-noise ratio.286-290 However, such blanking also eliminates part of the signal, and can

thus potentially distort the resulting FT spectrum. Previous scientists287 have noticed the

extended isotopic distribution but they didn’t do any further investigation of this problem.

Here, we simulate an FT-MS time-domain signal as a sum of noiseless exponentially

damped sinusoids whose amplitudes and frequencies are chosen to model a segment of the 57+

charge state isotopic distribution for an antibody (147 kDa). Comparison of the resulting Fourier

transform spectrum with recently acquired experimental data 103 provides a testbed to expose

artifacts due to the blanking. Addition of noise to the simulated time-domain signal reveals any

effect of blanking on mass accuracy. As explained below, we find that although blanking

improves FT spectral signal-to-noise ratio, mass accuracy does not improve, and blanking

generates additional spurious peaks that artificially broaden the isotopic distribution, thereby

making it more difficult to accurately measure protein mass and to identify post-translational

modifications and/or adducts. Identification of PTM is depend on the mass difference between

modified peptide segment and original peptide segment, any inaccurate mass will probably cause

assignment problem in complex database searching.

Materials and Methods

Materials

Recombinant, humanized IgG1k therapeutic antibody (1,324 amino acids,

C6,528H10,088N1,728O2,098S44) was expressed and purified by Pfizer Inc., and provided by Jason C.

Rouse.103

Experiments

All experimental data were acquired with a custom-built 9.4 T FT-ICR mass spectrometer.291

Page 121: Broadband Phase Correction Of Fourier Transform Ion Cyclotron Mass Spectra

104

[M+57H]57+ ions from electrosprayed humanized IgG1k therapeutic antibody were quadrupole-

isolated and externally accumulated84 prior to transfer to the ICR cell.103 The ions were then

excited by broadband linear sweep (190<Vp-p<240, 720-45 kHz) at 50 Hz/µs chirp rate. 125

conditionally co-added117 individual 11.66 s digitized transients were averaged, apodized with a

full-Hanning (magnitude-mode spectrum) or half-Hanning (absorption-mode spectrum) weight

function, and zero-filled once prior to fast Fourier transformation. Each FT-ICR mass spectrum

was externally calibrated105, 106 with respect to a prior sample containing Ultramark® and Met-

Arg-Phe-Ala (MRFA) peptide.

Simulations

In FT-ICR MS, the detected time-domain ICR signal, f(t), may be modeled by the sum of

exponentially damped sinusoids: 252, 80

f(t) = ∑Aicos(ω0,it+φi)exp(-t/τi) 0<t<T [7.1]

in which i represents each ion signal, i is a characteristic damping time constant, or relaxation

time, T is the time-domain data acquisition period, φi is the initial phase, ω0,i is the natural

angular frequency, and Ai is the amplitude (which is directly proportional to the number of ions

of that cyclotron frequency). Here, we define a simulated time-domain signal, f(t):

f(t) = [A1cos(ω1t+ φ1)+…+ A57cos(ω2t+ φ57)] exp(-t/τ) +N(t) 0<t<T [7.2]

in which ions of all frequencies have the same damping time constant, τ; the amplitudes are

chosen to match the calculated natural abundance isotopic distribution for the antibody, and N(t)

is Gaussian-distributed random noise. The isotopic distribution has been truncated to include

only the 57 highest experimental nominal mass absorption-mode peak heights for the antibody

57+ charge state (see below); and the remaining parameters are chosen to match the experimental

time-domain signal (T = 11.66 s, 8 MWord data) and absorption-mode spectrum (ICR

frequencies between 55600 and 55630 Hz, bandwidth = 359712.2 Hz).

Estimation of Relaxation Time τ

Page 122: Broadband Phase Correction Of Fourier Transform Ion Cyclotron Mass Spectra

105

Time (s)

f(t) = [A1cos(ω1t+φ1) + … + A57cos(ω57t+ φ57)] exp(-t/ּז) + N(t)

Imaginary Spectrum, Im(ω)Real Spectrum, Re(ω)

Magnitude Spectrum, M(ω)

Apodization

Zero-Filling

Fourier Transform

Phase Correction

Dispersion Spectrum, D(ω)Absorption Spectrum, A(ω)

M(ω) = ((Re(ω))2 + (Im(ω))2)1/2 = ((A(ω))2 + (D(ω))2)1/2

Figure 7.1. Generation of frequency-domain FT-ICR mass spectra from a simulated isotopic distribution with 57 frequency components. To a good approximation the ~1 Da separation between adjacent isotopic nominal masses in a

given charge state isotopic distribution corresponds to a single frequency spacing in the

frequency-domain FT-ICR spectrum for that distribution. Thus, we can derive the relaxation time

in Eq. 7.2 from the amplitudes of the 2nd and 4th time-domain "beat" peak heights, H2 and H4 as

follows:

H2=Aexp(-t2/τ) [7.3a]

H4=Aexp(-t4/ τ) [7.3b]

Eqs. 3a and 3b may be solved to yield τ:

=(t4-t2)/ln(H2/H4) [7.4]

Page 123: Broadband Phase Correction Of Fourier Transform Ion Cyclotron Mass Spectra

106

The simulated time-domain signal (with or without noise) was subjected to selective

blanking (or not) of the data between "beats", apodized, zero-filled and Fourier transformed to

real and imaginary (Re(ω) and Im(ω), absorption and dispersion (A(ω) and D(ω), and magnitude

(M(ω)) frequency spectra, as shown schematically in Figure 7.1. The frequency and maximum

height of each simulated absorption-mode peak are obtained by parabolic interpolation and the

peak frequency error is reported as the difference between the experimental and simulated values

of frequency and interpolated frequency. The procedure was programmed in C and C++

operating under labWindows/CVI.

Results and Discussion

Spectral Profile without Blanking

An experimental time-domain ICR signal from the isolated 57+ charge state isotopic distribution

from an electrosprayed antibody (Figure 7.2, (a)) exhibits characteristic "beats" corresponding to

the period of the frequency difference between ions differing by ~1 Da in mass. A simulated

noiseless time-domain signal modeled as a sum of 57 exponentially damped sinusoids (Figure

7.2, (b)) was apodized, zero-filled, and Fourier transformed to produce the absorption-mode and

magnitude-mode frequency spectra shown in Figure 7.2, (c), (d). Note that the simulated isotopic

distribution was intentionally truncated to include only the 57 highest-magnitude peaks, so as to

better expose the artifacts introduced by time-domain signal blanking (see below).

Spectral Profile after Blanking

Figure 7.3 (top) shows a simulated noiseless time-domain signal, in which data between the

"beats" have been replaced by zeroes. The resulting absorption-mode and magnitude-mode

frequency spectra (Figure 7.3, middle & bottom) now contain numerous artifact peaks at

frequencies below and above the frequency range for the simulated isotopic distribution. We

have observed more or less artifact peaks as we zeroed different ranges between the “beats”. The

explanation for appearance of artifacts is that zeroing datapoint in time domain will confuse

Fourier Transformation to find frequency components. Fourier transformation will find

Page 124: Broadband Phase Correction Of Fourier Transform Ion Cyclotron Mass Spectra

107

additional frequency components due to trying match all information from zeroed time domain

data.

0 1 2 3 4 5 6 7 8 9 10 11

Time (s)

Simulated

Absorption-Mode

Spectrum

0 1 2 3 4 5 6 7 8 9 10 11

Time (s)

a)

b)

Simulated

Magnitude-Mode

Spectrum

c)

d)

[M+57H]+

ExperimentalIgG1k Antibody

[M+57H]+

Simulated

Frequency (kHz)

55,596 55,604 55,612 55,620 55,628 55,636

Frequency (kHz)

55,596 55,604 55,612 55,620 55,628 55,636

Figure 7.2. a) Experimental time-domain ICR signal from the isolated 57+ charge state isotopic distribution from electrosprayed humanized IgG1k therapeutic antibody. b) Simulated noiseless time-domain transient with 57 frequency components. c) Absorption-mode frequency spectrum for b). d) Magnitude-mode frequency spectrum for b).

Page 125: Broadband Phase Correction Of Fourier Transform Ion Cyclotron Mass Spectra

108

0 1 2 3 4 5 6 7 8 9 10 11Time (s)

Absorption-Mode Spectrum

With Blanking

Magnitude-Mode Spectrum

With Blanking

[M+57H]+

Simulated (Noiseless)

Frequency (kHz)55,596 55,604 55,612 55,620 55,628 55,636

Frequency (kHz)55,596 55,604 55,612 55,620 55,628 55,636

Figure 7.3. Top: Simulated noiseless time-domain signal after blanking of the data between the "beats". Middle: Absorption-mode frequency spectrum. Bottom: magnitude-mode frequency spectrum. Effect of Noise

We added Gaussian-distributed white noise to the simulated time-domain signal (Figure 7.4,

top) Comparison of the resulting absorption-mode spectra without (Figure 7.4, middle) and with

(Figure 7.4, bottom) shows that the additional artifact peaks introduced by time-domain selective

blanking are still evident. Finally, although the average peak height-to-noise ratio for the 10

highest spectral peaks increases by two-fold after selective time-domain blanking (Figure 7.5,

top), there is no reduction in frequency and thus m/z error (Figure 7.5, bottom).

Page 126: Broadband Phase Correction Of Fourier Transform Ion Cyclotron Mass Spectra

109

0 1 2 3 4 5 6 7 8 9 10 11Time (s)

Absorption-Mode Spectrum

With Blanking

Absorption-Mode Spectrum

W/O Blanking

[M+57H]+

Simulated (Noise Added)

Frequency (kHz)55,596 55,604 55,612 55,620 55,628 55,636

Frequency (kHz)55,596 55,604 55,612 55,620 55,628 55,636

Figure 7.4. Top: Simulated time-domain ICR signal with noise. Middle: Absorption-mode frequency spectrum from the above time-domain signal. Bottom: Absorption-mode frequency spectrum after blanking was performed on the above time-domain signal.

Conclusion

Although selective blanking of the time-domain data between "beats" does increase FT-ICR

spectral signal-to-noise ratio, blanking does not improve frequency (and thus mass) accuracy,

and introduces numerous additional artifact peaks at each end of the isotopic distribution. The

artifact peaks distort and broaden the apparent isotopic distribution, thereby complicating

determination of molecular monoisotopic and/or average mass, and making it harder to resolve

and identify chemical modifications and/or adducts. We conclude that zeroing of the data

between time-domain "beats" offers no advantages, and introduces artifact peaks that actually

reduces the available mass spectral information.

Page 127: Broadband Phase Correction Of Fourier Transform Ion Cyclotron Mass Spectra

110

P

eak H

eig

ht

to N

ois

e R

ati

o

55,614 55,615 55,616 55,617 55,618

Frequency (kHz)

Fre

qu

en

cy E

rro

r (p

pb

)

400

200

0

-200

-400

W/O Blanking

With Blanking

80

60

40

20

055,614 55,615 55,616 55,617 55,618

Frequency (kHz)

W/O Blanking

With Blanking

Figure 7.5. Top: FT-ICR average peak height-to-noise ratio (S/N) for each of the 10 highest absorption-mode peaks (from the time-domain data of Figure 7.4 (top)) with (blue) and without (red) blanking. Bottom: Ion cyclotron frequency errors for the 10 highest absorption-mode peaks with (blue) and without blanking (red).

Page 128: Broadband Phase Correction Of Fourier Transform Ion Cyclotron Mass Spectra

111

REFERENCES

(1) Marshall, A. G.; Verdun, F. R.Fourier Transforms in NMR, Optical, and Mass

Spectrometry: A User's Handbook; Elsevier: Amsterdam, 1990;460 pp. (2) Marshall, A. G.; Hendrickson, C. L."High-Resolution Mass Spectrometers." Ann. Rev.

Anal. Chem. 2008, 1, 579-599. (3) Giancaspro, C.; Comisarow, M. B."Exact Interpolation of Fourier Transform Spectra." Applied Spectroscopy 1983, 37, 153-166. (4) Verdun, F. R.; Giancaspro, C.; Marshall, A. G."Effects of Noise, Time-Domain Damping, Zero-Filling and the FFT Algorithm on the "Exact" Interpolation of Fast Fourier Transform Spectra." Appl. Spectrosc. 1988, 42, 715-721. (5) Keefe, C. D.; Comisarow, M. B."Exact Interplation of Apodized, Magnitude-Mode Fourier-Transform Spectra." Applied Spectroscopy 1989, 43, 605-607. (6) Chen, L.; Cottrell, C. E.; Marshall, A. G."Effect of Signal-to-Noise Ratio and Number of Data Points upon Precision in Measurement of Peak Amplitude, Position, and Width in Fourier Transform Spectrometry." Chemometrics and Intelligent Laboratory Systems 1986, 1, 51-58. (7) Liang, Z.; Marshall, A. G."Time-Domain (Interferogram) and Frequency-Domain (Absorption-Mode and Magnitude-Mode) Noise and Precision in Fourier Transform Spectrometry." Applied Spectroscopy 1990, 44, 766-775. (8) Rodgers, R. P.; White, F. M.; Hendrickson, C. L.; Marshall, A. G.; Andersen, K. V."Resolution, elemental composition, and simultaneous monitoring by Fourier transfrom ion cyclotron resonance mass spectrometry of organosulfur species before and after diesel fuel processing." Analytical Chemistry 1998, 70, 4734-4750. (9) Savory, J. J.; Kaiser, N. k.; Mckenna, A. M.; Xian, F.; Blakney, G. T.; Rodger, R. P.; Hendrickson, C. L.; Marshall, A. G."Parts-Per-Billion Fourier Transform Ion Cyclotron Resonance Mass Measurement Accuracy with a "Walking" Calibration Equation." Analytical

Chemistry 2011, 83. (10) Mirsaleh-Kohan, N.; Robertson, W. D.; Compton, R. N."Electron Ionization Time-of-Flight Mass Spectromety: Historical Review and Current Applications." Mass Spectrometry

Reviews 2008, 27, 237-285. (11) Cameron, A. E.; Eggers, D. F."An Ion "Velocitron"." Rev. Sci. Instrum 1948, 19, 605-607.

Page 129: Broadband Phase Correction Of Fourier Transform Ion Cyclotron Mass Spectra

112

(12) Wiley, W. C.; Mclaren, I. H."Time-of-Flight Mass Spectrometer with Improved Resolution." Reviews of Scientific Instruments 1955, 26, 1150-1157. (13) Verentchlkov, A. N.; Ens, W.; Standlng, K. G."Reflecting Time-of-Flight Mass Spectrometer with an Electrospray Ion Source and Orthogonal Extraction." Analytical Chemistry 1994, 66, 126-133. (14) Kelly, R. T.; Tolmachev, A. V.; Page, J. S.; Tang, K.; Smith, R. D."The Ion Funnel: Theory, Implementation, and Applications." Mass Spectrometry Reviews 2010, 29, 294-312. (15) Ibrahim, Y. M.; Belov, M. E.; Liyu, A. V.; Smith, R. D."Automated Gain Control Ion Funnel Trap for Orthogonal Time-of-Flight Mass Spectrometry." Analytical Chemistry 2008, 80, 5367-5376. (16) Pringle, S. D.; Giles, K.; Wildgoose, J. L.; Williams, J. P.; Slade, S. E.; Thalassinos, K.; Bateman, R. H.; Bowers, M. T.; Scrivens, J. H."An Investigation of the Mobility Separation of Some Petide and Protein Ions Using a New Hybrid Quadrupole/Travelling Wave IMS/oa-TOF Instrument." International Journal of Mass Spectrometry 2007, 261, 1-12. (17) Satoh, T.; Sato, T.; Tamura, J."Development of High-Performance MALDI-TOF Mass Spectrometer Utilizing a Spiral Ion Trajectory." Journal of the American Society for Mass

Spectrometry 2007, 18, 1318-1323. (18) Toyoda, M."Development of Multi-Turn Time-of-Flight Mass Spectrometers and Their Applications." European Journal of Mass Spectrometry 2010, 16, 397-406. (19) Verentchikov, A. N.; Yavor, M. I. Quasi-Planar Multi-Reflecting Time-of-Flight Mass Spectrometer, 2011,United States Patent, US 2011/01086729 A1

(20) Mamyrin, B. A."Time-of-Flight Mass Spectrometry (Concepts, Achievements, and Prospects)." Internation Journal of Mass Spectrometry 2000, 206, 251-266. (21) Toyoda, M.; Okumura, D.; Ishihara, M.; Katakuse, I."Multi-Turn Time-of-Flight Mass Spectrometers with Electrostatic Sectors." Journal of Mass Spectrometry 2003, 38, 1125-1142. (22) Hazama, H.; Aoki, J.; Nagao, H.; Suzuki, R.; Tashima, T.; Fujii, K.-i.; Masuda, K.; Awazu, K.; Toyoda, M.; Naito, Y."Construction of a Novel Stigmatic MALDI Imaging Mass Spectrometer." Applied Surface Science 2008, 255, 1257-1263. (23) Shimma, S.; Nagao, H.; Aoki, J.; Takahashi, K.; Miki, S.; Toyoda, M."Miniaturized High-Resolution Time-of-flight Mass Spectrometer MULTUM-S II with an Infinite Flight Path." Analytical Chemistry 2010, 82, 8456-8463. (24) Nishiguchi, M.; Ueno, Y.; Toyoda, M.; Setou, M."Design of a New Multi-Turn Ion Optical System'IRIS' for a Time-of-Flight Mass Spectrometry." Journal of Mass Spectrometry 2008, 44, 594-604.

Page 130: Broadband Phase Correction Of Fourier Transform Ion Cyclotron Mass Spectra

113

(25) Shchepunov, V.; Berdnikov, A.; Lzumi, H.; Giles, R.; Gall, N., proceedings of the 58th

ASMS Conference on Mass Spectrometry & Allied Topics, Salt Lake City, Utah, May 23-27 2010. (26) Ioanoviciu, D."The Spiral Main Path Electric Deflector as a Time-of-Flight Mass Analyzer." Internation Journal of Mass Spectrometry 2010, 290, 145-147. (27) Park, J.; Qin, H.; Scalf, M.; Hilger, R. T.; Westphall, M. S.; Smith, L. M.; Blick, R. H."A Mechanical Nanomembrane Detector for Time-of-Flight Mass Spectrometry." Nano Letters 2011, 11, 3681-3684. (28) Chiu, H.-Y.; Hung, P.; Postma, H. W. C.; Bockrath, M."Atomic-Scale Mass Sensing Using Carbon Nanotube Resonators." Nano Letters 2008, 8, 4342-4346. (29) Lassagne, B.; Garcia-Sanchez, D.; Aguasca, A.; Bachtold, A."Ultrasensitive Mass Sensing with a Nanotube Electromechanical Resonator." Nano Letters 2008, 8, 3735-3738. (30) Frank, M.; Labov, S. E.; Westmacott, G.; Benner, W. H."Energy Sensitive Cryogenic Detectors for High Mass Biomolecule Mass Spectrometry." Mass Spectrometry Reviews 1999, 18, 155-186. (31) Wenzel, R. J.; Matter, U.; Schultheis, L.; Zenobi, R."Analysis of Megadalton Ions Using Cryodetection MALDI Time-of-Flight Mass Spectrometry." Anal. Chem. 2005, 77, 4329-4337. (32) Cotter, R. J.; Gardner, B. D.; IItchenko, S.; D., E. R."Tandem Time-of-Flight Mass Spectrometry with a Curved Field Reflectron." Analytical Chemistry 2004, 76, 1976-1981. (33) Iltchenko, S.; Cotter, R. J."Collision Energetics in a Tandem Time-of-Flight (TOF/TOF) Mass Spectrometer with a Curved-Field Reflectron." Internation Journal of Mass Spectrometry 2007, 265, 372-381. (34) Pittenauer, E.; Allmaier, G."The Renaissance of High-Energy CID for Structural Elucidation of Complex Lipids: MALDI-TOF/RTOF-MS of Alkali Cationized Triacylglycerols." Journal of the American Society for Mass Spectrometry 2009, 20, 1037-1047. (35) Pittenauer, E.; Allmaier, G."High-Energy Collision Induced Dissociation of Biomolecules: MALDI-TOF/RTOF Mass Spectrometry in Comparison to Tamdem Sector Mass Spectrometry." Combinatorial Chemistry & High Throughput Screening 2009, 12, 137-155. (36) Toyoda, M.; Giannakopulos, A.; Colburn, A.; Derrick, P."High-Energy Collision Induced Dissociation Fragmentation Pathways of Peptides, Probed Using a Multiturn Tandem Time-of-Flight Mass Spectrometer "MULTUM-TOF/TOF"." Review of Scientific Instruments 2007, 78, 074101-074109.

Page 131: Broadband Phase Correction Of Fourier Transform Ion Cyclotron Mass Spectra

114

(37) Shimma, S.; Nagao, H.; Giannakopulos, A.; Hayakawa, S.; Awazu, K.; Toyoda, M."High-Energy Collision-Induced Dissociation of Phosphopeptides Using a Multi-Turn Tandem Time-of-Flight Mass Spectrometer." Journal of Mass Spectrometry 2008, 43, 535-537. (38) Satoh, T.; Sato, T.; Kubo, A.; Tamura, J."Tandem Time-of-Flight Mass Spectrometer with High Precursor Ion Selectivity Employing Spiral Ion Traiectory and Improved Offset Parabolic Reflectron." Journal of the American Society for Mass Spectrometry 2011, 22, 797-803. (39) Kolmonen, M.; Leinonen, A.; Pelander, A.; Ojanpera, I."A General Screening Method for Doping Agents in Human Urine by Solid Phase Extraction and Liquid Chromatography/Time-of-Flight Mass Spectrometry." Analytica Chimica Acta 2007, 585, 94-102. (40) Kolmonen, M.; Leinonen, A.; Pelander, A.; Ojanpera, I."Generic Sample preparation and dual polarity liquid chromatography-Time-of-Flight Mass Spectrometry for High-Throughput Screening in Doping Analysis." Drug Testing and Analysis 2009, 1, 250-266. (41) Lee, H. K.; Ho, C. S.; Iu, Y. P. H.; Lai, P. S. J.; Shek, C. C.; Lo, Y.-C.; Klinke, H. B.; Wood, M."Development of a broad Toxicological Screening Technique for Urine Using Ultra-Performance Liquid Chromatography and Time-of-Flight Mass Spectrometry." Analytica

Chimica Acta 2009, 649, 80-90. (42) Williamson, L. N.; Zhang, G.; Terry Jr, A. V.; Bartlett, M. G."Comparison of Time-of-Flight Mass Spectrometry to Triple Quadrupole Tandem Mass Spectrometry for Quantitative Bioanalysis: Application to Antipsychotics." Journal of Liquid Chromatography & Related

Technologies 2008, 31, 2737-2751. (43) Badoud, F.; Grate, E.; Perrenoud, L.; Avois, L.; Saugy, M.; Rudaz, S.; Venthey, J.-L."Fast Analysis of Doping Agents in Urine by Ultra-High-Pressure Liquid Chromatography-Quadrupole Time-of-Flight Mass Spectrometry I: Screening Analysis." Journal of

Chromatography A. 2009, 1216, 4423-4433. (44) van der Heeft, E.; Bolck, Y. J. C.; Beurner, B.; Nijrolder, A. W. J. M.; Stolker, A. A. M.; Nielen, M. W. F."Full-Scan Accurate Mass Selectivity of Ultra-Performance Liquid Chromatography Combined with Time-of-Flight and Orbitrap Mass Spectrometry in Hormone and Veterinary Drug Residue Analysis." Journal of the American Society for Mass Spectrometry 2009, 20, 451-463. (45) Chughtai, K.; Heeren, R. M. A."Mass Spectrometric Imaging for Biomedical Tissure Analysis." Chemical Reviews 2010, 110, 3237-3277. (46) Tucker, K. R.; Serebryannyy, L. A.; Zimmerman, T. A.; Rubakhin, S. S.; Sweedler, J. V."The Modified-Based Stretched Sample Method: Development and Application to MALDI-MS Imaging of Protein Localization in the Spinal Cord." Chemical Science 2011, 2, 785-795.

Page 132: Broadband Phase Correction Of Fourier Transform Ion Cyclotron Mass Spectra

115

(47) Sanders, M. E.; Dias, E. C.; Xu, B. J.; Mobley, J. A.; Billheimer, D.; Roder, H.; Grigorieva, J.; Dowsett, M.; Arteaga, C. L.; Caprioli, R. M."Differentiating Proteomic Biomarkers in Breast Cancer by Laser Capture Microdissection and MALDI MS." Journal of

Proteome Research 2008, 7, 1500-1506. (48) Kang, S.; Shim, H. S.; Lee, J. S.; Kim, D. S.; Kim, H. Y.; Hong, S. H.; Kim, P. S.; Yoon, J. H.; Cho, N. H."Molecular Proteomics Imaging of Tumor Interfaces by Mass Spectrometry." Journal of Proteome Research 2010, 9, 1157-1164. (49) Burnum, K. E.; Tranguch, S.; Mi, D.; Daikoku, T.; Dey, S. K.; Caprioli, R. M."Imaging Mass Spectrometry Reveals Unique Protein Profiles During Embryo Implantation." Endocrinology 2008, 147, 3274-3278. (50) Koizumi, S.; Yamamoto, S.; Hayasaka, T.; Konishi, Y.; Yamaguchi-Okada, M.; Goto-Inoue, N.; Sugiura, Y.; Setou, M.; Namba, H."Imaging Mass Spectrometry Revealed the Production of Lyso-Phosphatidylcholine in the Injured Ischemica Rat Brain." Neuroscience 2010, 168, 219-225. (51) Dill, A. L.; Ifa, D. R.; Manicke, N. E.; Costa, A. B.; Ramos-Vara, J. A.; Knapp, D. W.; Cooks, R. G."Lipid Profiles of Canine Invasive Transitional Cell Carcinoma of the Urinary Bladder and Adjacent Normal Tissue by Desorption Electrospray Ionization Imaging Mass Spectrometry." Analytical Chemistry 2009, 81, 8758-8764. (52) Girod, M.; Shi, Y.; Cheng, J.-X.; Cooks, R. G."Desorption Electrospray Ionization Imaging Mass Spectrometry of Lipids in Rat Spinal Cord." Journal of the American Society for

Mass Spectrometry 2010, 21, 1177-1189. (53) Kimmel, J. R.; Farmer, D. K.; Cubison, M. J.; Sueper, D.; Tanner, C.; Nemitz, E.; Worsnop, D. R.; Gonin, M.; Jimenez, J. L."Real-Time Aerosol Mass Spectrometry with Millisecond Resolution." Internation Journal of Mass Spectrometry 2011, 303, 15-26. (54) Shellie, R.; Marriott, P. J."Opportunities for Ultra-High Resolution Analysis of Essential Oils Using Comprehensive Two-Dimensioan Gas Chromatography: a Review." Flavour and

Fragrance Journal 2003, 18, 179-191. (55) Rowland, S. J.; Scarlett, A. G.; Jones, D.; West, C. E.; Frank, R. A."Diamonds in the Rough: Identification of Individual Naphthenic Acids in Oil Sands Process Water." Environmental Science & Technology 2011, 45, 3154-3159. (56) Rowland, S. J.; West, C. E.; Scarlett, A. G.; Jones, D.; Frank, R. A."Identification of Individual Tetra- and Pentacyclic Naphthenic Acids in Oil Sand Process Water by Comprehensive Two-Dimensional Gas Chromatography/Mass Spectrometry." Rapid

Communications In Mass Spectrometry 2011, 25, 1198-1204.

Page 133: Broadband Phase Correction Of Fourier Transform Ion Cyclotron Mass Spectra

116

(57) Li, X.; Xu, Z.; Lu, X.; Ying, B.; Kong, H.; Yu, Y.; Xu, G."Comprehensive Two-Dimensional Gas Chromatography/Time-of-Flight Mass Spectrometry for Metabonomics: Biomarker Discovery for Diabetes Mellitus." Analytica Chimica Acta 2009, 633, 257-262. (58) Wojtowicz, P.; Zrostlikova, J.; Kovalczuk, T.; Schurek, J.; Adam, T."Evaluation of Comprehensive Two-Dimensional Gas Chromatography Coupled to Time-of-Flight Mass Spectrometry for the Diagnosis of Inherited Metabolic Disorders Using an Automated Data Processing Strategy." Journal of Chromatography A. 2010, 1217, 8054-8061. (59) Yvette Gardner, J.; Brillhart, D. E.; Benjamin, M. M.; Dixon, L. G.; Mitchell, L. M.; Dimandja, J.-M. D."The Use of GCxGC/TOF MS with Multivariate Analysis for the Characterization of Foodborne Pathogen Bacteria Profiles." Journal of Separation Science 2011, 34, 176-185. (60) Ozel, M. Z.; Ward, M. W.; Hamilton, J. F.; Lewis, A. C.; Raventos-Duran, T.; Harrison, R. M."Analysis of Organic Nitrogen Compounds in Urban Aerosol Samples Using GCxGC-TOF/MS." Aerosol Science and Technology 2009, 44, 109-116. (61) Goguet, A.; Hardacre, C.; Maguire, N.; Morgan, K.; Shekhtman, S. O.; Thompson, S. P."Time of Flight Mass Spectrometry for Quantitative Data Analysis in Fast Transient Studies Using a Temporal Analysis of Products (TAP) Reactor." Analyst 2011, 136, 155-163. (62) Li, X.; Casiano-Maldonado, M.; Zhang, D.; Wesdemiotis, C."Top-Down Multidimensional Mass Spectrometry Methods for Synthetic Polymer Analysis." Macromolecules 2011, 44, 4555-4564. (63) Polce, M. J.; Ocampo, M.; Quirk, R. P.; Wesdemiotis, C."Tandem Mass Spectrometry Characteristics of Silver-Cationized Polystyrenes: Backbone Degradation via Free Radical Chemistry." Analytical Chemistry 2008, 80, 347-354. (64) Szyszke, R.; Hanton, S. D.; Henning, D.; Owens, K. G."Development of a Combined Standard Additions/Internal Standards Method to Quantify Residual PEG in Ethoxylated Surfactants by MALDI TOFMS." Journal of the American Society for Mass Spectrometry 2011, 22, 633-640. (65) Nasioudis, A.; van Velde, J. W.; Heeren, R. M. A.; van den Brink, O. F."Electrospray Ionization Mass Spectrometry of the Non-Covalent Complexes of Ammonium Ions with High Molar Mass Polyethers." Internation Journal of Mass Spectrometry 2011, 303, 63-68. (66) Fouquet, T.; Humbel, S.; Charles, L."Tandem Mass Spectrometry of Trimethylsilyl-Terminated Poly (Dimethylsiloxane) Ammonium Adducts Generated by Electrospray Ionization." Journal of the American Society for Mass Spectrometry 2011, 22, 649-658. (67) Mahoney, C. M.; Fahey, A. J.; Steffens, K. L.; Benner, B. A.; Lareau, R. T."Characterization of Composition C4 Explosives Using Time-of-Flight Secondary Ion Mass

Page 134: Broadband Phase Correction Of Fourier Transform Ion Cyclotron Mass Spectra

117

Spectrometry and X-ray Photoelectron Spectroscopy." Analytical Chemistry 2010, 82, 7237-7248. (68) Zhou, J.-L.; Qi, L.-W.; Li, P."Herbal Medicine Analysis by Liquid Chromatography/Time-of-Flight Mass Spectrometry." Journal of Chromatography A. 2009, 1216, 7582-7594. (69) Hernandez, F.; Portoles, T.; Pitarch, E.; Lopez, F. J."Searching for Anthropogenic Contaminants in Human Breast Adipose Tissues Using Gas Chromatography Time-of-Flight Mass Spectrometry." Journal of Mass Spectrometry 2008, 44, 1-11. (70) Comisarow, M. B.; Marshall, A. G."Fourier Transform Ion Cyclotron Resonance Spectroscopy." Chem. Phys. Lett. 1974, 25, 282-283. (71) Shi, S. D.-H.; Hendrickson, C. L.; Marshall, A. G."Counting Individual Sulfur Atoms in a Protein by Ultrahigh-Resolution Fourier Transform Ion Cyclotron Resonance Mass Spectrometry: Experimental Resolution of Isotopic Fine Structure in Proteins." Proc. Natl. Acad.

Sci. U.S.A. 1998, 95, 11532-11537. (72) Makarov, A."Electrostatic Axially Harmonic Orbital Trapping: A High-Performance Technique of Mass Analysis." Analytical Chemistry 2000, 72, 1156-1162. (73) Hu, Q.; Noll, R. J.; Li, H.; Makarov, A.; Hardman, M.; Cooks, R. G."The Orbitrap: a New Mass Spectrometer." Journal of Mass Spectrometry 2005, 40, 430-443. (74) Marshall, A. G.; Hendrickson, C. L.; Jackson, G. S."Fourier Transform Ion Cyclotron Resonance Mass Spectrometry: A Primer." Mass Spectrom. Rev. 1998, 17, 1-35. (75) Marshall, A. G.; Wang, T. C. L.; Ricca, T. L."Tailored excitation for Fourier transform ion cyclotron mass spectrometry." Journal of the American Chemical Society 1985, 107, 7893-7897. (76) Chen, L.; Wang, T.-C. L.; Ricca, T. L.; Marshall, A. G."Phase-Modulated Stored Waveform Inverse Fourier Transform Excitation for Trapped Ion Mass Spectrometry." Analytical Chemistry 1987, 59, 449-454. (77) Guan, S.; Marshall, A. G."Stored Waveform Inverse Fourier Transform (SWIFT) Ion Excitation in Trapped-Ion Mass Spectrometry: Theory and Application." Internation Journal of

Mass Spectrometry and Ion Processes 1996, 157/158, 5-37. (78) Bartholdi, E.; Ernst, R. R."Fourier spectroscopy and the causality principle." J. Magn.

Reson. 1973, 11, 9-19. (79) Comisarow, M. B.; Marshall, A. G."Theory of Fourier Transform Ion Cyclotron Resonance Mass Spectroscopy. I. Fundamental Equations and Low-Pressure Line Shape." J.

Chem. Phys. 1976, 64, 110-119.

Page 135: Broadband Phase Correction Of Fourier Transform Ion Cyclotron Mass Spectra

118

(80) Guan, S.; Li, G.-Z.; Marshall, A. G."Effect of Ion-Neutral Collision Mechanism on Trapped-Ion Equation of Motion: A New Mass Spectral Line Shape for High-Mass Trapped Ions." Int. J. Mass Spectrom. Ion Processes 1997, 167/168, 185-194. (81) Mitchell, D. W."Realistic Simulation of the Ion Cyclotron Resonance Mass Spectrometer Using a Distributed Three-Dimensional Particle-In-Cell Code." Journal of the American Society

for Mass Spectrometry 1999, 10, 136-152. (82) Eugene N. Nikolaev; Ron M. A. Heeren; Alexander M. Popov; Alexander V. Pozdneev; Chingin., K. S."Realistic modeling of Ion Cloud Motion In a Fourier Transform Ion Cyclotron Resonance Cell by Use of a Particle-In-Cell Approach." Rapid Communications In Mass

Spectrometry 2007, 21, 3527-3546. (83) Aarstol, M.; Comisarow, M."Apodization of FT-ICR Spectra." International Journal of

Mass Spectrometry and Ion Processes 1987, 76, 287-297. (84) Senko, M. W.; Hendrickson, C. L.; Emmett, M. R.; Shi, S. D.-H.; Marshall, A. G."External Accumulation of Ions for Enhanced Electrospray Ionization Fourier Transform Ion Cyclotron Resonance Mass Spectrometry." J. Am. Soc. Mass Spectrom. 1997, 8, 970-976. (85) Perry, R. H.; Cooks, R. G.; Noll, R. J."Orbitrap Mass Spectrometrty: Instrumentation, Ion Motion and Applications." Mass Spectrometry Reviews 2008, 27, 661-699. (86) Marshall, A. G.; Guan, S."Advantage of High Magnetic Field for Fourier Transform Ion Cyclotron Resonance Mass Spectrometry." Rapid Communications In Mass Spectrometry 1996, 10, 1819-1823. (87) Schaub, T. M.; Hendrickson, C. L.; Horning, S.; Quinn, J. P.; Senko, M. W.; Marshall, A. G."High Performance Mass Spectrometry: Fourier Transform Ion Cyclotron Resonance at 14.5 Tesla." Anal. Chem. 2008, 80, 3985-3990. (88) Makarov, A.; Denisov, E.; Lange, O."Performance Evaluation of a High-field Orbitrap Mass Analyzer." Journal of the American Society for Mass Spectrometry 2009, 20, 1391-1396. (89) Tolmachev, A. V.; Robinson, E. W.; Wu, S.; Kang, H.; Lourette, N. M.; Pasa-Tolic, L.; Smith, R. D."Trapped-Ion Cell with Improved DC Potential Harmonicity for FT-ICR MS." Journal of the American Society for Mass Spectrometry 2008, 19, 586-597. (90) Boldin, I. A.; Nikolaev, E. N."Fourier Transform Ion Cyclotron Resonance Cell with Dynamic Harmonization of the Electric Field in the Whold Volume by Shaping of the Excitation and Detection Electrode Assembly." Rapid Commun. Mass Spectrom. 2010, 25, 122-126. (91) Gabrielse, G.; Mackintosh, F."Cylindrical Penning Traps with Orthogonalized Anharmonicity Compensation." internation Journal of Mass Spectrometry and Ion Processes 1984, 57, 1-17.

Page 136: Broadband Phase Correction Of Fourier Transform Ion Cyclotron Mass Spectra

119

(92) Gabrielse, G.; Haarsma, L.; Rolston, S. L."Open-Endcap Penning Traps for High-Precision Experiments." Internation Journal of Mass Spectrometry 1989, 88, 319-332. (93) Naito, Y.; Fujiwara, M.; Inoue, M."Improvement of the Electric Field in the Cylindrical Trapped Ion Cell." International Journal of Mass Spectrometry and Ion Processes 1992, 120, 179-192. (94) Guan, S.; Marshall, A. G."Ion Traps for Fourier Transform Ion Cyclotron Resonance Mass Spectrometry:Principles and Design of Geometric and Electric Configuration." Internation

Journal of Mass Spectrometry and Ion Processes 1995, 146, 261-296. (95) Vartanian, V. H.; Hadjarab, F.; Laude, D. A."Open Cell Analog of the Screened Trapped-Ion Cell Using Compensation Electrodes for Fourier Transform Ion Cyclotron Resonance Mass Spectrometry." Internation Journal of Mass Spectrometry 1995, 151, 175-187. (96) Jackson, G. S.; White, F. M.; Guan, S.; Marshall, A. G."Matrix-Shimmed Ion Cyclotron Resonance Ion Trap Simultaneously Optimized for Excitation, Detection, Quadrupolar Axialization, and Trapping." Journal of the American Society for Mass Spectrometry 1999, 10, 759-769. (97) Bruce, J. E.; Anderson, G. A.; Lin, C. Y.; Gorshkov, M.; Rockwood, A. L.; Smith, R. D."A Novel High-Performance Fourier Transform Ion Cyclotron Resonance Cell for Improved Biopolymer Characterization." Journal of Mass Spectrometry 2000, 35, 85-94. (98) Gooden, J. K.; Rempel, D. L.; Gross, M. L."Evaluation of Different Combinations of Gated Trapping, RF-only Mode, and Trap Compensation for in-Field MALDI Fourier Transform Mass Spectrometry." Journal of the American Society for Mass Spectrometry 2004, 15, 1109-1115. (99) Brustkern, A. M.; Rempel, D. L.; Gross, M. L."An Electrically Compenstaed Trap Designed to Eighth Order for FT-ICR Mass Spectrometry." Journal of the American Society for

Mass Spectrometry 2008, 19, 1281-1285. (100) Brustkern, A. M.; Rempel, D. L.; Gross, M. L."A Tuning Method for Electrically Compensated Ion Cyclotron Resonance Mass Spectrometer Traps." Journal of the American

Society for Mass Spectrometry 2010, 21, 451-454. (101) Kaiser, N. k.; Savory, J. J.; Mckenna, A. M.; Quinn, J. P.; Hendrickson, C. L.; Marshall, A. G."Electrically Compensated Fourier Transform Ion Cycltron Resonance Cell for Complex Mixture Mass Analysis." Analytical Chemistry 2011, 83, 6907-6910. (102) Tolmachev, A. V.; Robinson, E. W.; Smith, R. D.; Pasa-Tolic, L."Trapping Radial Electric Field Optimization in Compensated FTICR Cells." J. Am. Soc. Mass. Spectrom 2011, 22, 1334-1342.

Page 137: Broadband Phase Correction Of Fourier Transform Ion Cyclotron Mass Spectra

120

(103) Valeja, S. G.; Kaiser, N. K.; Xian, F.; Hendrickson, C. L.; Rouse, J. C.; Marshall, A. G."Unit Mass Baseline Resolution for an Intact 148 kDa Therapeutic Monoclonal Antibody by Fourier Transform Ion Cyclotron Resonance Mass Spectrometry." Anal. Chem. 2011, 83, 8391-8395. (104) Misharin, A. S.; Zubarev, R. A.; Doroshenko, V. M."Fourier Transform Ion Cyclotron Resonance Mass Spectrometer with Coaxial Multi-electrode Cell (O-Trap): First Experimental Demonstration." Rapid Communications In Mass Spectrometry 2010, 24, 1931-1940. (105) Ledford, E. B., Jr.; Rempel, D. L.; Gross, M. L."Space Charge Effects in Fourier Transform Mass Spectrometry. Mass Calibration." Anal. Chem. 1984, 56, 2744-2748. (106) Shi, S. D.-H.; Drader, J. J.; Freitas, M. A.; Hendrickson, C. L.; Marshall, A. G."Comparison and Interconversion of the Two Most Common Frequency-to-Mass Calibration Function for Fourier Transform Ion Cyclotron Resonance Mass Spectrometry." Int. J. Mass

Spectrom. 1999, 195/196, 591-598. (107) Jefferies, J. B.; Barlow, S. E.; Dunn, G. H."Theory of Space-Charge Shift of Ion-Cyclotron Resonance Frequencies." Internation Journal of Mass Spectrometry and Ion Processes 1983, 34, 169-187. (108) Easterling, M. L.; Mize, T. H.; Amster, I. J."Routine Part-per-Million Mass Accuracy for High Mass Ions: Space-Charge Effects in MALDI FT-ICR." Analytical Chemistry 1999, 71, 624-632. (109) Muddiman, D. C.; Oberg, A. L."Statistical Evaluation of Internal and External Mass Calibration Laws Utilized in Fourier Transform Ion Cyclotron Resonance Mass Spectrometry." Analytical Chemistry 2005, 77, 2406-2414. (110) Gorshkov, M.; Good, D. M.; Lyutvinshiy, Y.; Yang, H.; Zubarev, R. A."Calibration Function for the Oribitrap FTMS Accounting for the Space Charge Effect." Journal of the

American Society for Mass Spectrometry 2010, 21, 1846-1851. (111) Comisarow, M. B.; Marshall, A. G."Selective-Phase Ion Cyclotron Resonance Spectroscopy." Can. J. Chem. 1974, 52, 1997-1999. (112) Vining, B. A.; Bossio, R. E.; Marshall, A. G."Phase Correction for Collision Model Analysis and Enhanced Resolving Power of Fourier Transform Ion Cyclotron Resonance Mass Spectra." Anal. Chem. 1999, 71, 460-467. (113) Beu, S. C.; Blakney, G. T.; Quinn, J. P.; Hendrickson, C. L.; Marshall, A. G."Broadband Phase Correction of FT-ICR Mass Spectr via Simultaneous Excitation and Detection." Anal.

Chem 2004, 76, 5756-5761.

Page 138: Broadband Phase Correction Of Fourier Transform Ion Cyclotron Mass Spectra

121

(114) Xian, F.; Hendrickson, C. L.; Blakney, G. T.; Beu, S. C.; Marshall, A. G."Automated Broadband Phase Correction of Fourier Transform Ion Cyclotron Resonance Mass Spectra." Anal. Chem. 2010, 82, 8807-8812. (115) Qi, Y.; Thompson, C. J.; Van Ordan, S. L.; O'Connor, P. B."Phase Correction of Fourier Transform Ion Cyclotron Resonance Mass Spectra Using Matlab." J. Am. Soc. Mass. Spectrom 2011, 22, 138-147. (116) Lange, O.; Damoc, E.; Wieghaus, A.; Makarov, A., proceedings of the 59th ASMS

Conference on Mass Spectrometry & Allied Topics, Denver, Colorado 2011. (117) Blakney, G. T.; Hendrickson, C. L.; Marshall, A. G."Predator Data Station: A Fast Data Acquisition System for Advanced FT-ICR MS Experiments." Int. J. Mass Spectrom. 2011, 306, 246-252. (118) Mize, T. H.; Taban, I. M.; Duursma, M.; Seynen, M.; Konijnenburg, M.; Vijftigschild, A.; Doornik, C. V.; Rooij, G. V.; Heeren, R. M. A."A Modular Data and Control System to Improve Sensitivity, Selectivity, Speed of Analysis, Ease of Use, and Transient Duration in an External Source FTICR-MS." Internation Journal of Mass Spectrometry 2004, 235, 243-253. (119) Taban, I. M.; Burge, Y. E. M.; Duursma, M.; Takats, Z.; Seynen, M.; Konijnenburg, M.; Vijftigschild, A.; Attema, I.; Heeren, R. M. A."A Novel Workflow Control System for Fourier Transform Ion Cyclotron Resonance Mass Spectrometry Allows for Unique on-the-Fly Data-Dependent Decisions." Rapid Communications In Mass Spectrometry 2008, 22, 1245-1256. (120) Perry, R. H.; Hu, Q.; Salazar, G. A.; Cooks, R. G.; Noll, R. J."Rephasing Ion Packet in the Orbitrap Mass Analyzer to Improve Resolution and Peak Shape." Journal of the American

Society for Mass Spectrometry 2009, 20, 1397-1404. (121) Macek, B.; Waanders, L. F.; Olsen, J. V.; Mann, M."Top-Down Protein Squencing and MS3 on a Hybrid Linear Quadrupole Ion Trap-Orbitrap Mass Spectrometer." Molecular &

Cellular Proteomics 2006, 5, 949-958. (122) Frese, C. K.; Altelaar, A. F. M.; Hennrich, M. L.; Nolting, D.; Zeller, M.; Griep-Raming, J.; Heck, A. J. R.; Mohammed, S."Improved Peptide Identification by Targeted Fragmentation Using CID, HCD and ETD on an LTQ-Orbitrap Velos." Journal of Proteome Research 2011, 10, 2377-2388. (123) Hakansson, K.; Cooper, H. J.; Emmett, M. R.; Costello, C. E.; Marshall, A. G.; Nilsson, C. L."Electron Capture Dissociation and Infrared Mutiphoton Dissociation MS/MS of an N-Glycosylated Tryptic Peptide to Yield Complementary Sequence Information." Analytical

Chemistry 2001, 73, 4530-4536.

Page 139: Broadband Phase Correction Of Fourier Transform Ion Cyclotron Mass Spectra

122

(124) McAlister, G. C.; Phanstiel, D.; Good, D. M.; Berggren, T.; Coon, J. J."Implementation of Electron-Transfer Dissociation on a Hybrid Linear Ion Trap-Orbitrap Mass Spectrometer." Analytical Chemistry 2007, 79, 3525-3534. (125) Mann, M.; Kelleher, N. L."Precision Proteomics: The Case for High Resolution and High Mass Accuracy." Proceedings of the National Academy of Sciences 2008, 105, 18132-18138. (126) Xu, F.; Xu, Q.; Dong, X.; Guy, M.; Guner, H.; Hacker, T.; Ge, Y."Top-Down High-Resolution Electron Capture Dissociation Mass Spectrometry for Comprehensive Characterization of Post-Translational Modifications in Rhesus Monkey Cardiac Troponin I." Internation Journal of Mass Spectrometry 2011, 305, 95-102. (127) Ryan, C. M.; Souda, P.; Bassilian, S.; Ujwal, R.; Zhang, J.; Abramson, J.; Ping, P.; Durazo, A.; Bowie, J. U.; Saif Hasan, S.; Baniulis, D.; Cramer, W. A.; Faull, K. F.; Whitelegge, J. P."Post-Translational Modifications of Integral Membrane Proteins Resolved by Top-Down Fourier Transform Mass Spectrometry with Collisionally Activated Dissociation." 2010, 791-803. (128) Pesavento, J. J.; Yang, H.; Kelleher, N. L.; Mizzen, C. A."Certain and progressive Methylation of Histone H4 at Lysine 20 During the Cell Cycle." Molecular and Cellular Biology 2008, 28, 468-486. (129) Li, M.; Jiang, L.; Kelleher, N. L."Global Histone Profiling by LC-FTMS after Inhibition and Knockdown of Deacetylases in Human Cells." Journal of Chromatography B. 2009, 877, 3885-3892. (130) Ferguson, J. T.; Wenger, C. D.; Metcalf, W. W.; Kelleher, N. L."Top-Down Proteomics Reveals Novel Proteins Forms Expressed in Methanosarcina Acetivorans." Journal of the

American Society for Mass Spectrometry 2009, 20, 1743-1750. (131) Pesavento, J. J.; Bullock, C. R.; LeDuc, R. D.; Mizzen, C. A.; Kelleher, N. L."Combinatorial Modification of Human Histone H4 Quantitated by Two-Dimensional Liquid Chromatography Coupled with Top-Down Mass Spectrometry." Journal of Proteome Research 2008, 283, 14927-14937. (132) Lee, J. E.; Kellie, J. F.; Tran, J. C.; Tipton, J. D.; Catherman, A. D.; Thomas, H. M.; Ahlf, D. R.; Durbin, K. R.; Vellaichamy, A.; Ntai, I.; Marshall, A. G.; Kelleher, N. L."A Robust Two-Dimensional Separation for Top-Down Tandem Mass Spectrometry of the Low-Mass Proteome." J. Am. Soc. Mass. Spectrom. 2009, 20, 2183-2191. (133) Durbin, K. R.; Tran, J. C.; Zamdborg, L.; Sweet, S. M. M.; Catherman, A. D.; Lee, J. E.; Li, M.; Kellie, J. F.; Kelleher, N. L."Intact Mass Detection, Interpretation, and Visualization to Automate Top-Down Proteomics on a Large Scale." Proteomics 2010, 10, 3589-3597. (134) Sampson, J. S.; Murray, K. K.; Muddiman, D. C."Intact and Top-Down Characterization of Biomolecules and Direct Analysis Using Infrared Matrix-Assisted Laser Desorption

Page 140: Broadband Phase Correction Of Fourier Transform Ion Cyclotron Mass Spectra

123

Electrospray Ionization Coupled to FT-ICR Mass Spectrometry." Journal of the American

Society for Mass Spectrometry 2009, 20, 667-673. (135) Lee, J. E.; Atkins, N.; Hatcher, N. G.; Zamdborg, L.; Gillette, M. U.; Sweedler, J. V.; Kelleher, N. L."Endogenous Peptide Discovery of the Rat Circadian Clock." Molecular &

Cellular Proteomics 2010, 9, 285-297. (136) Warder, S. E.; Tucker, L. A.; McLoughlin, S. M.; Strelitzer, T. J.; Meuth, J. L.; Zhang, Q.; Sheppard, G. S.; Richardson, P. L.; Lesniewski, R.; Davidsen, S. K.; Bell, R. L.; Rogers, J. C.; Wang, J."Discovery, Identification, and Characterization of Candidate Pharmacodynamic Markers of Methionine Aminopeptidase-2 Inhibition." Journal of Proteome Research 2008, 7, 4807-4820. (137) Han, X.; Aslanian, A.; Yates III, J. R."Mass Spectrometry for Proteomics." Current

Opinion in Chemical Biology 2008, 12, 483-490. (138) Armirotti, A."Bottom-Up Proteomics." Current Analytical Chemistry 2009, 5, 116-130. (139) Brosch, M.; Swamy, S.; Hubbard, T.; Choudhary, J."Comparison of Mascot and X!Tandem Performance for Low and High Accuracy Mass Spectrometry and the Development of an Adjusted Mascot Threshold." Molecular & Cellular Proteomics 2008, 962-970. (140) Tolmachev, A. V.; Monroe, M. E.; Purvine, S. O.; Moore, R. J.; Jaitly, N.; Adkins, J. N.; Anderson, G. A.; Smith, R. D."Characterization of Strategies for Obtaining Cofident Identifications in Bottom-Up Proteomics Measurements Using Hybrid FTMS Instruments." Analytical Chemistry 2008, 80, 8514-8525. (141) Sweet, S. M. M.; Jones, A. W.; Cunningham, D. L.; Heath, J. K.; Creese, A. J.; Cooper, H. J."Database Search Strategies for Proteomic Data Sets Generated by Electron Capture Dissociation Mass Spectrometry." Journal of Proteome Research 2009, 8, 5475-5484. (142) Karpievitch, Y.; Stanley, J.; Taverner, T.; Huang, J.; Adkins, J. N.; Ansong, C.; Heffron, F.; Metz, T. O.; Qian, W.-J.; Yoon, H.; Smith, R. D.; Dabney, A. R."A Statistical Framework for Protein Quantitation in Bottom-Up MA-Based Proteomics." Bioinformatics 2009, 25, 2028-2034. (143) Bereman, M. S.; Egertson, J. D.; MacCoss, M. J."Comparison Between Procedures Using SDS for Shotgun Proteomic Analyses of Complex Samples." Proteomics 2011, 11, 2931-2935. (144) Porteus, B.; Kocharunchitt, C.; Nilsson, R. E.; Ross, T.; Bowman, J. P."Utility of Gel-Free, Lable-Free Shotgun Proteomics Approaches to Investigate Microorganisms." Applied

Microbiology and Biotechnology 2011, 90, 407-416. (145) Plazas-Mayorca, M. D.; Zee, B. M.; Young, N. L.; Fingerman, I. M.; LeRoy, G.; Briggs, S. D.; Garcia, B. A."One-Pot Shotgun Quantitative Mass Spectrometry Characterization of Histones." Journal of Proteome Research 2009, 8, 5367-5374.

Page 141: Broadband Phase Correction Of Fourier Transform Ion Cyclotron Mass Spectra

124

(146) Sweet, S. M. M.; Li, M.; Thomas, P. M.; Durbin, K. R.; Kelleher, N. L."Kinetics of Re-establishing H3K79 Methylation Marks in Global Human Chromatin." The Journal of Biological

Chemsitry 2010, 285, 32778-32786. (147) Bumpus, S. B.; Evans, B. S.; Thomas, P. M.; Ntai, I.; Kelleher, N. L."A Proteomics Approach to Discovering Natural Products and Their Biosynthetic Pathways." 2009, 27, 951-955. (148) Boyne II, M. T.; Garcia, B. A.; Li, M.; Zamdborg, L.; Wenger, C. D.; Babal, S.; Kelleher, N. L."Tandem Mass Spectrometry with Ultrahigh Mass Accuracy Clarifies Peptide Identification by Database Retrieval." Journal of Proteome Research 2009, 8, 374-379. (149) Garcia, B. A.; Eric Thomas, C.; Kelleher, N. L.; Mizzen, C. A."Tissue-Specific Expression and Post-Translational Modification of Histone H3 Variants." Journal of Proteome

Research 2008, 7, 4225-4236. (150) Ong, S.-E.; Blagoev, B.; Kratchmarova, I.; Kristensen, D. B.; Steen, H.; Pandey, A.; Mann, M."Stable Isotope Labeling by Amino Acids in Cell Culture, SILAC, as a Simple and Accurate Approach to Expression Proteomics." Molecular & Cellular Proteomics 2002, 1, 376-586. (151) Manes, N. P.; Estep, R. D.; Mottaz, H. M.; Moore, R. J.; Clauss, T. R.; Monroe, M. E.; Du, X.; Adkins, J. N.; Wong, S. W.; Smith, R. D."Comparative Proteomics of Human Monkeypox and Vaccinia Intracellular Mature and Extracellular Enveloped Virions." Journal of

Proteome Research 2008, 7, 960-968. (152) Wu, J.; Warren, P.; Shakey, Q.; Sousa, E.; Hill, A.; Ryan, T.; He, T."Integrating Titania Enrichment, iTRAQ Labeling, and Orbitrap CID-HCD for Global Identification and Quantitative Analysis of Phosphopeptides." Proteomics 2010, 10, 2224-2234. (153) Zhang, Y.; Ficarro, S. B.; Li, S.; Marto, J. A."Optimized Orbitrap HCD for Quantitative Analysis of Phosphopeptides." Journal of the American Society for Mass Spectrometry 2009, 20, 1425-1434. (154) Dayon, L.; Pasquarello, C.; Hoogland, C.; Sanchez, J.-C.; Scherl, A."Combining Low and High Energy Tandem Mass Spectra for Optimized Peptide Quantification with Isobaric Tags." Journal of Proteomics 2010, 73, 769-777. (155) Zybailov, B.; Coleman, M. K.; Florens, L.; Washburn, M. P."Correlation of Relative Abundance Ratios Derived from Peptide Ion Chromatograms and Spectrum Counting for Quantitative Proteomic Analysis Using Stable Isotope Labeling." Analytical Chemistry 2005, 77, 6218-6224. (156) Taylor, N. S.; Weber, R. J. M.; Southam, A. D.; Payne, T. G.; Hrydziuszko, O.; Arvanitis, T. N.; Viant, M. R."A New Approach to Toxicity Testing in Daphnia Magna:

Page 142: Broadband Phase Correction Of Fourier Transform Ion Cyclotron Mass Spectra

125

Application of High Throughput FT-ICR Mass Spectrometry Metabolomics." Metabolomics 2009, 5, 44-58. (157) Denkert, C.; Budczies, J.; Weichert, W.; Wohlgemuth, G.; Scholz, M.; Kind, T.; Niesporek, S.; Noske, A.; Buckendahl, A.; Dietal, M.; Fiehn, O."Metabolite Profiling of Human Colon Carcinoma-Deregulation of TCA Cycle and Amino Acid Turnover." Molecular Cancer 2008, 7. (158) Takahashi, H.; Kai, K.; Shinbo, Y.; Tanaka, K.; Ohta, D.; Oshima, T.; Altaf-UI-Amin, M.; Kurokawa, K.; Ogasawara, N.; Kanaya, S."Metabolomics Apporach for Determining Growth Specific Metabolites Based on Fourier Transform Ion Cyclotron Resonance Mass Spectrometry." Analytical and Bioanalytical Chemistry 2008, 391, 2769-2782. (159) Baidoo, E. E. K.; Benke, P. I.; Neususs, C.; Pelzing, M.; Kruppa, G.; Leary, J. A.; Keasling, J. D."Capillary Electrophoresis-Fourier Transform Ion Cyclotron Resonance Mass Spectrometry for the Identification of Cationic Metabolites Via a pH-Mediated Stacking -Transient Isotachophoretic Method." Analytical Chemistry 2008, 80, 3112-3122. (160) Dunn, W. B.; Broadhurst, D.; Brown, M.; Baker, P. N.; Redman, C. W. G.; Kenny, L. C.; Kell, D. B."Metabolic Profiling of Serum Using Ultra Performance Liquid Chromatography and the LTQ-Orbitrap Mass Spectrometry System." Journal of Chromatography B. 2008, 871, 288-298. (161) Kamleh, M. A.; Hobani, Y.; Dow, J. A. T.; Watson, D. G."Metabolomic Profiling of Drosophila Using Liquid Chromatograpphy Fourier Transform Mass Spectrometry." FEBS

Letters 2008, 582, 2916-2922. (162) Werner, E.; Croixmarie, V.; Umbdenstock, T.; Ezan, E.; Chaminade, P.; Tabet, J.-C.; Junot, C."Mass Spectrometry-Based Metabolomics Accelerating the Characterization of Discriminating Signals by Combining Statistical Correlations and Ultrahigh Resolution." Analytical Chemistry 2008, 80, 4918-4932. (163) Jin, Z.; Daiya, S.; Kenttamaa, H."Characterization of Nonpolar Lipids and Selected Steroids by Using Laser-Induced Acoustic Desorption/Chemical Ionization, Atmosperic Pressure Chemical Ionization, and Electrospray Ionization Mass Spectrometry." Internation Journal of

Mass Spectrometry 2011, 301, 234-239. (164) Murphy, R. C.; Gaskell, S. J."New Applications of Mass Spectrometry in Lipid Analysis." The Journal of Biological Chemsitry 2011, 286, 25427-25433. (165) Huan, H.; Emmett, M. R.; Nilsson, C. L.; Conrad, C. A.; Marshall, A. G."High Mass Accuracy and Resolution Facilitate Identification of Glycosphingolipids and Phospholipids." Internation Journal of Mass Spectrometry 2010, 305, 116-119. (166) Huan, H.; Nilsson, C. L.; Emmett, M. R.; Ji, Y.; Marshall, A. G.; Kroes, R. A.; Moskai, J. R.; Colman, H.; Lang, F. F.; Conrad, C. A."Polar Lipid Remodeling and Increased Sulfatide

Page 143: Broadband Phase Correction Of Fourier Transform Ion Cyclotron Mass Spectra

126

Expression are Associated with the Glioma Therapeutic Candidates Wild Type P53 Elevation and the Topoisomerase-1 Inhibitor, Irinotecan." Glycoconjugate Journal 2010, 27, 27-38. (167) Huan, H.; Rodger, R. P.; Marshall, A. G.; Hsu, C. S."Algae Polar Lipids Characterized by Online Liquid Chromatography Coupled with Hybrid Linear Quadrupole Ion Trap/Fourier Transform Ion Cyclotron Resonance Mass Spectrometry." Energy & Fuels 2011, 26, 4770-4775. (168) Lane, A. N.; Fan, T. W.-M.; Xie, Z.; Moseley, H. N. B.; Higashi, R. M."Isotopomer Analysis of Lipid Biosynthesis by High Resolution Mass Spectrometry and NMR." Analytica

Chimica Acta 2009, 651, 201-208. (169) Garnier, N.; Rolando, C.; Hotje, J. M.; Tokarski, C."Analysis of Archaeological Triacylglycerols by High Resolution nanoESI, FT-ICR MS and IRMPD MS/MS: Application to 5th Century BC-4th Century AD Oild Lamps From Olbia (Ukraine)." Internation Journal of

Mass Spectrometry 2009, 284, 47-56. (170) Komaniecka, I.; Choma, A.; Lindner, B.; Holst, O."The Structure of a Novel Neutral Lipid a from the Lipopolysaccharide of Bradyrhizobium Elkanii Containing Three Mannose Units in the Backbone." Chemistry - A European Journal 2009, 16, 2922-2929. (171) Vidova, V.; Pol, J.; Volny, M.; Novak, P.; Havlicek, V.; Wiedmer, S.; Holopainen, J. M."Visualizing Spatial Lipid Distribution in Porcine Lens by MALDI Imaging High-Resolution Mass Spectrometry." Journal of Lipid Research 2010, 51, 2295-2299. (172) Pol, J.; Vidova, V.; Hyotylainen, T.; Volny, M.; Novak, P.; Strohalm, M.; Kostiainen, R.; Havlicek, V.; Wiedmer, S.; Holopainen, J. M."Spatial Distribution of Glycerophospholipids in the Ocular Lens." Plos ONE 2011, 6, 1-9. (173) Zou, Y.; Tiller, P.; Chen, I.-W.; Beverly, M.; Hochman, J."Metabolite Identification of Small Interfering RNA Duplex by High-Resolution Accurate Mass Spectrometry." Rapid

Communications In Mass Spectrometry 2008, 22, 1871-1881. (174) Bahr, U.; Aygun, H.; Karas, M."Sequencing of Single and Double Stranded RNA Oligonucleotides by Acid Hydrolysis and MALDI Mass Spectrometry." Analytical Chemistry 2009, 81, 3173-3179. (175) Taoka, M.; Yamauchi, Y.; Nobe, Y.; Masaki, S.; Nakayama, H.; Ishikawa, H.; Takahashi, N.; Isobe, T."An analytical Platform for Mass Spectrometry-Based Identification and Chemical Analysis of RNA in Ribonucleoprotein Complexes." Nucleic Acids Res. 2009, 37, 1-14. (176) Turner, K. B.; Brinson, R. G.; Young Yi-Brunozzi, H.; Rausch, J. W.; Miller, J. T.; Le Grice, S. F. J.; Marino, J. P.; Fabris, D."Structural Probing of the HIV-1 Polypurine Tract RNA:DNA Hybrid Using Classic Nucleic Acid Ligands." Nucleic Acids Research 2008, 36, 2799-2810. (177) Turner, K. B.; Young, Y.-B., Hye.; Brinson, R. G.; Marino, J. P.; Fabris, D.; Le Grice, S. F. J."SHAMS: Combining Chemical Modification of RNA with Mass Spectrometry to Examine

Page 144: Broadband Phase Correction Of Fourier Transform Ion Cyclotron Mass Spectra

127

Polypurine Tract-Containing RNA/DNA Hybrids." Cold Spring Harbor Lab Press 2009, 15, 1605-1613. (178) Turner, K. B.; Kohlway, A. S.; Hagan, N. A.; Fabris, D."Noncovalent Probes for the Investigation of Structure and Dynamics of Protein-Nucleic Acid Assdmblies: The case of NC-Mediated Dimerization of Genomic RNA in HIV-a." Biopolymers 2008, 91, 283-295. (179) Yu, E. T.; Hawkins, A.; Eaton, J.; Fabris, D."MS3D Structural Elucidation of theHIV-1 Packaging Signal." Proceedings of the National Academy of Sciences 2008, 105, 12248-12253. (180) Weidner, S. M.; Trimpin, S."Mass Spectrometry of Synthetic Polymers." Anal. Chem. 2010, 82, 4811-4829. (181) Pernot, P.; Carrasco, N.; Thissen, R.; Schmitz-Afonso, I."Tholinomics-Chemical Analysis of Nitrogen-Rich Polymers." Anal. Chem. 2010, 82, 1371-1380. (182) Wienhofer, I. C.; Luftmann, H.; Studer, A."Nitroxide-Mediated Copolymerization of MMA with Styrene: Sequence Analysis of Oligomers by Using Mass Spectrometry." Macromolecules 2011, 44, 2510-2523. (183) Kaczorowska, M.; Cooper, H. J."Electron Capture Dissociation, Electron Detachment Dissociation, and Collision-Induced Dissociation of Polyamindoamine (PAMAM) Dendrimer Ions with Amino, Amidoethanol, and Sodium Carboxylate Surface Group." Journal of the

American Society for Mass Spectrometry 2008, 19, 1312-1319. (184) Kaczorowska, M.; Cooper, H. J."Characterization of Polyphosphoesters by Fourier Transform Ion Cyclotron Resonance Mass Spectrometry." Journal of the American Society for

Mass Spectrometry 2009, 20, 2238-2247. (185) Kaczorowska, M.; Cooper, H. J."Electron Capture Dissociation and Collision-Induced Dissociation of Metal Ion (Ag+, Cu2+, Zn2+, Fe2+, and Fe3+) Complexes of Polyamidoamine (PAMAM) Dendrimers." Journal of the American Society for Mass Spectrometry 2009, 20, 674-681. (186) Song, J.; van Velde, J. W.; Vertommen, L. L. T.; van der Ven, L. G. J.; Heeren, R. M. A.; Van den Brink, O. F."Investigation of Polymerization Mechanisms of Poly(n-Butyl Acrylate)s Generated in Different Solvents by LC-ESI-MS2." Macromolecules 2010, 43. (187) Miladinovic, S. M.; Kaeser, C. J.; Knust, M. M.; Wilkins, C. L."Tandem Fourier Transform Mass Spectrometry of Block and Random Copolymer." Internation Journal of Mass

Spectrometry 2011, 301, 184-194. (188) Nasioudis, A.; Heeren, R. M. A.; van Doormalen, I.; de Wijs-Rot, N.; Van den Brink, O. F."Electrospray Ionization Tandem Mass Spectrometry of Ammonium Cationized Polyethers." Journal of the American Society for Mass Spectrometry 2011, 22, 837-844.

Page 145: Broadband Phase Correction Of Fourier Transform Ion Cyclotron Mass Spectra

128

(189) Marshall, A. G.; Rodgers, R. P."Petroleomics: Chemistry of the underworld." Proceedings of the National Academy of Sciences of the United States of America 2008, 105, 18090-18095. (190) Rodgers, R. P.; McKenna, A. M."Petroleum Analysis." Analytical Chemistry 2011, 83, 4665-4687. (191) Fernandez-Lima, F. A.; Becker, C.; Mckenna, A. M.; Rodger, R. P.; Marshall, A. G.; Russell, D. H."Petroleum Crude Oild Characterization by IMS-MS and FTICR MS." Analytical

Chemistry 2009, 81, 9941-9947. (192) Betancourt, S. S.; Todd Ventura, G.; Pomerantz, A. E.; Viloria, O.; Dubost, F. X.; Zuo, J.; Monson, G.; Bustamante, D.; Purcell, J. M.; Nelson, R. K.; Rodgers, R. P.; Reddy, C. M.; Marshall, A. G.; Mullins, O. C."Nanoaggregates of Asphaltenes in a Reservoir Crude Oil and Reservoir Connectivity." Energy & Fuels 2009, 23, 1178-1188. (193) Mckenna, A. M.; Purcell, J. M.; Rodger, R. P.; Marshall, A. G."Identification of Vanadyl Porphyrins in a Heavy Crude Oil and Raw Asphaltene by Atmospheric Pressure Photoionization Fourier Transform Ion Cyclotron Resonance (FT-ICR) Mass Spectrometry." Energy & Fuels 2009, 23, 2122-2128. (194) Pinkston, D. S.; Duan, P.; Gallardo, V.; Habicht, S. C.; Tan, X.; Qian, K.; Gray, M.; Mullen, K.; Kenttamaa, H. I."Analysis of Asphaltenes and Asphaltene Model Compounds by Laser-Induced Acoustic Desorption/Fourier Transform Ion Cyclotron Resonance Mass Spectrometry." Energy & Fuels 2009, 23, 5564-5570. (195) Smith, D. F.; Rodgers, R. P.; Rahimi, P.; Teclemariam, A.; Marshall, A. G."Effect of Thermal Treatment on Acidic Organic Species from Athabasca Bitument Heavy Vacuum Gas Oil, Analyzed by Negative-Ion Electrospray Fourier Transform Ion Cyclotron Resonance (FT-ICR) Mass Spectrometry." 2009. (196) Purcell, J. M.; Merdrignac, I.; Rodgers, R. P.; Marshall, A. G.; Gauthier, T.; Guibard, I."Stepwise Structural Characterization of Asphaltenes During Deep Hydroconversion Processed Determined by Atmospheric Pressure Photoionization (APPI) Fourier Transform Ion Cyclotron (FT-ICR) Mass Spectrometry." Energy & Fuels 2010, 24, 2257-2265. (197) Mullins, O. C.; Sheu, E. Y.; Hammami, A.; Marshall, A. G., Eds. Asphaltenes, Heavy

Oils, and Petroleomics; Springer: New York, 2007. (198) Mckenna, A. M.; Blakney, G. T.; Xian, F.; Glaser, P. B.; Rodger, R. P.; Marshall, A. G."Heavy Petroleum Composition. 2. Progression of the Boduszynski Model to the Limit of Distillation by Ultrahigh-Resolution FT-ICR Mass Spectrometry." Energy & Fuels 2010, 24, 2939-2946. (199) Mckenna, A. M.; Purcell, J. M.; Rodgers, R. P.; Marshall, A. G."Heavy Petroleum Composition. 1. Exhaustive Compositional Analysis of Athabasca Bitumen HVGO Distillates by

Page 146: Broadband Phase Correction Of Fourier Transform Ion Cyclotron Mass Spectra

129

Fourier Transform Ion Cyclotron Resonance Mass Spectrometry: A Definitive Test of the Boduszynski Model." Energy & Fuels 2010, 24, 2929-2938. (200) Hughey, C. A.; S., M. C.; Galasso-Roth, S. A.; Paspalof, G. B.; Mapolelo, M.; Rodgers, R. P.; Marshall, A. G.; Ruderman, D. L."Naphthenic Acids as Indicators of Curde Oil Biodegradation in Soil, Based on Semi-Quantitative Electrospray Ionization Fourier Transform Ion Cyclotron Resonance Mass Spectrometry." Rapid Communications In Mass Spectrometry 2008, 22, 3968-3976. (201) Mapolelo, M.; Stanford, L. A.; Rodgers, R. P.; Yen, A. T.; Debord, J. D.; Asomaning, S.; Marshall, A. G."Chemical Speciation of Calcium and Sodium Naphthenate Deposits by Electrospray Ionization FT-ICR Mass Spectrometry." Energy & Fuels 2009, 23, 349-355. (202) Mapolelo, M.; Rodgers, R. P.; Blakney, G. T.; Yen, A. T.; Asomaning, S.; Marshall, A. G."Characterization of Naphthenic Acids in Crude Oils and Naphthenates by Electrospray Ionization FT-ICR Mass Spectrometry." Internation Journal of Mass Spectrometry 2011, 300, 149-157. (203) Headley, J. V.; Peru, K. M.; Mishra, S.; Meda, V.; Dalai, A. K.; McMartin, D. W.; Mapolelo, M.; Rodgers, R. P.; Marshall, A. G."Characterization of Oil Sand Naphthenic Acids Treated with Ultraviolet and Microwave Radiation by Negative Ion Electrospray Fourier Transform Ion Cyclotron Resonance Mass Spectrometry." Rapid Communications In Mass

Spectrometry 2010, 24, 3121-3126. (204) Headley, J. V.; Barrow, M. P.; Peru, K. M.; Derrick, P."Salting-out Effects on the Characterization of Naphthenic Acis from Athabasca Oild Sands Using Electrospray Ionization." Journal of Environmental Science and Health Part A 2011, 46, 844-854. (205) Headley, J. V.; Peru, K. M.; Janfada, A.; Fahlman, B.; Gu, C.; Hassan, S."Characterization of Oil Sands Acids in Plant Tissue Using Orbitrap Ultra-High Resolution Mass Spectrometry with Electrospray Ionization." Rapid Communications In Mass Spectrometry 2011. (206) Pomerantz, A. E.; Todd Ventura, G.; Mckenna, A. M.; Canas, J. A.; Auman, J.; Koerner, K.; Curry, D.; Nelson, R. K.; Reddy, C. M.; Rodgers, R. P.; Marshall, A. G.; Peters, K. E.; Mullins, O. C."Combining Biomarker and Bulk Compositional Gradient Analysis to Assess Reservoir Connectivity." Organic Geochemistry 2010, 41, 812-821. (207) Juyal, P.; Mckenna, A. M.; Yen, A. T.; Rodgers, R. P.; Reddy, C. M.; Nelson, R. K.; Ballard Andrews, A.; Atolia, E.; Allenson, S. J.; Mullins, O. C.; Marshall, A. G."Analysis and Identification of Biomarkers and Origin of Color in a Bright Blue Crude Oil." Energy & Fuels 2011, 25, 172-182. (208) Kellmann, M.; Muenster, H.; Zomer, P.; Mol, H."Full Scan MS in Comprehensive Qualitative and Quantitative Eesidue Analysis in Food and Feed Matrices: How Much Resolving

Page 147: Broadband Phase Correction Of Fourier Transform Ion Cyclotron Mass Spectra

130

Power is Required?" Journal of the American Society for Mass Spectrometry 2009, 20, 1464-1476. (209) Krauss, M.; Singer, H.; Hollender, J."LC-High Resolution MS in Environmantal Analysis from Target Screening to the Identification of Unknows." Analytical and Bioanalytical

Chemistry 2010, 397, 943-951. (210) Helbling, D. E.; Hollender, J.; Kohler, H.-P. E.; Singer, H.; Fenner, K."High Throughput Identification of Microbial Transformation Products of Organic Micropollutants." Environmental

Science & Technology 2010, 44, 6621-6627. (211) Liou, J. S.-C.; Szostek, B.; DeRito, C. M.; Madsen, E. L."Investigating the Biodegradability of Perfluorooctanoic Acid." Chemosphere 2010, 80, 176-183. (212) van Leerdam, J. A.; Hogenboom, A. C.; van der Kooi, M. M. E.; Voogt, P. d."Determination of Polar 1H-benzotrizoles and Benzothiazoles in Water by Solid-Phase Extraction and Liquid Chromatography LTQ FT Orbitrap Mass Spectrmetry." Internation

Journal of Mass Spectrometry 2009. (213) Nam, S.; Joo, S.; Kim, S.; Baek, N.-I.; Choi, H.-K.; Park, S."Induced Metabolite Changes in Myriophyllum Spicatum During Co-Existence Experiment with the Cyanobacterium Microcystis Aeruginosa." Journal of Plant Biology 2008, 51, 373-378. (214) Schmidt, F.; Elvert, M.; Koch, B. P.; Witt, M.; Hinrichs, K.-U."Molecular Characterization of Dissolved Organic Matter in Pore Water of Continental Shelf Sediments." Geochimica et Cosmochimica Acta 2009, 73, 3337-3358. (215) Gonsior, M.; Zwartjes, M.; Cooper, W. J.; Song, W.; Ishida, K. P.; Tseng, L. Y.; Jeung, M. K.; Rosso, D.; Hertkorn, N.; Schmitt-Kopplin, P."Molecular Characterization of Effluent Organic Matter Identified by Ultrahigh Resolution Mass Spectrometry." Water Research 2011, 45, 2943-2953. (216) Bhatia, M. P.; Das, S. B.; Longnecker, K.; Charette, M.; Kujawinski, E. B."Molecular Characterization of Dissolved Organic Matter Associated with the Greenland Ice Sheet." Geochimica et Cosmochimica Acta 2010, 74, 3768-3784. (217) Marshall, A. G.; Comisarow, M. B.; Parisod, G."Relaxation and Spectral Line Shape in Fourier Transform Ion Cyclotron Resonance Spectroscopy." J. Chem. Phys. 1979, 71, 4434-4444. (218) Chow, K. H.; Comisarow, M. B."Frequency Errors and Phase Dependence in Magnitude Mode Apodized Fourier Transform Ion Cyclotron Resonance Spectra." International Journal of

Mass Spectrometry and Ion Processes 1989, 89, 187-203. (219) Comisarow, M. B.; Marshall, A. G."Frequency-Sweep Fourier Transform Ion Cyclotron Resonance Spectroscopy." Chem. Phys. Lett. 1974, 26, 489-490.

Page 148: Broadband Phase Correction Of Fourier Transform Ion Cyclotron Mass Spectra

131

(220) Marshall, A. G.; Roe, D. C."Theory of Fourier Transform Ion Cyclotron Resonance Mass Spectroscopy: Response to Frequency-Sweep Excitation." J. Chem. Phys. 1980, 73, 1581-1590. (221) Marshall, A. G.; Roe, D. C."Dispersion versus absorption: spectral line shape analysis for radiofrequency and microwave spectrometry." Analytical Chemistry 1978, 50, 756-763. (222) Craig, E. C.; Santos, I.; Marshall, A. G."Dispersion vs. Absorption (DISPA) Method for Automatic Phase Correction of Fourier Transform Ion Cyclotron Resonance Mass Spectra." Rapid Commun. Mass Spectrom. 1987, 1, 33-37. (223) Craig, E. C.; Marshall, A. G."Automated Phase Correction of FT NMR Spectra by Means of Phase Measurement Based on Dispersion versus Absorption Relation (DISPA)." J. Magn.

Reson. 1988, 76, 458-475. (224) Marshall, A. G."Convolution Fourier Transform Ion Cyclotron Resonance Spectroscopy." Chem. Phys. Lett. 1979, 63, 515-518. (225) Grothe, R. A. Estimation of Ion Cyclotron Resonance Parameters in Fourier Trnsform Mass Spectrometry, 2009,United States Patent Application, US20090278037A1

(226) Hakansson, K.; Chalmers, M. J.; Quinn, J. P.; McFarland, M. A.; Hendrickson, C. L.; Marshall, A. G."Combined electron capture and infrared multiphoton dissociation for multistage MS/MS in a Fourier transform ion cyclotron resonance mass spectrometer." Analytical

Chemistry 2003, 75, 3256-3262. (227) Blakney, G. T.; Robinson, D. E.; Ly, N. V.; Kelleher, N. L.; Hendrickson, C. L.; Marshall, A. G., Proceedings of the 53rd Amer. Soc. Mass Spectrom. Ann. Conf. on Mass

Spectrometry & Allied Topics, San Antonio,TX June 5-9, 2005; TP220. (228) Kendrick, E."A Mass Scale Based on CH2 = 14.0000 for High Resolution Mass Spectrometry of Organic Compounds." Anal. Chem. 1963, 35, 2146-2154. (229) Hughey, C. A.; Hendrickson, C. L.; Rodgers, R. P.; Marshall, A. G.; Qian, K."Kendrick Mass Defect Spectroscopy: A Compact Visual Analysis for Ultrahigh-Resolution Broadband Mass Spectra." Anal. Chem. 2001, 73, 4676-4681. (230) Bracewell, R.The Fourier Transform and Its Applications; McGraw-Hill Book Company: New York, 1965pp. (231) Wang, M.; Marshall, A. G."Laboratory-Frame and Rotating-Frame Ion Trajectories in Ion Cyclotron Resonance Mass Spectrometry." Int. J. Mass Spectrom. Ion Proc. 1990, 100, 323-346. (232) Lee, J.; Comisarow, M. B."Advantageous apodization functions for absorption-mode Fourier transform spectroscopy." Applied Spectroscopy 1989, 43, 599-604.

Page 149: Broadband Phase Correction Of Fourier Transform Ion Cyclotron Mass Spectra

132

(233) Schaefer, J.; Stejskal, E. O."Baseline artifacts in high-resolution fourier transform NMR spectra." Journal of magnetic resonance (1969) 1974, 15, 173-176. (234) Stejskal, E. O.; Schaefer, J."Comparisons of quadrature and single-phase fourier transform NMR." J. Magn. Reson. 1974, 14, 160-169. (235) Plateau, P.; Dumas, C.; Gueron, M."Solvent-peak-suppressed NMR: Correction of baseline distortions and use of strong-pulse excitation." J. Magn. Reson. (1969) 1983, 54, 46-53. (236) Tang, C."An Analysis of Baseline Distortion and Offset in NMR Spectra." J. Magn.

Reson. 1994, 109, 232-240. (237) Hoult, D. I.; Chen, C. N.; Eden, H.; Eden, M."Elimination of baseline artifacts in spectra and their integrals." J. Magn. Reson. 1983, 51, 110-117. (238) Otting, G.; Widmer, H.; Wagner, G.; Wuthrich, K."Origin of ?2 and ?2 ridges in 2D NMR spectra and procedures for suppression." J. Magn. Reson. 1986, 66, 187-193. (239) Freeman, R.A Handbook of Nuclear Magnetic Resonance; Longman Scientific & Technical, 1988;328 pp. (240) Marion, D.; Bax, A."Baseline distortion in real-fourier-transform NMR spectra." J.

Magn. Reson. 1988, 79, 352-356. (241) Wider, G."Elimination of Baseline Artifacts in NMR Spectra by Oversampling." J.

Magn. Reson. 1990, 89, 406-409. (242) Moskau, D."Application of real time digital filters in NMR spectroscopy." Concepts

Magn. Reson. 2002, 15, 164-176. (243) Heuer, A.; Haeberlen, U."A new method for suppressing baseline distortions in FT NMR." J. Magn. Reson. 1989, 85, 79-94. (244) Stephenson, D. S.; Binsch, G."Automated analysis of high-resolution NMR spectra. I. Principles and computational strategy." J. Magn. Reson. 1980, 37, 395-407. (245) Stephenson, D. S.; Binsch, G."Automated analysis of high-resolution NMR spectra. II. Illustrative applications of the computer program DAVINS." J. Magn. Reson. 1980, 37, 409-430. (246) Pearson, G."A general baseline-recognition and baseline-flattening algorithm." J. Magn.

Reson. 1977, 27, 265-272. (247) Dietrich, W.; Rudel, C. H.; Neumann, M."Fast and precise automatic baseline correction of one- and two-dimensional nmr spectra." J. Magn. Reson. 1991, 91, 1-11. (248) Brown, D."Fully Automated Baseline Correction of 1D and 2D NMR Spectra Using Bernstein Polynomials." J.Magn. Teson. A 1995, 114, 268-270.

Page 150: Broadband Phase Correction Of Fourier Transform Ion Cyclotron Mass Spectra

133

(249) Bartels, C.; Guntert, P.; Wutherich, K."IFLAT-A New Automatic Baseline-Correction Method for Multidimensional NMR Spectra with Strong Solvent Signals." J. Magn. Reson. 1995, 117, 330-333. (250) Carlos Cobas, J.; Bernstein, M. A.; Martin-Pastor, M.; Tahoces, P. G."A New General-purpose Fully Automatic Baseline-Correction Procedure for 1D and 2D NMR Data." J. Magn.

Reson. 2006, 183, 145-151. (251) Chang, D.; Banack, C. D.; Shah, S. L."Robust Baseline Correction Algorithm for Signal Dense NMR Spectra." J. Magn. Reson. 2007, 187, 288-292. (252) Francis R. Verdun; Carlo Giancaspro; Alan G. Marshall."Effects of Noise, Time-Domain Damping, Zero-Filling and the FFT Algorithm on the "Exact" Interpolation of Fast Fourier Transfrom Spectra." Appl. Spectrosc 1988, 42, 715-721. (253) Purcell, J. M.; Hendrickson, C. L.; Rodgers, R. P.; Marshall, A. G."Atmospheric Pressure Photoionization Fourier Transform Ion Cyclotron Resonance Mass Spectrometry for Complex Mixture Analysis." Anal. Chem. 2006, 78, 5906-5912. (254) Robb, D. B.; Covey, T. R.; Bruins, A. P."Atmospheric Pressure Photoionization: An Ionization Method for Liquid Chromatography−Mass Spectrometry." Anal. Chem. 2000, 72, 3653-3659. (255) Golotvin, S.; Williams, A."Improved Baseline Recognition and Modeling of FT NMR Spectra." J. Magn. Reson. 2000, 146, 122-125. (256) Jirasek, A.; Schulze, G.; Yu, M. M. L.; Blades, M. W.; Turner, R. F. B."Accuracy and Precision of Mannual Baseline Determination." Appl. Spectroc. 2004, 58, 1488-1499. (257) Frerking, M.Digital Signal Processing In Communications Systems; Springer, 1994;644 pp. (258) Marshall, A. G.; Liang, Z. M."Time-domain and frequency-domain (absorption-mode and magnitude-mode) noise and precision in Fourier transform spectrometry." Applied

Spectroscopy 1990, 44. (259) Antoniou, A.Digital Filters: Analysis Design, and Applications (2 ed). McGraw-Hill, 1993;689 pp. (260) Oppenheim, A. V.; Schafer, R. W.Discrete-Time Signal Procesing (2ed); Prentice-Hall, 1999;870 pp. (261) Lindon, J. C.; Ferrige, A. G."Digitisation and Data Processing in Fourier Transform NMR." Progress in NMR Spectroscopy 1980, 14, 27-66.

Page 151: Broadband Phase Correction Of Fourier Transform Ion Cyclotron Mass Spectra

134

(262) Lee, J.; Chow, K.; Comisarow, M."Anomalous Intensities of Apodized and Unapodized Magnitude Spectra." American Chemistry Society 1988, 60, 2212-2218. (263) Antonio DM; Kurt Wuthrich."Digital Filtering with a Sinusoidal Window Function: an Alternative Technique for Resolution Enhancement in FTNMR." Journal of Magnetic

Resonance 1976, 24, 201-204. (264) Daniel Traficante; Masoumeh Rajabzadeh."Optimum Window Function for sensitivity Enhancement of NMR Signals." Concepts in Magnetic Resonance 2000, 12, 83-101. (265) Goto, Y."Interpolation of Hamming-Apodized DFT Spectra." Electronics and

Communications in Japan 2000, 3, 1510-1517. (266) Fuson, M."FT NMR in the Intstrumental Analysis Course." Journal of Chemical

Education 1994, 71, 126-129. (267) Comisarow, M.; Melka, J."Error Estimates for Finite Zero-Filling In Fourier Transform spectrometry." Analytical Chemistry 1979, 51, 2198-2202. (268) Marshall, A. G."Theoretical Signal-to-Noise Ratio and Mass Resolution in Fourier Transform Ion Cyclotron Resonance Mass Spectrometry." Analytical Chemistry 1979, 51, 1710-1714. (269) Brenna, J. T.; Creasy, W."Experimental Evaluation of Apodization Functions For Quantitative Fourier Transform Mass Spectrometry." International Journal of Mass

Spectrometry and Ion Processes 1989, 90, 151-166. (270) Senko, M.; Canterbury, J. D.; Guan, S.; Marshall, A. G."A high performance modular data system for Fourier transform ion cyclotron resonance mass spectrometry." Rapid Commun.

Mass Spectrom. 1996, 10, 1839-1844. (271) Senko, M. W.; Hendrickson, C. L.; Pasa-Tolic, L.; Marto, J. A.; White, F. M.; Guan, S.; Marshall, A. G."Electrospray Ionization FT-ICR Mass Spectrometry at 9.4 Tesla." Rapid

Commun. Mass Spectrom. 1996, 10, 1824-1828. (272) Guan, S."General Phase Modulation Method for Stored Waveform Inverse Fourier Transform Excitation for Fourier Transform Ion Cyclotron Resonance Mass Spectrometry." J.

Chem. Phys 1989, 91. (273) Guan, S.; McIver Jr, R. T."Optimal Phase Modulation in Stored Wave Form Inverse Fourier Transform Excitation for Fourier Transform Mass Spectrometry. I. Basic Algorithm." J.

Chem. Phys 1990, 92, 5841-5846. (274) Zhu, M.; Zhang, H.; Humphreys, W. G."Drug Metabolite Profiling and Identification by High-Resolution Mass Spectrometry." J. Biol. Chem. 2011, 286, 25419-25425.

Page 152: Broadband Phase Correction Of Fourier Transform Ion Cyclotron Mass Spectra

135

(275) Xian, F.; Hendrickson, C. L.; Marshall, A. G."High Resolution Mass Spectrometry." Anal. Chem. 2012, 84, 708-719. (276) Grosshans, P. B.; Marshall, A. G."J. Mass Spectrom. Ion Processes 1992, 115, 1-19. (277) Harris, F. J."on the Use of Windows for Harmonic Analysis with the Discreter Fourier Transfrom." Proc. IEEE 1978, 66, 51-83. (278) Nuttall, A. H."Some Windows with Very Good Sidelobe Behavior." IEEE Trans. Acoust.

Speech Signal Process. 1981, 29, 84-91. (279) Pajer, R. T.; Armitage, I. M."Method for Complex Interpolation of Spectral Segments." J.

Magn. Reson. 1976, 21, 485-489. (280) Senko, M. W.; Beu, S. C.; McLafferty, F. W."Automated assignment of charge states from resolved isotopic peaks for multiply charged ions." J. Am. Soc. Mass Spectrom. 1995, 6, 52-56. (281) Wood, T. D.; Chen, L. H.; Kelleher, N. L.; Little, D. P.; Kenyon, G. L.; McLafferty, F. W."Direct Sequence Data from Heterogeneous Creatine Kinase (43 kDa) by High-Resolution Tandem Mass Spectrometry." Biochemistry 1995, 34, 16251-16254. (282) Ge, Y.; Lawhorn, B. G.; EINaggar, M.; Strauss, E.; Park, J.-H.; Begley, T. P.; McLafferty, F. W."Top Down Characterization of Larger Proteins (45 kDa) by Electron Capture Dissociation Mass Spectrometry." J. Am. Chem. Soc. 2002, 124, 672-678. (283) Ge, Y.; Rybakova, I. N.; Xu, Q.; Moss, R. L."Top-down high-resolution mass spectrometry of cardiac myosin binding protein C revealed that truncation alters protein phosphorylation state." Proc. Natl. Acad. Sci. USA 2009, 106, 12658-12663. (284) Zhang, H.; Cui, W.; Wen, J.; Blankenship, R. E.; Gross, M. L."Native Electrospray and Electron-Capture Dissociation in FTICR Mass Spectrometry Provide Top-Down Sequencing of a Protein Component in an Intact Protein Assembly." J. Am. Soc. Mass Spectrom. 2010, 21, 1966-1968. (285) Hofstadle, S. A.; Bruce, J. E.; Rockwood, A. L.; Anderson, G. A.; Winger, B. E.; Smith, R. D."Isotopic Beat Patterns in Fourier Transform Ion Cyclotron Resonance Mass Spectrometry: Implications for High Resolution Mass Measurements of Large Biopolymers." Int. J. Mass

Spectrom. Ion Processes 1994, 132, 109-127. (286) Senko, M. W.; Guan, S.; Huang, Y.; Marshall, A. G.; Mclafferty, F. W., Proceedings of

the 43rd ASMS Conference on Mass Spectrometry & Allied Topics, Atlanta, Georgia, May 21-26 1995; 806. (287) Kelleher, N. L.; Senko, M. W.; Siegel, M. M.; McLafferty, F. W."Unit Resolution Mass Spectra of 112kDa Molecules with 3 Da Accuracy." J. Am. Soc. Mass. Spectrom 1997, 8, 380-383.

Page 153: Broadband Phase Correction Of Fourier Transform Ion Cyclotron Mass Spectra

136

(288) Pase-Tolic, L.; Anderson, G. A.; Smith, R. D.; Brothers II, H. M.; Spindler, R.; Tomalia, D. A."Electrospray Ionization Fourier Transform Ion Cyclotron Resonance Mass Spectyrometric Characterization of High Molecular Mass Starburst Dendrimers." International Journal of Mass

Spectrometry and Ion Processes 1997, 165/166, 405-418. (289) Pase-Tolic, L.; Bruce, J. E.; Paula Lei, Q.; Anderson, G. A.; Smith, R. D."In-Trap Cleanup of Proteins from Electrospray Ionization Using Soft Sustained off-Resonance Irradiation with Fourier Transform Ion Cyclotron Resonance Mass Spectrometry." Anal. Chem. 1998, 70, 405-408. (290) Muddiman, D. C.; Null, A. P.; Hannis, J. C."Precise Mass Measurement of a Double-stranded 500 Base-pair (309kDa) Polymerase Chain Reaction Product by Negative Ion Electrospray Ionization Fourier Transform Ion Cyclotron Resonance Mass Spectrometry." Rapid

Commun. Mass Spectrom. 1999, 13, 1201-1204. (291) Kaiser, N. k.; Quinn, J. P.; Blakney, G. T.; Hendrickson, C. L.; Marshall, A. G."A Novel 9.4 Tesla FTICR Mass Spectrometer with Improved Sensitivity, Mass Resolution an Mass Range." J. Am. Soc. Mass Spectrom. 2011, 22, 1343-1351.

Page 154: Broadband Phase Correction Of Fourier Transform Ion Cyclotron Mass Spectra

137

BIOGRAPHICAL SKETCH

EDUCATION

Florida State University, Tallahassee, FL, USA

Ph.D. Candidate, Analytical Chemistry, expected April 2012 Xi’an Jiaotong University, Xian, China

B.E. Chemical Engineering, July 1996

PROFESSIONAL EXPERENCES

Graduate Research Assistant August 2006 -present Florida State University, Tallahassee, FL National High Magnetic Field Laboratory and Analytical Chemistry Program Advisor: Prof. Alan G. Marshall Advanced broadband phase correction of FT-ICR MS Improved FT-ICR 40% mass resolving power and 30% mass accuracy for complex data

analysis Characterized the third generation bio-fuel (Algal oil) by LC/MS/MS Identified composition and detailed information of algal oil

Research Assistant August 2003-July 2005 Louisiana State University, Baton Rouge, LA

Center for Advanced Microstructures and Devices (CAMD) Advisor: Dr. Yohannes Desta Utilized techniques including photolithography and electro-deposition to design different

micro devices such as micro-sensor, micro-reactor and micro fluidic chip Improved the strength and the elasticity of Ni-Fe alloy by applying the micro electro-

deposition techniques

Instructor

Xi’an Jiaotong University, Xian, China August 1996-2002 Laboratory of Chemical Engineering Department Taught the general and analytical chemistry experiments Analyzed samples by different analytical equipments such as IR, UV and AA

spectrophotometers

Page 155: Broadband Phase Correction Of Fourier Transform Ion Cyclotron Mass Spectra

138

PUBLICATIONS (listed in reverse chronological order)

Xian, F., Hendrickson, C. L and Marshall, A. G. Broadband Phase Correction of FT-ICR mass spectra by simultaneous excitation and detection. Analytical Chemistry (to be submitted soon) (2012)

Xian, F., Hendrickson, C. L, and Marshall, A. G. Phase spectrum of Fourier Transform Ion Cyclotron Resonance Mass Spectra. Analytical Chemistry (to be submitted soon) (2012)

Xian, F., Hendrickson, C. L and Marshall, A. G. Baseline Correction of Absorption-mode Fourier Transform Ion Cyclotron Mass Spectra. Analytical Chemistry (to be submitted soon) (2012)

Xian, F., Hendrickson, C. L., Blakney, G. T., Beu, S. C., and Marshall, A .G. Effect of Zero–Filling and Apodization on Fourier Transform Ion Cyclotron Resonance Mass Spectral Accuracy, Resolution, and Signal-to-Noise Ratio. Analytical Chemistry (to be submitted soon) (2012)

Xian, F., Hendrickson, C. L, and Marshall, A. G. Artifacts Induced by Selective Blanking of Time-Domain Data in Fourier Transform Mass Spectrometry. Applied Spectroscopy (to be submitted soon) (2012)

Mao, Y., Xian, F., McKenna, A. M., Rodgers, R. P., Hendrickson, C. L., and Marshall, A. G. Ultra-High-Resolution Fourier Transform Ion Cyclotron Resonance Mass Spectrometry of Distinction of <1 Da Doublets in Complex Mixture. Journal of American Society of Mass

Spectrometry (to be submitted soon) (2012)

Xian, F., Hendrickson, C. L, and Marshall, A. G. High Resolution Mass Spectrometry. Analytical Chemistry, 84, 708-719 (2012)

Valeja, S.G., Kaiser, N.K., Xian, F., Emmett, M.R., Hendrickson, C.L., Rouse, J.C. and Marshall, A.G. Unit Mass Resolution for an Intact 148 kDa Therapeutic Monoclonal Antibody by FT-ICR Mass Spectrometry. Journal of American Society of Mass Spectrometry, 83(22), 8391-8395 (2011)

Savory, J. J., Kaiser N. K. McKenna, A. M., Xian, F, Blakney, G. T., Hendrickson, C. L., Rodgers, R. P., and Marshall, A. G. Parts-Per-Billion Fourier Transform Ion Cyclotron Resonance Mass Measurement Accuracy with a “Walking” Calibration Equation. Analytical

Chemistry, 83(5), 1732-1736 (2011)

Xian, F., Hendrickson, C. L., Blakney, G. T., Beu, S. C., and Marshall, A. G. Automated Broadband Phase Correction of Fourier Transform Ion Cyclotron Resonance Mass Spectra. Analytical Chemistry 82(21), 8807-8812 (2010)

Page 156: Broadband Phase Correction Of Fourier Transform Ion Cyclotron Mass Spectra

139

McKenna, A. M., Xian, F., Glaser, P. B., Rodgers, R. P., and Marshall A. G. Heavy petroleum Composition 2. Evolution of the Boduszynski Model to the Limit of Distillation by Ultrahigh-Resolution FT-ICR Mass Spectrometry. Energy & Fuels, 24, 2939-2946 (2010) Marshall, A. G., Blakney, G. T., Beu, S. C., Hendrickson, C. L., McKenna, A. M., Purcell, J. M., Rodgers, R. P. and Xian, F. Petroleomics: a Test Bed for Ultra-High-Resolution Fourier Transform Ion Cyclotron Resonance Mass Spectrometry. European Journal of Mass

Spectroscopy 16(3), 367-371 (2010)

PRESENTATIONS AND ABSTRACTS (listed in reverse chronological order)

Xian, F; Hendrickson, C. L.; Blakney, G. T.; Beu, S. C. and Marshall, A. G., Effects of Zero-Fill

and Apodization on Absorption-Mode FT-ICR Mass Spectra. 8th N. Amer. Fourier Transform Mass Spectrometry Conf., Key West, FL, May 1-5 (2011)

Mao, Y., Xian, F., McKenna, A. M., Rodgers, R. P., Hendrickson, C. L., and Marshall, A. G., How Much Mass Resolution is Necessary: Counting the Possible Common Mass Doublets for

CcHhNnOoSs Elemental Compositions. 8th N. Amer. Fourier Transform Mass Spectrometry Conf., Key West, FL, May 1-5 (2011) Valeja, S. G., Kaiser, N. K., Xian, F., Emmett, M. R., Hendrickson, C. L., Rouse, J. C. and Marshall, A. G., Unit Mass Resolution for an Intact 148 kDa Therapeutic Monoclonal Antibody

by FT-ICR Mass Spectrometry. 8th N. Amer. Fourier Transform Mass Spectrometry Conf., Key West, FL, May 1-5 (2011) Xian, F.; Hendrickson, C. L.; Blakney, G. T.; Beu, S. C. and Marshall, A. G., Effects of Zero-Fill

and Apdization on Absorption-Mode FT-ICR Mass Spectra, 59th ASMS Conf. on Mass Spectrometry & Allied Topics, Denver, CO, June 5-9 (2011) Mao, Y., Xian, F., McKenna, A. M., Rodgers, R. P., Hendrickson, C. L., and Marshall, A. G., How Much Mass Resolution is Necessary: Counting the Possible Common Mass Doublets for

CcHhNnOoSs Elemental Compositions. 59th Amer. Soc. for Mass Spectrometry. Annual Conf. on Mass Spectrometry & Allied Topics, Denver, CO, June 5-9 (2011) Valeja, S. G., Kaiser, N. K., Xian, F., Emmett, M. R., Hendrickson, C. L., Rouse, J. C. and Marshall, A. G., Unit Mass Resolution for an Intact 148 kDa Therapeutic Monoclonal Antibody

by FT-ICR Mass Spectrometry. 59th Amer. Soc. for Mass Spectrometry. Annual Conf. on Mass Spectrometry & Allied Topics, Denver, CO, June 5-9 (2011) Xian, F.; Hendrickson, C. L.; Blakney, G. T.; Beu. S. C. and Marshall, A. G., Automated

Broadband Phase Correction for Improved FT-ICR Mass Spectra of Complex Mixtures, 58th Amer. Soc. for Mass Spectrometry. Annual Conf. on Mass Spectrometry & Allied Topics, Salt Lake City, UT, May 23-27 (2010) Savory, J. J.; Kaiser, N. K.; McKenna, A. M.; Blakney, G. T.; Hendrickson, C. L.; Rodgers, R. P.; Xian, F. and Marshall, A. G., A Walking Mass Calibration Equation for Complex Mixture

Page 157: Broadband Phase Correction Of Fourier Transform Ion Cyclotron Mass Spectra

140

Analysis by FT-ICR Mass Spectrometry, 58th Amer. Soc. for Mass Spectrometry. Annual Conf. on Mass Spectrometry & Allied Topics, Salt Lake City, UT, May 23-27 (2010) Xian, F.; Hendrickson, C. L.; Blakney, G. T.; Beu, S. C. and Marshall, A. G., Absorption

Spectra for FT-ICR Mass Spectrometry. FSU, Mar, 19 (2010) Marshall, A. G.; Blakney, G. T.; Emmett, M. R.; Hendrickson, C. L.; Rodgers, R. P.; Tipton, J. D. and Xian, F., Ultrahigh Mass Resolution and Mass Accuracy: What's Possible, What's Not,

and What It's Good For, Symposium. on Accurate Mass Measurement: State of the Art, Uses,

and Limitations, Pittcon 2009, Chicago, IL, March 8-14 (2009) Marshall, A. G.; Xian, F.; Hendrickson, C. L. and Blakney, G. T., Broadband Absorption-Mode

FT-ICR MS: 30% Enhancement in Resolving Power, 7th N. Amer. Fourier Transform Mass Spectrometry Conf., Key West, FL, April 19-22 (2009) Xian, F.; Hendrickson, C. L.; Blakney, G. T.; Beu, S. C. and Marshall, A. G., Improved

Broadband Phase Correction of Complex FT-ICR Mass Spectra: Baseline Roll and Apodization, 57th Amer. Soc. Mass Spectrum. Annual Conf. on Mass Spectrometry & Allied Topics, Philadelphia, PA, May 31 - June 5 (2009) Xian, F; Hendrickson, C. L.; Blakney, G. T.; Beu, S. C. and Marshall, A. G., Improved

Broadband Phase Correction of Complex FT-ICR Mass Spectra. 7th N. Amer. Fourier Transform Mass Spectrometry Conf., Key West, FL, April 19-22 (2009)